hach chloride by thiocyanate-method no.8113-doc316.53.01017_ed7

6
Chloride Page 1 of 6 Chloride, 8113 How to use instrument-specific information The Instrument-specific information table displays requirements that may vary between instruments. To use this table, select an instrument then read across to find the corresponding information required to perform this test. Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 (0.1 to 25.0 mg/L Cl ) Scope and Application: For water and wastewater Test preparation Table 81 Instrument-specific information Instrument Sample cell Cell orientation DR 6000 2495402 Fill line faces right DR 5000 2495402 Fill line faces user DR 3900 2495402 Fill line faces user DR 3800, DR 2800, DR 2700 2495402 Fill line faces right Before starting the test: Filter turbid samples with moderately rapid filter paper and a funnel before analysis. Both the sample and the blank will contain mercury (D009) at a concentration regulated as a hazardous waste by the Federal RCRA. Do not pour these solutions down the drain. Refer to the MSDS sheet for safe handling and disposal of hazardous waste. Gloves are recommended. Collect the following items: Description Quantity Ferric Ion Solution 1 mL Mercuric Thiocyanate Solution 2 mL Deionized Water 10 mL Sample Cells (see Instrument-specific information) 2 Pipet, TenSette ® , 0.1 to 1.0 mL 1 Pipet tips for 0.1 to 1.0 mL TenSette pipet 2 See Consumables and replacement items for reorder information.

Upload: balas43

Post on 16-Dec-2015

18 views

Category:

Documents


2 download

DESCRIPTION

Chloride in water can be determined by the colorimetric Thiocyanate method

TRANSCRIPT

  • Chloride, 8113

    How to use Tinin

    Chloride DOC316.53.01017Mercuric Thiocyanate Method Method 8113(0.1 to 25.0 mg/L Cl)Scope and Application: For water and wastewater

    Test prep

    Instrument

    DR 6000

    DR 5000

    DR 3900

    DR 3800, DR 280

    Before starting t

    Filter turbid sampl

    Both the sample aRCRA. Do not pou

    Refer to the MSDS

    Collect the follow

    Description

    Ferric Ion Solution

    Mercuric Thiocyan

    Deionized Water

    Sample Cells (see

    Pipet, TenSette,

    Pipet tips for 0.1 to

    See ConsumablesChloridePage 1 of 6

    instrument-specific informationhe Instrument-specific information table displays requirements that may vary between struments. To use this table, select an instrument then read across to find the corresponding formation required to perform this test.

    aration

    Table 81 Instrument-specific information

    Sample cell Cell orientation

    2495402 Fill line faces right

    2495402 Fill line faces user

    2495402 Fill line faces user

    0, DR 2700 2495402 Fill line faces right

    he test:

    es with moderately rapid filter paper and a funnel before analysis.

    nd the blank will contain mercury (D009) at a concentration regulated as a hazardous waste by the Federal r these solutions down the drain.

    sheet for safe handling and disposal of hazardous waste. Gloves are recommended.

    ing items:

    Quantity

    1 mL

    ate Solution 2 mL

    10 mL

    Instrument-specific information) 2

    0.1 to 1.0 mL 1

    1.0 mL TenSette pipet 2

    and replacement items for reorder information.

  • ChloridePage 2 of 6

    Chloride

    Mercuric Thiocyanate method for Chloride

    1. Select the tesInsert an adapter required (Instrumespecific informatioRefer to the user for orientation.

    5. Swirl to mix.

    9. Within five miafter the timer expwipe the blank andinto the cell holde

    70 Chloride

    Stored Programs

    Startt.if nt-n).

    manual

    2. Prepared Sample: Fill a sample cell with 10 mL of sample.

    3. Blank Preparation: Fill another sample cell with 10 mL of deionized water.

    4. Pipet 0.8 mL of Mercuric Thiocyanate Solution into each sample cell. Note: Use 1.0 mL with DR 5000.

    6. Pipet 0.4 mL of Ferric Ion Solution into each sample cell. Note: Use 0.5 mL with DR 5000.

    7. Swirl to mix. An orange color will develop if chloride is present.

    8. Start the instrument timer.A two-minute reaction time will begin.

    nutes ires, insert it

    r.

    10. ZERO the instrument.The display will show:

    0.0 mg/L Cl

    11. Wipe the prepared sample and insert it into the cell holder.

    12. READ the results in mg/L Cl.

    Zero Read

  • Chloride

    Interferences

    Sample collCte

    Accuracy chR

    S

    1

    2

    3

    4

    5

    6

    SR

    1

    Table 82 Interfering substances and levelsInterfering substance Interference levels and treatments

    Extreme pH

    The pH should be about 2 after adding reagents.If the sample is strongly acidic or alkaline, adjust a portion of sample before testing to a pH of

    1 See Optional reagenChloridePage 3 of 6

    ection, preservation and storageollect samples in glass or plastic containers. Samples can be stored for at least 28 days at room mperature.

    eckequired for accuracy check:

    Chloride Standard Solution, 1000-mg/L

    50 mL mixing cylinders (3)

    TenSette Pipet and tips

    tandard additions method (sample spike)

    . After reading test results, leave the sample cell (unspiked sample) in the instrument.

    . Select Options>More>Standard additions from the instrument menu.

    . Default values for standard concentration, sample volume, and spike volumes can be accepted or edited. After values are accepted, the unspiked sample reading will appear in the top row. See the user manual for more information.

    . Prepare three sample spikes. Fill three 50 mL mixing cylinders with 50 mL of sample. Use the TenSette Pipet to add 0.1 mL, 0.2 mL and 0.3 mL of 1000-mg/L Chloride Standard Solution, respectively, to the cylinders and mix each thoroughly.

    . Analyze a 10 mL portion of each sample spike as described in the Mercuric Thiocyanate method for Chloride test, starting with the 0.1 mL sample spike. Each addition should reflect approximately 100% recovery.

    . Select GRAPH to view the results. Select IDEAL LINE (or best-fit) to compare the standard addition results to the theoretical 100% recovery.

    tandard Solution Methodequired for accuracy check:

    Chloride Standard Solution, 1000-mg/L

    Volumetric Flask, 500-mL

    Pipet, Class A, 10 mL

    . Prepare a 20.0-mg/L chloride standard solution.

    a. Pipet 10.00 mL of Chloride Standard Solution, 1000-mg/L, into a 500-mL volumetric flask.

    b. Dilute to the mark with deionized water. Follow the Mercuric Thiocyanate method for Chloride test.

    about 7. Use either 5.0 N Sodium Hydroxide Standard Solution1 or a 1:5 dilution of perchloric acid. Use pH paper; most pH electrodes will contaminate the sample with chloride.

    ts and apparatus.

  • ChloridePage 4 of 6

    Chloride

    2. To adjust the calibration curve using the reading obtained with the standard solution, select Options>More>Standard Adjust from the instrument menu.

    3. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate concentration is used, enter the concentration and adjust the curve to that value.

    Method performance

    Summary ofCthcm

    Consumable

    Program

    70 2

    Required reagen

    Description

    Chloride Reagent S

    1 This test requires a

    (1) Ferric Ion S

    (1) Mercuric T

    Water, deionized

    Required appara

    Description

    Pipet, TenSette, 0

    Pipet Tips, for TenS

    Sample cell, 10 mL

    Recommended s

    Description

    Chloride Standard methodhloride in the sample reacts with mercuric thiocyanate to form mercuric chloride and liberate iocyanate ion. Thiocyanate ions react with the ferric ions to form an orange ferric thiocyanate

    omplex. The amount of this complex is proportional to the chloride concentration. Test results are easured at 455 nm.

    s and replacement items

    StandardPrecision

    95% Confidence Limits of Distribution

    SensitivityConcentration changeper 0.010 Abs change

    Portion of curve

    0.0 mg/L Cl 17.922.1 mg/L Cl 0.1 mg/L Cl 1.0 mg/L

    0.3 mg/L Cl 10.0 mg/L

    0.6 mg/L Cl 20.0 mg/L

    ts

    Quantity/Test Unit Catalog number

    et, includes: 250 tests/pkg1

    reagent blank. The number of tests shown refers to any combination of samples and blanks.

    2319800

    olution 1 mL 100 mL 2212242

    hiocyanate Solution 2 mL 200 mL 2212129

    10 mL 4 L 27256

    tus

    Quantity Unit Catalog number

    .1 to 1.0 mL 1 each 1970001

    ette Pipet 1970001 varies 50/pkg 2185696

    square, matched pair 2 2/pkg 2495402

    tandards

    Unit Catalog number

    Solution, 1000-mg/L Cl 500 mL 18349

  • Chloride

    Optional reagents and apparatus

    Description Unit Catalog number

    Chloride Standard Solution, 10-mL Voluette Ampule, 12,500-mg/L Cl 16/pkg 1425010

    Chloride Standard Solution, 100-mg/L Cl 1 L 2370853

    Cylinder, mixing, 5

    Filter Paper, funne

    Funnel, poly, 75 m

    Gloves, chemical r

    1 Other sizes availabl

    Perchloric Acid, AC

    pH Paper, 1.011.0

    Pipet Tips, for TenS

    Pipet, volumetric, C

    Pipet Filler, safety

    Sodium Hydroxide

    Voluette Ampule BChloridePage 5 of 6

    0 mL each 189641

    l, 125 mm 100/pkg 69257

    m each 108368

    esistant, size 99.51

    e

    pair 2410104

    S 680 g 75765

    pH range, 15 foot roll 5/pkg 39133

    ette Pipet 1970001 1000/pkg 2185628

    lass A, 10 mL each 1451538

    bulb each 1465100

    Standard Solution, 5.0 N 50 mL 245026

    reaker, for 10-mL ampules each 2196800

  • Hach Company, 2HACH COMPANYWORLD HEADQUARTERSTelephone: (970) 669-3050FAX: (970) 669-2932

    FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:Call 800-227-4224

    Contact the HACH office or distributor serving you.www.hach.com [email protected]

    In the U.S.A. Outside the U.S.A. On the Worldwide Web ; E-mail

    toll-free

    007, 2010, 2012. All rights reserved. Printed in the U.S.A. Edition 7

    ChlorideMercuric Thiocyanate MethodMethod 8113(0.1 to 25.0 mg/L Cl)