flow ehandbook

Upload: kamy001

Post on 08-Aug-2018

224 views

Category:

Documents


1 download

TRANSCRIPT

  • 8/22/2019 Flow eHandbook

    1/12

    Fl Habk

    Mii FlStart u

  • 8/22/2019 Flow eHandbook

    2/123

    tabl f CsKp mi pblms fm bbbli p 4Entrained gas can aect Coriolis meters but you can take steps to optimize perormance.

    usa h Aaci f Mams 9Magnetic owmeters provide accuracy and can be used in a variety o applications and environments

    tak a diff Lk a Cifal Pmps 13An unconventional assessment can provide insights or eective control.

    tPrevious Page Next Pageu

    http://www.usa.siemens.com/
  • 8/22/2019 Flow eHandbook

    3/12 4

    Kp mi pblms fm bbbli pEntrained gas can aect Coriolis meters but you can take steps to optimize perormance.

    and environments

    By Tim Patten, Micro Motion, Inc.

    CorIoLIS MeterS have long been used verysuccessully on single-phase luids. However,

    liquids that contain bubbles (air or gas) cause

    dynamic changes to a Coriolis meter that are not

    present in a single-phase luid and that lead to

    measurement errors.

    A Coriolis meter operates by driv ing one or

    two tubes at a resonant, or natural, requency. In

    the meter, the electronics (or transmitter) send a

    drive signal to the sensor that tracks the requen-

    cy o the tube and maintains the proper vibration

    amplitude. Driving on the resonant requency is

    important because it enables luid density mea-

    surement and minimizes power requirements.All modern Coriol is meters are intrinsical ly

    sae (IS), which limits the amount o power that

    is allowed to drive the sensor. Bubbles moving

    around in the l iquid tremendously increase uid

    damping, which results in power requirements

    that ar exceed what IS restrict ions permit. So, the

    tube amplitude signicantly decreases. Tis condi-

    tion is sometimes called stall, although the tubes

    usually do continue to vibrate to some extent.

    When the tube amplitude decreases, the

    signal-to-noise ratio also alls, making it a chal-

    lenge to extract the mass ow signal rom the

    Figure 1. When uid velocity cant overcome buoyancy, bubbles get caught in

    inlet leg.

    tPrevious Page Next Pageu

  • 8/22/2019 Flow eHandbook

    4/12 5

    relatively high level o noise. Older analog signal-

    processing techniques are highly sensitive to

    entrained air because signa l amplitude is low and

    noise is high; no algorithm is available to enhance

    the measurement signal, thereby improving the

    signal-to-noise ratio. In contrast, digital signal

    processing (DSP) can eectively lter the noise

    to yield a good stable measurement so long as the

    tube is vibrating, even at reduced amplitudes.Note: Even with DSP, when gas is present in a liq-

    uid stream the meter can only provide total-product

    density (including the gas), not liquid-only density.

    tHe IMPACt oF FLuId dynAMICS

    Coriolis meters are not sensitive to ow prole

    and other disturbances that aect other metering

    technologies. For instance, since the undamental

    measurement o delta comes rom the relative

    values o each o two tubes in bent-tube designs,

    swirl upstream o the meter doesnt impact the mea-

    surement because it doesnt matter how much ow

    goes through one tube or the other. Accuracy is not

    degraded even when one tube is completely plugged.

    However, when gas is present in a liquid, the

    ow prole can become a concern. Although the

    undamental measurement is unaected (that is,

    the relative delta ), the tubes can become imbal-

    anced due to the large density dierence between

    them (air in one, liquid in the other, or instance).

    An imba lance can cause meter zero errors; there-ore, measuring low ow rates can be problematic.

    An equally signicant problem occurs at rates

    too low to sweep bubbles out o the tubes. I the

    uid velocity is less than approximately 0.6 m/s,

    air will hang up in tube regions where the ow

    is against gravity (Figure 1). Bubbles get caught in

    the inlet tube leg because uid velocity is not great

    enough to push the bubbles down and out against

    gravity orcing the bubbles up. Tis issue is present

    in any bent-tube meter design because at some loca-

    tion in the tube the uid velocity is ghting gravity.

    Te solution is to keep ow rate high enough

    Figure 2. Even at low ow rate, measurements or 10,000-cP toothpaste with 2-5% void raction are within specifcation.

    tPrevious Page Next Pageu

  • 8/22/2019 Flow eHandbook

    5/12 6

    such that uid velocity can purge the sensor o

    air. A rate o 20% o meter nominal ow (1 m/s in

    the ow tube) or higher is adequate to completely

    purge the meter o bubbles and give good peror-

    mance. In a U-shaped meter, mounting the sensor

    in a vertica l pipe run with ow going up helps to

    keep the bubbles moving through the meter.

    tHe roLe oF FLuId ProPertIeS

    Pressure, luid temperature and viscosity all

    impact how a Coriolis meter deals with varying

    levels o entrained air.

    As pressure increases or decrea ses , the appar-ent void raction changes, o course. his means,

    or instance, i two meters are piped in series, the

    downstream meter is at a distinct disadvantage

    because the pressure is lower and thereore the

    void raction is higher.

    emperature plays a minor role, in that it a-

    ects viscosity and surace tension. It also impacts

    void raction to a small degree (higher tempera-

    ture results in higher void raction).

    Viscosity is a very important uid parameter

    because it directly inuences the propensity o

    the uid to hold up air (or gas). In a low-viscosity

    liquid such as water, air bubbles coalesce rom nely

    distributed small bubbles into large ones that collect

    at high points in the line. In contrast, i the bubbles

    stay nely distributed, as happens in high-viscosity

    liquids, they will be purged rom the meter easily

    and not collect and metering will be accurate.

    Figure 2 shows results or toothpaste with a viscosity

    o 10,000 cP and entrained air level between 2 and

    5%. Rates are quite low or a 2-in. meter (

  • 8/22/2019 Flow eHandbook

    6/127

    be measured because o the previously described

    separation issues. However, noise rejection im-

    provement with new DSP techniques will allow

    the minimum rate to be pushed lower.

    A signicant problem with any two-phase

    ow (water/air, dog ood with solids suspended in

    water, oil/gas, etc.) is at zero ow. When the ow

    is stopped, the multiple phases separate by gravity,

    prompting an imbalance in the tube. Tis imbal-ance causes an apparent meter zero change. Work

    on signal processing improvements to address this

    problem is currently a signicant area o research.

    A SPeCIAL CASe

    Empty-ull-empty batching can pose a related measure-

    ment issue. Such batching is most common to avoid

    cross-contamination o products when lling large

    tanks such as rail cars or trucks. Tereore, the loading

    line is purged with air or other inert gas between loads,

    leaving the meter empty beore and ater the batch.

    Generally, this application is not too difcult

    because the batches tend to be long (greater than

    one minute). Any transient meter behavior at the

    beginning and end o the batch is sma ll compared

    to the whole batch, so errors a re washed out.

    However, when batches are short (less than one

    minute), the transient errors can account or a signi-

    cant raction o the total error. Air may be entrained

    or a brie period, but the main issue is the time ittakes to ll the meter with uid. For instance, an ap-

    plication running at 3 m/s will take about 0.1 s to ll

    i the tube length is 0.3 m; an application at 0.3 m/s

    will take a ull second simply to ll the meter. Experi-

    ence has shown that i the meter ll-time is less than

    0.1 sec., good batching perormance can be achieved,

    regardless o the meters tube geometry.

    tIM PAtten is director o measurement technology or

    Micro Motion, Inc., Boulder, Colo. E-mail him at Tim.Pat-

    [email protected].

    tPrevious Page Next Pageu

    mailto:[email protected]:[email protected]://www.emersonprocess.com/mailto:[email protected]:[email protected]:[email protected]
  • 8/22/2019 Flow eHandbook

    7/12

    usa h Aaci f MamsMagnetic owmeters provide accuracy and can be used in a variety o applications

    and environments

    By David W. Spitzer

    MAgnetIC FLowMeterS are among themost versatile o lowmeter technologies. hese

    meters measure liquid velocity, rom which the

    volumetric low rate is inerred. he measure-

    ment is linear with liquid velocity and exhibits a

    relatively large turndown. In addition, the range

    o accurate low measurement is relatively large

    and easy to change a ter installation.

    Straight-run requirements are relatively short,

    so magnetic lowmeter technology can be applied

    where l imited straight run is available. In addi-

    tion, the technology has no Reynolds number

    constraints, so it can be used or liquids with high

    or varying viscosity. However, liquid electricalconductivity constraints must be satisied or

    these lowmeters to unction.

    he only wetted parts o the lowmeters are

    the liner and electrodes, both o which can be

    made rom materials that can withstand cor-

    rosion. his makes the lowmeters suitable or

    use in chemical plants where corrosion may be

    a concern. wo-wire magnetic lowmeters are

    available that do not require power wiring. hese

    can replace an existing lowmeter using the exist-

    ing conduit or wiring with little or no electrical

    rework.

    PrInCIPLe oF oPerAtIon

    Magnetic owmeters use Faradays Law o electro-

    magnetic induction to determine the velocity o a

    liquid owing through a pipe. Following Faradays

    Law, ow o a conductive liquid through a magnetic

    eld will generate a voltage signal. Tis signal is sensed

    by electrodes located on the ow tube walls. When

    the coils are located externally, a non-conductive liner

    is installed inside the pipe to electrically isolate the

    electrodes and prevent the signal rom being shorted.

    For similar reasons, non-conductive materials are used

    to isolate the electrodes or internal coil designs.

    Te uid itsel is the conductor that will move

    through the magnetic eld and generate a voltagesignal at the electrodes. When the uid moves aster,

    it generates more voltage. Faradays Law states that the

    voltage generated is proportional to the movement o

    the owing liquid. Te transmitter processes the volt-

    age signal to determine liquid ow.

    SeLeCtIon FACtorS

    Many actors must be considered when selecting

    a owmeter, including the ambient conditions to

    which the owmeter primary and transmitter will be

    exposed. For the most part, the ambient temperature

    rating o the owmeter primary is higher than that

    9 Next PageutPrevious Page

  • 8/22/2019 Flow eHandbook

    8/12 10

    o the transmitter and does not limit applicability.

    Many primary and transmitter enclosures that are

    rated or NEMA 4X or IP67 provide adequate pro-

    tection against ambient humidity and precipitation

    encountered in outdoor installations.

    Operating conditions inside the pipe include

    pressure, temperature and liquid conductivity. In

    addition, the liquid can be corrosive or abrasive. Tese

    conditions are typically addressed using appropriatemechanical design and material selection. Pressure

    requirements are addressed by appropriate design o

    the ow tube or the application. One supplier makes

    a specially designed magnetic owmeter that can

    withstand 1,500 to 2,000 bar (more than 20,000 psi).

    Many primaries are available with polytetrauoro-

    ethylene (PFE) or peruoroalkoxy (PFA) liners that

    are rated to about 266 Degrees F and 356 Degrees F

    (130 Degrees C and 180 Degrees C), respectively.

    Less expensive liners rated to lower temperatures are

    oten available to handle less demanding applications.

    Appropriate electrode and liner material selection can

    reduce the eects o corrosion and abrasion. ake care

    when using ceramic liners because they can shatter

    when temperature gradient constraints are exceeded.

    Whereas the conductivity o the liquid in a typical

    magnetic owmeter must be maintained above about 5

    mixro-Siemens/cm (micro-S/cm), special low-conduc-

    tivity designs are available that operate as low as about

    0.01 micro-S/cm. Some owmeters require more than

    50 micro-S/cm, however, they are low-cost units that

    are oten applied to water or wastewater service where

    this conductivity is usually not a constraint.Te amount o straight-run pipe required to

    achieve the stated accuracy o the owmeter is a

    reection o the quality o the design and the tight-

    ness o the accuracy specication. In many applica-

    tions, these owmeters will unction accurately

    with about three nominal pipe diameters upstream

    and two nominal pipe diameters downstream o

    the electrode.

    Magnetic owmeter operation requires good

    electrical connections between the electrodes and

    the liquid. Te quality o this connection can

    degrade i an electrode becomes coated or corroded;

    this can compromise AC owmeter accuracy by

    shiting the zero, and may cause the owmeter to

    ail to operate. Te advent o DC-pulse excitation

    transmitters reduced much o the need to address

    this issue. In addition, some manuacturers have

    designed their transmitters to exhibit a relatively

    high input impedance to help decrease the eects o

    connection quality.

    Magnetic owmeter coils can use and store sig-nicant amounts o energy relative to the amount o

    energy needed to cause ignition. Most magnetic ow-

    meter transmitters are designed to be non-incendive, so

    normal transmitter operation will not cause ignition.

    However, when installed in some hazardous locations,

    ormal approval is required, and the transmitter must

    be designed and installed to address the hazard.

    A hazard may be present not only in the general

    location o the primary and transmitter, but also inside

    the pipe where the electrodes can provide a source o

    ignition. o mitigate this hazard, the circuits o some

    designs limit the energy available at the electrodes to an

    amount less than that required or ignition.

    Maintaining equipment is simplied when sel-

    diagnostics are available to help the user. Te extent

    and quality o the diagnostics and their ease o use

    varies by manuacturer. Changing ranges is easier

    and more accurately perormed in a digital manner.

    Potentiometer adjustments and step switches are more

    prone to problems.

    otHer ConSIderAtIonS

    Te market or magnetic owmeters is competitive,so prices are relatively low. Magnetic owmeters or

    water and wastewater service can be economical due

    to the economies o scale and the relatively low cost o

    liners and electrodes or this service. However, applying

    magnetic owmeters to corrosive or abrasive services

    can signicantly increase the cost o the meters.

    Magnetic owmeters or use in the chemical in-

    dustry are typically more expensive than vortex shed-

    ders. In some applications, the cost can rival that o

    turbine owmeter or orice-plate owmeter systems.

    Magnetic owmeters are typically more economical

    than Coriolis mass owmeters.

    tPrevious Page Next Pageu

  • 8/22/2019 Flow eHandbook

    9/12 11

    FLowMeter PerForMAnCe

    Te purpose o installing a owmeter system is to

    accurately measure ow in a reliable manner. Issues

    related to physical properties, process parameters,

    electronic eatures and interconnections are oten

    given much consideration. Relatively little empha-

    sis, however, is given as to how well the owmeter

    will perorm its intended purpose. Adding to the

    conusion are the dierences in how perormance isexpressed and the incomplete nature o the avail-

    able inormation. Nevertheless, the quality o ow

    measurement should be a concern.

    Te perormance o a owmeter is quantied by

    its accuracy statements. Te reader must under-

    stand not only which parameter is being described,

    but also the manner in which the statement is

    expressed. In ow measurement, parameters are

    commonly described in terms o a percentage o

    the actual ow rate, a percentage o the ull scale

    o rate, or a percentage o the meter capacity. Tese

    terms are mathematically related, so it is possible to

    convert one to another (able 1). Note that when

    compared on a common basis, such as percent o

    rate, these statements describe signicantly dier-

    ent perormance.

    Other terminology may be used to express these

    concepts. When this occurs, conrm exactly what

    the other terms mean so they can be understood.

    Perormance statements apply to a range o

    ow or, stated dierently, between a minimum and

    maximum ow velocity. It is important to identiy

    the range in which the statement applies becauseperormance can be signicantly degraded or

    undened when the owmeter operates outside o

    this range.

    Complicating the issue are some owmeters that

    have dierent perormance statements or dier-

    ent measurement ranges. For example, a owmeter

    may have a reerence accuracy o 0.25% o rate

    rom velocities o 1 to 10 m/s, and an absolute er-

    ror o 0.0025 m/s rom 0.1 to 1 m/s. Perormance

    is undened below 0.1 m/s. able 2 describes this

    perormance using the above inormation. Note how

    perormance degrades at low ows.

    PerForMAnCe CLAIMS

    For the most part, the claims made by suppliers

    regarding magnetic owmeters are true statements,

    even though they may seem extraordinary. Te

    problem is that the statement may be incomplete,

    and may not include certain acts and inormation

    that clariy the statement. Sometimes claims are

    simplied or convenience and easier understand-

    ing. However, in many cases, urther investigationmay reveal other motives or doing so.

    For example, consider a magnetic lowmeter

    that has a reerence accuracy o 0.25% o rate and

    a turndown o 1,000:1. he implication is that

    the lowmeter can measure within 0.25% o rate

    over a 1,000:1 range o low. aken individua lly,

    both parts o the claim are likely true statements.

    Yet when combined, t hey can be misle ading by

    omission. Further investigation will show that

    the reerence accuracy o 0.25% o rate applies

    only within a range o low rates. Below the

    minimum low rate o the range, the reerence

    accuracy becomes a ixed absolute error. So as the

    low rate decreases, the accuracy expressed as a

    percentage o rate will increase.

    Assuming that the reerence accuracy o 0.25%

    o rate applies between 5% and 100% o meter

    capacity, and that between 0.1% and 5% o meter

    capacity, the reerence accuracy is xed at the abso-

    lute error at 5% o meter capacity. able 3 calculates

    reerence accuracy throughout the range o ows.

    Tis illustrates that above 0.5 m/s, the reerence

    accuracy is 0.25% o rate and that the turndown is10/0.01 or 1,000:1, both a s claimed. What is not

    stated in the claim is that the reerence accuracy

    degrades below 0.5 m/s and can approach 12.5%

    o rate. Also not stated is that in actual installa-

    tions, ows near meter capacity would rarely be

    encountered, so the 1,000:1 turndown would rarely

    be achieved. Assuming a more reasonable ull-scale

    calibration range o 0 to 2 m/s, this owmeter

    would achieve a 0.25% o rate reerence accuracy

    rom 0.5 to 2 m/s, or a 4:1 turndown, and only a

    200:1, or 2/0.01, turndown when the stated peror-

    mance at low ow rates is included.

    tPrevious Page Next Pageu

  • 8/22/2019 Flow eHandbook

    10/1212

    In addition to high turndown, some suppliers

    claim that their lowmeter operates at extremely

    low low rates. Consider a claim to measure

    velocity as low as 0.01 m/s. For a meter with a

    capacity o 10 m/s, this corresponds to a 1,000:1

    turndown. Although the lowmeter may operate

    at this low rate, able 3 shows that it does so

    with a reerence accurac y o 12.5% o rate.

    Statements about magnetic owmeters otenclaim high reerence accuracy. What oten is not

    stated is that it may apply over a range o higher

    ows, and much o this range may not be encoun-

    tered in actual operation. Furthermore, the reer-

    ence accuracy a s a percentage o rate generally

    degrades or is undened at lower ow rates (see

    tables). When the ca librated ull-scale is low, and

    the high reerence accuracy statements are limited

    to a small range o high ow rates, the stated

    reerence accuracy may not be achieved.

    In general, reerence accuracy should be clear-

    ly and completely stated or all ow rates prior to

    perorming any analysis. Te range o applicabil-

    ity o the high accuracy statement and the actual

    operating ow range should be compared.

    Magnetic owmeters are among the most

    versatile o owmeter technologies. However, the

    user should be aware o the manner in which their

    application and operation are described in orderto ensure that the proper magnetic owmeter is

    selected and installed.

    dAvId w. SPItzer has more than 25 years o experience in

    speciying, building, installing, commissioning and trouble-

    shooting process-control instrumentation. Spitzer is a principal

    in Spitzer and Boyes LLC, which oers consulting services or

    the process industries in addition to product development,

    marketing and distribution consulting or manuacturing and

    automation companies.

    tPrevious Page Next Pageu

    http://www.burkert-usa.com/
  • 8/22/2019 Flow eHandbook

    11/12 13

    tak a diff Lk a Cifal PmpsAn unconventional assessment can provide insights or eective control.

    By Andrew Sloley, Contributing Editor

    oFten, CrItICAL understanding o a systemcomes rom turning the common analysis on its

    head. With centriugal pumps, this means thinking

    that ow results rom back-pressure on the pump

    discharge, not that pump discharge pressure varies

    with ow rate.

    Tis unconventional approach was crucial in

    addressing a troublesome control system or the

    overhead o a natural gas liquids plants debutanizer,

    which goes to an accumulator downstream o a reux

    pump (Figure 1). Varying the rate o liquid product

    controls tower pressure. A dual-range controller

    handles the reux drum pressure -- one valve lets in

    uel gas to pressurize the system when pressure drops,

    another vents the drum when pressure rises to too

    high a level.

    Tis rather odd system did not work. Te reux

    pump cavitated all the time and pressure control was

    erratic.

    Te owner, which acquired the unit during a

    company buy-out, lacked tower drawings, exchanger

    inormation, pump curves, control valve inorma-

    tion and historical operating data. Current operating

    personnel had no experience with the unit and neverhad seen it work stably.

    Lack o inormation doesnt justiy ignoring the

    problem. So, lets examine this systems undamentals

    and explore the most serious shortcomings.

    Centriugal analysis starts by looking at two

    things: the system curve and the pump curve. Te

    system curve is the head loss required versus ow rate

    through the system. Te pump curve is the dynamic

    head generated by the centriugal pump. Te intersec-

    tion o the system and pump curves denes the ow

    rate the system will get.

    We most commonly attain the required ow rate by

    adjusting the system curve by adding an extra pressuredrop via a control valve. Alternatively, we can change

    the pump curve using an adjustable speed drive.

    In Figure 1, the reux control valve is a hand-op-

    erated control valve (HCV). Te reux system doesnt

    include an automated pressure drop. It essentially has

    a xed system curve. Tis brings us to thinking about

    the pump operation: pump ow stems rom back-

    pressure on the pump.

    Now, lets consider the tower pressure-control

    system. PC3 adjusts the product ow out o the

    system with the intent o changing the liquid level

    in the condensers. Varying wetted condenser area

    on the process side allows or pressure control. Tis

    is a simple, ast-acting and eective system or total

    condensation services.

    Meanwhile, the product drum pressure-control

    system maintains a constant destination pressure or

    the net product rom the reux pump.

    Te static head to the top o the tower ar exceeds

    the pressure change between the tower and product

    drum.

    Te problem comes rom how the systems interact.

    Te pressure control system requires level to existsomewhere up in the heat exchangers. Tink o the

    exchangers and piping as a tall narrow vessel -- i more

    liquid exits a vessel than goes in, the level drops, and

    vice versa. Te pressure control system should perorm

    similarly to a tight level control system.

    Te system curve or the reux stream includes

    two components: static head and system pressure

    drop. Static head doesnt change with ow rate, but

    system pressure drop does. I static head makes up

    most o the system curve, the curve is relatively at.

    Flat system curves create large ow rate changes rom

    small pressure drop changes.

    tPrevious Page Next Pageu

  • 8/22/2019 Flow eHandbook

    12/12

    Te ow rate out o the vessel (heat exchangers

    and piping) is very sensitive to changes in the HCV

    position. Te HCVs purpose is to generate enough

    pressure drop in the reux line so the unit can operate

    in a sweet spot where the pressure control system will

    work. In this case, we suspect the sweet spot is toosmall. Te back-pressure on the pump imposed by

    the HCV usually is too low. At low back-pressure the

    pump capacity exceeds the liquid rate. More liquid

    is leaving the vessel (heat exchangers and piping)

    than going in. Liquid level drops quickly. Finally, the

    pump cavitates.

    How can we address this problem?

    One way is to try to put as much dynamic pres-

    sure drop on the HCV as tolerable. Tis makes the

    reux system curve steeper, which gives more stable

    control. Tis is cheap and quick.

    A second, and better, approach automates the

    HCV. Control systems should transer a disturbance

    rom where its important to where its not. Whats

    important here is the ow rate out o the vessel

    -- so we can maintain tight level control. We must

    move the disturbance to something unimportant.

    Many dierent congurations are possible. Tecloud in Figure 1 shows one o the simplest and

    easiest options. A strap-on ultrasonic ow meter

    along with a bolt-on actuator on the HCV enables

    ully automated control o the overhead system. Te

    disturbance now is in the valve pressure drop -- an

    unimportant spot.

    Tere are other ways to approach this problem.

    But a undamental understanding o the system

    comes rom looking at the pump backwards.

    Andrew SLoLeyis a Chemical Processing Contributing

    Editor. You can e-mail him at [email protected]

    Debutanizer OverheaD

    Figure 1. Lack o an automated pressure drop in the reux system made control difcult.

    mailto:[email protected]:[email protected]