Image Capture
![Page 2: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/2.jpg)
X-Ray
![Page 3: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/3.jpg)
X-Ray Generation
![Page 4: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/4.jpg)
Film
Can be exposed directly to x-ray, but sensitivity is very low
• Nitrate – 1910s to 1930s• Acetate – 1920s to mid 1960s• Polyester – mid 1950s to present
TopcoatEmulsionAdhesive
Base
![Page 5: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/5.jpg)
Intensifying Screens
Intensifying Screen
Intensifying Screen
Intensifies available x-ray light
X-rays
![Page 6: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/6.jpg)
Film PACS
Film is the detector and storage…..“The PACS”
![Page 7: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/7.jpg)
Digital Imaging
• Ultrasound
• Nuclear Medicine
• Computed Tomography
• Magnetic Resonance
![Page 8: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/8.jpg)
Computed Radiography
Digital Image Capture
![Page 9: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/9.jpg)
• Expose plate
• Drop into Reader
• PMT scans plate
• Converts to Digital
• Sends to monitor
Secondary Capture DR
CR is Digital
Digital Image Capture - CR
![Page 10: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/10.jpg)
Imaging PlateExcited Phosphor High Energy
Low Energy
X-ray Energy
Digital Image Capture - CR
![Page 11: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/11.jpg)
High Energy
Low Energy
Laser Scanning
Creating Visible Light, Digital Capture
Excited Phosphor
Digital Image Capture - CR
![Page 12: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/12.jpg)
High Energy
Low Energy
Bright White Light
Excited Phosphor
Digital Image Capture - CR
![Page 13: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/13.jpg)
Flat Panel Detector
Digital Image Capture
![Page 14: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/14.jpg)
Image Capture - DR
• Flat Panel Detector
• Image info captured
• Converts to Digital
• Sends to monitor
DR is Digital
![Page 15: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/15.jpg)
TFT Array
CsI
Photodiode
X-rays
TFT Array
GOS
Photodiode
X-rays
Discussion of CsI vs. GOS
Csi focuses scintillator light to TFT
High Image quality
GOs scatters produced light to TFT causing reduced image quality
Flat Panel Technology - Cross Section
![Page 16: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/16.jpg)
Technology Review
![Page 17: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/17.jpg)
Flat Panel Technology - Cross Section
![Page 18: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/18.jpg)
Digital Image Capture - Flat Panel
8 Million
X-ray Energy
High Energy
Low Energy
![Page 19: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/19.jpg)
High Energy
Low Energy
X-ray Energy
8 Million
Digital Image Capture - Flat Panel
![Page 20: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/20.jpg)
Digital Radiography
Performance
![Page 21: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/21.jpg)
4096
Imaging Performance
Digital R
eceptors
![Page 22: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/22.jpg)
4096
Speed
Digital R
eceptors
![Page 23: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/23.jpg)
Digital Radiography
![Page 24: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/24.jpg)
Image Generation - Today
Using 100 yr old technology……Vacuum Tubes
• Heats a negatively charged metal filament to1000º C• Combination of heat and charge releases electrons• Strikes a positively charged electrode• X-rays created by the energy of the impact
![Page 25: Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to](https://reader030.vdocument.in/reader030/viewer/2022032414/56649ef45503460f94c07448/html5/thumbnails/25.jpg)
Image Generation - Future
Electron Field Emission
• A myriad of carbon nano-tubes on a platform
• Compact source of x-rays that can be controlled with great precision, built into an array
• Each element programmed to fire when required