gc selection guide 2008 - brechbuehler ag: ch...injection port cleaning kit 71 polyimide resin70...
TRANSCRIPT
-
1
Zeb
ron
Ove
rvie
wIn
trod
ucti
on t
o C
olum
n S
elec
tion
Zeb
ron
GC
Col
umns
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
GC SELECTION GUIDECOLUMNS • ACCESSORIES • SAMPLE PREPARATION
-
2
Zeb
ron
Overview
Introduction to Colum
n Selection
Zebron G
C C
olumns
GC
Accessories
SP
E P
roductsA
pplications
The History of ZebronPhenomenex began as a liquid chromatography company in 1982
and soon became recognized for excellent quality, outstanding cus-
tomer service, and technical support. After becoming the leading
HPLC column company in the world, it began to expand business
into other related fields. Phenomenex acquired a GC column manu-
facturing company in northern California, that was originally founded
by Dr. Robert Wohleb, who was also the co-founder of J&W (the “W”
of J&W). From the outset, Phenomenex’s goal was to provide scien-
tists the highest quality line of GC columns, as it had with HPLC. In
1998, Phenomenex introduced Zebron GC columns.
Zebron’s quality was soon recognized throughout the world. Today,
Phenomenex employs many of the world’s top GC scientists, with
decades of cumulative experience in GC column chemistry and
manufacturing technology. This expertise, and our constant dedica-
tion to quality and service, has allowed Phenomenex to become one
of the leading column companies in the world. Phenomenex’s relent-
less pursuit of innovation and customer support is never ending.
*This guarantee is not valid in India and China.
*
If you are not completely satisfied with any of the products featured in this brochure, return within 45 days for a FULL REFUND.
PHENOMENEX | WEB: www.phenomenex.com
-
3
Zeb
ron
Ove
rvie
wIn
trod
ucti
on t
o C
olum
n S
elec
tion
Zeb
ron
GC
Col
umns
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
Table of ContentsZebron Overview 2-9 The History of Zebron 2 Experience the Difference with Zebron 4 Industry Leading Quality 5 Low Bleed 6
High Temperature 7 Longer Column Lifetime 8
Developing Award Winning Solutions for Challenging Analysis 9Introduction to Column Selection 10-30
General Considerations for Column Selection 10 Stationary Phases 10 Selectivity vs. Polarity 11 Stationary Phase Selectivity 11 Summary of Stationary Phase Selection 13
Column Length 14 Influence of Length on Resolution 14 Influence of Length on Analysis Time 14 Other Considerations 14 General Guidelines 15
Column Internal Diameter (ID) 16 Efficiency & Analysis Speed 16 Sample Capacity 18 Flow Rate & Column Head Pressure 18 General Guidelines 19
Column Film Thickness 20 Retention (Phase Ratio) 20 Influence on Efficiency 22 General Guidelines 23
GC Column Selection Chart 24GC Columns by Manufacturer 25GC Column Selection for EPA Methods 26-28GC Column Selection for ASTM Methods 29GC Column Selection for USP Methods 30
Zebron GC Columns 31-53 Integrated Guard Columns 52
Guard Columns & Retention Gaps 53GC Accessories 54-73
Capillary Column Mini Unions 70Capillary Unions & Splitters 70Ceramic Scoring Wafer 71Column Check Standards 54Consumables for GC Systems 56-67(Includes: Ferrules, Injection Port Inlet Liners, Septa, Syringes)
Agilent 56-61PerkinElmer 64-65
Shimadzu 66-67 Varian 62-63
Cool-Lock Nut 55Ferrule Remover Tool Kit 71Flame Detector Jet Cleaning Kit 71Gas Traps & Purifiers 68-69
Injection Port Cleaning Kit 71 Polyimide Resin 70
Vials 72-73SPE Products 74-77Applications 78-104 Environmental 78 Toxicology 88 Pharmaceuticals 91
Food, Flavors, Fragrances 93 Industrial Chemicals 96 Petroleum 101
Mixed-Polarity
General Analytes
ZB-5msZB-MR-1ZB-XLB
Pesticides
ZB-MR-1ZB-MR-2ZB-5msZB-XLB
PolarNon-Polar
Your Analytes
10 20 30 min
PHENOMENEX | WEB: www.phenomenex.com
-
4
Zeb
ron
Overview
Introduction to Colum
n Selection
Zebron G
C C
olumns
GC
Accessories
SP
E P
roductsA
pplications
Experience the Difference with Zebron…
• Don’t Get Put on Hold
• Get Answers When You Need Them
• Experienced, Specialized GC Technical Support
• Available Overnight Delivery, Worldwide
• Application Development Support
• Free On-site Training
Have you ever called some of the other column manufacturers and been put on hold for long
periods of time, only to be forced to leave a message and hope someone called you back?
At Phenomenex, we believe that our customers are the lifeblood of our business and that
they deserve the highest level of service.
We guarantee that you will always speak to an experienced GC representative each and
every time you call. We are also available for troubleshooting and method development
questions. Our full staff of technical support specialists and laboratory personnel will give
you the answer you need, when you need it. Experience the difference with Zebron, where
you, the customer, are our number one priority.
“At Phenomenex we believe that customers are the lifeblood of our business. Big or small, your business
is important to us. We want to grow with you and
become a valued part of your business’s success. Let us
show you how the Zebron experience puts you first.”
PHENOMENEX | WEB: www.phenomenex.com
-
5
Zeb
ron
Ove
rvie
wIn
trod
ucti
on t
o C
olum
n S
elec
tion
Zeb
ron
GC
Col
umns
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
Industry Leading Quality• Every Column is Individually QC Tested
• Excellent Batch-to-Batch Reproducibility
• Industry Leading Quality Control Standards
Our commitment to quality is what makes Zebron the number one fastest growing GC
column line on the market. Each and every column you receive undergoes an exhaustive QC
check process that ensures all columns provide the highest level of performance.
Individually QC Tested
To start, all columns are individually QC tested by an experienced technician for important
parameters such as bleed, activity, retention, and efficiency. Our QC test standards are so
stringent that we must spend hours qualifying our GC systems to ensure they are sensitive
enough to measure the small differences defined by our pass/fail windows.
Aggressive Text Mixes
Ensuring our columns meet the needs of your analysis often requires using special QC test
procedures. Many of our products utilize specially designed test mixes that are formulated
to mimic the worst compounds that you encounter on a daily basis. Holding ourselves to
such high standards gives you the confidence that Zebron columns will work for your tough-
est methods, every time.
EfficiencyCritical for measuring the separation power of your column. Good columns have been typicallyconsidered to be above 3,700 plates/meter. The average efficiency for a ZB-5ms column exceeds 4,000 plates/meter, giving you the maximum separation possible.
BleedColumn bleed gives an indication of column stabil-ity and lifetime. Zebron columns provide industryleading bleed levels providing maximum sensitivity and stability.
ActivityColumn performance for acid and base compounds is often critical for achieving method stability. We monitor several acids and bases chosen specifi-cally to mimic the most challenging compounds our customers face.
RetentionTo ensure you receive reproducible product, we monitor the retention characteristics of our columns very closely. Hydrocarbons serve as a reference point to polar compounds.
PHENOMENEX | WEB: www.phenomenex.com X
-
6
Zeb
ron
Overview
Introduction to Colum
n Selection
Zebron G
C C
olumns
GC
Accessories
SP
E P
roductsA
pplications
Low Bleed• Lower Detection Limits
• Improved Quantitation
• Improved Compound Identification
• Less System Maintenance
Column bleed is the loss of lower molecular weight stationary phase pieces (MS Ions 355,
281, 207, 73) that are either the result of impurities in the starting polymer or the decom-
position of the phase at elevated temperatures. Engineered Self Cross-linking (ESC)™
bonding technology reinforces the stationary phase by incorporating ladder bridges into
the polymer backbone. The process begins with a careful fractionation of the base polymer,
which eliminates low molecular weight impurities and enhances coating efficiencies. The
columns are then cross-linked and surface bonded using an aggressive catalyst forming
an interpenetrating network, which provides enhanced column durability and extremely low
bleed levels.
Column bleed can also be an indicator of the stability and lifetime a GC column will offer.
When using sensitive detectors such as MS, bleed can impact method detection limits
(MDLs) causing difficulties at the low calibration ranges. Zebron columns provide very low
bleed levels making them sensitive and stable for a long time.
Column Bleed Comparison Test by MS Detection
Dimensions: 30 meter x 0.25 mm x 0.25 μm
Injection: 225 ˚C with a constant flow of 1.1 mL/min of helium
Oven Program: 240 ˚C for 10 min to 280 ˚C @ 40 °/min for 10 min to 320 ˚C @ 40 °/min for 10 min to 340 ˚C @ 40 °/min for 10 min to 360 ˚C @ 40 °/min for 10 min
Detector: MS, Scan range: 36-450 amu
Testing conditions were similar for all columns. All columns were new and obtained from the original manufacturer or its authorized channels. Data provided by Phenomenex in-house evaluations and represents bleed values obtained after 2 hours of conditioning.
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 10 20 30 40 50 60 min
Thou
sand
s
240 ˚C 280 ˚C
320 ˚C
340 ˚C
360 ˚C
12
3
4
51 Zebron™ ZB-5
Equivalent column from leading competitors:
2 Varian® Factor Four™, VF-5ms
3 J&W® DB-5ms
4 Agilent® HP-5ms
5 Restek® Rtx-5Sil MS
MS C E R T I F I E D
PHENOMENEX | WEB: www.phenomenex.com
-
7
Zeb
ron
Ove
rvie
wIn
trod
ucti
on t
o C
olum
n S
elec
tion
Zeb
ron
GC
Col
umns
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
High Temperature• Analyze High Molecular Weight Compounds
• Bake Off Matrix Contaminants
• Shorten Run Times
People often ask why do Zebron columns have such a high upper temperature limit com-
pared to everyone else? The answer lies with our proprietary bonding process called En-
gineered Self Cross-Linking (ESC). By carefully fractionating our base polymer and using
aggressive cross-linking catalysts, we produce a column that resists thermal degradation.
What this means to you is that all Zebron columns will give you the flexibility to use high
temperature to improve your GC analysis. For ultra high-temperature analyses, our award-
winning Zebron-Inferno™ line will let you to take your GC oven to 430 °C and still maintain
superior performance.
Butter Triglycerides
Column: Zebron ZB-5HT Inferno
Dimensions: 15 meter x 0.32 mm x 0.10 μm
Part No.: 7EM-G015-02
Injection: On-Column @ 103 ºC, 2 μL
Carrier Gas: Helium @ 1.8 mL/min (constant flow)
Oven Program: 100 ºC to 400 ºC @ 14 ºC/min for 10 min
Detector: FID @ 400 ºC
10 20 30 min
App
ID 1
6054
430 ˚C TPGC Stability Test*
At temperatures above 380 ˚C, conventional fused silica
tubing will become brittle and randomly break. Our special
new tubing is manufactured using a novel High Temperature
Polyimide Resin material that shows minimal thermal degra-
dation even at programmed temperatures up to 430 ˚C.
*Evaluated by performing 185 programmed temperature runs, total 23 hours at 430 ˚C. Polyimide tubing wasstill flexible as shown here.
Zebron ZB-Inferno™Still Flexible After 23 Hours at 430 ˚C
2007 Winner
PHENOMENEX | WEB: www.phenomenex.com X
-
8
Zeb
ron
Overview
Introduction to Colum
n Selection
Zebron G
C C
olumns
GC
Accessories
SP
E P
roductsA
pplications
Longer Column Lifetime• Analyze Dirty Samples
• Reduce Instrument Downtime Due to Column Changes
• Save Time and Money
Our customers consistently tell us that our columns last much longer than other products
they have used. We received the following picture and email from one of our customers do-
ing environmental analysis. They operate a mobile testing lab, thus don’t have the facilities
to do a sophisticated sample preparation.
They were contracted to do pesticide analysis for a former department of defense facility.
Many of the samples were taken from the ocean floor and contained high levels of pesti-
cides and other contaminants. Since they did not use guard columns and sample prepa-
ration was minimal, they expected column lifetime to be very short. To their surprise, the
Zebron columns lasted for the entire project. They were so impressed that they sent us this
picture and email:
“We wanted to share with you and Phenomenex the great success we had using your Zebron columns. We used 4 of your ZB-35, 30 meter columns to perform detailed pesticide analysis of soils during environmental remediation at a former Dept. of Defense facility.
Over the course of 6 months, we analyzed over 6500 samples on 2 GCs, many of these samples had high levels of pesticides and we did not use a guard column. Dan Twomey, our senior chemist and Vice President, personally performed much of the analysis and had nothing but high praise for the durability and consistency of the columns.
We are attaching a photo of the extract vials for the complete project.”Bob Burger - ESN North Atlantic
PHENOMENEX | WEB: www.phenomenex.com
-
9
Zeb
ron
Ove
rvie
wIn
trod
ucti
on t
o C
olum
n S
elec
tion
Zeb
ron
GC
Col
umns
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
Developing Award Winning Solutions for Challenging AnalysisIn the field of gas chromatography not much has changed for the last 15 years. However
technology is our passion and Zebron is gaining international recognitions for its new prod-
ucts that are specially designed for today’s most challenging methods.
High Performance MS Certified Phases
Our line of high performance MS columns provides the ultimate low bleed and low activity
chromatography. The Zebron ZB-1ms, ZB-5ms, ZB-5MSi, ZB-WAX, and ZB-XLB phases
are guaranteed to improve your analysis, saving your lab time and money.
100 % Water Stable Phases
The use of water in GC has always been considered a taboo, especially when work-
ing with Polyethylene Glycol (WAX) phases. The Zebron ZB-WAXPLUS column was
specially formulated to provide stable performance even with 100 % aqueous samples. The
water stability of the ZB-WAXPLUS can greatly improve the analysis of wine, beer, and dis-
tilled spirits.
Improved Analysis of Pesticides, Herbicides and Insecticides
Pesticides are the subject of increasing international regulation. Since many different types
of pesticides can be used on the same food product, the need to properly separate and
identify these compounds is important. Regardless of the technique, the new ZB-Multi-
Residue™ GC columns have been optimized for pesticides, herbicide, or insecticide analy-
sis. The columns have orthogonal selectivity that allows them to be used together in dual
column confirmation analysis for EPA Methods. Both phases are also MS Certified, so they
can be used on MS to confirm results or for multi-pesticide residue screening methods.
Unsurpassed High Temperature GC Analysis
At temperatures above 380 ˚C, conventional fused silica tubing will become
brittle and randomly break. Our Inferno columns are manufactured using a
novel high temperature polyimide resin material and stationary phase polymer
that shows minimal thermal degradation even at programmed temperatures up
to 430 ˚C.
Zebron Inferno™ columns have been recognized by R&D Magazine as one of the 100 most
technologically significant products introduced into the marketplace in 2007. Further, they
are the only GC columns in history to ever receive this award. This is a significant accom-
plishment and a true testament to their stability and performance.
2007 R&D 100 Award Recipient
PHENOMENEX | WEB: www.phenomenex.com
-
10
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
Stationary PhasesResolution between two components is mainly determined by the selectivity ( ) of the sta-
tionary phase. Column selectivity depends on the nature of the stationary phase, the nature
of the components, and the oven temperature at the time of elution. More polar phases
(with different chemical groups) will have more possible ways of interaction.
When we consider the master resolution equation, optimizing column selectivity has the
largest impact on resolution. By increasing the resolution between two compounds, the
total analysis can often be reduced significantly. When looking to optimize your GC method,
we suggest you look to first optimize the selectivity of your stationary phase.
RS =
√—N - 1 k4 k + 1
Efficiency Term
SelectivityTerm
RetentionTerm
General Considerations for Column Selection
PHASE: The stationary phase is the most influential col-umn parameter. It not only determines the final resolution ob-
tained (it determines the relative retention of the solutes), but
due to its specific characteristics, it will influence virtually ev-
ery column selection parameter.
LENGTH: Length is directly related to overall efficiency of the column and to overall analysis time. The relation between
length (efficiency) and resolution is a square root relation, while
analysis time is directly related to column length.
INTERNAL DIAMETER: Column internal diame-ter has a major impact on column efficiency (and thus on reso-
lution) and on the sample capacity of the column. It can also
dictate limitations on the injection and detection techniques
used and vice versa.
FILM THICKNESS: Film thickness determines so-lute retention and thus solute elution temperatures. It will
also play an important role in the sample capacity of the
column. Thin films are faster with higher resolution, but offer
lower capacity.
Other criteria like instrument availability, feasibility, budget considerations, etc. can also impose limitations on column selection.
Experience, literature data, and your Phenomenex Technical Representative are invaluable tools to guide you through the labyrinth
of column selection.
PHENOMENEX | WEB: www.phenomenex.com
-
11
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
ZB-35
ZB-1701
Figure 1: Zebron ZB-35 and the ZB-1701, with polarity 18 and 19 respectively, measured by McReynolds constants
2 4 6 8 10 12 min
12
3
4
5
6 7
8
9
10
11 12
13 14 15 16 17
Selectivity vs. Polarity
Many people often confuse column polarity and selectivity. Column polarity does give a
general guideline for sample capacity and separation, which can affect peak shape and res-
olution. However, even though two columns may have similar polarity, they may show very
different separation profiles because of the differences in the stationary phase chemistry.
To demonstrate this point, consider the separation of various components using the Ze-
bron ZB-35 and the ZB-1701, with polarity 18 and 19 respectively as shown in Table 1,
measured by McReynolds constants. The columns have very similar polarity, however the
column chemistries are very different (Figure 1). The addition of the cyanopropyl group in
the ZB-1701 column gives the column a very unique selectivity that is very different from
the ZB-35.
Stationary Phase Selectivity
Column selectivity is based on differences in interaction between analytes and the station-
ary phase at a specific temperature. The way in which a compound interacts with the phase
is determined by the chemical nature of the stationary phase. Below are the three most
prevalent interactions in GC:
1. Dispersive Forces or Van der Waals Interactions
These are the weakest of all intermolecular forces and occur between non-polar compounds.
This force causes the separation of compounds based on boiling point. Even though this is the
weakest of all the forces, it is the most abundant, making it the dominant interaction with most
GC stationary phases. The separation of hydrocarbons using a standard simulated distillation
analysis is a classic example of dispersive forces. In this method, mixtures of hydrocarbons are
analyzed and elute in peaks corresponding to their carbon number.
App
ID 1
6217 ASTM Method D 2887-97: Boiling Range
Distribution of Petroleum Fractions
Column: Zebron ZB-1HT Inferno
Dimensions: 15 meter x 0.53 mm x 0.15 μm
Part No.: 7EK-G014-05
Injection: On-Column @ 38 ºC, 0.1 μL
Carrier Gas: Helium @ 10 mL/min (constant flow)
Oven Program: 35 ºC for 1 min to 360 ºC @ 25 ºC/min
Detector: FID @ 375 ºC
Sample: 1. Hexane2. Heptane3. Octane4. Nonane5. Decane6. Undecane7. Dodecane8. Tetradecane9. Hexadecane
10. Octadecane11. Eicosane12. Tetracosane13. Octacosane14. n-Dotriacontane15. n-Hexatriacontane16. Tetracontane17. Tetratetracontane
Si OSi O
CH3
CH3
65 % 35 %
Si O
CH3
CH3Si O
NC
(CH2 )3
86 % 14 %
PHENOMENEX | WEB: www.phenomenex.com
-
12
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
12
3
4
56
7 8 9
10
11
6.5 min5.54.53.52.51.5 Significant Tailingand Adsorption for Acids & Bases
2
1
2
3
4 5
6
7
8
9
10
4 6 8 10 min
1
2
3
4
5
67
8
9
10
DeactivatedFor Acids & Bases
46
2 4 6 8 10 min
2. Dipole-Dipole Interactions
These forces can either be present as a permanent dipole or can be
induced by stationary phase-analyte interactions. Choosing a col-
umn with a higher dipole-dipole interaction can help separate com-
pounds that have similar boiling points, but different chemical struc-
tures. An example that demonstrates this concept is the separation
of meta and para-Xylene isomers. When using a non-polar phase
such as a ZB-5 (figure 3), where dispersive forces are the primary
interaction, separation between these two isomers is not possible.
By choosing a phase with a very high dipole moment such as the
ZB-WAX (figure 2), the separation between the two isomers is
achieved.
3. Hydrogen Bonding or Acid-Base Interactions
In GC, these types of forces are usually bad and cause compounds
to exhibit poor peak shape or to irreversibly bind in either the injec-
tion port liner or the column itself. Zebron columns undergo special
deactivation procedures to minimize the acid-base interactions, which
improves peak shape and improves chromatographic separations.
1
2
3
4
56
7,8
9
10
11
9.5 min5.5 6.5 7.5 8.54.53.52.51.5
Zebron™ ZB-5ms
vs.Agilent® HP-5ms
Zebron ZB-5ms vs. Agilent HP-5ms Column Performance Evaluation
Dimensions: 30 meter x 0.25 mm x 0.25 μm
Part No.: 7HG-G010-11
Injection: Split @ 250 ˚C, 100 mL/min
Oven Program: 240 ˚C for 10 min to 280 ˚C @ 40 ˚C/min for10 min to 320 ˚C @ 40 ˚C/min for 10 min to340 ˚C @ 40 ˚C/min for 10 min to 360 ˚C @40 ˚C/min for 10 min
Detector: MS-Scan Range: 36-450 amu
Sample: 1. Propionic Acid2. Octane3. Nitrobutane4. 4-Picoline5. Trimethyl Phosphate
6. 1,2-Pentanediol7. Propylbenzene8. Heptanal9. 3-Octanone
10. Decane
Figure 2: BTEX solution and four alkane markers were separated using a ZB-WAX 30 m x 0.32 mm x 0.50µm column. 1 = pentane, 2 = heptane, 3 = solvent(methylene chloride), 4 = benzene, 5 = decane, 6 =toluene, 7 - ethylbenzene, 8 = p-xylene, 9 = m-xylene, 10 = dodecane, 11 =o-xylene
Figure 3: BTEX solution with markers on ZB-5 30 m x 0.32 mm x 0.25 µm column. 1 = pentane, 2 = solvent(methylene chloride), 3 = benzene, 4 = heptane, 5 = toluene, 6 = ethylbenzene, 7 = m-xylene, 8 = p-xylene,9 = o-xylene, 10 = decane, 11 = dodecane
PHENOMENEX | WEB: www.phenomenex.com X
-
13
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
Summary of Zebron Stationary Phase Selection Based on Phase Polarity
Table 1: Phase Polarity
ZB-1
ZB-1ms
ZB-5
ZB-XLB
ZB-MR-1
ZB-MR-2
ZB-624
ZB-35
ZB-1701
ZB-1701P
ZB-50
ZB-WAX
ZB-FFAP
ZB-5ms
ZB-5MSi
ZB-1HT Inferno
ZB-5HT Inferno
ZB-WAXPLUS
5
LOW-POLARITY MEDIUM-POLARITY HIGH-POLARITY
8 9 11 1513 52 57 5818 19 24
PHENOMENEX | WEB: www.phenomenex.com
-
14
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
Column efficiency is directly proportional to column length, so longer columns will provide
higher resolution. However, there is not a direct relationship between efficiency and resolu-
tion. Based on the master resolution equation, resolution is proportional to the square root
of efficiency. This means that large increases in efficiency will not necessarily result in a
significant increase in separation.
There is also a direct relationship between column length and analysis time. When using
longer columns to increase separation power, the total analysis time will also increase.
The effect is more pronounced when using isothermal runs, but even in temperature pro-
grammed analyses the run time difference can be significant. When choosing column
length, it is important to find a column that provides the necessary resolution within an
acceptable run time.
Influence of Length on Resolution
Increasing the length will increase resolution in a square root relation. Due to disadvantages
of very long columns, increasing resolution by increasing the length is limited.
Influence of Length on Analysis Time
In order to increase resolution or sample capacity by increasing the length, the total analysis
time will also increase. Decreasing analysis time by increasing carrier gas velocity will cause
the column efficiency to decrease.
Other Considerations
Longer columns are more expensive because there is additional material and labor neces-
sary to produce these columns. However, cost per meter might be less for a 60 m column
than a 15 meter column. In an attempt to cut costs, some customers are tempted to cut
down a longer column into several shorter ones. We do not recommend this!
The cost on the shorter column is higher because we have individually QC tested that prod-
uct. When you buy a long column, our QC test evaluates the average performance of that
column over the entire length. In theory, the performance of the column should be the same
over the entire length, but is not guaranteed. The data you generate is too important and the
cost of getting bad data too high to risk anything but the best product.
Column Length
√—L ~ R
L ~ tR
RS =
√—N - 1 k4 k + 1
Efficiency Term
SelectivityTerm
RetentionTerm
PHENOMENEX | WEB: www.phenomenex.com
-
15
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
General Guidelines
The length of the column should be chosen primarily to obtain an adequate efficiency. Most
GC methods have far more resolution than is necessary.
• In general, we recommend starting with 30-meter columns.
These are the ideal balance between resolution and runtime.
• Use shorter columns (15-meter) for sample screening or
samples that contain very few analytes.
• Use 60-meter or longer columns when better separation is
needed or for complex samples with closely eluting peaks.
Table 2: Uses and Trends by Column Length
15 m Applications Chromatographic Trends as Length Increases
105 m Applications
Rapid analysis Increased retention times Low boilers
High efficiency separations Increased efficiency More complex mixtures
Screening Greater resolution Less active samples
Simple mixtures Programmed temperature analysis
High molecular weight compounds
More chemically active components
15 m 150 m
PHENOMENEX | WEB: www.phenomenex.com
-
16
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
ColumnInternal Diameter (ID)
RS =
√—N - 1 k4 k + 1
Efficiency Term
SelectivityTerm
RetentionTerm
The new developments of capillary column technology, and specifically in column ID, are
toward smaller internal diameters. Column efficiency is related to 1/ID of the column, so
the smaller the internal diameter the higher the column efficiency. As with column length,
resolution only increases by the square root of the efficiency so resolution may only increase
by 1.41 (square root of 2) when the ID is reduced by half. But unlike column length, using
smaller ID columns can actually lead to shorter analysis times because the column length
is often shorter.
Efficiency & Analysis Speed
As a general rule, the smaller the internal diameter, the shorter the analysis time will be for
a given resolution. The increase in efficiency offered by narrow bore GC columns often im-
proves separation enough to allow the same separation to be done in much less time.
To show how this is possible, consider Table 3 which shows various column IDs and their
corresponding efficiencies in plates/meter. When we compare a 30 meter x 0.25 mm ID col-
umn with a 20 meter x 0.18 mm ID column, we see that they have similar column efficiencies
or similar resolving power. However, the 0.18 column has 1/3 less length and therefore will
elute compounds in 1/3 less time.
Table 3: Column IDs and their corresponding efficiencies in plates/meter
Figure 4A shows the separation of 17 priority Polyaromatic Hydrocarbons (PAHs) contami-
nants using a standard 30 meter x 0.25 mm x 0.25 µm column. By choosing a column with
a similar phase ratio but smaller ID, the method can be shortened by over 100 % while still
meeting resolution requirements for key analytes (Figures 4B & 4C).
Column ID
(mm)
Efficiency
(Plates/ Meter)
Typical Column
Length (Meters)
Column Efficiency
Plates/Column
0.10 7,500 10 75,000
0.18 5,700 20 114,000
0.20 5,000 20 100,000
0.25 4,100 30 123,000
0.32 3,350 30 100,500
0.53 1,500 30 45,000
PHENOMENEX | WEB: www.phenomenex.com
-
17
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
Figure 4: Separation of 17 priority Polyaromatic Hydrocarbon contaminants using a Zebron ZB-5ms column. By choosing a column with similar phase ratio, but smaller ID, the method can be shortened by over 100 % while still meeting resolution requirements for key analytes. A) A 30 meter x 0.25 mm x 0.25 μm Zebron ZB-5ms will give a run time of 18 min. B) 11 min. run time on a 20 meter x 0.18 mm x 0.18 μm Zebron ZB-5ms. C) A 10 meter x 0.10 mm x 0.10 μm Zebron ZB-5ms run time will be shortened by over 100 % to 8 minutes.
Sample: 1. Naphthalene2. 2-Methylnaphthalene3. Acenaphthalene4. Acenaphthene5. Fluorene6. Phenanthrene7. Anthracene8. Fluoranthene9. Pyrene
10. Benz[a]anthracene11. Chrysene12. Benzo[b]fluoranthene13. Benzo[k]fluoranthene14. Benzo[a]pyrene15. Indeno[1,2,3-
cd]pyrene16. Dibenz[a,h]anthracene17. Benzo[g,h,i]perylene
2 4 6 8 10 12 14 16 18 min
1
2
3
4
5 67
89
10
11
12
13 14 15
16
17
App
ID 1
6017
Figure 4:
A: Separation of PAHs on ZB-5ms 30 meter x 0.25 mm x 0.25 μm ( = 250)
B: Separation of PAHs on ZB-5ms 20 meter x 0.18 mm x 0.18 μm ( = 250)
1
2
3
4 5
6
7
8 9
10
11
12
1314
1516
17
1 3 5 7 9 11 min
App
ID 1
5560
m in2 3 4 5 6 7
1
2
3
45 6 7
8 9
10
11
12
13
14
15
16
17
App
ID 1
5806
C: Separation of PAHs on ZB-5ms 10 meter x 0.10 mm x 0.10 μm ( = 250)
PHENOMENEX | WEB: www.phenomenex.com
-
18
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
Sample Capacity
The smaller the internal diameter of a GC column, the lower the sample capacity, which can
affect method sensitivity. This is especially true when using column IDs < 0.25 mm. Contrary
to most people’s intuition, the difficulty is usually greatest on the higher concentration levels
not at the low concentrations. This is because peaks usually get sharper since they spend
less time in the column. As peaks get sharper, they also get taller so the low level sensitivity
is not usually a problem.
The mega-bore columns (0.53 mm ID) combine the general advantages of capillaries with
the high sample capacity of packed columns. Even a small increase of internal diameter
(e.g., from 0.25 to 0.32 mm) results in a significant increase in sample capacity. This can
be helpful when working with very high concentration samples. Adjusting the split ratio in
your inlet can sometimes allow a more concentrated sample to be injected into a smaller
ID column. See Table 4 below for the typical loading capacities for various ID columns. As
long as method sensitivity can be achieved, we recommend using the smallest internal
diameter possible. This allows you to take advantage of the increased resolving power and
high-throughput capabilities of the narrow bore columns.
Table 4: Typical Sample Capacity (max for single component)
Flow Rate & Column Head Pressure
The optimum flow rate for a given column is dependent on the column dimensions and
the carrier gas used. Higher flow rates usually result in shorter analysis times because the
analytes are being pushed through the column faster. The extent to which you can use flow
to shorten your analysis often depends on the system you are using.
When using GC/MS with larger ID columns and high flow rates, the system can have prob-
lems maintaining acceptable settings. The newer MS systems have better pumping and
flow control devices, which allows for larger ID columns to be used. The use of 0.53 mm ID
is still not possible without using some flow restriction system.
When working with smaller IDs, head pressure can become a problem because they require
a higher head pressure to maintain appropriate flow rates. Flow controlled systems main-
tain a constant flow rate by increasing head pressure as the oven temperature increases.
This can be especially problematic on a 0.10 mm ID column since the final pressure often
exceeds 100 psi. In most applications, we find that using column IDs ≥ 0.18 mm are com-
patible with the pressure limitations of most GC systems.
ID (mm) df (µm) Capacity (ng)
0.18 0.18 20-75
0.25 0.25 50-125
0.32 0.50 100-250
0.53 1.00 500-1,000
Note: Capacity represents maximum loading per component. Sample capacity increases with film thickness.
PHENOMENEX | WEB: www.phenomenex.com
-
19
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
General Guidelines
ID (mm) Instrument Selection / Constraints
0.10 High operating pressures are needed.Might require system modifications.Easily overloaded, requires skilled technicians.
0.18 Similar operating parameters to the 0.25 mm ID.Increased efficiency allows for shorter run times.Good first step when looking to use narrow bore columns.Suited for use with mass spectrometers.
0.20 Small increase in efficiency vs. 0.25 mm ID columns.Suited for mass spectrometers.
0.25 Most popular dimension.On-column injection possible.Suited for mass spectrometric detection.
0.32 Easy to work with; good compromise between resolution and stability.Routine on-column injection.Compatible with newer mass spectrometric detection.
0.53 Can be installed in an adapted packed gas chromatograph.Easy on-column injection.Good for use with dirty or highly concentrated samples.
PHENOMENEX | WEB: www.phenomenex.com
-
20
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
ColumnFilm Thickness (df)A column’s film thickness can affect several important chromatographic parameters including:
• Retention: thicker film columns provide higher retention of low boiling compounds
• Efficiency: thinner film columns provide higher efficiency
• Activity: thicker film columns have lower activity for acids-bases
• Loading Capacity: thicker film columns have higher loading capacity
• Bleed: thinner film columns have lower bleed
Retention (Phase Ratio)
The retention of the solute by the stationary phase largely depends on the temperature
and on the Phase Ratio ( ). Phase ratio for a given column is calculated using Equation 2;
smaller values result in greater retention. Chromatographically this means that when us-
ing columns of the same ID, the column with a thicker film will have greater retention for a
given analyte. For low k solutes (volatile components), retention can be increased through
the value (higher film thickness). Table 5 lists the values for common IDs and film thick-
nesses.
Table 5: Phase ratio ( ) value for common columns
When using columns of two different IDs, the same film thickness does not translate to
the same retention characteristics. The optimum phase ratio depends on the goal of the
separation. If analyte retention is low, a column with a low can be used to increase reten-
tion. If column provides good retention, can be reduced to increase column efficiency and
decrease run time.
Film Thickness
df (µm)
Column Diameter (mm)
0.10 0.18 0.20 0.25 0.32 0.53
0.10 250 450 500 625 800 1325
0.18 139 250 278 347 444 736
0.25 100 180 200 250 320 530
0.33 –– –– 151 –– –– ––
0.50 –– 90 100 125 160 265
1.00 –– –– 50 63 80 133
1.50 –– –– –– 42 53 88
3.00 –– –– –– 21 27 44
5.00 –– –– –– 13 16 27
Increasing Retention
Incr
easi
ng R
eten
tion
Equation 2: = –––––––– 4 x dfID Internal Diameter (µm)df Film Thickness (µm)
ID
RS =
√—N - 1 k4 k + 1
Efficiency Term
SelectivityTerm
RetentionTerm
PHENOMENEX | WEB: www.phenomenex.com
-
21
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
The separation of light hydrocarbon impurities found in butane is an example. On a column
with a high such as the Zebron ZB-1 60 meter x 0.32 mm x 0.25 µm ( = 320) the isomers
co-elute due to the lack of interaction with the stationary phase (Figure 5). By using a column
with a lower such as the Zebron ZB-1 60 meter x 0.32 mm x 3.00 µm ( = 27), we are able to
achieve separation (Figure 6).
Figure 5
Figure 6
There is a limited benefit to increasing column retention. When working at k < 2 min, retention
will tend to fall to 0. With these analytes, increasing the film thickness is a very good way to
improve retention and allow time for separation to take place. In some cases using thick film
columns can eliminate the need to use cryogenic cooling.
The optimal range for retention in most cases is 2-10 min (Figure 7); in complex samples 2-20
min may be necessary. When analysis time is above 20 min, the benefit of increasing reten-
tion time is decreased and the film thickness should be kept as low as possible to maximize
column efficiency.
App
ID 1
5426
0 1 2
2
1
3
4
3 4 5 6 7 min
App
ID 1
4821
0 1 2 3 4 5 6 7 8 min
1
2 3 4
Butane Purity
Column: Zebron ZB-1
Dimensions: 60 meter x 0.32 mm x 0.25 μm
Part No.: 7KM-G001-11
Injection: Split 10:1 @ 200 ˚C, 5 µL
Carrier Gas: Hydrogen @ 6.5 mL/min (constant flow)
Oven Program: 40 ˚C (Isothermal)
Detector: FID @ 200 ˚C
Sample: 1. Ethane2. Propane3. Isobutane4. Butane
Butane Purity
Column: Zebron ZB-1
Dimensions: 60 meter x 0.32 mm x 3.00 μm
Part No.: 7KM-G001-36
Injection: Split 10:1 @ 200 ˚C, 5 µL
Carrier Gas: Hydrogen @ 6.5 mL/min (constant flow)
Oven Program: 40 ˚C (Isothermal)
Detector: FID @ 200 ˚C
Sample: 1. Ethane2. Propane3. Isobutane4. Butane
PHENOMENEX | WEB: www.phenomenex.com
-
22
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
Figure 7: Influence of Retention Factor (k) on Resolution
Influence on Efficiency
Column efficiency is proportional to 1/df, which means thinner film columns will provide the
highest efficiency. For thick film columns (low value), the resistance to mass transfer in the
liquid phase (CL) can no longer be ignored. We recommend you use the thinnest film column
possible that still provides adequate retention.
The exception to this rule is when working with acidic-basic compounds. Thicker film col-
umns prevent secondary interaction between the analyte and silanol groups on the walls of
the capillary tubing. When working with active samples, using a slightly thicker film column
can significantly improve peak shape, calibration stability and can also lead to longer col-
umn lifetimes.
The drawback to using thicker film columns is that they will have slightly higher bleed, which
can affect sensitivity. For analysis where bleed and activity are important, we suggest using
our MS certified phases because they are specially designed to improve both.
1.0
0.9
0.8
Optimal range for retentionEffienciency
0.7
0.6
k/(1
+k)
0.5
0.4
0.3
0.2
0.1
0 2 4 6 8 10 12 14 16 18 20 min
PHENOMENEX | WEB: www.phenomenex.com
-
23
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
General Guidelines
• Thin film columns for the analysis of high boiling compounds.
• Medium film columns for:
– Solutes with wide boiling point range
– Medium boiling compounds
• Thick film columns for volatile compounds
Table 6: Uses and Trends by Column Film Thickness
0.10 µm Applications Chromatographic Trends as Film Thickness Increases
5.0 µm Applications
Rapid analysis Increased retention times Low boilers, e.g., gases, solvents, purgeables (BP < 25 ˚C)
High efficiency separations Increased resolution (for low boiling compounds)
Used more with wide ID columns
High MW compounds Upper temperature limit decreases More chemically active components
Less chemically active components Bleed increases
MS applications More narrow boiling point range for analytes
0.10 µm 5.0 µm
PHENOMENEX | WEB: www.phenomenex.com
-
24
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
Zebron GC Column Selection Chart
Mixed-Polarity
General Analytes
ZB-5msZB-MR-1ZB-XLB
Pesticides
ZB-MR-1ZB-MR-2ZB-5msZB-XLB
Polar
Semi-Volatile Volatile
ZB-624ZB-WAXPLUS
ZB-WAXNeutralBase Acidic
ZB-WAXZB-50ZB-35
ZB-XLBZB-MR-2
ZB-WAXZB-XLB
ZB-FFAPZB-XLB
ZB-MR-2
Non-Polar
Semi-VolatileVolatile
ZB-1*ZB-5*Plot
*Thick Film (≥1.00 µm)
NeutralActive (Acids/Bases)
ZB-5ZB-5msZB-1msZB-5MSi
ZB-1HT Inferno™
ZB-5HT Inferno™
ZB-XLB
ZB-5ZB-5msZB-5MSi
ZB-1HT Inferno™
ZB-5HT Inferno™
ZB-XLB
Your Analytes
PHENOMENEX | WEB: www.phenomenex.com
-
25
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
This section is, neither in terms of manufacturers nor in terms of their products, a complete list,
and the accuracy of the data is not guaranteed. Small differences in dimensions or performance
might be possible and slight adjustments to your application may be necessary.
Composition Phenomenex Restek J&W SupelcoAgilentTechnologies (HP) Alltech SGE
Varian(Chrompack) OV
100 % dimethyl-polysiloxane
ZEBRONZB-1
Rtx-1,Rtx-1PONA,Rtx-1 F&F
DB-1,DB-2887,DB-1 EVDX
SPB-1,SPB-1 TG, SE-30,MET-1,SPB-1 Sulfur,SPB-HAP
HP-1,HP-101,HP-PONA,Ultra 1
AT-1,AT-Sulfur,EC-1
BP1,BP1-PONA,BPX1-SimD,
CP-Sil 5 CB OV-1
100 % dimethyl-polysiloxane
ZEBRONZB-1ms
Rtx-1ms DB-1ms MDN-1Equity-1
HP-1ms AT-1ms SolGel-1ms CP-Sil 5 CB MSVF-1ms
100 % dimethyl-polysiloxane
ZEBRONZB-1HT Inferno
MXT-1 SimDist DB-1HT Petrocol 2887 CP-SimDist
5 %-phenyl-95 %-dimethylpolysiloxane
ZEBRONZB-5
Rtx-5 DB-5 MDN-5, SPB-5,PTE-5, SE-54, PTA-5, SAC-5, Equity-5
HP-5, Ultra 2,HP-PAS-5,
AT-5, EC-5
BP5,BPX5
CP-Sil 8 CB OV-5
5 %-phenyl-95 %-dimethylpolysiloxane
ZEBRONZB-5MSi
Rtx-5ms, Rtx-5Amine,Rxi-5ms
DB-5 MDN-5S HP-5ms, HP-5msi
5 %-phenyl-Arylene-95 %-dimethylpolysiloxane
ZEBRONZB-5ms
Rtx-5SilMS DB-5ms,DB-5.625,DB-5ms EVDX
VF-5ms,CP-Sil 8 CB MS
5 %-phenyl-95 %-dimethylpolysiloxane
ZEBRONZB-5HT Inferno
Stx-5HTXTI-5HT
DB-5HT HT-5 VF-5HT
35 %-phenyl-65 %-dimethylpolysiloxane
ZEBRONZB-35
Rtx-35Rtx-35ms
DB-35,DB-35ms
MDN-35, SPB-35, SPB-608
HP-35,HP-35ms
AT-35 BPX35,BPX608
OV-11
50 %-phenyl-50 %-dimethylpolysiloxane
ZEBRONZB-50
Rtx-50 DB-17,DB-17HT,DB-17ms,DB-17 EVDX
SP-2250,SPB-17,SPB-50
HP-50+ AT-50 BPX50 CP-Sil 24 CB
6 %-cyanopropyl-phenyl-94 %-dimethyl-polysiloxane
ZEBRONZB-624
Rtx-1301,Rtx-624
DB-1301,DB-624,DB-VRX
SPB-1301,SPB-624
HP-VOC AT-624,AT-1301
BP624 CP-1301CP-Select 624 CB
OV-624
14 %-cyanopropyl-phenyl-86 %-dimethylpolysiloxane
ZEBRONZB-1701
Rtx-1701 DB-1701 SPB-1701,Equity-1701
AT-1701 BP10 CP-Sil 19 CB OV-1701
14 %-cyanopropyl-phenyl-86 %-dimethylpolysiloxane
ZEBRONZB-1701P
DB-1701P
Polyethylene glycol ZEBRONZB-WAX
Rtx-WAX,Famewax,Stabilwax-DB
DB-WAXetr Met-Wax,Omegawax
HP-INNOWax EC-Wax SolGel-WAX CP-Wax 57 CB
Polyethylene glycol ZEBRONZB-WAXPLUS
Stabilwax DB-WAX, CAM
Supelcowax 10 HP-20MCarbowax-20M
AT-WaxAT-AquaWax
BP20 CP-Wax 52 CB Carbo-wax 20M
Nitroterephthalic acid modified polyethylene glycol
ZEBRONZB-FFAP
Stabilwax-DB DB-FFAP Nukol,SPB-1000
HP-FFAP AT-1000,EC-1000
BP21 CP-Wax 58 CB OV-351
Proprietary ZEBRONZB-Multi-Residue 1 (MR-1)
Rtx-CLPesticides,Stx-CLPesticides
Proprietary ZEBRONZB-Multi-Residue 2 (MR-2)
Rtx-CLPesticides2,Stx-CLPesticides2
Proprietary ZEBRONZB-XLB
Rtx-XLB DB-XLB MDN-12 VF-Xms
Zebron GC Columns by Manufacturer and Cross Reference
Exclusive Phenomenex
2007 Winner. Top 100 technologically significant products.
PHENOMENEX | WEB: www.phenomenex.com
-
26
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
GC Column Selectionby EPA SpecificationsListed below are a few of our recommended Zebron columns for EPA analyses. Other
possible columns can also be used for these analyses. Please contact Phenomenex for
your specific GC column needs.
EPA Drinking Water Test Methods
Method DescriptionPhenomenexRecommendation Page
501.3 Trihalomethanes by GC/MS and SIM ZB-624 40
502.2 Volatile Halogenated Organics in Water by Purge & Trap GC/PID/ELCD ZB-624 40
503.1 Volatile Aromatics & Unsaturated Organics by Purge & Trap GC ZB-624 40
504.1 1,2-Dibromoethane (EDB), 1,2-Dibromo-3-chloropropane (DBCP), and 1,2,3-Trichloropropane (123TCP) by GC/ECD
ZB-1, ZB-624 31, 40
505 Organohalide Pesticides & Aroclors by GC/ECD ZB-1, ZB-MR-1, ZB-MR-2 31, 50, 51
507 Nitrogen & Phosphorus Containing Pesticides in Water by GC/NPD ZB-MR-1, ZB-MR-2 50, 51
508 Chlorinated Pesticides in Water by GC/ECD ZB-MR-1, ZB-MR-2 50, 51
513 2, 3, 7, 8-Tetrachlorodibenzo-p-dioxin by GC w/ high resolution Mass Spec ZB-5ms 36, 37
515.2 Determination of Chlorinated Acids in Water Using Liquid-Solid Extraction & GC/ECD ZB-5ms, ZB-1701P 36, 37, 43
524.2 Measurement of Purgeable Organic Compounds in Water by Purge & Trap Capillary Column GC/MS
ZB-624 40
525 Organic Compounds in Drinking Waterby Liquid-Solid Extraction and Capillary Column GC/MS
ZB-5ms, ZB-5MSi 35, 36, 37
551.1 Chlorinated Solvents & Disinfection By-Productsin Drinking Water by Liquid-Liquid Extraction & GC/ECD
ZB-35 38
552.2 Haloacetic Acids & Dalapon in Drinking Waterby Ion-Exchange Liquid-Solid Extraction & GC/ECD
ZB-5ms, ZB-35, ZB-1701, ZB-MR-1, ZB-MR-2
36, 37, 38, 42, 50, 51
PHENOMENEX | WEB: www.phenomenex.com
-
27
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
EPA Waste Water Test Methods
Method DescriptionPhenomenexRecommendation Page
601 Purgeable Halocarbons ZB-624 40
602 Purgeable Aromatics ZB-624 40
603 Acrolein & Acrylonitrile ZB-624 40
604 Phenols ZB-5ms, ZB-5MSi 35, 36, 37
606 Phthalate Esters ZB-5ms, ZB-5MSi 35, 36, 37
607 Nitrosamines ZB-5ms, ZB-35, ZB-1701P, ZB-5MSi 35, 36, 37, 38, 43
608 Organochlorine Pesticides & PCBs ZB-MR-1, ZB-MR-2, ZB-XLB 44, 50, 51
609 Nitroaromatics & Isophorone ZB-5, ZB-5MSi 34, 35
610 Polynuclear Aromatic Hydrocarbons ZB-5ms 36, 37
611 Haloethers ZB-5ms, ZB-5MSi 35, 36, 37
612 Chlorinated Hydrocarbons ZB-5ms, ZB-5MSi 35, 36, 37
613 2, 3, 7, 8-Tetrachlorodibenzo-p-dioxin ZB-5ms 36, 37
615 Chlorinated Herbicides ZB-5, ZB-35, ZB-MR-1, ZB-MR-2, ZB-5ms, ZB-5MSi, ZB-XLB
34, 35, 36, 37, 38, 44, 50, 51
619 Triazine Herbicides ZB-5, ZB-50, ZB-MR-1, ZB-MR-2, ZB-5MSi, ZB-XLB
34, 35, 39, 44, 50, 51
622 Organophosphorus Pesticides ZB-5, ZB-35, ZB-MR-1, ZB-MR-2, ZB-5MSi, ZB-XLB
34, 35, 38, 44, 50, 51
624 Purgeable Volatiles ZB-624 40
625 Base/Neutral and Acids ZB-5ms, ZB-5MSi 35, 36, 37
1618 Organophosphorus Pesticides, OrganohalidePesticides, Phenoxyacid Herbicides
ZB-MR-1, ZB-MR-2, ZB-XLB 44, 50, 51
1624 Volatile Organic Compounds by Isotope Dilution GC/MS ZB-624 40
1625 Semivolatile Organic Compounds by Isotope Dilution GC/MS ZB-5ms 36, 37
1653 Chlorinated Phenols in Waste Water by In-Situ Acetylation and GC/MS ZB-5ms 36, 37
GC Column Selectionby EPA SpecificationsListed below are a few of our recommended Zebron columns for EPA analyses. Other
possible columns can also be used for these analyses. Please contact Phenomenex for
your specific GC column needs.
PHENOMENEX | WEB: www.phenomenex.com
-
28
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
Listed below are a few of our recommended Zebron columns for EPA analyses. Other
possible columns can also be used for these analyses. Please contact Phenomenex for
your specific GC column needs.
EPA Solid Waste Test Methods
Method DescriptionPhenomenexRecommendation Page
8010B Halogenated Volatile Organics ZB-624 40
8015B Nonhalogenated Volatile Organics ZB-5, ZB-624, ZB-5HT, ZB-5MSi
34, 35, 40, 48, 49
8021B Aromatic and Halogenated Volatiles ZB-624 40
8030A Acrolein, Acrylonitrile, Acetonitrile ZB-624 40
8041 Phenols ZB-5ms, ZB-5MSi 35, 36, 37
8061A Phthalate Esters ZB-5ms, ZB-5MSi 35, 36, 37
8081A Organochlorine Pesticides ZB-5, ZB-35, ZB-MR-1, ZB-MR-2, ZB-XLB
34, 38, 44, 50, 51
8082 Polychlorinated Biphenyls (PCBs) by GC ECD ZB-5, ZB-5MSi, ZB-MR-1, ZB-MR-2, ZB-XLB
34, 35, 44, 50, 51
8091 Nitroaromatics and Cyclic Ketones ZB-5ms 36, 37
8100 Polynuclear Aromatic Hydrocarbons ZB-5ms, ZB-5MSi 35, 36, 37
8121 Chlorinated Hydrocarbons ZB-5ms, ZB-5MSi, ZB-MR-1, ZB-MR-2, ZB-XLB
35, 36, 37, 44, 50, 51
8141A Organophosphorus Pesticides ZB-5, ZB-5MSi, ZB-MR-1, ZB-MR-2, ZB-XLB
34, 50, 51
8151A Chlorinated Herbicides ZB-5ms, ZB-35, ZB-MR-1, ZB-MR-2, ZB-XLB
36, 37, 38, 50, 51
8260B GC/MS for Volatile Organics: Capillary Column Techniques ZB-624 40
8270C GC/MS for Semivolatile Organics: Capillary Column Techniques ZB-5ms, ZB-5MSi 35, 36, 37
8280A Analysis of Polychlorinated Dibenzo-p-dioxins and Polychlorinated Dibenzofurans ZB-5, ZB-5ms, ZB-5MSi 34, 35, 36, 37
GC Column Selectionby EPA Specifications
PHENOMENEX | WEB: www.phenomenex.com
-
29
Intr
od
uct
ion
to
Co
lum
n S
elec
tio
nZ
ebro
n G
C C
olum
nsG
C A
cces
sori
esS
PE
Pro
duct
sA
pplic
atio
nsZ
ebro
n O
verv
iew
GC Column Selectionby ASTM SpecificationsListed below are a few of our recommended Zebron columns for ASTM methods. Other
possible columns can also be used for these analyses. Please contact Phenomenex for your
specific GC column needs.
ASTM Methods
Method DescriptionPhenomenexRecommendation Page
D 1945 Natural gas PLT-5A
D 1946 Reformed gas PLT-5A , ZB-1 31
D 1983 FAME analysis BPX70
D 2245 Oils and oil acids in solvent-reducible paints BPX70
D 2268 Analysis of n-heptane and iso-octane (high purity) ZB-1 31
D 2306-96
Xylene isomers ZB-WAX, ZB-WAXPLUS 45, 46, 47
D 2426 Butadiene and styrene in butadiene concentrates ZB-1 31
D 2504 Non-condensable gases in C1-C3 hydrocarbons PLT-5A
D 2580 Phenols in water ZB-5, ZB-5MSi 34, 35
D 2600 Aromatic traces in light saturated hydrocarbons ZB-WAX 45
D 2743 Oil and oil acids BPX70
D 2800 FAME analysis BPX70
D 2804 Purity of methyl ethyl ketone ZB-WAX 45
D 2887 SimDist analysis of petroleum fractions ZB-1 31
D 2908 Volatile organics in water ZB-624, ZB-WAXPLUS 40, 46, 47
D 2998 Volatile organics in water ZB-1 31
D 2999 Monopentaerythritol in commercial pentaerythritol ZB-1 31
D 3009 Composition of turpentine ZB-WAXPLUS 46, 47
D 3054 Purity and benzene content of cyclohexane ZB-1 31
D 3086 Organochlorine pesticides in water ZB-5, ZB-5MSi 34, 35
D 3168 Polymers in emulsion paints ZB-1 31
D 3271 Solvent analysis in paints ZB-WAXPLUS 46, 47
D 3304 PCBs in environmental materials ZB-5, ZB-5MSi 34, 35
D 3328 Comparison of waterborne petroleum oils ZB-1 31
D 3329 Purity of methyl isobutyl ketone ZB-WAXPLUS 46, 47
D 3416 Total hydrocarbons, methane and CO in air PLT-5A
D 3432 Toluene diisocyanates in urethane prepolymers ZB-1 31
D 3447 Purity of trichlorotrifluoroethane (CFC-113) ZB-1 31
D 3452 Identification of rubber ZB-1, ZB-1HT 31
Method DescriptionPhenomenexRecommendation Page
D 3457 FAME analysis BPX70
D 3465 Purity of monomeric plasticizers ZB-1 31
D 3524 Diesel fuel in lubricating oil (SAE 30) ZB-1, ZB-1HT 31
D 3534 PCBs in water ZB-5, ZB-5MSi 34, 35
D 3606 Benzene and toluene in gasoline ZB-1 31
D 3687 Volatile organic compounds ZB-WAXPLUS, ZB-WAX 45, 46, 47
D 3710 SimDist analysis of gasoline and gasoline fractions ZB-1, ZB-1HT 31
D 3725 Fatty acids in drying oils ZB-FFAP 41
D 3760 Analysis of cumene ZB-WAXPLUS, ZB-WAX 45, 46, 47
D 3797 Analysis of o-xylene ZB-WAXPLUS, ZB-WAX 45, 46, 47
D 3798 Analysis of p-xylene impurities ZB-WAXPLUS, ZB-WAX 45, 46, 47
D 3876 Methoxyl and hydroxypropyl substitution in cellulose ether products
ZB-1 31
D 3962 Impurities in styrene ZB-FFAP 41
D 4059 PCBs in insulating liquids ZB-5, ZB-5MSi 34, 35
D 4275 Butylated hydroxy toluene in ethylene and ethylenevin-ylacetate polymers
ZB-1 31
D 4367 Benzene in hydrocarbon solvent ZB-1 31
D 4420 Aromatics in gasoline ZB-1 31
D 4735 Thiophene impurities in benzene ZB-FFAP 41
D 4768 Phenol and cresol inhibitors in insulating oils ZB-FFAP 41
D 5060 Impurities in ethylbenzene ZB-WAXPLUS, ZB-WAX, ZB-FFAP
41, 45, 46, 47
D 5134 Petroleum naphthas through n-nonane ZB-1 31
D 5135-35
Analysis of styrene ZB-WAXPLUS, ZB-WAX 45, 46, 47
D 5580 Aromatics in finished gasoline ZB-1 31
D 6352 Extended SimDist ZB-1HT 48, 49
D 6584 Determination of glycerine in biodiesel ZB-5HT 48, 49
E 0202 Analysis of glycols ZB-WAXPLUS 46, 47
E 1100 Analysis of denaturated ethanol ZB-WAXPLUS 46, 47
PHENOMENEX | WEB: www.phenomenex.com
-
30
Intro
du
ction
to C
olu
mn
Selectio
nZ
ebron GC
Colum
nsG
C A
ccessoriesS
PE
Products
Applications
Zebron O
verview
Listed below are a few of our recommended Zebron columns for ASTM methods. Other
possible columns can also be used for these analyses. Please contact Phenomenex for your
specific GC column needs.
USP Column Classifications
GC Column Selectionby USP Specifications
Method DescriptionPhenomenexRecommendation Page
G1 Dimethylpolysiloxane oil ZB-1, ZB-1ms 31, 32, 33
G2 Dimethylpolysiloxane gum ZB-1, ZB-1ms 31, 32, 33
G3 50 % Phenyl-50 % methylpolysiloxane ZB-50 39
G5 3-Cyanopropylpolysiloxane BPS70
G7 50 % 3-Cyanopropyl-50 % phenylmethylsilicone 007-225
G14 Polyethylene glycol (average MW 950-1,050) ZB-WAX, ZB-WAXPLUS 45, 46, 47
G15 Polyethylene glycol (average MW 3,000-3,700) ZB-WAX, ZB-WAXPLUS 45, 46, 47
G16 Polyethylene glycol (average MW 15,000) ZB-WAX, ZB-WAXPLUS 45, 46, 47
G17 75 % Phenyl-25 % methylpolysiloxane ZB-50 39
G19 25 % Phenyl-25 % cyanopropylmethylsilicone 007-225
G20 Polyethylene glycol (average MW of 380-420) ZB-WAX, ZB-WAXPLUS 45, 46, 47
G25 Polyethylene glycol TPA (Carbowax 20M terephthalic acid) ZB-FFAP 41
G27 5 % Phenyl-95 % methylpolysiloxane ZB-5, ZB-5MSi 34, 35
G27 5 % Phenyl-Arylene-95 % methylpolysiloxane ZB-5ms 36, 37
G28 25 % Phenyl-75 % methylpolysiloxane ZB-35 38
G32 20 % Phenylmethyl-80 % dimethylpolysiloxane ZB-35 38
G35 Polyethylene glycol & diepoxide esterified with nitroterephthalic acid ZB-FFAP 41
G36 1 % Vinyl-5 % phenylmethylpolysiloxane ZB-5, ZB-5MSi 34, 35
G38 Phase G1 plus a tailing inhibitor ZB-1 31
G39 Polyethylene glycol (average MW 1,500) ZB-WAX, ZB-WAXPLUS 45, 46, 47
G41 Phenylmethyldimethylsilicone (10 % phenyl substituted) ZB-5, ZB-5MSi 34, 35
G42 35 % Phenyl-65 % dimethylpolysiloxane ZB-35 38
G43 6 % Cyanopropylphenyl-94 % dimethylpolysiloxane ZB-624 40
G46 14 % Cyanopropylphenyl-86 % methylpolysiloxane ZB-1701 42
PHENOMENEX | WEB: www.phenomenex.com
-
31
Intr
oduc
tion
to C
olum
n S
elec
tion
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
Zeb
ron
Ove
rvie
wZ
ebro
n G
C C
olu
mn
s
• Temperature Limits: -60 to 360/370 ˚C (Isothermal/TPGC)*
• Low polarity column
• Used for “fingerprinting” and routine quality control analyses (e.g., citrus oils)
• Equivalent to USP Phase G2
ZB-1 GC Columns
Ordering InformationID (mm) df (µm) Temp. Limits ˚C Part No.10-Meter0.53 2.65 -60 to 340/360 7CK-G001-35
15-Meter0.25 0.10 -60 to 360/370 7EG-G001-02
0.25 0.25 -60 to 360/370 7EG-G001-11
0.25 1.00 -60 to 340/360 7EG-G001-22
0.32 0.25 -60 to 360/370 7EM-G001-11
0.32 1.00 -60 to 340/360 7EM-G001-22
0.53 0.15 -60 to 360/370 7EK-G001-05
0.53 0.50 -60 to 360/370 7EK-G001-17
0.53 1.50 -60 to 340/360 7EK-G001-28
0.53 3.00 -60 to 340/360 7EK-G001-36
0.53 5.00 -60 to 340/360 7EK-G001-39
ID (mm) df (µm) Temp. Limits ˚C Part No.30-Meter0.25 0.10 -60 to 360/370 7HG-G001-02
0.25 0.25 -60 to 360/370 7HG-G001-11
0.25 0.50 -60 to 360/370 7HG-G001-17
0.25 1.00 -60 to 360/370 7HG-G001-22
0.32 0.25 -60 to 340/360 7HM-G001-11
0.32 0.50 -60 to 360/370 7HM-G001-17
0.32 1.00 -60 to 360/370 7HM-G001-22
0.32 5.00 -60 to 340/360 7HM-G001-39
0.53 0.10 -60 to 340/360 7HK-G001-02
0.53 0.50 -60 to 340/360 7HK-G001-17
0.53 1.50 -60 to 340/360 7HK-G001-28
0.53 3.00 -60 to 340/360 7HK-G001-36
0.53 5.00 -60 to 340/360 7HK-G001-39
ID (mm) df (µm) Temp. Limits ˚C Part No.50-Meter0.25 0.50 -60 to 360/370 7JG-G001-17
60-Meter0.25 0.10 -60 to 360/370 7KG-G001-02
0.25 0.25 -60 to 360/370 7KG-G001-11
0.25 1.00 -60 to 340/360 7KG-G001-22
0.32 0.10 -60 to 360/370 7KM-G001-02
0.32 0.25 -60 to 360/370 7KM-G001-11
0.32 1.00 -60 to 340/360 7KM-G001-22
0.32 3.00 -60 to 340/360 7KM-G001-36
0.53 1.50 -60 to 340/360 7KK-G001-28
100-Meter0.25 0.50 -60 to 360/370 7MG-G001-17
Test MixZebron ZB-1 AG0-5155
ApplicationsAmines Mercaptans Pesticides
Drugs of abuse MTBE Semi-volatiles
Essential oils Natural gas odorants Simulated distillation
Ethanol Oxygenates and GROs Solvent impurities
Gases (refinery) PCBs Sulfur compounds (light)
Hydrocarbons
App
ID 1
4943 Simulated Distillation Markers: ASTM Method 2887
Column: Zebron ZB-1
Dimensions: 10 meter x 0.53 mm x 2.65 µm
Part No.: 7CK-G001-35
Injection: Split 10:1 @ 360 ˚C, 1 µL
Carrier Gas: Helium @ 8.1 mL/min (constant flow)
Oven Program: 35 ˚C to 360 ˚C @ 15 ˚C/min for 5 min
Detector: FID @ 360 ˚C
Sample: Analytes are 0.1 % in carbon disulfide.
1. Pentane (C5)2. Hexane (C6)3. Heptane (C7)4. Octane (C8)5. Nonane (C9)6. Decane (C10)7. Undecane (C11)
8. Dodecane (C12)9. Tetradecane (C14)
10. Pentadecane (C15)11. Hexadecane (C16)12. Heptadecane (C17)13. Octadecane (C18)14. Eicosane (C20)
15. Tetracosane (C24)16. Octacosane (C28)17. Dotriacontane (C32)18. Hexatriacontane (C36)19. Tetracontane (C40)20. Tetratetracontane (C44)
App
ID 1
266 Hazardous Substances
Column: Zebron ZB-1
Dimensions: 30 meter x 0.25 mm x 0.25 μm
Part No.: 7HG-G001-11
Injection: Split 20:1 @ 250 ˚C
Carrier Gas: Helium @ 0.88 mL/min (constant flow)
Oven Program: 50 ˚C to 200 ˚C @ 6 ˚C/min
Detector: FID @ 300 ˚C
Sample: 1. Aniline2. Benzyl alcohol3. 2-Methylphenol4. 4-Methylphenol5. Benzoic acid6. 4-Chloroaniline
7. 2-Methylnaphthalene8. 2,4,5-Trichlorophenol9. 2-Nitroaniline
10. 3-Nitroaniline11. Dibenzofuran12. 4-Nitroaniline
0 11 22 min
12
11
108
7
6
43
2
1
5
9
If you need a 5 in. cage, simply add a (-B) after the part number,e.g., 7HG-G001-11-B.
*Thicker films (≥1.0 µm df) are rated to 340/360 ˚C (Isothermal/TPGC).
Alternative to Any 100 % Dimethylpolysiloxane Phase:DB-1 Rtx-1 Rtx-1PONA BP1 SE-30 DB-1 EVDX HP-PONA
HP-1 Rtx-1 F&F SPB-1 007-1 AT-1 DB-2887 HP-101
MET-1 CP-Sil 5 CB OV-1 Ultra 1
Polarity
Bleed
Temperature Limits
Stability
Low High
Column Profile
CERTIFIED
MSESCEQUIPPED
PROTECT YOUR GC COLUMN. TRY Z-GUARD WITH YOUR NEXT ZEBRON ORDER.
0 5 10 15 20 25 min
1
2
3
4 56 7 8
9 1011 1213 1415
1617
18
19
20
PHENOMENEX | WEB: www.phenomenex.com
-
32
Introduction to Colum
n Selection
GC
Accessories
SP
E P
roductsA
pplicationsZ
ebron Overview
Zeb
ron
GC
Co
lum
ns
Lower Overall Column Activity
Activity is a key measure of column quality. This is why Zebron™ ZB-1ms columns are
aggressively tested to ensure full deactivation. Below is an example of the aggressive
QC test mix we use, notice the low tailing for even the most active compounds like 2-
Ethylhexanoic Acid!
Enhanced ESC Bonding Technology
After years of research, we have been able to improve upon our original technology! Using
advanced polymer fractionation and curing techniques, we have been able to lower our
bleed specification by 50 % and still maintain the low activity surface you have come to
depend on! The new ZB-1ms is truly the only 100 % Dimethylpolysiloxane column you will
ever need!
• Temperature Limits: -60 to 360/370 ˚C (Isothermal/TPGC)
• Lowered bleed (MS Certified) especially suited to highsensitivity GC/MS
• Extremely inert for active compounds such as drugs or pesticides
• Improved signal-to-noise ratio for better sensitivity andmass spectral integrity
• Identical selectivity to the ZB-1
• Available with Guardian Integrated Guard Columns
• Equivalent to USP Phase G2
ZB-1ms GC Columns
ApplicationsAmines Diesel Fuel Pesticides
Acids Drugs of Abuse Flavors and Fragrances
Polychlorinated Biphenyls (EPA Method 1668)
Column: Zebron ZB-1ms
Dimensions: 30 meter x 0.25 mm x 0.25 μm
Part No.: 7HG-G011-11
Injection: Split Flow 100 mL/min @ 250 ˚C, 1.0 µL
Carrier Gas: Hydrogen @ 1.18 mL/min (constant flow)
Oven Program: 140 ˚C (Isothermal)
Detector: FID @ 325 ˚C
Sample: 1. Decane2. 2-Ethylhexanoic Acid3. 4-Chlorophenol4. Naphthalene
5. Tridecane 6. 1-Undecanol
7. Dicyclohexylamine 8. Pentadecane
Alternative to Any MS-Certified 100 % Dimethylpolysiloxane Phase:DB-1ms HP-1ms Rtx-1MS VF-1ms Equity-1 MDN-1 AT-1ms
SOLGEL-1ms CP-Sil 5 CB ms
360 ºC
200000
400000
5 10 15 20 25 30 35 40 min
The shaded area depicts the bleed criteria for MS certified columns on a MS detector. MS bleed certification values are typically read at 320 ºC. This demonstrates the low bleed capabilities of the ZB-1ms. It meets MS certification limits even at 360 ºC!
320 ºC
Column: Zebron ZB-1ms
Dimensions: 30 meter x 0.25 mm x 0.25 μm
Part No.: 7HG-G011-11
Injection: Null injection @ 250 ˚C
Carrier Gas: Helium @ 1.0 mL/min (constant flow)
Oven Program: 40 ˚C to 360 ˚C @ 10 ˚C/min hold for 10 min
Detector: MSD @ 40-500 amu
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
1
1
2
35
4
6 7
8
5 10 min
App
ID 1
5545
CERTIFIED
MSESCEQUIPPED
PHENOMENEX | WEB: www.phenomenex.com X
-
33
Intr
oduc
tion
to C
olum
n S
elec
tion
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
Zeb
ron
Ove
rvie
wZ
ebro
n G
C C
olu
mn
s
Fast GC Narrow Bore Columns
In today’s fast paced world of chemical analyses, the demand to produce results more quickly
is greater than ever. The analysis of orange oil below demonstrates the ability of Fast GC
columns to dramatically decrease analysis times. For important considerations on how to
convert an existing method to Fast GC, please contact your Phenomenex representative.
10 15 20 25 30 35 40 min
1 2 3,4
56
7
8
9 10
11
12
13
14
15
16
17
18
19
20
2122
Ordering InformationID (mm) df (µm) Temp. Limits ˚C Part No.10-Meter0.10 0.10 -60 to 360/370 7CB-G011-02
12-Meter0.20 0.33 -60 to 360/370 7DE-G011-14
15-Meter0.25 0.25 -60 to 360/370 7EG-G011-11
0.32 0.25 -60 to 360/370 7EM-G011-11
20-Meter0.18 0.18 -60 to 360/370 7FD-G011-08
25-Meter0.20 0.33 -60 to 360/370 7GE-G011-14
ID (mm) df (µm) Temp. Limits ˚C Part No.30-Meter0.25 0.10 -60 to 360/370 7HG-G011-02
0.25 0.25 -60 to 360/370 7HG-G011-11
0.25 0.50 -60 to 360/370 7HG-G011-17
0.25 1.00 -60 to 360/370 7HG-G011-22
0.32 0.25 -60 to 360/370 7HM-G011-11
0.32 1.00 -60 to 360/370 7HM-G011-22
ID (mm) df (µm) Temp. Limits ˚C Part No.60-Meter0.25 0.25 -60 to 360/370 7KG-G011-11
0.25 1.00 -60 to 360/370 7KG-G011-22
0.32 1.00 -60 to 360/370 7KM-G011-22
Test MixZebron ZB-1ms AG0-7805
Orange OilColumn: Zebron ZB-1ms
Dimensions: 60 meter x 0.25 mm x 0.25 μmPart No.: 7KG-G011-11
Injection: Split 75:1 @ 275 ˚C, 0.2 µLCarrier Gas: Helium @ 0.8 mL/min (constant flow)
Oven Program: 75 ˚C for 4 min to 250 ˚C @ 4 ˚C/min for 5 minDetector: MSD @ 20-350 amuSample: 1. -Pinene
2. -Phellandrine 3. -Myrcene 4. Octanal 5. 3-Carene 6. Limonene 7. Nonanal 8. Linalool 9. cis-Limonene Oxide 10. trans-Limonene Oxide 11. Citronellal
12. -Terpineol 13. Decanal 14. Neral 15. Carvone 16. Gerenial 17. -Cubenene 18. Dodecanal 19. -Cubenene 20. Valencene 21. Cadinine 22. Nootkatone
40 min
1 2 3,4 6
5
7
8
9
10
11
12
13
14
15
16
17
1819
20
21 22
3 5 7 9 11 13 15 min
FAST15 min
Orange OilColumn: Zebron ZB-1ms
Dimensions: 20 meter x 0.18 mm x 0.18 μmPart No.: 7FD-G011-08
Injection: Split 100:1 @ 180 ˚C, 1µLCarrier Gas: Helium @ 1.0 mL/min (constant flow)
Oven Program: 50 ˚C to 100 ˚C @ 6 ˚C/min to 275 ˚C @ 18 ˚C/min for 4 min
Detector: MSD @ 30-350 amuSample: 1. -Pinene
2. -Phellandrine 3. -Myrcene 4. Octanal 5. 3-Carene 6. Limonene 7. Nonanal 8. Linalool 9. cis-Limonene Oxide 10. trans-Limonene Oxide 11. Citronellal
12. -Terpineol 13. Decanal 14. Neral 15. Carvone 16. Gerenial 17. -Cubenene 18. Dodecanal 19. -Cubenene 20. Valencene 21. Cadinine 22. Nootkatone
App
ID 1
5580
App
ID 1
5571
If you need a 5 in. cage, simply add a (-B) after the part number,e.g., 7HG-G011-11-B.
PROTECT YOUR GC COLUMN. TRY Z-GUARD WITH YOUR NEXT ZEBRON ORDER.
PHENOMENEX | WEB: www.phenomenex.com
-
34
Introduction to Colum
n Selection
GC
Accessories
SP
E P
roductsA
pplicationsZ
ebron Overview
Zeb
ron
GC
Co
lum
ns
• Temperature Limits: -60 to 360/370 ˚C (Isothermal/TPGC)*
• Versatile low polarity column
• Low bleed (MS Certified) especially suited to high sensitivity work using GC/MS
• Extremely inert for active compounds such as drugs or pesticides
• Resilient to dirty samples - long column life
• Great column for unknown samples
• Equivalent to USP Phase G27
ZB-5 GC Columns
Ordering Information
App
ID 1
4876 Phthalate Esters: EPA Method 606
Column: Zebron ZB-5
Dimensions: 30 meter x 0.53 mm x 1.50 μm
Part No.: 7HK-G002-28
Injection: Split 20:1 @ 300 ˚C, 1 µL
Carrier Gas: Helium @ 12.9 mL/min (constant flow)
Oven Program: 40 ˚C for 6 min to 300 ˚C @ 10 ˚C/min for 15 min
Detector: FID @ 300 ˚C
Sample: 1. Dimethyl Phthalate2. Diethyl Phthalate3. Di-n-butyl Phthalate
4. Butyl Benzyl Phthalate5. Bis(2-ethylhexyl) Phthalate6. Di-n-octyl Phthalate
ID (mm) df (µm) Temp. Limits ˚C Part No.15-Meter0.25 0.10 -60 to 360/370 7EG-G002-02
0.25 0.25 -60 to 360/370 7EG-G002-11
0.25 0.50 -60 to 360/370 7EG-G002-17
0.25 1.00 -60 to 340/360 7EG-G002-22
0.32 0.10 -60 to 360/370 7EM-G002-02
0.32 0.25 -60 to 360/370 7EM-G002-11
0.32 1.00 -60 to 340/360 7EM-G002-22
0.53 0.50 -60 to 360/370 7EK-G002-17
0.53 1.50 -60 to 340/360 7EK-G002-28
0.53 3.00 -60 to 340/360 7EK-G002-36
20-Meter0.18 0.18 -60 to 360/370 7FD-G002-08
ID (mm) df (µm) Temp. Limits ˚C Part No.30-Meter0.25 0.10 -60 to 360/370 7HG-G002-02
0.25 0.25 -60 to 360/370 7HG-G002-11
0.25 0.50 -60 to 360/370 7HG-G002-17
0.25 1.00 -60 to 340/360 7HG-G002-22
0.32 0.10 -60 to 360/370 7HM-G002-02
0.32 0.25 -60 to 360/370 7HM-G002-11
0.32 0.50 -60 to 360/370 7HM-G002-17
0.32 1.00 -60 to 340/360 7HM-G002-22
0.53 0.50 -60 to 360/370 7HK-G002-17
0.53 1.50 -60 to 340/360 7HK-G002-28
0.53 3.00 -60 to 340/360 7HK-G002-36
0.53 5.00 -60 to 340/360 7HK-G002-39
ID (mm) df (µm) Temp. Limits ˚C Part No.60-Meter0.25 0.10 -60 to 360/370 7KG-G002-02
0.25 0.25 -60 to 360/370 7KG-G002-11
0.25 0.50 -60 to 360/370 7KG-G002-17
0.25 1.00 -60 to 340/360 7KG-G002-22
0.32 0.25 -60 to 360/370 7KM-G002-11
0.32 1.00 -60 to 340/360 7KM-G002-22
0.53 1.50 -60 to 340/360 7KK-G002-28
Test MixZebron ZB-5 AG0-5155
0
1 23 4
5 6
10 20 30 40 min
ApplicationsAlkaloids FAMEs Phenols
Dioxins Halo-hydrocarbons Residual Solvents
Drugs PCBs/Aroclors Semi-volatiles
Essential Oils/Flavors Pesticides/Herbicides
Alternative to Any 5 %-Phenyl- 95 %-Dimethylpolysiloxane Phase:DB-5 HP-PAS-5 SPB-5 007-5 SE-54 EC-5 Ultra 2
HP-5 MDN-5 OV-5 AT-5 BP5 HP-101 CP-SIL 8 CB
Rtx-5 Equity-5 MDN-5 AT-5 BPX5 HP-5 Trace Analysis
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
If you need a 5 in. cage, simply add a (-B) after the part number,e.g., 7HG-G002-11-B.
*Thicker films (≥1.0 µm df) are rated to 340/360 ˚C (Isothermal/TPGC).
CERTIFIED
MSESCEQUIPPED
PROTECT YOUR GC COLUMN. TRY Z-GUARD WITH YOUR NEXT ZEBRON ORDER.
PHENOMENEX | WEB: www.phenomenex.com
-
35
Intr
oduc
tion
to C
olum
n S
elec
tion
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
Zeb
ron
Ove
rvie
wZ
ebro
n G
C C
olu
mn
s
• Temperature Limits: -60 to 360/370 ˚C (Isothermal/TPGC)
• Highly inert - improved peak shape for acidic/basic compounds
• MS certified low-bleed levels provide maximum sensitivity
• Industry leading QC specifications ensure column-to- column performance
• ESC bonding results in phase stability and high temperature limits
• Traditional bonding chemistry provides the same selectivity as the ZB-5 columns
ZB-5MSi GC Columns
Ordering Information
ApplicationsDrugs of Abuse FAMEs Pesticides
Nitrosamines Phenols EPA Methods
App
ID 1
6410 Phenols
Column: Zebron ZB-5MSi
Dimensions: 30 meters x 0.25 mm x 0.25 μm
Part No.: 7HG-G018-11
Injection: Split 5:1@ 240 ˚C, 1 μL
Carrier Gas: Helium @ 1.2 mL/min (constant flow)
Oven Program: 60 ˚C to 140 ˚C @ 5 ˚C/min to 280 ˚C @ 10 ˚C/min
Detector: MSD @ 230 ˚C; 45-450 amu
Sample: 1. Phenol2. 2-Chlorophenol3. 2-Methylphenol4. 4-Methylphenol5. 3-Methylphenol6. 2,6-Dimethylphenol7. 2-Nitrophenol8. 2-Ethylphenol9. 2,4-Dimethylphenol
10. 3,5-Dimethylphenol11. 2,5-Dimethylphenol12. 4-Ethylphenol13. 3-Ethylphenol14. 2,4-Dichlorophenol
15. Benzoic Acid16. 2,3-Dimethylphenol17. 3,4-Dimethylphenol18. 2.6-Dichlorophenol19. 4-Chloro-3-methylphenol20. 2,4,6-Trichlorophenol21. 2,4,5-Trichlorophenol22. 2,4-Dinitrophenol23. 4-Nitrophenol24. 2,3,4,6-Tetrachlorophenol25. 4,6-Dinitro-2-methylphenol26. Pentachlorophenol27. Dinoseb
Alternative to Any 5 %-Phenyl- 95 %-Dimethylpolysiloxane Phase:DB-5 Rtx-5ms MDN-5S HP-5ms Rtx-5Amine HP-5msi Rxi-5ms
ID (mm) df (µm) Temp. Limits ˚C Part No.15-Meter0.25 0.25 -60 to 360/370 7EG-G018-11
ID (mm) df (µm) Temp. Limits ˚C Part No.30-Meter0.25 0.25 -60 to 360/370 7HG-G018-11
0.25 0.50 -60 to 360/370 7HG-G018-17
0.32 0.25 -60 to 360/370 7HM-G018-11
0.32 0.50 -60 to 360/370 7HM-G018-17
ID (mm) df (µm) Temp. Limits ˚C Part No.60-Meter0.25 0.25 -60 to 360/370 7KG-G018-11
Test MixZebron ZB-5MSi AG0-8362
Column Profile
Low High
6 8 10 12 14 16 18 20 22 min
1 23
4,5
67
812
9,10,11
13,14,15
161718 19 20 21
2223
24
25
2627
If you need a 5 in. cage, simply add a (-B) after the part number,e.g., 7HG-G018-11-B.
CERTIFIED
MSESCEQUIPPED
PROTECT YOUR GC COLUMN. TRY Z-GUARD WITH YOUR NEXT ZEBRON ORDER.
Polarity
Bleed
Temperature Limits
Stability
PHENOMENEX | WEB: www.phenomenex.com
-
36
Introduction to Colum
n Selection
GC
Accessories
SP
E P
roductsA
pplicationsZ
ebron Overview
Zeb
ron
GC
Co
lum
ns
• Temperature Limits: -60 to 325/350 ˚C (Isothermal/TPGC)
• Arylene Matrix Technology™ (AMT)
• Fully conditioned within 35 minutes
• High response for acids and bases = Very low activity
• Enhanced resolution of Polyaromatic Hydrocarbons (PAHs) and other multi-ring aromatic compounds
• The perfect choice for EPA Methods 525, 610, 625, 8100, and 8270
• Equivalent to USP Phase G27
Arylene Matrix Technology™
The Zebron ZB-5ms is the accumulation of more than 15 years of
GC column manufacturing experience! We have applied our propri-
etary Engineered Self Cross-linking™ (ESC) bonding to an Arylene
polymer chemistry and created the new Arylene Matrix Technology.
The resulting columns have long lifetime, enhanced selectivity and
lower bleed than traditional Arylene products.
Low Bleed / Fast Conditioning
Achieves MS-Certified bleed level after only 35 minutes! Little to no
change in bleed after an additional 30 minutes of conditioning!
ZB-5ms GC Columns
10 20 30 40 50 60min
0.006
0.008
0.010
0.012
0.014
0.016
Volts
Passing Bleed Level
MS-Certified Bleedafter only 35min!
Long Lifetime
Consistent response after more than 700 samples at pH 2!
Reproducible Results
0.5
1.5
0 140 280 420 560 700Sample No.
Peak Skew
2,4-DichlorophenolHexachlorobutadiene
2.0
1
23 4 5
6
7 89 10
3.0 4.0 5.0 6.0 7.0 8.0 minLot 1Lot 2Lot 3Lot 4
ApplicationsAcids EPA Methods Pesticides/Herbicides
Alkaloids Essential Oils/Flavors Phenols
Amines FAMEs Residual Solvents
Dioxines Halo-hydrocarbons Semi-volatiles
Drugs PCBs/Aroclors Solvent Impurities
Alternative to Any MS-Certified 5 %-Phenyl-Arylene- 95 %-Dimethylpolysiloxane Phase:DB-5ms DB-5ms EVDX VF-5ms CP-Sil 8 CB MS DB-5.625
Column: Zebron ZB-5ms
Dimensions: 30 meter x 0.25 mm x 0.25 μm
Part No.: 7HG-G010-11
GC: HP 5890
Injection: Split 1:100 @ 250 ˚C, 1.4 µL
Carrier Gas: Hydrogen @ 140 ˚C, 40 cm/sec
Oven Program: 140 ˚C (isothermal)
Detector: FID @ 325 ˚C
Sample: 1. Decane2. 2-Ethylhexanoic Acid3. 1,6-Hexanediol4. 4-Chlorophenol5. Tridecane
6. 1-Methylnaphthalene7. 1-Undecanol8. Tetradecane9. Dicyclohexylamine
10. Pentadecane
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
CERTIFIED
MSESCEQUIPPED
PHENOMENEX | WEB: www.phenomenex.com X
-
37
Intr
oduc
tion
to C
olum
n S
elec
tion
GC
Acc
esso
ries
SP
E P
rodu
cts
App
licat
ions
Zeb
ron
Ove
rvie
wZ
ebro
n G
C C
olu
mn
s
App
ID 1
4442
App
ID 1
4946
Ordering InformationID (mm) df (µm) Temp. Limits ˚C Part No.10-Meter0.10 0.10 -60 to 325/350 7CB-G010-02
15-Meter0.25 0.25 -60 to 325/350 7EG-G010-11
20-Meter0.18 0.18 -60 to 325/350 7FD-G010-08
0.18 0.32 -60 to 325/350 7FD-G010-51
0.18 0.36 -60 to 325/350 7FD-G010-53
ID (mm) df (µm) Temp. Limits ˚C Part No.30-Meter0.25 0.25 -60 to 325/350 7HG-G010-11
0.25 0.50 -60 to 325/350 7HG-G010-17
0.25 1.00 -60 to 325/350 7HG-G010-22
0.32 0.25 -60 to 325/350 7HM-G010-11
0.32 0.50 -60 to 325/350 7HM-G010-17
0.32 1.00 -60 to 325/350 7HM-G010-22
ID (mm) df (µm) Temp. Limits ˚C Part No.60-Meter0.25 0.25 -60 to 325/350 7KG-G010-11
0.32 0.25 -60 to 325/350 7KM-G010-11
Test MixZebron ZB-5ms AG0-7578
Optimized Run Times
The added resolution offered by the Arylene Matrix Technology™ allows run times to be shortened by at least 20-30 %.
33