influence on early osseointegration of dental implants installed with two different drilling...
TRANSCRIPT
Influence on early osseointegration ofdental implants installed with twodifferent drilling protocols: ahistomorphometric study in rabbit
Juan BlancoElena AlvarezFernando MunozAntonio LinaresAntonio Cantalapiedra
Authors’ affiliations:Juan Blanco, Elena Alvarez, Antonio Linares,Department of Stomatology – Periodontology, School ofDentistry, University of Santiago de Compostela,Santiago de Compostela, SpainFernando Munoz, Antonio Cantalapiedra, School ofVeterinary of Lugo, University of Santiago deCompostela, Santiago de Compostela, Spain.
Corresponding author:Juan BlancoDepartment of Stomatology – PeriodontologySchool of DentistryUniversity of Santiago de Compostelac/Entrerrıos s/n 15702Santiago de CompostelaSpainTel.: þ 34 981 571 826Fax: þ 34 981 571 620e-mail: [email protected]
Key words: bone implant contact, early healing, primary stability, resonance frequency analysis
Abstract
Objective: To evaluate early osseointegration of dental implants installed with two different drilling
protocols.
Material and methods: Thirty-six cylindrical shape Mozo Grau implants, with a diameter of 3.75 and
11 mm long, were placed into the distal condyle (submerged) of each femur of 18 New Zealand
rabbits. In the control group, a 3.3 mm diameter drill was used as the last one prior implant installation
(standard protocol). In the test group, the same procedure was carried out but an additional 3.5 mm
drill was used as the final one (oversized protocol) Thus, we could obtain different primary stability at
day 0 between groups. Sacrifice of the animals was after 2, 4 and 8 weeks. Histomorphometric analysis
(bone-to-implant contact ratio [BIC%]) and implant stability quotient (ISQ) values (Ostells
) were
registered at each sacrifice time.
Results: The ISQ values were statistically significant different between groups at day 0 (control: 69.65;
test: 64.81); and after 2 weeks (control: 77.93; test: 74). However, after 4 and 8 weeks the results were
similar. BIC% showed a similar tendency, with 58.69% for the control group and 40.94% for the test group
after 2 weeks, this difference being statistically significant. At 4- and 8-week interval, BIC% was similar.
Conclusion: At 2-week interval (early healing), osseointegration had been influenced by different
primary stability at implant installation, being slower in the oversized protocol (lower primary stability),
which could be especially risky in challenging clinical situations, such as soft bone (class 3 and 4) and early/
immediate loading. However, from 4 week on, these differences disappeared. Nevertheless, we have to
consider that a direct transfer of the results of this animal study (time bone repair mechanisms) into clinic
has to be done with caution.
The term ‘‘osseointegration’’ refers to the process
whereby alloplastic materials (dental implants)
and bone are joined in a rigid, clinically asympto-
matic union that withstands functional loading
(Zarb & Albrektsson 1991). This is a definition
based on the mechanical stability of the implant.
In order to understand the healing process, it is
necessary to know the biological mechanisms
that play a role in the integration of the implant
with the surrounding bone.
Dental implant primary stability has been
demonstrated to be a key factor for implant
survival rates. Primary mechanical stability is
directly related to the quality and quantity of
bone at the recipient site, the type of implant
used and the surgical technique used to place the
implant (Meredith 1998). Biologically, this pri-
mary stability is obtained if the marginal and/or
apical areas of the implant site hold a large enough
quantity of compact bone and if the spongy bone
contains a sufficient number of trabeculae
(Albrektsson Berglundh & Lindhe 2003). Clini-
cally, at the time of implant placement, this kind
of primary stability is achieved by ‘‘tight fitting’’
between the implant surface and the avascular
cortical bone in the marginal area of the implant
bed. This intimate bone-to-implant contact (BIC)
is also the effect of the minute lateral displace-
ment exerted in the bone tissue during implant
adaptation, where the trabeculae of the marginal
portion shift towards the medullar space and the
sectioned blood vessels bleed. As a consequence,
a blood clot forms and is trapped between the
implant surface and the bone. This blood clot will
mature over the next few days and eventually be
replaced by granulation tissue (Albrektsson et al.
1983), woven bone and lamellar bone, producing
secondary stability. Therefore, secondary stability
is the consequence of the formation of new bone
and the remodelling process in both the area of
most direct contact (the bone–implant interface
itself) and a more distant area (Meredith 1998).
Date:Accepted 12 June 2010
To cite this article:Blanco J, Alvarez E, Munoz F, Linares A, Cantalapiedra A.Influence on early osseointegration of dental implantsinstalled with two different drilling protocols: ahistomorphometric study in rabbitClin. Oral Impl. Res. 22, 2011; 92–99.doi: 10.1111/j.1600-0501.2010.02009.x
92 c� 2010 John Wiley & Sons A/S
A non-invasive intraoral method for evaluating
implant stability has been developed in recent
years. Meredith et al. (1994) described a new
clinical non-invasive approach that consisted in
evaluating bone anchorage around an implant by
measuring the resonance frequency of a transdu-
cer coupled to the implant (Meredith et al. 1994;
Meredith et al. 1996; Meredith et al. 1997a,
1997b). This new technique was named resonance
frequency analysis (RFA). Numerous in vitro and
in vivo studies have corroborated RFA as a tech-
nique designed to reflect the bone/implant inter-
face and hence may be useful in documenting
clinical implant stability (Meredith et al. 1996;
Meredith 1998; Zix et al. 2008). However, no
correlations between histological parameters of
osseointegration and implant stability quotient
(ISQ values) could be identified by other authors
(Huwiler et al. 2007; Abrahamsson et al. 2009).
When trying to achieve primary stability in soft
bone (class 3 and 4), clinicians have attempted to
improve the stability by using osteotome for
condensation or undersized drills before implant
installation (Bahat 1992, 1993; Summers 1994,
1995; Blanco et al. 2008). However, sometimes it
is not possible to achieve ideal primary stability,
due to a bad control of the drilling and/or wrong
drilling sequence during surgery, resulting in an
oversized bone preparation.
The objective of this study was to evaluate the
early osseointegration of dental implants in-
stalled with two different drilling protocols (stan-
dard and oversized) in a soft bone animal model.
Materials and methods
Once approval from the Ethics Committee of the
University of Santiago de Compostela had been
granted, this research was carried out using 18
New Zealand white rabbits weighing an average
of 5.5 kg. All animals were installed in the
animal experimentation service facility at the
Veterinary Hospital Rof Codina of Lugo (Spain).
All experiments were performed according to the
Spanish Government Guide and the European
Guide for Animal Care. The experimentation site
was located on both distal condyles of the femurs.
The implants used in this study were MG
Osseous implants (Mozo Graus
, Valladolid,
Spain) made of commercially pure grade-IV tita-
nium, featuring 2.5 mm of machined surface
(1 mm in the platform and 1.5 mm of threads in
the endo-osseous portion), followed by 8.5 mm of
surface treated with resorbable blast media (RBM
blasted with calcium phosphate ceramics) all the
way to the apical portion of the implant. The
implants selected were 3.75 mm (core: 3.3 mm)
in diameter/11 mm long, and cylindrical down to
the seventh thread, after which it adopts a conical
component, with a self-tapping feature closer to
the apex.
Surgical procedure
In accordance with the described standards for
surgery on experimental animals, the surgery was
performed in the surgical area of the Rof Codina
Veterinary Clinical Hospital, at the University of
Santiago de Compostela, School of Veterinary
Medicine at Lugo. All surgeries were performed
by the same operator (J. B.). The surgical ap-
proach occurred under general anaesthesia and
was under the supervision of the veterinary sur-
geon at all times.
An incision was made at the distal femoral
condyle with a scalpel fitted with a number 15
blade, following a continuous line. Subsequently,
the skin, the subcutaneous tissue and the muscle
were drawn back to gain access to the bone.
Implant bed preparations were carried out accord-
ing to the recommendations of the manufacturer
(Mozo Graus
). Study groups were formed in
relation to the last drill used before implant
installation. In one condyle (control group), the
last drill was 3.3 mm in diameter (standard pro-
tocol) and in the other condyle (test group) the
last drill was 3.5 mm in diameter (oversized
protocol). Lastly, the countersink drill was used,
aiming to avoid the potential effect on primary
stability of the cortical bone plate. Once the
implant bed was prepared, the implants were
inserted and the resonance frequency was imme-
diately measured in both groups. Thus, using a
split design, in one condyle, one implant was
placed in the control group, and in the other
condyle, one implant was placed in the test
group. Finally, the incision was closed in layers
with reabsorbable suture (Vicryls
, Ethicon, Som-
erville, NJ, USA), leaving the implants sub-
merged. A total of 36 implants were placed,
two per rabbit (Fig. 1a–e).
After surgery, the rabbits were moved to the
laboratory animal house of the University of
Santiago de Compostela’s Lugo Campus and
put into individual hutches, with the following
conditions: temperature of 22 � 21C and a rela-
tive humidity of 50–70%.
Histological analysis and resonancefrequency evaluation
The animals in the study were sacrificed at 2, 4
and 8 weeks after implant installation, so that
samples could be obtained and prepared for his-
tological analysis. After being euthanized, flaps
were raised to obtain access to the submerged
implants in order to measure the ISQ values. The
implants were separated from each femur using a
diamond saw (Exact 300CLs Apparatebeau, Nor-
destedt, Hamburg, Germany). The biopsies were
processed for ground sectioning in conformity
with the Donath method (Donath 1995). The
samples were dehydrated and infiltrated with
resin (Technovit 7200s, VLC-Heraus Kulzer
GmbH, Werheim, Germany). Finally, the sam-
ples were sectioned using a grinding technique
(Exact 400CSs Apparatebeau, Hamburg, Ger-
many) up to approximately 20mm using the
Levai Laczko staining method.
Histometric evaluation
For the histometric analysis, the samples were
processed with an Olympuss
DP12 image digi-
talization unit that was coupled to an Olympuss
CH30 microscope and an Olympuss
SZX9 stereo
microscope (Olympuss
DF PLAPO 1 � �2 lens,
Tokyo, Japan). Using Olympus MicroImage soft-
ware, version 4.0 for Windows, the points of
interest on the digitalized images of the histologi-
cal samples were identified.
One examiner, who was blinded to treatment
allocation of the specimens, identified the follow-
ing linear measurements in each section with a
magnification of � 40 of the original one.
BIC
It is defined as the length of the bone surface in
direct contact with the implant starting from the
shoulder of the implant.
Implant perimeter (IP)
This starts from the implant shoulder to the last
visible thread, disregarding the unthreaded apical
portion (Nkenke et al. 2003) (Fig. 2).
BIC ratio (BIC%)
Defined as BIC/IP (� 100 [%]).
The primary variables were BIC% and ISQ
values at different time intervals.
Statistical analysis
Descriptive statistics were produced for each one
of the variable and groups (mean values, standard
deviation, median and scatter plots). Confidence
intervals for the differences of the means with a
confidence level of 95% were presented. The
Wilcoxon test for non-parametric paired data
was applied to both the clinical results (ISQ values
obtained from the measurements made with the
Osstell Mentors
, Integration diagnostic, Goteborg,
Sweden) and the histological results (BIC%). The
probability level of Po0.05 was considered as the
level of statistical significance.
All statistical analysis was performed with
SPSS 15.0 for Windows (SPSS Inc., Chicago, IL,
USA).
Blanco et al �Healing of implants installed with two drilling protocols
c� 2010 John Wiley & Sons A/S 93 | Clin. Oral Impl. Res. 22, 2011 / 92–99
Results
Two of the animals were sacrificed prematurely
due to femoral fracture induced by an infectious
process. Therefore, the final sample was com-
prised of 16 rabbits and 32 implants. Five rabbits
were sacrificed after 2 weeks, five after 4 weeks
and six after 8 weeks.
ISQ values (Osstell Mentors
)
The ISQ values of the resonance frequency regis-
tered with the Osstell Mentors
are shown in
Table 1 and Fig. 3. At implant installation, the
ISQ for the control group was 69.65 � 6.17, and
64.81 � 6.39 for the test group. The difference in
ISQ values was statistically significant. This
shows that the control implants had a higher
primary stability than the test ones.
After 2 weeks, there was a significant differ-
ence between test and control for the ISQ values.
The control showed an ISQ value of 77.93 �3.74 and the test 74 � 3.94.
However, after 4 and 8 weeks of healing, the
difference ceased to be significant. Nevertheless,
the ISQ values increased over time in both groups
(Fig. 3).
Histomorphometric results (BIC%)
The results of the percentage of BIC at each time
interval are displayed in Table 2 and Fig. 4. The
percentage of BIC registered for the rabbits sacri-
ficed after 2 weeks (Figs 5 and 6) displayed
statistically significant differences between the
test group (3.5 mm) and the control group
(3.3 mm). The values found were 40.94 � 7.77
and 58.69 � 10.67, respectively. However, no
significant differences were found in the other
two sacrifice groups, after 4 (Figs 7 and 8) and 8
weeks (Figs 9 and 10).
Discussion
The main determinants of implant stability are
the mechanical properties of the bone tissue at
the implant site and the degree of implant en-
gagement with that bone tissue. The mechanical
properties of bone are determined by the compo-
sition of the bone at the implant site and may
increase during healing because soft trabecular
bone tends to undergo a transformation into
dense cortical bone at the vicinity of the implant
surface. The strength of the implant–bone inter-
face is also influenced by surgical technique and
implant design and surface. For instance, the use
of a narrower final drill or a wider or tapered
implant will force more of the implant threads
into contact with the surrounding bone (Sen-
nerby & Meredith 2008).
The objective of this experimental study was to
evaluate whether the primary stability achieved
at the time of implant placement with two
different drilling protocols, influences the speed
of implant osseointegration. The results we
found show a tendency towards increased BIC
over time in both groups. Primary stability was
observed to have a significant positive influence
on the early osseointegration process (2 weeks);
Fig. 1. (a) View of the surgical area at the distal condyle of the femur of the rabbit. (b) The implant installed with the transporter. (c) The implant without the transporter. (d) Implant stability
quotient value registered with the Ostell Mentors
system. (e) Radiographic view of the implant after installation.
Fig. 2. The contained zone represents the measured area of
bone implant contact, from the implant shoulder to the last
visible thread in the histological section. Levai Laczko
staining method. Original magnification � 12.5.
Table 1. Implant stability quotient (ISQ) values
ISQ values Control (3.3) Test (3.5) 95% CI Wilcoxon test
Mean (SD) Median Mean (SD) Median
Day 0 69.65 (6.17) 72 64.81 (6.39) 64 1.05 to 8.63 P¼ 0.018n
Week 2 77.93 (3.74) 77 74 (3.94) 73 1.08 to 6.91 P¼ 0.042n
Week 4 81.65 (3.15) 81 78.78 (4.83) 77 � 2.71 to 8.31 P¼ 0.225Week 8 82.79 (5.07) 83.25 82.33 (5.7) 85.75 � 3.99 to 4.82 P¼ 1
nP-valueo0.05: statistical significance.
CI, confidence interval for the differences of the means; SD, standard deviation.
Blanco et al �Healing of implants installed with two drilling protocols
94 | Clin. Oral Impl. Res. 22, 2011 / 92–99 c� 2010 John Wiley & Sons A/S
osseointegration was faster in the control group
(greater primary stability) than in the test group.
Early clinical work indicated a relationship
between bone density and primary implant sta-
bility. Friberg et al. (1999) correlated cutting
resistance (i.e. bone density) with primary stabi-
lity for maxillary implants. Follow-up measure-
ments performed at the time of abutment
connection (6–8 months later) and after 1 year
in function indicated that all implants, irrespec-
tive of initial stability, tended to reach a similar
level of stability. These facts are in accordance
with what we have observed in our study from 4
weeks on, where the implants in both groups
attained the same stability levels.
Andersson et al. (2008) examined 102 Neoss
implants and found an inverse relationship be-
tween cutting torque (bone density) and changes
in implant stability during a study period of 12
months . They also identified a correlation be-
tween bone hardness, measured according to
Lekholm & Zarb (1985), and primary stability.
Implants in soft bone with low primary stability
showed a marked increase in stability compared
with implants in dense bone. We found this
tendency in our study in both groups (soft bone
model), which is similar to the results of a study
in rabbits finding that resonance frequency in-
creases with time as a function of an increased
stiffness resulting from new bone formation and
remodelling (Meredith et al. 1997b).
Some studies have failed to show a correlation
between the degree of BIC and RFA measure-
ments (Meredith et al. 1997b; Akca et al. 2006;
Ito et al. 2007). This may have to do with the
Fig. 3. Implant stability quotient values scatter plot for control (3.3) and test (3.5) treatment.
Table 2. Bone-to-implant (BIC) contact percentage
BIC% Control (3.3) Test (3.5) 95% CI Wilcoxon test
Mean (SD) Median Mean (SD) Median
Week 2 (N¼ 5) 58.69 (10.67) 56.34 40.94 (7.77) 40.91 6.89 to 28.58 P¼0.043n
Week 4 (N¼ 5) 49.97 (10.16) 46.87 55.43 (6.18) 56.3 � 22.92 to 12.01 P¼0.686Week 8 (N¼ 6) 60.74 (10.03) 58.48 57.32 (7.8) 50.76 � 13 to 19.84 P¼0.753
nP-valueo0.05: statistical significant.
SD, standard deviation; CI, confidence interval for the differences of the means.
Fig. 4. Bone implant contact percentage scatter plot for control (3.3) and test (3.5) treatment.
Fig. 5. (a, b) Control group. These images represent the
2-week interval. Levai Laczko staining method. Original
magnification � 12.5, and the close up images � 40.
Blanco et al �Healing of implants installed with two drilling protocols
c� 2010 John Wiley & Sons A/S 95 | Clin. Oral Impl. Res. 22, 2011 / 92–99
nature of the test, because the degree of bone
contact does not necessarily reflect the stiffness
of the surrounding bone. In modern implant
dentistry using moderately rough implants, the
surface is often covered by a thin layer of bone,
which is probably not important for the biome-
chanical support of implants. The study con-
ducted by Abrahamsson et al. (2009) to
evaluate the relationship between BIC and ISQ
values during a 12-week healing period in the
animal model did not find any correlation be-
tween the two parameters, either; the results
were in line with those found by Huwiler et al.
(2007), Schliephake et al. (2006) and Ito et al.
(2008). However, we did find a positive
correlation between the increase in ISQ values
and BIC.
We found significant differences in BIC be-
tween the two groups after 2 weeks; these differ-
ences may be of major importance in terms of
loading time. Thus, when there is suitable pri-
mary stability (ISQ values), loading could be
carried out early and even immediately. Ostman
et al. (2005, 2008) reported low failure rates when
using an implant stability quotient of 60 as an
inclusion criterion for immediate loaded implants
in totally edentulous maxillae and in posterior
mandibles. Sennerby & Meredith (1998) found
the RFA technique to be helpful in deciding when
to replace an immediately loaded temporary
prosthesis with a permanent prosthesis after im-
plant placement.
The introduction of new implant surfaces may
let us reduce treatment time protocols, because it
may be able to diminish the time required for
Fig. 6. (a, b) Test group. These images represent the 2-week
interval. Levai Laczko staining method. Original magnifica-
tion � 12.5, and the close up images � 40.Fig. 7. (a, b) Control group. These images represent the
4-week interval. Levai Laczko staining method. Original
magnification � 12.5, and the close up images � 40.
Fig. 8. (a, b) Test group. These images represent the 4-week
interval. Levai Laczko staining method. Original magnifica-
tion � 12.5, and the close up images � 40.
Blanco et al �Healing of implants installed with two drilling protocols
96 | Clin. Oral Impl. Res. 22, 2011 / 92–99 c� 2010 John Wiley & Sons A/S
osseointegration. The evidence from histomor-
phometric data and clinical studies suggests that
rough implant surfaces exert a clinically signifi-
cant influence on osseointegration. Berglundh
et al. (2003) have carried out a histological study
investigating the sequential healing events asso-
ciated with the placement of SLA (sand blasted,
large grit, acid etched) surface titanium implants.
This study confirms previous reports (Cochran
et al. 1998; Schenk & Buser 1998) that initial
bone formation around rough titanium implants
occurs not only to the exposed bone wall of the
surgically created implant bed but also along the
osteoconductive implant surface. Furthermore, it
was demonstrated that there was a higher level of
organization in the wound and higher BIC during
the early healing (from 2 h) associated with the
SLA surface compared with the machined sur-
faces (Abrahamsson et al. 2004). In addition, a
modification of the SLA implant surface has been
introduced (SLActive). This new surface allows
for immediate cell reaction right after implant
placement and accelerates bone remodelling dur-
ing the osseointegration process. Pre-clinical re-
sults have shown that SLActive provides 60%
more bone integration after 2 weeks compared
with the SLA surface (Buser et al. 2004). Results
from a human clinical study measuring implant
stability with the Osstell device demonstrated a
statistically significant improvement in stability
with SLActive implants over those with SLA
during the critical early treatment period between
weeks 2 and 4 (Oates et al. 2007). It has been
recently shown that implants with a SLActive
surface would promote complete bone regeneration
of acute dehiscence defects in dogs within a period
of 12 weeks (Schwarz et al. 2007). Surface treat-
ments with a calcium phosphate coating promote
early bone healing as well as subsequent bone
apposition, generating rapid biological fixation
of the implant to the bone (Le Guehennec et al.
2007; Meirelles et al. 2008). The implant analysed
in this paper presents a surface blasted with cal-
cium phosphate ceramics. Such a surface may
influence the speedy early healing obtained in
this study.
How well an implant is fixated to the sur-
rounding bone tissue is dependent on different
implant and host-related factors. Six factors,
listed by Albrektsson et al. (1981) have gained a
general acceptance as being especially important:
biocompatibility, design, surface quality, status
of host tissue, surgical technique and loading
conditions. Therefore, the development of new
implant designs (tapered), surfaces and clinical
techniques has enabled a marked reduction of the
initial healing period, even to the point of an
immediate/early loading of implants that show
high primary stability (Attard & Zarb 2005;
Ostman 2008). However, in a recently published
randomized-controlled clinical trial, no differ-
ences between standard cylindrical and tapered
implants were noted in terms of primary stabi-
lity, yielding clinically equivalent short-term
outcomes (Lang et al. 2007). Thus, the success
of immediate/early loading implant techniques is
dependent on the ability of the clinician to
achieve and to determine, the degree of primary
implant stability and changes in stability along
with new bone formation and remodelling. In
addition, the results of our study reflect the
importance of careful management of the surgical
technique in terms of both the selection of the
drilling protocol and/or gentle drilling.
Conclusion
The most important finding of our study is that,
with the tapered implant design utilized here, the
speed of osseointegration after 2-week interval
(early healing) has been significantly influenced
by primary stability achieved at implant installa-
tion time, being slower in the oversized protocol
(lower primary stability), which could be espe-
cially risky in challenging clinical situations,
such as soft bone (class 3 and 4) and early/
immediate loading. However, from 4 weeks on,
these differences disappeared. Nevertheless, we
have to consider that a direct transfer of the
Fig. 9. (a, b) Control group. These images represent the 8-
week interval. Levai Laczko staining method. Original
magnification � 12.5, and the close up images � 40.
Fig. 10. (a, b) Test group. These images represent the 8-week
interval. Levai Laczko staining method. Original magnifica-
tion � 12.5, and the close up images � 40.
Blanco et al �Healing of implants installed with two drilling protocols
c� 2010 John Wiley & Sons A/S 97 | Clin. Oral Impl. Res. 22, 2011 / 92–99
results of this animal study (time bone repair
mechanisms) into clinic is not recommended.
Acknowledgements: This
investigation was supported by the Mozo Grau
Company. The authors declare that they have
no conflict of interest.
References
Abrahamsson, I., Berglundh, T., Linder, E., Lang, N.P.
& Lindhe, J. (2004) Early bone formation adjacent to
rough and turned endosseous implant surfaces. An
experimental study in the dog. Clinical Oral Im-
plants Research 15: 381–392.
Abrahamsson, I., Linder, E. & Lang, N.P. (2009)
Implant stability in relation to osseointegration:
an animal experimental study in the Labrador dog.
Clinical Oral Implants Research 20: 313–318.
Akca, K., Chang, T.L., Tekdemir, I. & Fanuscu, M.I.
(2006) Biomechanical aspects of initial intraosseous
stability and implant design: a quantitative micro-
morphometric analysis. Clinical Oral Implants Re-
search 17: 465–472.
Albrektsson, T., Berglundh, T. & Lindhe, J. (2003)
Osseointegration: historic background and current
concepts. In: Lindhe, J., Karring, T. & Lang, N.,
eds. Clinical Periodontology and Implant Dentistry,
809–820. Oxford, UK: Balckwell Munksgaard.
Albrektsson, T., Branemark, P.I., Hansson, H.A., Ka-
semo, B., Larsson, K., Lundstrom, I., McQueen, D. &
Skalak, R. (1983) The interface zone of inorganic
implants in vivo: titanium implants in bone. Annals
of Biomedical Engineering 11: 1–27.
Albrektsson, T., Branemark, P.I., Hansson, H.A. &
Lindstrom, J. (1981) Osseointegrated titanium im-
plants. Acta Orthopaedica Scandinavica 52: 155–
170.
Andersson, P., Verrocchi, D., Viinamaki, R. & Sen-
nerby, L. (2008) A one-year clinical,radiographic and
RFA study of Neoss implants. Applied Osseointegra-
tion Research 6: 23–26.
Attard, N.J. & Zarb, G.A. (2005) Immediate and early
implant loading protocols: a literature review of
clinical studies. Journal of Prosthetic Dentistry 94:
242–258.
Bahat, O. (1992) Osseointegrated implants in the max-
illary tuberosity: report on 45 consecutive patients.
The International Journal of Oral & Maxillofacial
Implants 7: 459–67.
Bahat, O. (1993) Treatment planning and placement of
implants in the posterior maxillae: report of 732
consecutive Nobelpharma implants. The Interna-
tional Journal of Oral & Maxillofacial Implants 8:
151–61.
Berglundh, T., Abrahamsson, I., Lang, N.P. & Lindhe,
J. (2003) De novo alveolar bone formation adjacent to
endosseous implants. Clinical Oral Implants Re-
search 14: 251–262.
Blanco, J., Suarez, J., Novio, S., Villaverde, G., Ramos,
I. & Segade, L. (2008) Histomorphometric assess-
ment in human cadavers of the peri-implant bone
density in maxillary tuberosity following implant
placement using osteotome and conventional
techniques. Clinical Oral Implants Research 19:
505–510.
Buser, D., Broggini, N., Wieland, M., Schenk, R.K.,
Denzer, A.J., Cochran, D.L., Hoffmann, B., Lussi, A.
& Steinemann, S.G. (2004) Enhanced bone apposition
to a chemically modified SLA titanium surface.
Journal of Dental Research 83: 529–533.
Cochran, D.L., Schenk, R.K., Lussi, A., Higginbottom,
F.L. & Buser, D. (1998) Bone response to unloaded
and loaded titanium implants with a sandblasted and
acid-etched surface: a histometric study in the canine
mandible. Journal of Biomedical Materials Research
40: 1–11.
Donath, K. (1995) Preparation of histological sections.
Norderstedt: EXAKT – Kulzer Publication.
Friberg, B., Sennerby, L., Meredith, N. & Lekholm, U.
(1999) A comparison between cutting torque and
resonance frequency measurements of maxillary im-
plants. A 20-month clinical study. The International
Journal of Oral and Maxillofacial Surgery 28:
297–303.
Huwiler, M.A., Pjetursson, B.E., Bosshardt, D.D., Salvi,
G.E. & Lang, N.P. (2007) Resonance frequency ana-
lysis in relation to jawbone characteristics and during
early healing of implant installation. Clinical Oral
Implants Research 18: 275–280.
Ito, Y., Sato, D., Yoneda, S., Ito, D., Kondo, H. &
Kasugai, S. (2007) Relevance of resonance frequency
analysis to evaluate dental animal experiments. Clin-
ical Oral Implants Research 18: 1–6.
Ito, Y., Sato, D., Yoneda, S., Ito, D., Kondo, H. &
Kasugai, S. (2008) Relevance of resonance frequency
analysis to evaluate dental implant stability: simula-
tion and histomorphometrical animal experiments.
Clinical Oral Implants Research 19: 9–14.
Lang, N.P., Tonetti, M.S., Suvan, J.E., Bernard, J.P.,
Botticelli, D., Fourmousis, I., Hallund, M., Jung, R.,
Laurell, L., Salvi, G.E., Shafer, D. & Weber, H.-P.
(2007) Immediate implant placement with transmu-
cosal healing in areas of aesthetic priority: a multi-
centre randomized-controlled clinical trial I. Surgical
outcomes. Clinical Oral Implants Research 18:
188–196.
Le Guehennec, L., Soueidan, A., Layrolle, P. &
Amouriq, Y. (2007) Surface treatments of titanium
dental implants for rapid osseointegration. Dental
Materials 7: 844–54.
Lekholm, U. & Zarb, G.A. (1985) Patient selection and
preparation. In: Branemark, P.I., Zarb, G.A. &
Albrektsson, T., eds. Tissue-Integrated Prostheses:
Osseointegration in Clinical Dentistry, 199–209.
Chicago: Quintessence Publishing.
Meirelles, L., Currie, F., Jacobsson, M., Albrektsson, T.
& Wennerberg, A. (2008) The effect of Chemical and
nanotopographical modifications on the early stages
of osseointegration. The International Journal of Oral
& Maxillofacial Implants 23: 641–47.
Meredith, N. (1998) Assessment of implant stability as
a prognostic determinant. International Journal of
Prosthodontics 11: 491–501.
Meredith, N., Alleyne, D. & Cawley, P. (1996) Quan-
titative determination of the stability of the implant-
tissue interface using resonance frequency analysis.
Clinical Oral Implants Research 7: 261–267.
Meredith, N., Book, K., Friberg, B., Jemt, T. & Sen-
nerby, L. (1997a) Resonance frequency measure-
ments of implants stability in vivo. Clinical Oral
Implants Research 8: 226–233.
Meredith, N., Cawley, P. & Alleyne, D. (1994)
The application of modal vibration analysis to study
bone healing in vivo. Journal of Dental Research
73: 793.
Meredith, N., Shagaldi, F., Alleyne, D. & Sennerby, L.
(1997b) Resonance frequency measurements of im-
plant stability of titanium implants during healing in
the rabbit tibia. Clinical Oral Implants Research 8:
234–243.
Nkenke, E., Lehner, B., Weinzierl, K., Thams, U.,
Neugubauer, J., Steveling, H., Radespiel-Troger, M.
& Neukam, F.W. (2003) Bone contact, growth, and
density around immediately loaded implants in the
mandible of mini pigs. Clinical Oral Implants Re-
search 14: 312–321.
Oates, T.W., Valderrama, P., Bischof, M., Nedir, R.,
Jones, A., Simpson, J., Toutenburg, H. & Cochran,
D.L. (2007) Enhanced implant stability with a che-
mically modified SLA surface: a randomized pilot
study. The International Journal of Oral & Maxillo-
facial Implants 22: 755–760.
Ostman, P.O., Hellman, M. & Sennerby, L. (2005)
Direct implant loading in the edentulous maxilla
using a bone density-adapted surgical protocol and
primary implant stability criteria for inclusion. Clin-
ical Implant Dentistry and Related Research 7:
60–69.
Ostman, P.O., Hellman, M. & Sennerby, L. (2008)
Immediate occlusal loading of implants in the par-
tially edentate mandible: a prospective 1-year radio-
graphic and 4-year clinical study. The International
Journal of Oral & Maxillofacial Implants 23:
315–22.
Ostman, P.O. (2008) Immediate/early loading of dental
implants. Clinical documentation and presentation
of a treatment concept. Periodontology 2000 47:
90–112.
Schenk, R.K. & Buser, D. (1998) Osseointegration: a
reality. Periodontology 2000 17: 22–35.
Schliephake, H., Sewing, A. & Aref, A. (2006)
Resonance frequency measurements of implant sta-
bility in the dog mandible: experimental comparison
with histomorphometric data. The International
Journal of Oral and Maxillofacial Surgery 35:
941–946.
Schwarz, F., Herten, M., Sager, M., Wieland, M., Dard,
M. & Becker, J. (2007) Bone regeneration in dehis-
cence-type defects at chemically modified (SLActive)
and conventional SLA titanium implants: a pilot
study in dogs. Journal of Clinical Periodontology
34: 78–86.
Sennerby, L. & Meredith, N. (1998) Resonance fre-
quency analysis: measuring implant stability and
osseointegration. Compendium of Continuing Edu-
cation in Dentistry 19: 500–502.
Sennerby, L. & Meredith, N. (2008) Implant stability
measurements using resonance frequency analysis:
biological and biomechanical aspects and clinical
implications. Periodontology 2000 47: 51–66.
Summers, R.B. (1994) A new concept in maxillary
implant surgery: the osteotome technique. Compen-
Blanco et al �Healing of implants installed with two drilling protocols
98 | Clin. Oral Impl. Res. 22, 2011 / 92–99 c� 2010 John Wiley & Sons A/S
dium of Continuing Education in Dentistry 15: 152–
158.
Summers, R.B. (1995) The osteotome technique:
part4 – future site development. Compendium
of Continuing Education in Dentistry 16: 1080–
1092.
Zarb, G.A. & Albrektsson, T. (1991) Osseointegration –
a requiem for the periodontal ligament? Editorial.
International Journal of Periodontics and Restorative
Dentistry 11: 88–91.
Zix, J., Kessler-Liechti, G. & Mericske-Stern, R. (2008)
Measurement of dental implant stability by resonance
frequency analysis and damping capacity assessment:
comparison of both techniques in a clinical trial. The
International Journal of Oral & Maxillofacial Im-
plants 23: 525–530.
Blanco et al �Healing of implants installed with two drilling protocols
c� 2010 John Wiley & Sons A/S 99 | Clin. Oral Impl. Res. 22, 2011 / 92–99