how to calibrate microscope in pax-it

5
How to do the Microscopy Calibration with PAX-it 1. Add the Magnification into the software. Go to Database, click Utilities and Select Setup Database tab. 2. Click “Hard Drive” button 3. Select the field “Field #3: Magnification”, click the Configure Lookup Table” button, and then click “Add” button and type in the Microscope Magnification you want to calibrate. Finally click OK. 4. Click the “PaxCam” button and choose the right camera you are going to use. Put the standard scalar on the sample

Upload: phthysyllysm

Post on 29-Oct-2015

415 views

Category:

Documents


32 download

DESCRIPTION

How to calibrate a microscope in PAX-it software

TRANSCRIPT

How to do the Microscopy Calibration with PAX-it

1. Add the Magnification into the software. Go to Database, click Utilities and Select Setup Database tab.

2. Click “Hard Drive” button

3. Select the field “Field #3: Magnification”, click the “Configure Lookup Table” button, and then click “Add” button and type in the Microscope Magnification you want to calibrate. Finally click OK.

4. Click the “PaxCam” button and choose the right camera you are going to use. Put the standard scalar on the sample stage and focus it to the best condition. And then click “Capture” button to take a picture. You could also Check “Preview” and Make sure you are choosing the right camera you are using.

5. Choose the right Magnification in the Magnification row and Type in the name you want to name your file. Finally Click OK.

6. Click the Calibrate button in the upper left corner. The full scale is 1000 um, so one unit is 10 um. Using the mouse to adjust the lines. In order to make more accurate, try to use the full scale under the microscopy. And then choose the measument unit and type in the number of units used duing calibration. Finally click OK.

7. The sentence “Would you like to copy the calibration to all writeable drivers?” will pop out. Click No since we save in the hard drive.

8. After each magnification calibration, need to double check. Close the image and then go to the “back to last folder” button. Double click the image to open it.

9. Click “Draw Horizontal” icon. Draw a line and check whether the measurement is correct.

10.Click the “close” in the upper right corner to close the image.