guide for successful insertion of viavalve safety iv catheters/media/m/smiths-medical_com/fil… ·...

2
Guide for Successful Insertion of ViaValve ® Safety IV Catheters ONE HANDED TECHNIQUE TWO HANDED TECHNIQUE Apply tourniquet and prepare site according to policy. Hold catheter by ribbed needle housing with thumb and fingers to insert needle into skin. Visually inspect to confirm that needle bevel and push-off tab are facing up. Anchor vein with gentle skin traction. Insert needle at appropriate angle. Observe for flashback. If needed, slightly advance catheter and needle together to achieve full catheter entry into vein lumen. Place index finger behind the primary push-off tab and PUSH catheter to thread to desired length. DO NOT REINSERT NEEDLE INTO CATHETER AT ANY TIME. Stabilize device at push-off tab with index finger. Holding ribbed needle housing, PULL needle into needle guard until you hear a CLICK. THE CLICK AND VISUAL INSPECTION INDICATE THAT SAFETY DEVICE HAS ENGAGED SUCCESSFULLY. Remove tourniquet. Apply digital pressure as needed beyond catheter tip. Hold catheter hub and needle housing. Disconnect needle housing by pulling backwards with a slight upward motion. Connect Luer lock or tubing to hub per manufacturer’s recommendation. Secure connection with firm push and twist. Stabilize and dress according to policy. Apply tourniquet and prepare site according to policy. Hold catheter by ribbed needle housing with thumb and fingers to insert needle into skin. Visually inspect to confirm that needle bevel and push-off tab are facing up. Anchor vein with gentle skin traction. Insert needle at appropriate angle. Observe for flashback. If needed, slightly advance catheter and needle together to achieve full catheter entry into vein lumen. Place finger or thumb of other hand behind the primary push-off tab and PUSH catheter to thread to desired length. DO NOT REINSERT NEEDLE INTO CATHETER AT ANY TIME. Stabilize device at push-off tab with finger or thumb. Holding ribbed needle housing, PULL needle into needle guard until you hear a CLICK. THE CLICK AND VISUAL INSPECTION INDICATE THAT SAFETY DEVICE HAS ENGAGED SUCCESSFULLY. Remove tourniquet. Apply digital pressure as needed beyond catheter tip. Hold catheter hub and needle housing. Disconnect needle housing by pulling backwards with a slight upward motion. Connect Luer lock or tubing to hub per manufacturer’s recommendation. Secure connection with firm push and twist. Stabilize and dress according to policy. IMPORTANT: Please see the Instructions for Use supplied with the product for a complete listing of the indications, contraindications, warnings and precautions. INSERT INSERT PUSH PUSH DISCONNECT DISCONNECT PULL & CLICK PULL & CLICK

Upload: others

Post on 29-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Guide for Successful Insertion of ViaValve Safety IV Catheters/media/M/Smiths-medical_com/Fil… · Safety IV Catheters ONE HANDED TECHNIQUE TWO HANDED TECHNIQUE • Apply tourniquet

55

55

Guide for Successful Insertion of ViaValve®

Safety IV Catheters

ONE HANDED TECHNIQUE

TWO HANDED TECHNIQUE

• Apply tourniquet and prepare site according to policy.

• Hold catheter by ribbed needle housing with thumb and fi ngers to insert needle into skin.

• Visually inspect to confi rm that needle bevel and push-off tab are facing up.

• Anchor vein with gentle skin traction.

• Insert needle at appropriate angle.

• Observe for fl ashback.

• If needed, slightly advance catheter and needle together to achieve full catheter entry into vein lumen.

• Place index fi nger behind the primary push-off tab and PUSH catheter to thread to desired length.

• DO NOT REINSERT NEEDLE INTO CATHETER AT ANY TIME.

• Stabilize device at push-off tab with index fi nger.

• Holding ribbed needle housing, PULL needle into needle guard until you hear a CLICK.

• THE CLICK AND VISUAL INSPECTION INDICATE THAT SAFETY DEVICE HAS ENGAGED SUCCESSFULLY.

• Remove tourniquet.

• Apply digital pressure as needed beyond catheter tip.

• Hold catheter hub and needle housing.

• Disconnect needle housing by pulling backwards with a slight upward motion.

• Connect Luer lock or tubing to hub per manufacturer’s recommendation.

• Secure connection with fi rm push and twist.

• Stabilize and dress according to policy.

• Apply tourniquet and prepare site according to policy.

• Hold catheter by ribbed needle housing with thumb and fi ngers to insert needle into skin.

• Visually inspect to confi rm that needle bevel and push-off tab are facing up.

• Anchor vein with gentle skin traction.

• Insert needle at appropriate angle.

• Observe for fl ashback.

• If needed, slightly advance catheter and needle together to achieve full catheter entry into vein lumen.

• Place fi nger or thumb of other hand behind the primary push-off tab and PUSH catheter to thread to desired length.

• DO NOT REINSERT NEEDLE INTO CATHETER AT ANY TIME.

• Stabilize device at push-off tab with fi nger or thumb.

• Holding ribbed needle housing, PULL needle into needle guard until you hear a CLICK.

• THE CLICK AND VISUAL INSPECTION INDICATE THAT SAFETY DEVICE HAS ENGAGED SUCCESSFULLY.

• Remove tourniquet.

• Apply digital pressure as needed beyond catheter tip.

• Hold catheter hub and needle housing.

• Disconnect needle housing by pulling backwards with a slight upward motion.

• Connect Luer lock or tubing to hub per manufacturer’s recommendation.

• Secure connection with fi rm push and twist.

• Stabilize and dress accordingto policy.

IMPORTANT: Please see the Instructions for Use supplied with the product for a complete listing of the indications, contraindications, warnings and precautions.

INSERT

INSERT

PUSH

PUSH

DISCONNECT

DISCONNECT

PULL & CLICK

PULL & CLICK

Page 2: Guide for Successful Insertion of ViaValve Safety IV Catheters/media/M/Smiths-medical_com/Fil… · Safety IV Catheters ONE HANDED TECHNIQUE TWO HANDED TECHNIQUE • Apply tourniquet

PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN CANADA AND OTHER COUNTRIES

Smiths Medical ASD, Inc.6000 Nathan Lane NorthMinneapolis, MN 55442, USATel: 1-614-210-7300Toll-Free USA: 1-800-258-5361www.smiths-medical.com

Find your local contact information at: www.smiths-medical.com/customer-support

Smiths Medical is part of the global technology business Smiths Group plc. Please see the Instructions for Use/Operator’s Manual for a complete listing of the indications, contraindications, warnings and precautions. ViaValve, Jelco and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Offi ce and certain other countries. All other names and marks mentioned are the trademarks or service marks of their respective owners. ©2016 Smiths Medical. All rights reserved. SS195382EN-092016

MSOUCA-1068

To learn more about ViaValve® safety IV catheters visit www.smiths-medical.com

Clinical Tips

Issue Possible Cause Tips for Success

Blown vein

Feels dull

Catheter feels flimsy or kinks during threading

1 2

Difficult to connect/blood leakageTough to thread catheter off needle

‘Click’

1. Catheter partially covering needle tip.

2. Catheter pushed forward over needle bevel by inadvertently moving push-off tab forward prior to insertion.

3. Manipulation of hub prior to insertion (Do not twist hub of catheter)

• Visually inspect catheter to ensure it is properly seated prior to insertion. Often an audible click is heard when hub is re-seated (see diagram).

• Avoid resting fi nger on push-off tab before insertion.

• Consider stabilizing index fi nger in front of push-off tab for insertion if needed.

Blown vein

Feels dull

Catheter feels flimsy or kinks during threading

1 2

Difficult to connect/blood leakageTough to thread catheter off needle

‘Click’

1. Only tip of needle in vessel(catheter not in vessel yet).

2. Grip pads held too tightly.

3. Catheter is hitting valve.

CORRECT FINGER PLACEMENT:Make sure fi ngers are on the grip pads.

• Drop angle after confi rming fl ash and slightly advance catheter and needle together as single unit to achieve full vein entry before threading catheter. “Stop, Drop and Advance.”

• Confi rm fi ngers are on grip pads. Squeezing not necessary and may prevent guard from sliding forward for safety activation.

• Assess for optimal catheter location prior to insertion to avoid bifurcations and valves.

• Consider fl oating catheter in with fl uids.

Blown vein

Feels dull

Catheter feels flimsy or kinks during threading

1 2

Difficult to connect/blood leakageTough to thread catheter off needle

‘Click’

1. Needle pulled back before catheter fully threaded into vessel.

2. Pushing and pulling simultaneously.

CORRECT PROCEDURE:1. PUSH catheter to skin line

fi rst, then

2. PULL needle back.

• Prematurely removing the needle can give the “feel” that the catheter is fl imsy. Catheter needs the needle in order to thread properly into vessel (needle acts like a guidewire).

• Fluids may be used to fl oat in the catheter if needed.

• With fi ngers on grip pads, anchor to skin to isolate push-pull motions.

• PUSH catheter to skin line, then PULL needle back. Do not PUSH and PULL at same time.

Blown vein

Feels dull

Catheter feels flimsy or kinks during threading

1 2

Difficult to connect/blood leakageTough to thread catheter off needle

‘Click’

1. With rapid insertion, fl ash may be missed.

2. Force and angle of insertion too high, piercing back wall of the vessel.

3. Needle advanced too far, piercing back wall of vessel.

• Slow insertion and watch for fl ash to confi rm initial vessel entry.

• Reduce force applied to puncture skin.

• Lower angle of insertion.

• Drop angle once fl ash observed, advance slightly as one unit, stabilize the needle, and then ONLY thread the catheter forward off the needle.

Blown vein

Feels dull

Catheter feels flimsy or kinks during threading

1 2

Difficult to connect/blood leakageTough to thread catheter off needle

‘Click’

1. Insuffi cient force being applied to properly secure the extension set to the hub.

2. Luer slip connector is used instead of Luer lock connector on the extension set.

• Stabilize catheter hub, pull back the spin collar on the extension set to expose the male luer. Fully insert the nose/tip into the catheter hub, then reseat the spin collar. Slight resistance may be felt as the male luer meets the valve component inside the hub and pushes it forward to actuate.

• Utilize Luer lock versus Luer slip connector.

• After initial Luer connection with the ViaValve® catheter, digital pressure will be needed when removing components to avoid blood refl ux, as the valve will remain open.

Smiths Medical ASD, Inc.Southington, CT 06489 USAPhone: +1-860-621-9111