fractionators - medica · max tmp (mmhg) 500 housing internal diameter (mm) : 32 (1.0 m2) and 44...

2
NEW CASCADE FILTRATION ADVANTAGES Wide range of application Minimal Loss of non-pathogenic substances like albumin Selective apheresis of plasma components based on molecular weight MEDICA hollow fiber membrane specially developed for plasma fractionation It is a registered trade mark of Medica S.p.A is the MEDICA S.p.A. Via degli Artigiani, 7 41036, Medolla (Modena) Italia Phone +39 0535 51159 Fax + 39 0535 52605 [email protected] www.medica.it CE 0476 UNI EN ISO 9001: 2015 UNI CEI EN ISO 13485: 2016 MANUFACTURER Distributed by Technical features are subject to change without prior notice This document does not have contractual value M27333 REV 02 CDP130418C MEMBRANE Material Versatile ® -PES Polyethersulfone (PES) + Polyvinylpyrrolidone (PVP) Filtration area (m2) 1.0 and 2.0 Inner diameter (μm) 220 Outer diameter (μm) 300 Wall thickness (μm) 40 Max TMP (mmHg) 500 HOUSING Internal diameter (mm) : 32 (1.0 m 2 ) and 44 (2.0 m 2 ) Potting material: Polyurethane (PU) Cartridge and blood port material: Polycarbonate (PC) T otal filter length (mm) : 290 STERILISATION Beta Rays TECHNICAL SPECIFICATIONS AND DATA FRACTIONATORS

Upload: others

Post on 30-Apr-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FRACTIONATORS - Medica · Max TMP (mmHg) 500 HOUSING Internal diameter (mm) : 32 (1.0 m2) and 44 (2.0 m2) Potting material: Polyurethane (PU) Cartridge and blood port material: Polycarbonate

NEW

CASCADE FILTRATION ADVANTAGES

Wide range of application

Minimal Loss of non-pathogenic substances like albumin

Selective apheresis of plasma components based on molecular weight

MEDICA hollow fiber membrane specially developed for plasma fractionationIt is a registered trade mark of Medica S.p.A

is the

MEDICA S.p.A. Via degli Artigiani, 7 41036, Medolla (Modena) Italia

Phone +39 0535 51159Fax + 39 0535 [email protected]

CE 0476 UNI EN ISO 9001: 2015UNI CEI EN ISO 13485: 2016

MANUFACTURER

Distributed by

Tech

nic

al fe

atu

res

are

sub

ject

to

ch

ange

wit

ho

ut

pri

or

no

tice

T

his

do

cum

ent

do

es n

ot

hav

e co

ntr

actu

al v

alu

eM

27

33

3 R

EV

02

CD

P1

30

41

8C

MEMBRANE

MaterialVersatile®-PES

Polyethersulfone (PES) + Polyvinylpyrrolidone (PVP)

Filtration area (m2) 1.0 and 2.0

Inner diameter (µm) 220

Outer diameter (µm) 300

Wall thickness (µm) 40

Max TMP (mmHg) 500

HOUSING

Internal diameter (mm) : 32 (1.0 m2) and 44 (2.0 m2)

Potting material: Polyurethane (PU)

Cartridge and blood port material: Polycarbonate (PC)

Total filter length (mm) : 290

STERILISATION Beta Rays

TECHNICAL SPECIFICATIONS AND DATA

FRACTIONATORS

Page 2: FRACTIONATORS - Medica · Max TMP (mmHg) 500 HOUSING Internal diameter (mm) : 32 (1.0 m2) and 44 (2.0 m2) Potting material: Polyurethane (PU) Cartridge and blood port material: Polycarbonate

APPLICATIONS OF DOUBLE FILTRATION

removes a high molecular weight components from plasma. Immonoglobulin, cryoglobulin, low-density-lipoprotein, viruses are involved in inflammatory, autoimmune diseases and similar forms with acute or chronic decourse. We recommend the use of with our Apheresys Unit.

DOUBLE CASCADE FILTRATION

INDICATIONS

In vitro data (1,5 L plasma perfusion) | Plasma flow rate: 30ml/min

FRACTIOsmartTM Small is indicated for IgGApheresis to treat immunological disorders.

FRACTIOsmartTM Large is our proposal for LDL Apheresis, LDL apheresis is a form of apheresis, resembling dialysis, to eliminate the cholesterol-containing particle low-density lipoprotein (LDL) from the bloodstream.It is used in diseases featuring high LDL, such as the rare homozygous familiar hypercholesterolemia, when the heterozygous form does not respond to medical treatment, or when the treatment has led to dangerous side-effects (such as rhabdomyolysis).

FRACTIOsmartTM Medium is indicated for immune complexes and IgM Apheresis.

M90149 FRACTIOsmart TM Small 1.0M90099 FRACTIOsmart TM Small 2.0

Product Code

M90150 FRACTIOsmart TM Medium 1.0 M90100 FRACTIOsmart TM Medium 2.0

M90151 FRACTIOsmart TM Large 1.0M90101 FRACTIOsmart TM Large 2.0

Product Code

Product Code

0,00

0,10

0,20

0,30

0,40

0,50

0,60

0,70

0,80

0,90

1,00

< SIE

VING

COE

FFIC

IENT

S >

TOTAL PLASMA PROTEINS

FRACTIOsmartTM LargeFRACTIOsmartTM MediumFRACTIOsmartTM Small

0,00

0,10

0,20

0,30

0,40

0,50

0,60

0,70

0,80

0,90

1,00

BSA 66 kD IgG 150 kD IgM 950 kD

< SIE

VING

COE

FFIC

IENT

S>

PLASMA PROTEINS

0,00

0,10

0,20

0,30

0,40

0,50

0,60

0,70

0,80

0,90

BSA 66 kD IgG 150 kD IgM 950 kD

0,00

0,10

0,20

0,30

0,40

0,50

0,60

0,70

0,80

0,90

BSA 66 kD IgG 150 kD IgM 950 kD

0,00

0,10

0,20

0,30

0,40

0,50

0,60

0,70

0,80

0,90

BSA 66 kD IgG 150 kD IgM 950 kD

Cascade Filtration (CF) consists of a double filter system with the advan-tages of a selective removal without infusion of exogenous fluids. The sepa-rated plasma is pumped into a hollow fibers filter with a different nominal cut-off. The CF is effective in the cast nephropathy, in Waldenstrom’s mac-roglobulinemia, in active cryoglobu-linemia. CF is used successfully in Guil-lain-Barrè Syndrome (GBS), Myasthe-nia Gravis (MG), hypercholester-olemia, Focal Segmental Glomerulo-sclerosis (FSCS) and recently in the Age-related Macular Degeneration (AMD), Lupus Erythematosus, rhabdo-myolysis, acute renal failure due to Multiple Myeloma and others.

FRACTIOsmartTM LargeFRACTIOsmartTM MediumFRACTIOsmartTM Small