informatics driven quality initiatives - using the...

30
1 ©2011 MFMER | slide-1 William Pavlicek PhD AAPM, Charlotte, NC Aug 2, 2012 2:50PM Informatics driven Quality Initiatives - Using the Tools of Six Sigma ©2011 MFMER | slide-2 Outline of Talk Six Sigma Concepts PET Patient Dose reduction CT Patient Dose reduction – positioning and Z axis coverage Fluoroscopy – Lowering Patient Skin Dose Questions ©2011 MFMER | slide-3 W. Edwards Deming Taught engineering, physics in the 1920s, finished PhD in 1928 Met Walter Shewhart at Western Electric Long career in government statistics, USDA, Bureau of the Census 3 W. Edwards Deming, 1900 – 1993

Upload: vuongminh

Post on 09-Sep-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

1

©2011 MFMER | slide-1

William Pavlicek PhD

AAPM, Charlotte, NC

Aug 2, 2012 2:50PM

Informatics driven Quality Initiatives -Using the Tools of Six Sigma

©2011 MFMER | slide-2

Outline of Talk

• Six Sigma Concepts

• PET Patient Dose reduction

• CT Patient Dose reduction – positioning and Z axis coverage

• Fluoroscopy – Lowering Patient Skin Dose

• Questions

©2011 MFMER | slide-3

W. Edwards Deming

• Taught engineering, physics in the 1920s, finished PhD in 1928

• Met Walter Shewhart at Western Electric

• Long career in government statistics, USDA, Bureau of the Census

3

W. Edwards Deming, 1900 – 1993

2

©2011 MFMER | slide-4

W. Edwards Deming

• Deming was asked by Japan to lecture on statistical quality control to management

• Japanese adopted many aspects of Deming’s management philosophy

• Deming stressed “continual never-ending improvement”

• “If I had to reduce my message for management to just a few words, I’d say it all had to do with reducing variation”. (Deming, as quoted in The Deming Dimension, Henry R. Neave (SPC Press, 1990), p. 57)

• Deming lectured widely in North America during the 1980s

4

©2011 MFMER | slide-5

Deming energized “Statistical Thinking”

• Integral aspect of Six-Sigma (3.4 per million)

• Six Sigma is a ‘goal’ not often achieved!

• Fundamentally different from Statistical Methods

• The definition is broadened to include:

• Operations Research tools

• Discrete event “Simulation’

• Other elements of ‘lean process’

5*ASQ Statistics Division, http://www.asqstatdiv.org/stats-everywhere.htm

©2011 MFMER | slide-6

Defects Per Million Opportunities

3

©2011 MFMER | slide-7

* From Montgomery, D. C. (2009), Introduction to Statistical Quality Control 6th edition, Wiley, New York

How Variation Hurts …

7

©2011 MFMER | slide-8

The Motorola Six-Sigma Concept

• The Motorola Six-Sigma concept is to reduce the variability in the process so that the specification limits are six standard deviations from the mean, or only about 2 parts per billion defective (Fig 1-11a).

• When the Six-Sigma concept was initially developed, an assumption was made that when the process reached the Six-Sigma quality level, the process mean was still subject to disturbances that could cause it to shift by as much as 1.5 standard deviations off target (Fig 1-11b).

• There is an apparent inconsistency in this. Process performance isn’t predictable unless the process behavior is stable.

• No process or system is ever truly stable, and even in the best of situations, disturbances occur. The concept of a Six-Sigma process is one way to model this behavior. Like all models, it’s probably not exactly right, but it has proven to be a useful way to think about process performance, and has direct application to manufacturing.* 8* From Montgomery, D. C. (2009), Introduction to Statistical Quality Control 6th

edition, Wiley, New York. Figure 1-12.

©2011 MFMER | slide-9

The Generations of Six Sigma

• Generation I

• Focus on defect elimination

• Motorola, 1987-1993

• Generation II

• Focus on cost reduction

• GE, Allied Signal/Honeywell, 1994-1999

• Generation III

• Focus on value creation

• DuPont, 2000-present

9

4

©2011 MFMER | slide-10

Process Variation

• Variation is present in all processes and every aspect of the workplace

• Excess variation reduces process performance, decreases customer satisfaction, and has a negative impact on the bottom line

• Applies to all aspects of a business

• Statistical methods, an integral component of six sigma, are useful in shifting the process target to the desired level and reducing the variation around this target

10

©2011 MFMER | slide-11

As has been said, “the mean never happens,” — a heroic 4-day delivery time on one

order, with an awful 20-day delay on another, and no real consistency. The customers

in this chart feel nothing. Their life experience hasn’t changed; one bit. These

customers hear the sounds of celebration coming from within GE and ask, “What’s the

big event; what did we miss?” The customer only feels the variance that we have not yet removed. … Variation is evil in any customer-touching process. *

*Jack Welch, General Electric Company 1998 Annual Report11

©2011 MFMER | slide-12

Patients feel the variation!

• Too long Waiting for their exam

• Lack of proper prep means waiting, poss rescheduled exam

• Repeated exam due to xyz

• Multiple x-ray exams due to lack of available priors

• Poss need of follow up exam due to confidence – due to IQ?

5

©2011 MFMER | slide-13

• Patient goes for Out Patient Radiology exam: order, transport, waiting room, exam prep, tech/device performance, x-ray room availability, image quality, interpretation, communication to clinician.

• The patient experience has 10 components - is 99% okay?

Why “Quality Improvement” is Important:

©2011 MFMER | slide-14

Propagation of Error-repeatability of the exam-reproduced with diff techs/equipment/rads

©2011 MFMER | slide-15

Repeatability (i.e. back to back Ca Scores, bone density, kVp 4X, etc)

6

©2011 MFMER | slide-16

Reproducibility: Different techs, devices, patients…Different Quality?

©2011 MFMER | slide-17

CT fluoro biopsy dose, breast compression, fluoroscopy dose by physician, etc.

©2011 MFMER | slide-18

Focus of Six Sigma is on Process Improvement with an Emphasis on Achieving Significant Impact

• A process is an organized sequence of activities that produces an output that adds value to the organization

• All work is performed in (interconnected) processes

• Easy to see in some situations (manufacturing)

• Harder in others

• Any process can be improved

• An organized approach to improvement is necessary

• The process focus is essential to applying Six-Sigma methods

18

7

©2011 MFMER | slide-19

Organized Approach:

DMAIC (duy–may–ick)

©2011 MFMER | slide-2020

‘D’efine the Goal – What is important!(you are what you measure!)

� Project Charter – Admin Directive (like an internal contract with yourself)

� Problem Statement

� Goal Statement

� In/Out of Scope

� Team & Time Commitments

� Timeline / Milestone

� Estimate Financial Benefits

� Risks, Constraints & Compliance Issues Identified

©2011 MFMER | slide-21

The Primary Six Sigma Tools

• Process map

• Cause and effect analysis

• Measurement systems analysis*

• Capability study*

• Failure mode and effects analysis

• Observational study (regression)*

• Designed experiments*

• Control charts and out-of-control-action-plans*

* Statistical Tools 21

8

©2011 MFMER | slide-22

Use of Reference ValuesMeasure, Analyze and Improve!

©2011 MFMER | slide-23

General Comments

• One doesn’t really achieve Six Sigma

• Informatics with control charts are enormously helpful

• Shining a light by sharing data is very useful!

• Change can be difficult but fun!

• Leadership is key: drive out fear!

©2011 MFMER | slide-24

9

©2011 MFMER | slide-25

Joint Commission SE #47

©2011 MFMER | slide-26

How to determine if doses are ‘out of range’ (reproducible)?

©2011 MFMER | slide-27

Lots of Software

Solutions!

10

©2011 MFMER | slide-28

System Architecture

Knowledge Base

ExamInformation Database

Dosimetry Analyzer

Emails

WAN

Dose Tracking System

DICOM, HL7 Gateway

DICOM DICOM SR HL7

Texts

Reporting & Alerting Module

RIMS

©2011 MFMER | slide-29©2011 MFMER | slide-29

#1. PET-CT Six Sigma driven lower dose

©2011 MFMER | slide-30©2011 MFMER | slide-30

PET Scanner ‘Blocks’ BGO - PMT

11

©2011 MFMER | slide-31©2011 MFMER | slide-31

PET F-18

γγγγ

γγγγ

Coincidence ? Register EventYes

©2011 MFMER | slide-32©2011 MFMER | slide-32

Types of Coincident Events

True Scatter Randoms

Gamma ray (Singles)

Line of Response (LOR)

Prompts = True + Scatter + Randoms

©2011 MFMER | slide-33©2011 MFMER | slide-33

2D PET – SEPTA!All events used to reconstruct a slice

are detected within that slice.

12

©2011 MFMER | slide-34©2011 MFMER | slide-34

3D Volume AcquisitionEvents which cross several transaxial

slices are collected.

These events are used in the

reconstruction of each of the slices crossed.

©2011 MFMER | slide-35©2011 MFMER | slide-35

8050 60 704010 20 30

0

20000

40000

60000

80000

100000

2D

3D

NECR kCounts per second

kBq/cc

Practical Clinical Range for Oncology

10 20 30 40 50 60 70 80 mCi

©2011 MFMER | slide-36

Why wouldn’t one lower dose from 15-20 to 10 mCi (or lower) for 3D?

• What if patient is late?

• What if patient is early?

• What if patient has low uptake?

• What if patient is non-compliant?

Variation of time/dose has greater effect with smaller injected dose!

13

©2011 MFMER | slide-37

Solution: Bulk F-18 Infusion SystemNo patient to patient variability of dose!

©2011 MFMER | slide-38

Problem: Uptake Room � RestroomTech gets patient to scanner ‘on time’?

After 45min

©2011 MFMER | slide-39

Solution: Monitor Uptake � Restroom

After 45min

14

©2011 MFMER | slide-40

Automated Report Summary by Room/Day

©2011 MFMER | slide-41

Problem: Muscle uptake when walkingTool: Spaghetti Diagram

©2011 MFMER | slide-42

Solution: Minimum distance!

15

©2011 MFMER | slide-43

Six Sigma Results

• Variation in administered dose removed!

• Variation in ‘uptake time’ removed!

• Variation in transport uptake removed!

• These are ‘monitored’!

• Dose is LOWERED!

• Patients still provide unpredictable variation

©2011 MFMER | slide-44

Variability still present!PET Muscle uptake(variability of counts in lesions)

©2011 MFMER | slide-45

#2. CT Dose IndexReduction

Define

MeasureAnalyze

Improve

Control

16

©2011 MFMER | slide-46

Goal: AAPM Notification Values

©2011 MFMER | slide-47

©2011 MFMER | slide-48

DMAI “C”: Control Chart

17

©2011 MFMER | slide-49

Non-Size-Adjusted DLP vs #CT Exams for MCA Scanners

January 2011

(0018,1030) protocol name: 6300_AbdPel_uro_w_fur

(0008,103e) series description: 10m AbdPel a300

3

0 0

4

8

5 5

2

1

2 2

0 0 0 0 0 0

1

2

0

4

0

3 3

2

1 1

2

111

0

4

5

2

4

3

1 1 1

0 0 0 0 0 0

1

0 0

1

0

2

4

6

8

10

25

0-2

99

30

0-3

49

35

0-3

99

40

0-4

49

45

0-4

99

50

0-5

49

55

0-5

99

60

0-6

49

65

0-6

99

70

0-7

49

75

0-7

99

80

0-8

49

85

0-8

99

90

0-9

49

95

0-9

99

10

00

-10

49

10

50

-10

99

11

00

-11

49

11

50

-11

99

12

00

-12

49

12

50

-12

99

≥ 1

30

0

DLP w/o patient size adjustment(mGy-cm)

Mean - 2SD

288 mGy-cmMean + 2SD

1122 mGy-cm

Nu

mb

er

of

CT

Ex

am

s

Mean

705 mGy-cm

???????

Standardized:

Variation:

©2011 MFMER | slide-50

Analyze - Why Variation of DLP?

• Really Large Patient?

• Really Tall Patient?

• Elevated CTDIvol prescription?

• Other cause(s)?

©2011 MFMER | slide-51

18

©2011 MFMER | slide-52

Same patient 30 days apartCan you see the mAs/centering?

©2011 MFMER | slide-53

40 cm Projected SCOUT 36 cm Projected SCOUT

23 mSv 13 mSv

Not Centered Centered

©2011 MFMER | slide-54

Variation in DLP seen with variation in centering

19

©2011 MFMER | slide-55

ANALYZE: Variation due to Scout Order(180 mAs)

8.09 mGy 9.45 mGy 9.1 mGy 8.09 mGy

©2011 MFMER | slide-56

ANALYZE new example

CT of Abdomen/Pelvis 1/16/2011120kVp

180 ref. mAs

14.9mGy

CT of Abdomen/Pelvis 4/15/2011120kVp

180 ref. mAs

18.4mGy

23% increase in dose between these scans on the same patient.

©2011 MFMER | slide-57

CT Center FOV Positioning

20

©2011 MFMER | slide-58

% Dose Increase Due to Position

%mGy Increase Isocenter:Displaced

-9.0%

0.8% 0% -1.0%

2.5% 8.9% 4.5%

%mGy Increase Isocenter:Displaced

0%

Position changes mAs lookup table.

©2011 MFMER | slide-59

Same Patient

CT of Abdomen/Pelvis 1/16/2011120kVp

180 ref. mAs

14.9mGy

CT of Abdomen/Pelvis 4/15/2011120kVp

180 ref. mAs

18.4mGy

23% increase in dose between these scans on the same patient.

Dose Affecting Parameters:1. Sliderboard

2. X-Y Position 3. Body Width4. Z-Axis Attenuation

5. Metal Monitor

©2011 MFMER | slide-60

0%

10%

20%

30%

40%

50%

60%

Board Attenuation Patient Too Low Patient Wider Z Axis Coverage Different

Pareto Graph: Rank Change in CTDIvol

21

©2011 MFMER | slide-61

Variation: Z axis: Abdomen & Pelvis

©2011 MFMER | slide-62

New ‘standard’ Z axis coverage

©2011 MFMER | slide-63

22

©2011 MFMER | slide-64

Height/Weight with AP and Lateral Size

Use Random Forest Model estimates potentially ‘excessive’ Z axis coverage

©2011 MFMER | slide-65

Six Sigma Routine Abdomen-Pelvis

Top 10 Overscans

Superior (cm) Inferior (cm)

6.3 11

4.5 7.4

4.2 6.5

4.2 5.3

4.2 4.2

4.2 3.8

3.9 3.5

3.9 3.2

3.6 3.1

3.6 2.9

N= 300+ pts Overscan

superior 1 cm

inferior .4 cm

©2011 MFMER | slide-66

CT Colonography

Good scan coverage

OK to exclude lung bases, just include colon

23

©2011 MFMER | slide-67

CT Colonography Top 10 Overscans

Superior (cm)

Inferior (cm)

22.12 9.5

13.6 8

9.1 6.5

8.3 4.2

8.3 4.6

8 2.6

7.9 2.6

7.5 2

5.9 2

5.6 1

N= 20 pts Overscan

superior 6.4 cm

inferior 2.3 cm

©2011 MFMER | slide-68

Six Sigma Results

• Standardize Protocol Names

• Standardize CT’s Techniques

• Standardize CT Radiograph

• Standardize Patient Centering

• Standardize Z Axis coverage

• Control Charts

©2011 MFMER | slide-69

#3 Six Sigma DMAICwith Fluoroscopic Procedures

24

©2011 MFMER | slide-70

NEW NORMAL!

• New arterial and endoscopic procedures save open surgery!

• Patients are having multiple episodes of fluorsocopy

• 18% of currently scheduled patients have had more than 1 fluoroscopy!

• Patients are bigger

• Most fluoroscopy is outside of Radiology

©2011 MFMER | slide-71

Technical vs Behaviorial

©2011 MFMER | slide-72

5 ALARA Technical Tasksfor reproducibility

• Standardize Protocol Names!

• Standardize Fluoro/Acquistions!

• Default with Low(est) Fluoro pulse rate?

• Default with Low(est) DSA/Cine frame rate?

• Default with Low(est) dose per pulse?

• Always use Copper (0.1mm is 40%!)

25

©2011 MFMER | slide-73

(70-75 bpm, flat panel)(70-75 bpm, II)

7.5 FPS cine 15 FPS cineCurrent ½ Dose Old Dose

©2011 MFMER | slide-74

Use copper filter! vary thickness Lucite 15 cm to 35 cm

20% Iodine solution for CNR/mGy measures

©2011 MFMER | slide-75

~ 10% Lower Contrast to Noise for 0.1 mmCopper

CNR vs mm Cu Acquisition

0

20

40

60

80

100

120

140

160

0 0.2 0.4 0.6 0.8 1

mm Cu

CN

R

15cm Lucite

20cm Lucite

25cm Lucite

27.5cm Lucite

30cm Lucite

35cm Lucite

26

©2011 MFMER | slide-76

40% Lower Dose with 0.1mm Copper!

Differential Dose Acquisition

-90%

-80%

-70%

-60%

-50%

-40%

-30%

-20%

-10%

0%

0.0 0.2 0.4 0.6 0.8 1.0

mm Cu

Pe

rce

nt

Ch

an

ge

in

Do

se

fo

r A

dd

itio

n o

f C

op

pe

r

15cm Lucite

20cm Lucite

25cm Lucite

27.5cm Lucite

30cm Lucite

35cm Lucite

©2011 MFMER | slide-77

Use at least 0.1 mm Copper!

• Every fixed room without exception

• Every protocol!!!! (especially obese pt technique!)

• Portable C-arms (check with vendor)

• ALARA -100% exams with copper.

©2011 MFMER | slide-78

8 ALARA Behaviorsfor reproducibility

• Proper collimation

• Minimize use of Hi Mag

• Return to Normal FOV Fluoro prior to DSA

• Use of Fluoro Save (with LIH)

• Optimize patient positioning

• Table distant from tube

• DSA and Tap Lightly

• Announce cumulative Gy levels

27

©2011 MFMER | slide-79

Collimation Is Key: improves IQ, lowers PSD

79

< 1 Gy

1 -2 Gy

2 - 3 Gy

3 - 4 Gy

4 - 5 Gy

5 - 6 Gy

6- 7 Gy

7-8 Gy

8-9 Gy

9-10 Gy

©2011 MFMER | slide-80

0

1000

2000

3000

4000

5000

6000

19 17 13 10.5 8 7 6

Low

Normal

High

Behavior: choosing small FOV Result: High dose fluoro AND high DSA

Diagonal Size (in)

Sk

in D

ose

Ra

te (

mR

/min

)

©2011 MFMER | slide-81

8in Fluoroscopy 7in Fluoroscopy 6in Fluoroscopy

Single frame of II fluoro – fluoro save

0.7R/m 1.2 R/m 2 R/m

28

©2011 MFMER | slide-82

8in Single Shot 7in Single Shot 6in Single Shot

II Image Quality and small FOV Spot

Normal Normal Normal

©2011 MFMER | slide-83

Control: Are Fluoro Saved images in PACS?Informatic tools can monitorXA radiation events

©2011 MFMER | slide-84

29

©2011 MFMER | slide-85

SET ALARA I and II Levels (AK is ok)

IR Radiation Dose in mGy - EVAR

0

2000

4000

6000

8000

10000

12000

14000

16000

5/6

/2011

5/1

3/2

011

5/2

0/2

011

5/2

7/2

011

6/3

/2011

6/1

0/2

011

6/1

7/2

011

6/2

4/2

011

7/1

/2011

7/8

/2011

7/1

5/2

011

7/2

2/2

011

7/2

9/2

011

8/5

/2011

8/1

2/2

011

8/1

9/2

011

8/2

6/2

011

9/2

/2011

9/9

/2011

9/1

6/2

011

9/2

3/2

011

9/3

0/2

011

10

/7/2

011

10/1

4/2

011

10/2

1/2

011

10/2

8/2

011

11

/4/2

011

11/1

1/2

011

11/1

8/2

011

11/2

5/2

011

12

/2/2

011

12

/9/2

011

12/1

6/2

011

12/2

3/2

011

12/3

0/2

011

1/6

/2012

1/1

3/2

012

1/2

0/2

012

1/2

7/2

012

2/3

/2012

2/1

0/2

012

2/1

7/2

012

2/2

4/2

012

3/2

/2012

3/9

/2012

3/1

6/2

012

3/2

3/2

012

3/3

0/2

012

4/6

/2012

4/1

3/2

012

4/2

0/2

012

4/2

7/2

012

5/4

/2012

5/1

1/2

012

5/1

8/2

012

5/2

5/2

012

Date of IR Procedure

Radiation Dose in mGy Yellow Red Sentinel Event

Sentinel Event (15,000 mGy)

+

-

©2011 MFMER | slide-86

Fluoroscopy Positioning Review

Skin below IRP with Fluoro!

ExcellentToo much!

©2011 MFMER | slide-87

Peak Skin Dose Calculator

30

©2011 MFMER | slide-88

©2011 MFMER | slide-89

Fluoroscopy Service

OLDExposure Rate

CURRENTExposure Rate

GI Endoscopy/ERCP 3 ½ Pulses per second 3 ½ Pulses per second

Urology 15 Pules per second 3 ½ Pulses per second *

IR Room 1 RadiologyIR Room 1 VascularIR Room 1 DSA Aorta, C-E

10 pulses per second10 pulses per second4 Frames per second

7 ½ Frames per second *7 ½ Frames per second *2 Frames per second **

IR Room 2 RadiologyIR Room 2 VascularIR Room 2 DSA Aorta, C-E

15 Frames per second15 Frames per second4 Frames per second

7 ½ Frames per second *7 ½ Frames per second *2 Frames per second **

Cardiac Cath Rm 2 and 3 15 Pulses persecond FLUORO15 Fr/second CINE

7 ½ Pulses per second FLUORO7 ½ Fr/second CINE **

Electrophysiology LabVenograms

15 Fr/second FLUORO30 Fr/second CINE

4 Fr/second FLUORO *4 Fr/second CINE ***

©2011 MFMER | slide-90

Thank You!

• Six Sigma is a way of problem solving

• DMAIC

• Drive out variation – measure reproducibility

• Processes are the problem, not people!

• Team effort