GenieTouch Infusion/Withdrawal Dual Syringe Pump Owner’s Manual

Upload: others

Post on 12-Jun-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

GenieTouch™

Infusion/Withdrawal Dual Syringe Pump

Owner’s Manual

Page 2: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

TableofContents

AbouttheGenieTouch PartsoftheGenieTouch—1 SettingUptheGenieTouch—3 MountingSyringes—3

ControllingtheGenieTouch TheTouchScreen—6

GenieTouchHelp—7 TheGenieTouchMenus—7 Setup—7 Syringe—7 Operation—10 AutoReverse—14 Advanced—15 Main—18

GenieTouchPCInterface—19

GeneralInformationSpecifications—27CleaningInstructions—27Warranty—28DecontaminationVerificationForm—29

Thisproductisnotdesigned,intendedorauthorizedforuseinhumanapplications.©2016KentScientificCorporation,DocumentationVersion3.01,ForHelpVersion3.1andFirmwareRev1.03GenieTouchisatrademarkofKentScientificCorporation.

Page 3: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

1

AbouttheGenieTouchTheGenieTouchTMisaninfusion/withdrawaldual-syringepumpwithmulti-syringecapabilityandaseriesofuser-configurablefunctionsthatallowittomeetavarietyofneeds.Itsupportsoptionalaudiblealarmsandoptionalforceadjustment.TheGenieTouchisnotdesigned,intendedorauthorizedforuseinhumanapplications.

PartsoftheGenieTouchItisusefultobecomefamiliarwiththepartsoftheGenieTouchbeforeyoubeginusingit.TheGenieTouchsupportsuseoftwosyringesatatime,bothfacingthesamedirection.Therightsyringemountandtheleftsyringemounteachaccepttwosyringesofupto60mlsize.Note:Theleftandrightsyringeplungerreversingclampsonthepusherblockareusedtograbtheflangeontheendofthesyringeplunger.Withthesyringeplungerfixedtothepusherblock,theplungercanbemovedinbothforward(infusingoremptying)andreverse(withdrawingorfilling)directions.Theseclampsarealsocalledanti-siphonclampsbecausetheykeepthesyringeplungerfrompassivelymovingunderlargehydrostaticloads.

Page 4: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

2

RearviewTheRJ-45networkingconnectorsareforfutureGenieTouchmodels.

ThisistheU.S.powersupply.YoursmaylookdifferentifweshippedyourGenieTouchoutsidetheU.S.

Figure 1b

Powerswitch

PoweradapterentryExternalcontrol/remoteconnector(AUX)USBconnector

RJ-45networkingconnectors

Page 5: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

3

SettingUptheGenieTouchTheonlyphysicalsetuptheGenieTouchrequiresisthatyouplugitinandmountyoursyringes.

1. Connectthesupplied12VDCpoweradaptertotheGenieTouchthroughthepowerentryconnectorontherearofthechassis.(SeeFigures1band1c.)

2. Plugtheadapterintoanappropriatepowersource.

3. Slidethepowerswitchon.YouwillseetheOpening screen,whichcontainstwobuttons—SetupandMain.TappingSetupleadstoaseriesofscreensforconfiguringyourGenieTouchtomeetyourspecificneeds.TappingMainleadstotheMainscreenforcontrollingtheGenieTouchwhileitisrunning.

MountingSyringesTherearetwosyringemountsontherightsideoftheGenieTouchchassisforthetraditionalindustry-standardright-facingsyringeplacement.Therearealsotwosyringemountsontheleftsideofthechassisifyouprefertouseleft-facingsyringes.TheRightSideTomountasyringeinoneoftheright-sidemounts:

1.Selecttheproperclamporientation.ThesyringebarrelclampsareL-shapedandworkforbothsmallandlargesyringesdependingupontheorientationoftheclamp:

Forsmallersyringes£ PlacethescrewintheholeinthebottomoftheshortlegoftheL.£ Placethespringovertheprotrudingendofthescrew.£ LaythelongerlegofLinthenotchinthesideofthechassis.£ Tightenthescrewintothenotchinthechassis.

Forlargersyringes£ PlacethescrewintheholeinthetopoftheshortlegoftheL.£ Placethespringovertheprotrudingendofthescrew.£ Insertthebottomoftheshorterlegintothenotchinthesideof

thechassis.£ Tightenthescrewintothenotchinthechassis.

Thespringallowstheclamptoriseautomaticallywhenyouloosenit,makingsyringeremovalmoreconvenient.

Page 6: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

4

2.Loosenthescrewsonthetwosyringebarrelreversingclamps.Boththerightsyringemountandthepusherblockhaveblackreversingclampswithgrayscrewsatthetop.(SeeFigure1a.)Loosenallfourscrews.Ifthereversingclampofthepusherblockisontheleftfaceoftheblock,removethegrayscrewsandreplacetheclamplooselyontherightfaceoftheblock.

3.Placethesyringe.Laythesyringeinthetroughfacingoutofthechassiswiththebarrelflangebetweenthechassisanditsreversingclampandtheplungerflangebetweenthepusherblockanditsreversingclamp.

4.Tightenthescrews.Youmayhavetotightenallthescrewspartwaydownandthenreturntoretightenthem.Holdingthesyringeinplace,screwthesyringebarrelclampdownloosely.Tightenthescrewspart-wayonthereversingclamps.Wheneverythingislooselyheldinplace,squeezeeachreversingclampagainsttheflangeitisholdingandtightentofingertight.Thensqueezetheothersideofeachreversingclampandtightenit.Finally,returntothesyringeclampandmakeitfingertightaswell.

TheLeftSideTomountasyringeinoneoftheleft-sidemounts:

1.Selecttheproperclamporientation.Therearetwosyringebarrelclamps,astraightclampandanM-shapedclamp.Yourchoiceofclampwilldependuponthesizeofyoursyringe:

Forthesmallestsyringes–UsetheMclampwiththepointedsectionsfacingdown.Securetheclampwiththecentralscrew.Formid-sizesyringes–Usethestraightclamp.Securetheclampwiththecentralscrew.

Forlargersyringes–UsetheMclampwiththepointedsectionsfacingup.Securetheclampwiththecentralscrew.

Usethespringinthesamewayyouuseitforright-sidemounts.

Page 7: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

5

2.Loosenthescrewsonthetwosyringebarrelreversingclamps.Boththerightsyringemountandthepusherblockhaveblackreversingclampswithgrayscrewsatthetop.(SeeFigure1c.)Loosenallfourscrews.Ifthereversingclampofthepusherblockisontheleftfaceoftheblock,removethegrayscrewsandreplacetheclamplooselyontherightfaceoftheblock.

3.Placethesyringe.Laythesyringeinthetroughfacingoutofthechassis,withthebarrelflangebetweenthechassisanditsreversingclamp,andtheplungerflangebetweenthepusherblockanditsreversingclamp.

4.Tightenthescrews.Youmayhavetotightenallthescrewspartwaydownandthenreturntoretightenthem.Holdingthesyringeinplace,screwthesyringebarrelclampdownloosely.Tightenthescrewspart-wayonthereversingclamps.Wheneverythingislooselyheldinplace,squeezeeachreversingclampagainsttheflangeitisholdingandtightentofingertight.Thensqueezetheothersideofeachreversingclampandtightenit.Finally,returntothesyringeclampandmakeitfingertight.

NOTE:Thefirsttimeyouuseanewsyringesizeorbrandmodel,youwillneedto:

£ mountthesyringe£ calibrateit£ removeit£ fillit£ mountitagain

Whenyousubsequentlyusethesamesyringe,youcansimplyselectthesyringefromtheGenieTouchsavedlook-uptableofsyringes,andusethe<<or>>buttonstopositionthepusherblockcorrectlybeforeyoumountthesyringe.Seepage18.

Page 8: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

6

ControllingtheGenieTouch

TheTouchScreenTheGenieTouchmenuscreensaretouch-sensitive.Herearesometipsforusingthescreens:

£ Whentwochoicesappearsidebysideonthesamebutton,onlyoneortheothercanbeactiveatonetime.Theactivechoicewillbeyellowandtheinactivechoicewillbegrayed.

£ IfascreencontainsaSavebutton,youmusttapit

ifyouwanttosavethesettingsyouhavemade.

£ SomescreenscontainaDonebuttonwhichsavesyoursettingstotemporarymemoryandexitstothepreviousscreen.Inthosecases,youwillreturntoascreenwithaSavebutton.Tapittosavethefullconfigurationofsettingsonthatscreentonon-volatilememory.

£ SomescreenscontainaCancelbuttonwhichexitstothepreviousscreenwithoutsaving

yoursettings.

£ Numerickeypadscreens,whichallowyoutosetvaluesforvariousparameters,allcontainabottomrowwiththesebuttons:

• decimalpoint(•)whereappropriate• Savetoleaveandretainyourchanges• Deltodeletethelasttappednumberor

decimalpoint• Quittoleavewithnochangessaved

£ Atnumerickeypadscreens,ifyouattemptto

enteranumericvaluelargerthantheallowedmaximum,youwillseeanerrormessage.UsetheDelbuttontodeleteyourlastentryandtryagain.Theallowedmaximumvalueisalwaysdisplayedintheupperleftcornerofthescreen.

£ EveryscreencontainsaHelpbuttonforcontextsensitivehelp.

AcomputerinterfaceisalsoavailablefortheGenieTouch.VisittheGenieTouchproductpageatourwebsiteorseepage19tolearnmoreabouttheASCIIcommandsandhowtousethem.

www.kentscientific.com

Page 9: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

7

GenieTouchHelpWhenyoutapHelpatanyscreen,youwillseecontext-sensitivehelpforthatscreen.Tapping TOC(TableofContents)atanycontext-sensitivehelpscreenproducesalistofOwner’sManualsections.Thelistitemsarenottouch-sensitive.HighlightthesectionyouwanttoviewandtapSelecttoviewit.TapDowntomovetowardtheendofthelistandUptomovetowardthetop.Atthehelppage,tapNextorPrevtopageforwardorbackwardthroughmulti-pageentries.TapReturntogobacktothemostrecentTableofContents.Tap TOC togobacktothemainTableofContents.

TheGenieTouchMenusWhenyouturntheGenieTouchon,theOpening Screenappears.Itcontainstwobuttons:

£ Setup: LeadstotheSetupscreenforconfiguringtheGenieTouch

£ Main: LeadstotheMainscreenforcontrollingtheGenieTouchwhileitisrunning£ SetupTheSetupscreencontainsfourmainbuttons:Syringe,Operation,Auto ReverseandAdvanced.Thefunctionofeachofthesebuttonsisdescribedbelow.

£ Setup > SyringeTheSyringeSelectionscreenThetopbuttonofthisscreenshowsthekindofsyringecurrentlyselected—GenericorBrand.Theselectedtypeappearsinyellow,andtheotherbuttonsonthescreendisplaythecurrentsettings,ifany,forthattypeofsyringe.TochangebetweenGenericandBrand,tapthebluechoice.Theotherbuttonsonthescreenwillchangetoreflectthetypeofsyringeyouhaveselected.

£ Syringe>GenericUseaGenericSyringeIfGenericisnotalreadythechosensyringetype,tapGeneric,thenfollowthesesteps,describedindetailbelow:

1. SetthesyringeVolume.

Page 10: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

8

2. Seteitherthesyringe DiameterorLength.

3. Choosethesyringeorientation.

4. CalibratetheEmpty Position.

Whenyouarefinished,pressSave.

SetthesyringeVolume TapVolumetosetthevolumeforyourgenericsyringe.Attheresultingnumerickeypad,youmayenteravalueupto150ml.TaptheUnitsbuttoninthebottomrowtoswitchbetweenmlandul.Seteitherthe syringe DiameterorLength Tosetthediameter,tapDiameterandusetheresultingnumerickeypadtoenteranexactinsidediameterforyoursyringeinmm.Tosetthelengthinstead,tapLengthandfollowthesameprocess.Lengthisdefinedasthedistancebetweenthefullandemptypositions.Whenyouspecifyoneofthesevalues,GenieTouchwillcalculatetheotherbasedonyourVolumesetting.Choosethesyringeorientation TapFacing LeftorFacing RighttoindicatehowyouplantoorientthesyringeintheGenieTouch.Rightandleftrefertoyourrightorleft.NotethisselectiondefinesthephysicaldirectionfortheInfuseandWithdrawfunctions.(Re)CalibratetheEmpty Position Thebuttonforcalibrating(orrecalibrating)theEmpty Positiondoesnotappearuntilyouhavespecifiedalltheotherparametersforyoursyringe.Ifyouareusinganewkindofsyringe,thebuttonwillreadCalibrate Empty Positionandyoumustcompletetheprocessdescribedbelowbeforeyoursyringeisfullydefined.

Ifyouareusingasyringewithidenticaldimensionstooneyouhaveusedbefore,GenieTouchhassaveditscalibrationdatainthelookuptableandwillretrievethedataautomatically.ThebuttonwillreadRecalibrate Empty Position.Youshouldtapitonlyifyouwanttomakeachange.OneofthechangesyoucanmakeispressingErasetoremovethecurrentsyringefromtheGenieTouchlookuptable.Thetablecanhold48

Syringe Lengthmustbebetween1cmand12cmandsyringeDiametermustbebetween0.02mmand3cm.Ifyoutrytoinputnumbersoutsidetheseranges,youwillseeanerrormessage.

IfyouspecifyaDiameter/Volume combinationthatwouldcauseacalculatedLength tobeoutsidetheallowedrange, oraLength/VolumecombinationthatwouldcauseacalculatedDiametertobeoutsidetheallowedrange,thecalculatedvaluewilldisplayas Out of range.

Page 11: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

9

differentGenericorBrandsyringes.Ifyouexceedthatnumber,theoldestentrywillbeerasedtomakeroomforthenewone.

Tocalibratetheemptypositionforyoursyringe:

1. Mountthesyringe.Seepage3fordirections.

2. TapCalibrate Empty Position.

3. Movethepusherblockuntilitiswithinabout0.2mmoftheendofthesyringeorlightlytouchingit.Usethe Fast << or Fast >>buttonstogetclosetothecorrectpositionandthenswitchtotheSlow Adjustbuttontofine-tunetheposition.Notethat Slow Adjustmovesonlytowardwhatyouhavedefinedasemptyforthissyringe.

4. PressSavetosaveyoursettingsorCanceltoreturntothepreviousscreenwithno

changes.

£ Syringe>BrandUseaBrandSyringeIfBrandisnotalreadythechosensyringetype,tapBrand.Thenfollowthesesteps,describedindetailbelow:

1. Choosethesyringebrand.

2. Selectthesyringevolume.

3. Calibratethe Empty Position.

Whenyouarefinished,pressSave.

CAUTION!Positiontheplungertobelightlytouchingtheendofthesyringebarrelortobeabout0.2mmfromtheendofthesyringe.Donotforcetheplungerintotheendofthesyringe.Thatwillcausethesyringeandplungertomoveandflexduringuseandresultininaccuraciesinplungermovement.Inaddition,ifyoudefinetheEmpty PositiontobetooclosetotheendofthebarrelandthenselectaMaximum Forceof100%(seepage16),theforcecouldeasilydamageorbreakadelicatesyringe.

Page 12: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

10

ChoosethesyringebrandTapBrand.Attheresultingscreen,tapthenameofthesyringemanufacturer.SelectthesyringeVolume TapSelect Volume.Attheresultingscreen,choosefromthevolumesavailableforyourselectedmanufacturer.Youmustselectasyringebrandbeforeyouselectthevolume.CalibratetheEmpty Position SeedirectionsandCAUTIONnoteonpage7underSyringe>Generic.

£ Setup>Operation

TheOperationScreenThetopbuttonofthisscreenshowsthetypeofsyringeusecurrentlyselected—InfuseorWithdraw.Theselectedtypeappearsinyellow.TochangebetweenInfuseandWithdraw,tapthegrayedchoice.TheotherbuttonsonthescreendisplaytheavailablechoicesofOperationmodewiththecurrentsyringeuseselectioninyellow.

Foreithertypeofsyringeuse—InfuseorWithdraw—thefourchoicesofOperationmodearethesame:£ Constant Flow—drivesthepusherblockata

constantspeed£ Ramped Flow—drivesthepusherblockatalinearly

increasingordecreasingspeed£ Stepped Flow—drivesthepusheratanincreasingor

decreasingspeed,withthetotalchangedividedintoanumberofequalsteps

£ Pulsed Flow—drivesthepusherinaseriesoftwoalternatingcontinuousflowpulses

Eachofthefourchoicesleadstoascreenthatallowsyoutodefinethedetailsforthemode.Eachofthescreensisdiscussedbelow.

IfyouplantouseAuto Reversetoenableinfusionand/orwithdrawalofavolumegreaterthanthevolumeofthesyringe,youmustenableAuto Reversebeforeyouspecifythecharacter-isticsofyourOperation.Seepage14.

SyringeAbbreviationsAT: Air-TiteAllPlastic

HSWNorm-JectBD: BectonDickinson

Interm,WW,Plastipak

BDG: BectonDickinsonGlass—alltypes

CA: CadenceScienceMICRO-MATEglass

HA Hamilton1000-SeriesGastight

ME: MedallionSG: ScientificGlassEng

SGESH: SherwoodMonoject

PlasticTE: TerumoTE

Page 13: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

11

£ Operation>ConstantFlowTapConstant FlowforascreenthatdisplayschoicesfordefiningyourOperation.TherearefourwaystodefineaConstant Flow Operation,eachofwhichrequiresyoutospecifytwoparameters.Inallcases,thepusherblockmovesataconstantspeed.

SpecifyVolumeandTimeTapBy Volume & TimeattheConstant Flowscreentospecifythetotalvolumeofliquidtobemovedandthelengthoftimeoverwhichtomoveit.Thecalculatedflowratewillthenbedisplayed.

TapVolumeandenterthevolumeofliquidtobemoved.Unitsswitchesbetweenul(microliters)andml(milliliters).

TapTimeandenterthetime.Unitscyclesbetweenseconds,minutesandhours.Usedecimalformat—1.5hror90min,forexample,setthesameTime.

SpecifyFlowandVolumeTapBy Flow & VolumeattheConstant Flowscreentospecifytheflowrateandthetotalvolumeofliquidtobemoved.Thecalculatedtimeforfulldeliverywillthenbedisplayed.

TapFlowandenterthevolumeofliquidtoflowperunittime.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.

TapVolumeandenterthetotalvolumeofliquidtobemoved.Unitscyclesbetweenul(microliters)andml(milliliters).

SpecifyFlowandTimeTapBy Flow & TimeattheConstant Flowscreentospecifytheflowrateandthetotaltimeoverwhichtheflowistobemaintained.Thecalculateddeliveryvolumewillthenbedisplayed.

TapFlowandenterthevolumeofliquidtoflowperunittime.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.

WhenyoutapanyofthebuttonsattheConstant Flowscreen,youwillseethenumerickeypadtoallowyoutoenteryourvalue.Enteryourvalueusingatmostfourdigitsandadecimalpoint.

Page 14: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

12

TapTimeandenterthetotaldurationoftheOperation.Unitscyclesbetweenseconds,minutesandhours.Usedecimalformat—1.5hror90min,forexample,setthesameTime.

SpecifyFlow TapBy Flowatthe Constant Flowscreentospecifyaflowratetobemaintaineduntilthesyringeisemptyfor InfuseorfullforWithdraw.Notethatthesecondparameteryouarespecifyinghereisactuallyatimeof"untilfullorempty."YoucanusethisOperationmethodalongwithAuto Reverseforconstantpumping.Seepage14.

£ Operation>RampedFlow

TapRamped FlowforascreenthatdisplayschoicesfordefiningyourOperation.Youmustspecifythreeparameters—Start Flow Rate, End Flow RateandeitherTotal TimeorTotal Volume.Theprocess—eitherInfuseorWithdraw,dependinguponwhichyouhavechosen—beginsattheStart Flow Rateandincreasesordecreaseslinearlytothe End Flow Rateuntilthe Total VolumeismovedortheTotal Timeelapses.SpecifyStartFlowRateTapStart Flow RateattheRamped FlowscreentospecifytheflowrateatwhichtheOperationistobegin.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.

SpecifyEndFlowRateTapEnd Flow RateattheRamped FlowscreentospecifytheflowrateatwhichtheOperationistoend.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.SpecifyEitherTotalVolumeorTimeYoumustspecifyeitherthetotalvolumeofliquidtobemovedduringyouroperationorthetotallengthoftimeoverwhichliquidistobemoved.Either:

TapTotal VolumeattheRamped Flowscreentoenterthetotalvolumeofliquidtobemoved.Genie-TouchwillcalculatetheTotal Timeforyouroperationanddisplayitatthe Ramped Flowscreen.Unitsswitchesbetweenul(microliters)andml(milliliters).

WhenyoutapanyofthebuttonsattheRamped FlowscreenortheStepped Flowscreen,youwillseethenumerickeypadtoallowyoutoenteryourvalue.Enteryourvalueusingatmostfourdigitsandadecimalpoint.

Notethat0(zero)isanallowedvalueanytimeyouareaskedtoenterabeginningorendingflowrateforRamped FloworPulsed Flow.

Page 15: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

13

orTapTotal TimeattheRamped Flowscreentoenterthelengthoftimeyouroperationistocontinue.GenieTouchwillcalculatetheTotal Volumeforyouroperationanddisplayitatthe Ramped Flowscreen.Units switchesbetweenseconds,minutesandhours.

£ Operation>SteppedFlow

TapStepped FlowforascreenthatdisplayschoicesfordefiningyourOperation.Youmustspecifyfourparameters—Start Flow Rate,End Flow Rate, Number of StepsandeitherTotal TimeorTotal Volume.Theprocess—eitherInfuseorWithdrawdependinguponwhichyouhavechosen—beginsattheStart Flow RateandincreasesordecreasestotheEnd Flow Rateinstepsofequaldurationuntilthe Total VolumeismovedortheTotal Timeelapses.SpecifyStartFlowRateTapStart Flow RateattheStepped FlowscreentospecifytheflowrateatwhichtheOperationistobegin.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.SpecifyEndFlowRateTapEnd Flow RateattheStepped FlowscreentospecifytheflowrateatwhichtheOperationistoend.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.SpecifyeitherTotalVolumeorTimeYoumustspecifyeitherthetotalvolumeofliquidtobemovedduringyouroperationorthetotallengthoftimeoverwhichliquidistobemoved.Either:

TapTotal VolumeattheStepped Flowscreentoenterthevolumetobemoved.GenieTouchwillcalculatetheTotal Timeforyouroperationanddisplayitatthe Stepped Flowscreen.Unitsswitchesbetweenul(microliters)andml(milliliters).

orTapTotal TimeattheStepped Flowscreentoenterthelengthoftimeyouroperationistocontinue.GenieTouchwillcalculatetheTotal Volumeforyouroperationanddisplayitatthe Stepped Flowscreen.Unitsswitchesbetweenseconds,minutesandhours.

SpecifyNumberofStepsTap Number of Stepsatthe Stepped Flowscreentospecifythenumberofstepsintowhichthe Operationwillbedivided.

Page 16: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

14

£ Operation>PulsedFlowTapPulsed Flowforascreenthatdisplaysthechoicesfordefiningyour Operation.SincePulsed Flowismadeupofaseriesoftwo-partConstant Flow pulses,youmustspecifytwoparametersforeachofthepulsepartsandyoumustspecifytheNumber of PulsestobecarriedoutbeforetheOperationends.ConfigureFirstPartofthePulseTapPulse: 1st PartatthePulsed Flowscreentodefinetheparametersforthefirstpartofthepulseyouwanttouse.TheprocessisidenticaltoconfiguringaConstant Flow Operationwithoneexception:theBy Flowchoiceisnotavailable.Forcompletedirections,seeConstant Flowonpage11.

ConfigureSecondPartofthePulseTapPulse: 2nd PartatthePulsed Flowscreentodefinetheparametersforthesecondpartofthepulseyouwanttouse.TheprocessisidenticaltoconfiguringaConstant Flow Operationwithoneexception:theBy Flowchoiceisnotavailable.Forcompletedirections,seeConstant Flowonpage11.

SpecifyNumberofPulsesTap Number of Pulsesatthe Pulsed FlowscreentospecifyhowmanypulseswillmakeuptheOperation.YoumayenteranumberofpulsesoryoumaychooseContinuouswhichtellstheGenieTouchtocontinuepulsinguntilthesyringeiseitheremptyorfull,dependinguponwhetheryouhavechosenInfuseorWithdraw.

£ Setup > Auto Reverse TheAutoReverseScreenAuto Reversealternatesinfusionandwithdrawal.Itallowstheinfusionorwithdrawalofmorefluidthanasyringecanholdinasinglefilling.Auto ReversemaintainsallthecharacteristicsofyourOperation,whetheritisConstant,Ramped,SteppedorPulsed.Yoursettingforsyringeuse—InfuseorWithdraw—determineswhichofthetwowilloccurfirstandwillbethefocusoftheprocess.Forexample,ifyouhavechosenInfuse,theprocesswillbeginwithaninfusionaccordingtoyourOperationsettings,followedbya

YoucanuseAuto ReversetogetherwiththeContinuouschoiceforNumber of Pulsestoachievecontinuousoperationbeyondthevolumeofthesyringe.SeeAutoReversebelowformore.

Page 17: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

15

constant-speedwithdrawtorefillthesyringe,andthenanotherinfusion.ThepumpwillchangedirectioneachtimetheendofthesyringeisreacheduntilitreachestheendofyourOperationasyouhavedefinedit.SeeOperationonpage10.

ThetopbuttonofthisscreenshowsthestatusofAuto Reverse—EnabledorDisabled.Thecurrentstatusappearsinyellow.Tochangethestatus,tapthegrayedchoice.Touse Auto Flow,youmustspecifyaVolumeandeitheraReverse FloworReverse Speed.SpecifyaFloworSpeedTapReverse Flowtoenteryourflowrate.Useatmostfourdigitsandadecimalpoint.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.TapReverse Speedtoenterthe%offullpumpspeedtobeused.Enteratmostthreedigitsforavaluebetween1and100.SpecifyVolumeTapReverse Volumetoenteravolumeperpumpstroke.Thisvolumemustnotexceedthevolumeofthesyringeyouplantouse.

£ Setup > Advanced

TheAdvancedFeaturesSelectionScreenThisscreencontainsthreechoices:

£ Audible Alarms £ Maximum Force £ External Trigger

Eachleadstoanotherscreenwithchoicesforsettingupthecorrespondingfeature.

Important:TouseAuto Reverse,yoursyringemustbeequippedwithcheckvalvesforunidirectionalflow.Otherwise,Auto Reversewillsimplypushthesamesyringecontentsintoandoutofthesyringewhichcouldhaveundesirableconsequences.

NotethatifyoubeginyourOperation withapartiallyfilledsyringe,theinitialwithdrawalorinfusionstrokemaybeapartialone.

Page 18: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

16

£ Advanced>AudibleAlarmsTapAudible AlarmsattheAdvancedscreentosetoneormoreconditionsthatwilltriggerthebeeperwhenyouarerunningtheGenieTouch.Buttonsforcurrentlyenabledalarmswillbeyellow.Thosefordisabledalarmswillbegrayed.WhenallfourAlarmsareconfiguredthewayyouwantthemtobe,tapSave.TurnAlarmOnwhenRunEndsTapRun Donetocyclebetweenenablinganddisablingthisalarm.Whenitison,aone-secondlongbeepsoundsattheendofanOperation.TurnAlarmOnwhenRunisPartiallyThroughTap% Run DonetoenterthepercentageofRuncompletionthatwilltriggerthreequarter-secondlongbeeps.Enteranumberbetween1and99.Toturnthisalarmoff,tap Alarm Off atthenumerickeypadscreen.TurnAlarmOnwhenMotorisStalledTapMotor Stalledtocyclebetweenenablinganddisablingthisalarm.Whenitison,fiveone-secondlongbeepswillsoundtoindicatethatthemotorisstalledandthusunabletomovethepusherblock.TurnAlarmOnwhenMotorSlipsTapMotor Slipping tocyclebetweenenablinganddisablingthisalarm.Whenitison,fourhalf-secondlongbeepswillsoundtoindicatethatthemotorisunderexcessiveloadandcannotmovethepusherblock.

£ Advanced>MaximumForce

TapMaximum ForceattheAdvancedscreentosetthelinearoperatingforceofthepusherblock.Enteryourselectionasapercentageofthemaximumforce,whichisover36kg(80lb).Fortypicaloperations,asettingof50%shouldbesatisfactory.

£ Advanced>ExternalTrigger

Tap External TriggerattheAdvancedscreentoconfiguretheuseofanoptionalexternalsignalasaneventtrigger.YoucanapplytheexternalsignalthroughtheAUXport(externalcontrol/remoteconnectorinFig1b)onthebackoftheGenieTouch.Thatsignalmightbe,forexample,afootpedal,aneventswitchoracomputerinterface.

TheminimumsettingforMaximum Forceis25%.ThisshouldbeusedwithcautiontoavoidstallingtheGenieTouchmotor.

Page 19: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

17

ThetopbuttonontheExternal Triggerscreenshowsthecurrentstateofthetrigger—EnabledorDisabled.Thecurrentstatusappearsinyellow.Tochangethestatus,tapthegrayedchoice.ThethreebuttonsbelowallowyoutoconfiguretheGenieTouchbehaviorinresponsetothesignal.Youmustdefinewhether"off"istobeinterpretedasastoporapause,sethowGenieTouchwillbehaveonachangeinthesignalandspecifythenatureoftheelectricalsignalfromyourexternaltrigger.SetmeaningofOffTapOff = StoptodefineOffasacompletestopofyourOperation, equivalenttotappingtheStopbuttonontheRunscreen.Thisresetsthedisplayofthemovedandremainingvolumesdisplayedthere.TapOff = PausetodefineOffasaPause, equivalenttotappingthePausebuttonontheRunscreen.SetbehavioronchangeinsignalTapHold for OfftocauseaheldtriggertoeitherStoporPauseyourOperation,dependingonyoursettingat Off =above.Tap Tap for Run/Off tocauseataptoaswitchorfootpedaltocyclebetweenOffandRun,withOffdefinedatOff =above.SpecifynatureofsignalTapClosed = Offtodefinetheclosedorshorted-to-groundstateofyoursignalasOff.TapOpen = Off todefinetheopenorfloatingstateofyoursignalasOff.

Page 20: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

18

£ MainTheMainscreenshowsthekindofsyringeinuseandthedetailsofthecurrentlydefinedOperation.Inaddition,itcontainsfivebuttons:RUN,Setup, << , >>,Help.Thefunctionofeachofthesebuttonsisdescribedbelow.

£ Main > RUN TapRUNtobeginyourOperation.TheRUNbuttonchangestoPauseandtheOperationstarts.TappingPausecausestheOperationtostoptemporarily.ThePausebuttonchangestoResume,allowingyoutopickupwheretheoperationleftoff.WhiletheGenieTouchisrunningorpaused,the << and >>buttonsdisappearandarereplacedwithStop.PressStoptoendtheOperationandresettheOperationparameterstotheirinitialstate.

£ Main > Setup TapSetupfortheSetupmenus.

£ Main > << and > > Tap<< tomovethepusherblocktotheleftand>> tomoveittotheright.Theblockbeginsmovingveryslowlyandacceleratesuntilitreachesasteadyspeed.Tokeeptheblockmoving,keepyourfingeronthebutton.Letgotostoptheblock.Youcanalsopressanddragyourfingerslowlyinthedirectionofthebuttonarrows.Whenyouliftyourfinger,thepusherwillcontinuemoving.Stopitbytappingeither<<or>>.

£ Main > Help TapHelpforhelpwiththecurrenttopic.TapNextorPrevtopageforwardorbackwardthroughmulti-pageentries.Whenyouarefinished,tapReturn.ToseealistofavailableOwner’sManualsectionstoviewontheGenieTouchscreen,tap TOC(TableofContents)atanycontext-sensitivehelpscreen.TapDownorUptohighlightthesectionyouwanttoviewandtapSelecttoviewit.Attheresultinghelppage(s),tapNextorPrevtopageforwardorbackwardthroughmulti-pageentries.TapReturntogobacktotheTableofContents.

ImportantTheRUNbuttondoesnotappearifyouhavenotcompletedtheSetupprocess.

Page 21: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

19

GenieTouchPCInterfaceYoucandownloadComputerDriversforyourGenieTouchat:

http://www.ftdichip.com/Drivers/VCP.htm.ThesedriverswillallowyourcomputertocommunicatewiththeGenieTouchviaavirtualCOMport(VCP)connection.ForassistanceinsettingupyourcomputertocommunicatewiththeGenieTouch,pleasecontactKentScientific.

Notes§ AllcommandsaresentasASCIIcharacters.§ Aone-timepowerupcommandpromptissuppliedbytheinjectorInjector nnn.§ AllcommandsthatproduceerrorsareignoredandreturnanErrormessage.§ Commandsandargumentsarecaseinsensitive.§ Onlyonecommandperlineispermitted.§ Partsofcommandsincapsmustbesent;lowercaseportionsareoptional.§ Whitespacecanbeanynumberofspacesortabs.§ Alltextonalineafter‘!’isignored.§ Mostcommandsthatspecifyavalueorastatehavearequestformofthecommand

thatisthesamecommandprecededby?.

UnitDefinitionsLengthum micrometermm millimetercm centimeter

Volumeul microliterml millilitercc sameasml

Times,sec secondsm,min minutesh,hr hoursTimevaluescanbecombinedaspairs,e.g.3h20m.

Flowvu/tu volumeunitpertimeunitvu canbeanyoftheabovevolumeunitsandtucanbeanyoftheabovetimeunits.

COMPortSettings

Bitspersecond:115200DataBits:8Parity:NoneStopbits:1

Page 22: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

20

CommandsAdHocASCIICommandsASCIIpacket:Rec:ASCIIchars,[CR|LF|CRLF]

:Xmit:‘>’,ASCIIchars,CR,LF

BEEpNoarg =>1secbeepBEEp Num Num=Numberof1secbeepsBEEp Num On ms Off ms

Num=NumberofbeepsofdurationOnmSandseparationofOffmS OnandOffaretruncatedinternallytoamultipleof10ms.

Example:bee 3 500 ms 1500msèThreebeeps,on0.5sec,off1.5sec CONtrol CONtrol [LOCK | UNLock]

LOCK:Enterlockedstate.AcceptnocommandsexceptUNLOCK UNLOCK:Releasefromlockedstate.

Example:conlockèPreventunintentionaltouchscreeninputREPort [RESet] [OFF | ON ] [ MOVing ] [ POS ] [ PERc ] [ VOL ] [ EVEnt ] [ period ] ?REPort [RESet] [ MOVing ] [ POS ] [ PERc ] [ VOL ] [ EVEnt ] ?REPort

§ Reportposition,percentage,volumeand/orevents.Reportingcanbeperiodic,byrequestortriggeredbyanevent.

§ The?formsrequestanimmediatesinglereporto Ifthereareargumentsthereportcontainstherequestedinformation.The

argumentsarenotretained.o Iftherearenoargumentsthereportcontainstheitemsthatwerelastsetupusing

thenon-? REPORTform.§ Thenon-?formspecifiesitemsthatareretainedinEEPROM.Ifanon-zeroperiodis

specifiedreportingwillbeautomaticandperiodic.§ POSistheabsolutecarriagepositioninxxx.xxmm§ PERCisthepercentcompletionasxx.xx%wheninRUNmode§ VOListhedeliveredvolumewheninRUNmode§ EVENTreportsignificantchangeinstatus

o Stallifmovementstoppedbystallo Slippingifmotorisnotabletomoveasfastasrequestedo Limitifmovementstoppedbylimitbeforereachingtargeto Endifmovementstoppedbyreachingtarget

Page 23: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

21

§ MOVING restrictsreportingtowhenunit’scarriageisinmotion.Hasnoeffectona?report.

§ ONandOFFturnson/offperiodicandeventreporting,Hasnoeffectona?report.§ periodcanbeinanyoftheformsoftimebutisroundedtointegerseconds(thatcannot

exceed65535).Hasnoeffectfor?reporting.§ Ifnoargumentisgivenasinglereportisgeneratedusingthelastspecifiedarguments.§ RESETclearsallreportsettings.Itmustbethefirstargumentgivenorelseitwillclear

theargumentsthatprecedeit.§ AllreportsettingsarestoredinEEPROM

Example:rep on mov pos 1 secèReportpositioneverysecondwhilemovingRUNBeginrunningusingthecurrentconfigurationPAUsePauserunning.RUNwillun-pause.STOpStoprunning.RUNwillstartfrombeginning.Example:runèBeginrunningCLEar arg

§ arg=ALL:ClearSyr1,Syr2,Dispense,AutoRev§ arg=SYRinge:ClearSyringe§ arg=OPEration:ClearDispense,AutoRev§ arg=AUTorev:ClearAutoRev§ AllCLEarcommandswillforcethemotortostopifitisrunning

Example:cle syrèClearsyringevaluesDirectControlNotesaboutDirectControlCommandsForbothMOVeandSPEedpreviousvaluesarepreservedunlessexplicitlychangedbycommand.ToexitDirectControlissueaSTOP command.DefaultvaluesforallDIRECTCONTROLcmds:

§ DACissetto50%§ DestArgissettoPHYSICAL§ Speedissetto0

Page 24: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

22

ABSpos § Ifmotornotmoving,updateabsolutemotorpositionfromcalibrationtableandrestart

backlashcorrection.§ Ifmotormoving,restartbacklashcorrection

Example:abs REFerence RefArg RefArg

§ WhereLEFtandRIGhtareimpliedandbothmustbepresent§ REFerence [LEFt RefArg] [RIGht RefArg] § RefArg=PHYsical|SYRinge|FULl|EMPty|#cm|#mm§ InthecaseofFULland EMPtytheLEFtandRIGhtareignoredandthesyringefacing

directiongovernsusage.Example: ref 6 cm 4 cmèSetleftreferenceto6cmandrightreferenceto4cmSPEed SpeedArgs

§ SpeedArgs=#%|#um/hr|stop[#F]§ #%,#um/hrandstopspecifyspeed§ #Fspecifiesforce,i.e.DACvaluewhere100>=DAC>=MinDAC

Example:spe 50% 100FèSetspeedto50%maximumandforceto100%maximumMOVe LEFt | RIGht | ABS DestArg [SpeedArgs]

§ IfLEFt orRIGht isgiventhen:o DestArg=UNLimited=>nolimittomotiono DestArg=PHYsical=>physicalendofallowedmotionwithouthittingendstopso DestArg=SYRinge=>syringefulloremptypositionbasedonsyringeorientationo DestArg=REFerence=>valuespecifiedbyREFerencecommand

§ IfABSisgiventhenDestArg=#mmor#cm=>absolutepositionrelativetorightEndstop.Absolutepositioncanbeoutsideofsyringelimitssocaremustbeexercisednottodamagethesyringe

Example:mov abs 5 cmèMovetolocation5cm(fromrightendstop)SyringeCommandThefollowingcommandspecifiessyringeselectionparameters.Allargumentsthathaveassociatedunitsmustbegivenwiththeirunits.NumericalValues

ForLengthorDiameterForm:xxxx,xxx.x,xx.xxorx.xxx

Page 25: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

23

Units:um,mm,cmMustbeprecededbyLENorDIA

ForVolumeForm:xxxx,xxx.x,xx.xxorx.xxxUnits:ml,ulIsnotprecededbytext.Useonlythevalue,eg,40mlnotVOL 40ml

SYRinge Name Volume [RIGht | LEFt] [EmptyPos]

§ Thisformisusedtospecifyabrandnamedsyringe.§ Nameisanabbreviationofthemanufacturer’sname(see

listatright).§ Volumemustmatchoneofthestandardvolumes

providedbythespecifiedmanufacturer.SYRinge DIAmeter dia | LENgth len Volume [RIGht | LEFt] [EMPty pos]

§ Thisformisusedtospecifyagenericsyringe.§ EitherDiameteror Length—notboth—mustbespecified.§ RIGhtorLEFtspecifiesthedirection,i.e.RIGhtfacingor

LEFtfacing.IfneitherisgivenRIGhtfacingisassumed.§ EMPty posspecifiestheabsolutepositionthat

correspondstothesyringebeingempty.Itisspecifiedwithoutunitsandisassumedtobein10umunits.Itisforadvanceduseonly.Ifitisabsentthecurrentstoredvalueforposwillbeused.Tochangepos,useEMPandtheabsolutecarriagepositionin10umincrements.Forexample,syr dia 15mm 8ml rig emp.

Example:syr bd 60ml rightèSelectBD60mlbrandmountedfacingright.OperationCommandsThefollowingcommandsspecifydispensingparameters.Allargumentsthathaveassociatedunitsmustbegivenwiththeirunits.NumericalValuesForTimeForm:xxxx,xxx.x,xx.xxorx.xxxUnits:s,m,h,sec,min,hrorcompositeforms#h#mor#m#sLimit:Minimumtimeis100mSandresolutionis10mS

SyringeAbbreviationsAT: Air-TiteAllPlastic

HSWNorm-JectBD: BectonDickinson

Interm,WW,Plastipak

BDG: BectonDickinsonGlass—alltypes

CA: CadenceScienceMICRO-MATEglass

HA Hamilton1000-SeriesGastight

ME: MedallionSG: ScientificGlassEng

SGESH: SherwoodMonoject

PlasticTE: TerumoTE

Page 26: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

24

ForVolumeForm:xxxx,xxx.x,xx.xxorx.xxxUnits:ml,ul

ForFlowForm:xxxx,xxx.x,xx.xxorx.xxxUnits:ml/min,ml/hr,ul/min,ul/hr

Definitions

§ “dir”indicatesoneofthekeywords:INFuseorWIThdraw.

§ GenVolindicatesaspecificvolume,Vol.

[WIThdraw | INFuse] Flow [WIThdraw | INFuse] Flow GenVol [WIThdraw | INFuse] Flow Time [WIThdraw | INFuse] Time GenVol

§ IfFlowistheonlyargumentthenContinuousdeliveryisassumed.§ Anytwoof Time,GenVolorFlowwillspecifyaspecificdeliveryamount.§ Argumentscanbespecifiedinanyorder.§ Allargsmusthaveassociatedunits.Theunitsareusedtobothspecifythequantityof

thevalueandtoidentifythetypeofthevalue.Example:inf in 10 ml/min 1 minèInfusewithaflowof10ml/minfor1min[WIThdraw | INFuse] RAMp {Time or GenVol} FlowBeg FlowEnd

§ DefinesalinearrampwithstartingflowofFlowBegandendingflowofFlowEnd.§ EitherTimeorGenVolcanbespecifiedtosetthetotalamountdelivered.§ FlowBegand FlowEndmustbegivenasapairandinthestatedorder.§ Time,GenVolandtheFlowpaircanbeingiveninanyorder.§ Oneoftheflowscanbezero.

Example:wit ram 50 sec 0ml/min 10ml/min èWithdrawfor50secwithalinearrampstartingataflowof0andendingwithaflowof10ml/min[WIThdraw | INFuse] STEp #Steps {Time or GenVol} FlowBeg FlowEnd

§ Definesasteppedramp.§ Theflowisuniformlydividedinto#stepsstartingatFlowBegandendingwithFlowEnd.§ The#Stepsvalueislimitedtoainteger2to9999anditmustbethe1stargumentafter

STEP.§ InallotherwaysthesyntaxisthesameasforRAMP.

Page 27: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

25

Example:wit ste 4 50 sec 10ml/min 5ml/minèWithdrawfor50secwith4stepsstartingataflowof10ml/minandendingwithaflowof5ml/min[WIThdraw | INFuse] PULse #pulses Part1 Part2

§ Definesapulseddelivery.§ Apulseisdividedintotwoparts:Part1andPart2.§ ThespecificationofaPartmustproduceaflowrateforanon-zeroperiodoftime.§ Theflowratecanbezero.§ APartcanbespecifiedbyanyofpairs:{Flow&Time},{Flow&Volume}or{Time&

Volume}.§ The#pulsesislimitedtoanintegerof1to9999.§ #pulsecanbereplacedbythekeyword FOREVERinwhichcasedeliverywillcontinue

untilmanuallystopped.§ TheorderoftheargumentsforthePULSEcommandisfixed.

Example:wit pul 20 0ml/min 5 sec 10ml/min 5mlèPulsedwithdrawwith20pulses.Thefirstpartofthepulseiszeroflowfor5sec.Thesecondpartofthepulseis10ml/minforavolumeof5ml.Requests?CONtrol Returns:Locked/Unlocked?SYRinge

§ volumeisvolumeofsyringewithunits§ lengthislengthofsyringewithunits§ diameterisdiameterofsyringewithunits§ Returns:

o Syringe volume Len|Dia length|diameter [Left] o Syringe volume name [Left] o Syringe No Empty Pos o Syringe Undefined

?OPEration

§ dirisInfuseorWithdraw§ flow,beginFlow,endFlow,flow1orflow2areflowswithunits§ timeistimewithunits§ genVolisvolumewithunits § numisthenumberofstepsinrampornumberofpulses§ Returns:?

o dir Continuous flow

Page 28: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

26

o dir Constant flow time o dir Constant flow genVol o dir Constant time genVol o dir Ramp beginFlow endFlow time o dir Ramp beginFlow endFlow volConc o dir Steps:num beginFlow endFlow time o dir Steps:num beginFlow endFlow genVolo dir Pulses:num flow1 flow2 time o dir Pulses:num flow1 flow2 genVol

§ dir Undefined?REPort [RESet] [ MOVing ] [ POS ] [ PERc ] [ VOL ] [ EVEnt ] ?REPort SeediscussionunderREPort ?RUN

§ percispercentcomplete.Sentasnnn.nn%.Omittedifcontinuous.§ Returns:RunState§ RunState=Stopped,Paused [perc],Running[perc],Direct,Fast,Undefined

?SN ReturnsSN: nnn.. Ver: v.vvwhere nnn..istheserialnumberandv.vv istheversionnumber

Page 29: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

27

GeneralInformation

SpecificationsGenieTouchSyringenumber Upto2Syringesize 0.5ulto60mlUserInterface TouchscreenMaxLinearforce >36kg(80lb)SelectableFunctions ConstantRate

TimeDispensingVolumeDispensingRampPulsedStepped

ProgrammableFill/Empty YesPusherBlockStallDetection YesComputerInterface USBAudibleAlarmIndication YesPusherBlockPositionDetection

±0.2mm

Multi-SyringeRack OptionalForceAdjustment YesFootswitchInterface YesExternalTriggers YesWarranty 2years

CleaningInstructionsCleanthesurfaceofyourGenieTouchusingasoftlint-freecloth.Ifneeded,cleanwithaclothdampenedwithasolutionofwaterandamilddetergent,orisopropylalcoholmixedwithwaterata3:1ratio.

Page 30: GenieTouch User's Guide (Kent) V3.02 · 2018-06-25 · When you tap Help at any screen, you will see context-sensitive help for that screen. Tapping TOC (Table of Contents) at any

28

WarrantyThankyouforpurchasingaGenieTouch.Wetrulyappreciateyourbusiness.WestronglyadvisethatyoureadandstudythisOwner’sManualtoappreciatefullyallthefeatures,benefits,andcapabilitiesoftheGenieTouch.

ContactInformation KentScientificCorporation Email:[email protected] Phone:860.626.1172Torrington,Connecticut06790 Fax:860.626.1179www.kentscientific.comDisclaimerKentScientificCorporationmakesnorepresentationsorwarranties,expressed,statutoryorimplied,regardingthefitnessormerchantabilityofthecomponentsofthissystemforanyparticularpurpose.Further,KentScientificisnotliableforanydamages,includingbutnotlimitedto,lostprofits,lostsavings,orotherincidentalorconsequentialdamagesarisingfromownershiporuseoftheseproducts,orforanydelayintheperformanceofitsobligationsunderthewarrantyduetocausesbeyonditscontrol.InnocaseshallKentScientific'sfinancialobligationextendbeyondthecostoforreplacementoftheKentproductinquestion.KentScientificalsoreservestherighttomakeanyimprovementsormodificationstotheseproductsdescribedinthismanualatanytime,withoutnoticeofthesechanges.Allbrandandproductnamesusedinthismanualarethetrademarksoftheirrespectiveowners.Thisproductisnotdesigned,intendedorauthorizedforuseinhumanapplications.ProductWarrantyTheGenieTouchhasatwo(2)yearwarrantyincludingallpartsandlaborcharges.Thiswarrantydoesnotcoverdamagebyanycauseincluding,butnotlimitedto,anymalfunction,defectorfailurecausedbyorresultingfromunauthorizedserviceorparts,impropermaintenance,operationcontrarytofurnishedinstructions,shippingortransitaccidents,modificationsorrepairbytheuser,harshenvironments,misuse,neglect,abuse,accident,incorrectlinevoltage,fire,flood,othernaturaldisasters,ornormalwearandtear.ChangesormodificationsnotapprovedbyKentScientificCorporationcouldvoidthewarranty.Theforegoingisinlieuofallotherexpressedwarranties.KentScientificdoesnotassume,orauthorizeanypartytoassumeforit,anyotherobligationorliability.SatisfactionGuaranteeShouldyouexperiencedifficultywiththeGenieTouch,ourTechnicalSupportGroupwillassistyouintrouble-shootinganddeterminingiftheproductneedstobereturnedtoourfacility.WewillissueyouaReturnManufacturerAuthorization(RMA)numberbeforetheproductisshippedbackforrepair.Itisatthediscretionofthemanufacturertoreplaceorrepairadefectivepartorproduct.PleasecallCustomerServiceat888.572.8887toobtainaReturnManufacturerAuthorizationnumber.ShipmentswithoutanRMAnumberwillnotbeaccepted.Pleasenotethatafterour30-dayreturnpolicyperiodends,wewillbehappytoassistyouwithyourapplication,butcannotissueanycreditorrefundforareturnedGenieTouch.Priortoshipment,pleasecleananddecontaminatetheproductofanychemical,biological,orisotopiccontamination.PleaseincludeacompletedProductReturnFormwiththeshipment.Theformcanbefoundonpage29ofthisOwner’sManual.AdditionalcopiesareavailablebycallingKentScientificCustomerServiceat888.572.8887.