graphitized mesoporous carbon modified glassy carbon electrode for selective sensing of xanthine,...

Upload: sunilove

Post on 03-Apr-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    1/10

    Graphitized mesoporous carbon modified glassy carbon electrode for selective

    sensing of xanthine, hypoxanthine and uric acidRajendiran Thangaraj and Annamalai Senthil Kumar*

    Received 10th January 2012, Accepted 19th April 2012

    DOI: 10.1039/c2ay25029b

    An efficient electrochemical sensor for simultaneous electrochemical sensing of three purine

    compounds, uric acid (UA), xanthine (X) and hypoxanthine (HX), using a graphitized mesoporous

    carbon (GMC) modified glassy carbon electrode (GCE/GMC) has been demonstrated in pH

    7 phosphate buffer solution without any enzyme, prior to electrode activation, surfactant and sample

    pre-concentration step. Electrochemical investigation of the GCE/GMC with [Fe(CN)6]3 indicates

    metallic conductor like surface features of the modified electrode. A diffusion controlled reaction

    mechanism was identified for the electro-oxidation of the three purine compounds with an

    electrocatalytic pathway, except for the UA, where it shows a surface area effect with a mixed-diffusion

    and adsorption controlled mechanism at higher scan rates (>70 mV s1). Calculated full-width of the

    half maximum values for the simultaneous detection of the three purine compounds are 42, 53 and

    64 mV respectively and these are the lowest values ever reported in the literature, suggesting effective

    electron-transfer behaviour of the modified electrode for the purine oxidations. Calibration plots for

    the simultaneous detection of the purine compounds were linear in the concentration range of 20

    400 mM, 20320 mM and 20240 mM for UA, X and HX with detection limit values of 110 nM, 388 nM,

    and 351 nM respectively. Selective sensing of the purine compounds in human blood-plasma, urine and

    fish samples was successfully demonstrated with recovery values $100%.

    1. Introduction

    Xanthine (X), hypoxanthine (HX) and uric acid (UA) are

    important purine bases, where X and HX are the intermediates

    and UA is an end product of the purine metabolism of the cells

    of humans, animals and plants.13 When cells die, the purines in

    their genetic material break down. Purine catabolism of a dead

    fish is as follows: adenosine triphosphate / adenosine

    diphosphate / adenosine monophosphate / inosine mono-

    phosphate/ inosine/ HX/ X/ UA.13 Xanthine oxidase

    (XOD) is the specific enzyme present in the biological system,

    which helps to catalyze the HX and UA oxidation reactions.4

    Analyses of the intermediates and the end product can give

    vital information about the freshness of the fish foods.4

    Forexample, muscles of the freshly dead fish contain higher

    concentrations of HX and X than the end product, UA; while

    for the aged and decomposed fish, the concentration of UA is

    higher than that of X and HX.5 Similarly, abnormalities of the

    purine compounds present in humans are the clinical symptomsfor some critical diseases. For instance, increased excretion of

    the hypoxanthine and xanthine in urine is the indication of the

    inherited disease, Xanthinuria.6,7 Uric acid also serves as an

    antioxidant and helps to prevent damage to our blood vessel

    linings.8 For healthy individuals, the concentration of the UA

    present in the plasma is approximately 116500 mM and in the

    range of 1.24.5 mM in urinary excretion.9 In diseases such as

    gout, hyperuricemia and LeschNyhan syndrome, high levels of

    the UA are found in the blood samples.10 Meanwhile, relatively

    few foods contain concentrated amounts of the purines with

    high-protein content. Examples are: organ meats like kidney,

    fish, herring, sardines, mussels and yeast.11 Hence selective,

    sensitive and simultaneous sensing of UA, X and HX in bio-logical, food and clinical samples is of significant research

    interest in analytical chemistry.

    Conventional analytical methods available for the purine

    analysis are based on separation (HPLC or capillary electro-

    phoresis) coupled with UV-Vis or fluorescence spectroscopic

    techniques.1220 These methods are time consuming and require

    several off-line pre-concentration steps for the real sample

    analyses. Several electrochemical methods are available for the

    separation-less and enzyme-less analysis of the three purine

    compounds. Following are the recently reported working elec-

    trodes for the simultaneous detection of the three compounds:

    Environmental and Analytical Chemistry Division, School of AdvancedSciences, Vellore Institute of Technology University, Vellore-632 014,India. E-mail: [email protected]; Fax: +91-416-2243091/93;Tel: +91-416-2202754

    Electronic supplementary information (ESI) available: Effect ofsolution pH on simultaneous analysis of three purine bases atGCE/GMC (Fig. S1). See DOI: 10.1039/c2ay25029b

    2162 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012

    Dynamic Article LinksC

  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    2/10

    unmodified glassy carbon electrode (GCE),21 preanodized

    carbon paste electrode22 and functionalized single walled

    carbon nanotubecarbon paste electrode23 in acid pH solutions,

    and Nafion modified lead-ruthenate pyrochlore,24 Ru(DM-

    SO)4Cl2 electrodes,25 preanodized nontronite clay coated

    screen-printed electrode,26 electropolymerized poly(bromocresol

    purple) modified GCE,27 and impure multi-walled carbon

    nanotube (MWCNT).28 Most of the reported procedures either

    require expensive chemicals or require complicated electrodepreparation steps which include extensive pre-anodization of

    the underlying or modified electrode and the use of expensive

    starting materials such as Nafion, carbon nanotubes and

    ruthenium.24 Few chemically modified electrodes were also

    reported for the simultaneous detection of two purine

    compounds (UA and X) with examples being: 2-amino-1,3,4,-

    thiadiazole modified GCE,29 mesoporous SiO2graphite paste

    electrode30 and purified MWCNTdihexadecyl hydrogen

    phosphate surfactant modified electrode.31 Recently, a XOD

    enzyme biosensor constituted with functionalized graphene

    oxide + conducting polypyrrole graft copolymer + poly-

    (styrenesulfonic acid-g-pyrrole) modified electrode was reported

    for HX sensing in a physiological solution.32 In this work, weare reporting a commercially received graphitized mesoporous

    carbon (GMC) modified GCE (designated as GCE/GME)

    prepared by a much simpler method than the previously

    reported procedures without any enzyme, surfactant and acti-

    vation or functionalization of the electrode. The GCE/GMC is

    highly useful for a sensitive, selective and simultaneous vol-

    tammetric detection of the three purine compounds UA, X and

    HX in a physiological solution.

    Mesoporous materials in general have pores with an average

    diameter size (250 nm) in between the range of micro-porous

    (50 nm) materials. GMC is

    a porous carbon containing surface graphitic frameworks,

    prepared using certain soft polymers as a precursor and meso-porous silica materials (MSM) as a template by a high temper-

    ature pyrolysis procedure (non-conventional method of

    preparation).33 Structural features of the GMCs include homo-

    geneity with significant graphite-like domains and stacking

    heights, low amounts of surface functional groups, and low

    occurrences of imperfections such as twists, non-aromatic links,

    and carbon valencies.34 GMC has been demonstrated as a highly

    efficient support for Pt catalyst immobilization and for the

    enhanced direct methanol oxidation and oxygen reduction

    reactions suitable for fuel cell applications.35,36 Some of the

    conventional carbons, which are rich in carbonyl, phenolic and

    alcoholic functional groups, often result in severe oxidative

    corrosion during the electro-catalysis.37,38 But no such compli-cations were reported with the GMC. The GMC has been used as

    a catalyst for several organic functional group transformations

    including dehydrogenation of isobutane.39 The GMC material

    was rarely used for chemically modified electrode preparation

    and electroanalytical applications.40 Interestingly in this work,

    the GCE/GMC showed an efficient sensing of the three purine

    compounds without any ascorbic acid (AA) interference in

    a physiological solution. Finally, the modified electrode was

    successfully applied for simultaneous or selective sensing of the

    purine compounds in different kinds of clinical (blood and urine)

    and fish samples.

    2. Experimental

    2.1. Chemicals

    GMC (50 nm and 99.95% purity), MWCNT (outer diameter: 10

    15 nm; inner diameter: 26 nm, length 0.110 mm and $90%

    purity), uric acid, xanthine and hypoxanthine and ascorbic acid

    were of analytical grade and purchased from Sigma-Aldrich.

    They were used as received without any further purification.

    Aqueous solutions were prepared using deionized and alkaline

    KMnO4 distilled water (designated as DD water). Phosphate

    buffer solution (PBS, pH 7.1 0.05) of 0.1 M ionic strength was

    used as a supporting electrolyte.

    2.2. Measurements

    Voltammetric measurements were carried out using a CHI660C

    Electrochemical work station (USA) with 10 mL working

    volume. The electrode system consists of a glassy carbon elec-

    trode (GCE) and its chemically modified electrode (CME) as

    a working electrode (0.0707 cm2), Ag/AgCl (3 M KCl) as

    a reference electrode and platinum wire as a counter electrode.

    The surface of the GCE was cleaned both mechanically (polished

    with 0.05 micron alumina powder in a BAS polishing kit, cleaned

    with acetone and washed with double distilled water) and elec-

    trochemically by performing cyclic voltammetry (CV) continu-

    ously for 20 cycles in a potential window of0.2 to 1.2 V vs.

    Ag/AgCl in PBS. The GMC surface morphology was examined

    by transmission electron microscopy (TEM, JEOL-3010). A

    Bruker D8 Advanced diffractometer X-ray diffraction (XRD,

    Germany) instrument with Cu Ka source (l 1.5418 A) was

    used to probe the crystallographic information of the GMC.

    2.3. Preparation of the GCE/GMC modified electrode

    A GMCethanol suspension stock was first prepared by mixing 1mg of the GMC powder with 500 mL of absolute ethanol (99.9%)

    followed by 10 minutes sonication at room temperature (T

    300(2) K). For the GCE/GMC preparation, 5 mL of the stock

    was drop coated on the cleaned GCE surface and air dried for 10

    minutes at room temperature. The procedure allows for

    preparing a near uniform layer of GMC on the GCE surface. The

    as-prepared GMC layer was further electrochemically pretreated

    by continuous potential cycling (n 20, n no. of cycles) in the

    potential window of0.2 to 0.8 V vs. Ag/AgCl at a scan rate (v)

    of 50 mV s1 in pH 7 PBS. The GCE/GMC did not show any

    redox feature in the blank base electrolyte solution. Similar to the

    above procedure, GCE/MWCNT was also prepared by a drop

    casting methodology. Since dissolved oxygen did not influencethe electrochemical response, experiments were all carried out

    without any deaeration, as a physiological solution. The differ-

    ential pulse voltammetric (DPV) technique was used as a quan-

    titative electrochemical tool for the simultaneous detection of the

    purine bases X, HX and UA in pH 7 PBS.

    2.4. Real samples preparation

    Human blood plasma (#1(A)a, #1(A)b, #1B), healthy human

    urine (#2#4), red blood cell (RBC) and pus cell contaminated

    urine (#5#7) and fresh fish flesh (#8#11) samples were

    This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2163

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    3/10

    subjected to real sample analyses. Three healthy blood samples of

    volume 10 mL, where two samples belonging to the same person

    were collectedat 60 daytimeintervals(#1(A)a andb) with thehelp

    of a staff nurse from the Vellore Institute of Technology (VIT)

    health care centre. The three healthy human urine samples (#2

    #4) and three RBC/pus cell contaminated urine samples (#5#7)

    were also collectedfrom theVIT university healthcare centre. The

    RBC/pus cell counts in the urine samples are 1015/510, 1015/

    1015 and1520/36 respectivelyfor samples #5,#6 and#7. Priorto the analysis, the urine samples were filtered with ordinary filter

    paper and diluted (dilution factor 100) using PBS. Fresh fish

    samples: seer fish #8 (Scomberomorus guttatus), red snapper #9

    (Himantura bleekeri), goldstripe sardinella #10 (Sardinella gib-

    bosa), and carp fish #11 (semisulcatus) were purchased from

    a local animal food market in the Vellore city. All these samples

    were stored in a refrigerator at $5 C until use. Following is the

    procedure for the blood plasma sample preparation: 10 mL of the

    blood sample with a pinch of ethylenediamine tetraacetic acid

    (EDTA) was first centrifuged at 1600 rpm for about 20 minutes

    and the concentric plasma that appeared as a supernatant was

    taken for further analysis after proper dilution with PBS (dilution

    factor 5). For the fish real sample preparation, about 2 g of thefish flesh was weighed and homogenized with 10 mL of PBS using

    a mortar and pestle. The homogenized solution was filtered and

    further diluted with PBS (dilution factor 5). For all the real

    sampleanalyses,a 10 mLaliquotwas takenand examinedwiththe

    differential pulse voltammetric (DPV) technique. A standard

    addition approach was adopted for the quantification of the

    purine concentrations in the real samples. Since there is a weak

    adsorption of xanthine on the GCE/GMC surface, after the

    experiment with the analyte, the electrode was removed from the

    cell and immersed in 10 mL blank PBS for 10 seconds at room

    temperature. Such an experimental procedure gives easy regen-

    eration of the GCE/GMC for the successive electroanalytical

    measurements.

    3. Results and discussion

    3.1. Physicochemical characterization of the GMC

    TEM images of the GMC at different magnifications are given in

    Fig. 1AC. Randomly spaced empty clusters of uniform size

    $53 nm (outer diameter) are seen in the photographs. The black

    spots observed in the photographs are agglomerated units of the

    GMC. Details of the graphitic layers could be identified from the

    XRD analysis.

    The XRD pattern of the GMC was compared with another

    well-known crystalline carbon, MWCNT. Fig. 2 shows a well

    defined and intense diffraction peak at 2q 26.05 along with

    a weak peak at 43.02 that were obtained from the XRD of the

    GMC powder sample. The values correspond to the hklindex of

    {002} and {101} respectively for the hexagonal graphitic struc-

    ture on the mesoporous carbon. The control XRD pattern of the

    MWCNT material also shows similar 2q values at 26.07 and

    43.13. The calculated d-spacing (d002) value for the {002} peakusing the Braggs equation, d002 nl/2sin(q), where d002

    interplanar distance, lwavelength and q angle of diffraction,

    is 3.41 A for both the GMC and MWCNT samples. The values

    quoted in the literature for the unmodified mesoporous and

    graphitized mesoporous carbon (GMC) materials are 3.71 and

    3.36 A respectively.35,41 The 0.34 A reduction in the d-spacing is

    an indication of the enhanced crystallinity of the GMC

    compared to the MWCNT.

    3.2. Electrochemical characterization of the GCE/GMC

    The ferricyanide/ferrocyanide redox couple is a bench mark one

    electron-transfer electrochemical system, useful to access theconductive and the electron-transfer features of a modified

    electrode system. Fig. 3(A)a is the CV response of the GCE/

    GMC electrode in the presence of 1 mM of potassium ferri-

    cyanide at a scan rate of 10 mV s1. A well defined redox

    couple with anodic (Epa) and cathodic (Epc) peak potential

    values at 240 and 155 mV vs. Ag/AgCl respectively was noticed.

    Fig. 1 TEM images of GMC at different magnifications.

    Fig. 2 XRD patterns of graphitized mesoporous carbon and a multi-

    walled carbon nanotube.

    2164 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    4/10

    Calculated peak-to-peak potential (DEp Epa Epc) and

    apparent standard electrode potential (Eo0 Epa + Epc/2)

    values are 85 mV and 198 mV vs. Ag/AgCl respectively. For

    comparison, the unmodified GCE is also subjected to the

    ferricyanide experiment as in Fig. 3(A)b. Corresponding DEp

    and Eo0 obtained on the GCE are 86 and 200 mV vs. Ag/AgCl.The similarity in the DEp and E

    o0 values between the GCE and

    GCE/GMC indicates appreciable electron-transfer behavior of

    the GCE/GMC closer to the metallic conductive electrode.

    From the reversible redox reaction, the electrochemically

    accessible surface area, Ae (cm2), value could be calculated

    using the standard RandleSevcik equation: ipa or ipc 2.69

    105AeD1/2n3/2v1/2C, where n is the number of electrons involved

    in the redox reaction, D is the diffusion coefficient (7.6 106

    cm2 s1) and C is the concentration of the redox couple. Fig. 3B

    shows the plots of ipa vs. v1/2 for the ferricyanide at GCE and

    GCE/GMC. The GCE shows a linear line up to v 500 mV

    s1, while for the GCE/GMC the linearity is maintained up to

    70 mV s1, thereafter the line tends to saturate. The calculatedslope (ipa/v

    1/2) value (1.32 mA mV1/2 s1/2) is the same for the

    GCE and GCE/GMC. After substitution of the slope value in

    the RandlesSevcik equation, the Ae value is calculated to be

    0.0562 cm2, which is equal to $80% of the geometric surface

    area of the working electrode (0.0707 cm2). Note that GCE is

    atomically smooth and possesses near ideal surface features,

    whereas the GCE/GMC is the modified electrode with graphitic

    and non-graphitic porous carbon units. Even though the GMC

    has high surface area (BET surface area 900 m2 g1),33,34 in

    consideration with the Ae values, it can be speculated that only

    part of the total surface area of the electrodes is accessible for

    the electron-transfer reaction. The nature of surface polishing

    methods including the mechanical polishing of the electrodeand electrochemical pretreatments can change the surface area

    of the GCE. On the other hand, it is expected that there will be

    some diffusion restriction of the reduced and oxidized forms of

    the ferrocyanide complex within the GMC. This observation

    could be reflected by an improper (non-ideal) tail-off peak

    current and non-linearity in the ipa vs. v1/2 plot at v > 70 mV s1

    for the redox couple on the GCE/GMC working electrode,

    unlike the GCE case (Fig. 3(A)b). Nevertheless, due to the

    unique structural feature of the GMC, the electrochemical

    response of the GCE/GMC is remarkable and effective over the

    unmodified electrode for the purine compounds oxidation.

    3.3. Electrochemical and catalytic behaviours of the GCE/

    GMC

    Fig. 4AC show comparative CV responses of the GCE/GMC

    and unmodified GCE with UA, X and HX discretely at a scan

    rate of 50 mV s1 in pH 7 PBS. As can be seen from the CV of

    UA, the unmodified GCE shows a marked anodic oxidation

    peak at 295 mV vs. Ag/AgCl. The GCE/GMC also showed

    a similar peak potential value but with two and half timesenhancement in the anodic peak current value over the GCE.

    The observation may be due to the surface area effect of the

    GCE/GMC. For the case of X (Fig. 4B), the GCE/GMC

    showed a profound oxidation peak at 650 mV and was three

    times higher in the peak current value than the unmodified

    GCE. Note that the X oxidation peak potential at GCE is

    750 mV, which is 100 mV higher in over-potential than the

    GCE/GMC system. For the HX case (Fig. 4C), the unmodified

    electrode shows the absence of any CV oxidation peak up to

    1100 mV, whereas the GCE/GMC yielded a clear oxidation

    peak at 980 mV. In general, the surface area effect can be

    referred as when the modified electrode showed an enhanced

    oxidation/reduction current response to the analyte over itscorresponding unmodified electrode without altering the

    oxidation/reduction peak potential.42 On the other hand if the

    peak potential is altered, for example, the analyte oxidation

    potential is significantly reduced by the modified electrode

    (over-potential reduction with the symbol h) and it is then

    referred to as the electrocatalytic effect. In our case, X and

    HX have shown significant reductions in the h values, while for

    the case of the UA there is no such reduction in the h; however,

    only enhancement in the peak current value was noticed. Based

    on the information it can be concluded that X and HX

    oxidations are due to the electrocatalytic behaviour, while for

    the UA oxidation referred, it is due to the surface area effect of

    GCE/GMC.In continuation of the CV studies, DPV was also performed on

    the above purine compounds comparatively with GCE and

    GCE/GMC as in Fig. 4DF. All the three purine compounds

    show well-defined and sharp peak currents respectively at 250,

    580 and 900 mV vs. Ag/AgCl for UA, X and HX on the GCE/

    GMC. Unlike the CV behaviours, DPV of the unmodified GCE

    shows significant decreases in the peak current signals and high

    oxidation potential values, 460 and 780 mV for the UA and X

    respectively, and nil response to HX. Calculated h reductions for

    the UA and X are 200 and 220 mV respectively, in the DPV. But,

    no such observations were noticed in the CV (Fig. 4AC).

    Presumably, the applied amplitude (50 mV) and pulse effects

    (pulse width

    0.2 s; pulse period

    0.5 s) may have somehindrance to the electro-oxidation of the purine compounds on

    the GCE, while for the case with GCE/GMC a positive effect by

    the DPV parameters and hence considerable sharp peaks and

    reduction in the h were observed. Hence, it is a clear advantage of

    using DPV as the qualitative tool for sensitive analyses of the

    compounds in this work. Note that we have optimized the DPV

    parameters viz., pulse width, amplitude and pulse period with

    respect to the GMC modified electrode in order to get sharp

    voltammetric responses to the analytes. This is the reason why

    the corresponding unmodified GCE showed broad peaks in

    Fig. 4D and E.

    Fig. 3 (A) CV responses of bare GCE and GCE/GMC with 1 mM of

    K3[Fe(CN)6] in 0.5 M KCl solution at v 10 mV s1. (B) Plot ofipa vs. v

    1/2

    for the ferricyanide redox process at two different electrodes.

    This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2165

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    5/10

    In order to compare the efficiency of the electron-transfer

    behaviours (and in turn to the sharp peak currents), DPV peaks

    full-width of the half maximum (FWHM) value is taken as

    a sensitive parameter. Calculated FWHM values are 42, 53 and

    64 mV respectively for the UA, X and HX. The FWHM values

    for the recently reported Ru(DMSO)4Cl2Nafion modified

    electrode are 75, 75 and 100 mV respectively.25 The lowered

    FWHM values with the GCE/GMC are the clear evidence for the

    efficient peak current feature of the present system.

    Simultaneous analysis of the UA, X and HX on a GCE/GMC

    was nextexaminedby CVand DPV techniquesas inFig. 5AandBrespectively. Interestingly, the GCE/GMC shows well-defined

    and well-separated oxidation peak current responses. But the

    unmodified GCE displays a 13-fold decrease in the voltammetric

    peak current at theoxidation potentials of 400 and800 mV for the

    UAand X, andan absence of HX signal. The current signals areill

    definedtoo. Appearance of these current signals maybe dueto the

    response of overlapped UA, X and HX analytes. Calculated

    FWHM values for the analyte peaks from the simultaneous DPV

    measurements are UA 54, X 50 and HX 50 mV and these

    values were considerably smaller than the previously reported

    values of 59, 74, 86 mV on Ru(DMSO)4Cl2Nafion,25 and 65, 70,

    140 mV on poly(bromocresol purple) modified electrodes for the

    simultaneous oxidations of the three purines in a pH 7 PBS.27 All

    these experimental observations evidence the superior function of

    theGCE/GMC forthe simultaneousoxidations ofUA, X andHX

    in a physiological solution in this work.

    In order to checkthe nature of the electron-transfer mechanism,

    effects of CV scan rate (v) on the discrete oxidations of the three

    purine compounds were examinedas in Fig. 6AC. An increase inthe v implies a systematic increase in the peak currents for all the

    three compounds. Peak current valueswere measured andplotted

    against the scan rate (v) as a doublelogarithmic plot as inFig.6D.

    Calculated slope (vlog ipa/vlog v) values are: UA 0.3 (v < 70 mV

    s1) and 0.85 (v > 80 mV s1), X 0.30 and HX 0.30. These

    values are closer to the ideal values of 0.5 and 1 corresponding to

    diffusion and adsorption controlled electron-transfer mecha-

    nisms.43 Since theX hassome weak adsorption effecton theGCE/

    GMC (see the Experimental section), a slope value near 1 would

    be expected rather than $0.5 in this work. But the measured

    experimental value is $0.5, which indicates a diffusion controlled

    mechanism of the adsorbed X within the GCE/GMC film. The

    trend in the mechanism of the UA changes from diffusion toadsorption controlled reaction at scan rate >70 mV s1 (slope

    0.85),whichresembles the trend of theferricyanide case (Fig. 3A).

    Presumably, like ferricyanide, there will be some diffusion

    restriction for the UA within the pores of the GMC at a high scan

    rate, which in turn leads to change in the reaction mechanism.

    Exactdetailsfor the occurrence of suchreactionare still unknown.

    On the basis of Epa ([balog(v)]/2) + constant, for a totally

    irreversible diffusion-controlled process44,45 the ba (i.e., Tafel

    slope) value is measured as 10, 40 and 43 mV per decade for UA,

    X and HX oxidation reactions respectively on a GCE/GMC

    (Fig. 6E). Assuming one-electron transfer in the rate

    Fig. 4 Discrete CV (AC) and DPV (DF) responses of the electrochemical oxidations of UA, X and HX at GCE and GCE/GMC in a pH 7 PBS. DPV

    parameters are: amplitude 50 mV; increment potential 4 mV; pulse width 0.2 s; pulse period 500 s.

    Fig. 5 Simultaneous CV (A) and DPV (B) responses of GCE and GCE/

    GMC for a mixture of UA, X and HX dissolved in pH 7 PBS. CV scan

    rate (v) 50 mV s1. Other DPV parameters are as in Fig. 4.

    2166 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    6/10

    determination step (i.e., na 1), since ba 2.303RT/naFaa, the

    anodic-transfer coefficient (aa) can then be calculated as 5.9, 1.48

    and 1.37 for the UA, X and HX respectively. Previously, Zen

    et al. reported an aa value of$1 for an irreversible oxidation of

    cysteine on a lead-ruthenate pyrochlore modified electrode.46

    Hence, the high values obtained in the present work are an

    indication of the effective catalytic performance of the GCE/

    GMC to the purine oxidation reactions. Fig. 6F is a typical

    illustration for the electrochemical oxidation of the three purine

    compounds on the GCE/GMC.Fig. S1(A) shows the effect of solution pH (in the range pH 3

    8) on the simultaneous oxidations of the three analytes by the

    DPV. Plots of the respective anodic peak current and peak

    potential against the solution pH are displayed in Fig. S1(B) and

    S1(C) respectively. The peak current values increased positively

    against the solution pH and attained the maximum at pH $7.

    The peak potential values were systematically decreased against

    increase in the pH values. A slope value of60 mV per pH was

    calculated for all the three compounds on the GCE/GMC, sug-

    gesting the proton-coupled electron-transfer mechanism with

    equal number of proton/electron involvement in the reaction

    pathway (Nernstian case). Usually, depending on the pKa value

    of an analyte, there is a marked alteration in the trend of theoxidation peak potential value.24 Unfortunately, considering the

    pKa values of UA (5.4 and 10.3), X (7.4 and 11.2) and HX (8.94

    and 12.10), there were no alterations in the peak potentials with

    respect to the pKa values in this work. This observation further

    suggested that the protonation reaction is not a rate determining

    step in the electrochemical oxidation processes.

    3.4. Analytical measurements

    Effects of UA, X and HX concentrations (calibration) on the

    simultaneous detection of the individual analytes are next studied

    by systematically increasing one analyte concentration with the

    other two fixed as in Fig. 7AC. The optimal DPV parameters

    used in this work are: increment 4 mV, amplitude 50 mV,

    pulse width 0.2 s, pulse period 0.5 s and sampling width

    0.0167 s. Interestingly, systematic variations in the peak currents

    upon increases in the analyte concentrations were noticed. As

    seen in Fig. 7A, DPV of the UA shows regular increases in the

    peak current signals up to 400 mM with current sensitivity and

    regression coefficient values of 75.3 nA mM1 and 0.996 respec-

    tively. Ten repetitive measurements of 25 mM of UA result ina relative standard deviation (RSD) value of 0.88%. For the UA,

    the calculated detection limit (S/N 3) value is 110 nM.

    Analytical parameters for the X are: linearity up to 320 mM;

    sensitivity 61.8 nA mM1; regression 0.988, RSD 1.29%

    ([X] 25 mM, n 10) and detection limit 388 nM. Similarly for

    HX, the results are: linearity up to 240 mM; sensitivity 112.7

    nA mM1; regression 0.996, RSD 1.87% ([HX] 25 mM, n

    10) and detection limit 351 nM. Obtained detection limit

    values in this work are comparable to or even better than those

    previously reported in the literature. For instance, reported

    detection limit values for UA, X and HX on a nontronite clay

    coated screen-printed electrode with 2 V pre-anodization and 40

    s of pre-concentration are 0.42 mM, 7.0 mM and 0.34 mM,26

    poly(bromocresol purple) modified GCE; 0.20 mM, 0.06 mM and

    0.12 mM,27 NafionRu(DMSO)4Cl2; 0.37 mM, 2.35 mM and

    2.37 mM,25 and impure MWCNT; 0.141 mM, 0.134 mM, and 2.87

    mM,28 respectively. Simple preparation method and using rela-

    tively low cost carbon materials (other than the MWCNT) as

    a working electrode without any enzyme, activation and pre-

    concentration procedures are the superior features of the present

    working electrode over the existing CMEs. Note that the

    advantage of elimination of the ascorbic acid interference, unlike

    the MWCNT, was achieved with the GMC material (see Section

    3.5). With regard to a stable response of the GCE/GMC surface,

    Fig. 6 Effect of CV scan rate from 5500 mV s

    1 on the electrochemical oxidation of (A) UA, (B) X and (C) HX at GCE/GMC in a pH 7 PBS. Plots oflog(ipa) vs. log v (D) and Epa vs. log v (E) for the purine electrochemical oxidation reactions. (F) Typical illustration for the electrochemical oxidations of

    the three purines on the GCE/GMC.

    This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2167

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    7/10

    we expected that micro- and nano-porous voids might exist on

    the GCE surface (after the mechanical and electrochemical

    pretreatments); when GMC was modified as ethanol dispersion

    on the GCE surface, some of the portions of GMC might have

    occupied the porous sites of the underlying surface and thereby

    stabilizing the GCE/GMC. In order to check the reproducibility

    of the modified electrode, four modified electrodes were

    discretely prepared and subjected to DPV analysis of UA. The

    calculated RSD value for the peak currents is 3.06%, which

    indicates good reproducibility of preparation of the modified

    electrode.

    3.5. Interference study

    With the aim to extend the GCE/GMC for blood plasma purines

    analyses, we have examined the effect of a co-existing biochem-

    ical, AA, on the simultaneous detection of the purine samples.

    For a comparison, GCE/MWCNT was also subjected to the AA

    interference study (Fig. 8A). In a whole blood sample, common

    interferences for the UA are glucose, cholesterol, triglycerides,

    proteins, sodium chloride and AA. All these compounds, except

    AA, are electro-inactive. As from the clinical diagnosis data, the

    concentration range of the UA and AA present in a healthy

    humans blood plasma is 112506 mM (2.08.5 mg dL1) and AA

    is 40 15 mM, respectively.4749 Fig. 8B is a typical DPV response

    of the GCE/GMC for the simultaneous measurements of a fixed

    UA, X and HX content against variable AA concentrations in

    the range of 20500 mM. As can be seen, in the absence of AA,

    both GCE/MWCNT and GCE/GMC showed qualitatively

    similar DPV simultaneous responses. But in the presence of

    dilute AA, the GCE/MWCNT shows distinctly different vol-

    tammetric purine oxidation peaks. Plots of the three purinecompounds anodic peak current (ipa, base-line corrected)

    against the AA concentration, [AA], are given as insets in

    Fig. 8A. Unfortunately, addition of a dilute concentration of

    AA 20 mM leads to about a three fold increase in the purine

    oxidation peak currents and these currents get saturated after

    that. With regard to the AA oxidation peak at 0.1 V vs.

    Ag/AgCl, a marked peak current signal was noticed only when

    [AA] > 100 mM. Below 100 mM AA, large background currents

    were observed instead of any AA oxidation feature. These

    observations ruled out the choice of the GCE/MWCNT for the

    precise simultaneous analyses of the purine compounds in

    Fig. 7 Typical DPV responses for the simultaneous detection of UA, X

    and HX using a GCE/GMC in a pH 7 PBS. Plots of ipa vs. analyte

    concentration, [analyte], were given as the respective insets. DPV

    parameters are as in Fig. 4. Fig. 8 Effect of AA interference for the simultaneous DPV detection of

    the three fixed purine compound concentrations on (A) GCE/MWCNTand (B) GCE/GMC in pH 7 PBS. Respective ipa vs. [AA] plots were given

    as insets in the figure.

    2168 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    8/10

    the presence of dilute AA in a physiological solution. On the

    other hand, the GCE/GMC shows tolerable voltammetric

    current signals for the AA concentration up to 100 mM. In

    addition, the GCE/GMC did not show any AA oxidation peak

    current up to 100 mM (inset, Fig. 8B). Presumably the meso-

    porous structure has some negative interactive effects with the

    five member non-aromatic cored AA. These observations

    evidence that GCE/GMC is the better choice for the detection of

    uric acid in human blood plasma.

    3.6. Real sample analyses

    Practical applications of the GCE/GMC for the simultaneous or

    individual analyses of the three purine compounds in human

    blood plasma, urine and fish flesh samples were tested (Fig. 9 and

    Table 1). Typical DPV responses of the GCE/GMC for healthy

    humans blood plasma samples collected at sixty day time

    intervals (1st day: #1(A)a and 60th day: #1(A)b), are displayed in

    Fig. 9(A)a and b. Real sample analyses data are presented in

    Table 1. A marked DPV peak at 0.25 V vs. Ag/AgCl was noticed

    corresponding to the presence of the UA in the blood plasma

    samples. Standard spikes of the UA result in a systematic

    increase in the peak currents on the real sample peak. This

    observation qualitatively confirms the presence of the UA in the

    blood plasma samples. Based on the standard addition method,the UA concentrations in the blood samples were calculated as

    35.74 and 35.66 mM respectively for #1(A)a and #1(A)b. After

    correction of the dilution factor, the values turned out as 178.7

    and 178.3 mM, which are in the range of the standard UA level.9

    The observation of similar UA content values in those real

    Fig. 9 GCE/GMC responses for the real samples: (A) healthy human blood plasma samples (A)a and (A)b from the 1st and 60th day respectively, (B)

    healthy and (C) RBC/pus cell contaminated urine samples, and (D) a fresh fish sample in a pH 7 PBS. DPV parameters are as in Fig. 4.

    Table 1 Various purine real sample analyses using a GCE/GMC modified electrode by the standard addition method

    S. no. Real samples Analyte

    Detected value

    Added (mM) Found (mM) Recovery (%)Original Totalb

    1 Blood plasma (#1Aa)a UA 35.7 178.5 mM 20 20.44 102.22 Blood plasma (#1Ab)a UA 35.7 178.5 mM 20 20.73 103.73 Blood plasma (#1B) UA 63.7 318.5 mM 20 20.53 102.74 Urine (#2) X 1.9 95.0 mM 20 20.16 100.8

    UA 34.3 3.4 mM 30 31.15 103.85 Urine (#3) UA 37.0 3.7 mM 30 29.30 97.76 Urine (#4) UA 16.6 1.7 mM 20 19.67 98.47 Urine (#5)c UA 81.5 4.1 mM 50 49.97 99.98 Urine (#6)c UA 44.7 2.2 mM 50 50.22 100.49 Urine (#7)c UA 118.9 5.9 mM 50 49.57 99.110 Fish (#8) X 39.9 199.5 mM (760 mg g1) 30 30.30 101.011 Fish (#9) X 27.3 136.5 mM (520 mg g1) 60 62.40 104.012 Fish (#10) X 64.4 322.0 mM (1200 mg g1) 30 29.30 102.313 Fish (#11) X 34.5 172.5 mM (655 mg g1) 25 24.71 99.0

    a Collected from the same person. b After dilution factor correction. c Pus cell contaminated urine samples.

    This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2169

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    9/10

    samples proves the reliability of the present electrode for prac-

    tical analyses. In another case, for healthy human urine, sample

    #2, two peaks at 0.34 and 0.68 V corresponding to the UA and X

    were observed (Fig. 9B). The standard addition method quali-

    tatively confirms the presence of UA and X in the urine samples.

    Calculated UA and X concentration values are 318.7 and 96 mM

    respectively. In continuation, individual detection of UA present

    in two healthy humans urine samples (#3 and #4) was also

    examined and the values are 3.7 and 1.66 mM respectively andthese values are within the standard UA limits only.9 Fig. 9B is

    a DPV of the RBC/pus cell containing urine sample #5. Entries

    79 in Table 1 are the measured UA concentration values of the

    urine samples, #5#7. Since there is no precise literature avail-

    able to refer the RBC/pus cell count versus UA level in urine,

    based on the experimental results we are concluding here that

    there is no dependence of the excreted UA level on the RBC/pus

    cell values of the human urine.

    Finally four fresh fish samples (#8#11) were subjected to the

    DPV analyses. The samples showed two peak responses at 0.6

    and 0.95 V vs. Ag/AgCl corresponding to the X and HX

    compounds. All fish samples show ill defined HX peaks except

    fish sample #11. The standard addition of HX does not showincrements in the peak current. We suspect some matrix effect

    with the HX analysis and hence it could not be detected for some

    kinds of fish samples. Nevertheless, the X analysis content in the

    fish is vital to judge the quality of the fish. As seen in Table 1,

    detected X contents were in the order of 1200 > 760 > 655 > 520

    mg g1 for samples #10, #8, #11 and #9 respectively. The absence

    of UA with the animal food suggests appreciable freshness of the

    fish foods. Amongst the fish samples, sample #10 (Sardinella

    gibbosa) is found to be the richest content of X and hence these

    are the richest protein diet foods. Finally, calculated recovery

    values for the standard spikes of the purine samples all fall at

    $100%, which indicates the efficiency of the GCE/GMC in the

    real sample analyses.

    4. Conclusions

    The GMC modified glassy carbon electrode was developed for

    efficient and simultaneous electrochemical sensing of three

    purine compounds: uric acid, xanthine and hypoxanthine in

    a physiological solution. The GMC modified electrode shows

    well defined well separated and very sharp purine oxidation

    signals at around 0.3, 0.6 and 0.95 V vs. Ag/AgCl respectively for

    the compounds. The purine oxidation mechanisms follow the

    electrocatalytic pathway for the X and HX, while for the UA case

    it was due to the surface area effect. Investigation with the

    standard bench mark redox system, Fe(CN)63, showedimproper tail-off peak current behaviours due to diffusion

    restriction of the redox couple within the mesopores of the

    material. The effect of solution pH on the purine oxidation

    reactions obeyed the Nernstian type of proton-coupled electron-

    transfer reaction mechanism. Constructed calibration plots were

    linear up to 400, 320 and 240 mM of UA, X and HX respectively.

    Corresponding detection limit values were 110 nM, 388 nM and

    351 nM. Simultaneous or selective detection of the three purine

    compounds in blood-plasma, healthy urine samples and fish

    samples was successfully demonstrated. A matrix effect was

    noticed with some real sample analyses. Calculated recovery

    values for the analyses were $100%. Overall, the novelty of the

    present electrochemical sensor system lies in the simple prepa-

    ration steps using relatively low cost GMC materials for the

    separation-less analyses of the three purine bases without any

    enzyme, the pre-concentration procedure, the pre-anodization

    steps and the extendibility of the sensor to various kinds of real

    samples.

    Acknowledgements

    The authors are grateful for the financial support from the

    Department of Science and Technology (DST) under Tech-

    nology System Development Programme, India. We also thank

    Material Science Division, Indian Institute of Technology,

    Madras for the TEM measurements.

    References

    1 W. N. Kelly, McGraw, Harrisons Principles of Internal Medicine,New York, 10th edn, 1983, vol. 7, pp. 516524.

    2 D. P. Wootton and H. Freeman, Churchill Livingstone, Microanalysisin Medical Biochemistry, New York, 6th edn, 1982, pp. 7677.

    3 F. Cami~na, M. I. Novo-Rodriguez, S. Rodriguez-Segade andM. Castro-Gago, Clin. Chim. Acta, 1995, 243, 151164.

    4 M. A. Carsol, G. Volpe and M. Mascini, Talanta, 1997, 44, 21512159.

    5 E. Watanabe, Y. Tamada and N. H. Sato, Biosens. Bioelectron., 2005,21, 534538.

    6 K. Safranow and Z. Machoy, J. Chromatogr., B: Anal. Technol.Biomed. Life Sci., 2005, 819, 229235.

    7 C. Bory, C. Chantin and R. Boulieu, Clin. Biochem., 1997, 30, 247248.

    8 Sz. I. Varga, Z. Novak, L. Pataki, M. Patocskai and B. Matkovics,Clin. Chim. Acta, 1992, 205, 241244.

    9 R. T. Kachoosangi, C. E. Banks and R. G. Compton, Electroanalysis,2006, 18, 741747.

    10 V. V. S. Eswara Dutt and H. A. Mottola, Anal. Chem., 1974, 46,17771781.

    11 http://www.ama-assn.org/.

    12 J. Wynants, B. Petrov, J. Nijhof and H. V. Belle, J. Chromatogr., A,1987, 386, 297308.

    13 S. P. Ferraris, H. Lew and N. M. Elsayed, Anal. Biochem., 1991, 195,116121.

    14 M. Czauderna and J. Kowalczyk, J. Chromatogr., Biomed. Appl.,1997, 704, 8998.

    15 K. Safranow and Z. Machoy, J. Chromatogr., B: Anal. Technol.Biomed. Life Sci., 2005, 819, 229235.

    16 M. Kito, R. Tawa, S. Takeshima and S. Hirose, J. Chromatogr.,Biomed. Appl., 1983, 278, 3542.

    17 H. Lin, D.-K. Xu and H.-Y. Chen, J. Chromatogr., A, 1990, 760, 227233.

    18 A. Wang, L. Li, F. Zang and Y. Fang, Anal. Chim. Acta, 2000, 419,235242.

    19 T. Grune, G. A. Ross, H. Schmidt, W. Siems and D. Perrett, J.Chromatogr., A, 1993, 536, 105111.

    20 J. Wang, G. Chena, A. Muck Jr, D. Shin and A. Fujishima, J.

    Chromatogr., A, 2004, 1022, 207212.21 T. Yao, Y. Taniguchi, T. Wasa and S. Musha, Bull. Chem. Soc. Jpn.,

    1978, 51, 29372941.22 X. Cai, K. Kalcher and C. Neuhold, Fresenius J. Anal. Chem., 1994,

    348, 660665.23 Z. Wang, X. Dong and J. Li, Sens. Actuators, B, 2008, 131, 411416.24 J.-M. Zen, Y.-Y. Lai, G. Ilangovan and A. S. Kumar, Electroanalysis,

    2000, 12, 280286.25 A. S. Kumar and P. Swetha, J. Electroanal. Chem., 2010, 642, 135

    142.26 J.-M. Zen, Y.-Y. Lai, H.-H. Yang and A. S. Kumar, Sens. Actuators,

    B, 2002, 84, 237244.27 Y. Wang and L. L. Tong, Sens. Actuators, B, 2010, 150, 4349.28 A. Senthil Kumar and R. Shanmugam, Anal. Methods, 2011, 3, 2088

    2094.

    2170 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b
  • 7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin

    10/10

    29 P. Kalimuthu and S. Abraham John, Anal. Chim. Acta, 2009, 647, 97103.

    30 D. Sun, Y. Zhang, F. Wang, K. Wua, J. Chen and Y. Zhou, Sens.Actuators, B, 2009, 141, 641645.

    31 Y. Y. Sun, J. J. Fei, K. B. Wu and S. S. Hu, Anal. Bioanal. Chem.,2003, 375, 544549.

    32 J. Zhang, J. Lei, R. Pan, Y. Xue and H. Ju, Biosens. Bioelectron.,2010, 26, 371376.

    33 A. B. Fuertes and S. Alvarez, Carbon, 2004, 42, 30493055.34 M. Kruk, Z. Li and M. Jaroniec, Langmuir, 1999, 15, 14351441.

    35 X. Cui, F. Cui, Q. He, L. Guo, M. Ruan and J. Shi, Fuel, 2010, 89,372377.

    36 Y. Shao, S. Zhang, R. Kou, X. Wang, C. Wang, S. Dai,V. Viswanathan, J. Liu, Y. Wang and Y. Lin, J. Power Sources,2010, 195, 18051811.

    37 C.-C. Hung, P.-Y. Lim, J.-R. Chen and H. C. Shih, J. Power Sources,2011, 196, 140146.

    38 Y. Ishigami, K. Takada, H. Yano, J. Inukai, M. Uchida, Y. Nagumo,T. Hyakutake, H. Nishide and M. Watanabe, J. Power Sources, 2011,196, 30033008.

    39 H. Xie, Z. Wu, S. H. Overbury, C. Liang and V. Schwartz, J. Catal.,2009, 267, 158166.

    40 X. Lu, Y. Xiao, Z. Lei, J. Chen, H. Zhang, Y. Ni and Q. Zhang, J.Mater. Chem., 2009, 19, 47074714.

    41 C. D. Liang, S. Dai and G. Guiochon, Anal. Chem., 2003, 75, 49044912.

    42 R. L. McCreery, Chem. Rev., 2008, 108, 26462687.43 A. S. Kumar and P. Swetha, Colloids Surf., A, 2011, 384, 597

    604.44 H. Xia and H.-L. Li, J. Electroanal. Chem., 1997, 430, 183187.

    45 S. M. Golabi and H. R. Zare, J. Electroanal. Chem., 1999, 465, 168176.

    46 J.-M. Zen, A. S. Kumar and J.-C. Chen, Anal. Chem., 2011, 73, 11691175.

    47 Laboratory Test Handbook, ed. S. David, Lexi-Comp Inc., Cleveland,OH, 1996.

    48 N. W. Tietz, Text of Clinical Chemistry, W.B. Saunders, Philadelphia,PA, 1986.

    49 J. Lykkesfeldt, M. Viscovich and H. R. Poulsen, Free Radical Biol.Med., 2003, 35, 14391447.

    This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2171

    View Article Online

    http://dx.doi.org/10.1039/c2ay25029b