impact - keith conover's home page · 2012. 3. 15. · uni-vent® eagle™ model 754 portable...
TRANSCRIPT
-
IMPACT™Uni-Vent®
Eagle™
Model 754Portable Ventilator
Uni-Vent®Eagle™
-
Uni-Vent® Eagle™Model 754 Portable Ventilator
OPERATOR TRAINING PROGRAM FOR:
Emergency Services
Critical Care Transport
Aero medicine
-
Uni-Vent® Eagle™Model 754 Portable Ventilator Features
Microprocessor control of all functions including automaticmonitoring of internal battery and external power sources,internal compressor and external gases.
Extensive alarm monitoring of operating, non-operatingand advisory conditions.
Contemporary design to facilitate transport and placement.
Gas-efficient electronic control circuitry eliminates allpneumatic-logic circuits, and any dependency on gas foroperating power.
Rechargeable batteries, fully compatible with vehicularelectrical systems and airborne environments.
Microprocessor control of all functions including automaticmonitoring of internal battery and external power sources,internal compressor and external gases.
Extensive alarm monitoring of operating, non-operatingand advisory conditions.
Contemporary design to facilitate transport and placement.
Gas-efficient electronic control circuitry eliminates allpneumatic-logic circuits, and any dependency on gas foroperating power.
Rechargeable batteries, fully compatible with vehicularelectrical systems and airborne environments.
-
Uni-Vent® Eagle™Model 754 Portable Ventilator Features
Fully self-contained, may be operated withoutattachment of external gas(es) and/or external power.
Numerical panel markings indicate sequence-of-operation steps to simplify and speed start-up.
Graphics display includes 12-second pressure waveform.Its amplitude is calibrated to the adjacent digital bargraph.
Automatically compensates pressure transducer toaltitude-related barometric pressure changes up to25,000 ft.
Cont….
Fully self-contained, may be operated withoutattachment of external gas(es) and/or external power.
Numerical panel markings indicate sequence-of-operation steps to simplify and speed start-up.
Graphics display includes 12-second pressure waveform.Its amplitude is calibrated to the adjacent digital bargraph.
Automatically compensates pressure transducer toaltitude-related barometric pressure changes up to25,000 ft.
-
Your Uni-Vent® Eagle™Portable Ventilator
• Internal or External Power
• Internal Compressor
• Internal Air/Oxygen Mixer
• Software-generated PEEP
• Single-Patient-Use Circuit
• Fast Start-Up
• For use with Adults, Children,Infants
• Internal or External Power
• Internal Compressor
• Internal Air/Oxygen Mixer
• Software-generated PEEP
• Single-Patient-Use Circuit
• Fast Start-Up
• For use with Adults, Children,Infants
-
Uni-Vent® Eagle™Internal or External Power
External PowerInternal Power
External Power Supply
-
Uni-Vent® Eagle™Uses internal Compressor and/or external
gas(es) with internal Blender
Air/OxygenMixer
Manifold
Air
InternalCompressor
Air/OxygenMixer
Manifold
CPU
To Patient
Oxygen
-
Getting started,the basics…
• Ventilator circuit set-up
• Selecting a mode of operation
• Setting a ventilatory rate
• Setting an inspiratory time,or I:E ratio default
• Setting a tidal volume
• Setting an FI02• Setting High and Low Limit
Pressure alarms
• Ventilator circuit set-up
• Selecting a mode of operation
• Setting a ventilatory rate
• Setting an inspiratory time,or I:E ratio default
• Setting a tidal volume
• Setting an FI02• Setting High and Low Limit
Pressure alarms
-
Ventilator circuit set-up
• Connect 22mm corrugated hose toventilator GAS OUT fitting
• Connect green TRANSDUCER HOSEto ventilator TRANSDUCER hose barb(Green)
• Connect clear EXHALATION VALVEHOSE to Ventilator EXHALATIONVALVE hose barb (clear)
-
The basics…
Select a Mode of operation
Set a Ventilation RateSet a Ventilation Rate
Set an Inspiration Time, or I:ERatio default
Set a Tidal Volume
Set an FI02Set the High and Low LimitPressure Alarms
-
Additional functions…
• External Air• Sigh
• PEEP• PEEP• Pressure Plateau
• Manual Breath/Trigger
• Alarm Mute/Cancel
• Apnea Backup (Automatic)
-
Additional functions…External Air
• External AirThe EXTERNAL AIR Pushbutton Switchpermits you to manually select externalcompressed air as your primary air source.When the ventilator is turned ON, its CPU"samples" the respective gas fitting and looksfor an air pressure, greater than 40-PSI. If alower pressure, or no pressure is sensed, theLCD will display "OFF" (default value) andthe CPU allows operation to begin from theinternal compressor. If a pressure exceeding40-PSI (+/- 2 PSI) is sensed, the LCD willdisplay "ON" and allow operation to beginfrom the external air supply. Manuallypressing the pushbutton will toggle the displayfrom "OFF" to "ON or from "ON" to "OFF".
The EXTERNAL AIR Pushbutton Switchpermits you to manually select externalcompressed air as your primary air source.When the ventilator is turned ON, its CPU"samples" the respective gas fitting and looksfor an air pressure, greater than 40-PSI. If alower pressure, or no pressure is sensed, theLCD will display "OFF" (default value) andthe CPU allows operation to begin from theinternal compressor. If a pressure exceeding40-PSI (+/- 2 PSI) is sensed, the LCD willdisplay "ON" and allow operation to beginfrom the external air supply. Manuallypressing the pushbutton will toggle the displayfrom "OFF" to "ON or from "ON" to "OFF".
-
Additional functions…SIGH
• SIGHThe SIGH OFF/ON Pushbutton Switch permitsthe ventilator to operate with or without SIGH.When the SIGH function is activated, the firstventilator-generated breath is always a SIGH.Afterwards, each additional SIGH breath isdelivered once every 100 breaths or 7-minutes,whichever occurs first. A SIGH breath equals150% of the INSPIRATION TIME setting (andthe exhalation period). This will increase thedelivered volume of that breath by 50%. Byproportionately increasing both the inspirationand exhalation time, Uni-VentTM maintains theI:E ratio for all breaths. As a safetyprecaution, Uni-VentTM will not allow theinspiratory portion of a SIGH breath to exceed3 seconds.
The SIGH OFF/ON Pushbutton Switch permitsthe ventilator to operate with or without SIGH.When the SIGH function is activated, the firstventilator-generated breath is always a SIGH.Afterwards, each additional SIGH breath isdelivered once every 100 breaths or 7-minutes,whichever occurs first. A SIGH breath equals150% of the INSPIRATION TIME setting (andthe exhalation period). This will increase thedelivered volume of that breath by 50%. Byproportionately increasing both the inspirationand exhalation time, Uni-VentTM maintains theI:E ratio for all breaths. As a safetyprecaution, Uni-VentTM will not allow theinspiratory portion of a SIGH breath to exceed3 seconds.
-
Additional functions…PEEP
• PEEPThe PEEP OFF/ON-SET Value PushbuttonSwitch activates Uni-VentTM's internal PEEPlearning software. When the ventilator isturned ON, PEEP has a default value of 0cmH2O (no PEEP). Each time thepushbutton is pressed, PEEP is increased by1 cmH2O up to a maximum value of 20cmH2O (then back to “0” again). It may takeseveral breaths for the learning process toconclude. During this period you will see aPEEP value successively added to thepressure waveform after each breath. Toprotect against high airway pressures, Uni-VentTM will “dump” PEEP (to 0 cmH2O) ifthe patient should cough.
The PEEP OFF/ON-SET Value PushbuttonSwitch activates Uni-VentTM's internal PEEPlearning software. When the ventilator isturned ON, PEEP has a default value of 0cmH2O (no PEEP). Each time thepushbutton is pressed, PEEP is increased by1 cmH2O up to a maximum value of 20cmH2O (then back to “0” again). It may takeseveral breaths for the learning process toconclude. During this period you will see aPEEP value successively added to thepressure waveform after each breath. Toprotect against high airway pressures, Uni-VentTM will “dump” PEEP (to 0 cmH2O) ifthe patient should cough.
-
Additional functions…Manual Breath/Trigger
• Manual Breath/TriggerThe MANUAL BREATH/TRIGGERPushbutton Switch delivers a manualbreath each time the pushbutton switchis pressed. The inspiratory time andtidal volume of each manual breath isequal to the current control settings. If aventilator failure occurs, the operatorcan manually trigger a continuous flowof gas can be triggered for as long as thepushbutton switch is pressed.
The MANUAL BREATH/TRIGGERPushbutton Switch delivers a manualbreath each time the pushbutton switchis pressed. The inspiratory time andtidal volume of each manual breath isequal to the current control settings. If aventilator failure occurs, the operatorcan manually trigger a continuous flowof gas can be triggered for as long as thepushbutton switch is pressed.
-
Additional functions…Mute/Cancel Pushbutton
• Mute/CancelThe MUTE/CANCEL PushbuttonSwitch allows the operator totemporarily mute the audiblecomponent of an Operating Alarm orcancel the audible and flashing LEDAlarm Indicator components of anAdvisory Alarm. Most mute periods are30-seconds long. An External PowerLow/Failure Alarm mute will last for 5-minute intervals until a Low BatteryAlarm occurs and for 30-secondintervals thereafter.
The MUTE/CANCEL PushbuttonSwitch allows the operator totemporarily mute the audiblecomponent of an Operating Alarm orcancel the audible and flashing LEDAlarm Indicator components of anAdvisory Alarm. Most mute periods are30-seconds long. An External PowerLow/Failure Alarm mute will last for 5-minute intervals until a Low BatteryAlarm occurs and for 30-secondintervals thereafter.
-
The LCD Screenwhat you can see…
• Control settings• EXTERNAL AIR, SIGH, PEEP,
PLATEAU and PRESSUREALARMS settings are displayedunder their respective controls.
• Control settings• EXTERNAL AIR, SIGH, PEEP,
PLATEAU and PRESSUREALARMS settings are displayedunder their respective controls.
• VENTILATION RATE,INSPIRATION TIME/I:E RATIO,TIDAL VOLUME and AIR /OXYGEN MIXER settings aredisplayed above their respectivecontrols.
-
The LCD Screenwhat you can see…
ALARM MESSAGE CENTER (AMC):The centralized location for displayingup to 4 lines of alarm messageinformation.
MODE and Vmin indicators:Displays operating mode and minutevolume (A/C Mode).Inhalation/Exhalation indicator:Alternately displays the inspiration andexhalation phase of mechanical and/orspontaneous breaths.
Cont….
BATTERY and EXTERNAL POWERsource indicators:2-line area that displays the currentstatus of external power, internalpower, and fuses.
MODE and Vmin indicators:Displays operating mode and minutevolume (A/C Mode).Inhalation/Exhalation indicator:Alternately displays the inspiration andexhalation phase of mechanical and/orspontaneous breaths.
-
The LCD Screenwhat you can see…
Peak & Mean inspiratory pressuresare displayed after each breath
The digital bar graph displays peakpressure during each breath
Cont….
High and Low Pressure Limit Alarmsetting “tics” are continuouslydisplayed
Set and Delivered tidal volume arealternately displayed
6-second’s of pressure wave formdata, including PEEP is displayed
The digital bar graph displays peakpressure during each breath
-
Thank You!Thank You!
Email: [email protected]
IMPACT™Uni-Vent® Eagle™ Model 754 Portable VentilatorUni-Vent® Eagle™ Model 754 Portable Ventilator FeaturesUni-Vent® Eagle™ Model 754 Portable Ventilator FeaturesYour Uni-Vent® Eagle™ Portable VentilatorUni-Vent® Eagle™ Internal or External PowerUni-Vent® Eagle™ Uses internal Compressor and/or external gas(es) with internal BlenderGetting started, the basics…Ventilator circuit set-upThe basics…Additional functions…Additional functions… External AirAdditional functions… SIGHAdditional functions… PEEPAdditional functions… Manual Breath/TriggerAdditional functions… Mute/Cancel PushbuttonThe LCD Screen what you can see…The LCD Screen what you can see…The LCD Screen what you can see…Thank You!