DESCRIPTION
Incubators - ASTEC IVFCube Instruction Manual rev.5 2013.6.6 supplied by RI UK and Ireland The following instructions are for your own protection and safety use. Please read this manual carefully before start using the instrument. Failure to observe these precautions can result in property damage or personal injuries. All Information Correct at time of publication.TRANSCRIPT
AR-3100GC
Operation Manual
Drawer Type Incubator
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
2
1. General Description ................................................................................................ 4
1-1 About This manual ............................................................................................ 4
1-2 Warranty and Liability........................................................................................ 4
1-3 Warnings ........................................................................................................... 4
1-4 Warning Labels ................................................................................................. 5
2. Bio Hazard............................................................................................................... 8
3. Important Notice...................................................................................................... 8
4. Copy Rights............................................................................................................. 8
5. Specifications .......................................................................................................... 9
6. Accessories............................................................................................................. 9
7. Parts Description................................................................................................... 10
8. Description of Control Panel ................................................................................. 11
9. Prior to Installation ................................................................................................ 13
9-1 Initial Cleaning and Disinfecting...................................................................... 13
9-2 Procedure........................................................................................................ 14
10. Humidity for the Incubation Chamber ................................................................. 15
10-1 Filling the Humidity Slot................................................................................. 15
11. Connecting gas ................................................................................................... 16
11-1 Connecting CO2/N2 Gas Supply.................................................................... 16
11-2 Gas Pressure................................................................................................. 16
12. Main Power ......................................................................................................... 17
13. Setting up AR-3100GC ....................................................................................... 17
14. Gas Concentration Check................................................................................... 18
15. Drawer type Incubation Chamber ....................................................................... 19
16. Automatic Calibrations ........................................................................................ 19
17. Basic Settings ..................................................................................................... 20
17-1 Temperature Setting...................................................................................... 20
17-2 Overtemp Setting .......................................................................................... 20
17-3 CO2 Setting ................................................................................................... 20
17-4 Gas Selection................................................................................................ 20
17-5 O2 Setting ...................................................................................................... 20
18. Adjustment and Calibration................................................................................. 22
18-1 Adjustment / Calibration Mode:..................................................................... 22
18-2 Lid Temperature Adjustment A-3 .................................................................. 22
18-3 CO2 ZERO Manual Calibration A-4 ............................................................... 22
18-4 CO2 SPAN Adjustment A-5 ........................................................................... 22
18-5 O2 ZERO Adjustment A-6.............................................................................. 23
18-6 O2 ZERO Manual Calibration A-7.................................................................. 23
Index of Contexts
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
3
18-7 Flow Chart of Adjustment and setting ......................................................... 24
19. Chamber Temperature Calibration...................................................................... 25
20. Alarms ................................................................................................................. 26
20-1 CO2 Alarm...................................................................................................... 26
20-2 Overtemp Alarm ............................................................................................ 26
20-3 CO2 Alarm...................................................................................................... 26
20-4 O2 Alarm ........................................................................................................ 26
21. Cleaning and Disinfectant ................................................................................... 27
22. Wiring Diagram...................................................................................................... 28
23. Maintenance .......................................................................................................... 30
Descriptions of primary components ........................................................................... 31
Descriptions of Gas Sensors and Tubing.................................................................... 33
Descriptions of Gas Control Box ................................................................................. 34
Descriptions of Drawers............................................................................................... 35
Descriptions of Gas supply unit ................................................................................... 36
Descriptions of Gas supply Flow ................................................................................. 37
23. Warranty Information........................................................................................... 38
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
4
1. General Description
1-1 About This manual
1-2 Warranty and Liability
We warrant the product you have purchased for one calendar year after the date of purchase. In case
of any malfunctions caused by the manufacturer during this period of time, ASTEC will be responsible for
repair or replacement of failed parts or even the instrument itself as circumstance allows for free of
charge.
However, this warranty excludes the following conditions.
Any damages due to fire, earthquake, thunderstorm, and other catastrophic events and pollution and
abnormal electrical supply.
Any damages due to unduly repair, adjustment, calibration and modification.
Any damages due to unduly usage or mishandling.
Any damages caused by moving, dropping, transportation of the instrument.
Replacement and repair of expendables and consumables.
For service, please contact our agent that you have purchased the instrument from.
1-3 Warnings
Connect the instrument to a line plug according to local or county standards and to the specifications.
Do not use the instrument for different purposes from those shown by the manufacturer. (See Chapter 1
for General Descriptions)
Do not switch on the instrument without having waited at least 20 seconds after switching off.
Do not open the instrument before reading the servicing specific conditions described in the manual.
Follow instruction contained in the manual in case of changing any parts
To move the instrument, at least two persons are required. We suggest to contact the distributor or
directly to the manufacturer for servicing.
Thank you for purchasing ASTEC Drawer type incubator AR-3100GC. The following
instructions are for your own protection and safety use. Please read this manual carefully
before start using the instrument. Failure to observe these precautions can result in property
damage or personal injuries. Make sure to keep this manual handy so that you can read this
whenever you may need it.
Warranty is Valid only if the installation is done by people trained, in conformity
to instructions provided in this manual.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
5
1-4 Warning Labels
AR-3100GC has an intrinsic danger due to the structure designed for the purpose. For the safe use of
the instrument, the following mark is labeled on the instrument. Please use this instrument under a
caution indicated on the labels.
Power inlet on the back of the incubator is labeled with the Warning sticker.
The following symbols are used in this manual.
Warning, Caution
Prohibited
Important Note
Ventilate the room in which you use carbon dioxide with the instrument. It can harm your
body without adequate air ventilation. Do not breath carbon dioxide directly.
Do not place flammable objects in the incubator, such as alcohol, benzene, ethanol or
propane gas. Especially when you use oxygen, do not use matches or lighters or
cigarettes into the laboratory.
Do not use this instrument outdoors. It could cause an electrical short or a lethal shock.
Ask the manufacturer or your sales agent for the setting up. This will prevent from the
events such as gas leaks, water leaks or any other unexpected problems caused by
mishandling.
Warning This may cause serious or lethal injury.
Caution This may cause serious injury or damage
WarningWarningWarningWarning
Electric Shock The earth terminal MUST be
earthed properly
Cautions
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
6
The location where the instrument will be placed must be steady, firm and leveled.
Do not leave this instrument in humid or wet conditions. It could cause an electrical short.
When double-stack the incubators, use the optional brackets. The top incubator may fall
by earthquake or other unexpected cause or accidents.
Connect only to a suitable power source specifically assigned for the instrument(s)
Using the instrument outside of Japan may require a transformer. Please ask the
manufacturer or your sales agent.
The power plug should be inserted all the way into the power outlet. Make sure that the
power plug is clean and dust-free when used.
Make sure that the instrument is securely earth-grounded.
Do not earth the instrument to the gas pipe, water pipe and lighting rod. This will cause
an electrical shock.
Never use the flammable objects with the instrument.
Make sure the type of gas and appropriate gas pressure are used with the instrument. All
the gas connections with gas tube(s) must be firmly connected and tightened with
cable-tie.
Do not plug any wire or pin into gas ports, sampling port.
Toxic, pathogenic and radioactive substances should be kept away from the instrument.
During maintenance and cleaning, make sure that power off the instrument and unplug the
power code.
Do not take apart or modify the instrument without first-hand supervising.
In case of malfunctions, turn off the power switch in the back of the instrument, and then,
unplug the power cord.
When the incubator is disposed, please be concerned for environment safety.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
7
Pedestal, bench, desk or table where the instrument will be set up MUST be solid and
leveled even to prevent leakage of the water in the Jacket.
Using the correct power source specifically assigned for the instrument is extremely
required.
Do not unplug the power cable with soaked hands.
In case any cuts or worn-out was found on the power cable of the instrument, stop using
the cable and replace it with brand-new power cable.
Do not operate the instrument with soaked hands.
Do not put heavy objects or bottles that contain liquid.
Do not get on the instrument.
When close the inner and outer doors please hold the knob and handle in appropriate
manner to avoid pinching your hands or fingers.
Do not intentionally step on, bend or twist the power cable.
Do not lean against the instrument.
Do not lean against the door of the instrument, especially when it is open.
Use appropriate gloves when cleaning the instrument, especially for cleaning the chamber.
Do not inhale any chemicals, organic compounds, dust or moisture when cleaning.
Power off and unplug the power cable when moving the instrument to different location.
Remove the water from the water pan when moving the incubator to the different location.
In case the instrument will be unused for a long period of time, remove the water in the
water pan and dry out the chamber completely.
Do not splash the water to the outside of the instrument especially on the top of it.
The receiver of O2 sensor should be kept away from condensation.
Used O2 sensor should be properly discarded in order to avoid contamination of the natural
environment.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
8
2. Bio Hazard
If the product has been used with any substances that may be hazardous to human health, and it needs
to be sent back for maintenance, the product must be completely and thoroughly decontaminated before
being sent back.
3. Important Notice
(Read this section to the end prior to use for your own safety)
AR-3100GC requires separate CO2 and N2 cylinders. Premixed gas (ex. CO2: 5% O2: 5% N2: 90%) is
prohibited.
A power outlet should be solely assigned to the incubator. Use of power strip or same kind of
connections is not recommended.
Acid, benzene, soap, toothpaste are not recommended to use for cleaning.
Use of brush to any part of the incubator for cleaning is not recommended.
Do not wipe the plastic and rubber area with volatile liquid such as benzene. Use neutral detergent and
wipe off completely with water.
Do not clean the incubator by poring water (especially on top panel) and wiping it off.
Pure distilled or even cleaner water is recommended for the humidification in the chamber in order to
avoid any chance of contamination. Check the volume of remaining water at least once in 3 days.
Water for the humidification must be replaced or refilled as needed. The water should be pre-warmed
to the same temperature as the set temperature of the incubator.
Pedestal, bench, desk or table where the incubator will be set up MUST be solid and leveled.
Do not place flammable substances and volatile compounds in the incubator, such as alcohol, benzene,
ethanol, adhesive or propane gas. Especially when you use oxygen, do not use matches or lighters
and/or smoke cigarettes into the laboratory.
CO2 cylinder, which is equipped with siphon inside, cannot be used with the incubator.
Gas pressure to supply into the incubator should be maintained in the ranges between 0.03~0.05Mpa
(0.3~0.5 bar / 4.35~7.25 PSI).
In case the incubator will be kept powered off, remove the water for the humidification and dry out the
chamber completely.
The incubator MUST be properly earthed to ground.
Worn out O2 sensor should be properly discarded in order to avoid contamination of the natural
environment.
4. Copy Rights
All rights are reserved by ASTEC Co., Ltd. Revision, modification of the contents of this manual prior to
permission is strictly prohibited.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
9
5. Specifications
Model Drawer Type Incubator (Cat/N AR-3100GC)
Exterior Dimension W566 x D563 x H533*(mm) / W22.3 x D22.2 x H21.0*(inch)
Inner Dimension W180 x D217 x H20 (mm) / W7.1 x D8.6 x H0.8 (inch)
Drawer Chamber Capacity 775ml
4 x 4-well dish / 6 x 60mm petri dish / 12 x 35mm petri dish
Drawer Chamber Weight 2.8kg (6.2lb) / per chamber
Drawer Chamber
Number of Drawers 4
Temperature Control Digital PID control
Temperature Range 7℃ + ambient to 40℃
Temperature Accuracy (+-)0.2℃
Temperature Uniformity (+-)0.4℃
CO2 sensor Infra Red
CO2 Control Range 0~10%
CO2 Uniformity (+-)0.3%
O2 Sensor Galvanic battery sensor
O2 Control Range 4~10%
O2 Uniformity (+-)0.5%
Humidity in Chamber 90%RH
Humidity method Natural evaporation
Gas Used CO2 / N2
Gas Flow CO2 : 40ml/min / N2 : 120ml/min
Alarms User Programmable on Temperature, CO2, N2
Alarm output DC12V, 0.3V (Optional)
Safety feature Temperature cut-off (Software controlled)
Analog output Temperature, CO2, N2 (optional)
Weight 80kg / 177lb
Power 110V, 117V, 220V, 230V 50/60Hz
*Without adjustors
6. Accessories
List of standard accessories associated with AR-3100GC
Instruction Manual 1 Warranty Card 1
Plastic gas tube (3m) 1 3P adapter 1
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
10
7. Parts Description
(1) Incubation Drawer Chamber
(2) Incubation Drawer Chamber
(3) Incubation Drawer Chamber
(4) Incubation Drawer Chamber
(5) Control Panel
(6) Analog Data Output
(7) Alarm Output
(8) Main Power
(9) Circuit Breaker
(10) AC adaptor
(11) CO2 Inlet Port
(12) N2 Inlet Port
(13) Gas Sampling Port
(14) Gas Sampling Port
(15) Chamber Lid Knob
(16) Drawer Knob
(17) Aluminum block chamber
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
11
8. Description of Control Panel
Continues on to the next page.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
12
(1) Temperature Display Displays Temperature and setting value for Temperature
(2) CO2 Display Displays CO2 concentration and setting value for CO2
(3) O2 Display Displays O2 concentration and setting value for O2
(4) Heating Indicator Lights on when the chamber being heated
(5) CO2 Inject indicator Lights on when the CO2 being injected into the chamber
(6) O2 Inject indicator Lights on when the O2 being injected into the chamber
(7) Over Temp Alarm Indicator Lights on when Over temp alarm goes off with a beep
(8) CO2 Alarm Indicator Lights on when CO2 alarm goes off with a beep
(9) O2 Alarm Indicator Lights on when O2 alarm goes off with a beep
(10) Temp Set Indicator Lights on when chamber temperature being set
(11) Over Temp Set Indicator Lights on when temperature upper limit being set
(12) CO2 Set Indicator Lights on CO2 concentration upper limit being set
(13) N2/O2 Set Indicator Lights on when gas selection N2/O2 being modified
(14) O2 Set Indicator Lights on O2 concentration upper limit being set
(15) UP arrow key Used to increase value for settings and calibrations
(16) Set key Used to set value for settings and calibrations
(17) Down arrow key Used to decrease value for settings and calibrations
(18) Silence key Used to mute alarm when activated
(19) Control Panel for 2nd Drawer Same configurations as the control panel for the 1
st drawer
(20) Control Panel for 3rd Drawer Same configurations as the control panel for the 1
st drawer
(21) Control Panel for 4th Drawer Same configurations as the control panel for the 1
st drawer
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
13
9. Prior to Installation
� The place (bench, table or stand) where the incubator will be set up MUST be steady, firm and
leveled. Make sure the surface supports the weight of the incubator in operational conditions.
� To move the incubator, two or more persons are needed.
� Please keep 10 inches or more of clearance in the back and both sides of the incubator.
� Avoid placing an incubator where cold air drifts directly to the incubator such as air conditioning or
air vent. It may cause condensation in the chamber.
� All the surface of the chamber should be cleaned and disinfected before operation.
� The incubator should be placed in a clean room environment.
When placing the drawer type incubator on the optional base stool (P/N: AR-3100STD) balance the level
of incubator adjusting the height of the foots located at each bottom corner of the incubator and secure it
to the shelf using M6 screw bolts included in the base stool package. Drawer Type incubator has
dedicated threaded holes on the bottom surface where indicated by red circles in the below diagram.
9-1 Initial Cleaning and Disinfecting
Cleaning and disinfecting the incubator before its use are critical steps. The initial process and
subsequent maintenance should be done properly and following the steps included in this manual.
The incubator should be cleaned and disinfected with substances that can be removed with proper
rinsing. Any toxic substance or its residue could be biologically toxic which potentially affecting the
culture conditions end specimens development.
Figure. Bottom Surface of AR-3100GC
Front
Back
Threaded Holes
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
14
9-2 Procedure
Wipe all incubator surfaces (internal and external) with a 3% solution of 7X detergent in distilled water
using a damp sponge or a soft cloth (Kim wipes).
Rinse with distilled water using a damp sponge or soft cloth.
Wipe all surfaces (internal and external) with OH Peroxide solution (3%).
Rinse with distilled water using a damp sponge or soft cloth.
Dry the incubator chamber and all incubator parts with a lint-free soft cloth.
Wrap aluminum incubation block for sterilization (use steam sterilizer or oven only - DO NOT USE
ETYLHENE OXIDE METHOD for this purpose).
Once the sterilization of the incubator parts is completed and the parts are at room temperature open the
wrapping and install it.
The power outlet
Please double check for the appropriate power voltage suitable for the IVF CUBE you have purchased.
The power outlet must be individually assigned to the unit; we do not recommend multiple connections
for any circumstance.
The product must be properly grounded to the earth to avoid electric shock when short circuit.
Connect the incubator to a grounded dedicated circuit only. The power cord connector is the
main disconnect device for the incubator. Position the unit so that it can be easily disconnected.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
15
10. Humidity for the Incubation Chamber
10-1 Filling the Humidity Slot
Water in the humidity water slot should be distilled water and supplied in the slot with approximately
20ml of the water. Humidity water, because it is stagnant, can be common source of contamination
within the incubator. So we strongly recommend that the humidity water should be replaced once in at
least 3 days or more often, and the water slot should be autoclaved as often as possible.
It is also recommended to warm up the humidity water to about 37C before it is poured into the slot for
faster evaporation.
When the incubator is powered off more than 30 minutes, make sure to remove the water
from the chamber. When incubator is powered back up, make sure there is NO humidity in
the chamber. Humidity can harm CO2 and O2 sensor when incubator is powered off.
Humidity Water Slot
Open Space
Do not fill water in the open space to avoid clogging the CO2 supply port by the water.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
16
11. Connecting gas
11-1 Connecting CO2/N2 Gas Supply
The Gas port is located on the back of the incubator as shown in the figure below. Connect plastic
tubing (supplied with the incubator) and firmly tie them up with cable ties or clamps. The other end of
the tubing needs to be connected to gas regulator of gas supply tank or gas manifold.
11-2 Gas Pressure
The optimal gas supply pressure is approximately 0.03 Mpa (0.3 bar / 4.35 psi).
Excess gas pressure may cause malfunction and/or damage of the incubator.
N2 Gas Port
CO2 Gas Port
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
17
12. Main Power
Main power switch is located in the back of the unit as shown in the figure below. Make sure that the
AC cable is plugged into the power outlet assigned individually for the incubator, then, turn on the unit.
13. Setting up AR-3100GC
Gas-Stop Mode
The Gas Stop Mode is recommended for setting up the incubator for the first time, or setting it up again
after a complete cleaning and disinfection process, or after a long-time unused period.
Gas-Stop Mode: Press and hold the SET key on the control panel for about 5 seconds.
The ON/OFF control of Gas Stop Mode is independent for each chamber.
During the Gas Stop Mode, the Temp display shows the actual temperature and “G.ST” intermittently,
and both CO2 and O2 solenoid valve operations completely stop but warm up the chamber(s) in order to
bring the temperature and humidity to the stable condition before the gas supply.
The Gas Stop Mode needs to be turned off manually. To return the chamber(s) into normal operation,
press and hold the SET key on the control panel for about 5 seconds while the chamber(s) is in Gas Stop
Mode.
Main Power Switch
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
18
14. Gas Concentration Check
CO2 / O2 Concentration check
The CO2 and O2 gas concentration must be checked frequently using an analyzer. It is always
recommended to use an analyzer that has air-circulating function between the device and the incubator
to minimize the environmental change in the incubator during the measurement.
There are 2 sampling ports on each drawer, and either port can be used for sampling. The left
sampling port is covered by a cap. Remove the cap and connect one of the ports to the sampling port
of the gas analyzer, and connect the other port to the outlet port of the gas analyzer. Let the analyzer
read the concentration continuously until you finish the calibration. Ambient air remaining in the analyzer
may dilute the chamber concentration and the display values slightly drop from the current values. Wait
about 2 minutes for both incubator’s display and analyzer’s reading to become stable and compare the
difference in the concentration.
If the displayed CO2 and/or O2 concentration differs from the concentration measured with calibrated
analyzer, adjust the concentration on the display of the incubator to the actual concentration measured
with the analyzer. For the CO2 display adjustment, change the CO2 Span parameter to the same value
as that of actual CO2 concentration at A-4 menu. For the O2 display adjustment, calculate the
difference between the two values (actual O2 concentration - O2 display) and add this number to the
current value in A-6 menu. Refer to page 22 and 23 for more detail of adjustment menu.
Note: It is strongly recommended to use an analyzer which can circulate air between the incubator and
the device to prevent significant gas concentration change in the drawer chamber.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
19
Sampling Mode: <UP> key pressed for 5 seconds *Factory Use Only
During the Sampling Mode, Temperature display shows “SP” and the current temperature for 3 minutes.
Concentration check must be done during this period. After 3 minutes, the Sampling Mode
automatically cancels and goes back to normal function.
15. Drawer type Incubation Chamber
Circulation of the gas in the Chamber
Regardless of the lid of the drawer is opened or not, once the drawer is pulled out and pulled back in,
AR-3100GC starts circulating the gas in the drawer chamber
As soon as the drawer is pushed back in, it starts circulating the gas in order to mix the gas in the
incubation chamber and CO2 sensors (and tubes). This helps the sensor to detect the correct gas
concentration and much faster concentration recovery. The gas circulation continues for 1 minute, and
during this circulation, O2 display shows “CYC”. During this 1 minute period gas supply into the
chamber is stopped.
16. Automatic Calibrations
AR-3100GC performs calibrations on CO2 and O2 sensors once in every week. During the calibration
of CO2 sensor, CO2 display shows “CAL” and actual concentration alternately. When O2 sensor is
calibrated, O2 display shows “CAL” and actual O2 concentration alternately.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
20
17. Basic Settings
Every operation and parameter can be changed by using <SET> key on the front control panel. Using
<UP> and <Down> cursors allows changing the value for the settings or changing the parameters.
After changing the value or parameter, if the operation is left unchanged about 7 seconds, it goes back to
the normal display, which indicates the current condition of the incubator.
17-1 Temperature Setting
Press <SET> key once, the TEMP SET indicator lights on. Use <UP> and <DOWN> cursors to set the
temperature to the value you wish to set.
Press <SET> key once more, the control panel brings you to Over Temp Setting. If the control is left
unchanged for about 7 seconds, it returns to normal display.
17-2 Overtemp Setting
Press <SET> key until the Over temp set point lights on. Use <UP> and <DOWN> cursors to set the
temperature to the value you wish to set. In case the incubator reaches the temperature you set here in
the chamber, it will activate the Over Temp alarm.
For detailed descriptions for Over Temp Alarm, see the Alarms section on Page 24.
Press <SET> key once more, the control panel brings you to CO2 concentration setting. If the control is
left unchanged for about 7 seconds, it returns to normal display.
17-3 CO2 Setting
Press <SET> key until the CO2 set point lights on. Use <UP> and <DOWN> cursors to set the CO2
concentration you wish to set.
Press <SET> key once more, the control panel brings you back to the Temperature Setting.
If the control is left unchanged for about 7 seconds, it returns to normal display.
17-4 Gas Selection
Always choose N2 by Use <UP> and <DOWN> cursors
17-5 O2 Setting
Press <SET> key until the O2 set point lights on. Use <UP> and <DOWN> cursors to set the O2
concentration you wish to set.
If the control is left unchanged for about 7 seconds, it returns to normal display.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
21
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
22
18. Adjustment and Calibration
Adjustment / Calibration Mode is utilized for modifications to Temperature and CO2 concentrations by
changing factory pre-set parameters as needed. The modifications should be made accordingly to
adjust to the environment in which the incubator is used.
Some of the pre-set parameters should be modified only by a qualified service technician, as they are
critical to overall performance of the incubator. These parameters are indicated here below in this
section.
By pressing <SET> key for about 0.5 sec with <UP> and <DOWN> keys pressed and held, the
Adjustment / Calibration Mode will be activated.
18-1 Adjustment / Calibration Mode:
<UP> + <DOWN> Pressed and Hold + SET pressed for 0.5 sec
Temperature Adjustment (In-chamber temperature) A-1
When Temperature display shows ST1 in Adjustment / Calibration Mode, press SET key to go to
section A-1. You will see a value in this section. Increasing or decreasing the value here allows
you to adjust the actual temperature in the chamber adjusted to the actual temperature.
18-2 Lid Temperature Adjustment A-3
When Temperature display shows ST1 in Adjustment / Calibration Mode, press SET key until you
see section A-3, Increasing or decreasing the value here allows you to adjust the temperature.
For example, when you need to increase the displayed temperature by 0.2 C, decrease the value
shown by 0.2 using <UP> and <DOWN> keys.
18-3 CO2 ZERO Manual Calibration A-4
This function allows the incubator to carry out manual ZERO Calibration CO2 concentration. As a
default, A-4 shows OFF on the display. Change this to ON with <UP> and <DOWN> cursors and
wait for 10 second until the incubator starts manual calibration.
18-4 CO2 SPAN Adjustment A-5
When measured CO2 concentration is different from the CO2 display concentration and it needs to
be adjusted, change the value on A-5 to actual CO2 concentration by using <UP> and <DOWN>
cursors. It will allow the incubator to automatically adjust the value to the measured value.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
23
18-5 O2 ZERO Adjustment A-6
When measured O2 concentration is different from the O2 display concentration and it needs to be
adjusted, calculate the difference between these two values (A) and change the current value in A-6
by the same amount using <UP> and <DOWN> cursors. After changing the A-6 value, go back to
the normal display and confirm that O2 display shows the same value as that of the measured O2
concentration.
(A) = (Measured O2 Concentration) – (O2 Display Concentration)
18-6 O2 SPAN Manual Calibration A-7
This function allows the incubator to carry out manual ZERO Calibration for O2 concentration. As a
default, A-7 shows OFF on the display. Change this to ON with <UP> and <DOWN> cursors and
wait for 10 second until the incubator starts manual calibration.
Make sure to use a calibrated thermistor when temperature in chamber is re-calibrated.
Do not change the value on A-2 and b1~Gb as these setting are only used by the
manufacturer.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
24
18-7 Flow Chart of Adjustment and setting
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
25
19. Chamber Temperature Calibration
1. Prepare a calibrated thermometer and attach the sensor at the center of the dish plate surface where
indicated as “CAL Point” (Figure A-1). An electronic thermometer with thin lead is recommended.
2. After setting a thermometer in the chamber, close the lid and wait at least 1 hour. Once lid is opened,
it should be left running at least another 1 hour until the temperature is stabilized completely.
3. If the temperature on the thermometer indicates different value from the Temp display, go to "ST1"
and "A-1" for adjustment. Increase the value on "A-1" brings the temperature down, and decrease
the value brings the temperature up. For example, when set temp is 37.0 C and the actual temp is
36.6, you need to set the control temp up by 0.4. In this case, decrease the value by 0.4 in order to
bring up the actual temperature by 0.4.
Lid Temperature Adjustment
When Temperature display shows ST1 in Adjustment / Calibration Mode, press SET key until you see
section A-3, Increasing or decreasing the value here allows you to adjust the temperature.
For example, when you need to increase the displayed temperature by 0.2 C, decrease the value shown
by 0.2 using <UP> and <DOWN> keys.
Figure. A-1
CAL Point
In case you need to carry out the Calibration more than once, make sure that you have
at least 1 hour between Calibrations for the chamber temperature to stabilize.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
26
20. Alarms
20-1 CO2 Alarm
Press <SET> key in Function suspending mode until Temperature display shows ALA. Then, CO2
display shows ON or OFF. Use <UP> and <DOWN> key to change the setting. This function only
suspends CO2 alarm.
20-2 Overtemp Alarm
OVERTEMP. Alarm activates when the temperature inside the chamber reaches the temperature set on
OVERTEMP setting. The alarm triggers a beeping sound that can be shut off by pressing the
<SILENCE> key. However, OVER TEMP setpoint indicator keeps blinking until the temperature in the
chamber actually decreases. While the alarm is being activated, incubator partially shuts down the
power to all chamber heaters for safety reasons.
20-3 CO2 Alarm
CO2 Alarm activates when CO2 concentration is off ±1.0% or more from the concentration that is set on
CO2 concentration setting for 15 consecutive minutes. The alarm triggers a beeping sound and CO2
Alarm indicator blinks. By pressing <SILENCE> key, beeping shuts down and the CO2 indicator turns
off.
20-4 O2 Alarm
O2 Alarm activates when O2 concentration is off ±1.0% or more from the concentration that is set on O2
concentration setting for 15 consecutive minutes. The alarm triggers a beeping sound and O2 Alarm
indicator blinks. By pressing <SILENCE> key, beeping shuts down and the O2 indicator turns off.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
27
21. Cleaning and Disinfectant
Disinfecting the incubator interior
The incubator should be cleaned and disinfected with substances that can be removed with proper
rinsing. Any toxic substance or its residue could be biologically toxic which potentially affecting the
culture conditions end specimens development. The instructions here included have demonstrated to
clean and disinfect the incubator and its components allowing proper culture conditions as demonstrated
by validation testing and Mouse Embryo Assay (MEA).
Preparation
It is recommended to use gloves, safety goggles as well as a protective coat. Make sure the incubator is
unplugged.
Procedure
1. Turn the incubator off and unplug the power cord from the outlet wall. Shut off the CO2 supply from
the gas cylinder or gas manifold.
2. Open the lid of the incubation drawers and remove water from the slot.
3. Remove the aluminum incubation block out of the drawers.
4. Wipe all incubation drawer surfaces (internal), lids with a 3% solution of 7X detergent in distilled
water using a damp sponge or a soft cloth (Kim wipes).
5. Rinse with distilled water using a damp sponge or soft cloth.
6. Wipe all incubator surfaces (internal), lids with OH Peroxide solution (3%).
7. Rinse with distilled water using a damp sponge or soft cloth.
8. Dry the incubator chamber out with a lint-free soft cloth.
9. Wrap the aluminum incubation blocks for sterilization (use steam sterilizer or oven only - DO NOT
USE ETYLHENE OXIDE METHOD for this purpose).
10. Once the sterilization of the aluminum incubation block is completed and the parts are at room
temperature open the wrapping and install it back in the drawers.
11. Plug in the incubator and turn the incubator on. Pour humidity-water to the water slot with water at
room temperature (or higher temperature) for humidification of the drawers.
12. Initiate a Mode. The incubator will start heating until it reaches the set temperature. The chamber will
increase its humidity gradually. Open the CO2 flow from the gas cylinder or manifold. The incubator
will not inject CO2 until the Set Up Mode cycle is completed.
13. After Set Up Mode is completed (12 hours) the incubator will initiate CO2 injection into the chamber
to a pre set 5%. Adjust the incubator to the desired setting if different than the factory setting (5%
CO2 and 37C).
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
28
22. Wiring Diagram
110V, 115V 50/60Hz
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
29
220 – 240V 50/60Hz
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
30
23. Maintenance
Panels on the top, backside and right side can be removed as show below for maintenance service.
Top panel Screws
Side panel Screws
Screws
Back panel
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
31
Descriptions of primary components
CPU Board Locations
Main Parts on the CPU Platform
*Note: Analog output is an optional feature.
CPU for the 1st Chamber
CPU for the 2nd Chamber
CPU for the 3rd Chamber
CPU for the 4th Chamber
Right Side
Platform
CPU Board
Step Down Transformer
Air Filter (For air intake)
N2 Valve
CO2 Valve
CO2 Sensor Board
*Analog Output Board
N2 Flow Volume Control
CO2 Flow Volume Control
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
32
Right Side
Top Side
In-Line Filter Step Down Transformer (AC200 – 240V only)
Switching Power Supply
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
33
Descriptions of Gas Sensors and Tubing
Heat Protected Box Replacement
� Remove 2 screws and gently pull out heat-protected box a few centimeters.
� Unplug the connectors and completely remove heat-protected box.
Main Parts in the Heat-Protected Box
Heat Protected Box
Connector to Heating Element (also one on the other side)
Back Side
Screw (also one on the other side)
IR CO2 Sensor O2 Sensor
Gas Control Box
Calibration Valve
Circulation Pump
Circulation (IN)/ N2 Supply Port
Circulation (OUT)
CO2 Supply Port
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
34
Descriptions of Gas Control Box
CO2 Sensor Replacement
� Remove gas control box from the incubator by loosening screws on each side.
� Unscrew 4 screws and pull out the CO2 sensor when you need to replace it.
O2 Sensor Replacement
� Remove gas control box from the incubator by loosening screws on each side.
� Pull out the O2 sensor from the gas control box. The O2 sensor can be replaced without removing
sensor holder and O-rings.
� Place a new O2 sensor to the original position. Make sure that sensor completely fits to the sensor
slot. Gas leakage may occur if the sensor is not properly positioned.
Screws (one on each corner)
IR CO2 Sensor
Rubber Gasket
O2 Sensor
O2 Sensor Holder
O Rings
Gas Control Box
Gas Control Box Bracket
Screw
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
35
Descriptions of Drawers
� Remove 4 screws located on the heating surface and remove the aluminum plate. Temperature sensor, heating element and magnet switch for incubation box can be replaced.
� Simply remove the screws that hold the magnet switch for drawer and incubation box when these switches need to be replaced.
� Use a hexagonal wrench to remove the hinge.
Heating Surface Temperature Sensor
Screws
(Beneath the heating surface)
Cables
Toothed Wheel
Toothed Wheel
Magnet Switch (Drawer)
Magnet Switch (Incubation Box)
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
36
Descriptions of Gas supply unit
Disconnect all the tubes to the Gas supply ports in the back of the unit (please see Descriptions of Gas
Sensors and Tubing), and remove the 2 screws holding the gas supply unit. Then pull out the whole
gas supply unit.
CO2 Port Circulation (OUT)
Circulation (IN) / N2 Port
Knobbed Nut
Silicon Gasket
Screws
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
37
Descriptions of Gas supply Flow
Check Valves
N2 Port Air Filter Flow Volume N2 Valve
IR CO2 Sensor
O2 Sensor
Gas Control Box
Three-way Valve
Circulation (IN)/ N2 Supply Port
Circulation (OUT)
CO2 Supply Port
Circulation Pump
CO2 Port Air Filter Flow Volume CO2 Valve
To the other Drawers
To the other Drawers
Air Filter Pump Ambient Air
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
38
23. Warranty Information
We warrant the product you have purchased for two years after the date of purchase. In case of any
malfunctions caused by the manufacturer during this period of time, ASTEC will be responsible for repair
or replacement of failed parts or even the instrument itself as circumstance allows for free of charge.
However, this warranty excludes the following conditions.
Any damages due to fire, earthquake, thunderstorm, and other catastrophic events and pollution and
abnormal electrical supply.
Any damages due to unduly repair, adjustment, calibration and modification.
Any damages due to mishandling.
Any damages caused by moving, dropping, transportation of the instrument.
Replacement and repair of expendables and consumables.
For service, please contact our agent that you have purchased the instrument from.
Warranty is Valid only if the installation is done by people trained, in conformity to
instructions provided in this manual.
Drawer Type Incubator Operation Manual Rev. 5 ASTEC CO., LTD.
39
Factory Settings
1st Drawer
A-1
A-3
b-2
ASTEC CO., Ltd
4-6-15, Minamizato, Shime, Kasuya, Japan 811-2207
Tel: +81-92-935-5585 Fax: +81-92-936-6613
Email: [email protected]
Website: http://www.astec-bio.com/global/
2nd Drawer
A-1
A-3
b-2
3rd Drawer
A-1
A-3
b-2
4th Drawer
A-1
A-3
b-2