ge healthcare project management a-1/media/downloads/us/support/...project number date drawn by...

3
GE Healthcare has made reasonable efforts to ensure this BIM file conforms to requirements of the applicable pre-installation manual. This BIM file was produced to assist with site planning and is not intended for construction or final building modeling. GE Healthcare is not responsible for any damages resulting from use of such file for construction or final building modeling purposes. Compliance with all applicable local, state, and federal laws and regulations is the responsibility of the customer or its designated architects and engineers. Please keep in mind the information in this BIM file is dynamic and is subject to change without notice. It is the responsibility of the customer or its designated architect(s) and engineer(s) to ensure they have the latest equipment or product information. Scale Project number Date Drawn by Checked by GE Healthcare Project Management 12" = 1'-0" A-1 OPTIMA CT 680 - RENDERING 6-112 25.SEP.17 GENPACT

Upload: others

Post on 05-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: GE Healthcare Project Management A-1/media/downloads/us/support/...Project number Date Drawn by Checked by GE Healthcare Project Management 1/8" = 1'-0" A-2 OPTIMA CT 680 PLAN & SCHEDULE

GE Healthcare has made reasonable efforts to ensure this BIM file conformsto requirements of the applicable pre-installation manual. This BIM file wasproduced to assist with site planning and is not intended for construction orfinal building modeling. GE Healthcare is not responsible for any damagesresulting from use of such file for construction or final building modelingpurposes. Compliance with all applicable local, state, and federal laws andregulations is the responsibility of the customer or its designated architectsand engineers. Please keep in mind the information in this BIM file isdynamic and is subject to change without notice. It is the responsibility ofthe customer or its designated architect(s) and engineer(s) to ensure theyhave the latest equipment or product information. Scale

Project numberDateDrawn byChecked by

GE HealthcareProject Management

12" = 1'-0"

A-1

OPTIMA CT 680 - RENDERING

6-11225.SEP.17GENPACT

Page 2: GE Healthcare Project Management A-1/media/downloads/us/support/...Project number Date Drawn by Checked by GE Healthcare Project Management 1/8" = 1'-0" A-2 OPTIMA CT 680 PLAN & SCHEDULE

[6.3

0 m

]

20' -

8"

[2.7

4 m

]

9' -

0"

[4.01 m]

13' - 2"

GE Healthcare has made reasonable efforts to ensure this BIM file conformsto requirements of the applicable pre-installation manual. This BIM file wasproduced to assist with site planning and is not intended for construction orfinal building modeling. GE Healthcare is not responsible for any damagesresulting from use of such file for construction or final building modelingpurposes. Compliance with all applicable local, state, and federal laws andregulations is the responsibility of the customer or its designated architectsand engineers. Please keep in mind the information in this BIM file isdynamic and is subject to change without notice. It is the responsibility ofthe customer or its designated architect(s) and engineer(s) to ensure theyhave the latest equipment or product information. Scale

Project numberDateDrawn byChecked by

GE HealthcareProject Management

1/8" = 1'-0"

A-2

OPTIMA CT 680 PLAN & SCHEDULE

6-11225.SEP.17GENPACT

Specialty Equipment Schedule

Description Weight (lbs) BTUCONTROL ROOMNARROW OPERATORS CONSOLE 313 3201EXAM ROOMPOWER DISTRIBUTION UNIT 740 5119OPTIMA CT 680 GANTRY 3990 18703CT TABLE 1543 1023MAIN DISCONNECT CONTROL 90 0INJECTOR HEAD ON OVERHEADCOUNTERPOISED SUSPENSION

79 0

MONITOR ONOVERHEADCOUNTERPOISEDSUSPENSION

61 300

STORAGE CABINET 99 0

Page 3: GE Healthcare Project Management A-1/media/downloads/us/support/...Project number Date Drawn by Checked by GE Healthcare Project Management 1/8" = 1'-0" A-2 OPTIMA CT 680 PLAN & SCHEDULE

Hold the Left Click for Rotating aroundUse the Scroll for Zooming in and out

Click for Site Planning Support

6­112­Optima 680