Page 1: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

FireLiteOperation Manual

Whip Mix Corporation • 361 Farmington Ave. • P.O. Box 17183 • Louisville, KY 40217-0183 USA 502-637-1451 • 800-626-5651 • Fax 502-634-4512 • www.whipmix.com

Page 2: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

2

FireLite Furnace Operations Manual

Table of Contents

Chapter Three — Entering and Editing Programs To Enter and Edit a Program ...............................................12

Changing a Program While it is Running .............................12

Available Program Parameters ............................................12

Chapter Four — Running a ProgramTo Start a Program ..............................................................13

Skipping a Program Step ....................................................13

Chapter Five — Maintenance Cleaning Your Furnace ........................................................13

Replacing the Muffle ............................................................13

If More Help is Needed ........................................................14

Replacement Parts Parts List .............................................................................15

FireLite Accessory High Efficiency Vacuum Pump ............................................14

Conversion Table Fahrenheit to Celsius ...........................................................15

Other Products from Whip Mix .......................16

Specifications ........................................................................2

Accessories ..........................................................................3

Chapter One — Getting Started Before Unpacking ..................................................................3

Unpacking .............................................................................3

Lifting and Carrying ...............................................................3

Furnace Requirements ..........................................................4

Mounting the Keyboard ........................................................5

Installing the Vacuum Pump ..................................................5

Volume Control ......................................................................6

Front Panel Controls and Indicators ......................................6

Menu Selection Buttons ........................................................6

Changing Units of Measurement Between English and Metric .................................................7

Night Mode ............................................................................8

Idle Temperature Operation ...................................................8

Chapter Two — Special Function Input Customize Program Finish Tune ...........................................9

Fast Cool with Vacuum ........................................................10

Adjusting Calibrations for Low-Fusing and High-Fusing Porcelains .......................................................10

Adjust High-Fusing Porcelain Calibration ............................10

Adjust Low-Fusing Porcelain Calibration .............................10

Displaying Muffle Hours ......................................................11

Displaying Software Version ................................................11

Error Messages ...................................................................11

SpecificationsPower: 110 VAC +/- 10% 50/60 Hz

Power Consumption: Furnace and Pump: 1,400 W Max

Outputs: Pump Accessory Power: 85–245 VAC, 50/60 Hz, 300 W Max

Dimensions: 131/2" wide x 131/2" deep x 201/2" high

Temperature: 1,200° C / 2,190° F Max

Weight: 50 lbs.

Muffle Chamber: 33/4" diameter x 21/2" high

Fuse ratings and characteristics: 115 Volt units: Type 326, 12 amp Slo Blo ceramic fuse, 3AB style. 230 Volt units: Type 218- T, 6.3 amp Slo Blo ceramic fuse, 5 x 20 mm style.

Minimum and Maximum ValuesMinimum entry temperature = 65° C/149° F

Maximum entry temperature = 650° C/1,202° F

Maximum entry time = 60 minutes

Minimum heat-rate = 1° C/minute

Maximum heat-rate = 100° C/minute/181° F

Maximum final temperature = 1,200° C/2,192° F

Maximum vacuum = 76 cm/30 inches of Mercury

Maximum idle temperature = 650° C/1,202° F

Page 3: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

3

FireLite Furnace Operations Manual

AccessoriesThe following accessories will be found with your FireLite furnace. If you are missing any of these please call Whip Mix at (800) 626-5651.

Cooling Tray Operator’s Manual

Firing Tray Dry-Erase Program Log Magnets*

Power Cord

*Only use dry-erase marker

Chapter One — Getting StartedYou are undoubtedly eager to unpack, set up, and begin using your new furnace. Getting started will be much easier if you carefully review the information in this chapter and follow the steps as outlined.

Before Unpacking • When you unpack your furnace, be sure to save the

carton and packing materials. These can be used again if there is a need to ship or return equipment.

• Be sure to read and save the printed shipping material packed with your furnace — it contains valuable information.

• DO NOT turn on the power to your furnace until you are instructed to do so.

Unpacking • If the packing materials and/or the furnace appear

to be damaged, call your dealer before continuing.

• Remove the furnace from the carton and place it on a flat surface.

• Look under the muffle, locate and remove the shipping hardware.

Lifting and CarryingNEVER ATTEMPT TO LIFT OR CARRY THE OVEN WITHIN ONE HOUR OF DISCONNECTING THE POWER AFTER IT HAS BEEN IN USE.

If you are preparing to ship the unit in its carton to a location outside your current facility, allow the furnace to cool with the platform down and, once it is cool to the touch, remove the white ceramic fiber firing tray from the table and pack it in the accessory box. Damage to the oven will result from shipment of the oven with the firing tray inside the muffle. Push the up button to run the table to the muffle. Install the shipping hardware underneath the table by reversing the above instructions.

Before the power is disconnected, use the front panel up arrow button to raise the platform to its full up position, then disconnect the power and wait at least one hour to allow the oven table and other surfaces to cool to room ambient temperature.

If you are relocating the oven inside the current facility or have used a protective glove to remove the firing tray from the table, continue below.

Remember to disconnect both the mains power cord and the pump accessory power cord and also remove the vacuum tubing from the vacuum inlet at the rear of the oven.

The oven weighs approximately 50 pounds (25 kg). A four-wheeled platform cart is recommended if the new location is more than a few feet away. Position the oven with the Mains Power Switch closest to the front edge of the work surface. Feel underneath the table and the table lift arm to ensure the surfaces are cool to the touch. Slide your right arm just underneath the table lift arm as close to the chassis as possible. Use your left hand to grasp the overhang of the chassis back just above the vacuum inlet. Raise the oven two inches, keeping it over the work surface until your back and arms confirm that you are able to control the oven without injuring yourself. Step back from the work surface and pivot slowly until you are headed in the desired direction. Reverse the above sequence to place the oven in the new location. Refer to page 5 to reconnect the power and vacuum ports.

THE OVEN SHOULD NEVER BE PLACED ON A FLAMMABLE WORK SURFACE. NEVER PLACE THE OVEN IN A LOCATION WITH OVERHEAD SHELVING OR CABINETS THAT COULD INTERFERE WITH THE VENTILATION SLOTS IN THE TOP OF THE MUFFLE HOOD.

Page 4: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

4

FireLite Furnace Operations Manual

Furnace Requirements• Turn on the furnace using the power switch located on

the left side of the furnace. The display will light up and the furnace will perform a series of internal self-tests. During these self-tests you will hear the pump turn on and off.

• The lift will lower. Place the firing tray on the center of the lift table.

• The furnace requires 120 VAC and 12 amps of current. A special heavy-duty power cord has been supplied with your furnace.

DO NOT OPERATE ON ANY OTHER POWER CORD. DO NOT OPERATE WITH AN EXTENSION CORD. OPERATING THIS FURNACE ON A CIRCUIT WITH OTHER FURNACES OR ELECTRICAL APPLIANCES THAT REQUIRE SIGNIFICANT POWER MAY CAUSE A CIRCUIT BREAKER TO TRIP. INSTALL THE HEAVY-DUTY SAFETY GROUNDED POWER CORD AND PLUG THE FURNACE INTO A GROUNDED OUTLET.

• Position the furnace so the front is facing you. Be sure to have at least eight inches on all sides of the furnace to allow sufficient air flow to keep the furnace cool. Do not place anything flammable near the furnace.

• Minimum ambient room temperature 60° F (15.5° C). Maximum ambient room temperature 80° F (26.6° C). Minimum relative humidity 45%. Maximum relative humidity 60%.

• The mains power On/Off switch is located on the left side of the chassis. It is a rocker-type switch with a “1” and a “0” symbol embossed on it. When the “1” side is depressed, the mains power will be applied to the oven. When the “0” side is depressed, the mains power will be removed from the oven. Never position the oven in such an orientation where the operator has to lean over the top of the oven to reach the mains power switch.

THE TRIANGULAR BLACK AND YELLOW STICKER ON THE FRONT CENTER EDGE OF THE MUFFLE HOOD IS A WARNING THAT THE METALLIC SURFACE CAN BECOME HOT TO THE TOUCH. NEVER PLACE YOUR HAND ON THIS SURFACE OR LEAN ON THE TOP OF THE HOOD WHEN THE UNIT IS IN OPERATION.

USE OF THIS OVEN IN ANY MANNER NOT PROSCRIBED IN THIS MANUAL MAY RENDER OPERATOR SAFETY FEATURES INEFFECTIVE.

Figure 1 Figure 2

Intended UseThe FireLite furnace is intended for the firing of dental porcelain materials only.

If the furnace is not used in a manner specified by Whip Mix Corporation, the protection provided by the product may be impaired.

CAUTION: NEVER OPEN THE INTERIOR OF THE CHASSIS BEFORE FIRST REMOVING THE AC ELECTRICAL POWER CORD. THERE ARE DEADLY VOLTAGES INSIDE.

If you experience a power outage while firing, the firing table can be raised or lowered manually.

1. Open the back cover by removing the screws on the left side with an Allen wrench as seen in Figure 1.

2. With pliers, turn the metal rod located to the right of the shaft as seen in Figure 2.

Page 5: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

5

FireLite Furnace Operations Manual

Installing the Vacuum Pump• If you have a Whip Mix supplied vacuum pump, simply

attach the pump power cord into the international standard outlet on the furnace marked Vacuum Pump, as seen in Figure 6.

• If you do not have a Whip Mix supplied vacuum pump, plug the vacuum pump’s power cord into the outlet end of the short power cord supplied with the furnace.

• Attach the other end of this short power cord to the international standard outlet on the rear of the furnace marked Vacuum Pump.

• Attach a 1/4 inch vacuum rated hose from the vacuum pump to the brass fitting marked Vacuum Pump at the rear of the furnace.

Note: The vacuum pump will only run continuously. Whenever a program requires vacuum, the pump will run continuously until vacuum is no longer required.

Figure 6

Mounting the KeyboardWhen you unpack your FireLite furnace, the keyboard will be located in the accessory box.

1. Remove keyboard from accessory box and loosen the thumbscrews.

2. Flip the keyboard around to the proper orientation as seen in Figure 3. One lock-washer goes on the outside of the brackets on each side and one lock-washer sits between the keyboard and the inside of the bracket. Adjust the keyboard to the desired position and tighten the set screw.

3. To complete the keyboard installation, simply plug the keyboard cable into the back of the keyboard until you hear a click as seen in Figures 4 and 5.

Figure 3

Figure 5

Figure 4

Page 6: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

6

FireLite Furnace Operations Manual

Menu Selection Buttons4-Digit LED Display

When a program is not running, the display will automatically show the muffle temperature. If a program is running, the default display is time remaining to finish the program. The 4-digit display indicates:

• the time remaining in minutes/seconds. If the time remaining for a program is greater than 99 minutes, 59 seconds the time will be in minutes only.

• the muffle temperature when the temperature light is illuminated. When entering the rate rise for a program, the rate rise will be displayed as degrees per minute.

• the vacuum level when the vacuum light is on.

• temperature and vacuum in Celsius or English units of measure. The °F/Inches light indicates that the furnace is operating in English units of measurement. The °C/Centimeters light indicates that the furnace is operating in Metric units of measure.

2-Digit LED Display

Indicates active program number. Except when a program is calling for vacuum, then the display shows the current vacuum level. The 2-digit window will also display:

• the Special Function number when entering a special function.

• ER to indicate an error condition.

• FR or DR to indicate frequency or duration of a tone when customizing a user-defined tone.

Display Button

Pressing Display will cause the four-digit display to cycle:

• from time remaining to temperature.

• from temperature to vacuum.

• from vacuum to time remaining.

Volume ControlAfter the self-test, the furnace will play a tune to let you know it is ready to run a program. You can adjust the volume of the tune by turning a potentiometer located in the furnace.

To adjust the volume of the furnace:

1. Locate the small white button just above the power switch as seen in Figure 7.

2. Remove the white button with a flat-head screwdriver.

3. Insert the screwdriver through the hole and make sure the screwdriver sits on the groove of the potentiometer.

4. Turn the screwdriver clockwise to increase the volume or counter-clockwise to decrease the volume.

Figure 7

Program Status LED Graph

4-Digit LED Display

Time Re-maining

Enter

Start Program

Special Functions

Night Mode

Program Edit

Program Select

LastTemperature

Vacuum Level

Table Down/Abort

Table Up/Skip

2-Digit LED Display

Vacuum Status LED

Graph

Display

10-Key Pad

Front Panel Controls and Indicators

Figure 8

Page 7: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

7

FireLite Furnace Operations Manual

After approximately one minute, the display will revert from the item being displayed to the time remaining if a program is running or to muffle temperature if the furnace is in idle mode.

Program and Vacuum Status LED Graphs

• Indicates the item being entered while entering a program.

• Indicates the status of a program while running.

Start Program Button

Pressing this button will start the selected program number displayed in the 2-digit display.

Special Function Button

Press Special Function to perform furnace setup. Special functions are covered in detail in Chapter Four — Running a Program.

Keypad

The keypad is used to enter numeric data.

Enter Button

Press the Enter button to:

• store the input data.

• proceed to the next item when entering input information.

• clear an error message on the display.

Program Select Button

Used to select a program to run. Valid program numbers are from 1 through 50.

• Press the Program Select button.

• Input the desired program number you wish to run. The program number will be displayed in the 2-digit display.

• Press the Start Program button to start the program shown in the 2-digit display.

Note: Program number 0 is a special program. DO NOT run or use program number 0 to edit or store your program.

Program Edit Button

• Pressing Program Edit allows you to edit a program. Editing programs is covered in detail in Chapter Two — Special Functions Input.

• Pressing Program Edit while a program is running will allow you to make temporary changes to the program currently running. After the changed program has run, the program will revert to its original values. Changing programs is covered in detail in Chapter Three — Entering and Editing Programs.

Up/Skip Button

• While in idle mode, press the /Skip once and the lift moves upward. Press again and the lift stops moving.

• While the lift is moving down, press the /Skip and the lift will begin moving up.

• While a program is running, press the /Skip and the program will skip to the next step. For example, pressing the /Skip button during the soak time will advance the program to the rise rate. Pressing /Skip again will advance to the final temperature.

Down/Abort Button

• While in idle mode, press the /Abort once and the lift moves downward. Press again and the lift will stop moving.

• While the lift is moving up, press the /Abort and the lift will begin moving down.

• While a program is running, pressing the /Abort will cause the program to abort and the lift to move to its lower position.

• Press the /Abort while changing a program or while entering a special function and the program entry or special function will abort without changing the previous contents of the program or special function.

• While in any mode, press the /Abort and the furnace will return to idle mode.

Changing Units of Measure Between English and MetricThe furnace can be changed from Metric to English units of measure or from English to Metric units while in idle mode. Programmed values are automatically converted.

Follow these steps to convert from Metric to English:

1. Press the Special Function button.

2. Press 2 on the keypad.

3. Press the Enter button.

Follow these steps to convert from English to Metric:

1. Press the Special Function button.

2. Press 1 on the keypad.

3. Press the Enter button.

Page 8: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

8

FireLite Furnace Operations Manual

Idle Temperature OperationThe furnace will be in the idle mode of operation after any of the other modes are completed or whenever the /Abort button has been pushed to abort an operation. The furnace comes from the factory with the idle temperature set to 0° C. The furnace will idle at idle temperature whenever a program is aborted or finished.

Setting the Idle temperature:

1. Press the Special Function button.

2. Press 22 on the keypad.

3. Press the Enter button.

4. Input the desired idle temperature using the keypad.

5. Press the Enter button.

Placing the furnace in night mode keeps the muffle idling at a low temperature, preventing a buildup of moisture during the night. This special function password allows the user to set the night mode temperature of the furnace.

The furnace comes from the factory with the night mode temperature set to 150° C. The furnace will remain at this temperature when night mode is in operation.

• Pressing the Night Mode button while a program is running will cause the furnace to enter the night mode after the program has finished running.

• Pressing Night Mode while the furnace is idling will cause the furnace to enter night mode immediately.

Important — Before firing porcelain for the first time, it is best to rid the muffle of all moisture. The easiest way to do this is to use the Night Mode for several hours.

Night Mode

Page 9: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

9

FireLite Furnace Operations Manual

Chapter Two — Special Function InputThe FireLite furnace has several special functions that allow the furnace operation to be customized to your needs.

Customize Program Finish TuneThe FireLite furnace has an unique feature to allow you to customize the tune that plays indicating a program is finished. This allows you to distinguish one furnace from another when a program is finished. FireLite allows you to program four tunes to play one after another. You need to enter the frequency and the duration for each tune. The furnace will play back all the tones right after they have been entered. You can reprogram the tones as many times as you want.

Setting Program Tunes:

1. Press the Special Function button.

2. Press 55 on the keypad.

3. Press the Enter button.

4. Use the keypad to enter the frequency of the first tune. The four-digit display will show a default number and the two digit display will show FR. This indicates the frequency of the first tune. The larger the number the lower the pitch. The lower the number, the higher the pitch. The acceptable range is from 0 to 1,000 however, the recommended range of values are between 400 and 1,000.

5. Press the Enter button.

6. The four-digit display will show a default number and the two-digit display will show DR. This indicates the duration of the first tune.

7. Use the keypad to enter the duration of the first tune. Generally 500 represents one-second duration. The higher the value, the longer the duration. The acceptable range is from 1 to 1,000 however, the recommended range of values are between 400 to 1,000.

8. Press the Enter button.

9. The two-digit display will show FR to indicate the frequency of the second tune. Use the keypad to enter the frequency of the second tune.

10. Press the Enter button.

11. The two-digit display will show DR to indicate the duration of the second tune. Use the keypad to enter the duration of the second tune.

12. Press the Enter button.

13. The two-digit display will display FR to indicate the frequency of the third tune. Use the keypad to enter the frequency of the third tune.

14. Press the Enter button.

15. The two-digit display will show DR to indicate the duration of the third tune. Use the keypad to enter the duration of the third tune.

16. Press the Enter button.

17. The two-digit display will display FR to indicate the frequency of the fourth tune. Use the keypad to enter the frequency of the fourth tune.

18. Press the Enter button.

19. The two-digit display will show DR to indicate the duration of the fourth tune. Use the keypad to enter the duration of the fourth tune.

20. Press the Enter button.

21. The furnace will play the customized tune once. After that, the four-digit display will show the temperature of the furnace and the two-digit display will show the program number.

Note: Entering 0 as frequency would represent a silent or pause between note segments for the specified duration.

Note: The recommended range of duration of the tone is from 400 to 1,000 and the recommended range of frequency is from 400 to 1,000. The higher the number of duration, the longer the length of the tone. The higher the number of frequency, the lower the pitch of the tone.

Page 10: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

10

FireLite Furnace Operations Manual

Fast Cool with VacuumThe FireLite comes from the factory with a Fast Cool Vacuum feature. This feature automatically turns on the vacuum pump between programs to cool the muffle to the entry temperature of the new program.

To use this feature:

1. Press the Start Program button.

Note: If the final temperature of the last program is higher than the start temperature of the new program, the vacuum pump will automatically come on.

To deactivate this feature:

1. Press the Special Function button.

2. Press 82 on the keypad.

3. Press the Enter button.

4. Press 0 (zero) to change the display from YES to NO

5. Press the Enter button.

To reactivate this feature:

1. Press the Special Function button.

2. Press 82 on the keypad.

3. Press the Enter button.

4. Press any number key between 1 and 9 to change the four-digit display from NO to YES.

5. Press the Enter button.

Adjusting Calibrations for Low-Fusing and High-Fusing PorcelainsAll porcelain furnaces fire slightly differently. Even porcelain furnaces from the same manufacturer can fire porcelains at slightly different temperatures. The FireLite addresses this problem by allowing you to customize the calibration. This feature can be used to adjust your furnace to fire porcelains at manufacturers’ recommendations, or to adjust multiple furnaces to fire identically.

The FireLite calibration can be adjusted for normal high-fusing porcelains and for the new low-fusing porcelains. These adjustments do not affect each other, so both adjustments can be used if you fire both types of porcelain. The adjustments can range from 0°C/0°F to 50°C/90°F.

Adjust High-Fusing Porcelain Calibration (800° C/1,470° F or above)

Overfires:

1. Press the Special Function button.

2. Press 18 on the keypad.

3. Press the Enter button.

4. Enter the number of degrees to adjust the calibration downward.

5. Press the Enter button.

Underfires:

1. Press the Special Function button.

2. Press 19 on the keypad.

3. Press the Enter button.

4. Enter the number of degrees to adjust the calibration upward.

5. Press the Enter button.

Adjust Low-Fusing Porcelain Calibration(800° C/1,470° F or below)

Overfires:

1. Press the Special Function button.

2. Press 16 on the keypad.

3. Press the Enter button.

4. Enter the number of degrees to adjust the calibration downward.

5. Press the Enter button.

Underfires:

1. Press the Special Function button.

2. Press 17 on the keypad.

3. Press the Enter button.

4. Enter the number of degrees to adjust the calibration upward.

5. Press the Enter button.

Page 11: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

11

FireLite Furnace Operations Manual

Error MessagesThe FireLite furnace contains several self-checks and error detection circuits. The following error messages may occur:

ER 1 and ER 9 — No Vacuum Error — The program called for a vacuum and after two minutes the vacuum didn’t reach an acceptable level. Clean the lift platform and the O-Ring seal. Check all vacuum connections of the pump and the furnace. Try the program again.

ER 2 — Thermocouple Error — The electronics have detected an open circuit in the thermocouple. Replace the thermocouple to correct this condition.

ER 3 — Temperature Calibration Error — The computer has detected a problem with the temperature calibration. Contact Whip Mix Corporation for calibration instructions.

ER 4 — Lift Calibration Error —The computer has detected a problem with the lift calibration. Contact Whip Mix Corporation for calibration instructions.

ER 5 — Vacuum Calibration Error — The computer has detected a problem with the vacuum calibration. Contact Whip Mix Corporation for calibration instructions.

ER 6 — Program Memory Error — The computer has detected a problem with the memory that stores your programs. Check all your programs and correct any that have changed.

ER 7 — Setup Error — The computer has detected a problem with the special functions setup values. Check all special functions as described in Chapter Four — Special Functions.

ER 8 — Maximum Temperature Error — The furnace has run away to a temperature beyond the maximum allowed. Turn the power off and allow the furnace to cool. This error may indicate a bad thermocouple. Test the furnace by running a program. If it fails again, contact Whip Mix Corporation for instructions.

ER 12 — Bottom Lift Fail Error — The computer has detected an error with the bottom lift mechanism. The lift hasn’t pushed the bottom limit switch after the allowed time. Contact Whip Mix Corporation for calibration instructions.

ER 13 — Top Lift Fail Error — The computer has detected an error with the top lift mechanism. The lift hasn’t pushed the top limit switch after the allowed time. Contact Whip Mix Corporation for calibration instructions.

ER 14 — Wrong Thermocouple Error — This error only appears when performing auto-calibration. Contact Whip Mix Corporation for calibration instructions.

Displaying Muffle HoursTo see the number of hours the muffle has been over 605° C:

1. Press the Special Function button.

2. Press 90 on the keypad.

3. Press the Enter button.

Displaying Software VersionTo display the version of the software running in your furnace:

1. Press the Special Function button.

2. Press 9 on the keypad.

3. Press the Enter button.

Page 12: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

12

FireLite Furnace Operations Manual

Chapter Three — Entering and Editing ProgramsPrograms can be entered or changed while in the idle mode. Changes to a program currently running can also be made for the remaining stages of the cycle. The procedure for entering and editing a program is the same.

Note: Pressing the /Abort button at any time while entering or editing a program will abort the procedure and the original program will not be changed.

Available Program ParametersStart Temp — The recommended range is 0° to 650° C (0° to 1,202° F).

Pre Dry — The lift will be raised in five evenly spaced steps during the programmed entry time. The time is entered in minutes:seconds. The acceptable range is from 0 to 60 minutes and seconds.

Soak Time — This is a preliminary low temperature soak. The acceptable range is from 0 to 60 minutes. The time is entered in minutes:seconds.

Heat Rate — The recommended range is 0° to 100° C (2° to 181° F) per minute.

Final Temp — This is the firing temperature of the program and ranges from the entry temperature for the program to 1200°C (2,192°F).

Hold Time — The recommended range is from 0 to 60 minutes and seconds. The time is entered in minutes: seconds.

Cool Time — This is the time it takes to lower the lift from the upper position to the lower position. The lift will be lowered in five evenly spaced steps during the programmed cool time. The acceptable range is from 0 to 60 minutes. Time is entered in minutes:seconds.

Vacuum Level — The furnace can only run continuous vacuum. All programs are pre-set with vacuum on. To turn

vacuum off, press the Zero (0) button on the keypad. To add vacuum to a program, press 1 on the keypad.

Start Time — If vacuum is desired during the heat soak, you can enter a value from 0 (when the soak time starts) to end the time of the heat soak.

Start Temperature — Enter the temperature at which the vacuum will turn on. The recommended range is from Start Temp to Final Temp.

Release Temp — If vacuum is used in a program, it is necessary to enter a temperature at which the vacuum pump will turn off. The recommended range is from the programmed entry temperature to the final temperature of the program.

Note: Vac Release Temp (Vac Start Temp).

Release Time — If a program has a hold time, it is possible to hold vacuum during the hold time. If vacuum is not desired during the hold time, enter a value of zero (0).

After the last value has been entered, the program is stored in permanent memory.

Note: If an invalid entry is made while entering the parameters of a program, the furnace will play an error tune and set the display to zero (0).

3. Press the Last button to scroll backwards through the program to verify or make changes.

Note: Pressing the Last button at the Start Temp (Entry Temp) while editing a program will set the temperature to zero.

To Enter/Edit a Program1. Press the Program Edit button, press Enter button.

The LED light on the left will blink showing the program stage to be set. Use the keypad to enter the program parameters.

2. By pressing the Enter key, you will move to the next parameter in the program.

Changing a Program While it is Running1. While a program is running, press the Program Edit

button. The program number will be displayed in the two-digit display. The four-digit display will be blank.

2. Press the Enter button.

3. The light for the start temperature will flash. Use the nu-meric keypad to enter the new values and press Enter to continue.

4. Repeat this step for each of the remaining steps.

5. After the last value has been entered, the furnace will continue running the program with the new values. When the program has finished running, it will revert back to the original values.

Page 13: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

13

FireLite Furnace Operations Manual

Loading Your Work Into the Furnace:Press the down arrow key to lower the work platform. Place your work on the firing tray, Press Start Program key, and it will raise automatically under program control.

Skipping a Program StepManually override a program step and skip to the next step by pressing the /Skip button.

Chapter Five — MaintenanceCAUTION: THE GASSES PRODUCED WHEN HEAT TREATING REFRACTORY MODELS WILL DAMAGE THE TRANSPARENCY AND STRUCTURE OF THE QUARTZ TUBES IN THE MUFFLE CHAMBER AND MAY CAUSE VACUUM LEAKAGE. TO HELP EXTEND THE LIFE OF THE MUFFLE, IT IS NECESSARY TO PREHEAT REFRACTORY MODELS IN ABURNOUT FURNACE TO PREVENT THE GASSES FROM DAMAGING THE QUARTZ GLASS IN THE MUFFLE CHAMBER.

Cleaning Your FurnaceYour new furnace has been painted with an epoxy-based paint and may be cleaned by using a soft cloth and mild kitchen cleaners such as Fantastic® or Formula 409®. The front panel should be cleaned with a window cleaner such as Windex®.

Never clean the display window with a dry cloth or tissue. Always moisten the cleaning cloth with a cleaner such as Windex®.

Note: If hazardous material is spilled on or inside the furnace, contact Whip Mix Technical Support if there is any doubt as how to safely remove the material.

Chapter Four — Running a ProgramTo Start a Program:1. Press the Program Select button.

2. Use numeric keypad to enter the desired program number.

3. Press the Start Program button to begin the new program.

To run the program currently on the two-digit display, simply press the Start Program button.

When a program is finished, the four-digit display and the Program Finish LED will flash and the furnace will play a tune to let you know the program is finished. The furnace will automatically cool down to idle temperature at this time.

The display will stop flashing and be ready for the next input by pressing the Enter button.

The “program finished” tune can be customized to distinguish one furnace from another. See Chapter Two — Special Functions for details.

Replacing the MuffleThe FireLite furnace contains a muffle module designed for simple service and replacement. The muffle module contains the heating chamber and the thermocouple.

The replacement procedure is as follows:

1. Turn the power switch located at the rear of the furnace to the off position. Remove the power cord from the wall socket to prevent any chance of a shock hazard. Allow the furnace to cool.

2. Remove the six Phillips-head screws holding the muffle cover to the two upright brackets. Remove the muffle cover by lifting it straight up until it clears the upright brackets.

3. Remove the thermocouple connector from the column by squeezing the two tabs and pulling away from the column. Remove the heating element connector from the bracket by squeezing the two tabs and pulling away from the bracket. Remove the four large Phillips-head screws on the top of the muffle. Grasp the handle on the top of the muffle and lift vertically to remove the muffle from the vacuum chamber.

4. Inspect the O-ring for damage and replace if required. Clean the O-ring area and the sealing area of the new muffle with a mild cleaning solution such as Windex®. Lower the new muffle into the vacuum chamber with the thermocouple and the heating element connectors. These are keyed so it is not possible to plug them in incorrectly.

5. Repeat step 3 in reverse order, reinstalling the muffle screws and connectors.

6. Install the muffle cover by lowering it over the upright column and aligning the screw holes. Install the six Phillips-head screws that attach the muffle cover to the upright column.

7. Plug the power cord back into the wall socket and turn the power on. Observe that the temperature begins to rise and stabilize at the entry temperature of the current program.

8. Fire a chip of porcelain to test the firing temperature of the new muffle. Adjust the calibration as described in Chapter Two — Adjusting Calibrations for Low-Fusing and High-Fusing Porcelains.

CAUTION: NEVER OPEN THE INTERIOR OF THE CHASSIS BEFORE FIRST REMOVING THE AC ELECTRICAL POWER CORD. THERE ARE DEADLY VOLTAGES INSIDE.

Page 14: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

14

FireLite Furnace Operations Manual

If More Help is NeededWe hope you have many years of trouble-free service from your furnace. If you do have problems with the furnace, or if you have questions about the furnace not covered in the manual, contact your dealer or Whip Mix at:

Phone: 800-626-5651 Fax: 502-634-4512 E-Mail: [email protected] Or visit our web site at www. whipmix.com

Be prepared to provide the following information:

1. Your name

2. Your lab’s name and address

3. Your lab’s phone number

4. Your lab’s fax number

5. Furnace model and serial number (serial number can be found on the rear of the furnace)

6. Your question/problem

When you call, it would be helpful if you are near the furnace. The technician will probably ask you to run tests and report the results, or read the display while the test is running.

High Efficiency Vacuum Pump

FireLite Accessory

Manufactured and serviced at:

Whip Mix CorporationP.O. Box 17183

Louisville, KY 40217-0183

Replacement PartsPart Number Description

96034 Muffle, 115 Volt

96312 Thermocouple

96286 Drive Motor

96287 Lift Cable

96288 Vacuum Valve Assembly

96070 Vacuum Valve Plunger

96289 Power Supply PCB

96290 Logic PCB

96025 Table

96291 Limit Switch Assembly

96293 Overlay

96294 Power Switch Assembly

96295 Speaker Assembly

96296 Vacuum Hose Assembly

96015 Alumina Firing Tray

96297 Cooling Shelf

96016 Fuse, 12 amp (115 Volt units only)

96017 Fuse, 6.3 amp (230 Volt units only)

96004 115 Volt Power Cord

96060 Pump & Heater SSR, 10 amp

96011 Rubber Feet

96302 Dry-Erase Program Log Magnet

IT11236 Operation Manual

Page 15: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table

15

FireLite Furnace Operations Manual

Fahrenheit to Celsius

° F ° C ° F ° C ° F ° C ° F ° C 100 38 630 332 1,160 627 1,690 921

110 43 640 338 1,170 632 1,700 927

120 49 650 343 1,180 638 1,710 932

130 54 660 349 1,190 643 1,720 938

140 60 670 354 1,200 649 1,730 943

150 66 680 360 1,210 654 1,740 949

160 71 690 366 1,220 660 1,750 954

170 77 700 371 1,230 666 1,760 960

180 82 710 377 1,240 671 1,770 966

190 88 720 382 1,250 677 1,780 971

200 93 730 388 1,260 682 1,790 977

210 99 743 393 1,270 688 1,800 982

220 104 750 399 1,280 693 1,810 988

230 110 760 404 1,290 699 1,820 993

240 116 770 410 1,300 704 1,830 999

250 121 780 416 1,310 710 1,840 1,004

260 127 790 421 1,320 716 1,850 1,010

270 132 800 427 1,330 721 1,860 1,016

280 138 810 432 1,340 727 1,870 1,021

290 143 820 438 1,350 732 1,880 1,027

300 149 830 443 1,360 738 1,890 1,032

310 154 840 449 1,370 743 1,900 1,038

320 160 850 454 1,380 749 1,910 1,043

330 166 860 460 1,390 754 1,920 1,049

340 171 870 466 1,400 760 1,930 1,054

350 177 880 471 1,410 766 1,940 1,060

360 182 890 477 1,420 771 1,950 1,066

370 188 900 482 1,430 777 1,960 1,071

380 193 910 488 1,440 782 1,970 1,077

390 199 920 493 1,450 788 1,980 1,082

400 204 930 499 1,460 793 1,990 1,088

410 210 940 504 1,470 799 2,000 1,093

420 216 950 510 1,480 804 2,010 1,099

430 221 960 516 1,490 810 2,020 1,104

440 227 970 521 1,500 816 2,030 1,110

450 232 980 527 1,510 821 2,040 1,116

460 238 990 532 1,520 827 2,050 1,121

470 243 1,000 538 1,530 832 2,060 1,127

480 249 1,010 543 1,540 838 2,070 1,132

490 254 1,020 549 1,550 843 2,080 1,138

500 260 1,030 554 1,560 849 2,090 1,143

510 266 1,040 560 1,570 854 2,100 1,149

520 271 1,050 566 1,580 860 2,110 1,154

530 277 1,060 571 1,590 866 2,120 1,160

540 282 1,070 577 1,600 871 2,130 1,166

550 288 1,080 582 1,610 877 2,140 1,171

560 293 1,090 588 1,620 882 2,150 1,177

570 299 1,100 593 1,630 888 2,160 1,182

580 304 1,110 599 1,640 893 2,170 1,188

590 310 1,120 604 1,650 899 2,180 1,193

600 316 1,130 610 1,660 904 2,190 1,199

610 321 1,140 616 1,670 910 2,200 1,204

620 327 1,150 621 1,680 916 2,210 1,210

Celsius to Fahrenheit

° C ° F ° C ° F ° C ° F ° C ° F ° C ° F 35 95 300 572 565 1,049 830 1,526 1,095 2,003

40 104 305 581 570 1,058 835 1,535 1,100 2,012

45 113 310 590 575 1,067 840 1,544 1,105 2,021

50 122 315 599 580 1,076 845 1,553 1,110 2,030

55 131 320 608 585 1,085 850 1,562 1,115 2,039

60 140 325 617 590 1,094 855 1,571 1,120 2,048

65 149 330 626 595 1,103 860 1,580 1,125 2,057

70 158 335 635 600 1,112 865 1,589 1,130 2,066

75 167 340 644 605 1,121 870 1,598 1,135 2,075

80 176 345 653 610 1,130 875 1,607 1,140 2,084

85 185 350 662 615 1,139 880 1,616 1,145 2,093

90 194 355 671 620 1,148 885 1,625 1,150 2,102

95 203 360 680 625 1,157 890 1,634 1,155 2,111

100 212 365 689 630 1,166 895 1,643 1,160 2,120

105 221 370 698 635 1,175 900 1,652 1,165 2,129

110 230 375 707 640 1,184 905 1,661 1,170 2,138

115 239 380 716 645 1,193 910 1,670 1,175 2,147

120 248 385 725 650 1,202 915 1,679 1,180 2,156

125 257 390 734 655 1,211 920 1,688 1,185 2,165

130 266 395 743 660 1,220 925 1,697 1,190 2,174

135 275 400 752 665 1,229 930 1,706 1,195 2,183

140 284 405 761 670 1,238 935 1,715 1,200 2,192

145 293 410 770 675 1,247 940 1,724 1,205 2,201

150 302 415 779 680 1,256 945 1,733 1,210 2,210

155 311 420 788 685 1,265 950 1,742 1,215 2,219

160 320 425 797 690 1,274 955 1,751 1,220 2,228

165 329 430 806 695 1,283 960 1,760 1,225 2,237

170 338 435 815 700 1,292 965 1,769 1,230 2,246

175 347 440 824 705 1,301 970 1,778 1,235 2,255

180 356 445 833 710 1,310 975 1,787 1,240 2,264

185 365 450 842 715 1,319 980 1,796 1,245 2,273

190 374 455 851 720 1,328 985 1,805 1,250 2,282

195 383 460 860 725 1,337 990 1,814 1,255 2,291

200 392 465 869 730 1,346 995 1,823 1,260 2,300

205 401 470 878 735 1,355 1,000 1,832 1,265 2,309

210 410 475 887 740 1,364 1,005 1,841 1,270 2,318

215 419 480 896 745 1,373 1,010 1,850 1,275 2,327

220 428 485 905 750 1,382 1,015 1,859 1,280 2,336

225 437 490 914 755 1,391 1,020 1,868 1,285 2,345

230 446 495 923 760 1,400 1,025 1,877 1,290 2,354

235 455 500 932 765 1,409 1,030 1,886 1,295 2,363

240 464 505 941 770 1,418 1,035 1,895 1,300 2,372

245 473 510 950 775 1,427 1,040 1,904 1,305 2,381

250 482 515 959 780 1,436 1,045 1,913 1,310 2,390

255 491 520 968 785 1,445 1,050 1,922 1,315 2,399

260 500 525 977 790 1,454 1,055 1,931 1,320 2,408

265 509 530 986 795 1,463 1,060 1,940 1,325 2,417

270 518 535 995 800 1,472 1,065 1,949 1,330 2,426

275 527 540 1,004 805 1,481 1,070 1,958 1,335 2,435

280 536 545 1,013 810 1,490 1,075 1,967 1,340 2,444

285 545 550 1,022 815 1,499 1,080 1,976 1,345 2,453

290 554 555 1,031 820 1,508 1,085 1,985 1,350 2,462

295 563 560 1,040 825 1,517 1,090 1,994 1,355 2,471

Page 16: FireLite - Whip Mixwhipmix.com/wp-content/uploads/FireLite-Manual_Old_0911.pdf · shipment of the oven with the firing tray inside the muffle. Push the up button to run the table