grimm model 180 - us epa€¦ · continuous pm mass instrument monday, november 2, 2009 –training...
TRANSCRIPT
![Page 1: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/1.jpg)
2009 National Ambient Air Monitoring Conference
Continuous PM Mass Instrument
MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS
Grimm Model 180
Using Optical Technology
![Page 2: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/2.jpg)
A look at
the technology
used in the
GRIMM 180
2009 National Ambient Air Monitoring Conference
![Page 3: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/3.jpg)
Main principles of optical particle detection
Components Main FunctionSample air inlet Reproducibility
Sample air pump Constant flow rate
Light source Generating a signal
Well defined optical volumeOptical Beam
Light trap Eliminate signal noise
Detection optic Known aperture / backward, forward, 90°optical or aerodynamic focusing
Detector scattering light / photo diode, multiplierSignal processing& data processing
rapid count processing andaccurate size classification
All components together determine the spectrometersAll components together determine the spectrometerscounting efficiency (coincidence concentration)counting efficiency (coincidence concentration)
and sizing accuracy (particle size resolution) and sizing accuracy (particle size resolution)
![Page 4: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/4.jpg)
Basic Principles of Particles DetectionThe Nephelometer
x
y
y
aerosol particles
detection volume
Signal from agroupgroup of particles!
Light source
Detector
Sample in
![Page 5: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/5.jpg)
Advanced principles of particle detectionSpectrometer with 90° detection
y
90°x
yzaerosol particles
aerosol focusingdetection volume
Signal from eachsinglesingle particle!
Light source
Detector
Sample in
Light trap
GRIMM can do it !!! GRIMM can do it !!!
![Page 6: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/6.jpg)
Single Particle Light ScatteringGrimm Spectrometer Principle
![Page 7: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/7.jpg)
aerosol focusingaerosol focusing
Single Particle Light ScatteringGrimm Spectrometer Principle
![Page 8: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/8.jpg)
Pump
Laser
Sample stream with particles
Particle filterMeasuringchamber
PTFE-Filter
Light trap
Clean air
Sheath Air to protect Optics
Outlet
Purge air
Sheath air
![Page 9: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/9.jpg)
Particle count distribution is the base for the dust mass calculation. We convert these optical count values to a mass distribution. The geometric mean value between two thresholds of each of the 31 different channels is multiplied by a densityfactor corresponding to the established “urban environment”factor. With this implementation we obtain the proper mass distribution over the full range. Finally we needs to correlate the sample air volume to the mass value to obtain mass/volume.
Mass Distribution Determination & PM Calculation
Based on the obtained mass distribution we can now implement the cut points of the sampling heads. The PM10, PM2.5 and PM1 are then applied.
![Page 10: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/10.jpg)
Grimm Model 180
Candidate forUS EPA PM 2.5Class III FEM
Approval
![Page 11: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/11.jpg)
US EPA PM 2.5 Class III FEM Approval
A: Bakersfield, CA
B: Denver, CO
C: St. Louis, MI
D: New Haven, CT
A: Winter & Summer (46 days)
B & C: Winter Only (23 days each)
D: Summer Only (23 days)
GRIMM Test Sites for Class III Continuous Monitors
![Page 12: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/12.jpg)
Our choice of reference instrument:Our choice of reference instrument:
![Page 13: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/13.jpg)
US EPA PM 2.5 Class III FEM Approval
Grimm Trailer
![Page 14: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/14.jpg)
US EPA PM 2.5 Class III FEM Approval
Grimm Trailer
![Page 15: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/15.jpg)
Results Bakersfield – Scatter Plots
Bakersfield Sitey = 0.977x - 0.079
R2 = 0.9812
3
13
23
33
43
53
63
73
83
3 13 23 33 43 53 63 73 83
FRM (ug/m3)
Gri
mm
180
(ug/
m3)
![Page 16: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/16.jpg)
Results Bakersfield – Time Series
Bakersfield Site
3
13
23
33
43
53
63
73
83
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45
DAY
Conc
entra
tion
(ug/
m3)
FRMGrimm 180
![Page 17: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/17.jpg)
Results Denver – Scatter Plots
Denver Site y = 0.965x + 0.0939R2 = 0.966
3
8
13
18
23
28
3 8 13 18 23 28
FRM (ug/m3)
Grim
m 1
80 (u
g/m
3)
![Page 18: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/18.jpg)
Results Denver – Time Series
Denver Site
3
8
13
18
23
28
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29
DAY
Conc
entra
tion
(ug/
m3)
FRMGrimm 180
![Page 19: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/19.jpg)
Results St-Louis – Scatter Plots
St-Louis Site y = 0.9936x + 0.1815R2 = 0.9466
3
8
13
18
23
28
33
38
3 8 13 18 23 28 33 38
FRM (ug/m3)
Grim
m 1
80 (u
g/m
3)
![Page 20: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/20.jpg)
Results St-Louis – Time Series
St-Louis Site
3
8
13
18
23
28
33
38
1 3 5 7 9 11 13 15 17 19 21 23 25
DAY
Con
cent
ratio
n (u
g/m
3)
FRMGrimm 180
![Page 21: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/21.jpg)
Results New Haven – Scatter Plots
New Haven Sitey = 0.9664x + 0.0377
R2 = 0.9139
3
5
7
9
11
13
15
17
19
3 5 7 9 11 13 15 17 19
FRM (ug/m3)
Gri
mm
180
(ug/
m3)
![Page 22: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/22.jpg)
Results New Haven – Time Series
New Haven Site
3
5
7
9
11
13
15
17
19
1 3 5 7 9 11 13 15 17 19 21 23 25
DAY
Conc
entra
tion
(ug/
m3)
FRMGrimm 180
![Page 23: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/23.jpg)
Results Summary
Bakersfield Denver
St-Louis New Haven
![Page 24: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/24.jpg)
The Grimm 180 can operate for one year un-attended.Annual Calibration at Grimm USA Office.On-site audits can be easily achieved with the compact Test-Kit available.
Flow CheckLeak-TestZero Particle Test
Operation & Maintenance
![Page 25: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/25.jpg)
CALIBRATION TOWER
![Page 26: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/26.jpg)
Benefit of the technology
No Consumable
Minimal Maintenance
Real-Time Measurement (6 sec)
PM2.5 + PM10 + PMc with a single instrument
Insensitive to vibration → Mobile Application
Easy Installation
![Page 27: Grimm Model 180 - US EPA€¦ · Continuous PM Mass Instrument MONDAY, NOVEMBER 2, 2009 –TRAINING SESSIONS Grimm Model 180 Using Optical Technology. A look at the technology used](https://reader034.vdocument.in/reader034/viewer/2022050518/5fa1ab9720b933237a1ec3a7/html5/thumbnails/27.jpg)
Come to our booth for more
information.