Fluoroscopic Radiation Exposure in Spine Surgery
Robert E. Isaacs, MDDirector of Spine Surgery
Associate ProfessorDivision of Neurosurgery
Duke University
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Disclosures
• Consultant– Baxano– Nuvasive– Vertera
• Stock Holder– SafeWire– VilaSpine– SafeRay Spine– Vertera– Providence
• Scientific Advisory Board– Providence
• Travel Reimbursement– Vertera
– ACSR
• Research Support– Nuvasive
• Royalties– Nuvasive
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
The ProblemIonizing radiation exposure during
interventional medical procedures using fluoroscopy is a potential health threat to
everyone near the x-ray source
5/27/2016
2
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Deterministic risks (e.g., cataract, osteonecrosis)
Radiation Induced Cataract
Threshold < 500 mGy
Long-term stochastic risks (e.g., cancer)
5/27/2016
3
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
But, What About Spine Surgery?
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Radiation is Bad…
“extent of radiation exposure in patients with AIS may elevate the lifetime risk of solid cancers to between 1.4% and 2.4%.”
“78% of the annual radiation exposure for operative patients occurs intraoperatively”
The Spine Journal 14 (2014) 1984–1990
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
• Prospective study measuring surgeon radiation exposure in MIS TLIF, – mean fluoroscopy time of 1.69 minutes per case– mean radiation exposure to the surgeon’s torso (under a
lead apron) of 27 mrem per case.
• Average to patient was 138 mGy
5/27/2016
4
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
A Single Dose of 100 mGy
HEALTH RISKS FROM EXPOSURE TO LOW LEVELS OF IONIZING RADIATIONCommittee to Assess Health Risks from Exposure to Low Levels of Ionizing Radiation - NATIONAL RESEARCH COUNCIL OF THE NATIONAL ACADEMIES
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Compton Scatter
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Increased cancer risk among surgeons in an orthopaedic hospital
• One Ortho Hospital in Italy, 1976-2000• Ortho – 29% risk• Exposed (not surgeons) – 6%• Not Exposed – 4%• Ortho surgeons – 5X increased risk of cancer
• Mastrangelo G, Fedeli U, Fadda E, Giovanazzi A, Scoizzato L, Saia B. Increased cancer risk among surgeons in an orthopaedic hospital. Occup Med (Lond). 2005;55:498-500.
5/27/2016
5
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
• Spinal surgeons can sustain a 10- to 12-fold increase in radiation exposure as compared with surgeons using fluoroscopy for nonspinal procedures.
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
But I’m Protected by Lead?
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
• Putting in OPEN fluoroscopically assisted pedicle screws gives the surgeon 13.49 mSv of whole body occupational exposure to ionizing radiation during a typical year
• A surgeon beginning his/her career at age 30 years would exceed the lifetime limit for nonclassified workers in less than 10 years.
5/27/2016
6
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
How Many Of You Know Your Dosimeter Readings?
In fact, how many of us wear a Dosimeter?
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Results
• 59 MIS fusions were performed over 12 months
• 376 screws were placed• Average radiation dose to the patient per
screw placed was 0.50 Rem• Average radiation dose to the surgeon per
screw placed was 1.01 mRem
Can physician radiation exposure during surgery be predicted based on the patient exposure? Idler, et al SMISS 2014
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Surgeon radiation exposure was found to have a linear relationship with patient exposure
Can physician radiation exposure during surgery be predicted based on the patient exposure? Idler, et al SMISS 2014
Pati
ent
Rem
/scr
ew
5/27/2016
7
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
What About Lead?
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
• Above the Thyroid Shield - 13.49 mSv/yr • Below the Thyroid Shield - 4.31 mSv/yr • So, lead gives a 2/3rds reduction
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
• Lead protects what is closer to the x-ray source, but the radiation that gets through is substantial
• At the waist PROTECTED by a lead apron was 27 mRem
• At the UNPROTECTED thyroid level was 32 mRem
5/27/2016
8
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
The SolutionUse Pulse and/or Low Dose Imaging to achieve ALARA – radiation dosing “As Low As Reasonably Achievable”
5/27/2016
9
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
1 PLD is Statistically Equivalent to Lead
0
1
2
3
4
5
6
7
1 2 3 4 5 6
0
5
10
15
20
25
30
35
40
45
50
PLD Under Lead Over Lead
Everyone in the room, including the patient!
5/27/2016
10
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
The American College of RadiologyThe Society for Pediatric Radiology
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Standard Dose
1 Pulse & Low Dose
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Radiologists, medical physicists, registered radiologist assistants, radiologic technologists, and all supervising
physicians have a responsibility for safety in the workplace by keeping radiation exposure to staff, and to society as a whole, “as low as reasonably achievable” (ALARA) _
5/27/2016
11
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Radiologists, medical physicists, registered radiologist assistants, radiologic technologists, and all supervising
physicians have a responsibility for safety in the workplace by keeping radiation exposure to staff, and to society as a whole, “as low as reasonably achievable” (ALARA) and to
assure that radiation doses to individual patients are appropriate, taking into account the possible risk from
radiation exposure and the diagnostic image quality necessary to achieve the
clinical objective.
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
“Detail of the pedicle is lost with the low‐dose pulsed fluoroscopy”
“For finer detail, specifically when determining the medial boundary of the pedicle on an AP view, digital spot is preferable.”
5/27/2016
12
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Confidential
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
What Is LessRay?Pulse and Low Dose degrade image quality - LessRay can improve it to
close to full dose imaging
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Components of LessRay
5/27/2016
13
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Set Up
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Standard Dose
1 Pulse & Low Dose
LessRay Enhanced1 Pulse & Low Dose Image
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
LessRay:In clinical practice, the amount of image quality improvement achieved when a Pulsed and/or Low Dose image is processed with LessRay is dependent on the clinical task, patient size, anatomical location, and clinical practice. The dose should be set at a level to which the physician is able to achieve the adequate image quality needed for the particular clinical task. A consultation with a radiologist and a physicist may aid in determining the appropriate dose settings.
5/27/2016
14
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
The Technology
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
5/27/2016
15
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
5/27/2016
16
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Low Dose
LessRay Conventional
Not 1 Pulse / Low DoseBut at
Pulse, 0.2 mA
Preservation of Image Quality with 92% Less Radiation
Images
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Spine Surgeon = Non-Classified Worker
NCRP
Defines as (Age x 3.33mSv)
0.4712 mSv per Case (Some were multilevel)
If You Began Performing Kyphoplasties at age 30 and You Did Only 30 Cases/year AND NO OTHER Fluoroscopically Guided Surgeries
Would Exceed Your Total Lifetime Limit by age 38!!
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Cadaveric Study • Conventional Fluoroscopy vs. LessRay
• 7 Fellowship Trained Orthopedic Spine Surgeons / Neurosurgeons
• OEC C-Arms set to 1 Pulse & Low Dose
•The Lowest Automatic Setting
Reducing Radiation an Order of Magnitude During X-Ray Intensive Spinal Procedures Using a Standard Fluoroscope I Karikari; C Brown; D.G. Anderson, SMISS 2014
5/27/2016
17
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
LessRay® Enhanced
Unenhanced 1 Pulse & Low Dose
Reducing Radiation an Order of Magnitude During X-Ray Intensive Spinal Procedures Using a Standard Fluoroscope I Karikari; C Brown; D.G. Anderson, SMISS 2014
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
LessRay® Enhanced
Unenhanced 1 Pulse & Low Dose
Reducing Radiation an Order of Magnitude During X-Ray Intensive Spinal Procedures Using a Standard Fluoroscope I Karikari; C Brown; D.G. Anderson, SMISS 2014
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Results In no case did any of the physicians abandon the
LessRay enhanced images.
The average duration of fluoroscopy time for the entire procedure was 65.6s vs. 11.0s (Conventional vs LessRay-Assisted, p<0.001).
Statistically similar number of images with both methods prior to cement injection (24 vs. 21, p=0.30)
The pulsed/low dosed procedures achieved an overall
88.8% radiation reduction over conventional imaging (36.1mGy vs. 4mGy, p< 0.001).
Reducing Radiation an Order of Magnitude During X-Ray Intensive Spinal Procedures Using a Standard Fluoroscope I Karikari; C Brown; D.G. Anderson, SMISS 2014
5/27/2016
18
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Radiation Per Procedure
mG
y
Reducing Radiation an Order of Magnitude During X-Ray Intensive Spinal Procedures Using a Standard Fluoroscope I Karikari; C Brown; D.G. Anderson, SMISS 2014
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Digitally improvement allows the physician to use the lowest radiation settings on a C-Arm
Significantly reduces the amount of radiation exposure during a radiation intensive medical procedure when compared with conventional fluoroscopy
With the aid of image enhancement software like LessRay®, a fluoroscope can provide high quality images
with an order of magnitude less radiationwe are currently being subjected to.
Conclusions:
Reducing Radiation an Order of Magnitude During X-Ray Intensive Spinal Procedures Using a Standard Fluoroscope I Karikari; C Brown; D.G. Anderson, SMISS 2014
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
LessRay
Metal Elimination
5/27/2016
19
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
5/27/2016
20
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Conclusions
Patented / Patent Pending© SafeRay Spine, LLC 2013; all rights reserved
Lessray: ‐Improve Low Radiation Images