Abstract Methods for LC-MS peptide analysis oſten use acidified mobile phases, which can alter retention and selectivity. Here, we demonstrate the effects of modifying acid type, acid concentration, temperature, and gradient slope using several test probes. A Raptor ARC-18 col- umn was employed for this work because the stationary phase ligand is sterically protected, making it extremely stable and resistant to acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis. Introduction With the influx of biotherapeutics in medical research and healthcare, LC- MS analysis of peptides and small proteins continues to grow. Many of these analyses utilize acid-modified mobile phases to improve peak shape [1]; however, the effects of acidic mobile phases on selectivity and retention are oſten not well understood. In this article, we will explore the effects of acid type, acid concentration, temperature, and gradient slope on the selectivity and retention of several peptide probes and one small protein probe (cytochrome C) using the Raptor ARC-18 LC column. is column was selected because the ligand is sterically protected from acid damage and is stable to pH 1.0. In addition, it contains superficially porous particles (SPP), which allow for faster analysis times than can be achieved with fully porous particle (FPP) columns. Experimental Several experiments were carried out to demonstrate the effects of acid- modified mobile phases, temperature, and gradient slope on LC-MS peptide analysis. For this work, the peptide and small protein standards in Table I and a bovine serum albumin (BSA) trypsin digest standard (Protea) were diluted in acidified water. ese standards were injected into a Waters ACQUITY UPLC H-Class system configured with a PDA and QDa for simultaneous UV and MS detection. A wavelength of 215 nm and electrospray ionization in positive ion mode was used for UV and MS detection, respectively. Reversed-phase separations were performed using water and acetonitrile mobile phases modified with 0.01, 0.05, and 0.1% formic acid (FA) or trifluoroacetic acid (TFA) under gradient conditions using a Restek Raptor ARC-18 LC column. Specific conditions for each experiment are presented in Figures 2-5. Influencing Selectivity in LC-MS Peptides Analysis on the Raptor ARC-18 Column By Sharon Lupo and Frances Carroll Figure 1: Raptor ARC-18 LC Column Properties. Si O Stationary Phase Category: C18, octadecylsilane (L1) Ligand Type: Sterically protected C18 Particle: 2.7 μm or 5 μm superficially porous silica (SPP or “core-shell”) Pore Size: 90 Å Surface Area: 150 m 2 /g (2.7 μm) or 100 m 2 /g (5 μm) Recommended Usage: pH Range: 1.0–8.0 Maximum Temperature: 80 °C Maximum Pressure: 600 bar / 8,700 psi (2.7 μm) or 400 bar / 5,800 psi (5 μm) Properties: • Well-balanced retention profile. • Sterically protected to resist harsh, low-pH mobile phases. • Ideal for use with sensitive detectors like mass spec.

Upload: others

Post on 28-May-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Influencing Selectivity in LC-MS Peptides Analysis on the ... · acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis. Introduction

AbstractMethods for LC-MS peptide analysis often use acidified mobile phases, which can alter retention and selectivity. Here, we demonstrate the effects of modifying acid type, acid concentration, temperature, and gradient slope using several test probes. A Raptor ARC-18 col-umn was employed for this work because the stationary phase ligand is sterically protected, making it extremely stable and resistant to acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis.

IntroductionWith the influx of biotherapeutics in medical research and healthcare, LC-MS analysis of peptides and small proteins continues to grow. Many of these analyses utilize acid-modified mobile phases to improve peak shape [1]; however, the effects of acidic mobile phases on selectivity and retention are often not well understood. In this article, we will explore the effects of acid type, acid concentration, temperature, and gradient slope on the selectivity and retention of several peptide probes and one small protein probe (cytochrome C) using the Raptor ARC-18 LC column. This column was selected because the ligand is sterically protected from acid damage and is stable to pH 1.0. In addition, it contains superficially porous particles (SPP), which allow for faster analysis times than can be achieved with fully porous particle (FPP) columns.

ExperimentalSeveral experiments were carried out to demonstrate the effects of acid-modified mobile phases, temperature, and gradient slope on LC-MS peptide analysis. For this work, the peptide and small protein standards in Table I and a bovine serum albumin (BSA) trypsin digest standard (Protea) were diluted in acidified water. These standards were injected into a Waters ACQUITY UPLC H-Class system configured with a PDA and QDa for simultaneous UV and MS detection. A wavelength of 215 nm and electrospray ionization in positive ion mode was used for UV and MS detection, respectively. Reversed-phase separations were performed using water and acetonitrile mobile phases modified with 0.01, 0.05, and 0.1% formic acid (FA) or trifluoroacetic acid (TFA) under gradient conditions using a Restek Raptor ARC-18 LC column. Specific conditions for each experiment are presented in Figures 2-5.

Influencing Selectivity in LC-MS Peptides Analysis on the Raptor ARC-18 Column

By Sharon Lupo and Frances Carroll

Figure 1: Raptor ARC-18 LC Column Properties.

SiO

Stationary Phase Category:C18, octadecylsilane (L1)

Ligand Type: Sterically protected C18

Particle: 2.7 μm or 5 μm superficially porous silica (SPP or “core-shell”)

Pore Size: 90 Å

Surface Area: 150 m2/g (2.7 μm) or 100 m2/g (5 μm)

Recommended Usage:pH Range: 1.0–8.0

Maximum Temperature: 80 °C

Maximum Pressure: 600 bar / 8,700 psi (2.7 μm) or 400 bar / 5,800 psi (5 μm)

Properties:• Well-balanced retention profile.

• Sterically protected to resist harsh, low-pH mobile phases.

• Ideal for use with sensitive detectors like mass spec.

Page 2: Influencing Selectivity in LC-MS Peptides Analysis on the ... · acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis. Introduction

www.restek.com 2

Results and DiscussionSeveral probes were used to aid in the understanding of selectivity changes that affect LC-MS peptide analyses when the type and concentration of the acidic modifier in the mobile phase is manipulated. The retention of peptides containing basic amino acids (e.g., arginine) exhibited considerably large changes in retention when switching between formic acid and TFA. In Figure 2, the peptide MET-ARG-PHE-ALA not only changed elution order but also displayed improved peak shape and was more retained due to the increased ion pairing capabilities of TFA over formic acid. Alternatively, VAL-TYR-VAL is a polar peptide with an overall neutral charge whose changes in selectivity result from the favorable decrease in mobile phase pH when TFA is used; in this case, the peptide was less ionized and therefore better retained. Variations in acid concentration had a comparable effect. In Figure 3, overall retention times increased with increasing concentrations of TFA, and improvements in peak shape were observed for the small protein, cytochrome C (12.3 kDa). For this reason, sterically protected columns that are stable at low pH, such as the Raptor ARC-18 column, are often preferred for LC-MS peptide analysis.

Table I: Probes used for LC-MS Peptide Analyses

Peak ID Compound Molecular Weight (Da) SIR (m/z)

1 GLY-TYR 238 239.3

2 VAL-TYR-VAL 380 380.5

3 MET-ARG-PHE-ALA 524 524.6

4 Methionine enkephalin 574 574.6

5 Leucine enkephalin 556 556.7

6 Angiotensin II 1,046 524.4

7 Neurotensin 1,700 837.9

8 Cytochrome C 12,300 1,020.6

Figure 2: Effects of Acid Type on Selectivity and Peak Shape.

Time (min)0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

1

2

3

45

67

8

1

2

34

56

7

8

7.50 8.00 8.50 9.00 9.50 10.00

0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00

0.1% FA

0.1% TFA

LC_BA0353

Column Raptor ARC-18 (cat.# 9314A12)Dimensions: 100 mm x 2.1 mm IDParticle Size: 2.7 µmPore Size: 90 ÅTemp.: 60 °CSample Custom peptide/small protein standard mixDiluent: 0.1% Formic acid in waterConc.: 0.2–0.4 mg/mLInj. Vol.: 2 µL

Mobile PhaseA: Water + acidB: Acetonitrile + acid Time (min) Flow (mL/min) %A %B 0.00 1 95 5 1.00 1 95 5 10.0 1 50 50 10.1 1 95 5 13.0 1 95 5

Detector MSIon Mode: ESI+Mode: SRMInstrument UHPLC

Page 3: Influencing Selectivity in LC-MS Peptides Analysis on the ... · acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis. Introduction

www.restek.com3

Figure 3: Effects of Acid Concentration on Retention and Selectivity.

LC_GN0566

Column Raptor ARC-18 (cat.# 9314A12)Dimensions: 100 mm x 2.1 mm IDParticle Size: 2.7 µmPore Size: 90 ÅTemp.: 60 °CSample Custom peptide/small protein standard mixDiluent: Water + 0.1% formic acidConc.: 0.2–0.4 mg/mLInj. Vol.: 2 µLMobile Phase A: Water + TFAB: Acetonitrile + TFA Time (min) Flow (mL/min) %A %B 0.00 1 95 5 1.00 1 95 5 10.0 1 50 50 10.1 1 95 5 13.0 1 95 5 Detector MSIon Mode: ESI+Mode: MS scanInstrument UHPLC

0.500.00 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00

0.500.00 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00

0.500.00 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00

1

2,34

6

57

8

12

3

4

6

5

8

7

1

2

3

45

67

8

0.01% TFA

0.05% TFA

0.1% TFA

Time (min)

Page 4: Influencing Selectivity in LC-MS Peptides Analysis on the ... · acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis. Introduction

www.restek.com 4

Figure 4: Effects of Temperature on the Selectivity of Peptides in a BSA Tryptic Digest Standard.

In a second set of experiments, resolution and elution order changes were examined when temperature and gradient modifications were made to the analysis of a bovine serum albumin (BSA) trypsin digest standard. In the resulting peptide maps shown in Figure 4, base peak mass 468 m/z was resolved from an adjacent peak when a lower column oven temperature (40 °C) was used. Similarly, the gradient slope was shown to impact the band spacing of closely eluting compounds. This is demonstrated in Figure 5 where the number of resolved peaks increased with decreasing gradient slope. In general, steeper gradients can be used for the separation of specific peptides in order to achieve short cycle times. However, if numerous peptides from a digest are to be analyzed, a shallower gradient is typically preferred.

LC_GN0567

396

40 °C

386 46

846

2

524

656

464

436

396

386

462

523

656

465

436

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00Time (min)

60 °C

Column Raptor ARC-18 (cat.# 9314A12)Dimensions: 100 mm x 2.1 mm IDParticle Size: 2.7 µmPore Size: 90 ÅSample Bovine serum albumin (BSA) tryptic digest standardDiluent: Water + 0.1% formic acidConc.: 5 pmolInj. Vol.: 2 µLMobile Phase A: Water + 0.1% formic acidB: Acetonitrile + 0.1% formic acid Time (min) Flow (mL/min) %A %B 0.00 0.3 98 2 40.0 0.3 75 25 40.1 0.3 98 2 50.0 0.3 98 2Detector MSIon Mode: ESI+Mode: MS scanInstrument UHPLC

Page 5: Influencing Selectivity in LC-MS Peptides Analysis on the ... · acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis. Introduction

www.restek.com5

Figure 5: Effects of Gradient Steepness on the Selectivity of Peptides in a BSA Tryptic Digest Standard.

535 53

5424

561

656

363

331

422

577

354 33

336

733

148

957

5 424

301

384

722

313

395

462

464

455

506

548

628

425

331

422

577

354 33

3 410

368

331

330

489

424

689

300

384

722

396

386 46

846

253

552

453

5 677

561

656

464

436

506 45

554

8

628

301

331

422

561

354 33

341

036

733

1 330

489

425

689

301

345

384

723

395

313

386

462

468

535

535

524

535

561

678

465

656

563

436

535 50

6 455

547

628

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00 36.00 38.00 40.00 42.00 44.00 46.00 48.00

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00

1.00 2.00 3.00 4.00 15.00 16.00 17.00 18.00 19.00 20.005.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00

Time (min)

20 min GradientSlope = 1.15%

40 min GradientSlope = 0.575%

60 min GradientSlope = 0.383%

LC_GN0568

Column Raptor ARC-18 (cat.# 9314A12)Dimensions: 100 mm x 2.1 mm IDParticle Size: 2.7 µmPore Size: 90 ÅTemp.: 40 °CSample Bovine serum albumin (BSA) tryptic digest standardDiluent: Water + 0.1% formic acidConc.: 5 pmolInj. Vol.: 2 µL

Mobile Phase A: Water + 0.1% formic acidB: Acetonitrile + 0.1% formic acid Flow (mL/min) %A %B 0.3 98 2 0.3 75 25 0.3 98 2 0.3 98 2Detector MSIon Mode: ESI+Mode: MS scanInstrument UHPLC

Page 6: Influencing Selectivity in LC-MS Peptides Analysis on the ... · acid damage at low pHs, which is an important characteristic when developing LC-MS methods for peptide analysis. Introduction

ConclusionRelatively small changes in chromatographic parameters can have a large impact on LC-MS peptide analysis. In this work, we demonstrated that significant shifts in selectivity and retention can occur in response to relatively small changes to mobile phase acid (type and concentration), temperature, and gradient slope. Sterically protected columns like the Raptor ARC-18 column are stable at low pH and provide the versatility required for the development of peptide applications.

References[1] E. Wang, D. Kalu Appulage, F. Carroll, K. Schug, Intact protein separation and quantitation in biological fluids using HPLC and triple quadrupole mass spectrometry, ASMS, St. Louis, MO, 2015.

Questions? Contact us or your local Restek representative (www.restek.com/contact-us).Restek patents and trademarks are the property of Restek Corporation. (See www.restek.com/Patents-Trademarks for full list.) Other trademarks in Restek literature or on its website are the property of their respective owners. Restek registered trademarks are registered in the U.S. and may also be registered in other countries. To unsubscribe from future Restek communications or to update your preferences, visit www.restek.com/subscribe To update your status with an authorized Restek distributor or instrument channel partner, please contact them directly.