hfp100/hfp200 high flow vacuum pump - aircuity · hfp100/hfp200 high flow vacuum pump the hfp...
TRANSCRIPT
HFP100/HFP200 High Flow Vacuum Pump
The HFP Series High Flow Vacuum Pump provides a continuous vacuum through
the OptiNet communications backbone, and is offered with a bracket assembly,
backup redundancy, and status notification options. The pump connects to
the Sensor Suite (SST), which sequences air samples from the Air Data Routers
(ADRs). Multiple areas can be monitored from the ADRs, and the routers can be
networked as part of a larger distributed system.
Includes• Factorysetpressurereliefvalve.
• 100'of1/2"OD-3/8"IDblackpolyethyleneflameretardanttubing.MeetsUL94V2andUL1820.
• Quickconnecttubingfittingsforeasyinstallation.
• Check-valveassemblytosupportthebackuppumpconfiguration.
• Mufflerassembly.
• Vibrationarrestingrubberfeet.
• Electricalcordandplug.
Options• 115Vac/220Vacoptions.
• Bracketassemblyforeasyinstallation.
• Backuppumpandpumpcontrolmoduleredundancyoptionandfaultsignalingfeatures.
FEATURES
• Oillessoperation
• Permanentlylubricatedbearings
• Long-lifediaphragm
• Balancedforsmooth,lowvibrationoperation
• Redundancyandotherfailsafeoptions
HFP100/HFP200 High Flow Vacuum Pump
www.aircuity.com HFP100/HFP200HighFlowVacuumPump Page2of4
BaseNumber
SeriesNumber101 Single115Vac,60HzPump102 Dual115Vac,60HzPumps201Single220Vac,50HzPump202 Dual220Vac,50HzPumps
PumpBracketBA2Includesabracketassemblytomountthepump,
outletsappropriateforthespecifiedpumpvoltage,andapumpinterconnectionkit.
PumpControlModuleOptionPCMIncludespumpcontrolmoduleforbackuppumpconfiguration. Requiresdualpumpselectioninseriesdesgination(102or202).BlankNopumpcontrolmodule.
HFPX-BA2-X
Pump Specifications
Model Number Voltage-Frequency HP VA Watts Net Wt. Per Pump Operating Environment
HFP101/102 115Vac,60Hz .125 450 191 14.5lbsperpump 41°–104°F(5°–40°C)
HFP201/202 220Vac,50Hz .125 420 171 14.5lbsperpump 41°–104°F(5°–40°C)
Regulatory Compliance
Ordering Guide
Page3of4 HFP100/HFP200HighFlowVacuumPump www.aircuity.com
Pump Bracket (BA2)
SpecificationsSize 24.187"Wx7.5"Hx7.25"DWeightapprox.12lbs.Mounting WallMountedContactor Energizing Power 4VA,approx.
ThePumpBracketprovidesastableplatformtomountbothaprimaryandbackuppump.The“BA2”,includesafactorymountedoutletandcontactorassemblydesignedtooperatewithaPumpControlModule(PCM),whenordered,toenablebackuppumpoperationwhenasustainedlossofvacuumisdetected.Initsdeenergizedstate,thecontactorappliesfieldwiredlinevoltage(115Vacor220Vac),basedonthespecifiedpumpvoltagetooneside(thetop)oftheduplexoutlettopowertheprimarypump.WhenalossofvacuumisdetectedbythePCM,itwillenergizethecontactorviaClass2–lowvoltagepowerwhich,inturndisconnectslinevoltagefromtheprimarypumpandappliesittothebackuppump,pluggedintothebottomsideoftheduplexoutlet.ThisisalatchedstatethatmustbemanuallyresetviathePCMresetfunction,orbycyclingpower.
WallMountBracket
CheckValves
SwitchedDuplexOutlet
PrimaryPumpBackupPump
EnclosureforPCMLowVoltageConnectionstoContactor
HFP102-BA2-PCM - shown above (PCM board not shown)
www.aircuity.com HFP100/HFP200HighFlowVacuumPump Page4of4
Pump Control Module Option (PCM)
SpecificationsSize 4.25"Wx4.00"HMounting MountsinsidetheSSTenclosureviaacard-guidebracketincludedwiththeoption.Power24Vac,±15%60HzPower Consumption 2.4VA,approx.Operating Environment40°–120°F(4.4°–49°C),0–90%RH(non-condensing)Regulatory Compliance
PumpControlModulePCBcontinuouslymonitorsvacuumlevelsprovidedbythepumpassemblyviaasensorelement.ItispoweredoffoftheSSTandprovidessignalingtotheSSTtosignifynormaloperation,alossofvacuum(belowafactorysetthreshold),ortheconditioninwhichthebackuppumphasbeenactivated.Whenalossofvacuumhasbeendetectedforasustaineddurationofabout1minutethePCMlatches24Vactoremotelyenergizethecontactorinthepumpbracketassembly(BA2)inordertoenablethebackuppump.Toreturnoperationtotheprimarypump,thePCMmustberesetviaitspush-buttonresetfunction,orbycyclingpower.OptiNetinturnhastheabilitytocommunicatethisconditiontotheKnowledgeCenterinordertopromptacallforservice.
Features• Provideslatchingcontroltoenablebackuppump.
• ContinuouslymonitorsvacuumlevelsatSST.
• Provideslocal0–10Vscaledgaugepressureoutputtoaidfielddiagnostics.
• LEDindicatorsclearlysignifypumpandvacuumstatus.
• Automatic/FixedModes.
• Test/ResetFunctionality.
• IsolatedForm-Ccontactstosignifylossofvacuumtoothersystems.RatedforClass2power(30VAmax).
SSTAssembly PCMMountedinCardGuideBracket
PumpControlModule
ConnectionstoPumpContactor
©2012Aircuity,Inc.Specificationsaresubjecttochangewithoutnotice.HFP100/HFP20009/2012
Part15ClassA
UL916AccessoryEnergyManagementEquipment