Group Delay and Phase Measurement on Frequency Converters Application Note Products: | R&S ZVA8 | R&S ZVA24 | R&S ZVA40 | R&S ZVA50 | R&S ZVT8 | R&S ZVT20 Frequency converters e.g. in satellite transponders need to be characterized not only in terms of amplitude transmission but also in terms of phase transmission or group delay, especially with the transition to digital modulation schemes. They often do not provide access to the internal local oscillators This application note describes a method using the R&S ZVA to measure group delay of mixers and frequency converters with an embedded local oscillator very accurately. The key aspect of this new technique is, that the network analyzer applies a 2-tone signal to the frequency converter. By measuring the phase differences between the two signals at the input and at the output, it calculates group delay and relative phase between output and input. Application Note Bednorz 05.2012-1EZ60_1E

Upload: others

Post on 04-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Group Delay and PhaseMeasurement on FrequencyConvertersApplication NoteProducts:| R&SZVA8| R&SZVA24| R&SZVA40

| R&SZVA50| R&SZVT8| R&SZVT20

Frequency converters e.g. in satellitetransponders need to be characterized notonly in terms of amplitude transmissionbut also in terms of phase transmission orgroup delay, especially with the transitionto digital modulation schemes. They oftendo not provide access to the internal localoscillators This application note describesa method using the R&S ZVA to measuregroup delay of mixers and frequencyconverters with an embedded localoscillator very accurately. The key aspectof this new technique is, that the networkanalyzer applies a 2-tone signal to thefrequency converter. By measuring thephase differences between the two signalsat the input and at the output, it calculatesgroup delay and relative phase betweenoutput and input.

Appli

catio

nNo

te

Bedn

orz

05.20

12-1

EZ60

_1E

Page 2: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Abstract

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 2

Table of Contents1 Abstract ............................................................................................... 3

2 Theoretical Background..................................................................... 3

3 The Two-Tone Method........................................................................ 63.1 Test Setup.....................................................................................................................7

3.1.1 The hardware setup.....................................................................................................8

3.2 The instrument settings ..............................................................................................9

3.2.1 Find optimum settings for low trace noise .............................................................12

3.2.1.1 Increase the power level of Source 1 ......................................................................12

3.2.1.2 Apply averaging.........................................................................................................13

3.2.1.3 Increase the aperture ................................................................................................13

3.2.1.4 Increase number of points and apply smoothing ..................................................14

3.3 Calibration ..................................................................................................................15

4 Test Results ...................................................................................... 174.1 Comparing the Two Tone Technique with S-Parameter technique......................17

4.1.1 Measurement of a filter .............................................................................................17

4.1.2 Measurement of a cable ............................................................................................17

4.2 Measurement of a converter.....................................................................................18

5 Appendix ........................................................................................... 195.1 Alternative setups for lower frequencies ................................................................19

5.1.1 Setup for input frequencies lower than 700 MHz with ZVA24, ZVA40 or ZVA50(up-conversion)..........................................................................................................19

5.1.2 Setup for output frequencies lower than 700 MHz with ZVA24, ZVA40 or ZVA50(down conversion).....................................................................................................21

5.2 Searching for the correct LO frequency..................................................................22

6 Ordering Information........................................................................ 23

Page 3: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Abstract

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 3

1 AbstractMixers are one of the fundamental components of many receivers ortransmitters especially in the microwave range. Any mixer-based receiving ortransmit system requires that the mixers have well-controlled amplitude, phaseand group-delay responses. Especially phase- and group-delay linearity areessential for low bit error rates of data transmission or high target resolution forphased antenna array modules of surveillance systems.A key measurement is the relative and/or absolute group delay for frequencyconverters.Relative phase and group delay can be measured using the so-calledreference or golden mixer technique, as long as the local oscillator isaccessible.However, due to increasing integration and miniaturization often neither thelocal oscillator (LO) nor a common reference frequency signal are accessible.

This application note describes a new technique for measurements onfrequency converters with an embedded LO source and without direct accessto a common reference signal. Key for this new technique is, that the networkanalyzer stimulates the device under test with a two tone signal. By measuringthe phase differences between the two signals at the input and the output thenetwork analyzer calculates the phase transfer function or the group delay ofthe device under test..

The measurement accuracy does not depend on the embedded LO’sfrequency stability as long as the deviation is within the measurement band-width of the network analyzer receivers.

2 Theoretical BackgroundGroup delay measurements are based on phase measurements. Themeasurement procedure corresponds to the definition of group delay gr as thenegative derivative of the phase (in degrees) with respect to frequency f:

For practical reasons, Vector Network Analyzers measure a differencecoefficient of the transmission parameter S21 instead of the differentialcoefficient, which yields a good approximation to the wanted group delay gr, ifthe variation of phase is not too nonlinear in the observed frequency rangef, which is called the aperture.

Page 4: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Theoretical Background

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 4

Fig. 1 shows the terms = 2- 1 and f = f2- f1 for linearly decreasingphase response, e.g. of a delay line.

For non frequency converting devices the measurements of S21 at twodifferent frequencies can happen sequently. This works fine for non frequencyconverting DUTs as filters and amplifiers.

With frequency converting devices like mixers, the phase between the inputand output signal cannot be measured directly, because the frequencies aredifferent. Also the phase is not only influenced by the component itself, but alsoby the phase of the local oscillator.Therefore, phase and group delay measurements on mixers used so-calledreference mixer technique. The reference mixer uses the same local oscillatoras the device under test to reconvert either the RF or IF signal in order to getidentical frequencies at the reference and measurement receiver of the VNA.This technique also helps to get rid of phase instabilities of the local oscillator.

LO

f1

f2

f1

f2

meas. channel

ref. channel

reference mixer

sig.generatorpower splitter

VNA

generator

ref. ports

f2

f2

PORT 1 PORT 2

RFREF

IFREF

This measurement delivers phase and group delay relative to a golden mixer,that was measured for calibration instead of the mixer under test (MUT).Normally, the golden mixer is assumed to be ideal, so that the measurementresult of the MUT shows the phase and group delay difference in respect ofthis golden mixer.

Page 5: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Theoretical Background

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 5

If the LO of the device under test is not accessible, group delay measurementsare not possible with a reference mixer, but use AM or FM modulated signals.Other methods try to reconstruct the LO. They use an external signal generatoras LO for the reference mixer and try to tune the generator frequency until thephase drift versus time of the IF is minimized.These techniques have limitations in terms of dynamic range measurementaccuracy and speed. In addition internal local oscillators of the device undertest often are not very stable, which makes it hard for the external generator tofollow.

The R&S ZVA uses a different approach, that overcomes the problems of theformer techniques.

Page 6: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6

3 The Two-Tone MethodThis new method offered with option ZVA-K9 uses a two tone signal, that issent to the device under test.

IF

LO

RF

f1 f1+ f f2 f2+ f

The R&S ZVA measures the phase differences between both carriers at theinput and the output of the device under test. Then, the group delay iscalculated as :

;12360

1 withf

The frequency difference f between both carriers is called the aperture.

To measure the phase difference of two carriers the R&S ZVA provides foreach analogue receiver channel two digital receivers that measure both signalssimultaneously.

This technique works also in case of a frequency converting DUT, becausefrequency and phase instabilities of the DUT’s LO are cancelled out, whencalculating .

NCO 1

NCO 2

A

DIF

IF

IFBW30 MHz

IFBW1Hz...5 MHz (30 MHz)

sampling rate80 MHz

a‘

a

Page 7: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 7

)))(((cos 11 LOLO outtff

)))(((cos 22 LOLO outtff IF

LO

RF

))((cos 11 intf

))((cos 22 intf

= (2out +LO- 1out-LO) - (2in - 1in)

Besides group delay, the R&S ZVA also calculates the relative phase of theDUT by integration of the group delay as well as the dispersion bydifferentiation of the group delay. Using a mixer with known group delay forcalibration, provides absolute group delay. If only relative group delay isnecessary, a golden mixer is sufficient for calibration.

3.1 Test Setup

The following accessories are required:

Vector Network Analyzer ZVA 4-port orZVT with at least 3 ports

Direct Generator / Receiver Access ZVA-B16Frequency Conversion ZVA-K4Embedded LO Mixer Delay Measurement ZVA-K9Cable Set for ZVA-K9 (recommended) ZVA-B9

This example measures a converterwith the following settings:

Swept RF and IF, fixed LO, IF=RF-LO

RF frequency: 5.37 GHz to 5.47 GHzLO frequency: 4.5 GHzIF frequency: 870 MHz to 970 MHz

Page 8: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 8

3.1.1 The hardware setup

For accurate measurements, it is necessary to generate a two tone signal withan accurate and stable frequency offset. The R&S ZVA can provide this signalby using both sources of a 4-port model.The two tone signal is generated by using an external combiner or using one ofthe ZVA’s internal couplers as combiner.

For that purpose perform the following connections:

Src out (Port 1) -> Meas out (Port 2)

Port 2 -> Src in (Port 1)

Rohde & Schwarz offers with theZVA-B9 a cable set for thedifferent types of ZVA (see orderinformation).

Thus the two tone signal runs viathe reference receiver of Port 1 tothe input of the DUT.This setup is recommended for allZVA models, as long as IF andRF frequencies are above 700MHz. If a microwave ZVA asZVA24, ZVA40 or ZVA50 is used,the attenuation at low frequenciesof the internal coupler may lead toan increased trace noise. If this istoo high, alternative setups arerecommended as indicated in theappendix.To increase the accuracy, well matched 6 dB to 10 dB attenuators arerecommended at both ports directly in the measurement plane.

ZVA 24 VECTOR NETWORK ANALYZER 10 MHz … 24 GHzROHDE&SCHWARZ

SOURCE REF MEAS PORTSOURCE REF MEAS PORTSOURCE REF MEAS PORTSOURCE REF MEAS PORT

OUT

IN IN

OUTOUT

IN IN

OUT OUT

IN IN

OUT OUT

IN IN

OUT 2413GND

IF

LO

RF

attenuators

ZVA-B9

Setup using ZVA-B9

Page 9: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 9

3.2 The instrument settings

Configure the mixermeasurement.

PressMODE - MIXER DELAY MEAS- “DEFINE MIXER DELAYMEAS..”

Press ”Define Mixer Measurement”

Select the LO source

Because the LO of the DUT isinternal, “NONE” is selected

Page 10: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 10

Configure the frequencies andconversion type.

Press “Set Frequencies” in thescalar mixer measurementdialog.

RF frequency: SweptLO frequency : FixedIF frequency: Auto

Fixed Frequency LO: 4.5 GHzSwept frequency:Start 5.37 GHz

Stop 5.47 GHz

Conversion:IF=RF-LO

Press “OK”

Set the power levels:

Press “Set Powers” in thescalar mixer measurementdialog.

CW Power: 0 dBm

The DUT should be measuredwith -10 dBm. Because 10 dBattenuators are applied to theports, 0 dBm is selected.

Press “OK” two times to closethe Set Power and the DefineMixer Measurement.. dialog

Page 11: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 11

Configure the settings for thegroup delay measurement

Upper Tone: Port 3Aperture: 3 MHzMeas Bandwidth: 1 kHzSelectivity: Normal

External generators can be used as well, but are not recommended, becausefrequency and phase stability are very critical.The selected aperture is 3 MHz. It means that the frequency offset f betweenboth carriers is 3 MHz. The aperture for the mixer group delay measurementcan be changed in this dialog, only, not in the format menu.The measurement bandwidth should be as wide as the embedded LO’sfrequency stability or deviation. Normal selectivity is sufficient, as long as theaperture is about 50 times wider than the measurement bandwidth. If theaperture is very small, selective filters have to be used.

Press Set frequencies andPowers in the “Mixer DelayMeasurement…” dialog.

Normally all settings arealready done in the previousdialogs, but can be checkedagain.

If the settings exceed thefrequency range of the ZVA,“Fit Frequency Range” willadjust it automatically

Press “OK”

Now you are back in the Mixer Delay Measurement dialog.Press “OK” to activate the mixer group delay, The measurement parameter ofthe active trace is switched to group delay automatically.

Page 12: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 12

-110

-100

-90

-80

-70

-60

-50

-40

-30

-40

1 of 2 (Max)

Pwr 0 dBmCh1 Mix Frq RF Center 5.42 GHz Span 100 MHz

Trc1Trc2

b2/a1MixDly

dB MagReal

10 dB /10 ns/

Ref -40 dBRef 30 ns MCal

• M 1M 1

5.4260005.427500

GHzGHz

-40.31131.632

dBns

b2/a1

M 1

M 1

3/22/2009, 11:15 PM

It is recommended to measureconversion loss first, to checkif the settings especially thefrequency of the embeddedLO is correct.Mixer group delay and con-version loss can be measuredin the same channel so simplyadd a trace for the conversionloss.

Press:

TRACE SELECT – ADDTRACEMEAS – RATIOS – b2/a1 SrcPort1

If the trace is noisy or themagnitude of the conversionloss changes when reducingthe measurement bandwidth,the LO frequency of the DUTis different to the settings ofthe ZVA.

3.2.1 Find optimum settings for low trace noise

The settings used above should produce (loss of DUT 6 dB) a group delaydeviation (that looks as trace noise) of 100 ps pk-pk. If the DUT has a higherloss and thus higher trace noise the following steps can help to reduce thenoise.

3.2.1.1 Increase the power level of Source 1

Using the cable set ZVA-B9 the source of port 1 is injected into the coupler orbridge of Port 2 via Meas Out. Due to the loss of the coupler or bridge, it isreduced by this attenuation. The attenuation is about 10 dB. (The coupler ofthe microwave ZVAs has 10 dB attenuation above 700 MHz and more forlower frequencies). That means, that the power of the carrier from source 1 isabout 10 dB lower than the carrier from source 2. Increasing the power ofsource 1 will improve the trace noise. This can be done in the portconfiguration.

Page 13: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 13

Press:

MODE – Port Config…Pb… (from Port 1)

The dialog to configure thepower of port 1 pos up.

Add 20 dB for ZVA8 and 10dB for the other ZVAmodels in “Port PowerOffset”

The power of source 1 isnow always 10 dB higherthan the power of source 2

Press “OK”

3.2.1.2 Apply averaging

Reducing the IF bandwidth might help, but can lead to worse results if theembedded LO is unstable. Applying averaging improves the trace noisewithout reducing the bandwidth.

3.2.1.3 Increase the aperture

Increasing the aperture will also improve the trace noise. But there arelimitations. Increasing the aperture, increases also the phase difference of thetwo tones. This phase difference should not exceed 180° otherwise the resultis incorrect.

.50180

f360-(f)

maxmax

f

The higher the group delay, the lower the maximum aperture. A DUT with 100ns limits the maximum aperture to 5 MHz. You find information how to select asuitable aperture in the App.-Note. 1EZ 35 “Group and Phase DelayMeasurements with Vector Network Analyzer ZVR”.

Page 14: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 14

3.2.1.4 Increase number of points and apply smoothing

Increasing the aperture is not only limited by the group delay of the DUT, butwill also hide information, especially if the group delay of the DUT changesversus frequency. Smoothing together with increasing the number of pointsimproves the trace noise furthermore without loosing information. Smoothingcan sometimes improve the trace noise more than applying averaging.

Example:

Press:

SWEEPNumber of Points…: 2001TRACE FUNCTSmoothing Aperture..: 0.5 %

Smoothing aperture of 0.5 meansthat ZVA takes the average of 10adjacent points (=0.5%*2001).

30

35

40

45

50

55

60

65

70

40

2 of 2 (Max)

Pwr 0 dBmCh1 Mix Frq RF Center 5.42538 GHz Span 50 MHz

Trc1Mem3[Trc1]

MixDlyMixDly

RealReal

5 ns/5 ns/

Ref 40 nsRef 40 ns

Smo

• M 1 5.425380 GHz 36.811 nsMixDly

M 1

3/22/2009, 10:10 PM

Page 15: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 15

3.3 Calibration

The calibration is done with an ideal or known calibration mixer. To get aknown calibration mixer, it can be characterized with ZVA-K5 before. ZVA-K5is an option to measure vector error corrected absolute phase and group delayof mixers.Often absolute group delay is not required, but only relative group delay andgroup delay ripple. In these cases it is sufficient to use a “golden mixer” withlinear phase and flat group delay for calibration.

Select a suitable aperture for themeasurement before starting thecalibration. Changing the aperturebetween calibration and deceasesthe measurement accuracy.

Connect the golden or calibrationmixer instead of the DUT.

Use an external signal generator asLO. Set the frequency and thepower level. Lock the referencefrequencies of the LO to the ZVAso that narrow bandwidths can beused.

To lock the ZVA to the sig. genconnect a suitable cable and press:

System Config - ExternalReferencePWR BW AVGMeas Bandwidth:-100 Hz

AVERAGE FACTOR: 10AVERAGE ON

Press:MODE - MIXER DELAY MEAS –CAL MIXER DELAY MEAS…

The calibration dialog pops up.

Page 16: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

The Two-Tone Method

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 16

For absolute group delay, the dataof the calibration mixer have to beloaded using the button “Load”

For relative group delay, a data setwith 0 s of delay is loaded, usingthe button “Zero Delay”

Press:Take Cal Sweep

Wait until the message “finished”appears and close the dialog.

If necessary, the calibration datacan be saved and recalled usingthe “save” and “Load” button.

400

Trc1 MixDly Real 100 ps/ Ref 0 s MCal

MixDly

MCAL In the trace line shows thatthe trace is calibrated

Page 17: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Test Results

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 17

4 Test Results

4.1 Comparing the Two Tone Technique with S-Parameter technique

4.1.1 Measurement of a filter

Blue:Full two port calibrated S-pa-rameter technique

Red:Normalized two tone technique

Green:Normalized S-parameter technique

4.1.2 Measurement of a cable

Green:Normalized S-parameter meas

Read:Normalized two tone technique

Blue:Full two port calibrated S-parameters

500

600

700

800

900

1000

1100

1200

1300

900

1 of 3 (Max)

PwrPwr

0 dBm0 dBm

S-Par-CalUCh3

CenterCenter

5.42 GHz5.42 GHz

—— —

SpanSpan

100 MHz100 MHz

Trc4Trc8Trc9

S21MixDlyS21

DelayRealDelay

100 ps/100 ps/100 ps/

Ref 100 psRef 900 psRef 900 ps

S-Par-CalUCh3Ch3

Ca? SmoMCal SmoCal Smo

• M 1M 2M 3M 4

899.37902.31903.00907.18

pspspsps

S21 M 1 902.76 psM 2 895.31 psM 3 895.06 psM 4 900.84 ps

M 1 912.26 psM 2 896.00 psM 3 902.40 psM 4 917.79 ps

M 3 M 4M 2M 1

M 2 M 3 M 4M 1 M 2 M 3 M 4M 1

3/8/2009, 12:06 PM

20

25

30

35

40

45

50

55

60

30

1 of 3 (Max)

PwrPwr

0 dBm0 dBm

S-Par-CalUtwo-tone-nor...

CenterCenter

5.42 GHz5.42 GHz

—— —

SpanSpan

50 MHz50 MHz

Trc4Trc8Trc9

S21MixDlyS21

DelayRealDelay

5 ns/5 ns/5 ns/

Ref 30 nsRef 30 nsRef 30 ns

S-Par-CalUtwo-tone-nor...two-tone-nor...

Cal intMCalCal

S21 M 1 44.653 nsM 2 30.271 nsM 3 30.998 nsM 4 35.679 nsM 5 48.388 ns

M 1 44.696 nsM 2 30.434 nsM 3 30.762 nsM 4 35.357 nsM 5 48.525 ns

M 1 44.690 nsM 2 30.157 nsM 3 30.752 nsM 4 35.435 nsM 5 48.492 ns

M 2 M 3

M 4

M 1

M 5

M 3

M 4

M 2

M 1

M 5

M 2 M 3

M 4

M 1

M 5

3/8/2009, 12:39 PM

Page 18: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Test Results

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 18

4.2 Measurement of a converter

The measurement results of thefilter and the combination of filterand mixer are compared.

The dark green trace shows thefull two port calibratedmeasurement of the filter (in nonfrequency converting mode).

The blue trace shows themeasurement result of the filterafter transmission normalization.

The purple trace shows the groupdelay of the entire converter, aftercalibrating with a golden mixer.

The maximum deviation of allmeasurements is about 200 ps

10

15

20

25

30

35

40

45

50

30

2

PwrPwrPwr

10 dBm10 dBm10 dBm

Ch1Ch2Ch5

Mix Frq RFMix Frq RFMix Frq RF

CenterCenterCenter

5.42 GHz5.42 GHz5.42 GHz

———

———

SpanSpanSpan

100 MHz100 MHz100 MHz

Trc1Mem3[Trc1]Trc6

b2/a1S21MixDly

DelayDelayReal

Trc2Mem4[Trc2]Mem7[Trc6]

MixDlyMixDlyMixDly

RealRealReal

MixDly

M 1 5.407426 GHz 37.422 nsM 2 5.420000 GHz 30.124 nsM 3 5.429911 GHz 32.865 ns

M 1 5.407426 GHz 38.058 nsM 2 5.420000 GHz 30.238 nsM 3 5.429911 GHz 32.905 ns

M 1 5.407426 GHz 37.970 nsM 2 5.420000 GHz 30.174 nsM 3 5.429911 GHz 33.037 ns

M 1

M 2M 3

M 1

M 2M 3

M 1

M 2M 3

1/16/2009, 3:44 PM

Page 19: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Appendix

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 19

5 Appendix

5.1 Alternative setups for lower frequencies

Loss of the coupler of the microwave instruments ZVA24, ZVA40 and ZVA50

5.1.1 Setup for input frequencies lower than 700 MHz with ZVA24, ZVA40or ZVA50 (up-conversion)

This case might have the biggest impact on the trace noise. The reason is, thatthe source from port 1 is attenuated by the loss of the coupler of port 3 that isused as combiner. At 100 MHz the loss is about 35 dB. Additionally, the twotone signal is attenuated again (e.g. 35 dB @ 100 MHz) before it can bemeasured by the a1 receiver of Port 1.

1E8 1E9

-70

-60

-50

-40

-30

-20

-10

0

10

0

1

Pwr 10 dBmCh1 Start 10 MHz Stop 10 GHz

Trc1 S42 dB Mag 10 dB / Ref 0 dB

M 1M 2M 3M 4M 5

20.00000050.00000070.000000100.00000

1.000000

MHzMHzMHzMHzGHz

-47.106-39.941-37.352-34.584-16.311

dBdBdBdBdB

S42

M 1

M 2M 3

M 4

M 5

1/30/2009, 4:08 AM

Page 20: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Appendix

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 20

Improvement of trace noise using Power Splitters instead of ZVA-B9

Black: Conversion loss with ZVA-B9

Blue: Conversion loss with powersplitters

Black: Group delay with ZVA-B9

Blue: Group Delay with power split-ters

1E8-2.0-1.5-1.0-0.50.00.51.01.52.0

0.0

1

Pwr 10 dBmCh1 MixFrq RF Start 12 MHz Stop 1 GHz

Trc1Mem6[Trc1]

b2/a1b2/a1

dB MagdB Mag

0.5 dB /0.5 dB /

Ref 0 dBRef 0 dB

Math Mem4[Trc1] b2/a1 dBMag 0.5 dB / Ref 0 dB

M1 13.324300 MHz 0.0055 dBb2/a1

M1

1E8-800-600-400-200

0200400600800

0

2

Pwr 10 dBmCh1 MixFrq RF Start 12 MHz Stop 1 GHz

Trc2Mem3[Trc2]Mem5[Trc2]

MixDlyMixDlyMixDly

RealRealReal

200 ps/200 ps/200 ps/

Ref 0 sRef 0 sRef 0 s

MCal MathInvisible

•M1M1

725.71994730.48366

MHzMHz

-25.248-44.288

psps

MixDly

M1M1

3/8/2009, 6:18 PM

Page 21: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Appendix

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 21

5.1.2 Setup for output frequencies lower than 700 MHz with ZVA24,ZVA40 or ZVA50 (down conversion)

To avoid the coupler loss at Port 2, the direct receiver access can be used.The output of the DUT then is connected to Meas In of Port 2 to avoid the lossof the coupler.

Page 22: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Appendix

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 22

Improvement of trace noise using Meas In instead of Port 2

Purple: Conversion loss using Port 2

Green: Conversion loss usingMeas IN

Red: Group delay using Port 2

Blue: Group Delay using Meas In

5.2 Searching for the correct LO frequency

Create a new channel

Press:CHAN SELECT – ADD CHAN-NEL+TRACE+ DIAG AERA

MEAS - WAVE QUANTITIES -b2 Src Port 1

MODE – PORT CONFIG

Push fb …(Frequency) of Port 1

Set the RF frequency to aconstant value in the pass bandof the DUT.

In this example 5.42 GHz.

Press OK

-0.8-0.6-0.4-0.20.00.20.40.60.8

0.0

1

Pwr 0 dBmCh1 MixFrq IF Start 12 MHz Stop 200 MHz

Trc1Mem5[Trc1]Mem6[Trc1]

b2/a1b2/a1b2/a1

dB MagdB MagdB Mag

0.2 dB /0.2 dB /0.2 dB /

Ref 0 dBRef 0 dBRef 0 dB

MathInvisible

M1 146.04734 MHz 0.0006 dBb2/a1

M1

-800-600-400-200

0200400600800

0

2

Pwr 0 dBmCh1 MixFrq IF Start 12 MHz Stop 200 MHz

Trc2Mem3[Trc2]Mem4[Trc2]

MixDlyMixDlyMixDly

RealRealReal

200 ps/200 ps/200 ps/

Ref 0 sRef 0 sRef 0 s

MCal MathInvisible

• M1 44.500000 MHz -23.627 psMixDly

M1

3/8/2009, 5:28 PM

Page 23: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

Ordering Information

1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 23

Set the stimulus axis to the IFfrequency

Press “Stimulus”

Select “All Receivers”

Press “OK” two times

-100

-90

-80

-70

-60

-50

-40

-30

-20

-90

2 of 3 (Max)

Pwr 0 dBmCh1 Arb Port All Rec Center 920 MHz Span 100 kHz

Trc1 b2 dB Mag 10 dB / Ref -90 dBm

b2 M 1 920.00800 MHz -52.791 dBm

M 1

3/22/2009, 11:03 PM

Reduce the span andmeasurement bandwidth

SWEEP - SPAN 100 kHz

Pwr BW AVG –MEAS BANDWIDTH: 1 kHz

SEARCH – MARKERTRACKING - SEARCH MAX

The marker automaticallysearches for the maximum. Themarker value shows the downconverted IF frequency.

The LO is RF-IFLO = 5.42 GHz-920.008 GHzLO = 4.499992 GHz

6 Ordering Information

Designation Type Order No.

Vector Corrected Mixer Measurements1) ZVA-K5 1311.3134.02Embedded LO Mixer Delay Measurements 1) ZVA-K9 1311.3128.02Frequency Conversion ZVA-K4 1164.1863.02Cable Set for ZVA-K9 for ZVA8 ZVA-B9 1305.6541.02Cable Set for ZVA-K9 for ZVA24 / ZVA40 2.9mm ZVA-B9 1305.6541.03Cable Set for ZVA-K9 for ZVA 50 ZVA-B9 1305.6541.04

1) requires ZVA-K4

Page 24: Group Delay and Phase Measurement on Frequency …...The Two-Tone Method 1EZ60_1E Rohde & Schwarz Group Delay and Phase Measurement of Converters 6 3 The Two-Tone Method This new method

About Rohde & SchwarzRohde & Schwarz is an independent groupof companies specializing in electronics. It isa leading supplier of solutions in the fields oftest and measurement, broadcasting, radiomonitoring and radiolocation, as well assecure communications. Established 75years ago, Rohde & Schwarz has a globalpresence and a dedicated service networkin over 70 countries. Company headquartersare in Munich, Germany.

Regional contactEurope, Africa, Middle East+49 1805 12 42 42* or +49 89 4129 137 [email protected]

North America1-888-TEST-RSA (1-888-837-8772)[email protected]

Latin [email protected]

Asia/Pacific+65 65 13 04 [email protected]

This application note and the suppliedprograms may only be used subject to theconditions of use set forth in the downloadarea of the Rohde & Schwarz website.

Rohde & Schwarz GmbH & Co. KGMühldorfstraße 15 | D - 81671 MünchenPhone + 49 89 4129 - 0 | Fax + 49 89 4129 – 13777

www.rohde-schwarz.com