hplc columnsuct’s selectra® pfpp hplc column is an excellent choice for the retention and...

16
SELECTRA ® PFPP PENTAFLUOROPHENYLPROPYL STATIONARY PHASE HPLC COLUMNS INNOVATION THROUGH CHEMISTRY UCT

Upload: others

Post on 28-Dec-2019

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

SELECTRA® PFPPPENTAFLUOROPHENYLPROPYL STATIONARY PHASE

HPLC COLUMNS

I N N O VAT I O N T H R O U G H C H E M I S T R Y

U C T

Page 2: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

Product Benefits• Can be used for Reverse Phase, Normal

Phase, or HILIC separations

• Excellent first-choice column for method development

• Excellent for LC/MS/MS• Strongly retentive for basic compounds• Special selectivity compared to C18 phases• May exhibit pi-pi overlap• Fully endcapped

215.781.9255 • 800.385.3153

S E L E C T R A ® P F P P H P L C C O L U M N S

Selectra® PFPP HPLC ColumnsDimensions Particle Size Part Numbers

50 x 2.1 mm 1.8 µm SLPFPP50ID21-18UM

100 x 2.1 mm 1.8 µm SLPFPP100ID21-18UM

50 x 4.6 mm 1.8 µm SLPFPP50ID46-18UM

100 x 4.6 mm 1.8 µm SLPFPP100ID46-18UM

10 x 2.1 mm guard (sold as a 2 pack) 1.8 µm SLPFPPGDC20-18UM

10 x 4.6 mm guard (sold as a 2 pack) 1.8 µm SLPFPPGDC46-18UM

50 x 2.1 mm 3 µm SLPFPP50ID21-3UM

100 x 2.1 mm 3 µm SLPFPP100ID21-3UM

50 x 4.6 mm 3 µm SLPFPP50ID46-3UM

100 x 4.6 mm 3 µm SLPFPP100ID46-3UM

150 x 4.6 mm 3 µm SLPFPP150ID46-3UM

10 x 2.1 mm guard (sold as a 2 pack) 3 µm SLPFPPGDC20-3UM

10 x 4.6 mm guard (sold as a 2 pack) 3 µm SLPFPPGDC46-3UM

50 x 2.1 mm 5 µm SLPFPP50ID21-5UM

100 x 2.1 mm 5 µm SLPFPP100ID21-5UM

50 x 4.6 mm 5 µm SLPFPP50ID46-5UM

100 x 4.6 mm 5 µm SLPFPP100ID46-5UM

150 x 4.6 mm 5 µm SLPFPP150ID46-5UM

250 x 4.6 mm 5 µm SLPFPP250ID46-5UM

10 x 2.1 mm guard (sold as a 2 pack) 5 µm SLPFPPGDC20-5UM

10 x 4.6 mm guard (sold as a 2 pack) 5 µm SLPFPPGDC46-5UM

HPLC Guard HolderHPLC Guard Cartridge Holder SLGRDHLDR

UHPLC Guard Cartridge Holder SLGRDHLDR-HP

Replacement Peek Tip for Holder SLGRDHLDR-TIP (2/pk)

Page 3: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

www.unitedchem.com

The Selectra® PFPP (pentafluorophenylpropyl) phase interacts strongly with basic, halogenated and nitrogen-containing analytes. The pentafluorophenyl ring exhibits higher selectivity for electron rich aromatic containing molecules as compared to other sorbents. This phase also shows affinity for molecules with fluorine and other halogen atoms. The PFPP phase uses multiple selectivity mechanisms including hydrogen bonding, dipole-dipole, pi-pi overlap, hydrophilic (HILIC), and hydrophobic interactions.

UCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes include nucleosides, nucleotides, purines, pyrimidines, [beta]-blockers, tricyclic antidepressants, fluorinated corticosteroids, and taxanes. Selectra® PFPP columns are ideal for pharmaceutical and natural products research.

UCT’s Selectra® PFPP is well suited to LC-MS instrumentation due to its reliability and efficiency with acidic mobile phases. The Selectra® PFPP shows excellent peak shape in neutral mobile phases as well. The Selectra® PFPP columns are available packed with 3 or 5 µm particles and are offered in various column dimensions.

Retention Profile of Imipramine on a

RP Region HILIC Region

C18 does not have HILIC character, whereas, PFPP does (one of PFPP’s major advantages).

Column 1 Selectra® PFPP, 5 cm x 2.1 mm, 3 µm

Column 2 Selectra® C18, 5 cm x 2.1 mm, 3 µm

Mobile Phase 10mM NH3HCOOH: ACN

(adjusted to pH = 4 with Formic Acid)

Detector UV@254nm

Flow Rate 0.2 mL/min

Sample Imipramine

Page 4: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

215.781.9255 • 800.385.3153

S E L E C T R A ® P F P P H P L C C O L U M N S

Comparison of Benzodiazepenes on a C18 vs. PFPP

C18

PFPP

M...1CozB41021040 fo )pm 002 elps 81CozB41021040( 3 elpmaS morf 1 mapezanolC onimA-7 :DI aD 3.222/1.682 :)sriap 03( MRM+ fo CIX ax. 2.5e4 cps.

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5Time, min

0.00

5000.00

1.00e4

1.50e4

2.00e4

2.50e4

3.00e4

3.50e4

4.00e4

4.50e4

5.00e4

5.50e4

6.00e4

6.50e4

7.00e4

7.50e4

8.00e4

8.50e4

9.00e4

9.50e4

1.00e51.03e5

Inten

sity,

cps

aM...ozB41021040 fo )pm 002 elps PPFPozB41021040( 4 elpmaS morf 1 mapezanolC onimA-7 :DI aD 3.222/1.682 :)sriap 03( MRM+ fo CIX x. 7090.0 cps.

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5Time, min

0.00

5000.00

1.00e4

1.50e4

2.00e4

2.50e4

3.00e4

3.50e4

4.00e4

4.50e4

5.00e4

5.50e4

6.00e4

6.50e4

7.00e4

7.50e4

8.00e4

8.50e4

9.00e4

9.50e4

1.00e5

1.05e5

1.10e5

1.15e51.18e5

Inten

sity,

cps

Analyte C18 RT (min) PFPP RT (min)1. 7 Aminoclonazepam 1.11 1.26

2. Clonazepam 2.02 2.86

3. Lorazepam 2.16 2.16

4. Oxazepam 2.20 2.36

5. Alpha Hydroxy Alprazolam 2.22 2.45

6. Nordiazepam 2.70 3.18

7. Temazepam 2.75 3.12

8. Alprozolam 2.79 3.41

9. Diazepam 3.74 4.50

Column C18 100 x 2.1 mm, 5 µm

Mobile Phase A = 40% H2O w/ 0.1% Formic Acid B = 60% MeOH w/ 0.1% Formic Acid

Flow Rate 0.5 mL/min

Column Temp 50 °C

Injection Volume

5 µL

Column Selectra® PFPP LC 100 x 2.1 mm, 5 µm

Part Number SLPFPP100ID21-5UM

Mobile Phase A = 40% H2O w/ 0.1% Formic Acid B = 60% MeOH w/ 0.1% Formic Acid

Flow Rate 0.5 mL/min

Column Temp 50 °C

Injection Volume

5 µL

1

2

3

4, 5

6, 7

8

9

1

2

3

4

5

67

8

9

Page 5: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

www.unitedchem.com

Gabapentin/Pregabalin/Baclofen

XIC of +MRM (6 pairs): 160.2/97.1 Da ID: pregabalin from Sample 1 (10232013GabaPregaBaclo std1) of 10232013GabaPregaBaclo3.wi... Max. 1.5e6 cps.

1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0Time, min

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

2.2e6

2.4e6

2.6e6

2.8e6

3.0e6

3.2e6

3.4e6

Inten

sity,

cps

Analyte Precursor ion

Product ion 1

Product ion 2

Gabapentin 172.1 137.1 67.1

Baclofen 214.01 150.8 65.0

Pregabalin 160.2 124.1 97.1

Instrument: Agilent 1200 Binary Pump SL

Detector: ABSciex 4000 Qtrap MS/MS

Mode: ESI +

Column: Selectra® PFPP, 100 x 2.1 mm, 5 µm (UCT p/n SLPFPPID21-5UM)

Flow Rate: 0.5 mL/min

Column Temp: 50 °C

Injection Volume: 10 µL

Mobile Phase A: 0.1% Formic Acid in H2O

Mobile Phase B: 0.1% Formic Acid in MeOH

Isocratic: A:B (85:15)

Gabapentin

Pregabalin Baclofen

Page 6: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

215.781.9255 • 800.385.3153

Cocaine / Benzoylecgonine / Cocaethylene

XIC of +MRM (21 pairs): 290.1/105.1 Da ID: BE from Sample 3 (10182013PFPPCoc std2) of 10182013PFPPCOCISOcratic2.wiff (Turbo ... Max. 1.4e6 cps.

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5Time, min

0.0

1.0e5

2.0e5

3.0e5

4.0e5

5.0e5

6.0e5

7.0e5

8.0e5

9.0e5

1.0e6

1.1e6

1.2e6

1.3e6

1.4e6

1.5e6

1.6e6

1.7e6

1.8e61.9e6

Inte

nsity

, cps

AnalytePrecursor

ionProduct

ion 1Product

ion 2

Benzoylecgnonine 290.1 168.1 105.1

Cocaine 304.1 182.1 105.1

Cocaethylene 318.2 196.4 82.7

Instrument: Agilent 1200 Binary Pump SL

Detector: ABSciex 4000 Qtrap MS/MS

Mode: ESI +

Column: Selectra® PFPP, 100 x 2.1 mm, 5 µm (UCT p/n SLPFPP100ID21-5UM)

Flow Rate: 0.5 mL/min

Column Temp: 50 °C

Injection Volume: 10 µL

Mobile Phase A: 0.1% Formic Acid in H2O

Mobile Phase B: 0.1% Formic Acid in MeOH

Isocratic: A:B (30:70)

S E L E C T R A ® P F P P H P L C C O L U M N S

Cocaine

Benzoylecgonine

Cocaethylene

Page 7: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

www.unitedchem.com

Bath Salts

Analyte

MRM Transitions

Relative Retention Time

(minutes)Q1 Q3

1. Flephedrone 182.1 164.2 4.78

2. Methylone 208.1 160.1 5.26

3. Ethylone 222.0 131.0 5.84

4. Mephedrone 178.1 145.0 6.10

5. Butylone 222.0 131.0 6.36

6. Methcathinone 192.2 144.0 6.93

7. MDPV 276.2 126.1 8.67

8. Pyravalerone 246.2 105.2 8.83

Time %A %B

0.00 98 2

1.00 65 35

5.00 65 35

7.00 0 100

10.00 0 100

10.20 98 2

15.00 STOP

Instrument: Thermo Scientific Dionex Ultimate 3000 LC

Detector: Thermo Scientific TSQ Vantage tandem mass spectrometer

Mode: ESI +

Column: Selectra® PFPP, 100 x 2.1 mm, 5 µm (UCT p/n SLPFPP100ID21-5UM)

Flow Rate: 0.3 mL/min

Injection Volume: 10 µL

Mobile Phase A: 0.1% Formic Acid in H2O

Mobile Phase B: 0.1% Formic Acid in MeOH

Gradient:

1

2

34

5

6

7

8

Page 8: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

Ephedra Alkaloids

RT: 0.00 - 24.99 SM: 7G

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24Time (min)

0

20

40

60

80

1000

20

40

60

80

1000

20

40

60

80

100

Rel

ativ

e Ab

unda

nce

0

20

40

60

80

1004.00

5.38 6.26 7.07 7.782.96

10.15

11.41

12.888.71 13.87

15.1117.03

19.7213.79

20.54

23.7018.66

NL: 1.82E4TIC F: + c ESI SRM ms2 168.104 [90.969-90.971, 106.969-106.971] MS 100ppb_std

NL: 1.27E4TIC F: + c ESI SRM ms2 152.141 [91.019-91.021, 115.019-115.021] MS 100ppb_std

NL: 2.83E4TIC F: + c ESI SRM ms2 166.173 [91.009-91.011, 115.019-115.021] MS 100ppb_std

NL: 1.51E4TIC F: + c ESI SRM ms2 180.175 [90.989-90.991, 147.039-147.041] MS 100ppb_std

S E L E C T R A ® P F P P H P L C C O L U M N S

AnalyteRT

(min)Precursor

ionProduct

ion 1Product

ion 2

Synephrine 4.00 168.1 91.0 107.0

Norephedrine 10.15 152.1 91.0 115.0

Norpseudoephedrine 11.41 152.1 91.0 115.0

Ephedrine 15.11 166.2 91.0 115.0

Pseudophedrine 17.03 166.2 91.0 115.0

Methylephedrine 20.54 180.2 91.0 147.0

Instrument: Thermo Scientific Dionex Ultimate 3000 LC

Detector: Thermo Scientific TSQ Vantage tandem mass spectrometer

Mode: ESI +

Column: Selectra® PFPP, 100 x 2.1 mm, 3 µm (UCT p/n SLPFPP100ID21-3UM)

Guard Column: Selectra PFPP, 10 x 2.0 mm, 3 µm (UCT p/n SLPFPPGDC20-3UM)

Flow Rate: 0.3 mL/min

Column Temp.: 40 °C

Injection Volume: 8 µL

Mobile Phase A: 10mM Ammonium Acetate

Mobile Phase B: MeOH

Isocratic: A:B (80:20)

Synephrine

Norephedrine

Norpseudoephedrine

Ephedrine Pseudophedrine

Methylephedrine

Page 9: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

Diquat/Paraquat

Norm.

120

100

80

60

40

20

0

2 4 6 8 10

DAD1 A, Sig=258,4 Ref=550, 100 (010914\PQDQ2.D)DAD1 B, Sig=308,4 Ref=550, 100 (010914\PQDQ2.D)

Paraquat Diquat

1.67

5

3.02

1

1.67

5

ParaquatDiquat

Instrument: HP 1050

Detector: HP Photo Diode Array

Column: Selectra® PFPP, 100 x 2.1 mm, 3 µm (UCT p/n SLPFPP100ID21-3UM)

Flow Rate: 0.2 mL/min

Injection Volume: 1 µL

Mobile Phase A: 10mM Ammonium Acetate (pH=4)

Mobile Phase B: Acetonitrile

Isocratic: A:B (99:1)

Monitoring Wavelength: Diquat: 308 nm

Paraquat: 258 nm

Page 10: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

S E L E C T R A ® P F P P H P L C C O L U M N S

Sympathomimetic Amines

Analyte

MRM Transitions Relative Retention Time

(minutes)Q1 Q3

1. Amphetamine 136.1 91.2 5.41

2. MDA 180.2 163.1 6.24

3. Methamphetamine 150.1 91.2 6.35

4. MDMA 194.2 163.1 7.29

Instrument: Agilent 1200 Binary Pump SL

Detector: ABSciex API 4000 Qtrap MS/MS

LC Column: Selectra® PFPP HPLC Column 50 x 2.1 mm, 5 µm (UCT P/N SLPFPP50ID21-5UM)

Mode: ESI +

Flow Rate: 0.3 mL/min

Injection Volume: 10 µL

Mobile Phase A: 0.1% Formic Acid in D.I. H2O

Mobile Phase B: 0.1% Formic Acid in Methanol

Isocratic: A:B (30:70)

SELECTRA® PFPP HPLC Column

1

2

3

4

Page 11: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

Column Care & Usage

Each UCT, LLC high performance liquid chromatography (HPLC) column is individually packed and tested to ensure superior performance. A HPLC CoA is included with each column. It contains a chromatogram, the column serial number, and the lot number of the packing material. Retain this information for as long as you have the column; it may be useful if troubleshooting is ever required.

Guard Columns & Filters

In-line filters and/or guard cartridges can extend the life of an analytical column. They are connected in-line prior to the analytical column. l Pressure Recommendation UCT, LLC HPLC columns are silica based. To ensure optimal column life, operating pressures of 3000psi or lower are recommended. Column pressures may increase as the column ages as particulates from the system accumulate on the column. Sudden increases in pressure are usually a result of a blocked frit. Pressure will vary with different mobile phases. For example Water/Methanol mixtures will generally give higher back pressure than Water/Acetonitrile mobile phases. (see Table I)

l Guard Cartridges Guard cartridges are used to capture impurities that may otherwise lodge on the HPLC column. Guard cartridges are especially useful with samples from biological sources as these may contain lipids and proteins that pass through frits can quickly block columns. Guard cartridges should have the same phase as the column they are protecting. Guard cartridges should be replaced when the chromatography begins to deteriorate or when the guard cartridge contributes excessive back pressure to the HPLC system.

Page 12: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

Mobile Phase

When shipped, the column contains the storage solvent listed on its HPLC CoA. Before use the first time, ensure that your initial planned mobile phase is compatible with this solvent (see Table II). If it is not, you must flush the column with an intermediate solvent that is compatible with both the storage solvent and your planned mobile phase. Be especially cognizant if you are using buffers; the storage solvent for most columns contains greater than 50% organic solvent, and contact with a buffer could cause a salting out effect. The resulting precipitate can plug the column. Flow Direction & Flow Rate

The arrows on the column label indicate the recommended flow direction.

Begin by connecting the inlet end of the column to the injector or autosampler and allow mobile phase to flow from the outlet end of the column into a beaker for 10–15 minutes. Gradually increase the flow rate. For recommended flow rates refer to Table III. Then, stop the mobile phase flow and connect the column to your detector. Because every LC system is unique, especially when used in gradient mode, your results may slightly differ from those obtained in our laboratory. UCT, LLC Technical Service can assist you in optimizing your separations. Be sure to record the operating pressure before calling.

Increasing Column Lifetime

Silica based UCT, LLC HPLC column packing materials have a pH operating range of 2-8. Extended use of any column at extreme pH can shorten column lifetime.

The upper temperature limit for silica based HPLC columns is 80 °C. Elevated temperatures can improve efficiency by lowering solvent viscosity, but column lifetime may be compromised.

Use of HPLC-Grade solvents is strongly recommended. Residue and chemical contaminants in non-HPLC grade solvents can alter a column’s selectivity and, potentially plug the inlet frit leading to an increased system pressure. Mobile phase filtering and degassing (either off-line or in-line) is highly recommended. Column lifetime is also governed by stationary phase type. Hydrocarbon phases, such as C18, are relatively chemically inert. Polar phases, such as cyano or amino, require somewhat more care as they can be chemically active.

S E L E C T R A ® P F P P H P L C C O L U M N S

Page 13: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

Column Maintenance

Columns should not be subjected to mechanical or pressure shock. This can cause irreversible damage to the column.

Columns should be run in the flow direction as marked on the column. The one exception is for column cleaning. The flow can be reversed to back-flush frits if blockages occur.

Do not store the column in an aqueous buffer, this will promote microbiological contamination. First flush the column with water and then with 50/50 organic solvent/water prior to storage.

Washing Procedure for Reverse-Phase Columns

Washing the column successively with non-polar eluents will usually remove accumulated impurities. Follow the washing sequence below, using 30mL of each solvent, to thoroughly clean the column.

1. Distilled water 90%, 10% Methanol

2. 0.5M H3P04 90%, 10% Methanol (Optional)*

3. Distilled water 90%, 10% Methanol (Optional)*

4. Methanol

5. Methanol/Chloroform (1:1) (Optional)

6. Methanol or Acetonitrile (Optional)

7. Distilled water 90%, 10% Methanol

8. Eluent to recondition column

*NOTE: Whenever step 2 is used, it must be followed by step 3.

It is recommended that columns be dedicated for the specific method when ion pairing reagents are used. This is because it is difficult to remove all of the ion pairing reagent.

l Protein Contamination

Proteins can adsorb onto columns causing loss of performance. In this situation rinse the column overnight with 20% 0.1M nitric acid/80% isopropanol at a flow rate of 1/5th the usual flow rate (i.e. at 0.2 mL/min for 4.6mm ID columns). Ensure that the rinse solution is directed directly to solvent waste and not through the detector.

l Lipid Contamination If lipids or other highly hydrophobic compounds have contaminated the column use the full washing procedure except replace step 5 with 100% chloroform or dichloromethane.

Page 14: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

2

1.8

1

1.6

1.4

1.2

0.8

0.6

0.4

0.2

0

2

1.8

1

1.6

1.4

1.2

0.8

0.6

0.4

0.2

00% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100%

MeOH/Water

MeCN/Water

MeCN/MeOH

III: Mobile Phase Viscosity ComparisonVi

scos

ity, c

PTable I: Mobile Phase Viscosity Comparison

S E L E C T R A ® P F P P H P L C C O L U M N S

Page 15: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

Table I: Solvent Miscibility Chart

Ace�c AcidAcetone

AcetonitrileBenzene

n-ButanolButyl Acetate

Carbon Tetrachloride Immiscible SolventsChloroform

Cyclohexane1,2-Dichloroethane

DichloromethaneDimethylformamide

Dimethyl SulfoxideDioxaneEthanol

Ethyl AcetateDi-Ethyl Ether

HeptaneHexane

MethanolMethyl-t-Butyl EtherMethyl Ethyl Ketone

Pentanen-Propanol

Iso-PropanolDi-Iso-Propyl Ether

TetrahydrofuranToluene

TrichloroethyleneWater

Xylene

Acet

ic Ac

idAc

eton

eAc

eton

itrile

Benz

ene

n-Bu

tano

lBu

tyl A

ceta

te

Carb

on Te

trach

lorid

eCh

loro

form

Cyclo

hexa

ne

1,2-

Dich

loro

etha

ne

Dich

loro

met

hane

Dim

ethy

lform

amid

e

Dim

ethy

l Sul

foxid

eDi

oxan

eEt

hano

lEt

hyl A

ceta

teDi

-Eth

yl Et

her

Hept

ane

Hexa

neM

etha

nol

Met

hyl-t

-But

yl Et

her

Met

hyl E

thyl

Keto

nePe

ntan

en-

Prop

anol

Iso-P

ropa

nol

Di-Is

o-Pr

opyl

Ethe

r

Tetra

hydr

ofur

anTo

luen

e

Trich

loro

ethy

lene

Wat

erXy

lene

Table II: Solvent Miscibility Chart

Page 16: HPLC COLUMNSUCT’s Selectra® PFPP HPLC column is an excellent choice for the retention and selectivity of amine-containing compounds, and charged bases. Typical compound classes

PRICES AND TERMSOur prices are subject to change without notice. The price in effect when we receive your order will apply. All prices are in US Dollars and are F.O.B. Terms of payment are net 30 days. MINIMUM ORDERSWe welcome all orders, therefore, we do not have a minimum order requirement. When ordering, please include your purchase order number, complete “Ship To” and “Bill To” address, catalog number, quantity, and description of product(s). Also include your name and a phone number where you can be reached should we have any questions concerning your order. SHIPMENTSNormal processing is within 24 hours after receipt of an order. Unless special shipping requests have been made, our trained staff will send all orders by UPS Ground service. The appropriate shipping charges (freight & insurance costs) will be added to the invoice, unless otherwise instructed by the customer.

SPECIAL PRICINGWe offer special pricing for volume purchases and standing orders. These discounts apply to bonded phase extraction column purchases only. Please call a sales representative for more information on special pricing qualifications.

RETURN POLICYOur Quality Manager will handle all returns. Before returning merchandise, please call to obtain a return authorization number from the quality manager. We will need to know the reason for the return, date of purchase, purchase order number and invoice number in order to issue a return authorization number. Return merchandise must be received before a credit can be issued. Returns will not be accepted after 90 days. A restocking fee of 25% of the price paid, or a minimum of $25.00 (whichever is greater) will be charged on all returns. WARRANTYAll products manufactured by UCT are guaranteed against defects in materials and workmanship for a period of 90 days after shipment. UCT will replace any items that prove to be defective during this time period. The exclusive remedy requires the end user to first advise UCT of the defective product by phone or in writing and must include order number, the lot number and the shipping date.

To initiate this action, photographs of the product, including packaging and labeling of the containers, must be submitted to the UCT Representative for approval. With approval a return authorization can be initiated, and must be received within 30 days. Once the materials arrive at UCT a further inspection of the materials must be completed and accepted by our Quality Manager prior to further action of credits or replacement. UCT’s total liability is limited to the replacement cost of UCT products.

This warranty does not apply to damage resulting from misuse.

4101-06-07 U C T

Contact Us

Phone: 215.781.9255 800.385.3153Fax: 215.785.1226

UCT, Inc. 2731 Bartram Rd. Bristol, PA 19007

Email: [email protected]: www.unitedchem.com

S E L E C T R A ® P F P P H P L C C O L U M N S