Monitoring of the Wavelength Calibration Lamps for the Hubble
Space Telescope
Pascucci, I. , Proffitt, C., Ghavamian, P., Sahnow, D., Oliveira, C., Aloisi, A., Keyes, T., Penton, S. V. – Proc. SPIE, Vol. 7731, 77313B (2010)
![Page 2: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/2.jpg)
STIS and COS Lamps
STIS : 3 hollow cathode Pt-Cr-Ne lamps
COS : 2 hollow cathode Pt-Ne lamps
~1130-3200Å
~5400-8000Å~3200-
5400Å
![Page 3: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/3.jpg)
When do we take a lamp spectrum?
STIS : every time we use a new grating or tilt position and generally after ~1 orbit
COS : lamp exposure contemporaneous to science observations
![Page 4: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/4.jpg)
1. Metals sputtered from the cathode trap Ne atoms➔ reduced gas pressure ➔ need higher turn-on voltage
2. Aging of the barium getter ➔ impurities mainly H ➔ H lines
3. COS: lamp beams become larger with time (not intrinsic fading)
Causes for lamps fading
typical lifetime 15,000 mA h or 1,500 h at 10 mA current
![Page 5: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/5.jpg)
STIS Lamps
STIS : 3 Pt-Cr-Ne lamps – LINE, HITM1 and HITM2
Different locations and optics: LINE brightest on the detector ➔ FUV echelle modesHITM1 and 2 ➔ optical low-res modes
![Page 6: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/6.jpg)
STIS – Lamps Fading with Time
![Page 7: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/7.jpg)
STIS – Lamps Fading in the FUV
![Page 8: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/8.jpg)
STIS – Lamp flux vs Usage
![Page 9: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/9.jpg)
STIS – Current Lamp Flux Ratios
PID 12079 : 4 internal orbits
![Page 10: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/10.jpg)
STIS – Early vs Recent Lamp Exposures
![Page 11: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/11.jpg)
STIS – The Most Affected Setting
➔ allow for the HITM2 lamp
current exp time with LINE = 46s
![Page 12: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/12.jpg)
COS and STIS Lamp comparison
when the COS sensitivity degradation is removed
![Page 13: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/13.jpg)
Aged-lamps Intensities
Nave et al. 2008
![Page 14: ILARIA PASCUCCI Space Telescope Science Institute Department of Physics & Astronomy, Johns Hopkins University Monitoring of the Wavelength Calibration](https://reader036.vdocument.in/reader036/viewer/2022070306/5519e357550346443e8b5088/html5/thumbnails/14.jpg)
Summary• The STIS LINE lamp, which was the brightest at the beginning of operation in 1997, has faded dramatically at the shortest FUV wavelengths (1150-1200Å) and is now fainter than the other two STIS lamps
• The STIS LINE lamp output remains higher than the HITM1 and HITM2 lamp outputs by factors of ~5-10 at all wavelengths > 1300Å
• The fading of the STIS LINE lamp between 1997 and 2010 in absolute terms is factor of ~15 at the very short FUV wavelengths, much greater than what expected from accelerated aging tests in the laboratory
• We have evidence that all STIS lamps fade even when they are not used indicating that on-orbit factors are impacting the lifetime of the lamps • The 6-month monitoring of the COS P1 lamp shows that the lamp output fades similarly to the STIS lamps as a function of lamp usage
Pascucci, I. , Proffitt, C., Ghavamian, P., Sahnow, D., Oliveira, C., Aloisi, A., Keyes, T., Penton, S. V. – Proc. SPIE, Vol. 7731, 77313B (2010)