Mechanical specifications R¹ R² scWheels: 4 with break Height (cm) 192 192 190
Width (cm) 70 70 71,5
Depth (cm) 91 91 94
Warmer weight with accessories (kg) 60 60 100
R¹ R² scBED NB bed height (since floor) 100 100 100
Bed weight capacity 10 10 10
Width (cm) 32 42 61
Depth (cm) 61 70 72
Electric height adjustment Electric height adjustment SC – Foldable acrylic bassinet walls
R1 – Removable crib
R2 – Foldable acrylic bassinet walls
Heater rotation (180º)
R1 – Removable crib with optional re-verse Phototherapy Model Medphoto 6
Voltage RequiReMentsVoltage automatic selection
100 -140V~
205 -245V~
Current 5,0A 3,0A
Network frequency
50//60Hz
Power consumption
1400 VA
Heating element
Quartz-coated electric
resistancePower: 580
Watts
LampHalogen lamp with dichroic
reflector – 12V/20W
Temperature control (Automatic mode)
From 25ºC – 38°C, with
0.1°C increments
Power heating control (manual mode)
0 – 100%, with 10%
increments
Temperature Indication
From 20ºC – 45°C, with ±
0,1°C accuracy
Maximum admissible error
± 0,2°C ± 0,3°C
iso13458
european authorizedRepresentative center
obelis s.a.www.obelis.net
0470
infant Radiant waRMeRs Model MatRix
NEONATAL
RODM
IDIA
The
info
rmat
ion
cont
aine
d he
rein
is s
ubje
ct to
alte
ratio
ns w
ithou
t pre
viou
s no
tice.
Further informations: www.olidef.com.br – +55 (16) 3919.9350 – [email protected] – Ribeirão Preto/SP – Brazil
technical especificationsControl unit MicroprocessadoTemperature adjustment range 25,0ºC - 38,0ºCSkin temperature measurement 20,0ºC - 45,0ºCHeat power (10% increments) 0 - 100%Display resolution 0,1ºCThermostatic precision 0,2ºCBrazilian ANVISA registration number 10227180023Standards NBR IEC 60.601-1
NBR IEC 60.601-1-2NBR IEC 60.601-2-21
Voltage 110/220V automático
geneRal chaRacteRistic R¹ R² scRadiant warmer system Radiant warmer systemHeater rotation (180º): access for X-rays devices
contRol systeMs R¹ R² scSafety electronic systemManual heater modeSkin mode (servo–controlled)
contRol Module R¹ R² scAlphanumeric LCD Skin temperature probeDigital power indicatorApgar timerRechargeable internal batteryAlarme silence buttonKeyboard blockSimultaneous temperature display¹Standby indicator LED
alaRMsPower failureManual mode warning each 10 min.
High skin temperature + 1ºC
High temperature safety 15 min.¹
Low skin temperature - 1ºC
Heating failureSensor failureSystem failureHeater out of positionSensor disconnected from the patient skinSensor disconnectAlarm silence 10 min.
¹ Manual mode: power adjust proportional heater indicator and Apgar timer Automatic mode: temperature adjust, skin temperature, proportional heater indication/Apgar timer
¹Available jars capacity: 2 L; 2,5 L; 3 L; 5 L – ²With flowmeter, suction unit with vacuum gauge and humidifier – ³ With flowmeter, suction unit and humidifier
BaBy´s coMpaRtMent R¹ R² scRemovable crib Foldable acrylic bassinet wallsX-rays transparent bedX-ray cassete trayMattress (impermeable/non-allergic/non-toxic)
acessóRios R¹ R² scVentilator tube support Temperature sensor coverTransport handlesCabinetWeighing scale (external display)Instrument trayIntermediate tray (45 x 60 cm)Autoclavable mattress cover Oxygen hood (three sizes)Cylindres/cylinders holderPhotoherapy model Maxiphoto halogenReverse phototherapy model Medphoto 6Collection jar with lid¹Large drawer (up to 2 units)Drawer set (up to 2 units)X-rays transparent bedBumpersManual resuscitatorRuler for installation of accessoriesTriple ruler AR²Triple ruler O2²Triple ruler O2³Régua tripla mista – Ar – O2 – Vacuômetro Continous bed tilting systemElectrical height adjustment Hydraulic bed tilting systemIV poleHigh adjustable IV poleMonitor trayPower outlets (4 or 8)
¹ in the maximum power
R1: with removable acrylic crib R2: with flat bed and foldable acrylic walls Standard Optional accessory N/A – Not available
linha neonatal
Newborns prematures babies have greater difficulties in maintaining their body temperature than full-term babies and adults because of their large body surface in relations to their volume. The use of the thermal therapy has been demonstrated as being very useful to increase the survival rate in low-weight babies. Several devices have been developed to actively warm and maintain appropriate temperature in neonates and infants, such as radiant warmers and hot air incubators. The neonatal infant radiant warmers models Matrix R e SC, are recommended for delivery rooms, nurseries and intensive care units and provide a warm and stable microenvironment for small babies, who need additional care in this critical period in the beginning of their lives. These are high-technology equipment whose resources may include control systems of temperature, along with factors such as quality, effectiveness and user – friendly operation, all specially designed to fully meet the user´s expectations, therefore providing greater comfort for the small patients.
Infant radiant warmers model Matrix
R2 - with flat bed and foldable acrylic walls