SmallSmall--Field DosimetryField Dosimetry
Indra J. DasIndiana University School of Medicine, Indianapolis, IN, USA
Maria M. Aspradakis
IJD/AAPM-10
Maria M. AspradakisCantonal Hospital of Lucerne, Lucerne Switzerland
Otto A. SauerUniversity of Wurzburg, Germany
Anders AhnesjöUppsala University & Nucletron Scandinavia AB, 751 47 ,Uppsala, Sweden
-
Challenges in Small Field MV Photon Dosimetry
IJD/AAPM-10
Maria Mania Aspradakis
Cantonal Hospital of Lucerne Switzerland
-
Klinik und Poliklinik für StrahlentherapieDirektor: Prof. Dr. Dr. M. Flentje
Small Field DosimetryRelative dosimetry: detector choice and corrections
Otto A SauerOtto A Sauer
-
Modelling of small fields in Treatment Planning Systems
Anders AhnesjöAdj. Prof. Uppsala University, Sweden
Senior Research Scientist, Nucletron, Uppsala
-
Treatment Fields
Magna-FieldsTraditional Fields
40x40 cm2 4x4 cm2200x200 cm2
IJD/AAPM-10
Advance Therapy FieldsSRS/SRTGamma KnifeCyber-KnifeTomotherapyIMRT
4x4 cm2 0.3x0.3 cm2
Small Field
-
What is a Small Field?
� Lack of charged particle■ Dependent on the range of secondary
electrons
IJD/AAPM-10
■ Photon energy
� Collimator setting that obstructs the source size
� Detector size comparable to the field size
[ Detail discussion by Maria in this talk ]
-
CPE & Electron Range
� Electron range= dmax in forward direction
� Electron range in lateral direction
■ Nearly energy independent
� CPE, Charged Particle Equilibrium
IJD/AAPM-10
■ Nearly energy independent
■ Nearly equal to penumbra (8-10 mm)
� Field size needed for CPE■ Lateral range
■ 16-20 mm
dmax
[ Detail discussion by Anders on treatment planning in this talk ]
-
Definition of Small Fields
An elegant presentation of this issue by Anders in this talk
IJD/AAPM-10Das et al, Med. Phys. 35, 206-215, 2008
-
�� Ion ChambersIon Chambers■■ Diameter, volume, and window thicknessDiameter, volume, and window thickness■■ W/e, S/W/e, S/ρρ, , µµen/en/ρρ■■ PerturbationPerturbation
�� FilmsFilms■■ Silver bromide filmsSilver bromide films
●● high Z, sensitivity, thicknesshigh Z, sensitivity, thickness■■ GAFchromicGAFchromic polymer filmspolymer films
Dosimeters & Associated ProblemsDosimeters & Associated Problems
IJD/AAPM-10
■■ GAFchromicGAFchromic polymer filmspolymer films●● Sensitivity, thickness, polymerization at low energySensitivity, thickness, polymerization at low energy
�� TLD/OSL (sensitivity at low energy, F center)TLD/OSL (sensitivity at low energy, F center)■■ Size 10Size 10--15 15 µµm, difficulty of handlingm, difficulty of handling
�� Fricke Dosimeter (G value)Fricke Dosimeter (G value)■■ (Size and dose)(Size and dose)
�� MOSFET (thickness, dose rate, burnout, life, cost)MOSFET (thickness, dose rate, burnout, life, cost)�� Monte Carlo (benchmarking, expertise, reliability)Monte Carlo (benchmarking, expertise, reliability)
[ Detail discussion by Otto in this talk ]
-
6 MV; Central Axis
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
Rel
ativ
e d
ose
at d
max
Scanditronix-SFDScanditronix-PFDExradin-A16PTW-PinpointPTW-0.125ccPTW-0.3ccPTW-0.6ccPTW-MarkusWellhofer-IC4
15 MV; Central Axis
1.0
1.1
Field Size Limit for Accurate Dose Measurements with Available Detectors
IJD/AAPM-10
0.0
0 1 2 3 4 5 6 7 8 9 10 11
Field Size (cm)
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
0 1 2 3 4 5 6 7 8 9 10 11
Field Size (cm)
Rel
ativ
e d
ose
at d
max
Scanditronix-SFD
Scanditronix-PFDExradin-A16
PTW-PinpointPTW-0.125cc
PTW-0.3ccPTW-0.6cc
PTW-MarkusWellhofer-IC4
Das et al, TG-106, Med Phys, 35, 4186, 2008
-