Fyrite® ProCombustion Gas Analyzer
Instruction 24-9388Operation & Maintenance
Rev. 6 – July 2009
Product Leadership • Training • Service • Reliability
WARRANTYBacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications. Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option, of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within one (2) years after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacharach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IM-PLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARIS-ING FROM BACHARACH’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies including, without limitation incidental or consequential damages. No agreement varying or extending the foregoing warranties, remedies or this limita-tion will be binding upon Bacharach, Inc. unless in writing, signed by a duly authorized officer of Bacharach.
A Instruction 24-9388
Register Your Warranty by Visiting www.mybacharach.com
Notice:
Product improvements and enhancements are continuous, therefore the specifications and information con-tained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in con-nection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Bacharach, Inc.
Copyright © 2002–2003, Bacharach, Inc., all rights reserved.BACHARACH and Fyrite are registered trademarks of Bacharach, Inc. All other trademarks,trade names, service marks and logos referenced herein belong to their respective companies.
Fyrite Pro
Instruction 24-9388 �
Contents
Ordering Information ...............................................................................iii
1.0 INTRODUCTION ..................................................................................11.1 TheFyr�tePro..................................................................................11.2 Operat�onalOverv�ew.....................................................................2
2.0 TECHNICAL CHARACTERISTICS ..................................................3
3.0 SETTING UP THE ANALYZER .........................................................53.1 Prel�m�narySteps............................................................................53.2 Check�ng&Replac�ngtheBatter�es..............................................53.3 Connect�ngtheProbeandThermocouple......................................63.4 FrontPanelPushButtons..............................................................83.5 SetupMode......................................................................................9
3.5.1 Enter�ngSetup..................................................................93.5.2 Select�ngTemperatureUn�ts............................................93.5.3 COChannelSetup(Model125Eonly)...........................103.5.4 DraftChannelSetup.......................................................103.5.5 T�meSetup.......................................................................113.5.6 DateSetup.......................................................................113.5.7 YearSetup.......................................................................123.5.8 Select�ngPr�nterProtocol...............................................123.5.9 Ex�t�ngtheSetupScreen................................................12
4.0 OPERATION .......................................................................................134.1 Perform�ngCOandDraftTest�ng................................................13
4.1.1 AnalyzerTurnOnandWarmUp...................................134.1.2 Measur�ngCO(Model125Eonly)..................................144.1.3 Measur�ngDraft..............................................................14
4.2 Perform�ngaCombust�onTest.....................................................154.2.1 AnalyzerTurnOnandWarmUp...................................154.2.2 Select�ngaFuel...............................................................154.2.3 Sampl�ngPo�nt................................................................164.2.4 Perform�ngaCombust�onEff�c�encyTest......................184.2.5 End�ngaCombust�onTest..............................................18
4.3 Operat�ngT�ps...............................................................................194.4 Us�ngtheBackl�ght.......................................................................204.5 Us�ngtheProbe.............................................................................204.6 Sav�ngTestData�nMemory........................................................204.7 Open�ngandV�ew�ngSavedTestData.......................................214.8 Pr�nt�ngTestData.........................................................................22
Fyrite Pro
Instruction 24-9388��
4.9 Clear�ngSavedTestData.............................................................244.10 Turn�ngOFFtheAnalyzer&COPurge.....................................244.11 Resett�ngtheM�croprocessor......................................................254.12 RunModeScreens........................................................................264.13 ScreenDescr�pt�ons.....................................................................28
4.13.1 Warm-UpScreensInclud�ngO2SensorOutput.........284.13.2 FuelScreen...................................................................284.13.3 TA(StackTemperature)&DraftScreen....................294.13.4 DraftZeroScreen.........................................................294.13.5 Stack&EFFScreen.....................................................304.13.6 Rat�oofCOtoCO2Screen(Model125Eonly)............304.13.7 qA&ExcessA�rScreen................................................314.13.8 TL&EtaScreen...........................................................314.13.9 CO&COA�rFreeScreen(Model125Eonly).............324.13.10 CO2&O2Screen.........................................................324.13.11 SensorErrorScreen...................................................33
5.0 CALIBRATION & MAINTENANCE ................................................355.1 Enter�ngtheCal�brat�onModeandTest�ng theD�splaySegments....................................................................355.2 Amb�entTemperatureCal�brat�on..............................................365.3 Combust�onA�rThermocoupleCal�brat�on.................................365.4 StackThermocoupleCal�brat�on..................................................375.5 OxygenSensorZero.......................................................................385.6 COSensorZero&Cal�brat�on(Model125Eonly).......................395.7 DraftCal�brat�on...........................................................................415.8 WaterTrap/F�lterMa�ntenance.................................................425.9 O2SensorReplacement.................................................................435.10 COSensorReplacement(Model125Eonly)...............................445.11 PumpAssemblyReplacement.....................................................45
6.0 PARTS & SERVICE ...........................................................................496.1 ReplacementParts........................................................................496.2 Accessor�es.....................................................................................496.3 Serv�ceCenters..............................................................................50
Appendix A – Formulas and Tables .......................................................51
Fyrite Pro
Instruction 24-9388 ���
Ordering InformationModel 122E 125E
Part Number 24-7276 24-7220
MeasurementsOxygen
Stack Temperature
Ambient Temperature
Second Thermocouple Channel (Combustion Air Temperature)
Carbon Monoxide (CO)
Draft
CalculationsqA (Stack Loss)
Combustion Efficiency
ETa
Excess Air (LAMBDA )
Carbon Dioxide (CO2)
CO Air Free
CO/CO2
Fyrite Pro
Instruction 24-9388�v
Notes:
Instruction 24-9388 1
Fyrite Pro
1.0 INTRODUCTION
1.1 The Fyrite ProTheFyr�teProser�esof�nstrumentsarehand-heldcombust�oneff�c�encyanalyzersthataredes�gnedforon-demandsampl�ngofres�dent�alfurnac-esandappl�ances.
Th�smanualconta�ns�nformat�ononallmodelsoftheFyr�tePro.Pleased�sregardany�nformat�onthatdoesnotperta�ntoyourmodel.
WARNINGS!
This analyzer is not intended to be used on a continuous basis.
This analyzer does not have an audible alarm, and is not intended to be used as a safety device.
Except for battery replacement, this analyzer should only be opened and serviced by authorized personnel.
When testing an appliance, a full visual inspection of the appliance should be carried out to ensure its safe operation.
Instruction 24-93882
Fyrite Pro
1.2 Operational OverviewPress�ngtheI/ObuttonturnstheanalyzerON.Notethatthere�sawarm-upper�odof60secondsbeforetheun�tcanbeusedfortest�ng.
Aftercomplet�onofthewarm-upper�od,chooseafuelcodethatcorre-spondstothefuelbe�ngburnedbytheappl�ancebe�ngtested,andthenpresstheeNTeR buttontoplacetheanalyzer�nto�tsRunMode.Beg�ntest�ngby�nsert�ngtheanalyzer’sprobetube�ntotheflue-gasstreamoftheappl�anceundertest.Theanalyzerw�llmon�torthefluegasandd�splaymeasuredandcalculatedvaluesthatarerelat�vetothecombust�onprocess.Thesevaluesared�splayedontheanalyzer’sLCD,andarechosenford�splaybypress�ngtheIncrement()andDecrement()buttons.
Press�ngthehOlD buttondur�ngatestw�llfreezeallmeasuredandcalculat-edvaluesatthe�rcurrentlevels.Press�ngtheeNTeR buttonresumestest�ng.
Press�ngthehOlD buttontwice dur�ngatestw�lld�splaythePr�nt/MemoryMenuScreen.Us�ngthe andeNTeR buttons,theoperatorcanchoosetopr�ntthecurrentread�ngs,savethecurrentread�ngs�nmemory,recallandv�ewsavedread�ngsfrommemory,clearallthememory,orpr�ntallsavedread�ngs.
Abackl�ghtenablesanoperatortoreadthed�splay�nd�mly-l�tareas.Press�ngthelIGhT buttonturnsthebackl�ghtONandOFF.Thebackl�ghtw�llautomat�callyturnOFFafter10m�nutesofkeyboard�nact�v�ty.
Apowersaverfunct�onw�llshuttheanalyzerOFFafter20m�nutesofkeyboard�nact�v�ty.Thepowersaverfunct�on�sd�sabled�fthedetectedO2value�slessthan17.9%,orforanalyzersw�thaCOsensortheCOvalue�smorethan50ppm.
PresstheI/ObuttontoturntheanalyzerOFF.Notethatthere�sa5sec-onddelaybeforetheanalyzeractuallyturnsOFF,dur�ngwh�cht�metheun�tcanbekeptONbypress�ngtheeNTeR button.Foranalyzersthatconta�naCOsensor,there�saCOpurgefeaturethatkeepstheanalyzer’spumprunn�ngunt�lthemeasuredCOleveldropsbelow50ppm.
Instruction 24-9388 3
Fyrite Pro
2.0 TeChNICAl ChARACTeRISTICSThe Analyzer Directly Measures and Displays:
•Combust�onA�r(TL)......................................0to100°C(32to212°F)•FlueGasO2content........................................0.0to20.9%O2•FlueGasTemperature(TA)...........................–18to537°C(0to999°F)•FlueGasCOcontent(Model125Eonly).............0to2,000ppmCO•Draft................................................................–9.99to67mb/HPa
(–9.99to27�nWC)
The Analyzer Computes and Displays:(whenthemeasuredoxygenlevel�sbelow17.9%)
•qA(StackLoss)...............................................0to99%•ETa..................................................................0to109.6%•Combust�onEff�c�ency....................................0.1to99.9%•FlueGasCOA�r-Freecontent(Model125Eonly).. 0to9,999ppm•FlueGasCO2content.....................................0.1toafueldependent
max�mumvalue�n%•ExcessA�r(LAMBDA)...................................0to9.99•CO/CO2(Model125Eonly)..................................0to0.9999
Fuel Selection:(F1thruF4denotesthefuelselectedasd�splayedontheLCD)
•NaturalGas(F1)•Propane(F2)•L�ghtO�l(F3)•C�tyGas(F4)
Normal Operating Conditions:Temperature:•Analyzer.........................................................0to40°C(32to104°F)•Probe...............................................................538°Cmax(1,000°F)at
125mm�nsert�on
Hum�d�ty............................................................15to90%RH,Non-Condens�ng
Instruction 24-93884
Fyrite Pro
Power Requirements........................ Fourd�sposable‘AA’Alkal�nebat-ter�es(anopt�onalACadapter�sava�lablethatel�m�natestheneedforbatter�es)
Operating Time (using batteries)... Upto18hourscont�nuous(pumprunn�ngandbackl�ghtoff)
Warm Up Time................................... 60seconds
Memory ............................................... Upto10setsofread�ngscanbesaved�nmemory.Whenmemory�sfull,thenextread�ngsavedw�lloverwr�tetheoldestread�ng.
Display................................................. 4D�g�t,2L�ne,7-segmentL�qu�dCrystalD�splay
Front Panel Controls....................... S�xpushbuttonsw�tches(RefertoSect�on3.4)
Accuracy:
•Oxygen............................................. ±0.3%O2w�thatyp�calfluegasconcentrat�onofCO2
•CarbonMonox�de............................ ±5%ofread�ngor±10ppm,wh�ch-ever�sgreater*
•FlueGasTemperature.................... ±2°Cbetween0&124°C(±4°Fbetween32&255°F)±3°Cbetween125&249°C(±6°Fbetween256&480°F)±4°Cbetween250&400°C(±8°Fbetween481&752°F)
•SecondThermocoupleChannel...... ±2°Cbetween0&100°C(±4°Fbetween32&212°F)
•Amb�entTemperature..................... ±2°Cbetween0&40°C(±4°Fbetween32&104°F)
•Draft................................................ ±1%ofread�ngor±0.05mB(±0.02�nWC),wh�chever�sgreater
* T�ghterCOaccuracy�nthelowerranges,upto±2ppm,maybeatta�ned�falowerrangecal�brat�ongas(e.g.100ppmCO)�sused.
Instruction 24-9388 5
Fyrite Pro
3.0 SeTTING UP The ANAlYZeR
3.1 Preliminary StepsBeforeus�ngtheanalyzer...
•Checkbatter�es(Sect�on3.2)•Connectprobetoanalyzer(Sect�on3.3)•Checksetup(Sect�on3.5)
3.2 Checking & Replacing the BatteriesInstallfreshbatter�esasdescr�bedbelow.Checktheanalyzerforsuff�c�entchargepr�ortoeachuse.Replacethebatter�es�fthelow-batterysymbol
appears�nthelowerr�ghtcornerofthescreen.Toreplacethebat-ter�es:
1. Removebatterycoverfrombackofanalyzer.
2. Ifoldbatter�esare�nstalled,removethemandproperlyd�scard.
3. Observ�ngthepolar�tymark�ngs�ns�dethebatterycompartment,�nstallfour‘AA’Alkal�nebatter�esasshown�nF�gure3-1.
4. Replacebatterycover.
Figure 3-1. Battery Installation
Remove battery cover bypushing down on coverand sliding it outward
Instruction 24-93886
Fyrite Pro
3.3 Connecting the Probe and ThermocoupleAttachtheprobeandhoseassemblytotheanalyzerasfollows(F�gure3-2):
1. Pushtheyellowbandedqu�ck-connectFlueGasHose(g�v�ngasl�ghttw�st)ontotheGAS�nletf�tt�ng.
2. Pushthebluebandedqu�ck-connectDraftHose(g�v�ngasl�ghttw�st)onthe“+”pressuref�tt�ng.
3. PushtheFlueGasThermocouple�ntotheT-STACKjack
Important:DO NOT force the thermocouple connector into its jack. The connection tabs are different sizes, allowing the connector to fit in only one way.
4. Pushtheopt�onalcombust�on-a�rthermocouple�ntotheT-AIRjack.
Note: The optional combustion-air thermocouple provides a convenient way to measure the burner’s combustion-air temperature when its temperature is not the same as room air. If this thermocouple is not used, then a separate mea-surement of the combustion-air temperature must be made using the probe’s thermocouple during the analyzer’s warm-up period.
Inspecttheflue-gashoseforcracks.Replacethehose�f�t�sdefect�ve.
Beforeus�ngtheanalyzer,checkthattheWaterTrap/F�lter�sdryandnotd�rty.Ifnecessary,dryoutthetrapandreplacethef�lterelementperSect�on5.8.
Instruction 24-9388 7
Fyrite Pro
Figure 3-2. Connecting the Probe and Hose Assembly
Instruction 24-93888
Fyrite Pro
3.4 Front Panel Push ButtonsNotethatapushbuttonmayperformseveralfunct�ons,depend�ngontheanalyzer’smodelnumberandwhatscreen�sbe�ngd�splayedatthet�me.
I/O • TogglestheanalyzerONandOFF. •Placestheanalyzer�ntoe�ther�tsSetuporCal�brat�onMode
whenused�nconjunct�onw�ththeeNTeR orhOlD button. •Scrollsupthroughthed�splayscreensdur�ngatest. •Causesthed�splayedvalueto�ncreaseorchangewh�le�nthe
Cal�brat�on,Setup,orPr�nt/MemoryMenuScreen. •Scrollsdownthroughthed�splayscreensdur�ngatest. •Causesthed�splayedvaluetodecreaseorchangewh�le�nthe
Cal�brat�on,Setup,orPr�nt/MemoryMenuScreen.
eNTeR •EnterstheRunMode(startsacombust�ontest)fromtheFuelScreen.
•Unfreezesthed�splayafterpress�ngthehOlD button. •D�splaystheFuelScreenwhenhelddownfor2secondswh�le�n
theRunMode. •Performstheact�onselected�nthePr�nt/MemoryMenuScreen. •Storesthed�splayedvalueandautomat�callystepstothenext
screenwhenpresseddur�ngcal�brat�onorsetup. •Placestheun�t�nto�tsRunModewhenhelddownfor2seconds
wh�le�ntheCal�brat�onMode. •Abortsturn-offandkeepstheanalyzerturnedONwhen
presseddur�ngthe5secondturn-off-delayper�od. •AbortstheCOpurgefunct�onatturn-offwhenthemeasured
COlevel�sabove50ppm. •Setsuptheanalyzertobeplaced�nto�tsCal�brat�onMode
whenhelddownw�ththeanalyzerOFF.(Used�nconjunct�onw�ththeI/Obutton.)
hOlD •Placestheanalyzeronholdandfreezesthevalues�nallRunModeScreensdur�ngatest,allow�ngtheoperatortoscrollthroughthed�splaysandv�ewalltestvaluesatthatpo�nt�nt�me.Press�ngeNTeR resumestest�ng.
•D�splaysthePr�nt/MemoryMenuScreenwhenpressedtwice fromtheRunMode.
• Setsuptheanalyzertobeplaced�nto�tsSetupModewhenhelddownw�ththeanalyzerOFF.(Used�nconjunct�onw�ththeI/Obutton.)
lIGhT Togglesthebackl�ghtONandOFF.
Instruction 24-9388 9
Fyrite Pro
3.5 Setup ModeTheanalyzer�spresetatthefactoryfortheparametersshownbelow,butcanbechangedasdescr�bed�nthe�rassoc�atedsect�ons.
Function Parameter To ChangeFuel NaturalGas(F1) Sect�on4.13.2TemperatureUn�t °C Sect�on3.5.2COChannel AutoZero(A2) Sect�on3.5.3Draft(Pressure) mB Sect�on3.5.4Clock NotIn�t�al�zed Sect�on3.5.5thru3.5.7Pr�nterProtocol IrDA Sect�on3.5.8
3.5.1 EnteringSetup1.W�ththeanalyzerturnedOFF,pressandholddownthehOlD
button.
2. Pressthe I/Obutton.
3.Releasebothbuttons.
4. Theanalyzer�snow�n�tsSetupMode.RefertoSect�ons3.5.2thru3.5.8for�nformat�ononhowtosetupeachparameter.
3.5.2 SelectingTemperatureUnitsTheTemperatureSetupScreen�slabeled“Un�t.”
1. EntertheSetupModeperSect�on3.5.1.Ifnecessary,repeatedlypresstheeNTeR buttonunt�l“Un�t”�sd�splayed.
2. Presse�thertheor buttonunt�lthedes�redtemperatureun�t(°For°C)�sd�splayed.
3. PresseNTeR tomovetothenextSetupScreen,orI/Otoex�tsetup.
Instruction 24-938810
Fyrite Pro
3.5.3 COChannelSetup(Model125Eonly)
TheCOChannelSetupScreen�slabeled“CO.”
1. EntertheSetupModeperSect�on3.5.1.Ifnecessary,repeatedlypresstheeNTeR buttonunt�l“CO”�sd�splayed.
2. Presse�therthe orbuttonunt�lthedes�redparameter�sd�splayed.-no Sensord�sabled.Useth�sfunct�ontoturnofftheCOchan-
nel�fthesensor�sbadornot�nstalledtoavo�dtheCOsensorerrormessagefromappear�ng.
-2Ero ManualZero.WhentheCOchannel�ssettomanualzero,theanalyzerdoesnotzerotheCOsensortoamb�entcond�-t�onsdur�ngstartup.
-A2 AutoZero.W�ththeCOchannelsettoautozero,theCOsensor�szeroedtotheamb�entCOleveldur�ngstartup.Important: When using this mode the analyzer must be turned ON in fresh air; otherwise, incorrect CO readings will occur.
3. PresseNTeR tomovetothenextSetupScreen,orI/Otoex�tsetup.
3.5.4 DraftChannelSetupTheDraftSetupScreen�slabeled“PrESUn�t”.
1. EntertheSetupModeperSect�on3.5.1.Ifnecessary,repeatedlypresstheeNTeR buttonunt�l“PrES”�sd�splayed.
2. Presse�therthe orbuttonunt�lthedes�redparameter�sd�splayed.-no Pressuresensord�sabledornot�nstalled.-WC �nchesofWaterColumn-mB M�ll�bars-HPa HectoPascals
3. PresseNTeR tomovetothenextSetupScreen,orI/Otoex�tsetup.
Instruction 24-9388 11
Fyrite Pro
3.5.5 TimeSetupTherearetwoT�meSetupScreens,oneforhoursandtheotherform�n-utes.Twobarsappearabovethesegmentsbe�ngchanged.T�me�sd�s-played�na24hourformat.
1. EntertheSetupModeperSect�on3.5.1.Ifnecessary,repeatedlypresseNTeRunt�lthef�rstT�meSetupScreen�sd�splayed—theonew�thtwobarsoverthehourd�g�ts.
2. Presse�thertheor buttonunt�lthecorrecthourvalue�sd�splayed.
3. PresseNTeR tomovetheselect�onbarsoverthem�nuted�g�ts.
4. Presse�thertheorbuttonunt�lthecorrectm�nutevalue�sd�splayed.
5. PresseNTeR tomovetothenextSetupScreen,orI/Otoex�tsetup.
3.5.6 DateSetupTherearetwoDateSetupScreens,eachlabeled“DAtE”.Thef�rstscreensetsthemonth(secondtwod�g�ts)wh�lethesecondscreensetstheday (f�rsttwod�g�ts).
1. EntertheSetupModeperSect�on3.5.1.Ifnecessary,repeatedlypresseNTeRunt�lthef�rst“DAtE”Screen�sd�splayed.
2. Presse�thertheorbuttonunt�lthecorrectmonth�sd�splayed.
3. PresseNTeR tochangethe‘Day’value.
4. Presse�thertheorbuttonunt�lthecorrectday�sd�splayed.
5. PresseNTeR tomovetothenextSetupScreen,orI/Otoex�tsetup.
Instruction 24-938812
Fyrite Pro
3.5.7 YearSetupTheYearSetupScreen�slabeled“yEAr”.
1. EntertheSetupModeperSect�on3.5.1.Ifnecessary,repeatedlypresseNTeRunt�l“yEAr”�sd�splayed.
2. Presse�thertheorbuttonunt�lthecorrectyear�sd�splayed.
3. PresseNTeR tomovetothenextSetupScreen,orI/Otoex�tsetup.
3.5.8 SelectingPrinterProtocolTheanalyzercanbesetupfore�theranHPorIrDAtypepr�nter.
1. EntertheSetupModeperSect�on3.5.1.Ifnecessary,repeatedlypresseNTeRunt�loneofthefollow�ngscreens�sd�splayed.
2. Presse�thertheorbuttontoselectthedes�redpr�nterprotocol.
3. PresseNTeR tomovetothenextSetupScreen,orI/Otoex�tsetup.
3.5.9 ExitingtheSetupScreenPresstheI/Obuttonatanyt�metoex�ttheSetupModeandturnOFFtheanalyzer.Notethatthelastd�splayedparameter�sautomat�callysaved�nmemory.
Instruction 24-9388 13
Fyrite Pro
4.0 OPeRATION
4.1 Performing CO and Draft Testing4.1.1 AnalyzerTurnOnandWarmUp
1. Connecttheprobeandmakesurethattheanalyzer�sproperlysetupperSect�on3.0.
2. TurnONtheanalyzerbypress�ng�sI/Obuttonandobservethatthefollow�ngWarm-UpScreensared�splayed.
Thesescreensshowtheanalyzer’smodelnumber(ProE122orProE125),theun�t’ssoftwarerev�s�on,andf�nallytherema�n�ngwarm-upt�mecounteddownfrom60seconds,�nclud�ngameasure-mentoftheO2sensor’soutputlevel.
Tip: An O2 error will occur when the oxygen sensor’s output drops to between 80 and 90. Consider replacing the oxygen sensor when its output level drops below 100.
3. Wa�tfortheanalyzertocountdown�tswarm-upper�od;afterwh�ch,�fnoerrorsweredetectedtheFuelScreen�sd�splayed.
If,however,anyerrorsweredetecteddur�ngwarm-up,theSensorStatusScreen�sd�splayed.Ifth�soccurs,refertoSect�on4.13.11for�nformat�ononhowtocorrecttheerror.
Instruction 24-938814
Fyrite Pro
4.1.2 MeasuringCO(Model125Eonly)
1. TurnONtheanalyzerandallow�ttowarm-upperSect�on4.1.1.
Important: If the CO channel is set up for auto zero, then the analyzer must be turned ON in fresh air; otherwise, incorrect CO readings will occur. Refer to Section 3.5.3. When the CO channel is set up for manual zero, and if the CO reading is higher than zero when sampling fresh air, then manually zero the CO channel per Section 5.6, Steps 1 thru 4.
2. PushtheeNTeR buttonthenpresse�therthe orbuttonasnec-essarytod�splaytheCO&COA�rFreeScreen.
3. Inserttheanalyzer’sprobe�ntotheareatobetestedandobservethedetectedCOlevel�nppmontheLCD.
4.1.3 MeasuringDraft1. TurnONtheanalyzerandallow�ttowarm-upperSect�on4.1.1.
2. PresseNTeR andthenpresse�therthe orbuttonasnecessarytod�splaytheTA&DraftScreen.
3. Thedraftread�ngshouldbezerowhensampl�ngrooma�r.Ifnot,pressthebuttononce tod�splaytheDraftZeroScreen,andthenpresstheeNTeR buttontozerothedraftchanneltoamb�entatmo-spher�ccond�t�ons.
4. Inserttheanalyzer’sprobe�ntotheareatobetestedandobservethedraftread�ngontheLCD.
Instruction 24-9388 15
Fyrite Pro
4.2 Performing a Combustion Test
4.2.1 AnalyzerTurnOnandWarmUpImportant: The probe must be at room temperature before performing the following steps.
1. Connecttheprobe,and�fused,thecombust�on-a�rthermocoupletothebottomoftheanalyzerperSect�on3.3.
2. Beforeturn�ngONtheanalyzer,pos�t�ontheprobeortheopt�onalcombust�on-a�rthermocoupletomeasurethetemperatureoftheburner’scombust�ona�rbyperform�ngoneofthefollow�ng:•When NOT USING the Combustion-Air Thermocouple–
Beforeturn�ngONtheanalyzer,�tsprobemustbelocated�ntheareaconta�n�ngtheburner’scombust�on-a�rsupply.Iftheburner�sus�ngrooma�r,thens�mplyplacetheprobew�th�ntheroom.Inthecaseofah�gh-eff�c�encyfurnacewherecombust�ona�r�sdrawn�nfromanouts�desource,�nserttheprobe�ntothecom-bust�on-a�rstreamso�tcanmeasure�tstemperaturew�ththeburneroperat�ng.
•When USING the Combustion-Air Thermocouple –Theanalyzer’ssecondthermocouplechannelcanbeusedto�nde-pendentlymon�torthetemperatureoftheburner’scombust�on-a�rsupplywh�letest�ng.Whenus�ngth�ssecondthermocouplechannel,�nsertthethermocouple�ntotheburner’scombust�on-a�rstream,andpos�t�ontheprobetomeasuretheamb�entroomtemperature.
3. TurnONtheanalyzerandallow�ttowarmupperSect�on4.1.1.
4.2.2 SelectingaFuel1. W�thFuelScreend�splayed,presse�thertheorbuttontostep
throughthefuelscodesunt�ltheproperfuel�sselected.Theana-lyzerdefaultstothemostrecentlyselectedfuel.
F1=NaturalGas F2=Propane F3=L�ghtO�l F4=C�tyGas
2. PresstheeNTeRbuttontoselectthed�splayedfuelcodeandentertheRunMode.
Instruction 24-938816
Fyrite Pro
Atmospheric Burner or Gravity Vented Forced Air hot Water Tank
4.2.3 SamplingPointForced Air Furnace–Whentest�ngatmospher�cburnerorgrav�tyvented,forceda�rheat�ngequ�pmentw�thaclamshellorsect�onalheatexchangerdes�gn,testeachoftheexhaustportsatthetopoftheheatex-changer.Theprobeshouldbe�nsertedback�ntoeachoftheexhaustportstoobta�nafluegassample,beforeanyd�lut�ona�r�sm�xed�n.
Hot Water Tank –Domest�chotwatertanksw�ththe‘bell’shapeddraftd�verterontopcanbeaccuratelytestedby�nsert�ngtheprobet�pd�rectly�ntothetopofthef�retubebelowthed�verter.
80% Efficiency Fan Assist or Power Vented–Combust�ontest�ngoffanass�storpowervented,furnaces/bo�lersshouldbedonethroughaholedr�lled�nthevent�mmed�atelyabovethe�nducerfan.
90% Efficiency Condensing–Condens�ngfurnaces/bo�lerscanbetestedthroughaholedr�lled�ntheplast�cventp�pe(whenallowedbythemanufac-turerorlocalauthor�tyofjur�sd�ct�on)ortakenfromtheexhaustterm�nat�on.
Atmospheric or Gravity Vented Boiler–Bo�lers,wh�chhavea‘bell’shapeddraftd�verterd�rectlyontop,shouldbetestedd�rectlybelowthed�verterthroughaholedr�lled�ntheventconnector.
Instruction 24-9388 17
Fyrite Pro
90% eff. Condensing Furnace/Boiler
Atmospheric or Gravity Vented Boiler
80% eff. Fan Assist or Power Vented Furnace/Boiler
Instruction 24-938818
Fyrite Pro
4.2.4 PerformingaCombustionEfficiencyTest1. Afterturn�ngONtheanalyzerandselect�ngtheappropr�atefuel,
theTA&DraftScreen(refertoSect�on4.13.3)shouldbed�splayed.
2. Pos�t�ontheprobe�ntheflue-gasstreamtoobta�nthehottestTA(stacktemperature)read�ng.Locating the highest stack temperature is very important for accurate calculations of qA, Eta, and efficiency.
3. Burner-serv�ceprocedurescannowbeg�n.Usethe andbuttonstoscrollthroughtheanalyzer’sotherd�splayscreens(refertoSec-t�on4.12).Theanalyzerread�ngsw�llupdatecont�nuouslyshow�ngchanges�nburnerperformance.
Note: When a calculation cannot be made because of im-proper data (i.e., oxygen level above 17.9%), four dashes “- - - -” will appear in place of the calculated value on both the screen and printout.
Tip: Pressing the hOlD button will freeze all readings and stop the pump, allowing the operator to scroll through the Run Mode Screens and examine the readings at any single point in time. Press eNTeR to restart the pump and resume testing.
4.2.5 EndingaCombustionTestWARNING! Burn Hazard. Do not touch the probe after removing it from the stack. Allow the probe to cool before handling (about 5 minutes).
1. Removeprobefromtheflue-gasstream.
2. Allowthepumptorununt�lallcombust�ongasesareflushedfromtheanalyzeras�nd�catedbytheO2read�ngreturn�ngto20.9%.
3. TurnOFFtheanalyzerperSect�on4.10.
Instruction 24-9388 19
Fyrite Pro
4.3 Operating Tips•Whenananalyzer�sbrought�nfromacoldveh�cle,let�twarmup
slowlytom�n�m�zecondensat�on.Temperaturesbelowfreez�ngw�llnotdamagetheanalyzer;however,br�ng�ngacoldanalyzer�ntoawarm,hum�denv�ronmentmaycausecondensatetoform�ns�dethecase.
•IftheCOchannel(Model125Eonly)�ssetupforAutoZero(refertoSect�on3.5.3),ensurethattheanalyzer�ssampl�ngfresha�rwhenturnedON.Pull�ngaflue-gassamplethroughtheanalyzerdur�ng�tswarm-upper�odw�llnotdamagetheanalyzer,but�tw�llresult�n�n-correctCOread�ngs.AlsonotethataCOsensorerrorw�lloccur�fthedetectedCOlevel�sabove50ppmdur�ngwarm-up.
•Whenmeasur�ngflue-gas,notethatflue-gascondensate�sac�d�candverycorros�ve.It�s�mportantnottoallowtheanalyzer’s�nternalcom-ponentstobecomesoaked�ncondensateforlongper�odsoft�me.
•Beforeeachuse,�nspectthef�lterelementofthewater-trap/f�lteras-sembly.Replacethef�lter�f�tlooksd�rty.RefertoSect�on5.8.
•Whensampl�ngflue-gas,keeptheanalyzerabovethewater-trap,andkeepthetrap�navert�calpos�t�on.Th�sw�llmax�m�zetheeffect�ve-nessofthetrapandkeepl�qu�dcondensatefrombe�ngdrawnd�rectly�ntotheanalyzer.
•Whenl�qu�dcondensate�sseen�ns�dethewatertrap,emptythetrapbefore�tbecomesfull.RefertoSect�on5.8.
•It�srecommendedthattheanalyzerbepurgedaftertak�ngaflue-gasmeasurementbeforeturn�ng�tOFF.Oncetheprobe�sremovedfromthestack,d�sconnectthehoseassemblyfromthebottomoftheanalyz-erandletthepumprunfor10m�nutesorsotocompletelyremoveanyrema�n�ngfluegasesanddryanycondensatefrom�ns�dethesensorchamber.
•Whenstor�ngtheanalyzer,�t’sagood�deatoemptythewatertrapandleave�topentofurtherdry�tout.
•Cal�bratetheanalyzerevery6monthstoensure�tsaccuracy.
Instruction 24-938820
Fyrite Pro
4.4 Using the BacklightTheLCDcanberead�nd�mly-l�tareasbypress�ngthelIGhT button.
Thebackl�ghtautomat�callyturnsOFFafter10m�nutesofkeyboard�nact�v-�ty,butcanbeturnedOFFatanyt�mebyaga�npress�ngthelIGhT button.
4.5 Using the ProbeAr�g�dsta�nlesssteelprobew�thhandle,connectedtoaflex�blehosew�th�ntegralwater-trap/f�ltercanbeusedtodrawagassample�ntotheana-lyzerfromtheroom,gr�lles,d�ffusers,andfurnaceflues.
Thehoseandprobeassemblycanbedetachedfromtheanalyzerwhentheoperatordes�restosamplew�thouttheprobe(�.e.,whensampl�ngforCO).
4.6 Saving Test Data in MemoryUpto10�nd�v�dualsetsoftestdatacanbesaved�nmemoryasfollows:
Note: When memory is full, the next reading saved will over-write the oldest reading.
Note: The analyzer's setup information as entered in Sec-tion 3.5 is stored along with the test data. For example, temperatures stored in °C are recalled and printed in °C even if the analyzer is currently set up for °F.
1. Iftheanalyzer�s�n�tsRunMode,pressthehOlD buttontwice toenterthePr�nt/MemoryMenuScreens.Iftheanalyzer�salready�n�tsHoldMode,pressthehOlD buttononlyonce.Thef�rstmenu�temd�splayed�sthePr�ntScreen.
Instruction 24-9388 21
Fyrite Pro
2. Pressthebuttononce tod�splaytheSaveScreen.Thenumbershown�nth�sscreenrepresentsthememorylocat�on(1thru10)towh�chthecurrenttestdataw�llbesaved.
3. PresseNTeR tosavethetestdataandreturntotheHoldMode,orpresshOlD toreturntotheHoldModew�thoutsav�ng.
4.7 Opening and Viewing Saved Test DataPerformthefollow�ngtoopenandv�ewsavedtestdata:
Note: If no test data has been saved, the option to open the memory for viewing will not be available.
1. Iftheanalyzer�s�n�tsRunMode,pressthehOlD buttontwice toenterthePr�nt/MemoryMenuScreens.Iftheanalyzer�salready�n�tsHoldMode,pressthehOlD buttononlyonce.Thef�rstmenu�temd�splayed�sthePr�ntScreen.
2. Presse�therthe orbuttonunt�ltheOpenScreen�sd�splayed,andthenpresseNTeRtoopenthememorylocat�onsforv�ew�ng.Thenumbershown�nthesecondscreenrepresentsthemostrecentmemorylocat�onwheredatawasstored.
3. Presse�therthe orbuttontoscrolltothedes�redmemoryloca-t�on,andthenpresseNTeR torecallthestoreddataandreturntotheHoldMode.Wh�le�ntheHoldMode,therecalleddatacanbev�ewedus�ngthe andbuttons,orpr�ntedperSect�on4.9.
Instruction 24-938822
Fyrite Pro
4.8 Printing Test DataTurnONthepr�nter.Refer to the printer’s instruction manual for detailed operation and maintenance information.
Ifnotalreadydone,setthepr�nterparametersasfollows:•Data:8b�ts•Baud:9600•Par�ty:None•Handshak�ng:X-on/X-off
Al�gnthepr�nterw�ththetopoftheanalyzerasshown�nF�gure4-1.
Thenextstep�nthepr�nt�ngproceduredependsonwhethertheoperatordes�restopr�ntonlythecurrenttestdata,orallstoredtestdatastart�ngw�ththemostrecent.Performtheappropr�ateprocedurebelow:
Print Current Test Data1. W�ththeanalyzer�n�tsRunMode,pressthehOlD buttontwice to
d�splaythePr�ntScreen.
2. PresseNTeR tostartpr�nt�ng.
Print All Test Data1. W�ththeanalyzer�n�tsRunMode,pressthehOlD buttontwice
tod�splaythePr�ntScreen.Thenpressthe buttontod�splaythePr�ntAllScreen.
2. PresseNTeRtostartpr�nt�ng.
Instruction 24-9388 23
Fyrite Pro
Figure 4-1. Printer Alignment & Sample Printout
Notes: When a calculation cannot be made because of improper data (i.e., oxygen level above 17.9%), four dashes “- - - -” will appear in place of the calculated value on both the screen and printout.
When one or more of the following sensors are not installed, their associated values do not print:
CO Sensor: CO, CO Undilute and CO/CO2
Draft Sensor: Draft
� � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � �� � � � � � � � � � � � �
� � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � �� � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � �
Instruction 24-938824
Fyrite Pro
4.9 Clearing Saved Test DataClearallsavedtestdataasfollows:
Note: If no test data has been saved, the option to clear memory will not be available.
1. W�ththeanalyzer�n�tsRunMode,pressthehOlD buttontwice tod�splaythePr�ntScreen.
2. Presse�therthe orbuttontoscrolltotheClearScreen,andthenpresseNTeRtod�splaytheClearAllScreen.
3. PresseNTeR aga�ntoclearmemoryandreturntotheHoldMode,orpresshOlD toreturntotheHoldModew�thoutclear�ngmemory.
4.10 Turning OFF the Analyzer & CO PurgePresstheI/ObuttontoturnOFFtheanalyzer.Theun�tw�llcountdownfrom5beforeturn�ngOFF,thusallow�ngt�mefortheoperatortoaborttheturnOFFprocessbypress�ngtheeNTeRbutton.
Ifah�ghCOlevel�sdetectedatturnOFF,theun�tw�llrema�nONw�th�tspumprunn�ngandd�splay“PUrGCO”.Thecountdownfrom5w�llnotbeg�nunt�lthedetectedCOleveldropsbelow50ppm.Althoughnotrecom-mended,thepurg�ngprocesscanbebypassedbypress�ngtheI/Obuttonasecondt�me.
Instruction 24-9388 25
Fyrite Pro
4.11 Resetting the MicroprocessorIftheanalyzer‘locksup’andcannotbeturnedOFF,resetthem�croproces-sorbyremov�ngoneofthebatter�esforatleast5seconds.
Instruction 24-938826
Fyrite Pro
Figure 4-2. Model 122E Run Mode Screens
4.12 Run Mode ScreensF�gures4-2and4-3showtheorder�nwh�chthescreensared�splayedbypress�ngtheeNTeR andbuttonsaftertheanalyzerwarms-upandenters�tsRunMode.RefertoSect�on4.13foradeta�leddescr�pt�onofeachscreen.
� � � � �
�
Fuel Select:F1 = Natural GasF2 = PropaneF3 = Light OilF4 = City Gas
Stack Temperature & Draft
Draft Zero
Stack Temperature & efficiency
qA & excess Air (lAMBDA)
Tl & eta
CO2 &O2
Instruction 24-9388 27
Fyrite Pro
Fuel Select:F1 = Natural GasF2 = PropaneF3 = Light OilF4 = City Gas
Stack Temperature & Draft
Draft Zero
Stack Temperature & efficiency
Ratio of CO to CO2
qA & excess Air (lAMBDA)
Tl & eta
CO & CO Air Free (Undilute)
CO2 &O2
� � � � �
�
Figure 4-3. Model 125E Run Mode Screens
Instruction 24-938828
Fyrite Pro
4.13 Screen Descriptions4.13.1 Warm-UpScreensIncludingO2SensorOutput
Assoonastheanalyzer�sturnedON,aser�esofWarm-UpScreensared�splayed.Thesescreensshowtheanalyzer’smodelnumber(ProE122orProE125),theun�t’ssoftwarerev�s�on,andf�nallytherema�n�ngwarm-upt�mecounteddownfrom60seconds,�nclud�ngameasurementoftheO2sensor’soutputlevel.Attheendofwarm-up,theFuelScreen�sd�splayed.
Note: If any errors were detected during warm-up, the Sensor Error Screen (Section 4.13.11) is displayed.
Tip: An O2 error will occur when the oxygen sensor’s output drops to between 80 and 90. Consider replacing the oxygen sensor when its output level drops below 100.
4.13.2 FuelScreen
TheFuelScreen�sd�splayed�mmed�atelyafterwarm-up,and�swhereanoperatorchooseswh�chfuel�stobeusedbytheanalyzertocalculatecombust�oneff�c�ency.Press�nge�therthe orbuttonscrollsthed�s-playthroughthefuelcodes.Press�ngeNTeRselectsthed�splayedfuelandplacestheanalyzer�nto�tsRunMode.Notethatthefuelcodew�llbed�splayed�nthetopr�ghtcornerofallscreenswh�le�ntheRunMode.
Fuel Codes: F1=NaturalGas F3=L�ghtO�l F2=Propane F4=C�tyGas
Front Panel Button Functions: – D�splaysnextfuelcode – D�splaysprev�ousfuelcode eNTeR – GotoRunMode(refertoSect�on4.12) hOlD – Noeffect lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
Instruction 24-9388 29
Fyrite Pro
4.13.3 TA(StackTemperature)&DraftScreen
TheTAread�ng�sthestacktemperaturemeasuredattheprobet�p.
Thedraftread�ng�sthepressuremeasuredattheprobet�pasreferencedtotheanalyzer’s“–”pressureport.
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Holdfor2secondstod�splayFuelScreen lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
4.13.4 DraftZeroScreen
TheDraftZeroScreenallowstheusertomanuallyzerothedraftchannelwhenthepressureread�ng�sanyth�ngotherthanzerowhenmeasur�ngamb�entroompressure.Tozerothedraftchannel,removealltub�ngfromtheanalyzer’s“+”and“–”portsandpresstheeNTeR button.
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Zerosthedraftchannel,andreturnstothescreenconta�n�ng
thedraftread�ng. lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
Instruction 24-938830
Fyrite Pro
4.13.5 Stack&EFFScreen
TheSTACKread�ng�sthetemperaturemeasuredattheprobet�p.
TheEFF(Eff�c�ency)read�ng�sacalculat�onofwhatpercentageofenergypresent�nthefuel�sbe�ngconverted�ntousableheatbasedonthefuel’sh�ghheat�ngvalue.
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Holdfor2secondstod�splayFuelScreen lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
4.13.6 RatioofCOtoCO2Screen(Model125Eonly)
TheCOC2Screend�splaysaread�ngthat�sthecalculat�onofppmCOd�v�dedby%CO2asdetected�ntheflue-gasstream.
Note: The display does not show a decimal point in front of the reading. The above reading is actually 0.0006.
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Holdfor2secondstod�splayFuelScreen lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
Instruction 24-9388 31
Fyrite Pro
4.13.7 qA&ExcessAirScreen
�
TheqAread�ng�sacalculat�onofstackloss�npercent.
TheEXCESSAIR(LAMBDA)read�ng�sthecalculat�onofthepercentageofextraa�rthat�sava�lable�nthecombust�onchamberabovethetheoret�-calamountneededforperfectcombust�on.
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Holdfor2secondstod�splayFuelScreen lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
4.13.8 TL&EtaScreen
TheTLread�ng�sthecombust�on-a�rtemperaturethat�susedforef-f�c�encyandqAcalculat�ons.Th�stemperature�smeasuredbye�thertheopt�onalT-AIRthermocoupleonacont�nuousbas�s,orbytheT-STACKthermocoupledur�ngstart-up(refertoSect�on4.2.1).Ifneither thermo-couple�s�nstalledwhentheanalyzer�sturnedON,thenth�stemperatureread�ngcomesfromatemperaturesensorlocated�ns�detheanalyzer.
TheEtaread�ng�sthecalculatedpercenteff�c�encyasreferencedtoalowheat�ngvalue.
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Holdfor2secondstod�splayFuelScreen lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
Instruction 24-938832
Fyrite Pro
4.13.9 CO&COAirFreeScreen(Model125Eonly)
TheCOread�ng�sthemeasurementofhowmuchcarbonmonox�de�nppm�spresent�ntheflue-gasstream.
TheCOAIRFREE(und�lute)read�ng�sacalculat�onoftheexactppmconcentrat�onofcarbonmonox�de�foxygenwere0.0%�naccordancew�thANSIstandardZ21.1.
COx20.920.9–O2measured
COA�rFree=
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Holdfor2secondstod�splayFuelScreen lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
4.13.10 CO2&O2Screen
TheCO2read�ng�sacalculat�onofthepercentageofcarbond�ox�dethat�sbe�ngproducedbythecombust�onprocess.
TheO2read�ng�sthemeasuredpercentageofoxygenpresent�ntheflue-gasstream.
Front Panel Button Functions: – D�splaysnextscreen(refertoSect�on4.12) – D�splaysprev�ousscreen(refertoSect�on4.12) hOlD – Freezesd�splay(presseNTeR tounfreeze) eNTeR – Holdfor2secondstod�splayFuelScreen lIGhT – Togglesbackl�ghtON/OFF I/O – TurnsanalyzerOFF(5seconddelay)
Instruction 24-9388 33
Fyrite Pro
4.13.11 SensorErrorScreen
AnO2sensorerror�sd�splayed�ftheanalyzerdeterm�nesdur�ngthewarm-upcyclethattheoxygensensor’soutput�stoolowfor�ttobeusable.However,�ntheextremecond�t�onwhentheO2sensorhasno output,asen-sorerrorw�llnotoccur.Instead,theO2read�ngw�llbe0.0asd�splayed�ntheCO2&O2Screen.
Tip: To avoid O2 errors from occurring, consider replac-ing the O2 sensor when its output level drop below 100 as displayed during warm-up (refer to Section 4.13.1).
WhentheCOchannel�ssetupforAutoZero(refertoSect�on3.5.3),aCOsensorerrorw�lloccur�fthedetectedcarbonmonox�delevel�sabove50ppmdur�ngthewarm-upcycle.Notethat�ftheCOchannel�ssetupformanualcal�brat�on,theanalyzerdoesnotauto-zerotheCOsensordur�ngwarm-up,andthusdoesnotgenerateaCOsensorerrorwhentheanalyzer�sturnedON�nanatmosphereconta�n�ngah�ghbackgroundlevelofCO.
Dothefollow�ngbeforereplac�ngasuspectedspentordefect�vesensor:1. TurnOFFtheanalyzerandturn�tbackON�nanareaoffresha�r
(conta�n�ng20.9%O2andnoCO).
2. PerformtheOxygenSensorZeroprocedure(Sect�on5.5)and,�fap-pl�cable,theCOSensorZero&Cal�brat�onprocedure(Sect�on5.6).
3. IfSteps1&2donotel�m�natetheerrorcond�t�on,replacesensor(s)(refertoSect�ons5.9or5.10).
Instruction 24-938834
Fyrite Pro
Notes:
Instruction 24-9388 35
Fyrite Pro
5.0 CAlIBRATION & MAINTeNANCeImportant:Fresh batteries should be installed, or use the optional AC adapter to power the analyzer during calibra-tion. The unit should be allowed to stabilize at room temper-ature for at least 2 hours before proceeding with calibration. To maintain accuracy as listed in the Technical Character-istics Section of this manual, the standards used must be at least four times as accurate as the stated accuracy of the Fyrite Pro.
5.1 entering the Calibration Mode and Testing the Display Segments
1. W�ththeanalyzerturnedOFF,placetheun�t�nfresh,amb�enta�r;thenpressandholddowntheeNTeRbutton.
2. PresstheI/O buttonandrelease�t.ObservethatallLCDsegmentsareturnedON.
3. ReleasetheeNTeR button.Observetheun�t’smodelnumberandsoftwarevers�onared�splayed.Theword“CAL”�sthend�splayedwh�letheun�twarmsupandcountsdownfrom60seconds.
Attheendof60seconds,thef�rstcal�brat�onscreen�sd�splayed.
Note: The order in which the calibration screens are displayed depends on whether the CO channel is set up for automatic or manual zero (Section 3.5.3). Calibration Screen Display Order CO Auto Zero:Amb�ent,TL,TA,O2,CO,Pressure CO Manual Zero:CO,Pressure,Amb�ent,TL,TA,O2
Instruction 24-938836
Fyrite Pro
5.2 Ambient Temperature Calibration Material Required: Cal�bratedThermometerProcedure:
1. EntertheCal�brat�onModeasdescr�bed�nSect�on5.1.Thenre-peatedlypresstheeNTeRbuttonunt�l“AMBIENT”appearsatthetopofthed�splay.
2. Usetheandbuttonstosetthed�splayedvaluetomatchtheread�ngofacal�bratedthermometeratroomtemperature.
3. PresseNTeRtostorethed�splayedvalueandmovetothenextcal�brat�onscreen,orholddowneNTeRfor2secondstostorethed�splayedvalueandentertheRunMode,orpresstheI/O buttontoex�ttheCal�brat�onModeandturnOFFtheanalyzerw�thoutsav-�ngthechanges.
5.3 Combustion Air Thermocouple Calibration
Material Required:ThermocoupleS�mulatorProcedure:
1. Attachthethermocouples�mulatortotheanalyzer’sCombust�onA�rThermocouple(T-AIR,seeF�gure3-2).
2. EntertheCal�brat�onModeasdescr�bed�nSect�on5.1.Thenre-peatedlypresstheeNTeRbuttonunt�l“TL”appearsatthetopofthed�splay.
3. Setthes�mulatorto0.0°C(32.0°F);thenusetheandbuttonstosetthed�splayedvaluetomatchthes�mulator’svalue.
4. Setthes�mulatorto100.0°C(212.0°F);thenusetheandbut-tonstoadjustthed�splayedvaluetomatchs�mulator’svalue.
Instruction 24-9388 37
Fyrite Pro
5. RepeatSteps3and4asnecessaryunt�ltheun�t�sread�ngcorrectlyatbothtemperatures.
6. PresseNTeRtostorethesecal�brat�onvaluesandmovetothenextcal�brat�onscreen,orholddowneNTeR for2secondstostorethesecal�brat�onvaluesandentertheRunMode,orpresstheI/Obuttontoex�ttheCal�brat�onModeandturnOFFtheanalyzerw�thoutsav�ngthechanges.
7. Removethermocouples�mulatorfromanalyzer.
5.4 Stack Thermocouple CalibrationMaterial Required:ThermocoupleS�mulator
Procedure:
1. Attachthethermocouples�mulatortotheanalyzer’sStackThermo-coupleconnector(T-STACK,seeF�gure3-2).
2. EntertheCal�brat�onModeasdescr�bed�nSect�on5.1.Thenre-peatedlypresstheeNTeRbuttonunt�l“TA”appearsatthetopofthed�splay.
3. Setthes�mulatorto0°C(32°F);thenusetheandbuttonstosetthed�splayedvaluetomatchthes�mulator’svalue.
4. Setthes�mulatorto300°C(572°F);thenusetheandbuttonstoadjustthed�splayedvaluetomatchs�mulator’svalue.
5. RepeatSteps3and4asnecessaryunt�ltheun�t�sread�ngcorrectlyatbothtemperatures.
6. PresseNTeRtostorethesecal�brat�onvaluesandmovetothenextcal�brat�onscreen,orholddowneNTeR for2secondstostorethesecal�brat�onvaluesandentertheRunMode,orpresstheI/Obuttontoex�ttheCal�brat�onModeandturnOFFtheanalyzerw�thoutsav�ngthechanges.
7. Removethermocouples�mulatorfromanalyzer.
Instruction 24-938838
Fyrite Pro
5.5 Oxygen Sensor ZeroMaterial Required: •Cyl�nderof100%N�trogen,P/N9550-0049 •Cal�brat�onK�t,P/N24-7059Procedure:
1. W�ththeanalyzersampl�ngfresha�r,entertheCal�brat�onModeasdescr�bed�nSect�on5.1.ThenrepeatedlypresstheeNTeRbuttonunt�l“O2”appears�nthelower-lefts�deofthed�splay.
2. Allowpumptorunandsamplefresha�rforatleast1m�nute.
3. Usethe andbuttonstosetthed�splayto20.9%.
4. SetuptheCal�brat�onK�tw�th100%N2asdescr�bed�nthe�nstruc-t�onssuppl�edw�ththek�t.
5. Connectthetub�ngoftheCal�brat�onK�ttotheGAS�nletoftheanalyzer;thenadjusttheregulatorforapprox�mately2SCFHofexcessflow(seeF�gure5-1).
6. Aftertheanalyzerhasstab�l�zed(2to3m�nutes,)usetheandbuttonstosetthed�splayedvalueto0.0%.
7. D�sconnecttub�ngfromanalyzerandturnoffgasflow.
8. Allowtheanalyzer’spumptorununt�ltheO2read�ngreturnsto20.9.Ifnecessary,usetheandbuttonstoreadjusttheread�ngto20.9.RepeatSteps3thru7tover�fythezeroadjustment.
9. PresseNTeRtostorethenewcal�brat�onvaluesandmovetothenextcal�brat�onscreen,orholddowneNTeRfor2secondstostorethenewcal�brat�onvaluesandentertheRunMode,orpresstheI/Obuttontoex�ttheCal�brat�onModeandturnOFFtheanalyzerw�thoutsav�ngthechanges.
Instruction 24-9388 39
Fyrite Pro
5.6 CO Sensor Zero & Calibration (Model 125e only)
Material Required: •Cyl�nderof100ppm(P/N51-1994)or 500ppm(P/N24-0492)COcal�brat�ongas •Cal�brat�onK�t,P/N24-7059
To�mprovetheaccuracyoftheCOread�ng,wesuggestthat�ftheanalyzerw�llbepr�mar�lyusedforfluegastest�ng,thencal�brateus�ng500ppmCO.Iftheanalyzer,however,w�llbepr�mar�lyusedforamb�enttest�ng,thencal�brateus�ng100ppmCO.
Procedure:
1. W�ththeanalyzersampl�ngfresha�r,entertheCal�brat�onModeasdescr�bed�nSect�on5.1.ThenrepeatedlypresstheeNTeR buttonunt�l“CO”appears�ntheupper-lefts�deofthed�splay.
0CAL
CO pp m
2. Allowthepumptorunandsamplefresha�rforatleast1m�nute.
3. Usetheandbuttonstosetthed�splayedvalueto0ppm.
4. Dooneofthefollow�ng:
a.Endth�sprocedureandsavethenewzerovaluebyhold�ngdowntheeNTeR buttonfor2seconds;afterwh�ch,theanalyzerenters�tsRunMode.
b.Cont�nuew�thStep5tospantheCOsensoronaknownconcen-trat�onofcarbonmonox�de.
5. SetuptheCal�brat�onK�tw�th100or500ppmCOasdescr�bed�nthe�nstruct�onssuppl�edw�ththek�t.
6. Connectthetub�ngoftheCal�brat�onK�ttotheGAS�nletoftheanalyzer;thenadjusttheregulatorforapprox�mately2SCFHofexcessflow(seeF�gure5.1).
7. Aftertheanalyzerhasstab�l�zed(2to3m�nutes),usetheandbuttonstosetthed�splayedvaluetomatchtheCOconcentrat�onstampedonthegascyl�nder.
Instruction 24-938840
Fyrite Pro
8. PresseNTeRtostorethenewcal�brat�onvaluesandmovetothenextcal�brat�onscreen,orholddowneNTeRfor2secondstostorethenewcal�brat�onvaluesandentertheRunMode,orpresstheI/Obuttontoex�ttheCal�brat�onModeandturnOFFtheanalyzerw�thoutsav�ngthechanges.
9. D�sconnecttub�ngfromanalyzerandturnoffgasflow.
Figure 5-1. Calibration Kit Hookup
1
2 3 4 5
LEAVE TOPPORT OPEN
Parts Shown:
1. Gas Cylinder2. Regulator*3. Tubing*4. Tee*5. Flowmeter*6. Fitting, Gas** Contained in Calibration Kit
3
6
Instruction 24-9388 41
Fyrite Pro
5.7 Draft CalibrationMaterial Required: •Bellows •M�cromanometer Range:±20mb(±8�n.H2Ocolumn) Accuracy:±0.025mB(±0.01�n.H2Ocolumn)Procedure:
1. Setthedraftun�tsto“mB”perSect�on3.5.4.2. EntertheCal�brateModeperSect�on5.1,andthen
presstheeNTeR buttonunt�l“mB”appearsonther�ght-hands�deofthed�splay.
3. Ifnecessary,zerothedraftchannelbyus�ngtheandbuttonstoadjusttheread�ngto0.00±0.02.
4. Setuptheanalyzerand�tstestequ�pmentperF�gure5-2.5. Usethebellowstoapplyapos�t�vepressureof10.00mB(4.00�n.
H2O),andthenusetheandbuttonstoadjusttheanalyzer’sd�splayedvaluetomatchtheread�ngonthemanometer.
6. RemovepressureandrepeatSteps3thru5unt�lthedes�redread-�ngsareobta�ned.
7. PresseNTeRtostorethenewcal�brat�onvaluesandmovetothenextcal�brat�onscreen,orholddowneNTeRfor2secondstostorethenewcal�brat�onvaluesandentertheRunMode,orpresstheI/Obuttontoex�ttheCal�brat�onModeandturnOFFtheanalyzerw�thoutsav�ngthechanges.
Figure 5-2. Draft Calibration Equipment
1
2
3
1
Manometer
Bellows
Parts Shown:
1. Tubing*2. Tee*3. Fitting, “+” Pressure Port* Contained in Calibration Kit
Instruction 24-938842
Fyrite Pro
5.8 Water Trap / Filter MaintenanceTheWaterTrap/F�lterAssemblyremoveswatercondensatefromthegassample,andalsopreventssootfromcontam�nat�ngthe�nternalcompo-nentsoftheanalyzer.
Drain the water condensate after every test.
Procedure:
1. PullofftheInletEndCapus�ngasl�ghttw�st�ngmot�on.
2. Pouroutallwatercondensate;thenreassembletrap.
Replace the Filter element when dirty.
Material Required: •F�lterElement,P/N07-1644
Procedure:
1. PullofftheOutletEndCapus�ngasl�ghttw�st�ngmot�on.
2. Removeandd�scardoldf�lter.
3. Installnewf�lter;thenreassembletrap.
Figure 5-3. Water Trap / Filter Assembly
OUTLET
FILTER ELEMENT
To replace Filter Element,pull off Outlet End Cap usinga slight twisting motion.
To empty Water Trap,pull off Inlet End Cap usinga slight twisting motion.
INLET
Instruction 24-9388 43
Fyrite Pro
5.9 O2 Sensor ReplacementBesuretoperformallofthechecksl�sted�nSect�on4.13.11toensurethattheO2sensorneedstobereplaced.
Material Required: •OxygenSensor,P/N24-8106 •#1Ph�ll�psScrewdr�ver
Procedure:
1. D�sassembletheanalyzerasfollows:a. Removethebatterycoverandthebatter�es,uncover�ngoneof
thecoverhold-downscrews.b. Removeandsetas�deallfourcoverhold-downscrews.
c. W�ththeanalyzeron�tsback,removethefrontcover,lay�ng�tfacedowntotheleftofthebody.
d. Carefullyremovethec�rcu�tboard,sl�pp�ngoffthebatterycon-nectorontop,andthenlay�ngthec�rcu�tboardfacedown�nthetopcover.
2. Sl�pofftheoxygensensor’selectr�calconnectorfromthec�rcu�tboard.Thenpushdown;tw�stcounterclockw�se;thenpulltheoxy-gensensoroutof�tssocket(seeF�gure5-4).Tip:To obtain a better grip on the oxygen sensor, it may be necessary to remove the screw that secures the sensor socket to the case.
3. Us�ngtheoldsensorasagu�de,removethepaperback�ngfromthenewsensorgasketconta�ned�nthereplacementk�t,andadhere�ttothenewsensor.
4. D�sposeoftheoldoxygensensor�napropermanner(seethe�n-struct�onsheetthatcomesw�ththenewsensor).
5. Mountthenewoxygensensor�n�tssocket.Ifthesensorsocketwasremoved�nStep2,re-attach�ttothecase.
6. Plugtheoxygensensor’selectr�calconnector�ntothepr�ntedc�rcu�tboard(observepolar�ty,seeF�gure5-5),andthenreassembletheanalyzer.
Note: The sensor may take several hours to stabilize after being connected to the printed circuit board.
Instruction 24-938844
Fyrite Pro
5.10 CO Sensor Replacement (Model 125e only)
Besuretoperformallofthechecksl�sted�nSect�on4.13.11toensurethattheCOsensorneedstobereplaced.
Material Required: •COSensor,P/N24-7265 •COSensorGasket,P/N24-1112 •#1Ph�ll�psScrewdr�verProcedure:
1. D�sassembletheanalyzerasfollows:
a. Removethebatterycoverandthebatter�es,uncover�ngoneofthecoverhold-downscrews.
b. Removeandsetas�deallfourcoverhold-downscrews.
c. W�ththeanalyzeron�tsback,removethefrontcover,lay�ng�tfacedowntotheleftofthebody.
d. Carefullyremovethec�rcu�tboard,sl�pp�ngoffthebatterycon-nectorontop,andthenlay�ngthec�rcu�tboardfacedown�nthetopcover.
2. GentlypullCOsensoroutof�tssocket(seeF�gure5-5).
3. Properlyd�sposeoftheoldCOsensor(seethe�nstruct�onsheetthatcomesw�ththenewsensor).
4. It�srecommendedthattheCOsensorgasketbereplacedatthesamet�measthesensor.Removetheoldgasketandd�scard.Re-movethepaperback�ngfromthenewgasketandadhere�ttothecase�nthesamepos�t�onastheoldone.
Important: Ensure that the new gasket does not restrict the inlet and exhaust holes in the case.
5. PlugthenewCOsensor�nto�tssocket.
6. Reassembletheanalyzer.
Instruction 24-9388 45
Fyrite Pro
5.11 Pump Assembly ReplacementMaterial Required: •Pump,P/N24-3048 •#1Ph�ll�psScrewdr�verProcedure:
1. D�sassembletheanalyzerasfollows:
a. Removethebatterycoverandthebatter�es,uncover�ngoneofthecoverhold-downscrews.
b. Removeandsetas�deallfourcoverhold-downscrews.
c. W�ththeanalyzeron�tsback,removethefrontcover,lay�ng�tfacedowntotheleftofthebody.
d. Carefullyremovethec�rcu�tboard,sl�pp�ngoffthebatterycon-nectorontop,andthenlay�ngthec�rcu�tboardfacedown�nthetopcover.
2. Sl�poffthepumpmotor’selectr�calconnectorfromthec�rcu�tboard.
3. Unscrewthepump’sholddownclampandremove�tfromthepump(seeF�gure5-4).Makenoteofhowthepumpw�r�ng�srouted.
4. Makenoteofhowthetub�ngconnectstothepump;thencarefullyremovetub�ngfrompump.
5. Removetheoldpumpandd�scard.
6. Installthenewpumpandre�nstallthetub�ng,tak�ngcarenottop�nchorcr�mpthetub�ng.Alsobesurepumpw�r�ng�sroutedaswasnoted�nStep3.
7. Reassembletheanalyzer.
Instruction 24-938846
Fyrite Pro
Figure 5-4. Inside Case Components
CO SENSORGASKET(Model 125E only)
OXYGEN SENSOR& SOCKET
PRESSURESENSORON PCB
PUMP
HOLDDOWNCLAMP
Instruction 24-9388 47
Fyrite Pro
Figure 5-5. PCB Components
COSENSOR(Model 125E only)
ToBATTERY
ToOXYGENSENSOR
ToPUMP
T-STACKT-AIR
PRESSURESENSOR
(REDWIRE)
(REDWIRE)
(REDWIRE)
Instruction 24-938848
Fyrite Pro
Notes:
Instruction 24-9388 49
Fyrite Pro
6.0 PARTS & SeRVICe
6.1 Replacement PartsDescription Part No.CarbonMonox�deSensor........................................................................... 24-7265CarbonMonox�deSensorGasket...............................................................24-1112OxygenSensorw/Gasket.......................................................................... 24-8106OxygenSensorGasket................................................................................24-1111ProbeStopw/Thumbscrew.......................................................................19-3037PumpAssembly..........................................................................................24-3048WaterTrap/F�lterAssembly,Complete....................................................24-1107WaterTrap/F�lterAssembly,F�lterElement(packof3).........................07-1644
6.2 AccessoriesSTANDARD:Battery,“AA”Alkal�ne............................................................................ 204-0004Carry�ngCase............................................................................................24-0865Instruct�onManual....................................................................................24-9388Probe,Hose,andWaterTrap/F�lterAssembly....................................... 24-7263Qu�ckStartGu�de......................................................................................24-9393
OPTIONAL:ACAdapter(BatteryEl�m�nator).............................................................. 24-1254Cal�brat�onK�t........................................................................................... 24-7059GasCyl�nder: 100ppmCO..........................................................................................51-1994 500ppmCO.........................................................................................24-0492 100%N�trogen.................................................................................9550-0049Pr�nter,IrDA(�ncludes120VACcharger)................................................ 24-1229Pr�nterPaper: 1Roll....................................................................................................06-8733 5RollPack............................................................................................24-1310Protect�veRubberBootw/Magnets.......................................................... 24-1127SoftCarry�ngCase.................................................................................... 24-1267Thermocouple,SecondChannel: 1�nch...................................................................................................104-1798 10feet..................................................................................................104-1797
Instruction 24-938850
Fyrite Pro
6.3 Service CentersReplacementpartsandserv�cecanbeobta�nedbycontact�ngoneofthefol-low�ngBacharachServ�ceCenters:
United States621 Hunt Valley CircleNew Kensington, PA 15068Phone: 724-334-5051Fax: 724-334-5723Email: [email protected]
CanadaBacharach of Canada, Inc.250 Shields Court Unit #3Markham, Ontario L3R 9W7 CanadaPhone: 905-470-8985Fax: 905-470-8963Email: [email protected]
MéxicoBacharach de MéxicoPlaya Regatas No. 473 Tercer PisoCol. Militar MarteDelegación Iztacalco, 08830México D.F. MéxicoPhones: +52-555-634-7740 +52-555-634-7741FAX: +52-555-634-7738Email: [email protected]
Instruction 24-9388 51
Fyrite Pro
Appendix A – Formulas and Tables
CO2 =CO2 max (20.9 – O2%)
20.9
qA = (TA – TL) x ( +B)A2
20.9 – O2
λ (Excess Air) = 20.920.9 – O2
Eta = (100% – qA) + [Kf, if TA < (T0 – 1.2 x O2)]
CO Air Free (Undilute) = CO x λ
CO/CO2 =COppm x 10–6
CO2% x 10–2
Where:O2 = Measured Oxygen in percentTA = Measured stack temperature in °CTL = Measured combustion air temperature in °CCO = Measured carbon monoxide level in ppmCO2 = Calculated carbon dioxide level in percentqA = Calculated stack loss in percentλ =CalculatedLAMBDAA2, B, CO2 max, T0, and Kf are constants (see table below):
Fuel Code
Fuel A2 B CO2 max T0 Kf
F1 Natural Gas 0.66 0.009 11.8 58 °C 9.6%F2 Propane 0.63 0.008 13.8 54 °C 7.6%F3 Light Oil 0.68 0.007 15.4 51 °C 5.3%F4 City Gas 0.63 0.011 13.1 0 °C 0%
Efficiency = Eta – Kf
Instruction 24-938852
Fyrite Pro
Notes:
Instruction 24-9388 53
Fyrite Pro
Notes: