gravity-operated pump for tube feeding

1
341 New Inventions GRAVITY-OPERATED PUMP FOR TUBE FEEDING THE conventional gravity drip apparatus for continuous tube feeding has two disadvantages : first, a thick feed may " sludge " behind the regulating clip ; and second, controlling a slow rate of flow is often difficult. Both these factors may prevent the administration of the correct daily volume of feed ; and uneven administration is a common cause of colic and diarrhoea, particularly in jejunal feeding. To meet these objections many types of expensive and complex forced-feed pumps have been devised. 1 2 After much experience of pumping jejunal feeds containing gastric and duodenal aspirate, a simple gravity-operated Clamo . apparatus has been designed which not only has the minimum of moving parts but also does not require a regulating clip or valve in the delivery system. The pump (see figure) consists of two vessels, A and B, containing a total of 250 ml. ’mercury, mounted on a platform which is pivoted at its midpoint. By inclining the plane, the mercury can be made to run from one vessel to the other at a rate determined by the pressure of a brass plate on a spiral of fine rubber tubing. Latex rubber balloons within A and B contain the feed. The 1. Snyder, C. C. Plast. reconstr. Surg. 1951, 8, 73. 2. Fallis, L. S., Barron, J. Bull. int. Chir. 1956, 15, 9. arrangement of the valve C is such that if A is filled with mercury, and the balloon in B with a suitable feed, the flow of mercury into B will displace the feed at a constant rate into the patient ; at the same time the balloon in A will be filled by the feed from the reservoir. When all the mercury has passed from A into B, the pressure in B will fall to approximately that of the atmosphere ; the balloon in A is then full of feed and the vessel B with mercury. If the platform and the valve C are now rotated anticlockwise through 90°, A will refill with mercury, displacing the feed from the reservoir. The complete cycle, which is illustrated, can be repeated as often as required. At a feed rate of 1 ml. per minute the platform is tipped every 4 hours. If a balloon should tear, mercury cannot reach the patient because the outlet tube is carried above the level of the upper container. Contamination of the feed by mercury passing across the balloon has not occurred after periods of exposure of up to 11 hours. To control the rate of rotation of the platform, and thus to prevent damage to the vessels A and B, a modified piston from a Potter-Buckey X-ray screen is mounted between the platform and the base so that the assembly can be tipped only slowly. Department of Surgery, University of Edinburgh D. C. SIMPSON B.SC., PH.D. Edin. External Staff, Medical Research Council H. A. F. DUDLEY M.B. Edin., F.R.C.S.E. A RESUSCITATION TABLE FOR THE NEWBORN RESUSCITATION of the newborn after anaesthesia for labour or csesarean section is becoming more and more of the anaesthetist. If this delicate procedure is to be carried out effici- ently, the tiny pa- tient must be in the correct posi- tion, at the right height for the operator, and with the head fixed so that suction and manual ventila- tion can be quickly and atraumatic- ally applied. Many tables and pieces of appara- tus have been designed for this purpose, but most have dis- advantages. The table illus- trated here (see figure) is of a con- venient height for general use with the anaesthetist standing ; it is equally satisfactory for laryngoscopy when he is sitting before it. Laryngoscopy is facilitated by fixing the baby’s head in the hole in the mattress-a most important factor since an anoxic child is always well relaxed and, owing to the presence of amniotic fluid, very slippery. The hole is oval and measures 11 X 9 cm. Fixation by this method is good enough to prevent a healthy baby from rolling off the table even when struggling and kicking. The top of the table is covered by a 5 cm. thick mattress made of sponge nylon, in which the hole for the child’s head is cut. It is also fixed in a 10° Trendelenburg slope, which can be altered, but for most cases 10° is satisfactory, provided efficient suction is available. The table is equipped with an oxygen cylinder which has both a suction unit and a flowmeter attached. The suction is effective for both pharyngeal and tracheal work, and the flowmeter is connected to a bag and mask of a special light-weight type designed for babies.1 The tube from the flowmeter to the bag incorporates a safety device in the form of a blow-off valve or water- seal set at a pressure of 15-20 cm. of water. This prevents any excessive pressure from reaching the baby when artificial respiration is being applied with the bag and mask. A clock on the table enables an accurate check to be made of the time which passes before normal respiration starts, and draws attention to the need to avoid delay in suction from the bronchial tree or in ventilation. Other equipment includes a water jug for clearing the suction catheter and a drawer beneath the table 1. Secher, O., Nyrop, M. Anesthesiology, 1955, 16, 808

Upload: haf

Post on 30-Dec-2016

215 views

Category:

Documents


1 download

TRANSCRIPT

341

New Inventions

GRAVITY-OPERATED PUMP FOR TUBE FEEDING

THE conventional gravity drip apparatus for continuoustube feeding has two disadvantages : first, a thick feedmay

" sludge " behind the regulating clip ; and second,

controlling a slow rate of flow is often difficult. Boththese factors may prevent the administration of thecorrect daily volume of feed ; and uneven administrationis a common cause of colic and diarrhoea, particularly injejunal feeding.To meet these objections many types of expensive and

complex forced-feed pumps have been devised. 12 Aftermuch experience of pumping jejunal feeds containinggastric and duodenal aspirate, a simple gravity-operated

Clamo .

apparatus has been designed which not only has theminimum of moving parts but also does not require aregulating clip or valve in the delivery system.The pump (see figure) consists of two vessels, A and B,

containing a total of 250 ml. ’mercury, mounted on aplatform which is pivoted at its midpoint. By incliningthe plane, the mercury can be made to run from onevessel to the other at a rate determined by the pressureof a brass plate on a spiral of fine rubber tubing. Latexrubber balloons within A and B contain the feed. The

1. Snyder, C. C. Plast. reconstr. Surg. 1951, 8, 73.2. Fallis, L. S., Barron, J. Bull. int. Chir. 1956, 15, 9.

arrangement of the valve C is such that if A is filled withmercury, and the balloon in B with a suitable feed, theflow of mercury into B will displace the feed at a constantrate into the patient ; at the same time the balloon in Awill be filled by the feed from the reservoir.When all the mercury has passed from A into B, the

pressure in B will fall to approximately that of theatmosphere ; the balloon in A is then full of feed and thevessel B with mercury. If the platform and the valve Care now rotated anticlockwise through 90°, A will refillwith mercury, displacing the feed from the reservoir.The complete cycle, which is illustrated, can be repeatedas often as required. At a feed rate of 1 ml. per minutethe platform is tipped every 4 hours.If a balloon should tear, mercury cannot reach the

patient because the outlet tube is carried above thelevel of the upper container.

Contamination of the feed by mercury passing acrossthe balloon has not occurred after periods of exposureof up to 11 hours.To control the rate of rotation of the platform, and

thus to prevent damage to the vessels A and B, a modifiedpiston from a Potter-Buckey X-ray screen is mountedbetween the platform and the base so that the assemblycan be tipped only slowly.

Department of Surgery,University of Edinburgh

D. C. SIMPSONB.SC., PH.D. Edin.

External Staff,Medical Research Council

H. A. F. DUDLEYM.B. Edin., F.R.C.S.E.

A RESUSCITATION TABLE FOR THE NEWBORN

RESUSCITATION of the newborn after anaesthesia forlabour or csesarean section is becoming more and more

of the anaesthetist.If this delicateprocedure is to becarried out effici-ently, the tiny pa-tient must be inthe correct posi-tion, at the rightheight for theoperator, and withthe head fixed sothat suction andmanual ventila-tion can be quicklyand atraumatic-ally applied.Many tables andpieces of appara-tus have beendesigned for thispurpose, butmost have dis-advantages.The table illus-

trated here (seefigure) is of a con-venient height forgeneral use withthe anaesthetiststanding ; it is

equally satisfactory for laryngoscopy when he is sittingbefore it. Laryngoscopy is facilitated by fixing thebaby’s head in the hole in the mattress-a mostimportant factor since an anoxic child is always wellrelaxed and, owing to the presence of amniotic fluid,very slippery. The hole is oval and measures 11 X9 cm. Fixation by this method is good enough to preventa healthy baby from rolling off the table even whenstruggling and kicking.The top of the table is covered by a 5 cm. thick mattress

made of sponge nylon, in which the hole for the child’shead is cut. It is also fixed in a 10° Trendelenburgslope, which can be altered, but for most cases 10° issatisfactory, provided efficient suction is available.The table is equipped with an oxygen cylinder which

has both a suction unit and a flowmeter attached. Thesuction is effective for both pharyngeal and trachealwork, and the flowmeter is connected to a bag and maskof a special light-weight type designed for babies.1The tube from the flowmeter to the bag incorporates asafety device in the form of a blow-off valve or water-seal set at a pressure of 15-20 cm. of water. This preventsany excessive pressure from reaching the baby whenartificial respiration is being applied with the bag andmask. -

A clock on the table enables an accurate check to bemade of the time which passes before normal respirationstarts, and draws attention to the need to avoid delayin suction from the bronchial tree or in ventilation.

Other equipment includes a water jug for clearingthe suction catheter and a drawer beneath the table

1. Secher, O., Nyrop, M. Anesthesiology, 1955, 16, 808