icru 58

40
ANJU.K.V. MSC RADIATION PHYSICS CALICUT UNIVERSITY ICRU 58 Dose and Volume Specification For Reporting Interstitial Therapy

Upload: anju-kv

Post on 12-Apr-2017

1.104 views

Category:

Education


3 download

TRANSCRIPT

Page 1: Icru 58

ANJU.K.V.MSC RADIATION PHYSICS CALICUT UNIVERSITY

ICRU 58Dose and VolumeSpecification ForReportingInterstitial Therapy

Page 2: Icru 58

INTRODUCTIONWith the introduction of miniaturized brachytherapy sources, the sophisticated 3-D source localization methods have been introduced, which can be linked to computerized methods for dose calculation. Therefore, a common language is valuable to provide a method of dose specification and reporting which can be used for implants of all types of interstitial brachytherapy procedures.

Page 3: Icru 58

1985: ICRU Report 38, Was written regarding the Dose and volume specification as well as reporting intracavitary Brachytherapy in GYN. 1997: ICRU Report 58, Was written to generate a guideline about the dose and volumes specification for reporting interstitial Brachytherapy.

Page 4: Icru 58

ICRU 58Aim: to develop a common language based on existing Concepts.1. usable to describe what has been done.2. closely related to the outcome of the treatment.3. generally understood.

Consistency with external beam radiotherapy (EBT)definition of volumes – the same.ICRU reference point for dosimetry in EBT – cannot betaken over directly for interstitial BT.

Page 5: Icru 58

ICRU 58There are several historic systems of prescribing brachytherapy• –Manchester system• –Paris system• Quimby system

Page 6: Icru 58

• The term “system” denotes a set of rules which takes into account the

source types , strengths, geometry and method of application to obtain suitable dose distributions over the volume(s) to be treated. The system also provides a means of calculating and specifying dose. • It is important to remember that while an implant may follow the source distribution rules of a system, it does not comply with the system unless the method of dose specification and prescription are also followed.

ICRU 58

Page 7: Icru 58

ICRUTwo important type of implants in ICRU 58 are,1. Temporary implant2. Permanent implant

Page 8: Icru 58

Temporary Implant• The radioactive sources are removed from the tissue after the treatment is

completed.• Radionuclide used have typically longer half life.

Permanent Implant• The brachytherapy sources remain in the patient, and they will not be

removed. • Radionuclide used have typically shorter half life.

ICRU 58

Page 9: Icru 58

ICRU 58

Traditionally, the temporary implants were performed with the linear (like wires) radioactive sources or seeds arranged in a linear fashion (i.e. ribbons, etc.).However, in a permanent implants, multiple loose sources were distributed in random orientations.

Page 10: Icru 58

With the design of the low energy radioactive sources in the form of stranded seeds or radioactive wires, now the permanent implant can also be performed with linear or prearranged ribbons.The loose seed implants areconsidered separately in this report.

ICRU 58

Page 11: Icru 58

For Planning permanent implantsThe number of sources depends on,a. Their initial strength.b. Type of the radioisotope.

For Planning temporary implantsThe total time of implantation depends on,a. Number of sourcesb. Strengths and sourcec. Pattern of distribution of sources

Page 12: Icru 58

Planning boost permanent and temporary implants,the number of seeds and the initial activity of the seeds are adjusted upon the external beam radiation dose.

Page 13: Icru 58

KERMA (Kinetic Energy Released in Matter)

The sum of the initial kinetic energie of all the ionizing charged particles

liberated by uncharged ionizing particles.

K = dEtr/ dm

ABSORBED DOSEThe mean energy imparted by

radiation to a unitmass of medium

D=dE/dm

1Gy = 1J/Kg

Page 14: Icru 58

Reference Air Kerma rateReference Air Kerma Rate = Air Kerma Rate at a reference point of 1 cm or 1 m.

ORIt is defined as the air kerma rate measured at one meter along the perpendicular bisector of the long axis of the source,corrected for air attenuation and scattering.Unit : 1 U = μGy/hThe term m^2 is not included inthe unit,since it is inherent in the definition itself.

Page 15: Icru 58

Total Reference Air Kerma (TRAK)TRAK= Σti . Si

Where;Si = Reference air kerma rate foreach source.ti = Irradiation time for each source.

Page 16: Icru 58

Description of Source Patternssingle plane: implant is defined as an implant containing two or more sources, which lie in the same plane.

two-plane (double-plane): implant contains two planes, which are generally parallel to each other.

Multi-plane: implant, contains three or more planes.

non-plane: sphere or cylinder.

Page 17: Icru 58

VOLUMES(Gross Tumor volume)

(Clinical Target Volume)(Planning

Target Volume)

(Treatment Volume)(Irradiated Volume)

Page 18: Icru 58

Gross Tumor Volume (GTV)The palpable or visible/demonstrable extentand location of the malignant growth.

Clinical Target Volume (CTV)The volume of the tissue that contains agross tumor volume and/or subclinicalmicroscopic malignant disease which has tobe eliminated.

Page 19: Icru 58

Planning Target Volume (PTV)The volume of tissue receiving theprescribed irradiation.For an interstitial brachytherapy, the PTV is,in general, identical to the CTV.

Treatment Volume (TV)The volume of tissue, which is encompassedby an isodose surface that has beenspecified by the radiation oncologist.The dose value at this isodose surface is theminimum target dose

Page 20: Icru 58

CTVThe maximum diameters measured in three orthogonal directions must be reported.

Page 21: Icru 58

CENTRAL PLANEThe plane that is perpendicular to the maindirection of the linear sources and passingthrough the estimated center of the implant.

Page 22: Icru 58

Central Plane (continue)

The plane that is perpendicular to the maindirection of the linear sources and passingthrough the estimated center of the implant.

Page 23: Icru 58

Description of Dose distributionInterstitial Brachytherapy Implants:1. Non Homogeneous dose Distribution .2. High Regions around Sources.3. Regions of plateau dose or local minimum doses.

The dose distribution in a plane perpendicular to linear and parallel sources, shows a plateau dose region of low dose gradient. In this example of three sources of 6 cm long and with 1.5 cm spacing, the dose varies by less than 2% between the sources.

Page 24: Icru 58

Dose Distribution in One or More Planes through the Implant1. The minimum information needed for the dose distribution of an implant is,The isodose curves in at least one plane either in tabular form or by graphical Presentation.2. The central plane of the implant should be used, If only one plane is chose.

Page 25: Icru 58

Prescription DoseThe prescribed dose is defined as the dose, which the physician intends to give, and enters in the patient’s treatment chart.Minimum Target Dose (MTD): The minimum dose at the periphery of the CTV = Minimum dose decided upon by the clinician as adequate to treat the CTV (minimum peripheral dose).MTD≅ 90% of the prescribed dose in the Manchsteer system for interstitial therapy.

Page 26: Icru 58

Prescription Dose (continue)Mean Central Dose (MCD) : DmThe minimum dose at the periphery of the CTV= Arithmetic mean of the local minimum doses between sources, in the central planes.

Page 27: Icru 58

Prescription Dose (continue)High Dose VolumeThe volume of tissue that is encompassed that will receive more than 150% of the mean central dose be reported.Low Dose VolumeA low dose volume should be defined as a volume within the clinical target volume, encompassed by an isodose corresponding to 90% of the prescribed dose. The maximum dimension of the low dose volume in any plane calculated should be reported.

Page 28: Icru 58

Prescription Dose (continue)Dose Uniformity ParametersICRU defines the following two methods for the dose uniformity

1. The spread of the individual minimum doses used to calculate the mean

central dose in the central plane (expressed as a percentage of themean central dose).2. The dose homogeneity index; defined as the ratio of minimum target dose to the mean central dose.

Page 29: Icru 58

Time Dose FactorsTime and Dose Rates for Temporary ImplantsTime = Prescribed dose/initial dose rate Time Dose Pattern for Temporary Implantsa. Continuous irradiationb. Non-continuous irradiationc. Pulsed irradiation

Page 30: Icru 58

Recommendations for Recording and ReportingParameters Required for Recording and Reporting. Description of Volumes : GTV, PTV,…….. Description of Sources :1. Radioisotope and filtration2. Pattern of source distribution• Description of Technique and Source pattern

Page 31: Icru 58

Table 1: Levels of priority for reportingtemporary interstitial implants

Page 32: Icru 58

Continue

Page 33: Icru 58

a. Priority1 . Concerned with doses in the central plan2 . Required calculations outside the central plan. If this is not available, then a more detailed description of source pattern under priority 1 is required.3 . Additional information mostly of clinical research interestb. Level of computation1. No computer needed2. Hand calculation and/or computer calculation in central plane.3. 3-D computation neededc. Essential to establish consistent reporting and related topast experience, necessary for comparison of brachytherapy data and for relating outcome to treatment.

Page 34: Icru 58

Table 3: Levels of priority for reporting implants with a single stationary source line.

Page 35: Icru 58

Table 4: Levels of priority for reporting implants with a moving source.

Page 36: Icru 58

Table 5: Levels of priority for reporting implants with a surface applicators

Page 37: Icru 58

CONCLUSION

It is very important to follow these recommendations to achieve the aim of brachytherapy.

In this presentation, the ICRU-58 recommended concepts and procedures for dose specifications and reporting of the interstitial brachytherapy implants has reviewed.

Page 38: Icru 58

Selected reference1. ICRU Report 38: Dose and Volume Specification for Reporting

Intracavitary Gynecology (1985)2. ICRU Report 58: Dose and Volume Specification for Reporting

Interstitial Therapy (1997)3. The GEC ESTRO Handbook of Brachytherapy, Editors: Alain

Gerbaulet, Richard Pötter, Jean-Jacques Mazeron, Erik van Limbergen (2002;107;3225-332

4. ESMP 2013 Reporting in Brachytherapy,Lecture 5 Brachytherapy Implantation.s

Page 39: Icru 58
Page 40: Icru 58