graphitized mesoporous carbon modified glassy carbon electrode for selective sensing of xanthine,...
TRANSCRIPT
-
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
1/10
Graphitized mesoporous carbon modified glassy carbon electrode for selective
sensing of xanthine, hypoxanthine and uric acidRajendiran Thangaraj and Annamalai Senthil Kumar*
Received 10th January 2012, Accepted 19th April 2012
DOI: 10.1039/c2ay25029b
An efficient electrochemical sensor for simultaneous electrochemical sensing of three purine
compounds, uric acid (UA), xanthine (X) and hypoxanthine (HX), using a graphitized mesoporous
carbon (GMC) modified glassy carbon electrode (GCE/GMC) has been demonstrated in pH
7 phosphate buffer solution without any enzyme, prior to electrode activation, surfactant and sample
pre-concentration step. Electrochemical investigation of the GCE/GMC with [Fe(CN)6]3 indicates
metallic conductor like surface features of the modified electrode. A diffusion controlled reaction
mechanism was identified for the electro-oxidation of the three purine compounds with an
electrocatalytic pathway, except for the UA, where it shows a surface area effect with a mixed-diffusion
and adsorption controlled mechanism at higher scan rates (>70 mV s1). Calculated full-width of the
half maximum values for the simultaneous detection of the three purine compounds are 42, 53 and
64 mV respectively and these are the lowest values ever reported in the literature, suggesting effective
electron-transfer behaviour of the modified electrode for the purine oxidations. Calibration plots for
the simultaneous detection of the purine compounds were linear in the concentration range of 20
400 mM, 20320 mM and 20240 mM for UA, X and HX with detection limit values of 110 nM, 388 nM,
and 351 nM respectively. Selective sensing of the purine compounds in human blood-plasma, urine and
fish samples was successfully demonstrated with recovery values $100%.
1. Introduction
Xanthine (X), hypoxanthine (HX) and uric acid (UA) are
important purine bases, where X and HX are the intermediates
and UA is an end product of the purine metabolism of the cells
of humans, animals and plants.13 When cells die, the purines in
their genetic material break down. Purine catabolism of a dead
fish is as follows: adenosine triphosphate / adenosine
diphosphate / adenosine monophosphate / inosine mono-
phosphate/ inosine/ HX/ X/ UA.13 Xanthine oxidase
(XOD) is the specific enzyme present in the biological system,
which helps to catalyze the HX and UA oxidation reactions.4
Analyses of the intermediates and the end product can give
vital information about the freshness of the fish foods.4
Forexample, muscles of the freshly dead fish contain higher
concentrations of HX and X than the end product, UA; while
for the aged and decomposed fish, the concentration of UA is
higher than that of X and HX.5 Similarly, abnormalities of the
purine compounds present in humans are the clinical symptomsfor some critical diseases. For instance, increased excretion of
the hypoxanthine and xanthine in urine is the indication of the
inherited disease, Xanthinuria.6,7 Uric acid also serves as an
antioxidant and helps to prevent damage to our blood vessel
linings.8 For healthy individuals, the concentration of the UA
present in the plasma is approximately 116500 mM and in the
range of 1.24.5 mM in urinary excretion.9 In diseases such as
gout, hyperuricemia and LeschNyhan syndrome, high levels of
the UA are found in the blood samples.10 Meanwhile, relatively
few foods contain concentrated amounts of the purines with
high-protein content. Examples are: organ meats like kidney,
fish, herring, sardines, mussels and yeast.11 Hence selective,
sensitive and simultaneous sensing of UA, X and HX in bio-logical, food and clinical samples is of significant research
interest in analytical chemistry.
Conventional analytical methods available for the purine
analysis are based on separation (HPLC or capillary electro-
phoresis) coupled with UV-Vis or fluorescence spectroscopic
techniques.1220 These methods are time consuming and require
several off-line pre-concentration steps for the real sample
analyses. Several electrochemical methods are available for the
separation-less and enzyme-less analysis of the three purine
compounds. Following are the recently reported working elec-
trodes for the simultaneous detection of the three compounds:
Environmental and Analytical Chemistry Division, School of AdvancedSciences, Vellore Institute of Technology University, Vellore-632 014,India. E-mail: [email protected]; Fax: +91-416-2243091/93;Tel: +91-416-2202754
Electronic supplementary information (ESI) available: Effect ofsolution pH on simultaneous analysis of three purine bases atGCE/GMC (Fig. S1). See DOI: 10.1039/c2ay25029b
2162 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012
Dynamic Article LinksC
-
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
2/10
unmodified glassy carbon electrode (GCE),21 preanodized
carbon paste electrode22 and functionalized single walled
carbon nanotubecarbon paste electrode23 in acid pH solutions,
and Nafion modified lead-ruthenate pyrochlore,24 Ru(DM-
SO)4Cl2 electrodes,25 preanodized nontronite clay coated
screen-printed electrode,26 electropolymerized poly(bromocresol
purple) modified GCE,27 and impure multi-walled carbon
nanotube (MWCNT).28 Most of the reported procedures either
require expensive chemicals or require complicated electrodepreparation steps which include extensive pre-anodization of
the underlying or modified electrode and the use of expensive
starting materials such as Nafion, carbon nanotubes and
ruthenium.24 Few chemically modified electrodes were also
reported for the simultaneous detection of two purine
compounds (UA and X) with examples being: 2-amino-1,3,4,-
thiadiazole modified GCE,29 mesoporous SiO2graphite paste
electrode30 and purified MWCNTdihexadecyl hydrogen
phosphate surfactant modified electrode.31 Recently, a XOD
enzyme biosensor constituted with functionalized graphene
oxide + conducting polypyrrole graft copolymer + poly-
(styrenesulfonic acid-g-pyrrole) modified electrode was reported
for HX sensing in a physiological solution.32 In this work, weare reporting a commercially received graphitized mesoporous
carbon (GMC) modified GCE (designated as GCE/GME)
prepared by a much simpler method than the previously
reported procedures without any enzyme, surfactant and acti-
vation or functionalization of the electrode. The GCE/GMC is
highly useful for a sensitive, selective and simultaneous vol-
tammetric detection of the three purine compounds UA, X and
HX in a physiological solution.
Mesoporous materials in general have pores with an average
diameter size (250 nm) in between the range of micro-porous
(50 nm) materials. GMC is
a porous carbon containing surface graphitic frameworks,
prepared using certain soft polymers as a precursor and meso-porous silica materials (MSM) as a template by a high temper-
ature pyrolysis procedure (non-conventional method of
preparation).33 Structural features of the GMCs include homo-
geneity with significant graphite-like domains and stacking
heights, low amounts of surface functional groups, and low
occurrences of imperfections such as twists, non-aromatic links,
and carbon valencies.34 GMC has been demonstrated as a highly
efficient support for Pt catalyst immobilization and for the
enhanced direct methanol oxidation and oxygen reduction
reactions suitable for fuel cell applications.35,36 Some of the
conventional carbons, which are rich in carbonyl, phenolic and
alcoholic functional groups, often result in severe oxidative
corrosion during the electro-catalysis.37,38 But no such compli-cations were reported with the GMC. The GMC has been used as
a catalyst for several organic functional group transformations
including dehydrogenation of isobutane.39 The GMC material
was rarely used for chemically modified electrode preparation
and electroanalytical applications.40 Interestingly in this work,
the GCE/GMC showed an efficient sensing of the three purine
compounds without any ascorbic acid (AA) interference in
a physiological solution. Finally, the modified electrode was
successfully applied for simultaneous or selective sensing of the
purine compounds in different kinds of clinical (blood and urine)
and fish samples.
2. Experimental
2.1. Chemicals
GMC (50 nm and 99.95% purity), MWCNT (outer diameter: 10
15 nm; inner diameter: 26 nm, length 0.110 mm and $90%
purity), uric acid, xanthine and hypoxanthine and ascorbic acid
were of analytical grade and purchased from Sigma-Aldrich.
They were used as received without any further purification.
Aqueous solutions were prepared using deionized and alkaline
KMnO4 distilled water (designated as DD water). Phosphate
buffer solution (PBS, pH 7.1 0.05) of 0.1 M ionic strength was
used as a supporting electrolyte.
2.2. Measurements
Voltammetric measurements were carried out using a CHI660C
Electrochemical work station (USA) with 10 mL working
volume. The electrode system consists of a glassy carbon elec-
trode (GCE) and its chemically modified electrode (CME) as
a working electrode (0.0707 cm2), Ag/AgCl (3 M KCl) as
a reference electrode and platinum wire as a counter electrode.
The surface of the GCE was cleaned both mechanically (polished
with 0.05 micron alumina powder in a BAS polishing kit, cleaned
with acetone and washed with double distilled water) and elec-
trochemically by performing cyclic voltammetry (CV) continu-
ously for 20 cycles in a potential window of0.2 to 1.2 V vs.
Ag/AgCl in PBS. The GMC surface morphology was examined
by transmission electron microscopy (TEM, JEOL-3010). A
Bruker D8 Advanced diffractometer X-ray diffraction (XRD,
Germany) instrument with Cu Ka source (l 1.5418 A) was
used to probe the crystallographic information of the GMC.
2.3. Preparation of the GCE/GMC modified electrode
A GMCethanol suspension stock was first prepared by mixing 1mg of the GMC powder with 500 mL of absolute ethanol (99.9%)
followed by 10 minutes sonication at room temperature (T
300(2) K). For the GCE/GMC preparation, 5 mL of the stock
was drop coated on the cleaned GCE surface and air dried for 10
minutes at room temperature. The procedure allows for
preparing a near uniform layer of GMC on the GCE surface. The
as-prepared GMC layer was further electrochemically pretreated
by continuous potential cycling (n 20, n no. of cycles) in the
potential window of0.2 to 0.8 V vs. Ag/AgCl at a scan rate (v)
of 50 mV s1 in pH 7 PBS. The GCE/GMC did not show any
redox feature in the blank base electrolyte solution. Similar to the
above procedure, GCE/MWCNT was also prepared by a drop
casting methodology. Since dissolved oxygen did not influencethe electrochemical response, experiments were all carried out
without any deaeration, as a physiological solution. The differ-
ential pulse voltammetric (DPV) technique was used as a quan-
titative electrochemical tool for the simultaneous detection of the
purine bases X, HX and UA in pH 7 PBS.
2.4. Real samples preparation
Human blood plasma (#1(A)a, #1(A)b, #1B), healthy human
urine (#2#4), red blood cell (RBC) and pus cell contaminated
urine (#5#7) and fresh fish flesh (#8#11) samples were
This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2163
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
3/10
subjected to real sample analyses. Three healthy blood samples of
volume 10 mL, where two samples belonging to the same person
were collectedat 60 daytimeintervals(#1(A)a andb) with thehelp
of a staff nurse from the Vellore Institute of Technology (VIT)
health care centre. The three healthy human urine samples (#2
#4) and three RBC/pus cell contaminated urine samples (#5#7)
were also collectedfrom theVIT university healthcare centre. The
RBC/pus cell counts in the urine samples are 1015/510, 1015/
1015 and1520/36 respectivelyfor samples #5,#6 and#7. Priorto the analysis, the urine samples were filtered with ordinary filter
paper and diluted (dilution factor 100) using PBS. Fresh fish
samples: seer fish #8 (Scomberomorus guttatus), red snapper #9
(Himantura bleekeri), goldstripe sardinella #10 (Sardinella gib-
bosa), and carp fish #11 (semisulcatus) were purchased from
a local animal food market in the Vellore city. All these samples
were stored in a refrigerator at $5 C until use. Following is the
procedure for the blood plasma sample preparation: 10 mL of the
blood sample with a pinch of ethylenediamine tetraacetic acid
(EDTA) was first centrifuged at 1600 rpm for about 20 minutes
and the concentric plasma that appeared as a supernatant was
taken for further analysis after proper dilution with PBS (dilution
factor 5). For the fish real sample preparation, about 2 g of thefish flesh was weighed and homogenized with 10 mL of PBS using
a mortar and pestle. The homogenized solution was filtered and
further diluted with PBS (dilution factor 5). For all the real
sampleanalyses,a 10 mLaliquotwas takenand examinedwiththe
differential pulse voltammetric (DPV) technique. A standard
addition approach was adopted for the quantification of the
purine concentrations in the real samples. Since there is a weak
adsorption of xanthine on the GCE/GMC surface, after the
experiment with the analyte, the electrode was removed from the
cell and immersed in 10 mL blank PBS for 10 seconds at room
temperature. Such an experimental procedure gives easy regen-
eration of the GCE/GMC for the successive electroanalytical
measurements.
3. Results and discussion
3.1. Physicochemical characterization of the GMC
TEM images of the GMC at different magnifications are given in
Fig. 1AC. Randomly spaced empty clusters of uniform size
$53 nm (outer diameter) are seen in the photographs. The black
spots observed in the photographs are agglomerated units of the
GMC. Details of the graphitic layers could be identified from the
XRD analysis.
The XRD pattern of the GMC was compared with another
well-known crystalline carbon, MWCNT. Fig. 2 shows a well
defined and intense diffraction peak at 2q 26.05 along with
a weak peak at 43.02 that were obtained from the XRD of the
GMC powder sample. The values correspond to the hklindex of
{002} and {101} respectively for the hexagonal graphitic struc-
ture on the mesoporous carbon. The control XRD pattern of the
MWCNT material also shows similar 2q values at 26.07 and
43.13. The calculated d-spacing (d002) value for the {002} peakusing the Braggs equation, d002 nl/2sin(q), where d002
interplanar distance, lwavelength and q angle of diffraction,
is 3.41 A for both the GMC and MWCNT samples. The values
quoted in the literature for the unmodified mesoporous and
graphitized mesoporous carbon (GMC) materials are 3.71 and
3.36 A respectively.35,41 The 0.34 A reduction in the d-spacing is
an indication of the enhanced crystallinity of the GMC
compared to the MWCNT.
3.2. Electrochemical characterization of the GCE/GMC
The ferricyanide/ferrocyanide redox couple is a bench mark one
electron-transfer electrochemical system, useful to access theconductive and the electron-transfer features of a modified
electrode system. Fig. 3(A)a is the CV response of the GCE/
GMC electrode in the presence of 1 mM of potassium ferri-
cyanide at a scan rate of 10 mV s1. A well defined redox
couple with anodic (Epa) and cathodic (Epc) peak potential
values at 240 and 155 mV vs. Ag/AgCl respectively was noticed.
Fig. 1 TEM images of GMC at different magnifications.
Fig. 2 XRD patterns of graphitized mesoporous carbon and a multi-
walled carbon nanotube.
2164 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
4/10
Calculated peak-to-peak potential (DEp Epa Epc) and
apparent standard electrode potential (Eo0 Epa + Epc/2)
values are 85 mV and 198 mV vs. Ag/AgCl respectively. For
comparison, the unmodified GCE is also subjected to the
ferricyanide experiment as in Fig. 3(A)b. Corresponding DEp
and Eo0 obtained on the GCE are 86 and 200 mV vs. Ag/AgCl.The similarity in the DEp and E
o0 values between the GCE and
GCE/GMC indicates appreciable electron-transfer behavior of
the GCE/GMC closer to the metallic conductive electrode.
From the reversible redox reaction, the electrochemically
accessible surface area, Ae (cm2), value could be calculated
using the standard RandleSevcik equation: ipa or ipc 2.69
105AeD1/2n3/2v1/2C, where n is the number of electrons involved
in the redox reaction, D is the diffusion coefficient (7.6 106
cm2 s1) and C is the concentration of the redox couple. Fig. 3B
shows the plots of ipa vs. v1/2 for the ferricyanide at GCE and
GCE/GMC. The GCE shows a linear line up to v 500 mV
s1, while for the GCE/GMC the linearity is maintained up to
70 mV s1, thereafter the line tends to saturate. The calculatedslope (ipa/v
1/2) value (1.32 mA mV1/2 s1/2) is the same for the
GCE and GCE/GMC. After substitution of the slope value in
the RandlesSevcik equation, the Ae value is calculated to be
0.0562 cm2, which is equal to $80% of the geometric surface
area of the working electrode (0.0707 cm2). Note that GCE is
atomically smooth and possesses near ideal surface features,
whereas the GCE/GMC is the modified electrode with graphitic
and non-graphitic porous carbon units. Even though the GMC
has high surface area (BET surface area 900 m2 g1),33,34 in
consideration with the Ae values, it can be speculated that only
part of the total surface area of the electrodes is accessible for
the electron-transfer reaction. The nature of surface polishing
methods including the mechanical polishing of the electrodeand electrochemical pretreatments can change the surface area
of the GCE. On the other hand, it is expected that there will be
some diffusion restriction of the reduced and oxidized forms of
the ferrocyanide complex within the GMC. This observation
could be reflected by an improper (non-ideal) tail-off peak
current and non-linearity in the ipa vs. v1/2 plot at v > 70 mV s1
for the redox couple on the GCE/GMC working electrode,
unlike the GCE case (Fig. 3(A)b). Nevertheless, due to the
unique structural feature of the GMC, the electrochemical
response of the GCE/GMC is remarkable and effective over the
unmodified electrode for the purine compounds oxidation.
3.3. Electrochemical and catalytic behaviours of the GCE/
GMC
Fig. 4AC show comparative CV responses of the GCE/GMC
and unmodified GCE with UA, X and HX discretely at a scan
rate of 50 mV s1 in pH 7 PBS. As can be seen from the CV of
UA, the unmodified GCE shows a marked anodic oxidation
peak at 295 mV vs. Ag/AgCl. The GCE/GMC also showed
a similar peak potential value but with two and half timesenhancement in the anodic peak current value over the GCE.
The observation may be due to the surface area effect of the
GCE/GMC. For the case of X (Fig. 4B), the GCE/GMC
showed a profound oxidation peak at 650 mV and was three
times higher in the peak current value than the unmodified
GCE. Note that the X oxidation peak potential at GCE is
750 mV, which is 100 mV higher in over-potential than the
GCE/GMC system. For the HX case (Fig. 4C), the unmodified
electrode shows the absence of any CV oxidation peak up to
1100 mV, whereas the GCE/GMC yielded a clear oxidation
peak at 980 mV. In general, the surface area effect can be
referred as when the modified electrode showed an enhanced
oxidation/reduction current response to the analyte over itscorresponding unmodified electrode without altering the
oxidation/reduction peak potential.42 On the other hand if the
peak potential is altered, for example, the analyte oxidation
potential is significantly reduced by the modified electrode
(over-potential reduction with the symbol h) and it is then
referred to as the electrocatalytic effect. In our case, X and
HX have shown significant reductions in the h values, while for
the case of the UA there is no such reduction in the h; however,
only enhancement in the peak current value was noticed. Based
on the information it can be concluded that X and HX
oxidations are due to the electrocatalytic behaviour, while for
the UA oxidation referred, it is due to the surface area effect of
GCE/GMC.In continuation of the CV studies, DPV was also performed on
the above purine compounds comparatively with GCE and
GCE/GMC as in Fig. 4DF. All the three purine compounds
show well-defined and sharp peak currents respectively at 250,
580 and 900 mV vs. Ag/AgCl for UA, X and HX on the GCE/
GMC. Unlike the CV behaviours, DPV of the unmodified GCE
shows significant decreases in the peak current signals and high
oxidation potential values, 460 and 780 mV for the UA and X
respectively, and nil response to HX. Calculated h reductions for
the UA and X are 200 and 220 mV respectively, in the DPV. But,
no such observations were noticed in the CV (Fig. 4AC).
Presumably, the applied amplitude (50 mV) and pulse effects
(pulse width
0.2 s; pulse period
0.5 s) may have somehindrance to the electro-oxidation of the purine compounds on
the GCE, while for the case with GCE/GMC a positive effect by
the DPV parameters and hence considerable sharp peaks and
reduction in the h were observed. Hence, it is a clear advantage of
using DPV as the qualitative tool for sensitive analyses of the
compounds in this work. Note that we have optimized the DPV
parameters viz., pulse width, amplitude and pulse period with
respect to the GMC modified electrode in order to get sharp
voltammetric responses to the analytes. This is the reason why
the corresponding unmodified GCE showed broad peaks in
Fig. 4D and E.
Fig. 3 (A) CV responses of bare GCE and GCE/GMC with 1 mM of
K3[Fe(CN)6] in 0.5 M KCl solution at v 10 mV s1. (B) Plot ofipa vs. v
1/2
for the ferricyanide redox process at two different electrodes.
This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2165
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
5/10
In order to compare the efficiency of the electron-transfer
behaviours (and in turn to the sharp peak currents), DPV peaks
full-width of the half maximum (FWHM) value is taken as
a sensitive parameter. Calculated FWHM values are 42, 53 and
64 mV respectively for the UA, X and HX. The FWHM values
for the recently reported Ru(DMSO)4Cl2Nafion modified
electrode are 75, 75 and 100 mV respectively.25 The lowered
FWHM values with the GCE/GMC are the clear evidence for the
efficient peak current feature of the present system.
Simultaneous analysis of the UA, X and HX on a GCE/GMC
was nextexaminedby CVand DPV techniquesas inFig. 5AandBrespectively. Interestingly, the GCE/GMC shows well-defined
and well-separated oxidation peak current responses. But the
unmodified GCE displays a 13-fold decrease in the voltammetric
peak current at theoxidation potentials of 400 and800 mV for the
UAand X, andan absence of HX signal. The current signals areill
definedtoo. Appearance of these current signals maybe dueto the
response of overlapped UA, X and HX analytes. Calculated
FWHM values for the analyte peaks from the simultaneous DPV
measurements are UA 54, X 50 and HX 50 mV and these
values were considerably smaller than the previously reported
values of 59, 74, 86 mV on Ru(DMSO)4Cl2Nafion,25 and 65, 70,
140 mV on poly(bromocresol purple) modified electrodes for the
simultaneous oxidations of the three purines in a pH 7 PBS.27 All
these experimental observations evidence the superior function of
theGCE/GMC forthe simultaneousoxidations ofUA, X andHX
in a physiological solution in this work.
In order to checkthe nature of the electron-transfer mechanism,
effects of CV scan rate (v) on the discrete oxidations of the three
purine compounds were examinedas in Fig. 6AC. An increase inthe v implies a systematic increase in the peak currents for all the
three compounds. Peak current valueswere measured andplotted
against the scan rate (v) as a doublelogarithmic plot as inFig.6D.
Calculated slope (vlog ipa/vlog v) values are: UA 0.3 (v < 70 mV
s1) and 0.85 (v > 80 mV s1), X 0.30 and HX 0.30. These
values are closer to the ideal values of 0.5 and 1 corresponding to
diffusion and adsorption controlled electron-transfer mecha-
nisms.43 Since theX hassome weak adsorption effecton theGCE/
GMC (see the Experimental section), a slope value near 1 would
be expected rather than $0.5 in this work. But the measured
experimental value is $0.5, which indicates a diffusion controlled
mechanism of the adsorbed X within the GCE/GMC film. The
trend in the mechanism of the UA changes from diffusion toadsorption controlled reaction at scan rate >70 mV s1 (slope
0.85),whichresembles the trend of theferricyanide case (Fig. 3A).
Presumably, like ferricyanide, there will be some diffusion
restriction for the UA within the pores of the GMC at a high scan
rate, which in turn leads to change in the reaction mechanism.
Exactdetailsfor the occurrence of suchreactionare still unknown.
On the basis of Epa ([balog(v)]/2) + constant, for a totally
irreversible diffusion-controlled process44,45 the ba (i.e., Tafel
slope) value is measured as 10, 40 and 43 mV per decade for UA,
X and HX oxidation reactions respectively on a GCE/GMC
(Fig. 6E). Assuming one-electron transfer in the rate
Fig. 4 Discrete CV (AC) and DPV (DF) responses of the electrochemical oxidations of UA, X and HX at GCE and GCE/GMC in a pH 7 PBS. DPV
parameters are: amplitude 50 mV; increment potential 4 mV; pulse width 0.2 s; pulse period 500 s.
Fig. 5 Simultaneous CV (A) and DPV (B) responses of GCE and GCE/
GMC for a mixture of UA, X and HX dissolved in pH 7 PBS. CV scan
rate (v) 50 mV s1. Other DPV parameters are as in Fig. 4.
2166 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
6/10
determination step (i.e., na 1), since ba 2.303RT/naFaa, the
anodic-transfer coefficient (aa) can then be calculated as 5.9, 1.48
and 1.37 for the UA, X and HX respectively. Previously, Zen
et al. reported an aa value of$1 for an irreversible oxidation of
cysteine on a lead-ruthenate pyrochlore modified electrode.46
Hence, the high values obtained in the present work are an
indication of the effective catalytic performance of the GCE/
GMC to the purine oxidation reactions. Fig. 6F is a typical
illustration for the electrochemical oxidation of the three purine
compounds on the GCE/GMC.Fig. S1(A) shows the effect of solution pH (in the range pH 3
8) on the simultaneous oxidations of the three analytes by the
DPV. Plots of the respective anodic peak current and peak
potential against the solution pH are displayed in Fig. S1(B) and
S1(C) respectively. The peak current values increased positively
against the solution pH and attained the maximum at pH $7.
The peak potential values were systematically decreased against
increase in the pH values. A slope value of60 mV per pH was
calculated for all the three compounds on the GCE/GMC, sug-
gesting the proton-coupled electron-transfer mechanism with
equal number of proton/electron involvement in the reaction
pathway (Nernstian case). Usually, depending on the pKa value
of an analyte, there is a marked alteration in the trend of theoxidation peak potential value.24 Unfortunately, considering the
pKa values of UA (5.4 and 10.3), X (7.4 and 11.2) and HX (8.94
and 12.10), there were no alterations in the peak potentials with
respect to the pKa values in this work. This observation further
suggested that the protonation reaction is not a rate determining
step in the electrochemical oxidation processes.
3.4. Analytical measurements
Effects of UA, X and HX concentrations (calibration) on the
simultaneous detection of the individual analytes are next studied
by systematically increasing one analyte concentration with the
other two fixed as in Fig. 7AC. The optimal DPV parameters
used in this work are: increment 4 mV, amplitude 50 mV,
pulse width 0.2 s, pulse period 0.5 s and sampling width
0.0167 s. Interestingly, systematic variations in the peak currents
upon increases in the analyte concentrations were noticed. As
seen in Fig. 7A, DPV of the UA shows regular increases in the
peak current signals up to 400 mM with current sensitivity and
regression coefficient values of 75.3 nA mM1 and 0.996 respec-
tively. Ten repetitive measurements of 25 mM of UA result ina relative standard deviation (RSD) value of 0.88%. For the UA,
the calculated detection limit (S/N 3) value is 110 nM.
Analytical parameters for the X are: linearity up to 320 mM;
sensitivity 61.8 nA mM1; regression 0.988, RSD 1.29%
([X] 25 mM, n 10) and detection limit 388 nM. Similarly for
HX, the results are: linearity up to 240 mM; sensitivity 112.7
nA mM1; regression 0.996, RSD 1.87% ([HX] 25 mM, n
10) and detection limit 351 nM. Obtained detection limit
values in this work are comparable to or even better than those
previously reported in the literature. For instance, reported
detection limit values for UA, X and HX on a nontronite clay
coated screen-printed electrode with 2 V pre-anodization and 40
s of pre-concentration are 0.42 mM, 7.0 mM and 0.34 mM,26
poly(bromocresol purple) modified GCE; 0.20 mM, 0.06 mM and
0.12 mM,27 NafionRu(DMSO)4Cl2; 0.37 mM, 2.35 mM and
2.37 mM,25 and impure MWCNT; 0.141 mM, 0.134 mM, and 2.87
mM,28 respectively. Simple preparation method and using rela-
tively low cost carbon materials (other than the MWCNT) as
a working electrode without any enzyme, activation and pre-
concentration procedures are the superior features of the present
working electrode over the existing CMEs. Note that the
advantage of elimination of the ascorbic acid interference, unlike
the MWCNT, was achieved with the GMC material (see Section
3.5). With regard to a stable response of the GCE/GMC surface,
Fig. 6 Effect of CV scan rate from 5500 mV s
1 on the electrochemical oxidation of (A) UA, (B) X and (C) HX at GCE/GMC in a pH 7 PBS. Plots oflog(ipa) vs. log v (D) and Epa vs. log v (E) for the purine electrochemical oxidation reactions. (F) Typical illustration for the electrochemical oxidations of
the three purines on the GCE/GMC.
This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2167
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
7/10
we expected that micro- and nano-porous voids might exist on
the GCE surface (after the mechanical and electrochemical
pretreatments); when GMC was modified as ethanol dispersion
on the GCE surface, some of the portions of GMC might have
occupied the porous sites of the underlying surface and thereby
stabilizing the GCE/GMC. In order to check the reproducibility
of the modified electrode, four modified electrodes were
discretely prepared and subjected to DPV analysis of UA. The
calculated RSD value for the peak currents is 3.06%, which
indicates good reproducibility of preparation of the modified
electrode.
3.5. Interference study
With the aim to extend the GCE/GMC for blood plasma purines
analyses, we have examined the effect of a co-existing biochem-
ical, AA, on the simultaneous detection of the purine samples.
For a comparison, GCE/MWCNT was also subjected to the AA
interference study (Fig. 8A). In a whole blood sample, common
interferences for the UA are glucose, cholesterol, triglycerides,
proteins, sodium chloride and AA. All these compounds, except
AA, are electro-inactive. As from the clinical diagnosis data, the
concentration range of the UA and AA present in a healthy
humans blood plasma is 112506 mM (2.08.5 mg dL1) and AA
is 40 15 mM, respectively.4749 Fig. 8B is a typical DPV response
of the GCE/GMC for the simultaneous measurements of a fixed
UA, X and HX content against variable AA concentrations in
the range of 20500 mM. As can be seen, in the absence of AA,
both GCE/MWCNT and GCE/GMC showed qualitatively
similar DPV simultaneous responses. But in the presence of
dilute AA, the GCE/MWCNT shows distinctly different vol-
tammetric purine oxidation peaks. Plots of the three purinecompounds anodic peak current (ipa, base-line corrected)
against the AA concentration, [AA], are given as insets in
Fig. 8A. Unfortunately, addition of a dilute concentration of
AA 20 mM leads to about a three fold increase in the purine
oxidation peak currents and these currents get saturated after
that. With regard to the AA oxidation peak at 0.1 V vs.
Ag/AgCl, a marked peak current signal was noticed only when
[AA] > 100 mM. Below 100 mM AA, large background currents
were observed instead of any AA oxidation feature. These
observations ruled out the choice of the GCE/MWCNT for the
precise simultaneous analyses of the purine compounds in
Fig. 7 Typical DPV responses for the simultaneous detection of UA, X
and HX using a GCE/GMC in a pH 7 PBS. Plots of ipa vs. analyte
concentration, [analyte], were given as the respective insets. DPV
parameters are as in Fig. 4. Fig. 8 Effect of AA interference for the simultaneous DPV detection of
the three fixed purine compound concentrations on (A) GCE/MWCNTand (B) GCE/GMC in pH 7 PBS. Respective ipa vs. [AA] plots were given
as insets in the figure.
2168 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
8/10
the presence of dilute AA in a physiological solution. On the
other hand, the GCE/GMC shows tolerable voltammetric
current signals for the AA concentration up to 100 mM. In
addition, the GCE/GMC did not show any AA oxidation peak
current up to 100 mM (inset, Fig. 8B). Presumably the meso-
porous structure has some negative interactive effects with the
five member non-aromatic cored AA. These observations
evidence that GCE/GMC is the better choice for the detection of
uric acid in human blood plasma.
3.6. Real sample analyses
Practical applications of the GCE/GMC for the simultaneous or
individual analyses of the three purine compounds in human
blood plasma, urine and fish flesh samples were tested (Fig. 9 and
Table 1). Typical DPV responses of the GCE/GMC for healthy
humans blood plasma samples collected at sixty day time
intervals (1st day: #1(A)a and 60th day: #1(A)b), are displayed in
Fig. 9(A)a and b. Real sample analyses data are presented in
Table 1. A marked DPV peak at 0.25 V vs. Ag/AgCl was noticed
corresponding to the presence of the UA in the blood plasma
samples. Standard spikes of the UA result in a systematic
increase in the peak currents on the real sample peak. This
observation qualitatively confirms the presence of the UA in the
blood plasma samples. Based on the standard addition method,the UA concentrations in the blood samples were calculated as
35.74 and 35.66 mM respectively for #1(A)a and #1(A)b. After
correction of the dilution factor, the values turned out as 178.7
and 178.3 mM, which are in the range of the standard UA level.9
The observation of similar UA content values in those real
Fig. 9 GCE/GMC responses for the real samples: (A) healthy human blood plasma samples (A)a and (A)b from the 1st and 60th day respectively, (B)
healthy and (C) RBC/pus cell contaminated urine samples, and (D) a fresh fish sample in a pH 7 PBS. DPV parameters are as in Fig. 4.
Table 1 Various purine real sample analyses using a GCE/GMC modified electrode by the standard addition method
S. no. Real samples Analyte
Detected value
Added (mM) Found (mM) Recovery (%)Original Totalb
1 Blood plasma (#1Aa)a UA 35.7 178.5 mM 20 20.44 102.22 Blood plasma (#1Ab)a UA 35.7 178.5 mM 20 20.73 103.73 Blood plasma (#1B) UA 63.7 318.5 mM 20 20.53 102.74 Urine (#2) X 1.9 95.0 mM 20 20.16 100.8
UA 34.3 3.4 mM 30 31.15 103.85 Urine (#3) UA 37.0 3.7 mM 30 29.30 97.76 Urine (#4) UA 16.6 1.7 mM 20 19.67 98.47 Urine (#5)c UA 81.5 4.1 mM 50 49.97 99.98 Urine (#6)c UA 44.7 2.2 mM 50 50.22 100.49 Urine (#7)c UA 118.9 5.9 mM 50 49.57 99.110 Fish (#8) X 39.9 199.5 mM (760 mg g1) 30 30.30 101.011 Fish (#9) X 27.3 136.5 mM (520 mg g1) 60 62.40 104.012 Fish (#10) X 64.4 322.0 mM (1200 mg g1) 30 29.30 102.313 Fish (#11) X 34.5 172.5 mM (655 mg g1) 25 24.71 99.0
a Collected from the same person. b After dilution factor correction. c Pus cell contaminated urine samples.
This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2169
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
9/10
samples proves the reliability of the present electrode for prac-
tical analyses. In another case, for healthy human urine, sample
#2, two peaks at 0.34 and 0.68 V corresponding to the UA and X
were observed (Fig. 9B). The standard addition method quali-
tatively confirms the presence of UA and X in the urine samples.
Calculated UA and X concentration values are 318.7 and 96 mM
respectively. In continuation, individual detection of UA present
in two healthy humans urine samples (#3 and #4) was also
examined and the values are 3.7 and 1.66 mM respectively andthese values are within the standard UA limits only.9 Fig. 9B is
a DPV of the RBC/pus cell containing urine sample #5. Entries
79 in Table 1 are the measured UA concentration values of the
urine samples, #5#7. Since there is no precise literature avail-
able to refer the RBC/pus cell count versus UA level in urine,
based on the experimental results we are concluding here that
there is no dependence of the excreted UA level on the RBC/pus
cell values of the human urine.
Finally four fresh fish samples (#8#11) were subjected to the
DPV analyses. The samples showed two peak responses at 0.6
and 0.95 V vs. Ag/AgCl corresponding to the X and HX
compounds. All fish samples show ill defined HX peaks except
fish sample #11. The standard addition of HX does not showincrements in the peak current. We suspect some matrix effect
with the HX analysis and hence it could not be detected for some
kinds of fish samples. Nevertheless, the X analysis content in the
fish is vital to judge the quality of the fish. As seen in Table 1,
detected X contents were in the order of 1200 > 760 > 655 > 520
mg g1 for samples #10, #8, #11 and #9 respectively. The absence
of UA with the animal food suggests appreciable freshness of the
fish foods. Amongst the fish samples, sample #10 (Sardinella
gibbosa) is found to be the richest content of X and hence these
are the richest protein diet foods. Finally, calculated recovery
values for the standard spikes of the purine samples all fall at
$100%, which indicates the efficiency of the GCE/GMC in the
real sample analyses.
4. Conclusions
The GMC modified glassy carbon electrode was developed for
efficient and simultaneous electrochemical sensing of three
purine compounds: uric acid, xanthine and hypoxanthine in
a physiological solution. The GMC modified electrode shows
well defined well separated and very sharp purine oxidation
signals at around 0.3, 0.6 and 0.95 V vs. Ag/AgCl respectively for
the compounds. The purine oxidation mechanisms follow the
electrocatalytic pathway for the X and HX, while for the UA case
it was due to the surface area effect. Investigation with the
standard bench mark redox system, Fe(CN)63, showedimproper tail-off peak current behaviours due to diffusion
restriction of the redox couple within the mesopores of the
material. The effect of solution pH on the purine oxidation
reactions obeyed the Nernstian type of proton-coupled electron-
transfer reaction mechanism. Constructed calibration plots were
linear up to 400, 320 and 240 mM of UA, X and HX respectively.
Corresponding detection limit values were 110 nM, 388 nM and
351 nM. Simultaneous or selective detection of the three purine
compounds in blood-plasma, healthy urine samples and fish
samples was successfully demonstrated. A matrix effect was
noticed with some real sample analyses. Calculated recovery
values for the analyses were $100%. Overall, the novelty of the
present electrochemical sensor system lies in the simple prepa-
ration steps using relatively low cost GMC materials for the
separation-less analyses of the three purine bases without any
enzyme, the pre-concentration procedure, the pre-anodization
steps and the extendibility of the sensor to various kinds of real
samples.
Acknowledgements
The authors are grateful for the financial support from the
Department of Science and Technology (DST) under Tech-
nology System Development Programme, India. We also thank
Material Science Division, Indian Institute of Technology,
Madras for the TEM measurements.
References
1 W. N. Kelly, McGraw, Harrisons Principles of Internal Medicine,New York, 10th edn, 1983, vol. 7, pp. 516524.
2 D. P. Wootton and H. Freeman, Churchill Livingstone, Microanalysisin Medical Biochemistry, New York, 6th edn, 1982, pp. 7677.
3 F. Cami~na, M. I. Novo-Rodriguez, S. Rodriguez-Segade andM. Castro-Gago, Clin. Chim. Acta, 1995, 243, 151164.
4 M. A. Carsol, G. Volpe and M. Mascini, Talanta, 1997, 44, 21512159.
5 E. Watanabe, Y. Tamada and N. H. Sato, Biosens. Bioelectron., 2005,21, 534538.
6 K. Safranow and Z. Machoy, J. Chromatogr., B: Anal. Technol.Biomed. Life Sci., 2005, 819, 229235.
7 C. Bory, C. Chantin and R. Boulieu, Clin. Biochem., 1997, 30, 247248.
8 Sz. I. Varga, Z. Novak, L. Pataki, M. Patocskai and B. Matkovics,Clin. Chim. Acta, 1992, 205, 241244.
9 R. T. Kachoosangi, C. E. Banks and R. G. Compton, Electroanalysis,2006, 18, 741747.
10 V. V. S. Eswara Dutt and H. A. Mottola, Anal. Chem., 1974, 46,17771781.
11 http://www.ama-assn.org/.
12 J. Wynants, B. Petrov, J. Nijhof and H. V. Belle, J. Chromatogr., A,1987, 386, 297308.
13 S. P. Ferraris, H. Lew and N. M. Elsayed, Anal. Biochem., 1991, 195,116121.
14 M. Czauderna and J. Kowalczyk, J. Chromatogr., Biomed. Appl.,1997, 704, 8998.
15 K. Safranow and Z. Machoy, J. Chromatogr., B: Anal. Technol.Biomed. Life Sci., 2005, 819, 229235.
16 M. Kito, R. Tawa, S. Takeshima and S. Hirose, J. Chromatogr.,Biomed. Appl., 1983, 278, 3542.
17 H. Lin, D.-K. Xu and H.-Y. Chen, J. Chromatogr., A, 1990, 760, 227233.
18 A. Wang, L. Li, F. Zang and Y. Fang, Anal. Chim. Acta, 2000, 419,235242.
19 T. Grune, G. A. Ross, H. Schmidt, W. Siems and D. Perrett, J.Chromatogr., A, 1993, 536, 105111.
20 J. Wang, G. Chena, A. Muck Jr, D. Shin and A. Fujishima, J.
Chromatogr., A, 2004, 1022, 207212.21 T. Yao, Y. Taniguchi, T. Wasa and S. Musha, Bull. Chem. Soc. Jpn.,
1978, 51, 29372941.22 X. Cai, K. Kalcher and C. Neuhold, Fresenius J. Anal. Chem., 1994,
348, 660665.23 Z. Wang, X. Dong and J. Li, Sens. Actuators, B, 2008, 131, 411416.24 J.-M. Zen, Y.-Y. Lai, G. Ilangovan and A. S. Kumar, Electroanalysis,
2000, 12, 280286.25 A. S. Kumar and P. Swetha, J. Electroanal. Chem., 2010, 642, 135
142.26 J.-M. Zen, Y.-Y. Lai, H.-H. Yang and A. S. Kumar, Sens. Actuators,
B, 2002, 84, 237244.27 Y. Wang and L. L. Tong, Sens. Actuators, B, 2010, 150, 4349.28 A. Senthil Kumar and R. Shanmugam, Anal. Methods, 2011, 3, 2088
2094.
2170 | Anal. Methods, 2012, 4, 21622171 This journal is The Royal Society of Chemistry 2012
View Article Online
http://dx.doi.org/10.1039/c2ay25029b -
7/28/2019 Graphitized Mesoporous Carbon Modified Glassy Carbon Electrode for Selective Sensing of Xanthine, Hypoxanthin
10/10
29 P. Kalimuthu and S. Abraham John, Anal. Chim. Acta, 2009, 647, 97103.
30 D. Sun, Y. Zhang, F. Wang, K. Wua, J. Chen and Y. Zhou, Sens.Actuators, B, 2009, 141, 641645.
31 Y. Y. Sun, J. J. Fei, K. B. Wu and S. S. Hu, Anal. Bioanal. Chem.,2003, 375, 544549.
32 J. Zhang, J. Lei, R. Pan, Y. Xue and H. Ju, Biosens. Bioelectron.,2010, 26, 371376.
33 A. B. Fuertes and S. Alvarez, Carbon, 2004, 42, 30493055.34 M. Kruk, Z. Li and M. Jaroniec, Langmuir, 1999, 15, 14351441.
35 X. Cui, F. Cui, Q. He, L. Guo, M. Ruan and J. Shi, Fuel, 2010, 89,372377.
36 Y. Shao, S. Zhang, R. Kou, X. Wang, C. Wang, S. Dai,V. Viswanathan, J. Liu, Y. Wang and Y. Lin, J. Power Sources,2010, 195, 18051811.
37 C.-C. Hung, P.-Y. Lim, J.-R. Chen and H. C. Shih, J. Power Sources,2011, 196, 140146.
38 Y. Ishigami, K. Takada, H. Yano, J. Inukai, M. Uchida, Y. Nagumo,T. Hyakutake, H. Nishide and M. Watanabe, J. Power Sources, 2011,196, 30033008.
39 H. Xie, Z. Wu, S. H. Overbury, C. Liang and V. Schwartz, J. Catal.,2009, 267, 158166.
40 X. Lu, Y. Xiao, Z. Lei, J. Chen, H. Zhang, Y. Ni and Q. Zhang, J.Mater. Chem., 2009, 19, 47074714.
41 C. D. Liang, S. Dai and G. Guiochon, Anal. Chem., 2003, 75, 49044912.
42 R. L. McCreery, Chem. Rev., 2008, 108, 26462687.43 A. S. Kumar and P. Swetha, Colloids Surf., A, 2011, 384, 597
604.44 H. Xia and H.-L. Li, J. Electroanal. Chem., 1997, 430, 183187.
45 S. M. Golabi and H. R. Zare, J. Electroanal. Chem., 1999, 465, 168176.
46 J.-M. Zen, A. S. Kumar and J.-C. Chen, Anal. Chem., 2011, 73, 11691175.
47 Laboratory Test Handbook, ed. S. David, Lexi-Comp Inc., Cleveland,OH, 1996.
48 N. W. Tietz, Text of Clinical Chemistry, W.B. Saunders, Philadelphia,PA, 1986.
49 J. Lykkesfeldt, M. Viscovich and H. R. Poulsen, Free Radical Biol.Med., 2003, 35, 14391447.
This journal is The Royal Society of Chemistry 2012 Anal. Methods, 2012, 4, 21622171 | 2171
View Article Online
http://dx.doi.org/10.1039/c2ay25029b