GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5. Easy to Change k’ (Temperature Program)

Upload: christian-grant

Post on 12-Jan-2016

213 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

GC Advantages

1. Very Large N (Very Long Columns)2. No Packing Material (A=0)3. Simple Mobile Phase (Compressed Gas)4. Universal Detectors (FID)5. Easy to Change k’ (Temperature Program)

Page 2: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

GC Limitations

1. Analytes must be Thermally Stable2. Analytes must be Relatively Volatile3. MW < 4004. Not possible to operate at biological conditions

If these limitations are not critical, then GC is probably the BEST means for analyzing a complex sample

Page 3: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

GC Limitations

1. Analytes must be Thermally Stable2. Analytes must be Relatively Volatile3. MW < 4004. Not possible to operate at biological conditions

If at least one of these is a serious concern, then another technique must be employed.

Page 4: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Modern Liquid Chromatography

(post 1969)

HPLC

High Performance Liquid Chromatography

(originally High Pressure Liquid Chromatography)

Page 5: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

LIQUID CHROMATOGRAPHY

Page 6: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

HPLC INSTRUMENTATION

1. Mobile Phase Supply2. Sample Injector3. Column (Stationary Phase)4. Detector

Page 7: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

HPLC Stack Configuration

Page 8: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Mobile Phase Supply: Solvent Reservoirs

Page 9: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Mobile Phase Supply: Solvent Reservoirs

To Pump

Page 10: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Mobile Phase Supply: Syringe Pump

Provides a constant, smooth, high pressure flow.

Difficult to mix or change solvent (reservoir is inside the pump)

Page 11: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Mobile Phase Supply: Reciprocating Pump

Draws solvent(s) from an external reservoir.

Flow is not as uniform, dissolved gas can be troublesome.

Pump only works if liquid is in chamber (must be primed)

Page 12: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Mobile Phase Composition

Reversed Phase ≡ Retention decreases as mobile phasepolarity decreases

Aqueous Mobile Phases

Normal Phase ≡ Retention decreases as mobile phasepolarity increases

Organic Mobile Phases

Page 13: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Advantages of Reversed Phase HPLC

1. Weak Attractive Forces

2. Aqueous mobile phase, sometimes with added liquid organic modifiers (MeOH, Acetonitrile), dissolved salts, and/or buffers.

3. Wide scope: may separate polar, non-polar, ionizeable, and ionic compounds (perhaps at the same time).

4. Elution occurs in order of decreasing polarity (but not at predictable as GC Retention Index).

Page 14: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5
Page 15: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Effect of Organic Modifiers(separation of common analgesics)

Mobile phase: Water + % MeOH + 0.5% H3PO4

Page 16: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Effect of pH(separation of sulfa drugs)

Mobile phase: 20 mM KH2PO4 : acetonitrile (95:5)

Page 17: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Separation of aromatic carboxylic acids

Isocratic:constant 0.055 M sodium nitrate

Gradient:0.01 to 0.1 M sodium nitrate in 25 min

Page 18: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Separation of amino acids with pH Gradient Elution

Page 19: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Sample Injection

Page 20: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Sample Injection

Page 21: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Auto-injection

Page 22: GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5

Stop Here for now!