    METTLER TOLEDO MultiRangeID1 Plus / ID1 Plus-A weighing terminals

    BildBild

    METTLER ID1 Plus

  • 1Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals Contents

    Contents Page

    1 Checklists ......................................................................................................................................... 2

    1.1 Maintenance checklist ......................................................................................................................... 2

    1.2 Service checklist ................................................................................................................................. 3

    2 Spare parts ....................................................................................................................................... 4

    2.1 ID1 Plus exploded view diagram, with interfaces .................................................................................... 4

    2.2 ID1 Plus-A exploded view diagram, analog connection ........................................................................... 5

    2.3 Spare parts list ................................................................................................................................... 6

    3 Error displays .................................................................................................................................... 8

    3.1 Operating errors .................................................................................................................................. 8

    3.2 Error messages .................................................................................................................................. 8

    4 Troubleshooting ................................................................................................................................ 9

    5 Corrective maintenance ..................................................................................................................... 10

    5.1 Opening the terminal ........................................................................................................................... 10

    5.2 Checking the voltages ......................................................................................................................... 10

    5.3 Removing the main board ................................................................................................................... 10

    5.4 Changing the microfuses ..................................................................................................................... 11

    5.5 Changing the program ........................................................................................................................ 11

    5.6 Changing the main board .................................................................................................................... 11

    5.7 Checking and changing the transformer, setting the power supply voltage ................................................. 12

    5.8 Changing the keypad .......................................................................................................................... 12

    5.9 Changing the analog board ................................................................................................................. 13

    5.10 Installing the main board ..................................................................................................................... 13

    5.11 Closing the terminal ............................................................................................................................ 13

    6 Settings ............................................................................................................................................ 14

    6.1 Configuring terminal and weighing platform ........................................................................................... 14

    6.2 Preparing the identcard (ID1 Plus only) ................................................................................................. 14

    7 Software compatibility ....................................................................................................................... 15

    8 Pin assignment ................................................................................................................................. 15

  • 2ID1 Plus / ID1 Plus-A weighing terminals

    Service manual 00705863D 08/02

    1 Checklists

    1.1 Maintenance checklist

    1.1.1 Visual inspection

    Check condition of the following scale components:

    Housing Keypad Protective hood Weighing platform Peripherals

    Check condition of the following cables:

    Power cable Weighing platform connection cable Data transfer cable (if used)

    Check for protected position of cables.

    1.1.2 Function check

    Press power switch.

    For verified scales, check the ID code (setting on ID card and display in the terminal must match after pressing the "Test key";see installation instructions or operating instructions, Installation section).

    Carry out calibration and display checks with the "Test key".

    Check functions by making entries via the keypad (see operating instructions).

    Check the following settings of the weighing platform (see service manual of the connected weighing platform):

    Calibration Cornerload Linearity Hysteresis

    Check the following plug-in connections for firm seating:

    Power cable Weighing platform connection cable Data transfer cable (if used) Peripheral connection (if installed)

    Checklists

  • 3Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals

    1.2 Service checklist

    Carry out the following check procedure at the terminal and the weighing platform before troubleshooting and after servicing.

    1.2.1 At the terminal

    Check whether an operating error is present.

    Check power and weighing platform connection cables.

    Check calibration via the "Test key".

    Check functions entries via keypad (see operating instructions).

    Carry out display, keypad and printer tests; for this purpose:

    Unplug terminal. Unplug weighing platform. Switch terminal on. PLUG IN appears in the display. Press SET TO ZERO key until SERVICE MODE appears in the display. Press Enter key until all display elements illuminate. Press all keys in sequence from left to right. If a GA46 printer is connected, a test printout is printed automatically after the last key is pressed. "END?" appears in the display. The test routine can be repeated by pressing the Tare key. Press the Enter key to return to weighing mode.

    Check all plug-in connections and program EPROM for firm seating.

    Check connected peripherals.

    1.2.2 At the weighing platform

    Check that transport locks of the weighing platform have been released properly and are in the correct position.

    Check supports of the weighing platform.

    Check that the load plate is free and has no contact with the surroundings.

    Check maximum capacity and linearity of the weighing platform.

    Check play of all stops and limits.

    Check that the lever system is free on all sides.

    Check knife edges, bearings and flexible bearings.

    Checklists

  • 4ID1 Plus / ID1 Plus-A weighing terminals

    Service manual 00705863D 08/02

    H

    G

    F

    C

    0

    T

    Enter

    F

    13

    16

    B

    L1

    N

    17

    14

    12

    13

    11

    10

    br.

    bl.

    8

    97

    5

    6

    6

    4

    3

    2

    1

    19

    18

    15

    H

    J

    E

    D

    C

    J

    A

    B

    N

    A

    ED

    L1

    G

    F

    27

    Spare parts

    Cable connectionsfor example cable A to connector A

    2 Spare parts

    2.1 ID1 Plus exploded view diagram, with interfaces

  • 5Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals

    2.2 ID1 Plus-A exploded view diagram, analog connection

    Spare parts

    26

    23

    25

    21

    20

    22

    24

    28

  • 6ID1 Plus / ID1 Plus-A weighing terminals

    Service manual 00705863D 08/02

    2.3 Spare parts list

    Spare parts

    Item Designation Part No. ID1 Plus Part No. ID1 Plus-A

    1 Cover, compl., with display window and keypad 00 505 326 00 505 326

    2 Keypad 00 203 585 00 203 585

    3 Identcard S, short 00 504 468

    4 Cover seal 00 203 594 00 203 594

    5 Power cable

    - Euro 00 504 247 00 504 247- GB 00 504 248 00 504 248- DK 00 504 249 00 504 249- CH 00 504 250 00 504 250- USA 00 504 251 00 504 251- ZA 00 504 252 00 504 252- AUS, RA 00 504 253 00 504 253

    6 Ring seal 00 203 654 00 203 654

    7 Leveling foot 00 504 186 00 504 186

    8 O-ring 16 x 1.25 00 203 656

    9 IDNet connector 00 203 598

    10 Transformer (set to 230 V) 00 203 587 00 203 587

    11 Insulating plate 00 203 655 00 203 655

    12 Spacer 00 504 470 00 504 470

    13 Main board ID1 Plus 00 203 578 00 203 578

    14 Program ID1 Plus 00 505 303 00 505 303

    15 Microfuse slow-blow 315 mA - L 250 V 00 087 431 00 087 431

    16 Microfuse slow-blow 1.25 A - L 250 V 00 087 805 00 087 805

    17 Microfuse slow-blow 2.5 A - H 250 V 00 089 922 00 089 922

    18 Interface 101 CL20 mA 00 505 237 00 505 237Interface 102 RS232 00 505 238 00 505 238Interface 104 RS422/485 00 506 847 00 506 847

    19 Interface 103 digital I/O 00 505 239 00 505 239

  • 7Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals

    Item Designation Part No. ID1 Plus Part No. ID1 Plus-A

    20 Analog board (point ADC board) without Eprom 00 507 000

    21 Point cable 00 507 003

    22 Minispreader 00 204 130

    23 Reducing ring 00 506 860

    24 Cable gland, heavy gauge 7 00 201 887

    25 Sealing sleeve 00 507 008

    26 Grub screw M3 x 6 DIN 913 00 205 958

    27 Verification set (for updating Identcard bracket) 22 000 386 22 000 386

    28 Eprom for analog board 00 507 274

    Set of labels 00 507 010

    Spare parts

  • 8ID1 Plus / ID1 Plus-A weighing terminals

    Service manual 00705863D 08/02

    3 Error displays

    3.1 Operating errors

    Incorrect operations at the terminal and operating conditions of the weighing platform under which a weight determinationis not admissible or not possible are displayed by the terminal in clear text, see operating instructions.

    3.2 Error messages

    If an operational fault is identified internally by the terminal, the faulty component is shown in the display.

    Error messages ID1 Plus / ID1 Plus-A terminals

    Error displays

    Error messages Meaning Testing Component

    ERR ID1 Error terminal Check all plug-in connec- EPROM main boardtions and EPROM for firm ID1 Plusseating.

    ERR I-F Error interface Check attached peripherals EPROM main boardand connections. ID1 PlusCheck EPROM.

    ERR BRU Error weighing platform Check connection cable of Measuring cellmeasuring cell.

    CAL ERR Calibration error Perform calibration again. Measuring cell

    Clearing error message: Switch terminal off and then on again.

    If error is redisplayed: Replace faulty component, see section 5, Corrective maintenance or service manual of attachedweighing platform.

  • 9Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals Troubleshooting

    4 Troubleshooting

    Fault

    Terminal can not be switched on

    After terminal switched on,- PLUG IN - appears

    After terminal switched on, noweight display

    No data transfer to measuring cell

    No entry via keypad possible

    Fault rectification

    Establish connection to powersupply

    Check power cable Check microfuses on main board

    and replace if necessary (see 5.4) Switch on display Change main board (see 5.6)

    Attach weighing platform

    Replace connection cable

    Check microfuses on main boardand replace if necessary (see 5.4)

    Replace measuring cell(see service manual of attachedweighing platform)

    Plug in weighing platform cable

    Change analog board (see 5.9) Change point cable

    Release transport safeguards(see operating instructions ofattached weighing platform)

    Adjust frame and lever system (seeservice manual of attached weighingplatform)

    Change main board (see 5.6) Change analog board (see 5.9) Change point cable

    Check microfuses on main boardand replace if necessary (see 5.4)

    Change main board (see 5.6) Replace measuring cell

    (see service manual of attachedweighing platform)

    Change analog board (see 5.9) Change point cable

    Replace keypad (see 5.8) Change main board (see 5.6) Change analog board (see 5.9) Change point cable

    Possible causes

    Power cable not plugged in

    No line voltage Fuse blown

    Display switched off Main board faulty

    Weighing platform not attached toterminal

    Connection cable of weighingplatform faulty

    Fuse blown

    Measuring cell faulty

    Ribbon cable for weighing platformconnection on main board (item Hin exploded view diagram) notattached

    Analog board faulty Point cable faulty

    Transport safeguards of weighingplatform not released

    Mechanical system touchingsomewhere

    Main board faulty Analog board faulty Point cable faulty

    Fuse blown

    Main board faulty Measuring cell faulty

    Analog board faulty Point cable faulty

    Keypad faulty Main board faulty Analog board faulty Point cable faulty

    Note

    See also operating instructions of ID1 Plus or ID1 Plus-A, section "What if...?".

  • 10

    ID1 Plus / ID1 Plus-A weighing terminals

    Service manual 00705863D 08/02

    Open terminal, see section 5.1.

    Measure voltages at ribbon cable on main board atbottom right.

    Pin Assignment Value

    1, 2 GND3 VDIS 47 VDC4 VNOR 16 VDC5 VCC 5 VDC6 VOPT27 VOPT18 Vheat29 Vheat110 VBA2 3.2 VDC

    If the voltages are not correct, check microfuses andtransformer, see sections 5.4 and 5.7.

    5 Corrective maintenance

    Caution

    Before opening the instrument:

    Disconnect power plug.

    Unplug attached weighing platform and peripherals.

    5.1 Opening the terminal

    Unscrew 2 screws on underside of the cover at the front and lift off cover.

    Caution

    Do not scratch coating on the inside of the display window, if need be simply wipe carefully using a soft cloth.

    Unplug keypad cable (J) at main board.

    5.2 Checking the voltages

    5.3 Removing the main board

    Open terminal, see section 5.1.

    Unplug weighing platform cable (H) on right of main board next to the display.

    Unscrew heavy gauge screw fitting and push in power cable approx. 5 cm.

    Unscrew 2 screws on main board and lift main board out of guide.

    Unplug interface cable (A).

    Unplug 6 connectors on underside and lift out board.

    10 1STPBRC1

    Corrective maintenance

    }}

    12 VAC

    7 VAC

  • 11Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals

    5.4 Changing the microfuses

    Remove main board, see section 5.3.

    Check microfuses and replace if necessary.

    Fuse ratings

    SI 1 slow blow 2.5 A - H 250 V

    SI 2 slow-blow 1.25 A - L 250 V

    SI 3 slow-blow 315 mA - L 250 V

    SI 5 slow-blow 315 mA - L 250 V

    Reinstall main board, see section 5.10.

    SI3

    SI5

    SI2

    SI1

    CautionThe battery must not be changed as improper handling can lead to an explosion hazard.Disposal of the board with battery only by the manufacturer!

    Corrective maintenance

    5.5 Changing the program

    Open terminal, see section 5.1.

    Remove old EPROM program from socket using IC extractor.

    Insert new EPROM program in the IC socket paying heed to the marking on the IC and the socket. Carefully bend IC contactsstraight.

    Close terminal, see section 5.11.

    5.6 Changing the main board

    Remove main board, see section 5.3.

    Remove 5 spacers; insulating plate, top and bottom are separate parts.

    Break off new main board into top and bottom.

    Mount insulating plate between top and bottom with the 5 spacers.

    Pull off program from old main board socket using IC extractor and insert in the IC socket of the new main board paying heedto the marking on the IC and the socket. Carefully bend IC contacts straight.

    Reinstall main board, see section 5.10.

  • 12

    ID1 Plus / ID1 Plus-A weighing terminals

    Service manual 00705863D 08/02

    P7

    P5

    P3

    S1S2S7S9S10S8S4S5S6

    5.7 Checking and changing the transformer, setting the power supply voltage

    Checking the transformer

    Change main board, see section 5.3.

    Check voltages on secondary side.

    Pin Rated Voltage at Open-circuitcurrent rated current voltage

    S1, S2 0.95 A 12 V 5 % 13.8 VS4, S5 0.225 A 3.2 V 5 % 4.0 VS5, S6 0.225 A 3.2 V 5 % 4.0 VS7, S9 0.13 A 32 V 5 % 36.5 VS8, S10 0.17 A 10 V 5 % 11.6 V

    If the tolerances are not complied with, change trans-former.

    Changing the transformer, setting the power supplyvoltage

    Remove transformer.

    With new transformer, check connections on primary side:

    for 230 V connections 3 and 7 (factory setting)

    for 115 V connections 3 and 5

    To switch to 115 V, carefully undo and lift off cover,unsolder and resolder connection accordingly.

    Install transformer.

    Reinstall main board, see section 5.10.

    5.8 Changing the keypad

    Open terminal, see section 5.1.

    Pull off affixed foilsheet keypad from display window.

    Remove residual adhesive and dirt from adhesion surface of the display window.

    Pull protective film off new keypad.

    Lead connection cable through the cutout in the display window.

    Affix keypad and smooth down firmly.

    Close terminal, see section 5.11.

    Corrective maintenance

    S1

    S2

    S9

    S10

    250

    V

    115

    V

    6 5 4 3 2 17

  • 13Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals

    2 1

    5.9 Changing the analog board

    Remove main board, see section 5.3.

    Disconnect weighing platform cable and point cable.

    Undo nut and screw and lift out analog board.

    Disengage minispreader from old analog board and lockin place on new board.

    Plug in analog board and install with screw and nut.

    Plug in weighing platform cable (1) and point cable (2).

    Reinstall main board, see section 5.10.

    5.10 Installing the main board

    ST9ST4

    ST3ST6

    ST5

    ST7

    Plug 6 connectors into bottom part of main board.

    Plug weighing platform connector and interface connec-tor, if used, into top part of main board.

    Route power cable outward and tighten heavy gaugescrew fitting.

    Insert main board in the guide and fix with 2 screws.

    Close terminal, see section 5.11.

    Pin assignment

    ST 3 L1 (brown)ST 4 N (blue)ST 5ST 6 transformerST 7ST 9 shield bonding

    Corrective maintenance

    5.11 Closing the terminal

    Caution

    If necessary, wipe display window on the inside very carefully using a soft cloth!

    Plug in keypad cable.

    Engage cover at top in screw heads and fasten at front using 2 screws on underside of cover.

    }

  • 14

    ID1 Plus / ID1 Plus-A weighing terminals

    Service manual 00705863D 08/02

    6 Settings

    6.1 Configuring terminal and weighing platform

    6.1.1 ID1 Plus-A

    All other information on the ID1 Plus-A terminal is contained in the operating instructions for the ID1 Plus-A.

    6.1.2 ID1 Plus

    For the ID1 Plus terminal, the service mode settings are described in the service manual for the service mode of the connectedweighing platform.

    3

    1

    p

    p

    p p

    p p p

    p p p p p

    p p

    p p p

    p p

    p p

    p p

    p p

    p

    p p

    p p

    p

    p

    p p

    p p p

    p p p p p

    p p

    p p p

    p p

    p p

    p p

    p p

    p

    p p

    p p

    1

    2

    1

    Settings

    Putting into operation after changing the weighingplatform configuration

    Remove identcard.

    Affix appropriate measurement data plate (1).

    Adjust identcode disk (2) so that it matches the displayedidentcode.

    Assemble identcard.

    With certified scales, seal screw (3).

    6.2.3 Updating from old identcard (bracket) to new identcard (socket)

    Therefore you need the calibration set 22 000 386.The procedure is described in the installation information for the identcard.

    = YES = NO

    Service mode operation

    Program blocks are selected the same way in service modeas in master mode.

    Entering service mode

    Press and hold the Enter key until MASTER appears in the display.

    Press the NO key; SERVICE appears in the display.

    Press the YES key to enter service mode; the CODE prompt appears.

    Enter the following code: YES NO NO YES. The first service mode block, RETURN, appears.

    6.2 Preparing the identcard (ID1 Plus only)

    6.2.1 New identcard (socket)

    In the factory, the identcard is provided with the measurement data plate of the standard configuration of the weighing platform,the identcode shown is "1".For mounting the identcard, see installation information for the identcard.

    6.2.2 Old identcard (bracket)

  • 15Service manual 00705863D 08/02

    ID1 Plus / ID1 Plus-A weighing terminals Software compatibility / Pin assignment

    7 Software compatibility

    The ID1 Plus weighing terminal is compatible with all weighing platforms of the F, K, N and D series.

    With the ID1 Plus-A terminal: see ID1 Plus-A operating instructions, section "Technical data".

    Front view

    8 Pin assignment

    Weighing platform connection

    A TXD+, transmitting loop of the weighing platform

    B VDIS 40 V

    C VNOR 13 V

    D RXD+, receiving loop of the weighing platform

    F RXD, receiving loop of the weighing platform

    G Protective ground

    H Signal ground

    J TXD, transmitting loop of the weighing platformThe pin assignment of the ID1 Plus-A is described in the ID1Plus-A operating instructions.

    AB

    C

    D

    E F G

    H

    J

    K

    L

    M

  • 00705863D