Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10064 Mesa Ridge Ct #1208 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Drip Club – 8-30-17 September 5th, 2017 September 6th, 2017
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Strudelhaus N/A ND 50.66ug/mL ± 0.507ug/mL
165.9ug/mL ± 1.66ug/mL
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10054 Mesa Ridge Ct #132 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Milkman / Drip Club November 1st, 2016 November 3rd, 2016
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Churrios N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10054 Mesa Ridge Ct #132 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Milkman / Drip Club November 1st, 2016 November 3rd, 2016
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Crumbleberry N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10054 Mesa Ridge Ct #132 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Milkman / Drip Club November 1st, 2016 November 3rd, 2016
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Milky O’s N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10054 Mesa Ridge Ct #132 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Milkman / Drip Club November 1st, 2016 November 3rd, 2016
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Moonies N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10054 Mesa Ridge Ct #132 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Milkman / Drip Club November 1st, 2016 November 3rd, 2016
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Original Milkman N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10054 Mesa Ridge Ct #132 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Milkman / Drip Club November 1st, 2016 November 3rd, 2016
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Pudding N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10064 Mesa Ridge Ct #1208 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Drip Club – 8-30-17 September 15th, 2017 September 17th, 2017
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Hazel N/A 42.4ug/mL ± 0.424ug/mL
202.3ug/mL ± 2.023ug/mL
2328ug/mL ± 23.28ug/mL
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10064 Mesa Ridge Ct #1208 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Drip Club – 8-30-17 September 5th, 2017 September 6th, 2017
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman – Little Dipper N/A ND 99.5ug/mL ± 1.00ug/mL
184.3ug/mL ± 1.84ug/mL
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10064 Mesa Ridge Ct #1208 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Drip Club – 8-30-17 September 5th, 2017 September 6th, 2017
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman Delights – Pink2 N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10064 Mesa Ridge Ct #1208 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Drip Club – 8-30-17 September 5th, 2017 September 6th, 2017
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman Delights – Truffleberry N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10064 Mesa Ridge Ct #1208 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Drip Club – 8-30-17 September 5th, 2017 September 6th, 2017
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman Delights – Bubble Milk N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein
Document Rev. 1.7 - September 2016 2
Notes: Samples were received in good condition with factory seals intact. Chain of custody was properly maintained during the entirety of the sampling and receipt process. This document was drafted as a .pdf and contains two pages.
Document Rev. 1.7 - September 2016 1
Nude Nicotine Analytics
10064 Mesa Ridge Ct #1208 San Diego, CA 92121 www.nudenicotine.com [email protected] (858) 216-2044
Gas Chromatography Tandem Mass Spectrometry (GC/MS) Analysis Results:
Customer Name: Date of Sample Receipt: Date of Analysis:
Drip Club – 8-30-17 September 15th, 2017 September 17th, 2017
Analytical Compound: Description: CAS #: Limit of Detection:
2,3-Butanedione (diacetyl) Diketone 431-03-8 1.08ug/mL
2,3-pentanedione (acetyl propionyl) Diketone 600-14-6 1.26ug/mL
3-Hydroxy-2-butanone (acetoin) Ketone 513-86-0 1.01ug/mL
Scope of Method to be Performed: Diketone Trio Analysis:
Aliquot is diluted 10:1 with methanol containing a known amount of butyric-d7 acid as an internal standard. The aliquot is then capped, mixed, and injected into the GC/MS system. Analyte identification is confirmed by no less than 3 unique ion fragments by the mass selective detector. Results are quantitated utilizing a calibration curve generated from certified reference materials.
Product Name: External Lot Number:
Results – Diacetyl:
Results – Acetyl Propionyl:
Results – Acetoin:
Milkman Delights – Melon Milk N/A ND ND ND
<LOQ = Below limit of Quantitation / ND = Not Detected / 1ug/mL = 1ppm / 1000ug/mL = 1mg/mL
The above results are guaranteed accurate and precise to the best of our ability. Data has been checked and in accordance with Nude Nicotine’s approved analytical protocols. I certify that all results in this report are adherent to this statement:
GC/MS Operator and Analytical Report
Prepared by: Signature & Date:
J. Rubenstein