how accurate can a custom-made implant be positioned in large acetabular defects? marieke baauw gijs...
TRANSCRIPT
![Page 1: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/1.jpg)
How accurate can a custom-made implant be positioned in large
acetabular defects?
Marieke BaauwGijs van Hellemondt
Bart SwierstraMiranda van Hooff
Maarten Spruit
![Page 2: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/2.jpg)
Disclosure
• Two authors (GH and MS) are paid faculty for Mobelife, Leuven, Belgium.
![Page 3: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/3.jpg)
![Page 4: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/4.jpg)
Defect analysis and classification
![Page 5: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/5.jpg)
Defect analysis and classification
![Page 6: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/6.jpg)
Defect analysis and classification
![Page 7: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/7.jpg)
Pre-op planning & implant design– Custom porous augment and custom plate: monoblock or
modular– Inclination, anteversion angle and center of rotation
![Page 8: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/8.jpg)
Surgery
![Page 9: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/9.jpg)
AP radiographs pre- and post-op
![Page 10: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/10.jpg)
Methods
• 16 cases with Paprosky type 3 defects• Pre- and post operative CT-scans• Comparison of INCL, AV, ROT and COR
![Page 11: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/11.jpg)
Results
• 9 implants perfectly positioned• 7 implants were malpositioned in one or more
parameters – 1 with respect to INCL (> 10°)– 3 with respect to AV (> 10°)– 4 with respect to rotation (> 10°)– 5 with respect to the COR (> 5mm )
• One intra-operative complication: anterior wall acetabular fracture.
![Page 12: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/12.jpg)
Discussion
• No literature to directly compare with• Free hand cup positioning is inaccurate even
in pimary hip surgery• With Patient specific implants in primary hips
slightly better• Compared to existing literature on acetabular
component placement in less challenging defects our results are at least comparable.
![Page 13: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/13.jpg)
Discussion
• Clinical results:–26 patients– FU 3-36 months–1 dislocation–No infection–No implant removal
![Page 14: How accurate can a custom-made implant be positioned in large acetabular defects? Marieke Baauw Gijs van Hellemondt Bart Swierstra Miranda van Hooff Maarten](https://reader035.vdocument.in/reader035/viewer/2022062715/56649d745503460f94a53f77/html5/thumbnails/14.jpg)