i.'.' la'-ur- 96-3451 - unt digital library/67531/metadc... · radiation protection...
TRANSCRIPT
LA'-UR-
I . ' . '
96-3451 Title:
Author(s) :
Submitted tor
THE USE OF MICROIXSIMETRIC TECHNIQUES I N RADIATION PROTECTION
Jing Chen, ESH-4
W.H. Casson, ESH-4 D.G. Vas i l ik , ESH-4
H.H. HSU, ESH-4
QCT 3 0 1996 O S T I
Health Physics of Radiation Generating Machines 30th Midyear Topical Meeting of t h e Health Physics Society, San Jose, CA, January 5-8, 1997
Los Alamos National Laboratory, an affirmative actionlequal opportunity empldyer, is operated by the University of California for the US. Department of Energy under contract W-7405-ENG-36. By acceptance of this article, the publisher recognizes that the US. Government retains a nonexclusive, royalty-free license to publish or rqrOduCe the published form of this contribution, or to allow others to do so, for US. Government purposes. The LoS Alamos National Laboratory requests that the publisher identify this article as work performed under the auspices of the US. Department of Energy. Form No. 836 R5
Si2629 1W91
DISUAXMER
Portions of this document may be illegible in electronic image products. Images are produced from the best available original document.
DISCLAIMER
This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or use- fulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any spe- cific commercial product, process, or service by trade name, trademark, manufac- turer, or otherwise does not necessarily constitute or imply its endorsement, recom- mendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof.
The Use of Micmdosimetric Techniques in Radiation Protection
Measurvmen ts
Jing Chen, Hsiao-Hua Hsu, William H. Casson, and Dennis G. Vasilik
MS G761
Los Alamos National Laboratory
Los Alamos, NM 87545
Abstract
A major objective of radiation protection is to determine the dose equivalent for routine
radiation protection applications. As microdosimetry has developed over approximately
three decades, its most important application has been in measuring radiation quality,
especially in radiation fields of unknown or inadequately known energy spectra. In these
radiation fields, determination of dose equivalent is not straightforward; however, the use
of microdosimetric principles and techniques could solve this problem. In this paper, we
discuss the measurement of lineal energy, a microscopic analog to linear energy transfer,
and demonstrate the development and implementation of the variance-covariance method,
a novel method in experimental microdosimetry. This method permits the determination
of dose mean lineal energy, an essential parameter of radiation quality, in a radiation field
of unknown spectrum, time-varying dose rate, and high dose rate. Real-time monitoring of
changes in radiation quality can also be achieved by using microdosimetric techniques.
Introduction
A major objective of radiation protection is to determine the dose equivalent for routine
radiation protection applications. Dose equivalent, H, is defined as a product of the
absorbed dose, D, and the radiation quality factor, Q, where Q takes into account the
biological effectiveness of different ionizing radiations. The quality factor is currently
defined as a function of linear energy transfer (LET), which, in a strict sense, is a
computational, not measurable quantity.
Microdosimetry is based on experimentally measurable quantities in small irradiated
volumes. Lineal energy, y, is a microscopic analog to LET. The measurement of lineal
energy can provide, in many cases, good approximations to determine quality factors and
thus the dose equivalent, as described in ICRU 40 (ICRU 1986).
It is well known that microdosimetric techniques are pertinent for measurements with low
doses and low dose rates. In this paper, we will present microdosimetric techniques that
can be equally well used in radiation fields of high dose rate.
In radiation fields of high dose rate, the determination of dose equivalent becomes difficult
because of the pile-up effect. This is especially the case in high-LET radiation fields in
which the quality factor is significantly larger than unity and also in radiation fields of
unknown or inadequately known energy spectra in which a conservative mean quality
factor should normally be used. Although modern digital data processing techniques can
help to some extent to correct for the pile-up effect numerically, many instruments are
2
t
still limited to a given dose rate. The use of the variance-covariance method, a novel
microdosimetric technique, could help us to come out of this dilemma. To enable better
understanding of the variance-covariance method, we first discuss the variance method.
The Variance Method
Dose mean lineal energy, z, is defined as the dose-weighted average of the single-event
spectrum, f(y) (ICRU 1983). There is, however, a direct method that one can use to
determine dose mean lineal energy without having to know the single-event spectrum and
without the restriction of very low dose rates (as in the case of a single-event spectrum
measurement). This is the variance method devised by Bengtsson (Bengtsson 1970) and
successfully employed under laboratory conditions by Bengtsson and Lindborg
(Bengtsson and Lindborg 1974). Mathematically the variance method can be expressed as
where r is the frequency mean value of the multi-event spectrum of energy imparted E,
and V, is the relative variance of the multi-event spectrum.
In a constant radiation field, the measured signal variance results from the microdosirnetric
fluctuations of energy imparted in the detector. In a series of measurements of dose
- increments, the original signals are E ~ , c2, ..., I, and they are the basis for cal~ulations of
and V,. Correspondingly, the radiation dose and dose mean lineal energy are
3
*
where m the mass and 7 the mean chord length of the detector volume.
The variance method was subject to the requirement of a constant radiation field, Le., the
dose increments per measurement interval had to remain exactly the same throughout a
series of measurements. Often, this condition cannot be met. The further development of
this method to meet more practical situations has resulted in the variance-covariance
method.
The Variance-Covariance Method
The limitation of the variance method was removed by the variance-covariance method
(Kellerer and Rossi 1984). Its mathematical form is
(3) - - E , = (v, - C r A B ) . E A *
The method uses two detectors (detector A and detector B) that determine energy
imparted in a series of synchronous measurement intervals @e., the two measuring
channels are triggered exactly at the same time). The basic measured signals are &Al, em, ...,
eAn for detector A, and cBl, eB2, ..., eBn, for detector B. In this method, the two detectors do
not need to be identical, but the ratio of doses in the two detectors should be constant.
The microdosimetric variance of energy imparted is obtained as the difference of two
terms. The first term, V,, is the relative variance of measured signals from one detector
(for instance, detector A), which includes microdosimetric and dose-rate fluctuations, Le.,
4
the total fluctuation of detector signals. The second term, C,, is the relative covariance
between synchronous signals of the two detectors. It merely represents the dose-rate
fluctuation that effects the two detectors correlatively, while the microdosimetric
fluctuations in the two detectors are statistically independent. The difference of V,-C,
gives the microdosimetric fluctuation only. The microdosimetric fluctuation of energy
imparted in the detectors is radiation specific and reflects radiation quality. In a constant
radiation field, the observed covariance should be zero, and the variance-covariance
method reduces to the variance method.
The variance-covariance method has been successfully used in numerous micrudosimetric
investigations such as in clinical radiation fields (Honore et al. 1990; Chen et al. 1992;
Chen et al. 1994) and in space activities (Lindborg et al. 1995). Recently, the variance-
covariance method has been extended to the situation in which there are slow changes to
the ratio of dose rates in the two detectors (Kellerer 1996). This extension is important
for personal rem meters. In personal rem meters, the ratio of dose rate can change when
the person who carries the detectors moves in a radiation field.
The following is the new form of the variance-covariance method applied to variable dose
rates compounded by changes of the ratio of dose rates in the two detectors,
5
I
’ -
It differs from equation 2 by a factor, k, and the ratio of mean energy imparted in the two
-- detectors in the covariance term. This factor, k - E , / E A , reflects the changes of the ratio
of dose rates in the two detectors. We can calculate k from successive measured signals:
* ‘ A i E A ( i + l ) k = (5)
The signal processing as given in equations 1 to 5 is easy to perform with digital signal
processing devices.
Conclusion
We have discussed the advantages of using microdosimetric parameters in operational
radiation protection and have presented well-established microdosimetric techniques. We
also have presented a new development that can help determine dose mean lineal energy
in various practical- radiation measurements. Although the exact relationship between the
definition of radiation quality factor in terms of LET and in terms of y is still under
discussion, we find that microdosimetric techniques should be employed, at least in
situations when conventional methods used in radiation protection become difficult or
even impossible. Two examples follow.
Example 1: Radiation field characteristics are neither known nor well known.
Because many conventional detection methods are developed for radiation fields of given
energy spectra, choosing a suitable rem meter in unknown radiation fields for the purpose
of radiation protection is difficult. The variance-covariance method is developed to
6
b - -
determine the radiation dose and radiation quality, i.e., dose equivalent without
preknowledge of the radiation fields. The method is especially suitable for mixed radiation
fields, such as in the case of neutron dosimetry and dosimetry related to space activities.
Example 2: Radiation fields of high-dose rates. These fields occur when conventional
techniques fail due to the high pile-up effect and when the pile-up problem cannot be
solved even with an available flash analog digital converter and modern digital data-
processing techniques.
The variance-covariance method has been developed as a method suitable for a high dose
rate and for time-varying dose rate as well. Future activities in operational radiation will
definitely benefit from the use of microdosimetric techniques.
References :
Bengtsson, L. G. Assessment of dose equivalent from fluctuations of energy deposition.
In: Proceedings of the Second Symposium on Microdosimetry. EUR 4452 HG;
Ebert ed.; Euroatom, Brussels; 1970: 375-395
Bengtsson, L. G.; and Lindborg, L. Comparison of pulse height analysis and variance
measurement of the determination of dose mean specific energy. In: Proceedings
of the Fourth Symposium on Microdosimetry. EUR 5 122; J Booz ed.; Euroatom,
Luxembourg, 1974: 823-84 1
Chen, J.; Roos, H.; and Kellerer, A. M. Microdosimetry of diagnostic x-rays: applications
of the variance-covariance method. Radiat. Res. 132:271-276; 1992.
7
I c
t
Chen, J.; Hahn, K.; Roos, H.; and Kellerer, A. M. Microdosimetry of therapy electron
beams-measurements and Monte-Carlo simulations. Radiat. Prot. Dosim. 52:
435-438; 1994
Honore, H. B.; Jessen, K. A.; and Nielson, H. H. Variance-covariance measurement of the
dose mean lineal energy in beams for radiotherapy. Radiat. Prot. Dosim. 3 1 : 453-
455; 1990.
ICRU. The quality factor in radiation protection. Report 40; International Commission on
Radiation Units and Measurements; Bethesda, Maryland 1986.
ICRU. Microdosimetry. Report 36; International Commission on Radiation Units and
Measurements; Bethesda, Maryland 1983.
Kellerer, A. M.; and Rossi, H. H. On the determination of microdosimetric parameters in
time-varying radiation fields: the variance-covariance method. Radiat. Res. 97:
237-245; 1984.
Kellerer, A. M. Generalization of the variance-covariance method for microdosimetric
measurements 11. Formulae for varying dose-rate ratio in the detectors and
synopsis of results. Radiat. Environ. Biophys. 35: 1 17-1 19; 1996.
Lindborg, L.; Grindborg, J. E.; Gullberg, 0.; Nilsson, U.; Samuelson, G.; and Uotila, P.
TEPC measurements with the variance-covariance method on board aircraft.
Radiat. Prot. Dosim. 61:119-124; 1995.
8