IES LM-79-08
MEASUREMENT AND TEST REPORT For
ATG Electronics Corp. 9020 Rancho Park Court Rancho Cucamonga, CA 91730
Test Model: AL25
Report Type:
Electrical and Photometric tests including: Luminous Flux,
Color, Luminous Intensity Distribution, THD, Power
Factor
Test Engineer: Daniel Duan
Report Number: RSZ160304501-10
Test Date: 2016-03-10
Report Date: 2016-03-11
Reviewed By: Jeanne Han/Safety Manager
Prepared By:
Bay Area Compliance Laboratories Corp. (Shenzhen)
6/F, the 3rd Phase of WanLi Industrial Building,
ShiHua Road, FuTian Free Trade Zone
Shenzhen, Guangdong, China
Tel: +86-755-33320018
Fax: +86-755-33320008
Test Facility: Test facility was located at Pu Long Cun 69, Puxinghu Industrial
Area, Tangxia Town, Dongguan, Guangdong, P.R.China.
Accreditation: The NVLAP Lab Code is 200707-0.
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 2 of 13 The NVLAP Lab Code is 200707-0.
1. Product Description
General Information:
One sample was received on 2016-03-04 and used for testing. Sample No.: RSZ160304501-S01 Model: AL25 Model Tested: AL25
Manufacturer: ATG Electronics Co.,Ltd
Brand Name: ATG
Product Designation: Outdoor Pole/Arm-mounted Area and Roadway Luminaires
Dimmable: Continuous Dimming
Dimming Range: 10% to 100%
Burning Time Before Test: 0 hour(For New Products)
Rated Values:
Rated Voltage/Frequency: 120-277V AC 50/60Hz
Rated Power: 255W
Nominal Light Output: 25500 lm
Nominal CCT: 5000K
Nominal CRI: 70
2. Standards Used
IESNA LM-79-08: Approved Method: Electrical & Photometric Measurement of Solid-state Lighting Products
ANSI C82.77-2002: Harmonic Emission Limits – Related Power Quality Requirements for Lighting
3. Description of Test Equipment
Device Manufacture Model No Serial No Test Range Calibration
date
Calibration
due date
Integrating
Sphere SENSING SPR-600 S09008 25~50℃ 2015-03-25 2016-03-24
Spectral
photometer SENSING SPR3000 90902027 350nm~800nm 2015-03-25 2016-03-24
Power Meter YOKOGAWA WT-210 91j926132 15/30/60/150/300/600 V 2016-03-04 2017-03-03
AC Power Supply ALL Power APW-
105N 970663 220V±10% 50HZ 2016-03-04 2017-03-03
Standard Light
Source EVERFINE D204 01331191 24V/100W 2015-08-27 2016-08-26
Thermal Meter SENSING N/A N/A 25、50℃ 2016-03-04 2017-03-03
DC Power
Supply ITECH IT6154
0061 0417 6471 0010
19 0~32V 2016-03-04 2017-03-03
AC Power Supply EVERFINE VPS1030
PWM 1012017 0-150V, 0-300V 2016-03-04 2017-03-03
DC Power
Supply EVERFINE WY12010 1009009 30V/5A 2016-03-04 2017-03-03
Power Meter YOKOGAWA WT-210 91KB35700 15/30/60/150/300/600 V 2016-03-04 2017-03-03
Goniophotometer EVERFINE GO-R5000 YG108492N10120001 1600mm,3000W/10A 2015-03-20 2016-03-19
Wireless Remote
Sensor N/A 433MHz N/A 0℃~50℃;-20℃~60℃ 2015-03-24 2016-03-23
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 3 of 13 The NVLAP Lab Code is 200707-0.
Device Manufacture Model No Serial No Test Range Calibration
date
Calibration
due date
Standard Light
Source EVERFINE D908 1012003 N/A 2015-09-08 2016-09-07
Statement of Traceability: Bay Area Compliance Laboratories Corp. (Shenzhen) attested that all calibration has been
performed using suitable standards traceable to National Primary Standards and International System of Units (SI).
4. Test Method
Product was tested with no seasoning. All stabilization and measurements were made in compliance with IES LM-79-08. The product was operated at rated voltage or at voltage required by manufacturer. The ambient temperature of the sample was maintained at 25°C±1°C during measurement. And relative humidity is less than 65%.
Integrating Sphere System
The system includes AC power source, digital power meter, DC power supply, spectrophotometer, and integrating sphere. The integrating sphere system is calibrated by standard light source before measurement. The system and standard light source has been calibrated regularly and traceable to the National Primary Standards.
4π geometry was used during measurement. The product was operated in its intended orientation in application and was recorded in this report.
The uncertainty of the light output (luminous flux) measurements is U=2.1% (K=2), at the 95% confidence level. The uncertainty of the correlated color temperature measurements is U=32K (K=2), at the 95% confidence level. The uncertainty of the CRI is U=2.1 (K=2) , at the 95% confidence level.
The uncertainty of power meter AC current U=0.19 % of rdg, AC Voltage U=0.15% of rdg, Power U=0.20%) (K=2), at the 95% confidence level.
Goniophotometer System
The goniophotometer system is calibrated by standard light source before measurement. The standard light source has been calibrated regularly and traceable to the National Primary Standards.
Type C goniophotometer was used for measuring total luminous flux, luminous intensity distribution, and color spatial uniformity. The product was operated in its intended orientation in application and was recorded in this report.
The uncertainty of the luminous intensity is U=2.82% (K=2) , at the 95% confidence level.
Additional Test
The Additional Test item may not be covered by IESNA LM-79-2008. Additional test including power factor, off-state power and THD, was measured by Digital Power Meter after stabilized at 25°C±1°C.Test voltage for THD and power factor test would be equal to rated voltage or, in case of a voltage range, maximum value of that range.
The uncertainty of power meter AC current U=0.19 % of rdg, AC Voltage U=0.15% of rdg, Power U=0.20%) (K=2), at
the 95% confidence level.
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 4 of 13 The NVLAP Lab Code is 200707-0.
5. Test Result
[Integrating Sphere System]
Total operating time for integrating sphere test: 1.0 hour
Test orientation: Downward
Electrical Measurement
Voltage (V) Frequency (Hz) Current (A) Power (W) Power Factor
120.09 60 2.1598 258.59 0.997
Photometric Measurement
Luminous Flux (lm) Radiant Flux (W) Efficacy (lm/W) CCT (K) Duv
25695.56 77.778 99.368 4939 0.00268
Chromaticity Coordinate
x y u v u' v'
0.3474 0.3589 0.2102 0.3257 0.2102 0.4885
Color Rendering Index
Ra
70.3
R1 R2 R3 R4
69 74 76 72
R5 R6 R7 R8
69 64 79 60
R9 R10 R11 R12
-22 37 69 39
R13 R14 R15
69 86 65
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 5 of 13 The NVLAP Lab Code is 200707-0.
Relative Spectral Power Distribution
nm mW
nm mW
nm mW
nm mW
nm mW
380 3.305E-02 465 4.551E-01 550 1.487E+00 635 8.098E-01 720 1.065E-01
385 2.189E-02 470 3.054E-01 555 1.493E+00 640 7.408E-01 725 9.329E-02
390 1.763E-02 475 2.276E-01 560 1.500E+00 645 6.701E-01 730 8.233E-02
395 1.433E-02 480 2.018E-01 565 1.485E+00 650 6.078E-01 735 7.214E-02
400 1.414E-02 485 2.092E-01 570 1.484E+00 655 5.455E-01 740 6.368E-02
405 2.617E-02 490 2.570E-01 575 1.461E+00 660 4.847E-01 745 5.584E-02
410 6.157E-02 495 3.486E-01 580 1.445E+00 665 4.315E-01 750 5.077E-02
415 1.481E-01 500 4.840E-01 585 1.412E+00 670 3.847E-01 755 4.461E-02
420 3.114E-01 505 6.476E-01 590 1.378E+00 675 3.401E-01 760 3.999E-02
425 5.741E-01 510 8.167E-01 595 1.341E+00 680 2.975E-01 765 3.551E-02
430 9.587E-01 515 9.821E-01 600 1.285E+00 685 2.633E-01 770 3.211E-02
435 1.458E+00 520 1.125E+00 605 1.229E+00 690 2.336E-01 775 2.847E-02
440 2.047E+00 525 1.246E+00 610 1.173E+00 695 2.045E-01 780 2.622E-02
445 2.355E+00 530 1.332E+00 615 1.099E+00 700 1.812E-01
450 1.862E+00 535 1.391E+00 620 1.031E+00 705 1.571E-01
455 1.109E+00 540 1.437E+00 625 9.601E-01 710 1.364E-01
460 6.950E-01 545 1.462E+00 630 8.823E-01 715 1.198E-01
0.0
0.2
0.4
0.6
0.8
1.0
1.2
380 440 500 560 620 680 740 800
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 6 of 13 The NVLAP Lab Code is 200707-0.
CIE 1931 x y Chromaticity Diagram
7-Step & 4-Step Chromaticity Quadrangles
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8
0.30
0.35
0.40
0.45
0.30 0.35 0.40 0.45 0.50 0.55
sample
7-STEP 2200K
7-STEP 2500K
7-STEP 2700K
7-STEP 3000K
7-STEP 3500K
7-STEP 4000K
7-STEP 4500K
7-STEP 5000K
7-STEP 5700K
7-STEP 6500K
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 7 of 13 The NVLAP Lab Code is 200707-0.
[Goniophotometer System]
Total operating time for luminous intensity distribution: 1.0 hour
Test orientation: Downward
Electrical Measurement
Input Voltage (V) Frequency (Hz) Input Current (A) Power (W) Power Factor
120.09 60 2.1584 258.5 0.9973
Photometric Measurement
Luminous Flux (lm) Efficacy (lm/W) Imax (cd) S/MH (C0/180) S/MH (C90/270)
25721.1 99.50 16635 2.45 1.63
Luminous Intensity Distribution
C0/180 C45/225 C90/270 C135/315 AVG.
Beam Angle (50% Imax): 154.0 27.6 22.6 24.4 57.2
Field Angle (10% Imax): 170.3 143.7 149.4 145.4 152.2
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 8 of 13 The NVLAP Lab Code is 200707-0.
Luminous Intensity (cd) Distribution Data
C
0° 22.5° 45° 67.5° 90° 112.5° 135° 157.5° γ
0.0° 3222 3222 3222 3222 3222 3222 3222 3222
5.0° 3270 3303 3324 3331 3324 3309 3287 3255
10.0° 3364 3423 3458 3461 3443 3418 3381 3325
15.0° 3531 3606 3628 3615 3570 3549 3513 3452
20.0° 3799 3907 3853 3787 3706 3702 3693 3643
25.0° 4334 4433 4174 3990 3862 3888 3924 3950
30.0° 5013 5544 4731 4214 4011 4072 4277 4681
35.0° 5581 6683 6088 4844 4481 4560 5288 5349
40.0° 6420 8519 7207 5806 4809 5276 5941 6751
45.0° 7077 10737 9383 5997 5039 5412 7615 8978
50.0° 7328 12069 12225 7110 5990 6447 10451 11146
55.0° 7137 11367 13991 8960 7539 8175 12916 11536
60.0° 6771 9512 14167 11425 9816 10509 13281 9553
65.0° 6417 7826 16226 12660 8926 12419 14595 7548
70.0° 4986 5020 10210 7349 3653 7389 10576 5110
75.0° 2327 1626 2032 2521 1414 2555 2100 1845
80.0° 787 1040 1175 1030 630 1036 1221 1358
85.0° 301 544 477 336 290 323 484 651
90.0° 3 52 47 46 2 33 29 2
95.0° 3 2 1 1 1 1 2 2
100.0° 4 3 2 1 1 1 2 3
105.0° 4 3 2 2 1 2 2 3
110.0° 4 3 2 2 2 2 2 3
115.0° 4 3 2 2 2 2 2 3
120.0° 4 3 2 2 2 2 2 3
125.0° 4 3 2 2 2 2 2 3
130.0° 4 3 2 2 2 2 3 3
135.0° 4 3 2 2 2 2 3 4
140.0° 4 3 3 2 2 3 4 4
145.0° 4 3 3 3 3 4 4 5
150.0° 4 4 4 4 4 4 5 5
155.0° 4 4 4 4 4 5 5 6
160.0° 4 4 4 4 4 5 6 6
165.0° 4 4 4 4 4 5 5 5
170.0° 4 4 4 4 4 5 5 4
175.0° 4 4 4 4 4 4 5 5
180.0° 4 3 3 3 4 4 4 4
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 9 of 13 The NVLAP Lab Code is 200707-0.
Luminous Intensity (cd) Distribution Data (cont.)
C
180° 202.5° 225° 247.5° 270° 292.5° 315° 337.5° γ
0.0° 3222 3222 3222 3222 3222 3222 3222 3222
5.0° 3215 3176 3146 3131 3133 3150 3182 3225
10.0° 3248 3164 3097 3064 3069 3105 3177 3270
15.0° 3338 3189 3074 3023 3033 3076 3186 3356
20.0° 3477 3242 3083 3026 3042 3088 3219 3492
25.0° 3751 3340 3133 3081 3104 3157 3313 3747
30.0° 4123 3467 3229 3197 3224 3288 3435 3938
35.0° 4499 3676 3388 3351 3377 3460 3648 4255
40.0° 5170 3968 3597 3455 3456 3589 3913 4651
45.0° 5890 4311 3767 3425 3361 3573 4118 5001
50.0° 6428 4615 3737 3170 3074 3300 4066 5261
55.0° 6597 4736 3366 2719 2598 2810 3639 5331
60.0° 6419 4458 2725 2104 1670 2186 2943 5019
65.0° 6246 3713 1967 1349 1336 1412 2122 4352
70.0° 5608 2550 1285 943 1084 957 1374 3136
75.0° 2824 1436 999 844 861 831 1006 1643
80.0° 978 867 622 642 574 658 667 913
85.0° 390 363 263 230 270 228 310 398
90.0° 29 53 3 2 2 2 49 32
95.0° 2 2 3 3 3 3 3 3
100.0° 2 3 4 3 3 4 3 3
105.0° 2 4 4 4 4 4 4 4
110.0° 3 4 4 4 4 4 4 4
115.0° 3 4 4 4 4 4 4 4
120.0° 3 4 4 5 5 5 4 4
125.0° 4 5 5 5 5 5 5 5
130.0° 4 5 6 6 5 5 5 5
135.0° 4 5 6 6 5 5 6 5
140.0° 4 5 6 6 5 5 5 5
145.0° 4 4 4 5 4 4 4 4
150.0° 3 3 4 4 4 4 4 4
155.0° 3 3 3 3 3 4 3 4
160.0° 3 3 3 3 3 3 3 4
165.0° 3 3 3 3 3 3 3 4
170.0° 3 3 3 3 3 3 3 4
175.0° 3 3 3 3 3 3 4 4
180.0° 4 4 4 4 3 4 4 4
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 10 of 13 The NVLAP Lab Code is 200707-0.
Average Area Illumination Figure
Height (m) Diameter
(cm) Eavg(lx) Emax(lx)
0.5 54.52 11527.0 14207.0
1.0 109.04 2882.0 3552.0
1.5 163.57 1281.0 1579.0
2.0 218.09 720.4 887.9
2.5 272.61 461.1 568.3
3.0 327.13 320.2 394.6
3.5 381.65 235.2 289.9
4.0 436.17 180.1 222.0
4.5 490.70 142.3 175.4
5.0 545.22 115.3 142.1
Angle: 57.2°. Flux out: 2782.0 lm.
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 11 of 13 The NVLAP Lab Code is 200707-0.
Zonal Lumen Density Measurement
Deg Flux (lm) %
Deg Flux (lm) %
0-5 77.2 0.30 0-5 77.2 0.30
5-10 233.0 0.91 0-10 310.2 1.21
10-15 393.7 1.53 0-15 703.9 2.74
15-20 563.7 2.19 0-20 1267.6 4.93
20-25 752.2 2.92 0-25 2019.8 7.85
25-30 977.5 3.80 0-30 2997.3 11.65
30-35 1255.7 4.89 0-35 4253.0 16.54
35-40 1619.1 6.29 0-40 5872.1 22.83
40-45 2036.2 7.92 0-45 7908.4 30.75
45-50 2532.4 9.84 0-50 10440.8 40.59
50-55 3009.9 11.70 0-55 13450.6 52.29
55-60 3270.2 12.72 0-60 16720.8 65.01
60-65 3387.1 13.17 0-65 20108.0 78.18
65-70 3004.6 11.68 0-70 23112.6 89.86
70-75 1492.1 5.80 0-75 24604.6 95.66
75-80 662.2 2.57 0-80 25266.8 98.23
80-85 338.0 1.32 0-85 25604.8 99.55
85-90 92.9 0.36 0-90 25697.7 99.91
90-95 3.7 0.01 0-95 25701.4 99.92
95-100 1.3 0.01 0-100 25702.7 99.93
100-105 1.5 0.00 0-105 25704.2 99.93
105-110 1.6 0.01 0-110 25705.9 99.94
110-115 1.6 0.01 0-115 25707.5 99.95
115-120 1.6 0.00 0-120 25709.1 99.95
120-125 1.6 0.01 0-125 25710.7 99.96
125-130 1.6 0.01 0-130 25712.3 99.97
130-135 1.6 0.00 0-135 25713.9 99.97
135-140 1.5 0.01 0-140 25715.4 99.98
140-145 1.3 0.00 0-145 25716.7 99.98
145-150 1.2 0.01 0-150 25717.9 99.99
150-155 1.0 0.00 0-155 25718.9 99.99
155-160 0.8 0.00 0-160 25719.7 99.99
160-165 0.6 0.01 0-165 25720.3 100.00
165-170 0.4 0.00 0-170 25720.8 100.00
170-175 0.3 0.00 0-175 25721.0 100.00
175-180 0.1 0.00 0-180 25721.1 100.00
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 12 of 13 The NVLAP Lab Code is 200707-0.
[Additional Test]
Test Item Test Voltage (V) Frequency (Hz) Test Result
Power Factor: 277.0 60 0.9551
Total Harmonic Distortion: 277.0 60 6.06%
Total Harmonic Distortion: 120.0 60 4.75%
6. Product Photo
Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian
Free Trade Zone Shenzhen, Guangdong, China
Report No. RSZ160304501-10 Page 13 of 13 The NVLAP Lab Code is 200707-0.
7. Product Test orientation in the Goniophotometer
***********END OF REPORT***********