FT-ICR MS
An introduction and possible applications
Ariane Kahnt
Fourier transform ion cyclotron resonance mass spectrometry
![Page 2: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/2.jpg)
The technique
• Offers 10-100 time higher mass resolution, resolving power and mass accuracy than any other mass analysis technique
Balough 2004 (LC GC Europe)
Orbitrap
![Page 3: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/3.jpg)
• Hybrid instrument available in SBS: LTQ FT (i.e. ion trap - Fourier transform ion cyclotron resonance mass spectrometer from Thermo Scientific)
![Page 4: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/4.jpg)
• Basic set-up for all types of mass spectrometers
The set-up
www.premierbiosoft.com
![Page 5: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/5.jpg)
Sample introduction
• Hyphenation to liquid chromatography (LC) or direct infusion possible
LC
![Page 6: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/6.jpg)
Sample ionisation
• Electrospray Ionisation (ESI)
www.lamondlab.com
www.newobjective.com
![Page 7: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/7.jpg)
Ion transfer optics
• After the introduction and formation of gas-phase ions, charged analytes are transferred by various octapole and quadrupole lenses to the mass analyser
ESI
Q00
Q0 Q1
3 sets of octapoles
• Focussing and guiding of ions
![Page 8: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/8.jpg)
Ion trap
• The linear ion trap is a fully operational MS detector but is at the same time also an ion preparation and injection system for the ion cyclotron MS
ESI
• Ions can be stored, isolated and fragmented in the IT
![Page 9: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/9.jpg)
Ion cyclotron resonance analyser
• ICR cell is located in the centre of the magnetic field and traps gas phase ions in the centre using electric and magnetic fields
• Excitation of ions to a larger cyclotron radius
Barrow et al., 2004 (The Analyst)
![Page 10: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/10.jpg)
Ion cyclotron resonance analyser
Marshall and Hendrickson, 2008 (Annu. Rev. Anal. Chem.)
time-domain image-current signal
frequency spectrum mass spectrum
excited ion cyclotron rotation
![Page 11: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/11.jpg)
![Page 12: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/12.jpg)
Mass resolution versus resolving power
• Mass analyser must resolve adjacent peaks in complex samples
• Ability of an instrument to separate closely spaced peaks is called resolving power
• Resolution is calculated from the acquired data
• FWHM (full width at half maximum) definition
R
http://masspec.scripps.edu/Marshall and Hendrickson, 2008 (Annu. Rev. Anal. Chem.)
![Page 13: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/13.jpg)
Some more things to consider…
• Mass defect– Characteristic for each atom – Arises from nuclear binding energy
• Charge state
𝑚𝑑𝑒𝑓𝑒𝑐𝑡❑=𝑚𝑛𝑜𝑚𝑖𝑛𝑎𝑙−𝑚𝑒𝑥 𝑎𝑐𝑡❑
• Isotopic patterns in MS
http://masspec.scripps.edu/
![Page 14: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/14.jpg)
FT-ICR MS applications in biological sciences
• Protein characterisation
Thelen and Miernyk, 2012 (Biochem. J.)
• Metabolomics • Lipidomics
![Page 15: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/15.jpg)
Intact protein characterization
• Non-routine analysis for proteins >50 kDa • Usually performed on single proteins or modest mixtures• Can help in measuring variations of coding
polymorphisms, alternative splicing, diverse post-translational modifications
• “Top-down” proteomics
• Tandem mass spectrometry
![Page 16: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/16.jpg)
Protein isoforms and posttranslational modifications
• E.g. non-enzymatic deamidation of asparagine (and to a lesser extent glutamine) in-vivo and in-vitro have important biological effects
• Such as on enzyme activity, folding, proteolytic degradation
http://www.bumc.bu.edu/
![Page 17: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/17.jpg)
Capability of electron capture dissociation (ECD)
• ECD involves capture of a thermal electron by the protonated peptide/protein
Kelleher, 2004 (Anal. Chem.)
• Causes peptide fragmentation from N-Ca bond
- Production of N-terminal c-type and C-terminal z-type ions
![Page 18: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/18.jpg)
http://www.bumc.bu.edu/ftms/research/isoaspartome/
• Location of the deamidated Asn49 residue could be shown applying ECD
![Page 19: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/19.jpg)
Soldi et al., 2013 (Int. J. Mol. Sci.)
• Combinational approach
![Page 20: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/20.jpg)
Peptide and protein tagging methodologies
• Mass defect labelling for intact or digested proteins
Þ H is the dominant element responsible for the mass defect of a protein - BUT peptides cluster tightly in each individual mass unit
• Incorporation of mass defect labels to alter the mass defect and to obtain a shift in mass plus using specific isotope pattern of labels
Hernandez et al., 2006 (Anal. Chem.)
![Page 21: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/21.jpg)
• A well prepared sample, • well-defined analytical goal,• appropriate use of accurate mass,• reproducible retention times and • good instrument control generates
unassailable data! • “Make it run in triplicate, and [you] get real
data from which to draw conclusions.”
But most importantly….
Roy Martin from Waters
![Page 22: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/22.jpg)
Thank you!
![Page 23: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/23.jpg)
• Nominal mass: integer mass value
• Average mass:
• Monoisotopic mass: exact mass of the most abundant isotopes of each element
average atomic mass valuee.g. 1H and 2H (D) with R.A. of 100% and 0.0115% (H = 1.00794)
e.g. H =1
e.g. H =1.007825
Chemical formula: C153H224N43O49S
www.wikipedia.org
• Accurate mass: measured mass
• Exact mass: theoretical mass
![Page 24: FT-ICR MS An introduction and possible applications Ariane Kahnt Fourier transform ion cyclotron resonance mass spectrometry](https://reader036.vdocument.in/reader036/viewer/2022062304/56649ea05503460f94ba2610/html5/thumbnails/24.jpg)
• Instrument calibration!
• Error of mass accuracy measurements
• Mass accuracy (in general) depends on several parameters :– Resolving power– S/N ratio of peaks, peak shape, overlapping species?– Scanning method, scan rate
Balough, 2004 (LC GC Europe)