Headspace Extraction of Volatile Oils
PHCH 137 (Lab) – Sec 3 Grp 4
![Page 2: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/2.jpg)
Contents of the Report
I. Basic Principles of Headspace AnalysisII. InstrumentationIII. Headspace Application
![Page 3: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/3.jpg)
Basic Principles of Headspace Extraction
![Page 4: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/4.jpg)
Definition
Volatile components from complex sample mixtures can be extracted from non-volatile sample components and isolated in the headspace or vapor portion of a sample vial.
![Page 5: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/5.jpg)
1. Static Headspace– Uses a gas-tight syringe and injection of a portion of
the headspace gas above a sample in a sealed container.
2. Purge and Trap– The sample is sparged w/ a stream of gas which
carries analytes to the solid trap. The trap is quickly heated and the analyte is released w. the stream of gas to the GC port.
Types of Headspace Sampling
![Page 6: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/6.jpg)
Volatile analytes
Sample, dilution sample, & matrix
modifiers
Gas phase
Sample phase
Phases of the headspace vial
![Page 7: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/7.jpg)
![Page 8: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/8.jpg)
Partition Coefficient and Phase Ratio
Partition Coefficient (K)
• The equilibrium distribution of an analyte between the sample phase and the gas phase
K = Cs/Cg
Where: Cs=concentration of analyte in sample phaseCg=concentration of analyte in gas phase
![Page 9: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/9.jpg)
Partition Coefficient and Phase Ratio
![Page 10: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/10.jpg)
Phase Ratio (β)
• The relative volume of the headspace compared to volume of the sample in the sample vial
β = Vg/Vs
Where: Vs=volume of sample phaseVg=volume of gas phase
Partition Coefficient and Phase Ratio
![Page 11: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/11.jpg)
• Partition coefficients and phase ratios work together to determine the final concentration of volatile compounds in the headspace of sample vials.
Cg = Co / (K + β)
Where: Cg = Concentration of volatile analytes in the gas phaseCo = Original concentration of volatile analytes
Partition Coefficient and Phase Ratio
![Page 12: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/12.jpg)
• Lowest values for both K and β will result in higher concentrations of volatile analytes in the gas phase.
• Lowering of K:– Change in temperature– Introducing salt in the aqueous matrix
• Common salts used: NH4Cl, (NH4)2SO4, NaCl, Na3C6H5O7, (Na)2SO4, K2CO3
Low Partition Coefficient and Phase Ratio
![Page 13: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/13.jpg)
Instrumentation
![Page 14: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/14.jpg)
Sample Vials
• Should match the type and size of the sample being analyzed.
• Always use pre-cleaned vials for sample preparation and storage.
• Some septa are available with a Teflon® face to eliminate bleed from the rubber portion of the septa. These septa should not be re-used.
![Page 15: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/15.jpg)
Heater and Mixer
• Temperature, time, and mixing can be used to improve the transfer of volatile analytes from the sample into the headspace of the vial.
![Page 16: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/16.jpg)
Sampling System
Types of Sampling:
1. Manual sampling2. Automated sampling:
a.Gas-tight Syringe Injectionb. Balanced-pressure Systemc. Pressure-loop System
![Page 17: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/17.jpg)
Sampling System
![Page 18: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/18.jpg)
Sampling System
![Page 19: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/19.jpg)
Sampling System
![Page 20: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/20.jpg)
Applications
![Page 21: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/21.jpg)
Organic Volatile Impurities
USP General Chapters <467>
![Page 22: Headspace Extraction of Volatile Oils](https://reader033.vdocument.in/reader033/viewer/2022061304/54ff8b824a7959b8508b549c/html5/thumbnails/22.jpg)
References:RESTEK CORP. 2000. A Technical Guide for Static
Headspace Analysis Using GC. Restek Corp. www.restek.com [14 July 2011].
UNITED STATES PHARMACOPOEIAL CONVENTION INC. 2007. The United States Pharmacopoeia 30th Revision and National Formulary 25th Edition [CD-ROM]. USPCI: Rockville, MD. Chapter <467>.