hypofractionation physics · motion-induced artifacts real life aapm tg-76 2006 the management of...
TRANSCRIPT
Hypofractionation – Physics Todd Pawlicki, Ph.D.Professor & Vice-Chair of Medical Physics
Objectives
• Understand special technical issues for hypofractionated treatments
• Be able to highlight relevant technical reports
• Discuss safety issues and mitigation strategies with hypofractionated treatments
Acknowledgements:
• UCSD physics team• William Song, PhD
The Issue
Dose deviation from the prescription for an error in a single fraction.
Two Categories
• Cases with stationary/rigid targets– Brain– Spine– etc
• Cases with moving/deforming targets– Lung– Liver– etc
Simulation
Planning
Treatment
Prescription
Guidance documents
Solberg et al PRO 2011
Potters et al IJROBP 2010
AAPM TG-101: Table of Contents
• Introduction and Scope• History and Rationale for SBRT• Current Status of SBRT-Patient Selection Criteria• Simulation Imaging and Treatment Planning• Patient Positioning, Immobilization, Target Localization,
and Delivery• Special Dosimetry Considerations• Clinical Implementation of SBRT• Future Directions
Safety and Quality
• Multidisciplinary team – May be everyone for smaller centers
• Choose a treatment site– Identify special issues for that site
• Team members express concerns, find solutions
• Everyone on the same page– Move forward only when preparation is done
Patient Selection
• Cooperative and understands instructions
• Can tolerate prolonged setup and treatments
• No resting tremors, uncontrolled pain, etc.
• Non-urgent cases only
Hamilton et al. Neurosurgery 1995Immobilization (body)
Framed Frame-less
AAPM TG-101 (2010)
SBRT Immobilization Accuracy
Accuracy summary:2 – 3 mm
Immobilization (head)
Framed Frame-less
Head immobilization accuracy ~ 1-2 mm
Immobilization – Other issues
• Patient comfort– Treatment delivery may also requires longer sessions– Simulation can take longer than usual (e.g., 4DCT)
• Treatment beams– Multiple beam angles– Arms over head even for lower spine targets
CT Simulation
Static targets• Can be similar as for non-
hypofractionated cases – Head vs. body
• MR can be helpful for target and normal tissue contouring (e.g. cord)
Moving targets• Should use a motion
management strategy– Breathing control,
compression, 4DCT
• Know the limitations of your strategy
Motion-Induced Artifacts
Example
Rietzel et al., Med Phys 2005;32(4):874-889
Motion-Induced Artifacts
Real Life
AAPM TG-76 2006 The Management of Respiratory Motion in Radiation Oncology
No general pattern of respiratory behavior can be assumed for each patient.
Respiratory Signal: TermsPeriod
Amplitude 0%
50%
100%
30% 70%
100% duty cycle = Beam ON between 0%-100% window50% duty cycle = Beam ON between 30%-70% window
Time
Am
plitu
de
Duty CycleGate 100 Gate 3070 Static Case
PTV PTV PTV
100% duty cycle0-100% window
50% duty cycle30-70% Window
4DCT Simulation
4DCTRespiratory signal
from system
X-ray ON
First couch position Second couch position Third couch position
Problems With Phase Sorting
Period Amplitude
Baseline
*Phase-to-amplitude relationship changes…
ITV ContouringMaxIP100(MIP100)
MaxIP3070(MIP3070)
AvgIP MinIP
30%40%50%60%70%80%90%
20%10%
0%Range:
30%40%50%60%70%
Range:
30%40%50%60%70%80%90%
20%10%
0%Range:
30%40%50%60%70%80%90%
20%10%
0%Range:
Contouring/Planning
• MaxIPs useful in lung (tumor density is higher than surrounding tissue)– If GATE100 treatment, then use MIP100 CT for ITV– If GATE3070 treatment, then use MIP3070 CT for ITV
• MinIPs useful in liver (tumor density is lower than surrounding tissue)
• Fuse the ITV onto AvgIP
• Plan on AvgIP
Treatment Planning
• Understand basic planning guidelines– Include entire dose calculation region
• Always use 3D planning– Conventional or intensity modulation– Achieve high dose conformality
• Multiple and/or non coplanar beams/arcs– Collision avoidance procedures in place
Multiple Beams
B Hoppe et al, IJROBP, 2008
5 – 11 static beam OR 1 – 3 arcs
To Gate or Not to Gate?
• Relevant questions– Motion > 0.5 cm?– Breathing regular? Period > 4 sec?– Patient compliance? – Patient tolerate longer treatment
time with gating?
• Decision at planning stage– GATE100 or GATE3070
Figure 9: AAPM TG-76 (Report 91), 2006
Planning • Decide on criteria for gating
– Note, most cases are not going to be gated
• ITV-PTV margin– At least 5 to 10 mm– Motion management does not mean zero margin
• Generally, prescription isodose covers 100% ITV– Adequate PTV coverage is also of concern
• Hotspot inside/outside PTV?– If inside, typically ≤ 110%– If outside, re-plan necessary
Isodose Coverage
• Isodose covers tightly around the ITV/PTV • No hotspots should be outside the PTV
Planning
• Normal tissues and constraints– Lungs, aorta, spinal cord, brachial plexus, esophagus, heart, rib cage– Tolerances are still evolving
• Generally, 5-11 fields (spacing ~ 20°, non-opposed)– Typically more beams are used with complex shaped targets
• 2.0 - 2.5 mm dose grid is adequate
• Use heterogeneity corrections and convolution/superposition algorithm
T10 (18 Gy x 1): IMRT Plan
T10 (18 Gy x 1): Dose Distribution
VMAT SRS18Gy x 1
6F
Multiple PTVs
1
CBCTGate100
1 1 1
1
3
2
Gate100 Setup Strategy
Step 1: Acquire kV/kV – register to boneStep 2: Acquire CBCT – register to superior border
Geometric center
Gate100 Match
Gate3070 Setup Strategy
1
CBCTGate3070
1 1 1
End-of-Expiration = Target at Superior Border
1
3
23
2
Step 1: Acquire kV/kV – register to boneStep 2: Acquire CBCT – register to superior border
Matched at the superior edge
CBCTReference CT
Gate3070 Match
Patient Number2019181716151413121110987654321
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0
vector by Patient Number
GYN – bony anatomy
Prostate – Fiducials
Image Interpretation
Efficient Treatment is Important
• Mean intra-fractional tumor position– Measured as a function of the interval between localization and repeated CBCTs
• 5.3 mm if the time > 34 min• 2.2 mm if the time < 34 min
Quality Assurance Overview
• Acceptance
• Commissioning
• Clinical implementation
Commissioning & Implementation• SBRT equipment
– Ensure geometric precision/accuracy for immobilization devices
• Check the inter-operation of all equipment– It should not be assumed one component that works in one clinical
situation will work for SBRT treatments
• Treatment planning system (TPS)– TPS accuracy for small fields– End-to-end testing
• Establish independent checks– e.g. RPC dosimetry– Audits for process checks
Thoughts on Commissioning
• Allow enough time for commissioning – do not rush to treat a patient
• Regular QA should verify equipment/process changes
• Ensure overall accuracy using end-to-end tests for each new treatment site
• Phone a friend, e.g., on-line collaborations
• Mechanical and radiation isocenter verification– E.g., Winston-Lutz test
• MV and kV isocentercoincidence verification
• AAPM TG-142, 2009– for other routine linac tests
On-going QA
TG-142 Linac QA Example
• Daily, Monthly, Annual• EDWs, MLCs, Imaging systems
Other Relevant AAPM Reports
• TG-42 (1995)– Stereotactic cranial radiosurgery
• TG-66 (2003)– CT and the CT simulation process
• TG-135 (2011)– Robotic radiosurgery
Patient-Specific Physics QA
• Planning requirements are different– Is the treatment plan acceptable?
• Appropriate imaging sequence
• Patient-specific measurements
• Directly supervise each treatment fraction– Appropriate setup– Appropriate beam-delivery parameters– Motion management input
Patient-Specific Physics QA cont.
• Independent review of setup and treatment parameters should be completed
• Validate initial setup instructions and check plan against prescription
• Review the plan with your physician– Contours Ok– Dose distribution makes sense
Summary and Final Caveats
• Correct images from simulation to planning• Contour on correct images• Multiple fields • Correct procedures for treatment setup• Don’t miss the big picture
– TPS commissioning, Linac output, end-to-end tests
• Use quality management program– Practice patterns and technology evolve