GKT-008 EMI Near Field Probe

USER MANUAL GW INSTEK PART NO. 82KT-00800EA1

ISO-9001 CERTIFIED MANUFACTURER

Page 2: GKT-008 EMI Near Field Probe User Manual

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will Corporation.

The information in this manual was correct at the time of printing. However, Good Will continues to improve its products and therefore reserves the right to change the specifications, equipment, and maintenance procedures at any time without notice.

Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.

Page 3: GKT-008 EMI Near Field Probe User Manual

TABLE OF CONTENTS

1

Table of Contents

SAFETY INSTRUCTIONS .................................. 2

MAIN FEATURES ............................................. 3 High Sensitivity ................................................................... 3 Ultra-Low Directionality ....................................................... 5 Good Far Field Frequency Response .................................. 10

EMI NEAR FIELD PROBE USES ...................... 11 Magnetic Field Probe ........................................................ 11 Electric Field Probe ........................................................... 12 AC Voltage Probe ............................................................... 13

APPLICATIONS .............................................. 14 EMI Test ............................................................................ 16 EMI M probe ..................................................................... 17 EMI Eprobe ....................................................................... 20 Voltage Probe .................................................................... 23 EMS Test ........................................................................... 25

SPECIFICATIONS ........................................... 26 PR-01 ................................................................................. 26 PR-02 ................................................................................. 26 ANT-04 .............................................................................. 27 ANT-05 .............................................................................. 27

Page 4: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

2

SAFETY INSTRUCTIONS Read the following before any operation to ensure your safety and to keep the device in the best condition.

Safety Symbols

These safety symbols may appear in this manual or on the device.

Attention: Refer to the Manual

DANGER High Voltage

Protective Conductor Terminal

DANGER Hot Surface

Earth (ground) Terminal

Page 5: GKT-008 EMI Near Field Probe User Manual

MAIN FEATURES

3

MAIN FEATURES

High Sensitivity The GKT-008 probe has a high sensitivity to quickly analyze a source of noise. The probe can be used with the dedicated EMI Pretest function on the GW Instek GSP-9300 spectrum analyzer. When used with the default pre-amplifier function, the sensitivity of the EMI probe can be significantly enhanced to distinguish subtle noise.

Test Results from a Common EMI Probe

Page 6: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

4

Test Results from the GKT-008 Probe

Page 7: GKT-008 EMI Near Field Probe User Manual

MAIN FEATURES

5

Ultra-Low Directionality Typically the design of near field ring probes cause measurement results to vary greatly when using the probes in a horizontal or vertical orientation. This often leads to incorrectly diagnosing a source of noise and therefore nullifying any corrective action. The special structure of the GKT-008 is designed to reduce the differences between the vertical and horizontal measurements, allowing noise to be measured accurately and effectively regardless of whether results are obtained in a horizontal or vertical position.

Measurement Differences in the Horizontal and Vertical Orientation of General Ring Probes(Case 1)

Page 8: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

6

Measurement Differences in the Horizontal and Vertical Orientation of General Ring Probes(Case 2)

Page 9: GKT-008 EMI Near Field Probe User Manual

MAIN FEATURES

7

Page 10: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

8

Small Measurement Differences in the Horizontal and Vertical Orientation When Using the GKT-008

Page 11: GKT-008 EMI Near Field Probe User Manual

MAIN FEATURES

9

Page 12: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

10

Good Far Field Frequency Response In general, test specifications are limited to 3M far field tests, but with tests performed using EMI probes to measure the near field electromagnetic field. However, the frequency response between far field & near field electromagnetic fields differ greatly, and thus when EMI near field probes are used to diagnose far field noise problems, test results can cause false positives and other misdiagnoses. Due to the structural design and electrical features of the GKT-008 probe, the measured electromagnetic wave in the near field has a strong correlation to the frequency response of the emitted far field electromagnetic wave.

Comparison of Measurement Results of Using the GKT-008 Probe in the Near Field and in 3M Far Field

Page 13: GKT-008 EMI Near Field Probe User Manual

EMI NEAR FIELD PROBE USES

11

EMI NEAR FIELD PROBE

USES

Magnetic Field Probe The magnetic field probe can analyze and diagnose the problems listed below:

Spatial noise of the device and equipment

Spatial noise of PCB boards

PCB trace and component radiated noise

Cable and wire radiated noise

Confirmation of shielding effectiveness

Page 14: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

12

Electric Field Probe The electric field probe can analyze and diagnose the problems listed below:

Ground and power noise

PCB trace conducted noise

IC pin output noise

I/O pin output noise

Confirmation of filter effectiveness

Page 15: GKT-008 EMI Near Field Probe User Manual

EMI NEAR FIELD PROBE USES

13

AC Voltage Probe The AC voltage probe can analyze and diagnose the problems listed below:

AC conducted Noise

Switching power supply noise and component noise

Page 16: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

14

APPLICATIONS This probe set should be used with GW Instek spectrum analyzers. When used in conjunction with the analyzer’s built-in application functions (EMI pretest and EMS Debug), the probes are able to diagnose EMI near field problems, estimate far fields as well as diagnose and analyze EMS.

When the EMC pretest button on the GSP-9300 is selected, the EMC mode will appear on the screen, and five common EMC functions will become available to select from.

Page 17: GKT-008 EMI Near Field Probe User Manual

APPLICATIONS

15

Page 18: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

16

EMI Test When the EMI test soft-key is selected, different test frequency bands will appear on the right of the screen. Select the desired test frequency band according to the EMI test requirements. If the desired test frequency band doesn’t appear on the screen, you can manually set the Start and Stop frequencies yourself.

Connect the probe to the “RF Input” terminal when performing EMI tests

The GSP-9300 has various EMI test functions available, including the ability to test in high noise areas by pressing the Amb Noise soft-key, the ability to enter the antenna factor settings according to the type of antenna by pressing the Correction soft-key as well as applying limit line specifications by pressing the Recall Limit soft-key.

Page 19: GKT-008 EMI Near Field Probe User Manual

APPLICATIONS

17

EMI M probe When you need to diagnose the energy distribution and source of spatial noise in products, the EMI probe as shown in the photo below can be used.

Note The measurement category and corresponding maximum measurement voltage for the ANT-04 or ANT-05 probe is CAT I 50Vdc.

They can’t be used for other measurement categories other than CAT I.

There are two EMI probe sizes in this probe set. According to the size of the area to be measured, the large probe can be used to measure over a large area and the small probe can be used to

Page 20: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

18

measure small areas of radiated emissions from IC components. Both probes have the same features for near field measurement.

When the EMI probe soft-key is selected, three commonly used frequency bands appear on the right of the screen. Select the desired frequency band according to the needs of the test. At this time you can sweep each area or each component one at a time to test for the highest sources of noise in the test product.

The GSP-9300 spectrum analyzer has a near field to far field conversion function that can be used when used in conjunction with these EMI probes. When the Correction soft-key is selected, noise in the near field can be converted to radiated field strength in the far field accordingly. This can this be used to determine whether the test product will be able to comply with the 3M far field test.

Page 21: GKT-008 EMI Near Field Probe User Manual

APPLICATIONS

19

Page 22: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

20

EMI Eprobe When you need to know the energy distribution and the source of noise in the PCB traces, power, ground or component pin conductors, you can use an electric field probe as shown below:

Note The measurement category and corresponding maximum measurement voltage for the PR-02 probe is CAT I 50Vdc.

It can’t be used in other measurement categories other than CAT I.

When the EMI EProbe soft-key is selected, two frequency bands will appear on the right of the screen for selection. Select a frequency band according the frequency of the noise to be analyzed. Touch the points to be measured and accessed, such as different ground points on the PCB board, with the tip of the electric field probe. The existing noise on each pin of the components as well as the size of noise on each I/O pin can all be easily diagnosed with the sensitivity of this electric field probe.

Page 23: GKT-008 EMI Near Field Probe User Manual

APPLICATIONS

21

When the EMI EProbe soft-key is selected, two frequency bands will appear on the right of the screen for selection. Select a frequency band according the frequency of the noise to be analyzed. Touch the points to be measured and accessed, such as different ground points on the PCB board, with the tip of the electric field probe. The existing noise on each pin of the components as well as the magnitude of noise on each I/O pin can all be easily diagnosed with the sensitivity of this electric field probe.

Select PCB trace Pin or I / O Cable Pin

Page 24: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

22

Turn Correction On

If PCB trace Pin is selected, select the parameter which best matches the trace length.

Page 25: GKT-008 EMI Near Field Probe User Manual

APPLICATIONS

23

Voltage Probe There is also an AC Voltage Probe in the GKT-008 probe set, shown in the photo below, which can be used as an effective and rapid tool for diagnosing and analyzing AC power conduction noise.

Note The measurement category and corresponding maximum measurement voltage for the PR-01 probe is CAT I/CAT II 300Vac.

Note When holding the probe during measurement, be sure to hold the probe below where the safety symbol is marked on the probe (the tip end of the probe)

In addition, you can perform pretesting of AC conduction noise when the probe is used with the built-in software on the GSP-9300.

When the Pretest soft-key is selected, the spectrum analyzer will automatically convert the AC noise measurements into an analog conduction value of a LISN. When the Debug soft-key is selected, each point or pin of the components on the PCB that are touched by the tip of the Voltage Probe can be output to find out the source of noise quickly. When diagnosing with the probe, please pay attention to its withstanding voltage. It can’t be used to touch points of AC voltage exceeding 300V, otherwise the probe and the spectrum analyzer can be damaged easily.

Note To further protect the spectrum analyzer from transient voltage surges, be sure that a transient limiter is used at the GSP-9300 RF input

Page 26: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

24

Transient limiter GPL-5010

Two kinds of axes are available

Page 27: GKT-008 EMI Near Field Probe User Manual

APPLICATIONS

25

EMS Test If the probe is used with the GSP-9300’s tracking generator, it allows you to perform debugging and diagnosis of EMS. The magnetic or electric field probe is connected to the TG (tracking generator) Output terminal of the spectrum analyzer. Select EMS Test and then select the desired EMS diagnostic frequency range. For more sensitive and vulnerable components on a PCB board, signal traces, ground lines or power lines, directly blanket the components with the signal to recognize and analyze where these products are subject to noise interference.

Connect the probe to the “Tracking Generator Output” terminal when performing EMS tests

Page 28: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

26

SPECIFICATIONS

PR-01

Frequency Range 150 kHz to 30 MHz

Primary Sensor E-field

Insertion Loss S21 / Frequency Response 5 dB / -3.6 dB (Avg.)

Deviation of Insertion Loss S21 ± 1 dB

Connector 50 Ω (SMA)

Input Resistance 10 MΩ

Rated Voltage 300 VAC

PR-02

Frequency Range 30 MHz to 3 GHz

Primary Sensor E-field

Insertion Loss S21 / Frequency Response 10 dB / -13.5 dB (Avg.)

Deviation of Insertion Loss S21 ± 3 dB

Connector 50 Ω (SMA)

Input Resistance 1 MΩ

Maximum Permissible Voltage at Probe Tip 50 VDC

Page 29: GKT-008 EMI Near Field Probe User Manual

SPECIFICATIONS

27

ANT-04

Frequency Range 30 MHz to 3 GHz

Primary Sensor H-field

Correction Factor / Frequency Response 10 dB / -26.5 dB (Avg.)

Deviation of Correction Factor ± 4 dB

Connector 50 Ω (SMA)

Maximum Input Power 0.5 W

ANT-05

Frequency Range 30 MHz to 3 GHz

Primary Sensor H-field

Correction Factor / Frequency Response 10 dB / -26.5 dB (Avg.)

Deviation of Correction Factor ± 4 dB

Connector 50 Ω (SMA)

Maximum Input Power 0.5 W

Page 30: GKT-008 EMI Near Field Probe User Manual

GKT-008 EMI Near Field Probe User Manual

28

Declaration of Conformity

We

GOOD WILL INSTRUMENT CO., LTD.

declare that the below mentioned product

Type of Product: EMI Probe Kit Set Model Number: GKT-008 (PR-01, PR-02, ANT-04, ANT-05)

are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to the EMC: 2014/30/EU, WEEE: 2012/19/EU and RoHS: 2011/65/EU. For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Directive, the following standards were applied: ◎Safety

Low Voltage Equipment Directive 2014/35/EU Safety Requirements EN 61010-031: 2002+A1: 2018