QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

FIRST CQ MEASUREMENTS AND SKIN DOSE

MEASUREMENTS WITH RADIOCHROMIC

FILMSSYRMA-3D Collaboration Meeting

Anna Veronese

17 February 2017

1 / 42

Page 2: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MATERIALS

2 / 42

Page 3: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM

Acquisition and Reconstruction Parameters:

3 / 42

Page 4: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM IMAGES35 keV High Dose:

4 / 42

Page 5: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM IMAGES35 keV High Dose:

5 / 42

Page 6: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS

Average CT number of water:

6 / 42

Page 7: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTSAverage CT number of water:

7 / 42

Page 8: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS

Noise and Uniformity:

8 / 42

Page 9: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS

Noise and Uniformity:

9 / 42

Page 10: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS

CT Number Linearity:

10 / 42

Page 11: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS

CT Number Linearity:

11 / 42

Page 12: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS RESULTS

Synthesis of the Results:

12 / 42

Page 13: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS RESULTSSynthesis of the Results:

13 / 42

Page 14: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

QC PHANTOM MEASUREMENTS RESULTS

What we learned:• Nylon isn’t Nylon... It’s Delrin!• Plexiglas absorbs ≈water @ 35 keV;• Low Contrast Details are unresolved;• Ring artefacts affect uniformity especially in the central

area;• PVC generated streaking artefacts: is it too absorbing?

14 / 42

Page 15: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

SKIN DOSE MEASUREMENTS USING RADIOCHROMIC

FILMS

Aim:• Measure the dose to the

patient’s skin;• Different irradiation

modalities.

Radiochromic Films:• GafChromicTM XR-QA2;• Sensitive 1-200 mGy;• Properly Calibrated.

15 / 42

Page 16: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

MEASUREMENTS AT THE SYRMEP BEAMLINESYRMEP Beamline @Elettra:• Elettra @ 2.4 GeV;• Laminar SR beam 3 mm height (@slits);• Energy 38 keV;• IOC gives the air-Kerma.

(a) (b)

16 / 42

Page 17: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

MEASUREMENTS AT THE SYRMEP BEAMLINE

Cylindrical Phantom 10 cm in diameter on patient’s support:• diameter, symmetry and µ similar to breast;• two pieces of films tightly wrapped around it to simulate

patient’s skin.

17 / 42

Page 18: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

MEASUREMENTS AT THE SYRMEP BEAMLINE

Four different irradiation modalities on different pieces of film:

• 180◦

• 360◦

• 0◦-180◦- step - 0◦-180◦

• 0◦-180◦- step - 180◦-360◦

18 / 42

Page 19: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 0◦-180◦

Radiochromic film scan:

(c)

(d)

19 / 42

Page 20: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 0◦-180◦Profile of horizontal dose distribution:

• Minimum dose: 18% of the maximum dose;• MGDt = 6.9 mGy→ Dose up to 205% of MGDt.

20 / 42

Page 21: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 0◦-180◦

Profile of vertical dose distribution:

(e)

(f)

21 / 42

Page 22: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 0◦-180◦

Profile of vertical dose distribution:

• Maximum peak-to-valley difference: 33% of the maximumdose.

22 / 42

Page 23: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 180◦-360◦

Radiochromic film scan:

(g)

(h)

23 / 42

Page 24: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 180◦-360◦Profile of horizontal dose distribution for odd slices:

• Minimum dose: 20% of the maximum dose.24 / 42

Page 25: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 180◦-360◦Profile of horizontal dose distribution for even slices:

• Minimum dose: 20% of the maximum dose.• MGDt = 6.9 mGy→ Dose up to 170% of MGDt.

25 / 42

Page 26: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 180◦-360◦Profile of vertical dose distribution: we see three differentpatterns

(i)

(j)

26 / 42

Page 27: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 180◦-360◦Profile of vertical dose distribution:

• Maximum peak-to-valley difference: 80% of the maximumdose.

27 / 42

Page 28: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 180◦-360◦Profile of vertical dose distribution:

• Maximum peak-to-valley difference: 34% of the maximumdose.• MGDt = 6.9 mGy→ Dose up to 120% of MGDt.

28 / 42

Page 29: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

0◦-180◦- STEP - 180◦-360◦Profile of vertical dose distribution:

• Maximum peak-to-valley difference: 77% of the maximumdose.

29 / 42

Page 30: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

CONCLUSIONS AND COMMENTS

We were able to measure the dose to the skin in differentirradiation conditions:• In 180◦the non irradiated skin receives > 10% of the

maximum dose;• 0◦-180◦- step - 0◦-180◦gives high areas of superposition

(with this step);• 0◦-180◦- step - 180◦-360◦is a good intermediate irradiation

(less superposition).

30 / 42

Page 31: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

FUTURE MEASUREMENTS

In the future we would like to repeat the measure with...• Different step between slices to study the superposition;• Different energies;• Different doses.

31 / 42

Page 32: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

THE END

Thank you for you kind attention!

32 / 42

Page 33: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

WHAT IS THE MEAN GLANDULAR DOSE (MGD)?• For the whole volume irradiated, MGD(mGy) is calculated

from the total energy deposited (Eg) in the total glandularmass of the breast (Mg), as:

MGD = Eg/Mg

• For a partial irradiation, MGDv (mGy) can be calculatedfrom the energy (eg) deposited only in the glandular massof the directly irradiated breast volume (mg), as:

MGDv = eg/mg

• Considering the scattered radiation coming outside of theirradiated volume, MGDt is the ratio between the totalenergy deposited in the whole breast (Eg) and theglandular mass in the irradiated volume (mg):

MGDt = Eg/mg

33 / 42

Page 34: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

180◦

Radiochromic film scan:

(k)

(l)

34 / 42

Page 35: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

180◦

Profile of horizontal dose distribution:

• Minimum dose: 9.5% of the maximum dose;• MGDv = 7,8 mGy→ Dose up to 240% of MGDv.

35 / 42

Page 36: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

VERTICAL BEAM PROFILE

Profile of vertical dose distribution:

(m)

(n)

36 / 42

Page 37: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

VERTICAL BEAM PROFILEProfile of vertical dose distribution:

• 1200 dpi scan: 1 pixel=2 µm;• FWHM=3,61 mm;• Magnification: 20%.

37 / 42

Page 38: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

360◦

Radiochromic film scan:

(o)

(p)

38 / 42

Page 39: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

360◦

Profile of horizontal dose distribution:

• Maximum variation: 24% of the maximum dose;• MGDv = 7.8 mGy→ Dose 175% - 130% of MGDv.

39 / 42

Page 40: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

DOSE EVALUATION

The dose is proportional to the change in reflectance of the film:

• Films have to be scanned both before and after irradiation;• PV (pixel value) change is evaluated as:

∆R = Rbefore − Rafter =PVbefore

216 − PVafter

216

• A piece of control film is never irradiated:

∆Rcontrol = Rcontrolbefore − Rcontrol

after =PVcontrol

before216 −

PVcontrolafter216

• The net change in reflectance is:

net∆R = ∆R−∆Rcontrol

40 / 42

Page 41: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

CALIBRATION PROCEDUREWe want a dose-response curve for the specific energy (38 keV):

• Obtained plotting the air Kerma Kfilm as a function of thecorresponding film response net∆R;• 3 4×4 cm film pieces exposed at the same time;• Different values if air-Kerma (1-20 mGy);

• Best fitting funtion logarithm one: Kfilm = a +b net∆R

ln (net∆R)

Net DeltaR (a.u.)0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16 0.18

Air

Ker

ma

(mG

y)

0

2

4

6

8

10

12

14

16

18

20 / ndf 2χ 4.61 / 6

Prob 0.5947a 0.1166± 0.6549 b 5.681±169 −

/ ndf 2χ 4.61 / 6Prob 0.5947a 0.1166± 0.6549 b 5.681±169 −

41 / 42

Page 42: FIRST CQ M SKIN DOSE M R F - agenda.infn.it€¦ · QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS FIRST CQ MEASUREMENTS AND SKIN DOSE MEASUREMENTS WITH RADIOCHROMIC FILMS SYRMA-3D

QUALITY CONTROL RADIOCHROMIC FILM MEASUREMENTS

OTHER QC PROTOCOL MEASUREMENTS

Ideas for other measurements:• Air Slits and Vacuum Slits control;• µ curve to verify energy;• Phantom to verify bed movement;• CTDI to check dose.