GenieTouch™
Infusion/Withdrawal Dual Syringe Pump
Owner’s Manual
TableofContents
AbouttheGenieTouch PartsoftheGenieTouch—1 SettingUptheGenieTouch—3 MountingSyringes—3
ControllingtheGenieTouch TheTouchScreen—6
GenieTouchHelp—7 TheGenieTouchMenus—7 Setup—7 Syringe—7 Operation—10 AutoReverse—14 Advanced—15 Main—18
GenieTouchPCInterface—19
GeneralInformationSpecifications—27CleaningInstructions—27Warranty—28DecontaminationVerificationForm—29
Thisproductisnotdesigned,intendedorauthorizedforuseinhumanapplications.©2016KentScientificCorporation,DocumentationVersion3.01,ForHelpVersion3.1andFirmwareRev1.03GenieTouchisatrademarkofKentScientificCorporation.
1
AbouttheGenieTouchTheGenieTouchTMisaninfusion/withdrawaldual-syringepumpwithmulti-syringecapabilityandaseriesofuser-configurablefunctionsthatallowittomeetavarietyofneeds.Itsupportsoptionalaudiblealarmsandoptionalforceadjustment.TheGenieTouchisnotdesigned,intendedorauthorizedforuseinhumanapplications.
PartsoftheGenieTouchItisusefultobecomefamiliarwiththepartsoftheGenieTouchbeforeyoubeginusingit.TheGenieTouchsupportsuseoftwosyringesatatime,bothfacingthesamedirection.Therightsyringemountandtheleftsyringemounteachaccepttwosyringesofupto60mlsize.Note:Theleftandrightsyringeplungerreversingclampsonthepusherblockareusedtograbtheflangeontheendofthesyringeplunger.Withthesyringeplungerfixedtothepusherblock,theplungercanbemovedinbothforward(infusingoremptying)andreverse(withdrawingorfilling)directions.Theseclampsarealsocalledanti-siphonclampsbecausetheykeepthesyringeplungerfrompassivelymovingunderlargehydrostaticloads.
2
RearviewTheRJ-45networkingconnectorsareforfutureGenieTouchmodels.
ThisistheU.S.powersupply.YoursmaylookdifferentifweshippedyourGenieTouchoutsidetheU.S.
Figure 1b
Powerswitch
PoweradapterentryExternalcontrol/remoteconnector(AUX)USBconnector
RJ-45networkingconnectors
3
SettingUptheGenieTouchTheonlyphysicalsetuptheGenieTouchrequiresisthatyouplugitinandmountyoursyringes.
1. Connectthesupplied12VDCpoweradaptertotheGenieTouchthroughthepowerentryconnectorontherearofthechassis.(SeeFigures1band1c.)
2. Plugtheadapterintoanappropriatepowersource.
3. Slidethepowerswitchon.YouwillseetheOpening screen,whichcontainstwobuttons—SetupandMain.TappingSetupleadstoaseriesofscreensforconfiguringyourGenieTouchtomeetyourspecificneeds.TappingMainleadstotheMainscreenforcontrollingtheGenieTouchwhileitisrunning.
MountingSyringesTherearetwosyringemountsontherightsideoftheGenieTouchchassisforthetraditionalindustry-standardright-facingsyringeplacement.Therearealsotwosyringemountsontheleftsideofthechassisifyouprefertouseleft-facingsyringes.TheRightSideTomountasyringeinoneoftheright-sidemounts:
1.Selecttheproperclamporientation.ThesyringebarrelclampsareL-shapedandworkforbothsmallandlargesyringesdependingupontheorientationoftheclamp:
Forsmallersyringes£ PlacethescrewintheholeinthebottomoftheshortlegoftheL.£ Placethespringovertheprotrudingendofthescrew.£ LaythelongerlegofLinthenotchinthesideofthechassis.£ Tightenthescrewintothenotchinthechassis.
Forlargersyringes£ PlacethescrewintheholeinthetopoftheshortlegoftheL.£ Placethespringovertheprotrudingendofthescrew.£ Insertthebottomoftheshorterlegintothenotchinthesideof
thechassis.£ Tightenthescrewintothenotchinthechassis.
Thespringallowstheclamptoriseautomaticallywhenyouloosenit,makingsyringeremovalmoreconvenient.
4
2.Loosenthescrewsonthetwosyringebarrelreversingclamps.Boththerightsyringemountandthepusherblockhaveblackreversingclampswithgrayscrewsatthetop.(SeeFigure1a.)Loosenallfourscrews.Ifthereversingclampofthepusherblockisontheleftfaceoftheblock,removethegrayscrewsandreplacetheclamplooselyontherightfaceoftheblock.
3.Placethesyringe.Laythesyringeinthetroughfacingoutofthechassiswiththebarrelflangebetweenthechassisanditsreversingclampandtheplungerflangebetweenthepusherblockanditsreversingclamp.
4.Tightenthescrews.Youmayhavetotightenallthescrewspartwaydownandthenreturntoretightenthem.Holdingthesyringeinplace,screwthesyringebarrelclampdownloosely.Tightenthescrewspart-wayonthereversingclamps.Wheneverythingislooselyheldinplace,squeezeeachreversingclampagainsttheflangeitisholdingandtightentofingertight.Thensqueezetheothersideofeachreversingclampandtightenit.Finally,returntothesyringeclampandmakeitfingertightaswell.
TheLeftSideTomountasyringeinoneoftheleft-sidemounts:
1.Selecttheproperclamporientation.Therearetwosyringebarrelclamps,astraightclampandanM-shapedclamp.Yourchoiceofclampwilldependuponthesizeofyoursyringe:
Forthesmallestsyringes–UsetheMclampwiththepointedsectionsfacingdown.Securetheclampwiththecentralscrew.Formid-sizesyringes–Usethestraightclamp.Securetheclampwiththecentralscrew.
Forlargersyringes–UsetheMclampwiththepointedsectionsfacingup.Securetheclampwiththecentralscrew.
Usethespringinthesamewayyouuseitforright-sidemounts.
5
2.Loosenthescrewsonthetwosyringebarrelreversingclamps.Boththerightsyringemountandthepusherblockhaveblackreversingclampswithgrayscrewsatthetop.(SeeFigure1c.)Loosenallfourscrews.Ifthereversingclampofthepusherblockisontheleftfaceoftheblock,removethegrayscrewsandreplacetheclamplooselyontherightfaceoftheblock.
3.Placethesyringe.Laythesyringeinthetroughfacingoutofthechassis,withthebarrelflangebetweenthechassisanditsreversingclamp,andtheplungerflangebetweenthepusherblockanditsreversingclamp.
4.Tightenthescrews.Youmayhavetotightenallthescrewspartwaydownandthenreturntoretightenthem.Holdingthesyringeinplace,screwthesyringebarrelclampdownloosely.Tightenthescrewspart-wayonthereversingclamps.Wheneverythingislooselyheldinplace,squeezeeachreversingclampagainsttheflangeitisholdingandtightentofingertight.Thensqueezetheothersideofeachreversingclampandtightenit.Finally,returntothesyringeclampandmakeitfingertight.
NOTE:Thefirsttimeyouuseanewsyringesizeorbrandmodel,youwillneedto:
£ mountthesyringe£ calibrateit£ removeit£ fillit£ mountitagain
Whenyousubsequentlyusethesamesyringe,youcansimplyselectthesyringefromtheGenieTouchsavedlook-uptableofsyringes,andusethe<<or>>buttonstopositionthepusherblockcorrectlybeforeyoumountthesyringe.Seepage18.
6
ControllingtheGenieTouch
TheTouchScreenTheGenieTouchmenuscreensaretouch-sensitive.Herearesometipsforusingthescreens:
£ Whentwochoicesappearsidebysideonthesamebutton,onlyoneortheothercanbeactiveatonetime.Theactivechoicewillbeyellowandtheinactivechoicewillbegrayed.
£ IfascreencontainsaSavebutton,youmusttapit
ifyouwanttosavethesettingsyouhavemade.
£ SomescreenscontainaDonebuttonwhichsavesyoursettingstotemporarymemoryandexitstothepreviousscreen.Inthosecases,youwillreturntoascreenwithaSavebutton.Tapittosavethefullconfigurationofsettingsonthatscreentonon-volatilememory.
£ SomescreenscontainaCancelbuttonwhichexitstothepreviousscreenwithoutsaving
yoursettings.
£ Numerickeypadscreens,whichallowyoutosetvaluesforvariousparameters,allcontainabottomrowwiththesebuttons:
• decimalpoint(•)whereappropriate• Savetoleaveandretainyourchanges• Deltodeletethelasttappednumberor
decimalpoint• Quittoleavewithnochangessaved
£ Atnumerickeypadscreens,ifyouattemptto
enteranumericvaluelargerthantheallowedmaximum,youwillseeanerrormessage.UsetheDelbuttontodeleteyourlastentryandtryagain.Theallowedmaximumvalueisalwaysdisplayedintheupperleftcornerofthescreen.
£ EveryscreencontainsaHelpbuttonforcontextsensitivehelp.
AcomputerinterfaceisalsoavailablefortheGenieTouch.VisittheGenieTouchproductpageatourwebsiteorseepage19tolearnmoreabouttheASCIIcommandsandhowtousethem.
www.kentscientific.com
7
GenieTouchHelpWhenyoutapHelpatanyscreen,youwillseecontext-sensitivehelpforthatscreen.Tapping TOC(TableofContents)atanycontext-sensitivehelpscreenproducesalistofOwner’sManualsections.Thelistitemsarenottouch-sensitive.HighlightthesectionyouwanttoviewandtapSelecttoviewit.TapDowntomovetowardtheendofthelistandUptomovetowardthetop.Atthehelppage,tapNextorPrevtopageforwardorbackwardthroughmulti-pageentries.TapReturntogobacktothemostrecentTableofContents.Tap TOC togobacktothemainTableofContents.
TheGenieTouchMenusWhenyouturntheGenieTouchon,theOpening Screenappears.Itcontainstwobuttons:
£ Setup: LeadstotheSetupscreenforconfiguringtheGenieTouch
£ Main: LeadstotheMainscreenforcontrollingtheGenieTouchwhileitisrunning£ SetupTheSetupscreencontainsfourmainbuttons:Syringe,Operation,Auto ReverseandAdvanced.Thefunctionofeachofthesebuttonsisdescribedbelow.
£ Setup > SyringeTheSyringeSelectionscreenThetopbuttonofthisscreenshowsthekindofsyringecurrentlyselected—GenericorBrand.Theselectedtypeappearsinyellow,andtheotherbuttonsonthescreendisplaythecurrentsettings,ifany,forthattypeofsyringe.TochangebetweenGenericandBrand,tapthebluechoice.Theotherbuttonsonthescreenwillchangetoreflectthetypeofsyringeyouhaveselected.
£ Syringe>GenericUseaGenericSyringeIfGenericisnotalreadythechosensyringetype,tapGeneric,thenfollowthesesteps,describedindetailbelow:
1. SetthesyringeVolume.
8
2. Seteitherthesyringe DiameterorLength.
3. Choosethesyringeorientation.
4. CalibratetheEmpty Position.
Whenyouarefinished,pressSave.
SetthesyringeVolume TapVolumetosetthevolumeforyourgenericsyringe.Attheresultingnumerickeypad,youmayenteravalueupto150ml.TaptheUnitsbuttoninthebottomrowtoswitchbetweenmlandul.Seteitherthe syringe DiameterorLength Tosetthediameter,tapDiameterandusetheresultingnumerickeypadtoenteranexactinsidediameterforyoursyringeinmm.Tosetthelengthinstead,tapLengthandfollowthesameprocess.Lengthisdefinedasthedistancebetweenthefullandemptypositions.Whenyouspecifyoneofthesevalues,GenieTouchwillcalculatetheotherbasedonyourVolumesetting.Choosethesyringeorientation TapFacing LeftorFacing RighttoindicatehowyouplantoorientthesyringeintheGenieTouch.Rightandleftrefertoyourrightorleft.NotethisselectiondefinesthephysicaldirectionfortheInfuseandWithdrawfunctions.(Re)CalibratetheEmpty Position Thebuttonforcalibrating(orrecalibrating)theEmpty Positiondoesnotappearuntilyouhavespecifiedalltheotherparametersforyoursyringe.Ifyouareusinganewkindofsyringe,thebuttonwillreadCalibrate Empty Positionandyoumustcompletetheprocessdescribedbelowbeforeyoursyringeisfullydefined.
Ifyouareusingasyringewithidenticaldimensionstooneyouhaveusedbefore,GenieTouchhassaveditscalibrationdatainthelookuptableandwillretrievethedataautomatically.ThebuttonwillreadRecalibrate Empty Position.Youshouldtapitonlyifyouwanttomakeachange.OneofthechangesyoucanmakeispressingErasetoremovethecurrentsyringefromtheGenieTouchlookuptable.Thetablecanhold48
Syringe Lengthmustbebetween1cmand12cmandsyringeDiametermustbebetween0.02mmand3cm.Ifyoutrytoinputnumbersoutsidetheseranges,youwillseeanerrormessage.
IfyouspecifyaDiameter/Volume combinationthatwouldcauseacalculatedLength tobeoutsidetheallowedrange, oraLength/VolumecombinationthatwouldcauseacalculatedDiametertobeoutsidetheallowedrange,thecalculatedvaluewilldisplayas Out of range.
9
differentGenericorBrandsyringes.Ifyouexceedthatnumber,theoldestentrywillbeerasedtomakeroomforthenewone.
Tocalibratetheemptypositionforyoursyringe:
1. Mountthesyringe.Seepage3fordirections.
2. TapCalibrate Empty Position.
3. Movethepusherblockuntilitiswithinabout0.2mmoftheendofthesyringeorlightlytouchingit.Usethe Fast << or Fast >>buttonstogetclosetothecorrectpositionandthenswitchtotheSlow Adjustbuttontofine-tunetheposition.Notethat Slow Adjustmovesonlytowardwhatyouhavedefinedasemptyforthissyringe.
4. PressSavetosaveyoursettingsorCanceltoreturntothepreviousscreenwithno
changes.
£ Syringe>BrandUseaBrandSyringeIfBrandisnotalreadythechosensyringetype,tapBrand.Thenfollowthesesteps,describedindetailbelow:
1. Choosethesyringebrand.
2. Selectthesyringevolume.
3. Calibratethe Empty Position.
Whenyouarefinished,pressSave.
CAUTION!Positiontheplungertobelightlytouchingtheendofthesyringebarrelortobeabout0.2mmfromtheendofthesyringe.Donotforcetheplungerintotheendofthesyringe.Thatwillcausethesyringeandplungertomoveandflexduringuseandresultininaccuraciesinplungermovement.Inaddition,ifyoudefinetheEmpty PositiontobetooclosetotheendofthebarrelandthenselectaMaximum Forceof100%(seepage16),theforcecouldeasilydamageorbreakadelicatesyringe.
10
ChoosethesyringebrandTapBrand.Attheresultingscreen,tapthenameofthesyringemanufacturer.SelectthesyringeVolume TapSelect Volume.Attheresultingscreen,choosefromthevolumesavailableforyourselectedmanufacturer.Youmustselectasyringebrandbeforeyouselectthevolume.CalibratetheEmpty Position SeedirectionsandCAUTIONnoteonpage7underSyringe>Generic.
£ Setup>Operation
TheOperationScreenThetopbuttonofthisscreenshowsthetypeofsyringeusecurrentlyselected—InfuseorWithdraw.Theselectedtypeappearsinyellow.TochangebetweenInfuseandWithdraw,tapthegrayedchoice.TheotherbuttonsonthescreendisplaytheavailablechoicesofOperationmodewiththecurrentsyringeuseselectioninyellow.
Foreithertypeofsyringeuse—InfuseorWithdraw—thefourchoicesofOperationmodearethesame:£ Constant Flow—drivesthepusherblockata
constantspeed£ Ramped Flow—drivesthepusherblockatalinearly
increasingordecreasingspeed£ Stepped Flow—drivesthepusheratanincreasingor
decreasingspeed,withthetotalchangedividedintoanumberofequalsteps
£ Pulsed Flow—drivesthepusherinaseriesoftwoalternatingcontinuousflowpulses
Eachofthefourchoicesleadstoascreenthatallowsyoutodefinethedetailsforthemode.Eachofthescreensisdiscussedbelow.
IfyouplantouseAuto Reversetoenableinfusionand/orwithdrawalofavolumegreaterthanthevolumeofthesyringe,youmustenableAuto Reversebeforeyouspecifythecharacter-isticsofyourOperation.Seepage14.
SyringeAbbreviationsAT: Air-TiteAllPlastic
HSWNorm-JectBD: BectonDickinson
Interm,WW,Plastipak
BDG: BectonDickinsonGlass—alltypes
CA: CadenceScienceMICRO-MATEglass
HA Hamilton1000-SeriesGastight
ME: MedallionSG: ScientificGlassEng
SGESH: SherwoodMonoject
PlasticTE: TerumoTE
11
£ Operation>ConstantFlowTapConstant FlowforascreenthatdisplayschoicesfordefiningyourOperation.TherearefourwaystodefineaConstant Flow Operation,eachofwhichrequiresyoutospecifytwoparameters.Inallcases,thepusherblockmovesataconstantspeed.
SpecifyVolumeandTimeTapBy Volume & TimeattheConstant Flowscreentospecifythetotalvolumeofliquidtobemovedandthelengthoftimeoverwhichtomoveit.Thecalculatedflowratewillthenbedisplayed.
TapVolumeandenterthevolumeofliquidtobemoved.Unitsswitchesbetweenul(microliters)andml(milliliters).
TapTimeandenterthetime.Unitscyclesbetweenseconds,minutesandhours.Usedecimalformat—1.5hror90min,forexample,setthesameTime.
SpecifyFlowandVolumeTapBy Flow & VolumeattheConstant Flowscreentospecifytheflowrateandthetotalvolumeofliquidtobemoved.Thecalculatedtimeforfulldeliverywillthenbedisplayed.
TapFlowandenterthevolumeofliquidtoflowperunittime.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.
TapVolumeandenterthetotalvolumeofliquidtobemoved.Unitscyclesbetweenul(microliters)andml(milliliters).
SpecifyFlowandTimeTapBy Flow & TimeattheConstant Flowscreentospecifytheflowrateandthetotaltimeoverwhichtheflowistobemaintained.Thecalculateddeliveryvolumewillthenbedisplayed.
TapFlowandenterthevolumeofliquidtoflowperunittime.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.
WhenyoutapanyofthebuttonsattheConstant Flowscreen,youwillseethenumerickeypadtoallowyoutoenteryourvalue.Enteryourvalueusingatmostfourdigitsandadecimalpoint.
12
TapTimeandenterthetotaldurationoftheOperation.Unitscyclesbetweenseconds,minutesandhours.Usedecimalformat—1.5hror90min,forexample,setthesameTime.
SpecifyFlow TapBy Flowatthe Constant Flowscreentospecifyaflowratetobemaintaineduntilthesyringeisemptyfor InfuseorfullforWithdraw.Notethatthesecondparameteryouarespecifyinghereisactuallyatimeof"untilfullorempty."YoucanusethisOperationmethodalongwithAuto Reverseforconstantpumping.Seepage14.
£ Operation>RampedFlow
TapRamped FlowforascreenthatdisplayschoicesfordefiningyourOperation.Youmustspecifythreeparameters—Start Flow Rate, End Flow RateandeitherTotal TimeorTotal Volume.Theprocess—eitherInfuseorWithdraw,dependinguponwhichyouhavechosen—beginsattheStart Flow Rateandincreasesordecreaseslinearlytothe End Flow Rateuntilthe Total VolumeismovedortheTotal Timeelapses.SpecifyStartFlowRateTapStart Flow RateattheRamped FlowscreentospecifytheflowrateatwhichtheOperationistobegin.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.
SpecifyEndFlowRateTapEnd Flow RateattheRamped FlowscreentospecifytheflowrateatwhichtheOperationistoend.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.SpecifyEitherTotalVolumeorTimeYoumustspecifyeitherthetotalvolumeofliquidtobemovedduringyouroperationorthetotallengthoftimeoverwhichliquidistobemoved.Either:
TapTotal VolumeattheRamped Flowscreentoenterthetotalvolumeofliquidtobemoved.Genie-TouchwillcalculatetheTotal Timeforyouroperationanddisplayitatthe Ramped Flowscreen.Unitsswitchesbetweenul(microliters)andml(milliliters).
WhenyoutapanyofthebuttonsattheRamped FlowscreenortheStepped Flowscreen,youwillseethenumerickeypadtoallowyoutoenteryourvalue.Enteryourvalueusingatmostfourdigitsandadecimalpoint.
Notethat0(zero)isanallowedvalueanytimeyouareaskedtoenterabeginningorendingflowrateforRamped FloworPulsed Flow.
13
orTapTotal TimeattheRamped Flowscreentoenterthelengthoftimeyouroperationistocontinue.GenieTouchwillcalculatetheTotal Volumeforyouroperationanddisplayitatthe Ramped Flowscreen.Units switchesbetweenseconds,minutesandhours.
£ Operation>SteppedFlow
TapStepped FlowforascreenthatdisplayschoicesfordefiningyourOperation.Youmustspecifyfourparameters—Start Flow Rate,End Flow Rate, Number of StepsandeitherTotal TimeorTotal Volume.Theprocess—eitherInfuseorWithdrawdependinguponwhichyouhavechosen—beginsattheStart Flow RateandincreasesordecreasestotheEnd Flow Rateinstepsofequaldurationuntilthe Total VolumeismovedortheTotal Timeelapses.SpecifyStartFlowRateTapStart Flow RateattheStepped FlowscreentospecifytheflowrateatwhichtheOperationistobegin.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.SpecifyEndFlowRateTapEnd Flow RateattheStepped FlowscreentospecifytheflowrateatwhichtheOperationistoend.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.SpecifyeitherTotalVolumeorTimeYoumustspecifyeitherthetotalvolumeofliquidtobemovedduringyouroperationorthetotallengthoftimeoverwhichliquidistobemoved.Either:
TapTotal VolumeattheStepped Flowscreentoenterthevolumetobemoved.GenieTouchwillcalculatetheTotal Timeforyouroperationanddisplayitatthe Stepped Flowscreen.Unitsswitchesbetweenul(microliters)andml(milliliters).
orTapTotal TimeattheStepped Flowscreentoenterthelengthoftimeyouroperationistocontinue.GenieTouchwillcalculatetheTotal Volumeforyouroperationanddisplayitatthe Stepped Flowscreen.Unitsswitchesbetweenseconds,minutesandhours.
SpecifyNumberofStepsTap Number of Stepsatthe Stepped Flowscreentospecifythenumberofstepsintowhichthe Operationwillbedivided.
14
£ Operation>PulsedFlowTapPulsed Flowforascreenthatdisplaysthechoicesfordefiningyour Operation.SincePulsed Flowismadeupofaseriesoftwo-partConstant Flow pulses,youmustspecifytwoparametersforeachofthepulsepartsandyoumustspecifytheNumber of PulsestobecarriedoutbeforetheOperationends.ConfigureFirstPartofthePulseTapPulse: 1st PartatthePulsed Flowscreentodefinetheparametersforthefirstpartofthepulseyouwanttouse.TheprocessisidenticaltoconfiguringaConstant Flow Operationwithoneexception:theBy Flowchoiceisnotavailable.Forcompletedirections,seeConstant Flowonpage11.
ConfigureSecondPartofthePulseTapPulse: 2nd PartatthePulsed Flowscreentodefinetheparametersforthesecondpartofthepulseyouwanttouse.TheprocessisidenticaltoconfiguringaConstant Flow Operationwithoneexception:theBy Flowchoiceisnotavailable.Forcompletedirections,seeConstant Flowonpage11.
SpecifyNumberofPulsesTap Number of Pulsesatthe Pulsed FlowscreentospecifyhowmanypulseswillmakeuptheOperation.YoumayenteranumberofpulsesoryoumaychooseContinuouswhichtellstheGenieTouchtocontinuepulsinguntilthesyringeiseitheremptyorfull,dependinguponwhetheryouhavechosenInfuseorWithdraw.
£ Setup > Auto Reverse TheAutoReverseScreenAuto Reversealternatesinfusionandwithdrawal.Itallowstheinfusionorwithdrawalofmorefluidthanasyringecanholdinasinglefilling.Auto ReversemaintainsallthecharacteristicsofyourOperation,whetheritisConstant,Ramped,SteppedorPulsed.Yoursettingforsyringeuse—InfuseorWithdraw—determineswhichofthetwowilloccurfirstandwillbethefocusoftheprocess.Forexample,ifyouhavechosenInfuse,theprocesswillbeginwithaninfusionaccordingtoyourOperationsettings,followedbya
YoucanuseAuto ReversetogetherwiththeContinuouschoiceforNumber of Pulsestoachievecontinuousoperationbeyondthevolumeofthesyringe.SeeAutoReversebelowformore.
15
constant-speedwithdrawtorefillthesyringe,andthenanotherinfusion.ThepumpwillchangedirectioneachtimetheendofthesyringeisreacheduntilitreachestheendofyourOperationasyouhavedefinedit.SeeOperationonpage10.
ThetopbuttonofthisscreenshowsthestatusofAuto Reverse—EnabledorDisabled.Thecurrentstatusappearsinyellow.Tochangethestatus,tapthegrayedchoice.Touse Auto Flow,youmustspecifyaVolumeandeitheraReverse FloworReverse Speed.SpecifyaFloworSpeedTapReverse Flowtoenteryourflowrate.Useatmostfourdigitsandadecimalpoint.Unitscyclesthroughml/min,nl/min,ul/minandml/hr.TapReverse Speedtoenterthe%offullpumpspeedtobeused.Enteratmostthreedigitsforavaluebetween1and100.SpecifyVolumeTapReverse Volumetoenteravolumeperpumpstroke.Thisvolumemustnotexceedthevolumeofthesyringeyouplantouse.
£ Setup > Advanced
TheAdvancedFeaturesSelectionScreenThisscreencontainsthreechoices:
£ Audible Alarms £ Maximum Force £ External Trigger
Eachleadstoanotherscreenwithchoicesforsettingupthecorrespondingfeature.
Important:TouseAuto Reverse,yoursyringemustbeequippedwithcheckvalvesforunidirectionalflow.Otherwise,Auto Reversewillsimplypushthesamesyringecontentsintoandoutofthesyringewhichcouldhaveundesirableconsequences.
NotethatifyoubeginyourOperation withapartiallyfilledsyringe,theinitialwithdrawalorinfusionstrokemaybeapartialone.
16
£ Advanced>AudibleAlarmsTapAudible AlarmsattheAdvancedscreentosetoneormoreconditionsthatwilltriggerthebeeperwhenyouarerunningtheGenieTouch.Buttonsforcurrentlyenabledalarmswillbeyellow.Thosefordisabledalarmswillbegrayed.WhenallfourAlarmsareconfiguredthewayyouwantthemtobe,tapSave.TurnAlarmOnwhenRunEndsTapRun Donetocyclebetweenenablinganddisablingthisalarm.Whenitison,aone-secondlongbeepsoundsattheendofanOperation.TurnAlarmOnwhenRunisPartiallyThroughTap% Run DonetoenterthepercentageofRuncompletionthatwilltriggerthreequarter-secondlongbeeps.Enteranumberbetween1and99.Toturnthisalarmoff,tap Alarm Off atthenumerickeypadscreen.TurnAlarmOnwhenMotorisStalledTapMotor Stalledtocyclebetweenenablinganddisablingthisalarm.Whenitison,fiveone-secondlongbeepswillsoundtoindicatethatthemotorisstalledandthusunabletomovethepusherblock.TurnAlarmOnwhenMotorSlipsTapMotor Slipping tocyclebetweenenablinganddisablingthisalarm.Whenitison,fourhalf-secondlongbeepswillsoundtoindicatethatthemotorisunderexcessiveloadandcannotmovethepusherblock.
£ Advanced>MaximumForce
TapMaximum ForceattheAdvancedscreentosetthelinearoperatingforceofthepusherblock.Enteryourselectionasapercentageofthemaximumforce,whichisover36kg(80lb).Fortypicaloperations,asettingof50%shouldbesatisfactory.
£ Advanced>ExternalTrigger
Tap External TriggerattheAdvancedscreentoconfiguretheuseofanoptionalexternalsignalasaneventtrigger.YoucanapplytheexternalsignalthroughtheAUXport(externalcontrol/remoteconnectorinFig1b)onthebackoftheGenieTouch.Thatsignalmightbe,forexample,afootpedal,aneventswitchoracomputerinterface.
TheminimumsettingforMaximum Forceis25%.ThisshouldbeusedwithcautiontoavoidstallingtheGenieTouchmotor.
17
ThetopbuttonontheExternal Triggerscreenshowsthecurrentstateofthetrigger—EnabledorDisabled.Thecurrentstatusappearsinyellow.Tochangethestatus,tapthegrayedchoice.ThethreebuttonsbelowallowyoutoconfiguretheGenieTouchbehaviorinresponsetothesignal.Youmustdefinewhether"off"istobeinterpretedasastoporapause,sethowGenieTouchwillbehaveonachangeinthesignalandspecifythenatureoftheelectricalsignalfromyourexternaltrigger.SetmeaningofOffTapOff = StoptodefineOffasacompletestopofyourOperation, equivalenttotappingtheStopbuttonontheRunscreen.Thisresetsthedisplayofthemovedandremainingvolumesdisplayedthere.TapOff = PausetodefineOffasaPause, equivalenttotappingthePausebuttonontheRunscreen.SetbehavioronchangeinsignalTapHold for OfftocauseaheldtriggertoeitherStoporPauseyourOperation,dependingonyoursettingat Off =above.Tap Tap for Run/Off tocauseataptoaswitchorfootpedaltocyclebetweenOffandRun,withOffdefinedatOff =above.SpecifynatureofsignalTapClosed = Offtodefinetheclosedorshorted-to-groundstateofyoursignalasOff.TapOpen = Off todefinetheopenorfloatingstateofyoursignalasOff.
18
£ MainTheMainscreenshowsthekindofsyringeinuseandthedetailsofthecurrentlydefinedOperation.Inaddition,itcontainsfivebuttons:RUN,Setup, << , >>,Help.Thefunctionofeachofthesebuttonsisdescribedbelow.
£ Main > RUN TapRUNtobeginyourOperation.TheRUNbuttonchangestoPauseandtheOperationstarts.TappingPausecausestheOperationtostoptemporarily.ThePausebuttonchangestoResume,allowingyoutopickupwheretheoperationleftoff.WhiletheGenieTouchisrunningorpaused,the << and >>buttonsdisappearandarereplacedwithStop.PressStoptoendtheOperationandresettheOperationparameterstotheirinitialstate.
£ Main > Setup TapSetupfortheSetupmenus.
£ Main > << and > > Tap<< tomovethepusherblocktotheleftand>> tomoveittotheright.Theblockbeginsmovingveryslowlyandacceleratesuntilitreachesasteadyspeed.Tokeeptheblockmoving,keepyourfingeronthebutton.Letgotostoptheblock.Youcanalsopressanddragyourfingerslowlyinthedirectionofthebuttonarrows.Whenyouliftyourfinger,thepusherwillcontinuemoving.Stopitbytappingeither<<or>>.
£ Main > Help TapHelpforhelpwiththecurrenttopic.TapNextorPrevtopageforwardorbackwardthroughmulti-pageentries.Whenyouarefinished,tapReturn.ToseealistofavailableOwner’sManualsectionstoviewontheGenieTouchscreen,tap TOC(TableofContents)atanycontext-sensitivehelpscreen.TapDownorUptohighlightthesectionyouwanttoviewandtapSelecttoviewit.Attheresultinghelppage(s),tapNextorPrevtopageforwardorbackwardthroughmulti-pageentries.TapReturntogobacktotheTableofContents.
ImportantTheRUNbuttondoesnotappearifyouhavenotcompletedtheSetupprocess.
19
GenieTouchPCInterfaceYoucandownloadComputerDriversforyourGenieTouchat:
http://www.ftdichip.com/Drivers/VCP.htm.ThesedriverswillallowyourcomputertocommunicatewiththeGenieTouchviaavirtualCOMport(VCP)connection.ForassistanceinsettingupyourcomputertocommunicatewiththeGenieTouch,pleasecontactKentScientific.
Notes§ AllcommandsaresentasASCIIcharacters.§ Aone-timepowerupcommandpromptissuppliedbytheinjectorInjector nnn.§ AllcommandsthatproduceerrorsareignoredandreturnanErrormessage.§ Commandsandargumentsarecaseinsensitive.§ Onlyonecommandperlineispermitted.§ Partsofcommandsincapsmustbesent;lowercaseportionsareoptional.§ Whitespacecanbeanynumberofspacesortabs.§ Alltextonalineafter‘!’isignored.§ Mostcommandsthatspecifyavalueorastatehavearequestformofthecommand
thatisthesamecommandprecededby?.
UnitDefinitionsLengthum micrometermm millimetercm centimeter
Volumeul microliterml millilitercc sameasml
Times,sec secondsm,min minutesh,hr hoursTimevaluescanbecombinedaspairs,e.g.3h20m.
Flowvu/tu volumeunitpertimeunitvu canbeanyoftheabovevolumeunitsandtucanbeanyoftheabovetimeunits.
COMPortSettings
Bitspersecond:115200DataBits:8Parity:NoneStopbits:1
20
CommandsAdHocASCIICommandsASCIIpacket:Rec:ASCIIchars,[CR|LF|CRLF]
:Xmit:‘>’,ASCIIchars,CR,LF
BEEpNoarg =>1secbeepBEEp Num Num=Numberof1secbeepsBEEp Num On ms Off ms
Num=NumberofbeepsofdurationOnmSandseparationofOffmS OnandOffaretruncatedinternallytoamultipleof10ms.
Example:bee 3 500 ms 1500msèThreebeeps,on0.5sec,off1.5sec CONtrol CONtrol [LOCK | UNLock]
LOCK:Enterlockedstate.AcceptnocommandsexceptUNLOCK UNLOCK:Releasefromlockedstate.
Example:conlockèPreventunintentionaltouchscreeninputREPort [RESet] [OFF | ON ] [ MOVing ] [ POS ] [ PERc ] [ VOL ] [ EVEnt ] [ period ] ?REPort [RESet] [ MOVing ] [ POS ] [ PERc ] [ VOL ] [ EVEnt ] ?REPort
§ Reportposition,percentage,volumeand/orevents.Reportingcanbeperiodic,byrequestortriggeredbyanevent.
§ The?formsrequestanimmediatesinglereporto Ifthereareargumentsthereportcontainstherequestedinformation.The
argumentsarenotretained.o Iftherearenoargumentsthereportcontainstheitemsthatwerelastsetupusing
thenon-? REPORTform.§ Thenon-?formspecifiesitemsthatareretainedinEEPROM.Ifanon-zeroperiodis
specifiedreportingwillbeautomaticandperiodic.§ POSistheabsolutecarriagepositioninxxx.xxmm§ PERCisthepercentcompletionasxx.xx%wheninRUNmode§ VOListhedeliveredvolumewheninRUNmode§ EVENTreportsignificantchangeinstatus
o Stallifmovementstoppedbystallo Slippingifmotorisnotabletomoveasfastasrequestedo Limitifmovementstoppedbylimitbeforereachingtargeto Endifmovementstoppedbyreachingtarget
21
§ MOVING restrictsreportingtowhenunit’scarriageisinmotion.Hasnoeffectona?report.
§ ONandOFFturnson/offperiodicandeventreporting,Hasnoeffectona?report.§ periodcanbeinanyoftheformsoftimebutisroundedtointegerseconds(thatcannot
exceed65535).Hasnoeffectfor?reporting.§ Ifnoargumentisgivenasinglereportisgeneratedusingthelastspecifiedarguments.§ RESETclearsallreportsettings.Itmustbethefirstargumentgivenorelseitwillclear
theargumentsthatprecedeit.§ AllreportsettingsarestoredinEEPROM
Example:rep on mov pos 1 secèReportpositioneverysecondwhilemovingRUNBeginrunningusingthecurrentconfigurationPAUsePauserunning.RUNwillun-pause.STOpStoprunning.RUNwillstartfrombeginning.Example:runèBeginrunningCLEar arg
§ arg=ALL:ClearSyr1,Syr2,Dispense,AutoRev§ arg=SYRinge:ClearSyringe§ arg=OPEration:ClearDispense,AutoRev§ arg=AUTorev:ClearAutoRev§ AllCLEarcommandswillforcethemotortostopifitisrunning
Example:cle syrèClearsyringevaluesDirectControlNotesaboutDirectControlCommandsForbothMOVeandSPEedpreviousvaluesarepreservedunlessexplicitlychangedbycommand.ToexitDirectControlissueaSTOP command.DefaultvaluesforallDIRECTCONTROLcmds:
§ DACissetto50%§ DestArgissettoPHYSICAL§ Speedissetto0
22
ABSpos § Ifmotornotmoving,updateabsolutemotorpositionfromcalibrationtableandrestart
backlashcorrection.§ Ifmotormoving,restartbacklashcorrection
Example:abs REFerence RefArg RefArg
§ WhereLEFtandRIGhtareimpliedandbothmustbepresent§ REFerence [LEFt RefArg] [RIGht RefArg] § RefArg=PHYsical|SYRinge|FULl|EMPty|#cm|#mm§ InthecaseofFULland EMPtytheLEFtandRIGhtareignoredandthesyringefacing
directiongovernsusage.Example: ref 6 cm 4 cmèSetleftreferenceto6cmandrightreferenceto4cmSPEed SpeedArgs
§ SpeedArgs=#%|#um/hr|stop[#F]§ #%,#um/hrandstopspecifyspeed§ #Fspecifiesforce,i.e.DACvaluewhere100>=DAC>=MinDAC
Example:spe 50% 100FèSetspeedto50%maximumandforceto100%maximumMOVe LEFt | RIGht | ABS DestArg [SpeedArgs]
§ IfLEFt orRIGht isgiventhen:o DestArg=UNLimited=>nolimittomotiono DestArg=PHYsical=>physicalendofallowedmotionwithouthittingendstopso DestArg=SYRinge=>syringefulloremptypositionbasedonsyringeorientationo DestArg=REFerence=>valuespecifiedbyREFerencecommand
§ IfABSisgiventhenDestArg=#mmor#cm=>absolutepositionrelativetorightEndstop.Absolutepositioncanbeoutsideofsyringelimitssocaremustbeexercisednottodamagethesyringe
Example:mov abs 5 cmèMovetolocation5cm(fromrightendstop)SyringeCommandThefollowingcommandspecifiessyringeselectionparameters.Allargumentsthathaveassociatedunitsmustbegivenwiththeirunits.NumericalValues
ForLengthorDiameterForm:xxxx,xxx.x,xx.xxorx.xxx
23
Units:um,mm,cmMustbeprecededbyLENorDIA
ForVolumeForm:xxxx,xxx.x,xx.xxorx.xxxUnits:ml,ulIsnotprecededbytext.Useonlythevalue,eg,40mlnotVOL 40ml
SYRinge Name Volume [RIGht | LEFt] [EmptyPos]
§ Thisformisusedtospecifyabrandnamedsyringe.§ Nameisanabbreviationofthemanufacturer’sname(see
listatright).§ Volumemustmatchoneofthestandardvolumes
providedbythespecifiedmanufacturer.SYRinge DIAmeter dia | LENgth len Volume [RIGht | LEFt] [EMPty pos]
§ Thisformisusedtospecifyagenericsyringe.§ EitherDiameteror Length—notboth—mustbespecified.§ RIGhtorLEFtspecifiesthedirection,i.e.RIGhtfacingor
LEFtfacing.IfneitherisgivenRIGhtfacingisassumed.§ EMPty posspecifiestheabsolutepositionthat
correspondstothesyringebeingempty.Itisspecifiedwithoutunitsandisassumedtobein10umunits.Itisforadvanceduseonly.Ifitisabsentthecurrentstoredvalueforposwillbeused.Tochangepos,useEMPandtheabsolutecarriagepositionin10umincrements.Forexample,syr dia 15mm 8ml rig emp.
Example:syr bd 60ml rightèSelectBD60mlbrandmountedfacingright.OperationCommandsThefollowingcommandsspecifydispensingparameters.Allargumentsthathaveassociatedunitsmustbegivenwiththeirunits.NumericalValuesForTimeForm:xxxx,xxx.x,xx.xxorx.xxxUnits:s,m,h,sec,min,hrorcompositeforms#h#mor#m#sLimit:Minimumtimeis100mSandresolutionis10mS
SyringeAbbreviationsAT: Air-TiteAllPlastic
HSWNorm-JectBD: BectonDickinson
Interm,WW,Plastipak
BDG: BectonDickinsonGlass—alltypes
CA: CadenceScienceMICRO-MATEglass
HA Hamilton1000-SeriesGastight
ME: MedallionSG: ScientificGlassEng
SGESH: SherwoodMonoject
PlasticTE: TerumoTE
24
ForVolumeForm:xxxx,xxx.x,xx.xxorx.xxxUnits:ml,ul
ForFlowForm:xxxx,xxx.x,xx.xxorx.xxxUnits:ml/min,ml/hr,ul/min,ul/hr
Definitions
§ “dir”indicatesoneofthekeywords:INFuseorWIThdraw.
§ GenVolindicatesaspecificvolume,Vol.
[WIThdraw | INFuse] Flow [WIThdraw | INFuse] Flow GenVol [WIThdraw | INFuse] Flow Time [WIThdraw | INFuse] Time GenVol
§ IfFlowistheonlyargumentthenContinuousdeliveryisassumed.§ Anytwoof Time,GenVolorFlowwillspecifyaspecificdeliveryamount.§ Argumentscanbespecifiedinanyorder.§ Allargsmusthaveassociatedunits.Theunitsareusedtobothspecifythequantityof
thevalueandtoidentifythetypeofthevalue.Example:inf in 10 ml/min 1 minèInfusewithaflowof10ml/minfor1min[WIThdraw | INFuse] RAMp {Time or GenVol} FlowBeg FlowEnd
§ DefinesalinearrampwithstartingflowofFlowBegandendingflowofFlowEnd.§ EitherTimeorGenVolcanbespecifiedtosetthetotalamountdelivered.§ FlowBegand FlowEndmustbegivenasapairandinthestatedorder.§ Time,GenVolandtheFlowpaircanbeingiveninanyorder.§ Oneoftheflowscanbezero.
Example:wit ram 50 sec 0ml/min 10ml/min èWithdrawfor50secwithalinearrampstartingataflowof0andendingwithaflowof10ml/min[WIThdraw | INFuse] STEp #Steps {Time or GenVol} FlowBeg FlowEnd
§ Definesasteppedramp.§ Theflowisuniformlydividedinto#stepsstartingatFlowBegandendingwithFlowEnd.§ The#Stepsvalueislimitedtoainteger2to9999anditmustbethe1stargumentafter
STEP.§ InallotherwaysthesyntaxisthesameasforRAMP.
25
Example:wit ste 4 50 sec 10ml/min 5ml/minèWithdrawfor50secwith4stepsstartingataflowof10ml/minandendingwithaflowof5ml/min[WIThdraw | INFuse] PULse #pulses Part1 Part2
§ Definesapulseddelivery.§ Apulseisdividedintotwoparts:Part1andPart2.§ ThespecificationofaPartmustproduceaflowrateforanon-zeroperiodoftime.§ Theflowratecanbezero.§ APartcanbespecifiedbyanyofpairs:{Flow&Time},{Flow&Volume}or{Time&
Volume}.§ The#pulsesislimitedtoanintegerof1to9999.§ #pulsecanbereplacedbythekeyword FOREVERinwhichcasedeliverywillcontinue
untilmanuallystopped.§ TheorderoftheargumentsforthePULSEcommandisfixed.
Example:wit pul 20 0ml/min 5 sec 10ml/min 5mlèPulsedwithdrawwith20pulses.Thefirstpartofthepulseiszeroflowfor5sec.Thesecondpartofthepulseis10ml/minforavolumeof5ml.Requests?CONtrol Returns:Locked/Unlocked?SYRinge
§ volumeisvolumeofsyringewithunits§ lengthislengthofsyringewithunits§ diameterisdiameterofsyringewithunits§ Returns:
o Syringe volume Len|Dia length|diameter [Left] o Syringe volume name [Left] o Syringe No Empty Pos o Syringe Undefined
?OPEration
§ dirisInfuseorWithdraw§ flow,beginFlow,endFlow,flow1orflow2areflowswithunits§ timeistimewithunits§ genVolisvolumewithunits § numisthenumberofstepsinrampornumberofpulses§ Returns:?
o dir Continuous flow
26
o dir Constant flow time o dir Constant flow genVol o dir Constant time genVol o dir Ramp beginFlow endFlow time o dir Ramp beginFlow endFlow volConc o dir Steps:num beginFlow endFlow time o dir Steps:num beginFlow endFlow genVolo dir Pulses:num flow1 flow2 time o dir Pulses:num flow1 flow2 genVol
§ dir Undefined?REPort [RESet] [ MOVing ] [ POS ] [ PERc ] [ VOL ] [ EVEnt ] ?REPort SeediscussionunderREPort ?RUN
§ percispercentcomplete.Sentasnnn.nn%.Omittedifcontinuous.§ Returns:RunState§ RunState=Stopped,Paused [perc],Running[perc],Direct,Fast,Undefined
?SN ReturnsSN: nnn.. Ver: v.vvwhere nnn..istheserialnumberandv.vv istheversionnumber
27
GeneralInformation
SpecificationsGenieTouchSyringenumber Upto2Syringesize 0.5ulto60mlUserInterface TouchscreenMaxLinearforce >36kg(80lb)SelectableFunctions ConstantRate
TimeDispensingVolumeDispensingRampPulsedStepped
ProgrammableFill/Empty YesPusherBlockStallDetection YesComputerInterface USBAudibleAlarmIndication YesPusherBlockPositionDetection
±0.2mm
Multi-SyringeRack OptionalForceAdjustment YesFootswitchInterface YesExternalTriggers YesWarranty 2years
CleaningInstructionsCleanthesurfaceofyourGenieTouchusingasoftlint-freecloth.Ifneeded,cleanwithaclothdampenedwithasolutionofwaterandamilddetergent,orisopropylalcoholmixedwithwaterata3:1ratio.
28
WarrantyThankyouforpurchasingaGenieTouch.Wetrulyappreciateyourbusiness.WestronglyadvisethatyoureadandstudythisOwner’sManualtoappreciatefullyallthefeatures,benefits,andcapabilitiesoftheGenieTouch.
ContactInformation KentScientificCorporation Email:[email protected] Phone:860.626.1172Torrington,Connecticut06790 Fax:860.626.1179www.kentscientific.comDisclaimerKentScientificCorporationmakesnorepresentationsorwarranties,expressed,statutoryorimplied,regardingthefitnessormerchantabilityofthecomponentsofthissystemforanyparticularpurpose.Further,KentScientificisnotliableforanydamages,includingbutnotlimitedto,lostprofits,lostsavings,orotherincidentalorconsequentialdamagesarisingfromownershiporuseoftheseproducts,orforanydelayintheperformanceofitsobligationsunderthewarrantyduetocausesbeyonditscontrol.InnocaseshallKentScientific'sfinancialobligationextendbeyondthecostoforreplacementoftheKentproductinquestion.KentScientificalsoreservestherighttomakeanyimprovementsormodificationstotheseproductsdescribedinthismanualatanytime,withoutnoticeofthesechanges.Allbrandandproductnamesusedinthismanualarethetrademarksoftheirrespectiveowners.Thisproductisnotdesigned,intendedorauthorizedforuseinhumanapplications.ProductWarrantyTheGenieTouchhasatwo(2)yearwarrantyincludingallpartsandlaborcharges.Thiswarrantydoesnotcoverdamagebyanycauseincluding,butnotlimitedto,anymalfunction,defectorfailurecausedbyorresultingfromunauthorizedserviceorparts,impropermaintenance,operationcontrarytofurnishedinstructions,shippingortransitaccidents,modificationsorrepairbytheuser,harshenvironments,misuse,neglect,abuse,accident,incorrectlinevoltage,fire,flood,othernaturaldisasters,ornormalwearandtear.ChangesormodificationsnotapprovedbyKentScientificCorporationcouldvoidthewarranty.Theforegoingisinlieuofallotherexpressedwarranties.KentScientificdoesnotassume,orauthorizeanypartytoassumeforit,anyotherobligationorliability.SatisfactionGuaranteeShouldyouexperiencedifficultywiththeGenieTouch,ourTechnicalSupportGroupwillassistyouintrouble-shootinganddeterminingiftheproductneedstobereturnedtoourfacility.WewillissueyouaReturnManufacturerAuthorization(RMA)numberbeforetheproductisshippedbackforrepair.Itisatthediscretionofthemanufacturertoreplaceorrepairadefectivepartorproduct.PleasecallCustomerServiceat888.572.8887toobtainaReturnManufacturerAuthorizationnumber.ShipmentswithoutanRMAnumberwillnotbeaccepted.Pleasenotethatafterour30-dayreturnpolicyperiodends,wewillbehappytoassistyouwithyourapplication,butcannotissueanycreditorrefundforareturnedGenieTouch.Priortoshipment,pleasecleananddecontaminatetheproductofanychemical,biological,orisotopiccontamination.PleaseincludeacompletedProductReturnFormwiththeshipment.Theformcanbefoundonpage29ofthisOwner’sManual.AdditionalcopiesareavailablebycallingKentScientificCustomerServiceat888.572.8887.