Application News No. A471 Spectrophotometric Analysis High Resolution Analysis of Carbon Monoxide (CO) Gas LAAN-A-FT-E053 FTIR is being used for analysis of gases in various industries, such as the gas manufacturing industry where it is used for production management, and for gas monitoring in such fields as chemical manufacturing and semiconductor manufacturing, etc. When conducting gas analysis by FTIR, the resolution and gas cell path length must be selected depending on the concentration and peak shape of the gas component of interest. Here, using the IRTracer-100, we introduce an example of high resolution measurement of the low-molecular weight gas, carbon monoxide (CO). To compare spectral differences due to resolution, we measured the infrared spectra of water vapor at resolutions of 0.25 cm -1 , 0.5 cm -1 and 1 cm -1 , respectively. The analytical conditions used are shown in Table 1. Fig. 1 shows an enlarged view of the infrared spectra in the vicinity of 1620 cm -1 . As can be seen here, the peaks that were not separated at 0.5 cm -1 and 1 cm -1 were clearly separated when measured with a resolution of 0.25 cm -1 . By conducting measurement at high resolution, the spectral peak intensity is noticeably increased, making it easier to distinguish between two peaks that are closely adjacent to one another. This is especially useful in component identification in cases where the mixed gas spectra include similarly shaped, overlapping peaks. Further, since the lower the molecular weight of the gas, the more pronounced the rotation and vibration of the spectrum, it is more effective to conduct measurement of low-molecular weight gases at high resolution. Using CO as an example of a low-molecular weight gas, we measured the spectra of CO gas samples at different concentrations (95 ppm, 191 ppm, 1207 ppm, 2415 ppm), and then generated a calibration curve. In this case, the resolution of 0.25 cm -1 was used. The other analytical conditions were as shown in Table 1. For the calibration curve, we used the height of the peak in the vicinity of 2170 cm -1 , and primary linear calculations were conducted using the multi-point calibration curve method. Fig. 2 shows the overlaid spectra of the CO gasses at their respective concentrations, and Fig. 3 shows the calibration curve. The correlation coefficient r = 0.999 for the calibration curve indicated excellent linearity. Use of a 10 cm gas cell in the quantitation of carbon monoxide gas using a resolution of 0.25 cm -1 and a DLATGS detector makes quantitation possible over a wide range of concentrations, from tens to thousands of ppm. When the 0.5 cm -1 resolution is used, the lower limit of quantitation will increase due to the lower peak intensity, assuming that the noise remains constant. In actuality, the resolution and number of accumulations must be determined while monitoring the state of the obtained spectrum in order to obtain the optimum lower limit of quantitation. n Introduction n Advantage of High Resolution n Quantitation of Carbon Monoxide (CO) Gas Instrument : IRTracer-100 Resolution : 0.25, 0.5, 1.0 cm -1 Accumulation : 30 Apodization : Happ-Genzel Detector : DLATGS Accessories : Gas Cell (10 cm) / NaCl Window Table 1 Instrument and Analytical Conditions 1614 1616 1618 1620 1622 1624 1626 cm -1 0.00 0.01 0.02 0.03 Abs 1616.84 1616.59 DLATGS 0.25cm -1 DLATGS 0.5cm -1 DLATGS 1cm -1 Fig. 1 Overlaid Spectra of Water Vapor Measured at Different Resolutions

Upload: others

Post on 26-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: High Resolution Analysis of Carbon Monoxide (CO) Gas€¦ · When conducting gas analysis by FTIR, the resolution and gas cell path length must be selected depending on the concentration

ApplicationNews

No.A471

Spectrophotometric Analysis

High Resolution Analysis of Carbon Monoxide (CO) Gas

LAAN-A-FT-E053

FTIR is being used for analysis of gases in various industries, such as the gas manufacturing industry where it is used for production management, and for g a s m o n i t o r i n g i n s u c h f i e l d s a s c h e m i c a l manufacturing and semiconductor manufacturing, etc.When conducting gas analysis by FTIR, the resolution and gas cell path length must be selected depending on the concentration and peak shape of the gas component of interest. Here, using the IRTracer-100, we in t roduce an examp le o f h igh re so lu t ion measurement of the low-molecular weight gas, carbon monoxide (CO).

To compare spectral differences due to resolution, we measured the infrared spectra of water vapor at reso lut ions of 0.25 cm-1, 0 .5 cm-1 and 1 cm-1, respectively. The analytical conditions used are shown in Table 1. Fig. 1 shows an enlarged view of the infrared spectra in the vicinity of 1620 cm-1.As can be seen here, the peaks that were not separated at 0.5 cm-1 and 1 cm-1 were clearly separated when measured with a resolution of 0.25 cm-1.By conducting measurement at high resolution, the spectral peak intensity is noticeably increased, making it easier to distinguish between two peaks that are closely adjacent to one another. This is especially useful in component identification in cases where the mixed gas spectra include similarly shaped, overlapping peaks.Further, since the lower the molecular weight of the gas, the more pronounced the rotation and vibration of the spect rum, i t i s more effect ive to conduct measurement of low-molecular weight gases at high resolution.

Using CO as an example of a low-molecular weight gas, we measured the spectra of CO gas samples at different concentrations (95 ppm, 191 ppm, 1207 ppm, 2415 ppm), and then generated a calibration curve. In this case, the resolution of 0.25 cm-1 was used. The other analytical conditions were as shown in Table 1. For the calibration curve, we used the height of the peak in the vicinity of 2170 cm-1, and primary linear calculations were conducted using the multi-point calibration curve method.Fig. 2 shows the overlaid spectra of the CO gasses at their respective concentrations, and Fig. 3 shows the calibration curve. The correlation coefficient r = 0.999 for the calibration curve indicated excellent linearity.Use of a 10 cm gas cell in the quantitation of carbon monoxide gas using a resolution of 0.25 cm-1 and a DLATGS detector makes quantitation possible over a wide range of concentrations, from tens to thousands of ppm. When the 0.5 cm-1 resolution is used, the lower limit of quantitation will increase due to the lower peak intensity, assuming that the noise remains constant. In actuality, the resolution and number of accumulations must be determined while monitoring the state of the obtained spectrum in order to obtain the optimum lower limit of quantitation.

n Introduction

n Advantage of High Resolution n Quantitation of Carbon Monoxide (CO) Gas

Instrument : IRTracer-100Resolution : 0.25, 0.5, 1.0 cm-1

Accumulation : 30Apodization : Happ-GenzelDetector : DLATGS Accessories : Gas Cell (10 cm) / NaCl Window

Table 1 Instrument and Analytical Conditions

1614161616181620162216241626cm-1

0.00

0.01

0.02

0.03

Abs

1616.84

1616.59

DLATGS 0.25cm-1DLATGS 0.5cm-1DLATGS 1cm-1

Fig. 1 Overlaid Spectra of Water Vapor Measured at Different Resolutions

Page 2: High Resolution Analysis of Carbon Monoxide (CO) Gas€¦ · When conducting gas analysis by FTIR, the resolution and gas cell path length must be selected depending on the concentration

ApplicationNews

No.

For Research Use Only. Not for use in diagnostic procedures.The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice.

© Shimadzu Corporation, 2013www.shimadzu.com/an/

A471

First Edition: Oct. 2013

0.20

0.15

0.10

0.05

0.00

2178 2174 2170 2166 2162cm-1

Abs

DLATGS-CO95 ppm 025 cm-1

DLATGS-CO191 ppm 025 cm-1DLATGS-CO1207 ppm 025 cm-1DLATGS-CO2415 ppm 025 cm-1

Correction Height CO

0.125

0.100

0.075

0.050

0.025

0.000

0 500 1000 2000ppm25001500

Calibration Curve EquationCorrection height = 4.944e-005×Concentration + 3.605e-003, r = 0.999

2130214021502160cm-1

Abs DLATGS-CO2415_025 cm-1MCT-CO2415ppm_025 cm-1

Fig. 2 Overlaid Spectra of CO at Four Concentrations (95, 191, 1207, 2415 ppm)

Fig. 3 Calibration Curve of CO Using Four Concentrations

Fig. 4 Spectra of CO Measured with a DLATGS Detector and an MCT Detector

n Advantage of Optional MCT Detector n SummaryThe standard detector for the IRTracer-100 is the DLATGS detector, which provides stable detection over a wide wave number range. However, for detection with higher sensitivity, the liquid nitrogen-cooled MCT detector is available as an option. In measurement at high resolution, the aperture between the interferometer and the light source must be reduced to increase the parallelism of incident light on the interferometer. This results in a reduction in the amount of light reaching the detector. Further, in analysis of a low-concentration gas, there are cases in which a long optical path length cell is required, but as the optical path length increases, the amount of light in

Here, using analysis of CO gas as an example of low-molecular weight gas analysis, we introduced the excellent performance of the IRTracer-100, with its high 0.25 cm-1 resolution and quantitative performance, as well as the advantages of using the optional MCT detector.Using the IRTracer-100 with the LabSolutions IR software offers stable measurement and simple quantitative operations.As always, when conducting gas analysis by FTIR, it is necessary to carefully consider various conditions, including the detection limit, corrosiveness, coexisting substances, etc. Please contact your Shimadzu representative for information such as cell selection, etc.

the cell is reduced. Therefore, when using a gas cell with a long optical path length for high-resolution measurement, a better spectrum can be obtained with an MCT detector, which can provide high sensitivity detection with low noise.Fig. 4 shows overlaid spectra of CO gas measured with an MCT detector and a DLATGS detector. Measurement of the 2415 ppm CO gas was conducted using a resolution of 0.25 cm-1, and a gas cell having an optical path length of 10 cm. The red-trace spectrum was obtained using the DLATGS detector, while the black-trace spectrum was measured with the MCT detector.Due to the high sensitivity of the MCT detector, the light intensity used for this measurement was reduced. However, when compared with the DLATGS detector under the same conditions, a spectrum with less noise was obtained using the MCT detector. Another advantage of the MCT detector is the faster mirror speed compared to that of the DLATGS detector, due to its higher optimum frequency response, which permits a shorter measurement time.