hoklas supplementary criteria no. 2

Upload: geroldo-rollie-l-querijero

Post on 14-Apr-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    1/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 1 of 54

    HOKLAS Supplementary Criteria No. 2

    All Test Categories Equipment Calibration and Verification

    1 Introduction

    1.1 The HOKLAS criteria on calibration and traceability for non-medical laboratories

    stated in Sections 5.5 and 5.6 of HOKLAS 003 and on calibration for medical

    laboratories stated in Section 5.3 of HOKLAS 015 apply to all laboratory

    equipment. All equipment used for tests and/or calibrations, including equipment

    for subsidiary measurement having a significant effect on the accuracy or validity

    of the result of the test, calibration or sampling shall be calibrated. This document

    sets out the requirements and recommendations for calibration requirements and

    calibration intervals for selected types of equipment in Appendices C to I as

    follows :-

    Appendix C provides general calibration requirements for common equipment

    applicable to all Test Categories. Where specific requirements are

    given in Appendices D, E, F, G, H and I for specific test areas of

    testing or measurement, those requirements should take precedence.

    Appendix D covers calibration of standards of measurement and instrumentation

    applicable to electrical and electronic testing.

    Appendix E covers equipment relevant to engineering metrology.

    Appendix F covers calibration using comparative techniques, of analytical

    instruments used predominantly in chemical testing laboratories.

    Appendix G covers equipment and calibration requirements applicable to

    construction materials testing.

    Appendix H covers calibration requirements of toy testing equipment.

    Appendix I covers calibration requirements for electrical and electronic product

    testing

    Appendix J covers calibration, performance check and preventive maintenance

    requirements for common medical testing equipment

    For equipment not included in the above appendices and equipment peculiar to

    particular test methods, this document explains how the calibration requirements

    and calibration intervals should be determined.

    1.2 General recommendations for determining calibration programmes and operation ofthe calibration systems are given in Section 5.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    2/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 2 of 54

    1.3 This document also explains the available sources of calibration.

    2 Calibration Requirements

    2.1 Parameters of equipment to be calibrated

    Where the values of parameters are specified in the relevant standard specifications,

    they shall be calibrated. Where tolerances are given, the measured values should

    be checked against them. Where tolerances are not specified, the requirements

    given in section 3 of this document should be followed.

    2.2 Calibration procedures and intervals

    Where the standard specifications state the calibration procedures and the

    calibration intervals, these shall be followed.

    Where they are not specified, the laboratory should determine appropriate

    calibration procedures and intervals after considering all relevant factors. For

    dimensions and weights, a period of six months is generally acceptable. Section 5

    gives more detailed recommendations for determining calibration programmes.

    2.3 Unspecified parameters

    For parameters which may affect equipment operation but are not specified, theeffects of change (for example, due to wear) should be assessed during regular

    equipment inspection and functional checks. Where necessary, specific

    procedures should be derived to assess the effects. It is a good practice to record

    initial values as a reference for later comparison.

    2.4 All equipment calibrations should be recorded.

    3 Parameters With No Stated Tolerance

    3.1 Where the tolerances of specified parameters are not given, the laboratory shoulddetermine whether the parameters are critical and how critical they are.

    3.2 For critical parameters, justification for adopting particular tolerances should be

    documented and HOKLAS will examine them during assessments.

    For dimensional and angular measurements, a laboratory may select from

    ISO 2768-1: 1989 a suitable tolerance class. As a rough guide, the medium

    tolerance class in Table 1 and Table 3, for dimensional and angular measurements

    respectively, is acceptable for most equipment.

    3.3 In the case of non-critical dimensions, HOKLAS will normally accept that theimplied tolerance is that the next significant figure should not result in rounding to

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    3/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 3 of 54

    a value other than the stated dimension. For example :-

    stated parameter implied tolerance6 mm

    3.2 mm

    1.0 kg

    1.000 kg

    6 mm 0.5 mm

    3.2 mm 0.05 mm

    1.0 kg 0.05 kg

    1.000 kg 0.0005 kg

    4 Source of Calibration

    4.1 If calibration of equipment is required, there are two choices :-(a) internal calibration; or

    (b) external calibration.

    4.2 HOKLAS encourages laboratories to develop their internal calibration capability

    and to calibrate their own equipment. A calibration can be performed internally if

    the laboratory has the necessary reference standards and materials and the

    laboratory staff are competent to perform the calibration. Internal calibration

    procedures should be documented.

    4.3 Calibrations which require equipment not available to the laboratory or demand

    specialist techniques outside the capabilities and experience of the laboratory staff

    (see HOKLAS 003, Clause 5.6.H (d)) should be performed externally.

    4.4 External calibration shall be performed by a competent calibration body defined in

    HOKLAS 003. A laboratory should take the following approaches to find a

    competent calibration body :-

    (a) Check the HOKLAS Directory of Accredited Laboratories (the up-to-date

    scope of accreditation can be found from the HOKLAS internet website at

    http://www.itc.gov.hk/en/quality/hkas/hoklas/directory/directory.htm) to see if

    a laboratory is accredited for the calibration.

    (b) Contact HKSAR Standards and Calibration Laboratory to see if they can

    perform the calibration.

    (c) Contact HOKLAS calibration Mutual Recognition Agreement (MRA) partners

    to see if there is a laboratory which can perform the calibration accredited by

    them.

    (d) Contact nominated primary standard institutions listed in HOKLAS internet

    website at

    http://www.itc.gov.hk/en/doc/quality/hkas/supplementary%20criteria/SC2-Ap

    pendixB.pdfto see if they can perform the calibration.

    HKAS Executive should be consulted if no suitable calibration laboratory could be

    found using the above approaches.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    4/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 4 of 54

    5 Calibration Programme and Operation of Calibration System

    5.1 Determination of calibration programmes

    The calibration programme applicable to an individual laboratory should be

    determined by the laboratory management with due regard to the following factors:

    (a) Which items of equipment can be calibrated internally and the interval

    between recalibrations.

    (b) Which items of equipment need to be calibrated by a competent calibration

    body (refer to HOKLAS 003, Section 5.6.H) and the interval betweenrecalibrations.

    (c) For each item of equipment, what ranges and cardinal points should be

    calibrated, the calibration uncertainty requirements and the conditions under

    which calibrations should be performed. These should be determined based

    on the requirements of the tests or measurements for which that item of

    equipment is being used. Laboratories should document such requirements

    and provide them to external calibration laboratories when soliciting

    calibration services.

    (d) When determining the recalibration intervals, the following should be takeninto consideration:

    - HOKLAS recommendations;

    - manufacturer recommendations;

    - frequency of use;

    - equipment condition;

    - equipment stability and previous calibration and maintenance history;

    - the accuracy and precision requirements of the tests for which that item of

    equipment is being used (Where the standard specification states the

    calibration intervals, these should be followed);

    - the likely influence of the working environment, e.g. corrosion, dust,

    vibration, frequent transportation and rough handling;- the availability and reliability of in-service verifications and functional

    checks, e.g. the availability of equipment, reference standards or materials

    and expertise for conducting in-service verifications and functional

    checks (Where in-service checks are not carried out between calibrations,

    the calibration interval must be shortened).

    For new test instruments, it is advisable to specify a shorter recalibration interval.

    When successive calibration results demonstrate that the characteristic of an

    instrument is stable, its recalibration interval may be extended. To facilitate the

    determination of calibration intervals, laboratories are recommended to plot curves

    of instrument drifts against time.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    5/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 5 of 54

    For more information for determination of equipment recalibration intervals, please

    refer to OIML International Document No. 10.

    The calibration programme should be documented and reviewed regularly.

    5.2 Designated officers of the laboratory shall be assigned the responsibility for the

    calibration of equipment and the management of reference materials. Laboratories

    should operate a system to alert staff of calibration or verification due dates for all

    items of equipment and reference materials. There should also be a mechanism to

    ensure that equipment or reference materials due for recalibrations can be released

    promptly from routine operation.

    5.3 Equipment overdue for calibration should generally be taken out of serviceimmediately. Under exceptional circumstances, the laboratory management may

    consider temporarily extending the equipment calibration validity for a short period

    of time. Extension of calibration validity should only be allowed when the

    equipment calibration history shows that the drift of the instrument is acceptable

    and in-service verifications and functional checks demonstrate that the instrument

    is functioning normally in every aspect. Nevertheless, the measurement

    uncertainty of the instrument may have to be increased according to the drift trend.

    5.4 Calibrations should be conducted according to the calibration requirements as

    explained in 5.1(c). Where an external calibration laboratory or a different section

    of the same laboratory is contracted to perform the calibration, they should beclearly and unambiguously informed of the calibration requirements. These

    requirements would generally include the calibration range, the cardinal points, the

    required calibration uncertainties and the conditions under which calibrations are to

    be performed. It is recommended that a copy of the documented calibration

    requirements be provided to the calibration laboratory accompanying the

    instrument to be calibrated.

    5.5 After an instrument is calibrated, the laboratory should assign a staff member to

    check whether the characteristics of the instrument as shown by the calibration

    results are still within the acceptable tolerance range. Out of tolerance instruments

    should be repaired and recalibrated before use. Alternatively, such instrumentsmay be downgraded for less demanding purposes provided that they are

    appropriately labelled.

    5.6 The suitability of the calibration programme and the adequacy of the calibration

    system will be examined and discussed during assessments.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    6/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 6 of 54

    6. Address

    6.1 Addresses of the various HOKLAS calibration MRA partners and their nominatedprimary standards institutions can be found from the HKAS website at

    (http://www.itc.gov.hk/en/quality/hkas/hoklas/agreement.htm).

    HKAS Executive

    Issue Date: 30 November 2005

    Implementation Date: 1 June 2006

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    7/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 7 of 54

    APPENDIX A

    ACCREDITATION BODIES WHICH HAVE MUTUAL RECOGNITION

    AGREEMENTS WITH HOKLAS ON CALIBRATION

    Please refer to the website http://www.itc.gov.hk/en/quality/hkas/hoklas/agreement.htm for

    up-to-date information.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    8/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 8 of 54

    APPENDIX B

    ADDRESSES OF NOMINATED PRIMARY STANDARDS INSTITUTIONS

    Please refer to the website

    http://www.itc.gov.hk/en/doc/quality/hkas/supplementary%20criteria/SC2-AppendixB.pdf

    for up-to-date information.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    9/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 9 of 54

    APPENDIX C

    CALIBRATION AND PERFORMANCE CHECK REQUIREMENTS - GENERAL

    This Appendix lists the current recommended calibration requirements for specific items of

    equipment. These are general recommendations applicable to all Test Categories for which

    a laboratory is accredited.

    Where specific requirements are given in Appendices D, E, F, G, H, I and J for specific areas

    of testing or measurement, those requirements should take precedence.

    Type of equipment

    Recommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    ACCELEROMETERS

    (i) Reference(ii) Working

    Five years (check annually)Two years

    ACOUSTIC CALIBRATORS One year

    Intercompare every six months

    ANEMOMETERS Two years

    AUTOCLAVES Initial (on-commissioning) and after

    significant repair*Monthly check*

    Each use*

    Each month*

    Every 6 months*

    Temperature measurement at

    strategic sitesEffectiveness of sterilisation (withbiological indicators)Pressure

    Timer accuracy

    Heating profiles of typical loads

    Effectiveness of sterilization bysuitable means such as chemical

    indicators, temperature and timerecord

    Effectiveness of sterilisation (withbiological indicators)

    Temperature controller calibrationTimer accuracy

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    10/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 10 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    BALANCES and SCALES Three years

    orThree years*

    Both alternative accompanied by

    (i) Each weighing*,(ii) Daily*, and

    (iii) Six months*

    By a competent calibration bodyas defined in clause 5.6.H (d) ofHOKLAS 003

    Calibrate using calibrated masses.Refer The Calibration of Weightsand Balances published by

    CSIRO

    (i) zero check(ii) one-point check using a

    known mass.(iii) repeatability check.

    BAND PASS FILTER SETS Two years

    BAROMETERS (Fortin) Five years (cleanness of mercury,vacuum space and one-point)

    BEAT FREQUENCY

    OSCILLATORS

    One year

    CALIBRATION BATHS

    AND FURNACES

    Initial complete temperature survey.Check temperature distribution at

    one temperature every five yearsCALIBRATION UNIT FOR

    AUDIO-FREQUENCYVOLTMETERS

    Check annually. Calibrate the

    ac/dc transfer error of thermalelement and multiplier every fiveyears

    CENTRIFUGES One year (where operating speedspecified)*

    Tachometer (mechanicalstroboscope or light cell type)

    Temperature calibration (where

    applicable)

    DENSITY BOTTLES,

    PYCNOMETERS

    One year* AS 2378; BS 733 App A; IP 190

    DENSITY METERS Initial and whenever testtemperature is changed, thereafter*One week checks with air anddouble distilled water*

    ASTM D4052

    DIAL GAUGES Two years or less, depending onuse*

    AS 2103; BS 907

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    11/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 11 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    DIES AND CUTTERSFor preparation of test specimenssuch as dumb-bell rubber specimens

    Frequent examination for damage.Full dimensional check whenever

    resharpened

    EXTENSOMETERS

    (a) Level and mirror types

    (b) Micrometer screw types

    (c) Dial indicator types

    (d) Recording types with electricaloutput

    Five years

    Five years

    Two years

    Two years

    AS 1545 Grade D (for proof stresstests load-extension curves for

    pre-stressing wires)

    AS 1545 Grade B (for modulus ofelasticity determinations)(BS EN 10002-4)

    FLOWMETERS

    (a) Rotameters (Reference)High flow,i.e. greater than 1 L/minLow flow,i.e. less than 1 L/min

    Two years*

    Two years*

    ASTMD3195

    Soap bubble flowmeter

    (b) Rotameters (Working) Each time on use* Soap bubble flowmeter

    (c) Orifice plates,Venturi nozzles

    (d) Wet test meters

    (e) Anemometers

    (f) Pitot tubes

    Initial

    Six months*

    One year*

    Two years

    Initial*On use*

    BS 1042: Section 1.2: 1989ISO 5167-1: 1991ISO 9300: 1990

    Visual inspect tip and other parts for

    damage, wear or contamination

    ASTM D1071

    Check dimensional compliance withBS 1402 Sec. 2.1 Annex A

    Inspect tip for damage, blockage,etc., as requested by BS 1042 Sec.2.1

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    12/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 12 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    FORCE TESTING MACHINES(tension, compression, universal)

    TYPE 1 - Mechanical ForceMeasuring Systems

    (a) Dead weight(b) Knife edge, lever and

    steelyard(c) Pendulum dynamometer

    (d) Elastic dynamometer(e.g. spring, ring with dialgauge)

    Note: Chain testing and similarmachines in frequent use

    TYPE 2 - Hydraulic or PneumaticForce MeasuringSystems

    (a) Mechanical systemincorporating a pneumatic or ahydraulic link, e.g.,

    proportional cylinder(b) Bourdon tube or diaphragm

    pressure gauge as forceindicator

    Five yearsFive years

    Two years

    Two years

    One year

    Two years

    Six months

    AS 2193BS 1610: Part 1BS 1610: Part 3

    BS 5214 (rubbers and plastics only)EN 10002-2

    (c) Type (b) fitted also with amaster gauge which can be

    disconnected during normaltesting

    (d) Bourdon tube or diaphragmgauge used only as a nulldetector for a mechanicalsystem

    (e) Bourdon tube with electrical

    transducer

    TYPE 3 - Electrical ForceMeasuring Systems

    One year (plus frequent checks byuser of working gauge against

    master gauge)

    Two years

    Two years

    Two years

    FREQUENCY ANALYSERS(Acoustic Measurements) Five years

    FREQUENCY RESPONSETRACERS(Acoustic Measurements) Check annually

    FREQUENCY STANDARDS(a) For acoustic measurements

    (b) For time and frequencymeasurements

    Two years

    One year

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    13/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 13 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    FURNACESFor use at specified temperatures One Year* BS 4309, AS 2853

    GAS STERILISERS Effectiveness of sterilisation.(Regular checks with biologicalindicators at representativelocations in typical loads)*

    GAUGE BLOCKS

    (a) Used as reference standards

    (b) Used as working equipment

    Four years

    Two years*

    BS 4311: Parts 1, 2 & 3

    HARDNESS TESTERS FOR

    METALS

    (a) Brinell, Vickers and Rockwellmachines including portable

    testers(b) Portable Brinell microscopes

    (c) Diamond indenters

    (See Note 1 at end of this

    Appendix)

    One year (partial)Three years (complete)

    One year (with calibrated graticule)

    One year (inspection) *

    Routine check by user each day fortester in use.

    AS 1816 (Brinell)AS 1817 (Vickers)

    AS 1815 (Rockwell)

    HARDNESS TESTERS FOR

    PLASTICS AND EBONITE

    (a) Dead weight testers forplastics

    (b) Meters (durometer) forrubber

    Three years

    Frequent checks by user usingreference hardness blocks

    ISO 48 : 1994Method N, H, L & M

    ISO 7619 : 1997

    HYDROMETERS

    (a) Reference(b) Working - glass

    - metal

    InitialOne year against reference or withfreshly prepared solutions of known

    density*Six months*

    AS 2026; IP 160; ASTM E126

    ISO R649

    HYGROMETERS

    (a) with electric impedance sensor(b) with chilled mirror sensor

    Two yearsSix years

    IMPACT TESTING MACHINES

    (Pendulum type)

    (a) Charpy, Izod and Universaltesters for metals

    (b) Charpy and Izod testers forplastics

    (c) Notching tools

    AS 1544.4; BS 131

    AS 1146.3

    Check regularly and whenever

    reground

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    14/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 14 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    IMPEDANCE MATCHINGNETWORKS

    (Acoustics measurements)

    5 years.Check annually

    INCUBATORS and CONSTANT

    TEMPERATURE CHAMBER

    for BIOCHEMICAL OXYGEN

    DEMAND

    Every six months against referencethermometer*Constant daily monitoring oftemperature variation duringtesting*

    Maximum/minimum thermometersshould also be used to ensuretemperature stays within rangeduring testing.

    LENGTH MEASURING

    DEVICES

    (a) Reference steel rules(b) Working rules and vernier

    callipers

    InitialMonthly or less depending on use*(zero and general condition)

    For making measurements whichwill not affect test results, rules andvernier callipers from reputablemanufacturers may be used without

    calibration

    LINEAR VARIABLE

    DIFFERENTIAL

    TRANSDUCERS (LVDTS)

    Daily or whenever used*

    Two years* (complete calibration)

    Check against length standard such

    as a micrometer

    MANOMETERS(a) Reference

    (b) Working

    Ten years

    Three years*

    Check fluid every three years

    Check against reference

    MICROMETERS One month (zero, one point andcondition of anvils)*

    Against gauge blocks. See alsoAS 2102; BS 870 and BS 959

    MICROPHONES

    (LS1P or LS2P)

    Check every three months.

    Calibrate annually or when 1 dBchange is detected, whichever is thesooner.

    MICROPHONE AMPLIFIERS Check response annually

    NEUTRAL DENSITY FILTERS Ten years

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    15/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 15 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    OVENS (Temperature of load spaceshould be monitored by appropriatetemperature sensor throughout use)

    (a) Ageing

    (b) Drying

    (c) Vacuum

    Five years (temperature variation,recovery time, rate of ventilation)

    Three years (temperature variationand evaporation rate in working

    space)*

    Five years (temperature variation,evaporation rate, pressure inworking space)*

    BS 2648, AS 2853

    AS 2853 and AS 1289.0

    PENETRATION CONES ANDNEEDLES

    Five years ASTM D217; IP 50; ASTM D5;ASTM D1321. Visually inspectneedle tips prior to use.

    PHOTOCELLS Check linearity of response everysix months. Check spectralresponse annually with colourfilters; calibrate every five years orwhen apparent filter transmittances

    change significantly.

    PHOTOMETRIC TEST PLATEFor luminance measurements

    Five years

    PRESSURE AND VACUUM

    GAUGES

    (a) Test gauges for calibration of

    working gauges(b) Working gauges subject to

    shock loading(c) Working gauges not subject

    to shock loading

    One year

    Six months or less depending onuse*One year*

    AS 1349; BS 1780: Part 2

    PRESSURE GAUGE TESTERS

    (a) Dead weight(b) Manometers

    Ten yearsTen years

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    16/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 16 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    PROVING DEVICES FORCALIBRATION OF FORCE

    TESTING MACHINES

    TYPE 1 - Elastic devices(a) Dial gauge for deflection(b) Micrometer screw for

    deflection measurement(mechanical or optical

    indication)(c) Electrical deflection

    measurement

    TYPE 2 - Proving levers

    TYPE 3 - Weights

    Two yearsFive years

    Two years

    Five years

    Ten years

    PSYCHROMETERS, ASSMANN

    HYGROMETERS and SLING

    TYPE PSYCHROMETERS

    Ten years (complete)Six months (compare thermometersat room temperature with wickdry)*

    AS 2001.1 Appendix C

    PYRHELIOMETERS Five years

    PYROMETERS

    (a) Reference(b) Working

    Three yearsSix months*

    BS 1041: Part 5

    QUARTZ CONTROL PLATES Ten years

    RADIOMETERS Calibrate after one hundred tests

    REFERENCE BALLASTS Five years

    REFERENCE GLASS FILTERS

    (a) Spectrophotometry

    (b) Colorimetry(d) Luminous transmittance

    Five years

    Five yearsFive years

    REFRACTIVE INDEX

    STANDARDS

    (a) Liquid(b) Solid

    Five years10 measurements or 10 years,whichever is the sooner.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    17/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 17 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    SIEVES Six months Sieves should comply with theGrade B requirements of AS 1152unless the test method specifies

    otherwise.Sieves supplied with a record cardas prescribed in BS 410 areaccepted as complying with

    AS 1152 requirements. (SeeAppendix B, AS 1152).

    Laboratories requiring more thanone set of sieves should have areference set meeting the aboverequirements and one or moreworking sets.Working sieves should be given a

    performance check against thereference set using material typicalof the samples normally subjectedto sieve analysis in the laboratory.

    SOUND LEVEL METERS One year

    Check every three months*

    SOUND POWER SOURCES Five years

    STANDARD LAMPSLuminous flux, luminous intensity,illuminance, spectral radiance,spectral irradiance

    Five years or after each 20-hourperiod of burning

    STOPWATCHES(timers)

    Three months* Comparisons against time andfrequency broadcasts fromshort-wave stations such as JJY

    (Japan) at 2.5, 5, 8, 10, and 15 MHzfor at least ten minutes.Comparisons against timing signalsbroadcasted by Radio TelevisionHong Kong is also acceptable.

    TACHOMETERS

    (a) Reference(b) Working

    Five yearsOne year*

    BS 3403

    TAPE RECORDERS(for Acoustic Measurements)

    Five yearsCheck annually*

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    18/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 18 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    THERMOHYGROGRAPHS One week against calibratedhygrometer or psychrometer whenused to determine actual RH and

    temperature*One month against calibratedhygrometer or psychrometer whenused only as indicator of RH and

    temperature variations*

    AS 2001.1 Appendices C & D

    THERMOCOUPLES

    (a) Reference(Preferably noble metal type)

    Three years (complete)Six months (one-point)*

    BS 1041: Part 4; BS EN 60584-1;ASTM E220; ASTM E230

    (b) Working In general, same as for referencethermocouples. Working

    thermocouples subject to frequentstress and strain should be checkedmore frequently depending on theirphysical condition and theinhomogeneity effect.

    Thermocouples used in hightemperatures have definite life

    spans that must be taken intoconsideration.

    THERMOMETERS

    (a) Liquid-in-glass

    (1) Reference

    (2) Working

    (b) Electronic

    (c) Platinum resistance

    (d) A.C. resistance bridge

    Ten years (complete)Six months (zero-point)*Initial check at sufficient points tocover the expected working rangefollowed by single-point checks at

    six-monthly intervals*

    Three years (complete)Six months (one-point)*Ten years (complete)Check ice-point before use or at

    least every 6 monthsFive years (resistance ratios)

    NML Technical Papers No. 1 and 5;ASTM E77Check against a calibrated referencethermometer; NML TechnicalPapers No. 1 and 5;

    IANZ Technical Guide No. 3NML Technical Paper No. 5

    THICKNESS GAUGES(for compressible materials) Two years Dial gauge, dimensions and

    pressure of plunger base

    VELOCITY TRANSDUCERS Calibrate sensitivity and frequencyresponse every two years.

    Check every six months*

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    19/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 19 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    VISCOMETERS

    (a) U-tube

    (1) Reference(2) Working

    Ten years (against standard oils) *Two years (using quality oil againstreference tubes or standard oils)*

    ASTM D2162; BS 4708ASTM D2162/D445; IP 71

    (b) Others

    (1) Brookfield

    (2) Ferranti(3) Zahn

    Two years (using standard oils) plusone-months checks using

    manufacturers oils, covering thenormal range of operation of theinstrument*Three months (using standard oils)*One year (using standard oils)*

    ASTM D2556

    VOLUMETRIC APPARATUS

    (a) Volumetric glassware(pipettes, burettes, volumetricflasks, distillation receivers)when used for tests where the

    error contribution fromglassware to the overallaccuracy required is significantor where methods specifyglassware meeting nominatedstandards.

    (b) Specialised glassware (watertraps, sulphonation flasks,centrifuge tubes, etc.)

    (c) Piston operated volumetricapparatus (micropipettes)

    Initial (on commissioning)*

    Initial*

    Initial*3 months*

    AS 2162.1; BS 1797; BS 6696;ISO 4787

    AS 2162.1; BS 1797

    Check volume delivered. Foradjustable devices check volume

    delivered at several settings (refer toAS 2162.2, ISO 8655 Parts 1-6)

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    20/54

    Items marked * in the table are those which can be carried out by the staff of a laboratory if it is suitably

    equipped and the staff is competent to perform such calibrations.

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 20 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents andequipment requirements

    WEIGHTS(a) Reference weights of integral

    construction, stainless steel or

    nickel chromium alloy

    Five years initial;Ten years subsequent

    ASTM E617

    (b) Working weights, Class B,stainless steel or nickel

    chromium alloy

    Three years* ASTM E617

    (c) Working weights, Class B, notstainless steel or nickelchromium alloy

    One year*

    Note: weights used for routine calibration of balance are considered working weights.

    NOTE 1 - ADDITIONAL REQUIREMENTS FOR HARDNESS TESTING ON METALS

    (1) Laboratories shall carry out a performance check test at the start of each day on which

    hardness tests are to be performed.

    (2) Hardness blocks with assigned values from established block manufacturers (such as Vickers,

    Avery, Wilson and Yamamoto) may be used.

    (3) The check should be carried out using forces close to those involved in the tests to be

    conducted on that day. So far as is practicable, the hardness values should also match those

    involved on that day. Laboratories shall have an adequate range of blocks to cover the range

    of hardness and the forces normally encountered.

    (4) Laboratories are encouraged to have an accredited metrology laboratory do a limited

    calibration on their blocks at the forces applied in the laboratory but which are not included in

    their original calibration, but it is recognised that in practice there will be occasions when a

    laboratory has to use a force for which the block has not been calibrated.

    (5) For Vickers and Rockwell hardness tests, laboratories may use uncalibrated indenters,

    provided those indenters have been checked by performance tests on hardness blocks withassigned values and by inspection at a magnification of a least 50x.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    21/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 21 of 54

    APPENDIX D

    CALIBRATION REQUIREMENTS FOR ELECTRICAL AND ELECTRONIC

    MEASURING EQUIPMENT

    The following table sets out recommended maximum periods between successive calibrations

    for a number of reference standards and measuring instruments. It must be stressed that

    these periods are generally considered to be the maximum appropriate in each case providing

    that the other criteria as specified below are met :-

    that the equipment is of good quality and of proven adequate stability, and

    that the laboratory has both the equipment capability and staff expertise to performadequate internal checks, and

    that if any suspicion or indication of overloading or mishandling arises, the

    equipment will be checked immediately and thereafter at fairly frequent intervals

    until it can be shown that stability has not been impaired.

    Where the above criteria cannot be met, appropriately shorter intervals may be specified.

    The nominal maximum period may be extended in special cases where the laboratory has

    demonstrated exceptional self-checking capability or successful participation in proficiency

    testing programmes involving portable standards.

    Type of equipmentRecommended maximum period

    between successive calibrations

    ATTENUATORS Three years (frequency response)

    BRIDGES Five years (full calibration)Check against laboratory standard annually

    CAPACITORS Five yearsIntercompare annually

    DIGITAL METERS One year

    INDUCTORS Five yearsIntercompare annually

    INSTRUMENTS, INDICATING

    AND RECORDING

    Five years

    Intercompare every six months, or more frequently as required

    INSTRUMENT and RATIO

    TRANSFORMERS

    Ten years

    INSTRUMENT TRANSFORMER TEST

    SETS

    Five years (full calibration)

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    22/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 22 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    POTENTIOMETERS Five years

    RESISTORS Five yearsIntercompare annually

    RF POWER MEASURING EQUIPMENT Three years

    SIGNAL GENERATORS One year (frequency accuracy, output level and attenuatorratio)

    STANDARD CELLS Two yearsIntercompare at least six-monthly

    TIME, TIME INTERVAL and

    FREQUENCY STANDARDS

    Calibration intervals dependent on equipment frequency, typeand accuracy required. This may be as frequently as daily ifthe highest possible performance is required.

    TRANSFER STANDARDS, AC-DC Eight yearsIntercompare immediately after calibration and then every four

    years

    VOLTAGE CALIBRATING

    TRANSFORMERS

    Ten years

    VOLT RATIO BOXES Five yearsIntercompare annually

    WATTHOUR METERS

    (ELECTRO-MECHANICAL)

    Two yearsIntercompare every three months

    WATTMETERS AND WATTHOUR

    METERS (ELECTRONIC)

    One year with regular intercomparisons - interval to be basedon history of performance

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    23/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 23 of 54

    APPENDIX E

    CALIBRATION REQUIREMENTS FOR ENGINEERING METROLOGY

    INSTRUMENTS

    The following table sets out the maximum recommended period between initial calibration

    and the first recalibration, and the maximum period between subsequent recalibrations

    provided that the first two calibrations indicate that the item is stable. The recalibration

    interval for each specific item will be examined during the assessments of the laboratory.

    Account will be taken of factors such as general condition and frequency of use.

    Item

    Maximum period between

    initial calibration and thefirst recalibration (years)

    Maximum period between

    subsequent recalibrations(years)

    ANGLE GAUGES(a) reference(b) working

    42

    84

    DIVIDING HEADS 5 8

    GAUGE BLOCKS(a) reference

    (b) working

    4

    2

    8

    4

    HEIGHT SETTING MICROMETERS AND

    RISER BLOCKS

    3 3

    LENGTH BARS(a) reference(b) working

    42

    84

    LEVELS (precision) 4 4

    LINEAR (precision) 5 10

    OPTICAL FLATS 3 6

    OPTICAL PARALLELS 3 6

    PRECISION POLYGONS 5 10

    ROLLERS AND BALLS 4 8

    ROUNDNESS STANDARDS 5 10

    ROUGHNESS STANDARDS 4 4

    SCREW CHECK PLUGS FOR RING

    GAUGES

    3 6

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    24/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 24 of 54

    Item

    Maximum period between

    initial calibration and thefirst recalibration (years)

    Maximum period between

    subsequent recalibrations(years)

    SCREW PITCH REFERENCE STANDARDS

    3 6

    SETTING CYLINDERS 3 6

    SETTING RINGS 3 6

    SQUARES

    trysquares block squares

    24

    58

    SURFACE PLATES cast iron granite

    34

    58

    THREAD MEASUREMENT CYLINDERS 4 6

    THREAD MEASUREMENT VEE PIECES 2 5

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    25/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 25 of 54

    APPENDIX F

    CALIBRATION AND PERFORMANCE CHECKS REQUIREMENTS FOR

    ANALYTICAL INSTRUMENTS FOR CHEMICAL TESTING

    1. Introduction

    1.1 Analytical instruments used in chemical testing are frequently calibrated by comparative

    techniques, that is, the response signals of the instrument are related to the

    concentrations or amounts of the analyte through the use of reference materials. Many

    types of analytical instruments require calibration before use and calibration thus forms

    an integral part of the analytical procedures.

    1.2 Correct use combined with proper calibration may not necessarily ensure an instrument

    is performing adequately. Where appropriate, periodic checks on the performance

    characteristics should be carried out. These instrument performance characteristics

    include, for example, response, stability and linearity of sources, sensors and detectors,

    the separation efficiency of chromatographic systems, the alignment and wavelength

    accuracy of spectrometers, etc. The purpose of these checks is to verify the performance

    of the instruments against the specified requirements of the tests for which the

    instruments are being used.

    1.3 This Appendix sets out specific calibration and performance checks requirements for

    analytical instruments used for chemical testing. Laboratories should refer to Sections

    5.5 and 5.6 of HOKLAS 003 and other sections of this supplementary criteria for

    general requirements on calibration of equipment. Equipment in common with other

    testing fields are listed in Appendix C.

    1.4 The laboratory should document and implement a calibration and/or performance checks

    programme appropriate to the circumstances involved. The performance characteristics

    to be checked as well as the acceptance criteria for and frequency of checks should be

    based on the requirements of the tests concerned taking into consideration the

    recommendations given in Table 1 and factors given in clause 3.1 of this appendix. It is

    the responsibility of the laboratory to demonstrate that the programme is suitable for thecircumstances involved.

    1.5 Since standard practice for operation of analytical instruments are generally unavailable

    or highly specific to a particular application, the laboratory shall document its practice

    for use of analytical instruments. This should include, as appropriate, a description of the

    operation of the instrument, procedures and acceptance criteria for calibration and

    performance checks, frequency of use and nature of quality control samples, and

    maintenance procedures.

    1.6 Calibration and performance checks can normally be conducted by the staff of a

    laboratory. The laboratory should, however, ensure that sufficient reference materialsand required equipment are available, and the persons involved should possess the

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    26/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 26 of 54

    required skills, experience and training.

    1.7 The laboratory shall ensure that all calibration and/or verification of analyticalinstruments are traceable to national or international standards of measurement or to

    reference materials produced by competent producers (see HOKLAS Supplementary

    Criteria No. 1).

    2. Calibration

    2.1 Normally, either the manufacturers of the instruments and/or the test standards

    themselves specify the calibration procedures and requirements. These instructions

    should be followed, if available. In some cases, check samples are used during analyses

    to confirm the validity of the calibration. Re-calibration of instruments should be carriedout when such checking procedure indicates that previous calibrations are no longer

    valid.

    2.2 Acceptance criteria for calibration and criteria for re-calibration should be based on the

    test standards, if available, or determined by the requirements of the tests concerned.

    These criteria shall be documented and strictly adhered to.

    2.3 Acceptance criteria for calibration should include a criterion for the acceptability of the

    calibration function and, when applicable, a criterion for the performance of instruments.

    An example of the former criterion is correlation coefficient of a linear calibration graph

    whilst the latter may include, for example, the slope of the calibration graph or themagnitude of the detector response at a specific level of analyte. The latter instrument

    performance check (IPC) during calibration should not be confused with that described

    in Clause 3 below. This IPC should be regarded as a routine start-up check on the

    usability of instruments as distinguished from the verification of equipment against

    requirements of tests.

    3. Performance check

    3.1 It is inappropriate to specify a rigid frequency for performance checks since the

    frequency depends on a number of factors including :-- type of equipment;

    - manufacturers recommendation;

    - trend data obtained from previous performance checks records;

    - recorded history of maintenance and servicing;

    - extent and severity of use;

    - tendency to wear and drift;

    - frequency of cross-checking against other reference standards;

    - environmental conditions (temperature, humidity, vibration, etc.);

    - accuracy and permissible limits of errors.

    3.2 Acceptance criteria for performance checks should be based on the requirements of test

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    27/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 27 of 54

    standards, if available, taking into consideration recommendations given by the

    manufacturers of the instruments. Test standards may sometimes explicitly state the

    acceptance criteria for various performance characteristics but, frequently, suchacceptance criteria have to be derived from the requirements of test standards. For

    instance, the detection limits of an instrument may have to be derived from the

    specification limits of the standards.

    3.3 Instruments shall be taken out of service when results of performance checks fall outside

    the acceptance criteria. Suitable corrective measures and servicing should be taken when

    results of performance checks indicate that there is a system degeneration. An

    instrument should not be put back into service unless subsequent checks indicate that its

    performance fulfils the acceptance criteria.

    3.4 The performance characteristics to be checked depends on the type of the instrument.Table 1 lists those performance characteristics of common analytical instruments that

    require to be checked.

    4. Calibration and Performance Check Frequency

    4.1 Table 1 sets out the normal frequencies for calibrations and performance checks of

    common analytical instruments. The frequencies, when given, are recommended

    maximum periods. They serve as a starting point for initial selection of

    re-calibration/performance checks intervals.

    4.2 Intervals of calibration/performance checks should be reviewed regularly and

    adjustments of these intervals may be made based on experience. However, any revision

    of frequency shall be justified (e.g. changes of usage, environment, required accuracy,

    etc.). The guidelines given in OIML International Document No 10 for the

    determination of recalibration period should be followed.

    4.3 Test standards may sometimes specify a shorter calibration/performance check intervals

    than that listed in Table 1. In such cases, the frequency recommended by the test

    standards shall be followed. In addition, if major mechanical or electronic maintenance

    is carried out, the appropriate checks should be conducted when the instrument is

    re-commissioned.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    28/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 28 of 54

    Table 1 Calibration and Performance Checks Requirements

    for Analytical Instruments

    Type of equipmentRecommended period between

    successive calibrations

    Calibration procedure or guidance

    documents and equipment

    requirements

    BOMB CALORIMETERS 6 months or after any significantpart of the system has been changed

    Determine the effective heat capacityusing certified benzoic acid. Referto ISO1928.

    CONDUCTIVITY METERS 6 months Determine cell constant usingpotassium chloride standard

    solutions appropriate to each desiredmeasuring range. Refer to ISO7888/OIML R56 & 68.

    DISSOLVED OXYGEN

    METERS

    Calibrate before useCheck linearity before using theinstrument for routine purposes andmonthly thereafter

    Calibrate at zero and a value nearsaturation. Check linearity. Refer toISO 5814.

    pH METERS Calibrate on useCheck performance for new

    electrodes, and once every 2 weeksthereafter

    Calibrate the instrument with twostandard buffers on use, appropriate

    to the anticipated pH of the samplebeing measured. Check performancecharacteristics such as liquid junction

    error, shift on stirring and noise.Refer to ISO 10523.

    TURBIDIMETERS Calibrate on use Calibrate the apparatus using

    formazine standard solutions. Atleast five points with each measuringrange should be used for plotting thecalibration graph. Refer to ISO 7027.

    CHROMATOGRAPHS Calibrate on use. Checkperformance yearly

    (a) Gas chromatographs check thefollowing, as appropriate:

    i. Overall system checks,

    precision of repeat sampleinjections, carry-over;

    ii. Column performance(capacity, resolution,retention);

    iii. Detector performance(detection limit and

    linearity);iv. System heating/

    thermostatting (accuracy,precision, stability, rampingcharacteristics);

    v. Autosampler (accuracy and

    precision of time routines).Refer to OIML R82/83/113

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    29/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 29 of 54

    Type of equipmentRecommended period between

    successive calibrations

    Calibration procedure or guidance

    documents and equipment

    requirements

    CHROMATOGRAPHS(contd)

    Calibrate on use. Checkperformance yearly

    (b) Liquid chromatographs checkthe following, as appropriate :

    i. Overall system performance;ii. Detector sensitivity;iii Mobile phase delivery system

    (precision, accuracy,pulse-free)

    Refer to OIML R112.

    SPECTROMETERS AND

    SPECTROPHOTOMETER

    Calibrate on use. Check

    performance yearly

    (a) UV/Visible - Check wavelength

    accuracy, stray radiation,absorbance accuracy and zero

    absorbance line flatness. Referto AS3753.

    (b) Infra-red - Check wavenumber

    accuracy, wavenumberreproducibility and beambalance. Refer to ASTM E932.

    (c) Atomic Absorption - Checkprecision, sensitivity and

    detection limit of metals ofinterest. Refer to OIML R100.

    (d) Atomic Emission usingarc/spark discharge - Checkprecision and background

    equivalent concentration. Referto AS 2883.

    (e) Inductively Coupled PlasmaAtomic Emission - Checkrepeatability, backgroundequivalent concentration and

    detection limit. Refer toAS3641.2/OIML R116.

    (f) X- Ray Fluorescence - Checkprecision, counter resolution,dead time. Refer to

    AS2563/OIML 123

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    30/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 30 of 54

    Annex I

    (Informative)

    Bibliography

    Part 1 General Principles of Equipment Calibration

    Laboratory staff responsible for equipment calibration and verification are strongly advised to

    consult the following references.

    1. CITAC/EURACHEM Guide 1 International Guide to Quality in Analytical Chemistry.

    An Aid to Accreditation.

    2. ISO10012 Measurement management systems - requirements for measurement process

    and measuring equipment.

    3. ISOGuide 32 Calibration in analytical chemistry and use of certified reference

    materials.

    4. ISOGuide 33 Uses of certified reference materials.

    5. ISO/TR 13530 Water quality - Guide to analytical quality control for water analysis.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    31/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 31 of 54

    Part 2 Recommended Practices for Specific Equipment

    The following references contain description of the operation, calibration and/or performancecheck procedures. Operation staff of laboratories are strongly advised to familiarize

    themselves with these references.

    A. UV/Visible Spectrophotometers

    1. AS3753 Recommended practice for chemical analysis by ultraviolet/visible

    spectrophotometry.

    2. ASTME131 Standard terminology relating to molecular spectroscopy

    3. ASTME169 Standard practices for general techniques of ultraviolet quantitative

    analysis

    4. ASTME275 Standard practice for describing and measuring performance of ultraviolet,

    visible and near-infrared spectrophotometers

    5. ASTME925 Standard practice for monitoring the calibration of ultraviolet-visible

    spectrophotometer whose spectral slit width does not exceed 2 nm

    6. ASTME958 Standard practice for measuring practical spectral bandwidth of

    ultraviolet-visible spectrophotometers.

    B. Infra-red Spectrophotometers

    1. ASTME168 Standard practice for general techniques of infrared quantitative analysis

    2. ASTME275 Standard practice for describing and measuring performance of ultraviolet,

    visible and near-infrared spectrophotometers

    3. ASTME932 Standard practice for describing and measuring performance of dispersiveinfrared spectrophotometers

    C. Atomic Absorption Spectrophotometers

    1. AS2134 Recommended practice for chemical analysis by atomic absorption

    spectrometry: AS2134.1 Flame atomic absorption spectrometry, AS2134.2 Graphite

    furnace spectrometry and AS2134.3 Vapour generation AAS

    2. AS3550.6 Determination of filtrable calcium - Flame atomic absorption spectrometric

    method

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    32/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 32 of 54

    3. ASTME1184 Standard practice for electrothermal (graphite furnace) atomic absorption

    analysis

    4. APHA3111 Metals by flame atomic absorption spectrometry

    5. APHA3112 Metals by cold-vapour atomic absorption spectrometry

    6. APHA3113 Metals by electrothermal atomic absorption spectrometry

    7. APHA3114 Arsenic and selenium by hydride generation/atomic absorption

    spectrometry

    D. Atomic Emission and X-Ray Fluorescence

    1. AS1502 Glossary of terms used in X-ray spectroscopy

    2. AS2563 Wavelength dispersive X-ray fluorescence spectrometers - determination of

    precision

    3. AS2883 Analysis of metals - Procedures for the setting up, calibration and

    standardization of atomic emission spectrometers using arc/spark discharge

    4. AS3641.1 Recommended practice for atomic emission spectrometric analysis Part 1Principles and techniques

    5. ASTME135 Standard terminology relating to analytical chemistry for metals, ores and

    related materials

    6. ASTME158 Standard practice for fundamental calculations to convert intensities into

    concentrations in optical emission spectroscopy analysis

    7. ASTMD305 Standard establishing and controlling spectrochemical analytical curves

    8. ASTME826 Standard practice for testing homogeneity of materials for the developmentof reference materials

    E Inductively Coupled Plasma

    1. AS3641.2 Recommended practice for atomic emission spectrometric analysis Part 2

    Inductively coupled plasma excitation

    2. APHA3120 Metals by plasma emission spectroscopy using inductively coupled plasma

    source

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    33/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 33 of 54

    F. Gas and liquid chromatographs

    1. AS3741 Recommended practice for chemical analysis by ion-chromatography

    2. ASTMD1945 Standard test method for analysis of natural gas by gas chromatography

    3. ASTMD4626 Standard practice for calculation of GC response factors

    4. ASTME260 Standard practice for packed column chromatography

    5. ASTME355 Standard practice for gas chromatography terms and relationships

    6. ASTME516 Standard practice for testing thermal conductivity detectors used in gas

    chromatography

    7. ASTME594 Standard practice for testing flame ionization detectors used in gas or super

    critical fluid chromatography

    8. ASTME682 Standard practice for liquid chromatography terms and relationships

    9. ASTME685 Standard practice for use of fixed wavelength photometric detectors used

    in LC

    10. ASTME697 Standard practice guide for use of EC detectors in gas chromatography

    11. ASTME840 Standard practice guide for use of FP detectors in gas chromatography

    12. ASTME958 Standard practice for measuring spectral bandwidth of UV/Vis

    spectrophotometers

    13. ASTME1151 Standard practice for ion chromatography terms and relationships

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    34/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 34 of 54

    APPENDIX G

    CALIBRATION REQUIREMENTS FOR CONSTRUCTION MATERIALS TESTING

    EQUIPMENT

    This appendix lists the items of equipment and their calibration requirements within the

    Construction Materials Test Category in the test areas covering aggregates, bituminous

    materials, concrete and steel. Specific calibration requirements for other test areas are

    included in other Supplementary Criteria listed as follows:

    Supplementary Criteria No. 15 for non-destructive tests of welding

    Supplementary Criteria No. 16 for foundation testing

    Supplementary Criteria No. 17 for building components and related testing

    Supplementary Criteria No. 18 for ground investigation testing and sampling

    Supplementary Criteria No. 19 for soil and rock testing

    For construction materials testing equipment, the criteria in this appendix should be read first

    over any other appendices in this supplementary criteria. Where conflicting criteria exist,

    the requirements of this appendix shall take precedence. The calibration periods listed here

    are the maximum acceptable, and shorter time periods may be required due to frequency of

    use and storage conditions.

    Environment

    General policy on environment is covered under section 5.3 of HOKLAS 003.

    In addition to the comments therein it is possible for an accreditation to be granted for tests

    conducted under more than one set of environmental conditions provided that these are :

    In accordance with the standard concerned and

    The test report indicates clearly which criteria have been adopted.

    Equipment

    Criteria are contained elsewhere in this Supplementary Criteria in respect of other items

    under different Test Categories. Several of the construction materials standards do however

    contain details of items of equipment peculiar to the test method concerned. Calibration and

    maintenance of these items should generally satisfy one of the following :-

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    35/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 35 of 54

    Where the standard concerned gives a calibration procedure, this procedure shall be

    followed and the maximum period between successive calibrations shall be that

    specified in the standard.

    Where no calibration procedure is given in the standard but tolerances on

    dimensions or mass are provided, a check against the specified values shall be

    made and recorded at least once every 12 months unless a calibration interval is

    given in the standard.

    Equipment for which no detailed criteria or tolerances are given should be regularly

    inspected.

    Solutions

    There are many instances where solutions are required in connection with both the physical

    and chemical testing of materials.

    It is not practical to give detailed guidance on the storage and shelf life of these but

    laboratories involved should either;

    Make up fresh solutions as and when required, where the standard concerned gives

    a calibration procedure but without a calibration interval, the calibration interval

    shall be 6 months, or

    Initiate a programme for ascertaining the shelf life of the solutions concerned.

    In the case of distillation and/or de-ionizing of water, the equipment concerned should be

    checked for satisfactory performance at least once every 6 months.

    Reference Materials

    Reference materials called up in standards shall be obtained from the sources stated in the

    standard and stored in the prescribed manner. Specific HOKLAS policy on referencematerials is covered under clause 5.6.H (f) of HOKLAS 003.

    Specific HOKLAS Calibration Criteria

    The following table sets out the maximum periods between successive calibrations for

    equipment to be used in connection with testing of aggregates, bituminous materials, concrete,

    steel, and soils within the Construction Materials Test Category. Specific HOKLAS policy

    for calibration is covered under section 5.6 and 5.6.H of HOKLAS 003.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    36/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 36 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipment

    requirements

    2 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003.

    6 months Repeatability check using referencemasses of higher grades.

    BALANCE(Reference)

    1 month One-point check using reference massesto balance capacity

    (a) 3 years or immediately

    following servicing

    Calibrate using reference masses in

    accordance with the national standardmethod or other equivalent standards.

    (b) 6 months Repeatability check using referencemasses in accordance with a standard

    method. Adjust the balance and carryout a full calibration if the requiredtesting accuracy is exceeded.

    (c) Daily before use One-point check using a known mass ator close to a frequently encounteredvalue or the balance capacity for a

    specified range.

    BALANCE

    (Working)

    (d) Each weighing Zero check after tare.

    CALLIPERS

    (Vernier or digital readout)

    (Accuracy equal to and betterthan 0.01mm)

    2 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS

    003

    (Accuracy between 1 mm and

    0.01mm)

    1 year Calibrate using reference gauge blocks

    or callipers checker. A sufficient

    number of readings shall be takencovering the expected working range.

    (Accuracy not better than 1mm) Initial only Callipers from reputable manufacturersmay be used.

    Callipers of any accuracies Before use Zero check.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    37/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 37 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    DIAL GAUGE

    (Accuracy equal to and better

    than 0.01mm)

    1 year By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003

    (Accuracy not better than0.01mm)

    1 year Calibrate using reference gauge blocksor a micrometer type calibrator forcompliance with BS907 or AS2103.

    LINEAR DISPLACEMENT

    MEASURING DEVICE

    (LVDT, digimatic gauge, Demecgauge or any other transducertypes)

    (Accuracy equal to and betterthan 0.01mm)

    1 year

    Before use

    By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003

    One-point check using gauge block.

    1 year Calibrate the measuring device togetherwith the same readout unit as is used for

    test using reference gauge blocks or a

    micrometer type calibrator. Thereadout unit can be an electrical displayunit, digital display unit or data logging

    system.

    (Accuracy not better than0.01mm)

    Before use One-point check using gauge block.

    MASS(Working-hanger weights)

    1 year Calibrate using reference balance.

    MICROMETER

    (Accuracy equal to and better

    than 0.01mm)

    2 years By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003

    (Accuracy not better than0.01mm)

    1 year Calibrate using reference gauge blocks.A sufficient number of readings shall betaken covering the expected workingrange.

    Micrometer of any accuracies Before Use Zero Check.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    38/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 38 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    OVEN (a) 3 years or immediatelyfollowing servicing or eachchange in temperature setting

    Check temperature variation in workingspace using different referencethermocouples at the same time and also

    check evaporation rate for compliancewith the requirements of AS 2853 andAS 1289 or other relevant standardmethods.

    (b) 6 months Check temperature at centre of usable

    oven space inside an empty oven using areference thermocouple. Ten readingsshall be taken at 3-minute intervals with

    the oven set at the working temperaturerange.

    PRESSURE GAUGECALIBRATOR

    (Reference standard)

    Dead weight tester 5 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003

    Manometer 5 years By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS

    003

    PRESSURE GAUGE(Reference)(Electrical transducer type)

    2 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003

    PRESSURE GAUGE

    (Working gauge)

    Bourdon tube or hydraulictype

    1 year By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003

    Transducers types 1 year By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003

    RULE 2 years Check against reference length standardwith a higher accuracy. Check at aminimum of 5 points over the entire

    length.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    39/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 39 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    SIEVE

    (Reference woven wire cloth

    sieves, for performance checkingof test sieves)

    Initial only Sieves with recognized manufacturers

    certificates which certify conformancewith BS410: 1986 shall be used.Reference sieves shall be downgraded totest sieves after 30 passes.

    (Perforated plate test sieves

    apertures larger than or equal to4mm)

    300 passes but not exceeding 6

    months

    Check aperture and bridge width using

    calibrated callipers, optical projection, orother appropriate devices in accordancewith BS410 : 1986.

    (Woven wire cloth test sieves apertures less than 4mm)

    300 passes but not exceeding 6months

    Check aperture using optical projectionor other appropriate equipment in

    accordance with BS410 : 1986.Alternatively, check performance ofapertures using certified referencesamples or performance check sampleswith reference sieves in accordance withBS 812:Part 100: 1990 or BS 1377:Part1:1990. Sieves with manufacturers

    certificates which certify conformancewith BS410:1986 may be used directlywithout initial checking.

    THERMOCOUPLE(for checking ovens)

    3 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS

    003

    6 months One-point check, at ice point or someother appropriate temperature.

    THERMOMETER -(Mercury-in-glass)

    (Accuracy equal to and betterthan 0.5C)

    5 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS

    003

    6 months One-point check, at ice point or someother appropriate temperature.

    (Accuracy not better than 0.5C) 5 years Calibrate over the expected workingrange using a reference thermometer and

    suitable calibration bath.

    6 months One-point check, at ice point or someother appropriate temperature.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    40/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 40 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    THERMOMETER (Electronic, digital and platinumresistance)

    3 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS

    003

    6 months One-point check, at ice point or someother appropriate temperature.

    TIME MEASURING DEVICE

    (a) Accuracy equal to or betterthan 0.5 second

    (b) Accuracy not better than 0.5

    second

    1 year

    3 months

    Calibrate against reference time standardin accordance with relevant calibration

    standard.

    Check against time given by radio time

    (RTHK) signal for at least half an hour.

    VACUUM GAUGE

    (a) Mechanical or hydraulic

    (b) Transducer

    2 years

    1 year

    By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003

    By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003

    VOLUMETRIC

    GLASSWARE(burettes, pipettes, volumetricflasks including measuring

    cylinders)

    Initial only Check by weighing the amount ofdistilled water that the vessel contains ordelivers at a measured temperature,applying any temperature corrections

    necessary using the Tables inBS1797:1987. The mass of water shallbe determined to within 0.01%. Thechecking shall be repeated a sufficientnumber of times to obtain the averagevolume. Where a vessel is to be used at

    a particular graduation mark only,checking may be limited to that mark butthe vessel must be labelled to indicateclearly the limited checking.

    AGGREGATES

    ABRASION MACHINE 1 year Check dimensions of moulds, trays and

    plates; speed of rotation of machine andmass of weights in accordance with BS812:Part 113: 1990

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    41/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 41 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    ACCELERATED

    POLISHING MACHINE

    1 year Check rate of flow of flowmeter,rotational speed of rad wheel, planes ofrotation of tyred and road wheels, free

    force of rubber wheels, and rubber tyredand road wheel dimensions inaccordance with BS 812:Part 114:1990.

    ELONGATION GAUGE 1 year Check with calibrated callipers ormicrometer

    FRICTION TESTER 1 year Check dimensions and mass of rubberslider in accordance with BS 812:Part

    114:1990.

    Before use Check specimen contact angle of rubberslider and verticality of column inaccordance with BS 812:Part 114:1990.

    IMPACT TESTING DEVICE 1 year Check dimensions of cylindrical steelcup and metal hammer, mass of circular

    metal base and metal hammer, and fallof hammer in accordance with BS812:Part 112: 1990.

    THICKNESS GAUGE

    (for determination of flakinessindex)

    1 year Check with calibrated callipers or

    micrometer.

    BITUMINOUS MATERIALS

    CENTRIFUGE(for extraction of binder)

    TACHOMETER

    1 year

    5 years

    Check the operating speeds using acalibrated tachometer

    By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS

    003

    MARSHALL TESTING

    MACHINE

    (a) Stability load

    measurement

    6 months By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003 in accordance with BS 1610 or BSEN 10002 or other equivalent standards.

    (b) Flow displacementmeasurement

    1 year Refer to previous section on dial gaugeor linear displacement measuring device.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    42/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 42 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    CONCRETE AND ITS CONSTITUENT MATERIALS

    CEMENT FINENESS

    TESTING APARATUS

    (a) Lea & Nurse 1 year By use of certified reference materialas defined in clause 5.6.H.(f) of

    HOKLAS 003.

    (b) Rigden or Blaine 3 months By use of certified reference materialas defined in clause 5.6.H.(f) of

    HOKLAS 003.

    CHLORIDE ION

    PENETRATION TESTER

    (a) Voltage measuring device 1 year By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevant

    standard.

    (b) Current measuring device

    (c) Reagent NaCl and NaOH

    1 year

    Freshly made

    By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevant

    standard.

    Reagents to be at least analytical grade

    COMPACTING FACTOR

    APPARATUS

    1 year Check compliance with requirement inaccordance with CS1 : 1990.

    COMPRESSION TESTING

    MACHINE

    (a) Load rate 1 month By use of calibrated timing device

    against criteria for tests in accordance

    with CS1: 1990.

    (b) Load verification 6 months or after relocation of the

    machine (Note: An additional checkat 3 months shall be performed if

    the two previous checks haveindicated a change of machinegrading)

    By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003 in accordance with BS 1610 or BS

    EN 10002 or other equivalent standards.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    43/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 43 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    COMPRESSION TESTING

    MACHINE(Contd)

    (c) Machine plattens auxiliaryplattens and spacer blocks

    (i) Surface condition 1 month Visual inspection check for signs ofwear and damage.

    (ii) Flatness, parallelism 1 year Using reference checking devices inaccordance with CS1 : 1990.

    (d) Performance 6 months or after relocation of themachine

    By a competent calibration body asdefined in the clause 5.6.H (d) of

    HOKLAS 003 in accordance with BS1881 or CS1:1990.

    (e) Strain cylinders(for checking theperformance of the machine)

    2 years By a competent calibration body asdefined in the clause 5.6.H (d) ofHOKLAS 003 in accordance with NIS0409 or BC415

    (f) Proving devices(for calibration of force

    testing machines)

    2 years By a competent calibration body asdefined in the clause 5.6.H (d) of

    HOKLAS 003 in accordance with BS1610 or BS EN 10002 or otherequivalent standards.

    CREEPTESTINGAPPARATUS(Load verification with aservo-controlled system)

    Each test By a competent calibration body asdefined in the clause 5.6.H (d) ofHOKLAS 003

    CURING TANK

    (CUBE OR OTHER SAMPLE) 6 months

    Daily

    Check the efficiency of circulation andtemperature distribution at 60%

    capacity, with different referencethermometers at minimum 9 points at thesame time.

    Check the max/min temperature.

    CUBE MOULD

    Dimensions, flatness, squarenessand parallelism

    6 months Check compliance with requirement inaccordance with CS1 : 1990..

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    44/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 44 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    CYLINDER (Reference-check

    the cylindricity of concrete

    core)

    5 years By a competent calibration body asdefined in the clause 5.6.H (d) ofHOKLAS 003 or by means of

    appropriate reference devices

    FLOW CONE 1 year Check apparatus in accordance withASTM C939 Clause 8 using a referencetime measuring device

    FLOW TABLE 1 year Check apparatus in accordance withBS1881:Part 105:clause 3.1

    MIST CURING ROOM 6 months

    Daily

    Check temperature using differentreference thermometers at a minimum of

    9 points at the same time or checkmaximum temperature fluctuation using

    a reference thermometer at minimum of9 points, and check humidity distributionto ensure no dry surfaces. Compare theeffectiveness with equivalent tank cured

    specimens

    Check the max/min temperatureCheck wetness of surfaces

    PFA FINENESS TESTING

    APARATUS(check aperture size of standardsieve)

    6 months By use of certified reference materialsas defined in clause 5.6.H.(f) ofHOKLAS 003.

    SLUMP

    (a) Cone 1 year Check dimensions in accordance withCS1:1990.

    1 week Check ovality of base and top to

    CS1:1990.

    (b) Rod 1 year Check dimensions in accordance withCS1:1990.

    STEEL

    ANCHOR BOLT PULL OUT

    AND PROOF LOAD TESTERLoad verification or forcetransducer

    1 year By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevantstandard

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    45/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 45 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    BENDING TESTING

    MACHINE

    (a) Load rate 1 month By use of calibrated timing deviceagainst criteria for tests in accordancewith CS1:1990

    (b) Formers 1 month Visual inspection for wear

    EPOXY COATED BAR

    (a) Coating thickness gauge 1 year By use of reference device in accordance

    with relevant standards

    (b) Holiday detector 1 year By use of reference device in accordancewith relevant standards

    EXTENSOMETER

    (for rebars and structural steel) 1 year By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003 in accordance with BS 3846:1970.

    EXTENSOMETER(for coupler and strand)

    (a) Mechanical type 1 year By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevantstandard.

    (b) Optical and transducer types 1 year By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevant

    standard.

    HARDNESS TESTER

    (a) Brinell, Vickers and

    Rockwell Machines

    1 year By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevantstandard.

    (b) Indenter 3 years By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevantstandard.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    46/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 46 of 54

    Type of equipmentRecommended maximum period

    between successive calibrations

    Calibration procedure or

    guidance documents and equipmentrequirements

    IMPACT TESTER

    Charpy V or U notch

    machine

    1 year By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevantstandard.

    TORQUE WRENCH(For checking of power-operated

    and hand-operated friction bolts)

    1 year By a competent calibration body as

    defined in clause 5.6.H (d) of HOKLAS003 in accordance with a relevantstandard

    TENSILE TESTING

    MACHINE

    (a) Load rate 1 month By use of calibrated timing device

    against criteria for tests in accordancewith BS 18.

    (b) Load verification 6 months or after relocation of themachine (Note: An additional checkat 3 months shall be performed iftwo previous checks have indicated

    a change of machine grading.1 month)

    By a competent calibration body asdefined in clause 5.6.H (d) of HOKLAS003 in accordance with relevantstandards.

    (c) Specimen gaugingequipment

    Spot check on prepared samples.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    47/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 47 of 54

    APPENDIX H

    CALIBRATION REQUIREMENTS FOR SELECTED PHYSICAL AND

    MECHANICAL TOY TESTING EQUIPMENT

    The general principles in Appendix C apply to common equipment such as thermometers,

    balances, mass, thermohygrographs, stopwatches, pressure gauges, ovens, vernier callipers,

    micrometers, glassware, rules, furnaces, etc. For chemical testing, the general principles for

    the use of analytical instrumentation as explained in Appendix F are relevant.

    The calibration requirements for selected equipment used in toy testing are detailed below.

    Type of equipmentRecommended period between

    successive calibrationsCalibration procedure

    and requirements

    FORCE AND TORQUE

    MEASURING EQUIPMENT

    USED FOR TOY TESTING

    SUCH AS PUSH PULL GAUGE,

    TORQUE GAUGE AND

    TENSION GAUGE

    (a) Portable spring type with dialgauge

    (b) Non-portable spring type withdial gauge

    One year (complete)*Six months*

    (repeatability of zero)

    Two years* (complete)Six months*

    (repeatability of zero)

    Dial gauge mechanism should bechecked for repeatability of zero with

    fast and slow movement in tensionand compression, or clockwise andanticlockwise directions whereappropriate

    FIXTURESincluding accessibility probes, smallparts cylinder, bite test clamp, rattletest fixture, flexure tester,compression test disc, tyre removal

    metal hooks, metal cylinders, straightrods, pins, steps construction for

    tumble tests, etc.

    Initial calibration for dimensionalcompliance *

    Yearly inspection for damage and

    wear *

    IMPACT MEDIUM FOR DROP

    TEST

    Initial calibration for hardness ofmedium *

    Yearly inspection for damage and

    wear *

    PROJECTILE VELOCITY

    TESTER

    Two years * The distance between the starting andstopping sensor and the accuracy ofthe timer should be calibrated. Thesoftware for calculating the velocityfrom the distance and time readings

    should be verified.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    48/54

    HOKLASSupplementary

    Criteria No.2

    Issue No. 4

    Page 48 of 54

    Type of equipmentRecommended period between

    successive calibrationsCalibration procedure

    and requirements

    SHARP EDGE TESTER Initial calibration*

    (complete)

    One month*(force calibration)Six months *(inspection of mandrel condition)

    Calibration should be in accordance

    with the requirements of relevant toytesting standards, such as those formandrel diameter, rotation speed,surface hardness, surface roughnessand force. Mandrel force calibrationmay be performed using a calibrated

    dead weight.

    SHARP POINT TESTER Initial calibration *Six months *(inspect the end cap for wear)

    Calibration should be in accordancewith the requirements of relevant toytesting standards, such as those for

    dimensions of gauging slot, thicknessof end cap, micrometer graduation

    and force of spring.

    After a piece of equipment is calibrated, a staff member of the laboratory should check

    whether the characteristics of the equipment as shown by the calibration results are within the

    tolerance range specified by the test standard. Use of out of tolerance equipment is not

    recommended. In exceptional cases where it is impractical to repair or replace the

    equipment, the out of tolerance equipment may be used only if the following conditions are

    met :

    (a) The laboratory has at least one set of working equipment with characteristics withinthe specified tolerance range which can be used to determine compliance for critical

    cases.

    (b) The laboratory has evaluated the effects of the out of tolerance equipment and

    documented the conditions under which the out of tolerance equipment will not affect

    test results.

    (c) The out of tolerance equipment is properly labelled to alert test operators that such

    equipment is out of tolerance and should only be used under the conditions

    documented in (b). Under all other conditions, the equipment mentioned in (a)

    should be employed.

    (d) Test operators are properly trained and fully aware of the meaning of the label

    mentioned in (c), the limitations of the use of the out of tolerance equipment and the

    actions to take when the defect of the equipment may affect test results.

  • 7/27/2019 HOKLAS Supplementary Criteria No. 2

    49/54