FOCAS-2000, MEPC107(49)

MAINTENANCE & CHECKING

Authorized by:

Page 2: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

SPECIFICATIONS

ALARM CONTACT OUTPUT ,ALARM LAMP MOVEMENT VIEW

1

Page 3: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING TEST OPERATION METHOD BY THE AUTOMATIC

OPERATION FUNCTION

2

Page 4: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING TEST OPERATION METHOD BY THE AUTOMATIC

OPERATION FUNCTION

3

Page 5: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

SPECIFICATIONS

TAMBULATED LIST OF MENU ITEMS

4

Page 6: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

SETTINGS DATA AND RANGES

5

Page 7: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

OPERATION & MAINTENANCE

It is recommended for you to periodically check and maintain FOCAS-2000 as shown below to assure its performance for a long

time.

6

Page 8: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING ALARM RELAY TEST

1. Screen must be in stand-by mod

2. Press ▲ or ▼ button until you will see on the screen the below.

3. Press “Enter” and “A1=OFF” become “A1=ON”

(If alarm is connected to Alarm 1 terminal, visual & audio alarm will energize)

4. Press “Enter” and “A1=ON” -> “A1=OFF” & “A2=OFF” -> “A2=ON”

5. (If alarm is connected to Alarm 2 terminal, visual & audio alarm will energize)

6. Press “Enter” and “A1=OFF” -> “A1=ON” & “A2=ON” -> “A2=ON”

7. Press “Enter” and “A1=ON” -> “A1=OFF” & “A2=ON” -> “A2=OFF”

8. Press “Esc.” to return to Stand-by screen

Kindly revert with Visual (on the screen) & Audio (Engine room) alarm condition after the

test.

7

Page 9: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

1. Clean unit’s cell thoroughly.

2. Drain sensor and fill in with fresh mineral water.

3. Screen must be in stand-by mode

4. Press ▲ or ▼ button until you will see on the screen the below values.

5. Take a clear photo of the values and revert to us.

NOTE: It is very important to fill the Cell with a Fresh Mineral water (drinkable).

8

Page 10: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING CLEANING THE SENSOR

9

Page 11: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING CLEANING THE FILTERS

10

Page 12: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING REPLACING THE SILICA GEL

REPLACING THE CLOCK BACK UP BATTERY

11

Page 13: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING REPLACING THE OPERATIONAL LOG MEMORY CARD

12

Page 14: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING REPLACING THE OPERATIONAL LOG MEMORY CARD

13

Page 15: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

MAINTENANCE & CHECKING REPLACING THE OPERATIONAL LOG MEMORY CARD

14

Page 16: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

FOCAS-2000 Sensor unit Replacement kit Replacement procedure manual l

1

Replacement Procedure Manual 2000-210

15 ppm Bilge Alarm FOCAS-2000 SENSOR UNIT REPLACEMENT KIT Rev. 1.01

Thoroughly read this manual to properly use this kit.

Cautions -1 If a 10 copper pipe is currently used for the detection section drainage outlet, first replace

the pipe

with a 10 joint having been used and then replace the whole detection section.

-2 Drain and discard water inside the sensor unit into the waste solution vessel separately

prepared.

-3 Do not apply excessive impact to the kit in order to avoid damage.

Content of the Kit

Sensor unit

(with hex cap nut

 and copper ball)

The following articles are

 separately requied.

 ・Cross slot screwdriver(♯2)

 ・Vessel for waste solution

 ・Unused waste cloths

 ・Open end wrench in 15 mm width across flats

  or adjustable end wrench (200 mm)

・Precision certificate

for sensor unit

・Replacement procedure manual

(in Japanese/in English)

Sensor unit replacement

Front view

Open

Esc ▼ ▲ Enter

⇒AL.2POWER

MODEL FOCAS-2000

AL.1

15ppm Bilge Alarm

+ +

+

+

+

+

+ + 2,Sensor unit

4,Solenoid valve unit

5,Drain

12,Detection cable

7,Fresh water valve

6,Sample water valve

1,Upper lib

3,Detection section- solenoid valve hex

cap nut

14,Detection section fixing screw

15,Solenoid valve section fixing screw

13,Knurled screw

※Accuracy certificate label

11,Operation switch

10,Display

9,Power Lamp

8,Sample water outlet

15

Page 17: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

FOCAS-2000 Sensor unit Replacement kit Replacement procedure manual l

2

(1) Shut off the supply power to FOCAS-2000 and verify the Power lamp (Item 9 in the front view) gone

off.

(2) Remove two detection cables (Item 12 in the front view).

* For removal, release the connector lock.

(3) Stop supplying the sample water and fresh water by closing corresponding valves (Item 6, 7 in the

front view).

Remove the piping to the Sample water outlet (Item 8 in the front view),

(4) Remove screw from the drain (Item 5 in the front view) by using a #2 cross-slot screwdriver, remove

the upper lid (Item 1 in the front view), and then drain water inside the sensor unit.

Upon completing draining, restore the screw into the drain..

Set the detection section – solenoid valve hex cap nut (Item 3 in the front view) such that

the nut is ready to be removed by using an open end wrench (15 mm width across flats) or adjustable

end wrench. (5) Loosen the detection section fixing screws (Item 14 in the front view) by using a #2 cross-slot

screwdriver for being ready to be removed, and remove the sensor unit by loosening the solenoid

valve section fixing screws (Item 15 in the front view).

If the sensor unit is difficult to be removed, first tighten the knurled screw (Item 13 in the front

view) inside the casing for making ready to remove the sensor unit and then remove the unit by

loosening said knurled screw.

* The fixing screws do not drop, thanks to the drop prevention structure.

(6) Upon verifying the replacing sensor unit (Item 2 in the front view) being mounted with the hex cap

nut and copper ball, mount the section at the location in alignment with the solenoid valve joint.

Upon verifying that four detection section fixing screws can be mounted without any difficulty,

tighten them together with four solenoid valve section fixing screws.

Upon the completion of tightening total eight screws, lightly tighten the knurled screw. (7) Mount two detection section cables.

(8) Turn on the supply power, verify the power lamp lighting, and then

open respective valves corresponding to the sample water and fresh water.

(9) Upon replacement warning being displayed (figure on the right) on the display (Item 10 in the

front view), press the [Enter] key of the operation switch (Item 11 in the front view) to return

to standby, establishing the status ready for operation.

(10)When verifying operation, please use the『Kit for Operation Verification』

(provided in the same package)

Head Office: 6-13-6, Nishinippori, Arakawa-ku, Tokyo 116-0013, Japan

Phone: 03-3800-9777 Website: http://www.fellow-co.com

Fax: 03-3800-9770 E-mail: [email protected]

Tokyo Factory: 6-13-9, Nishinippori, Arakawa-ku, Tokyo 116-0013, Japan

Rev 1.01: December 2014

Connector locking function

Warning〔178〕Sensor was Changed

16

Page 18: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

TROUBLESHOOTING

17

Page 19: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

SERVICE PARTS

18

Page 20: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

DISPLAY OF GROUPS AND DESCRIPTION STATUSES

19

Page 21: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

DISPLAY OF GROUPS AND DESCRIPTION STATUSES

20

Page 22: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

EDIT AUTOMATION LTD 15 & 17 Akti Kountouriotou, Piraeus, 18534, Greece

Tel.: +210 4110007 [email protected] www.edit.com.gr

Notice of Discontinuation

Internal Unit for FOCAS-1500C and FOCAS-1500E:

Oil Concentration Meter (15PPM Bilge Alarm)

21

Page 23: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

FOCAS-1500 TO FOCAS-2000 ELECTRICAL CONNECTIONS MANUAL

All Focas 15PPM units usually have 3 sets of cable connections.

1. Unit’s power supply. 2. Visual and sound alarms. 3. a) As per MEPC 60(33), one NC contact to stop the Oil Bilge Pump.

b) As per MEPC 107(49), terminal with output voltage which energize the 3Way solenoid valve.

Depending on the vessel’s system please proceed to the connections of Focas-2000 according Drawing 1.3.

If your vessel is:

3a) As per MEPC 60(33) cables which stop the Bilge Pump are connected to “OIL ALARM 2 COM & N.C.” In that case in Focas-2000 the cables should be connected to “Alarm 2 COM & NC”.

3b) As per MEPC 107(49) cables for the solenoid valve connection are in terminal “PS S1 & S2”. In that case in Focas-2000 the cables should be connected to “Valve AC OUT”.

Note 1: For your convenience, probably your good-vessel’s system is as per MEPC 60(33).

Note 2: For any further help you may need with the connection of the unit, please send us a photo of your old unit terminal connection so we can guide you with more specific information about the connections.

22

Page 24: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

23

Page 25: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

24

Page 26: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

25

Page 27: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

ABS Quality Evaluations Certificate Of Conformance

This is to certify that the Quality Management System of:

E.D.I.T Automation Ltd

15 & 17 AKTI KOUNTOURIOTOU STR.,

Kastela

Piraeus, Attiki 185 34

Greece

hasbeen assessed by ABSQuality Evaluations, Inc. and found to be in conformance with the requirementsset forth by:

ISO 9001:2015

The Quality Management System is applicable to:

MANUFACTURE, REPAIRS AND SALES OF MARINE & INDUSTRIAL AUTOMATION SYSTEMS. SALES,

MAINTENANCE AND CALIBRATION OF MARINE ENVIRONMENTAL PROTECTION SYSTEMS. TRAINING ON THE

USE OF AUTOMATION AND ENVIRONMENTAL PROTECTION SYSTEMS.

This certificate may be found on the ABS QE Website (www.abs-qe.com). For certificates issued in the People's Republic of China information may also be

verified on the CNCA website (www.cnca.gov.cn).

Certificate No: 42956

Certification Date: 02 February 2012

Effective Date: 15 February 2019

Expiration Date:

Revision Date:

11 March 2022

05 July 2019

Dominic Townsend, President

Validity of this certificate is based on the successful completion of the periodic surveillance audits of the management system defined by the above scope and is contingent upon

prompt ,written notification to ABS Quality Evaluations, Inc. of significant changes to the management system or components thereof.

ABS Quality Evaluations, Inc. 1701 City Plaza Drive, Spring, TX 77389, U.S.A.

Validity of this certificate may be confirmed at www.abs-qe.com/cert_validation. Copyright 2011-2019 ABS Quality Evaluations, Inc. All rights reserved.

Page 28: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING

ABS Quality Evaluations Certificate Of Conformance

This is to certify that the Environment Management System of:

E.D.I.T Automation Ltd

15 & 17 AKTI KOUNTOURIOTOU STR.,

Kastela

Piraeus, Attiki 185 34

Greece

hasbeen assessed by ABSQuality Evaluations, Inc. and found to be in conformance with the requirementsset forth by:

ISO 14001:2015

The Environment Management System is applicable to:

MANUFACTURE, REPAIRS AND SALES OF MARINE & INDUSTRIAL AUTOMATION SYSTEMS. SALES,

MAINTENANCE AND CALIBRATION OF MARINE ENVIRONMENTAL PROTECTION SYSTEMS. TRAINING ON THE

USE OF AUTOMATION AND ENVIRONMENTAL PROTECTION SYSTEMS.

This certificate may be found on the ABS QE Website (www.abs-qe.com). For certificates issued in the People's Republic of China information may also be

verified on the CNCA website (www.cnca.gov.cn).

Certificate No: 57133

Effective Date: 25 May 2018

Expiration Date:

Revision Date:

24 May 2021

05 July 2019 Dominic Townsend, President

Validity of this certificate is based on the successful completion of the periodic surveillance audits of the management system defined by the above scope and is contingent upon

prompt ,written notification to ABS Quality Evaluations, Inc. of significant changes to the management system or components thereof.

ABS Quality Evaluations, Inc. 1701 City Plaza Drive, Spring, TX 77389, U.S.A.

Validity of this certificate may be confirmed at www.abs-qe.com/cert_validation. Copyright 2011-2019 ABS Quality Evaluations, Inc. All rights reserved.

Page 29: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING
Page 30: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING
Page 31: FOCAS-2000, MEPC107(49) MAINTENANCE & CHECKING