Installation and
Setup GuideRev 5December 2, 2010Software version 4.0x
FMS-1630LIsolation Monitor
TRIATEKPhone: 770-242-1922
Fax: 770-242-1944http://www.triatek.com
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 2
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
General
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Mounting / Wiring
FMS-1630L Internal Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
FMS-1630L Remote Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-14
Function Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Wiring Diagram - Internal Sensor to Electric Damper Actuator . . . . . . . 16
Wiring Diagram - Internal Sensor to Pneumatic Damper Actuator. . . . . 17
Wiring Diagram - Internal Sensor to Variable Speed Drive. . . . . . . . . . . 18
Wiring Diagram - Remote Sensor to Electric Damper Actuator . . . . . . . 19
Wiring Diagram - Remote Sensor to Pneumatic Damper Actuator . . . . 20
Wiring Diagram - Remote Sensor to Variable Speed Drive . . . . . . . . . . 21
Wiring Diagram - Analog Input to Remote Pressure Sensor with
Internal Pressure Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-23
Wiring Diagram - Analog Input to 2 Remote Pressure Sensors . . . . 24-25
Wiring Diagram - Analog Input to Temperature Sensor . . . . . . . . . . . . . 26
Wiring Diagram - Digital Input to Door Switch . . . . . . . . . . . . . . . . . . . . 27
Wiring Diagram - Digital Input to Flow Switch . . . . . . . . . . . . . . . . . . . . 28
Wiring Diagram - Digital Input to Occupancy Sensor. . . . . . . . . . . . . . . 29
Wiring Diagram - Relay Output to Alarm . . . . . . . . . . . . . . . . . . . . . . . . 30
Wiring Diagram - Relay Output to Warning . . . . . . . . . . . . . . . . . . . . . . 31
Wiring Diagram - Relay Output to Heating Stage 1 . . . . . . . . . . . . . . . . 32
Wiring Diagram - Relay Output to Heating Stage 2 . . . . . . . . . . . . . . . . 33
Isolated Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Table of Contents
FMS-1630L INSTALLATION GUIDE
page 3 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Communications
General, Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-37
Reference
Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Programming
Touchpad Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Seltting Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Setting Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Exiting Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Setpoints1 Occupied & Unoccupied . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Setpoints2 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Setpoints2 Dead Band. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Setpoints2 Exaccel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
PID Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Setting Alarm Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Enable Door Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Buzzer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Display Refresh Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Damper Position. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Filter Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
Isolation Mode Change . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Sensor Range Settings Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
Alarm Line Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 4
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
GeneralSpecifications
Part Numbers
ElectricalOptional External Remote Sensor Distance . . . . . . . . . . . . . . . . Up to 4000 feetOptional External Remote Sensor Wiring 18-22 AWG, twisted shielded copper pairPower Supply. . . . . . . . . . . . . . . . . . . . . . . . . . Class2, 24 VAC/DC ±10%, 10 VA (Step-Down Isolation Transformer provided,120 to 24 VAC, 60 Hz,)System Accuracy Pressure Range Accuracy @ 72°F ± 5°F 0 - .0100 ± .0001’’ WC 0 - .0500 ± .0005’’ WC 0 - .1000 ± .001’’ WC 0 - .2000 ± .002’’ WC NIST Traceable / Individual certification available as option
4 Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . 4-20mADC, 0-5VDC or 0-10VDC4 Analog Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-5VDC or 0-10VDC4 Digital Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-5 or 0-24 VDC4 Relay Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A@24VDC, .2A@120VACInput Impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10k - 0 Output Impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 -10k
MechanicalLCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-Line x 16 characters per lineLCD Character Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.19 inches
Dimensions FMS-1630L Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.0”W x 6.0”H x 2.0”D Optional External Remote Sensor Housing . . . . . . . . . 2.0”W x 3.0”H x 2.7”D Stainless Steel Cover Plate for Flow Tube . . . . . . . . . . 2.7”W x 4.5”H x 0.2”D Stainless Steel Cover Plate for Remote Sensor. . . . . . . 2.7”W x 4.5”H x 0.2”DShipping Weight FMS-1630L w / Flow Tube Cover Plate. . . . . . . . . . . . . . . Approximately 3.5 lb FMS-1630L w / Optional External Remote Sensor . . . . . . Approximately 4.0 lbMounting FMS-1630L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Surface Flow Tube Cover Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Flush Optional External Remote Room Pressure Sensor . . . . . . . . . . . . . . . . . . Flush
EnvironmentalOperating Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 32° to 125° F Operating Operating Humidity . . . . . . . . . . . . . . . . . . . . 10% - 95% RH, Non-condensing
No special ventilation requirements are required for the proper operation of this product.
Using this product in a manner not specifi ed by the manufactuer may result in injury or hazardous conditions and should be avoided.
Analog Output Pressure Range Internal or Remote Sensor
1 = 0-5 VDC (> 500 001 = -0.0100 to +0.0100”WC (-2.4910 to + 2.4910 Pa) 0 = Standard Internal Sensor
2 = 0-10VDC (> 1K 005 = -0.0500 to +0.0500” WC (-12.46 to +12.46 Pa) 1 = Optional External Remote Sensor
010 = -0.1000 to +0.1000” WC (-24.91 to +24.91 Pa)
020 = -0.2000 to +0.2000” WC (-49.82 to +49.82 Pa)
Custom Ranges Available
FMS-1630L - ❑ - ❑ - ❑
FMS-1630L INSTALLATION GUIDE
page 5 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
GeneralOverview
The TRIATEK FMS-1630L Series Isolation Monitor is an ultra-sensi-
tive instrument used to control and/or monitor air pressure in hospital rooms,
labs, and clean rooms. This unit is a precision measuring system capable of
measuring and displaying air pressures down to 0.0001” WC. It can be used
to control pressures down to 0.001” WC. Features of the unit include:
• digital display of pressure with a
programmable descriptor
• six-key touchpad for convenient operation
• three LED indicator lights that show room
status at a glance
• audible and visual alarm annunciation
• Four relay outputs for transmitting alarm
conditions to a remote location such as a
central monitoring station
• auxiliary analog inputs for use with
optional sensors
• analog outputs used in control applications
• password protection of programmed setup
• keylock switch for isolation select protection
• optional LON communications.
The FMS-1630L is equipped with a 4-line 16-character per line alpha-
numeric LCD Display and can be wiped clean using a soft cloth without the
use of harsh chemicals. No decontamination requirements are required for the
proper operation of this product.
There are three Room Status lights to indicate “Normal” when pressure
is within defined limits, “Caution” when pressure is nearing an out-of-limits
condition, and “Alarm” when pressure is outside defined acceptable limits.
The pressure ranges for these conditions are easily set by the user for the
specific installation when necessary. The status lights provide overview of
room pressure conditions at a glance.
Alarm conditions can be defined by the user, in terms of desired
pressure settings for the room being monitored. When an alarm condition
occurs, it can be annunciated in three user-definable ways: 1) on the dis-
play, 2) with an audible alarm, and 3) transmitted via contacts to a remote
monitoring station and over network. The alarm can be set to automatically
reset when the unit has sensed that the room pressure has returned to
proper limits, or it may be set to remain on until manually reset. In either
case, the attendant can mute the audible alarm by pressing the Silence
Alarm touchpad.
The FMS-1630L provides four Relay outputs that can be used for
remote alarm annunciation or pilot control functions. The operation of each
output can be configured by the user to define the exact room air pressure
values above and below which the output will operate. Delay times before
activation can be specified to minimize nuisance activation.
In many installations, it is important to have other room variables such as
temperature or relative humidity to be displayed along with room pressure.
The FMS-1630L provides for this by means of Additional Analog Inputs
which can be configured for 4-20 ma current or for voltage input signals.
The input can be scaled as needed to display correct values, and a suitable
descriptor can be applied.
The FMS-1630L provides Analog Outputs which can be set up for
voltage. It can be programmed to be proportional output for providing a
linear signal to an automation system, or programmed for PID floating point
output for direct control of damper actuators, speed drives, etc.
The user can set up a Password in the programming sequence to pro-
tect against unauthorized or casual access to the FMS-1630L programmed
variables.
Room pressure selection of Positive, Negative, or No Isolation is set
using a Keylock Switch.
An LON Communications Interface enables the unit to communi-
cate room status information to the building automation system.
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 6
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting/Wiring
FMS-1630L with Internal Sensor
Introduction:
FMS-1630L display/processor units having an internal sensor should be installed outside of the room on the
corridor wall. Sensor tubing should be between this corridor and isolation room in the most direct and shortest
path.
The FMS-1630L is calibrated with settings programmed in the factory according to the customer specifications
(provided by the model number). The user can change the settings by using the instructions provided in the User/
Programming section of this manual. This product should be calibrated periodically to ensure the accuracy of the
measurement readings and overall performance.
The electrical connections to the FMS-1630L are made via a convenient removable terminal block as shown on
page 10. All wiring should conform to the Local Regulations and National Electrical Code. Take care not to run
sensor wiring in the same conduit as line voltage or other conductors that supply highly inductive loads such as
generators, motors, solenoids, contactors, etc. Use 22 AWG or larger.
Suggestion:
When mechanical and electrical installations are complete, apply power to unit and observe display. The room
pressure should be indicated correctly for whatever it is. If parameters need to be changed, check the table of
contents to locate the specific item of interest.
FMS-1630L INSTALLATION GUIDE
page 7 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting/WiringFMS-1630L with Internal Sensor
Mounting Steps:
1. Choose a location free of airflow obstructions,
and minimal air turbulence. The installation
should also avoid stagnant air or “short circuit-
ing” of the supply air to the exhaust. Typically, the
FMS should be located at eye level. It should not
be mounted over the door since the LCD display
cannot be read from that angle and since the
alarm annunciator SILENCE touchpad cannot be
reached to silence an alarm if the local alarm is
enabled.
2. Remove the front cover from the base. Use the
base as a template to mark the wall for the four
mounting holes, wire conduit hole and the pres-
sure tube (see page 9).
3. Run the conduit and all wiring according to
local regulations and National Electrical Code.
IMPORTANT: Use the 24 VAC stepdown isola-
tion transformer, provided with the unit, to power
the FMS 1630; this will prevent ground loop and
other problems that might otherwise be encoun-
tered.
4. Drill a 3/4” hole through the walls for the sen-
sor tube fitting. Refer to page 9. Typically, the
hole on the opposite side should be either level
or below the sensor clearance hole on the FMS
side to reduce the potential for liquid from enter-
ing the sensor line.
6. Pull the air tube through the wall. Slide the hose
over a long thin screwdriver or similar rod to help
pull the tube through the wall.
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 8
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting/WiringDisplay/Processor Unit with Electrical Conduit
7. Slide the tube tightly onto the sensor fitting.
Caulk all openings in the wall and around the
conduit, and apply duct sealant putty in the con-
duit around the electrical wires so that there can
be no air flow through the back of the case or
through the wall into the case. It is very important
that no air is able to bypass the flow path through
the sensor tube. Do not caulk the opening
between the front cover and the mounting
base (or back cover). This space is a 1/16”
opening between the wall and the cover and
also 1/16” between the mounting base and the
cover. It is necessary for this path to be open for
air flow when monitoring room pressurization.
Remember: The front cover is designed to
be spaced away from the wall so that the
reference air can flow without obstruction.
This is superior to a more conventional vent
opening which is subject to being clogged
by pencils or other trash being stuffed into
the opening. It is important that caulking not
be applied to the gap between the wall and
the back edges of the front cover.
Pull slack from the tube on the opposite side of
the wall as the display/processor case is moved
toward the wall for mounting. On the opposite
side the return air stainless base and louvered
plate must be mounted as shown on page 9.
Caulk the tube and anchor holes in the wall and
caulk behind the stainless mounting plate so that
there can be no air flow through the wall. Mount
the base to the wall with the appropriate screws
and anchors.
8. Connect the wires to the removable connector.
9. Plug the removable connector onto the board
in the unit (See page 10), then carefully fold the
cable and install the front cover of the FMS-
1630L to the base.
10. On the opposite side of the wall, a length of tube
will be protruding from the wall. Cut the tube as
needed to leave approximately two inches pro-
truding. Press the tube onto the barbed fitting of
the Flow Tube Mounting Plate (see page 9). Push
the mounting plate and tube toward the wall,
forcing the excess into the wall space. Fasten
the mounting plate to the wall using appropriate
hardware.
11. Mount the louvered stainless steel cover plate to
the mounting plate with the screws supplied.
12. With power applied, the unit should display room
pressure.
FMS-1630L INSTALLATION GUIDE
page 9 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting/Wiring
Display/Processor Unit with Electrical Conduit
IsolationRoom
Flow Tube Sensor
Gasket
Conduit-1/2''minimum
1 Pair Shielded Cablewith ground wire16 AWG or smaller
UL ClassifiedFireSealantapply around tube and between stainless steel plate and wall to seal unit penetration
Stainless SteelReturn Air
Cover Plate
Unlock
Lock
Corridor
Wall Section(cutaway view)
FMS-1630L Display/ Processor
Corridor
Isolation Room
* See Note
*Note:Flow tube must not have bends that can close off the air flow.
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 10
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting/WiringInternal Sensor
The FMS-1630L is calibrated with settings programmed in the factory according to the customer speci-
fications (provided by the model number). The user can change the settings by using the instructions
provided in the Programming section of this manual, beginning on page 28.
FMS-1630L Series with Standard Internal Sensor
Ground Terminal
No Link When Internal Sensor
Used
FMS-1630L INSTALLATION GUIDE
page 11 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting/WiringRemote Sensor Option
Introduction
FMS-1630L’s ordered with a remote sensor have no internal sensor; that is, no sensor is included in the dis-
play case. You must install a remote sensor unit in the wall between the isolation room and the adjoining corridor.
Port P1 must be oriented towards the isolation room and Port P2 toward the corridor. Please see illustration on
page 12. With this sensor orientation a positive pressure value indicates that the isolation room is positive with
respect to the corridor. A single pair (two conductor) shielded cable must be connected between the remote
sensor and the display processor unit. Maximum length of this cable is 4000 feet. The display unit can be installed
outside the room, at the nurses station, in the engineering office, or at any other location as needed.
Mounting Steps:
1. Cut opening in the wall of the isolation room to receive the mounting enclosure for the remote sensor elec-
tronics. Nominal hole dimensions are 3 1/8” h x 2” w.
2. Drill a 7/16” hole through the opposite wall for the flow tube as shown.
3. Bring the 2-conductor signal wire through the cut-out.
4. Knock out the back and one bottom knock-out, then with the signal wire pulled through the bottom hole
into the box. Mount the enclosure box in the opening provided and secure with screw and anchors. Seal
around the edges of the box with UL Classified Fire Sealant.
5. Push a length of flow tube through the back hole on through the 7/16” hole in the opposite wall.
6. Attach the flow tube to the sensor port, then push the tube and sensor module into place and secure to
the mounting enclosure with two 6-32 x 3/4 screws supplied.
7. Install the louvered cover plate.
8. On the opposite side (corridor) attach the flow tube to the barbed fitting of the flow tube mounting plate.
9. Press the mounting plate into place, allowing the excess tube length to go into the wall space. Secure with
the screws and anchors.
10. Install the louvered cover plate.
11. At the FMS-1630L connect the two leads from the remote sensor to the terminals marked +V “+” and AI1
“-”. Remote Sensor (+) goes to terminal +V (+) and Remote sensor IN (-) goes to terminal AI1 (-).
Install jumper on pin 1 JP6 and JP3 see diagram on page 14. (If not installed at the factory.)
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 12
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Wall Section(cutaway view)
Corridor
Gasket
Stainless Steel Mounting Plate
Gasket
UL ClassifiedFireSealantapply around tubeand between stainless steel plate and wall to seal unit penetration
Stainless SteelFlow TubeMountingPlate
Sensor Port P1
Sensor Port P2
Printed Circuit Boardon back ofremotesensor
Louvered Cover Plate
IsolationRoom
Use 18-22 AWGtwisted shielded
copper pair(up to 4000 feet)
Plug-in Terminal Blockfor connection of transmitterto FMS-1600Series
PP22PP11
~~UL Classified
FireSealant
apply around tube and between stainless steel plate and wall
to seal unit penetration
Flow Tube*See Note
Mounting/WiringRemote Sensor Option (must be specified at time of purchase)
To FMS Display/Processor
Remote Sensor Input
(see “Mounting/Wiring
Remote Sensor Option) Corridor
Isolation Room
External Remote Sensor Installation Detail
(Side View)
*Note:When Flow Tube mounting plate is located directly opposite the sensor, Flow Tubing must be cut as short as possible to prevent a kink from forming after installation
FMS-1630L INSTALLATION GUIDE
page 13 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting/WiringRemote Sensor Option
InsideWall
IsolationRoom
Corridor
P2P1
Gasket
Stainless Steel Mounting Plate
use 18-22 AWGtwisted shielded copper pair(up to 4000 feet)
~~
Stainless SteelFlow TubeMountingPlate
Sensor Port P1
Sensor Port P2
Louvered Cover Plate
Plug-in Terminal Blockfor connection of transmitterto FMS-1600Series
UL ClassifiedFireSealantapply around tube and between stainless steel plate and wall to seal unit penetration
Corridor
Isolation Room
Remote Sensor Installation Detail
(Bottom View)
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 14
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+V
ANALOG IN
123
EXT.SNSRSELECT
EXT.SNSR
GND
111
TPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
P3
1
1
1
1111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
SELECT
G5v+V
DI
SRVCSRVC
FLASH_WP
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
LED1
JP2
LED
LED2
1
P4
TxLED3
LED4Rx
INT.SNSRGND
+10v
+5v
GND
GND
AUX PWR O
UTP
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
-+
Remote Sensor/ Transmitter
Mounting / WiringRemote Sensor Connector Strip
The electrical connections to the FMS-1630L are made via a convenient removable terminal block as shown below. All wiring should conform to the Local
Regulations and National Electrical Code. Take care not to run Sensor wiring in the same conduit as line voltage or other conductors that supply highly
inductive loads such as generators, motors, solenoids, contactors, etc. Use 22 AWG or larger.
Removable Connector for easy installation of field wiring.
Note:The FMS remote sensor is an option which must be specified at time of purchase from factory.
Remote Sensor Connector
FMS-1630L INSTALLATION GUIDE
page 15 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringFunction Descriptions
AUX Power Output
LON Communications
Analog Output
Analog Input
Power
Digital Input
Relay Output
Power supply outputs used for power auxiliary devices. Each is supply limited to 100
ma.
LON FTT-10A Free Topology Network building automation system can access the sta-
tus and the configuration of the unit. Also used by configuration software to configure
unit.
Unit has 4 Analog Outputs which can be configured for 0-5 VDC or 0-10VDC.
Unit has 4 Analog Inputs which can be configured for 4-20 mA, 0-5VDC, or 0-
10VDC.
The FMS-1630L Series can be powered by either 24 VAC or 24 VDC.
A 120 VAC to 24 VAC Step Down Isolation Transformer is provided and recommended.
This power must be from a Class 2 supply only.
The unit has 4 Digital Inputs with selectable pull-up voltages of 0V, +5V and +18V.
The unit has 4 SPDT Relay outputs with normally open contacts.
Function Description
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 16
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Internal SensorAnalog Output to Electric Damper Actuator
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
ANALOG IN
123
EXT.SNSR
GND
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7Ain
JP6
1
1
1
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
GND
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
LED1
JP2
LEDLED2
1
P4
TxLED3
LED4Rx
1
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
+V
TPUT
+10v
+5v
GND
GND
AUX PWR O
UTPTP-+
+ -+-
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
Electric Damper Actuator
Optional ExternalTransmitter
Power Supply
CONNECTIONS FOR POWERED OUTPUT
110 VAC In 24 VAC In
Step Down Isolation Transformer
FMS-1630L INSTALLATION GUIDE
page 17 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Internal SensorAnalog Output to Pneumatic Damper Actuator
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
ANALOG IN
123
EXT.SNSR
GND
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
1
1
1
+10v
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
+5v
GND
GND
SHLD
NETWRK IN
NT_2
NT_1AUX PW
R
PWR
Vsel
1
SW1
LED1
JP2
LED
LED2
1
P4
TxLED3
LED4Rx
1
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
+V
GND
WR O
UTPUT
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
CONNECTIONS FOR POWERED OUTPUT
~~
WR
~~110 VAC In 24 VAC In
CP-3240
P G AI FB
M B
GageM B
to Pneumatic Actuator for Damperor Venturi Valve
Main Air Supply20-30 PSI
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 18
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring - Internal SensorAnalog Output to Variable Speed Drive
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
ANALOG IN
123
EXT.SNSR
GND
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
1
1
1
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
GND
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
SW1
LED1
JP2
LEDLED2
1
P4
TxLED3
LED4Rx
1
1
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
+V
TPUT
+10v
+5v
GND
GND
AUX PWR O
UTUTP
AI
G
+ -+-
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
Variable Speed Drive
Optional ExternalTransmitter
Power Supply
CONNECTIONS FOR POWERED OUTPUT
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
FMS-1630L INSTALLATION GUIDE
page 19 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Remote SensorAnalog Output to Modulated Air Controller
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+V
ANALOG IN
123
EXT.SNSRSELECT
EXT.SNSR
GND
1
TPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
P3
1
1
1
1111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
CONNECTIONS FOR POWERED OUTPUT
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
SELECT
G5v+V
DI
SRVCSRVC
FLASH_WP
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
LED1
JP2
LED
LED2
1
P4
TxLED3
LED4Rx
24VIN
+
-
INT.SNSRINTGND
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
+10v
+5v
GND
GND
AUX PWR O
UTP
INT
~~
-+
Air Control Modulator
-+
Remote Sensor/ Transmitter
Air Flow To and From Unit Between Room and Corridor
+ -+-Optional External
TransmitterPower Supply
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 20
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringAnalog Input to Remote Pressure Sensor with Internal Sensor0-10V Out
0-10V OUT
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
123
EXT.SNSR
+5v
GND
GND
GND
GND
AUX PWR O
UTPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
1
1
1
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
1
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
LED1
JP2
LED
LED2
1
P4
TxLED3
LED4Rx
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
-IN
+
Pressure Sensor
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
FMS-1630L INSTALLATION GUIDE
page 21 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringAnalog Input to Remote Pressure
Sensor with Internal Sensor
4-20mA Out
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
123
EXT.SNSR
+5v
GND
GND
GND
GND
AUX PWR O
UTPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
1
1
1
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
1
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
LED1
JP2
LED
LED2
1
P4
TxLED3
LED4Rx
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
4-20mA OUT
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
-+
Pressure Sensor
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 22
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringAnalog Input to 2 Remote Pressure Sensors 0-10V Out
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
123
EXT.SNSRSELECT
EXT.SNSR
+5v
GND
GND
GND
GND
AUX PWR O
UTPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
1
1
1
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
T.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
LED1
JP2
LED
LED2
1
P4
TxLED3
LED4Rx
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111INT.S
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
T.S
~~
-
+
Pressure Sensor 1 IN
~~
~~
-IN
Pressure Sensor 2
+
0-10V OUT
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
FMS-1630L INSTALLATION GUIDE
page 23 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Analog Input to 2 Remote Pessure
Sensors4-20mA Out
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+V
ANALOG IN
123
EXT.SNSRSELECT
EXT.SNSR
GND
TPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
Ain
JP6
1
1
1
1111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
+10v
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
INT.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
GND
GND
GND
SHLD
NETWRK IN
NT_2
NT_1AUX PW
R OUTP
PWR
Vsel
1
SW1
LED1
JP2
LED
LED2
1
P4
TxLED3
LED4Rx
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
+5v
-+
Pressure Sensor 1
~~
-+
Pressure Sensor 2
4-20mA OUT
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 24
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringAnalog Input to Temperature Sensor
0-5v (JP7)
Jumper Setting
AI 1AI 2AI 3AI 4
1234 = JUMPER ON
Analog Input
0-10v (JP7)
AI 1AI 2AI 3AI 4 JP7
1234
Analog Input
4-20mA (JP6)
AI 1AI 2AI 3AI 4
Analog Input
JP6
0-5v (JP5)
AO 1AO 2AO 3AO 4
Analog Output 0-10v (JP5)
AO 1AO 2AO 3AO 4 JP5
Analog Output
Note: 4-20mA can only be used if thecorresponding JP7 Analog Input Jumperis set to OFF.
Note: Each Analog Input or Output value can be set independent of one another.Example:AI 1 can be set to 0-5v jumper OFF (JP7 PIN 1)AI 2 can be set to 0-10v jumper ON (JP7 PIN 2)AI 3 can be set to 4-20mA jumper ON (JP6 PIN 3) and jumper OFF (JP7 PIN 3)AO 1 can be set to 0-10v jumper ON (JP5 PIN 1)AO 2 0-5v can be set to 0-5v jumper OFF (JP5 PIN 2)
= JUMPER OFF
1234
1234
1234
1234
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
123
EXT.SNSR
+5v
GND
GND
GND
GND
AUX PWR O
UTPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
JP4
2
JP5
4
RNG
1mA
3
JP7
AinJP6
1
1
1
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
SRVCSRVC
FLASH_WP
1
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
LED1
JP2
LEDLED2
1
P4
TxLED3
LED4Rx
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1LED3LEDLEDLED
LED4LEDLEDLED
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
~~-+
Temperature Sensor
-+
PowerSupply
FMS-1630L INSTALLATION GUIDE
page 25 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Digital Input to Door Switch
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
Door Switch
Door SwitchMagnet
DOOR
DOORFRAME
= ON
JP4 is set at factory to +5 setting. User can change voltage setting to suit application requirement.
Jumper Block ON CPU Board
G1
5v
V+
1
JP4
5 62
DI_1
DI_2
DI_4
GND
GND
GND
DIGITAL IN
DI_3
G5v+V
DIVsel
11
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
NT_1
NT_2
NETWRK O
UT
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSR
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLD
123
EXT.SNSRSELECTINT.SNSR
SRVC
EXT.
SRVC
SNSR
FLASH_WP
+5v
GND
GND
GND
GND
1
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
UTPUT
PWR
ANALOG O
UT
1
SW1
RNG
4
123
LED1
JP2
LED
Ain
4
Aout
LED2
1JP4
2
JP5
4
RNG
1Tx
mA3
LED3
LED4
JP7
Ain
JP6 Rx
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111JP2FLAJP2FLA
LED2
1
1
LED3LEDLEDLED
LED4LEDLEDLED
1
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
A switch having normally-open or normally-closed contacts may be used with the FMS-1630L to serve as a timed alarm buzzer inhibit, when the room
door has been opened. An optional door switch (part number SWD-100) may be purchased from TRIATEK for this specific purpose.
Door switch connections are located inside the front cover of the FMS-1630L. After the switch has been installed at the door and connected to the FMS
unit, its operation may be programmed as described on page 44 of this manual.
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 26
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Digital Input to Flow Switch
DI_1
DI_2
DI_4
GND
GND
GND
DIGITAL IN
DI_3G5v+V
DIVsel
11
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
NT_1
NT_2
NETWRK O
UT
GND
PH1
PH2
+VS
SYN
+
-
24VIN
SASH SNSR
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLD
123
EXT.SNSRSELECTINT.SNSR
SRVC
EXT.
SRVC
SNSR
FLASH_WP
+5v
GND
GND
GND
GND
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
UTPUT
PWR
ANALOG O
UT
SW1
RNG
4
123
LED1
JP2
LED
Ain
4
Aout
LED2
1JP4
2
JP5
4
RNG
1
Tx
mA3
LED3
LED4
JP7
Ain
JP6 Rx
1
1
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1
1
LED3LEDLEDLED
LED4LEDLEDLED
1
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
S
FLOW SWITCH
= ON
JP4 is set at factory to +5 setting. User can change voltage setting to suit application requirement.
Jumper Block ON CPU Board
G
5v
+V
1
JP4
5 62
FMS-1630L INSTALLATION GUIDE
page 27 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringDigital Input to Occupancy Sensor
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
EXT.SNSR
FLASH_WP
+5v
GND
GND
GND
GND
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
UTPUT
ANALOG O
UT
Vsel
RNG
LED
Aout
JP4
JP5
RNG
JP7
Ain
JP6
JP2
1
1
1
1
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
NT_1
NT_2
NETWRK O
UT
GND
PH1
PH2
+VS
SYN
24VIN
SASH SNSR
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLD
EXT.SNSRSELECTINT.SNSR
EXT.SNSR
FLASH_WP
+5v
GND
GND
GND
GND
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
UTPUT
ANALOG O
UT
RNG
LED
Aout
JP4
JP5
RNG
JP7
Ain
JP6
JP2FLAFLA
1
1
1
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
OCCUPANCYSENSOR
= ON
JP4 is set at factory to +5 setting. User can change voltage setting to suit application requirement.
Jumper Block ON CPU Board
G
5v+V
1
JP4
5 62
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 28
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Relay Output to Alarm
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
AO_1
AO_2
AO_3
GND+10v
+V
PH1
PH2
+VS
SYN
SASH SNSR
EXT.SNSRSELECTINT.SNSRGND
GND
GND
1
AUX PWR O
UTPUT
ANALOG O
UT
JP3111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
AO_4
AI_1
AI_2
AI_3
AI_4
GND
ANALOG IN
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
+
-
24VIN
DIGITAL IN
SHLDDI_3
123
G5v+V
DI
SRVC
EXT.
SRVC
SNSR
FLASH_WP
+5v
GND
GND
SHLD
NETWRK IN
NT_2
NT_1
PWR
A
Vsel
1
SW1
RNG
4
123
LED1
JP2
LED
Ain
4
Aout
LED2
1
JP4
2
JP5
4
RNG
1
Tx
mA3
LED3
LED4
JP7
Ain
JP6 Rx
******
***
SW1
LED1D1D1D1
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1
1
1
LED3LEDLEDLED
LED4LEDLEDLED
1
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
+VMonitor Output 1
2A Max
30VDC Max
NOTE: Relay 3 & 4 Share a common contact terminal block pin, (C3/4).
FMS-1630L INSTALLATION GUIDE
page 29 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringRelay Output to Warning
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
AO_1
AO_2
AO_3
AO_4
GND
GND
+10v
+V
DI_1
GND
PH1
PH2
+VS
SYN
SASH SNSR
EXT.SNSRSELECTINT.SNSR
+5v
GND
GND
GND
1
AUX PWR O
UTPUT
ANALOG O
UT
JP3111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
AI_1
AI_2
AI_3
AI_4
ANALOG IN
NT_1
NT_2
NETWRK O
UT
DI_2
DI_4
GND
GND
+
-
24VIN
DIGITAL IN
SHLDDI_3
123
G5v+V
DI
SRVC
EXT.
SRVC
SNSR
FLASH_WP
GND
SHLD
NETWRK IN
NT_2
NT_1
PWR
Vsel
1
SW1
RNG
4
123
LED1
JP2
LED
Ain
4
Aout
LED2
1
JP4
2
JP5
4
RNG
1
Tx
mA3
LED3
LED4
JP7
Ain
JP6 Rx
******
***
SW1
LED1D1D1D1
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1
1
1
LED3LEDLEDLED
LED4LEDLEDLED
1
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
+VMonitor Output 2
2A Max
30VDC Max
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 30
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Relay Output 1
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
EXT.SNSRSELECTINT.SNSRGND
1JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
+-
24VIN
SASH SNSRDIGITAL IN
SHLDDI_3
123
EXT.SNSR
G5v+V
DI
SRVC
EXT.
SRVC
SNSR
FLASH_WP
+5v
GND
GND
GND
GND
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
UTPUT
PWR
ANALOG O
UT
Vsel
1
SW1
RNG
4
123
LED1
JP2
LED
Ain
4
Aout
LED2
1
JP4
2
JP5
4
RNG
1
Tx
mA3
LED3
LED4
JP7
Ain
JP6 Rx
******
***
SW1
LED1D1D1D1
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1
1
1
LED3LEDLEDLED
LED4LEDLEDLED
1
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
+VHeating Stage 1
2A Max
30VDC Max
TTemperatureSensor
FMS-1630L INSTALLATION GUIDE
page 31 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / WiringRelay Output 2
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
AO_1
AO_2
+V
PH2
+VS
EXT.SNSRSELECTINT.SNSRGND
GND
1
OUTPUT
JP3111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
ANALOG IN
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
SYN
+
-
24VIN
SASH SNSRDIGITAL IN
SHLDDI_3
123
G5v+V
DI
SRVC
EXT.
SRVC
SNSR
FLASH_WP
+5v
GND
GND
GND
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
U
PWR
ANALOG O
UT
Vsel
1
SW1
RNG
4
123
LED1
JP2
LED
Ain
4
Aout
LED2
1
JP4
2
JP5
4
RNG
1
Tx
mA3
LED3
LED4
JP7
Ain
JP6 Rx
******
***
SW1
LED1D1D1D1
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1
1
1
LED3LEDLEDLED
LED4LEDLEDLED
1
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
+VHeating Stage 2
2A Max
30VDC Max
TTemperatureSensor
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 32
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Step Down Isolation Transformer(provided with FMS-1630L)
110 VAC24 VAC
+
-
24VIN
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_4
GND
GND
+10v
+V
ALOG IN
NT_1
NT_2
NETWR
DI_1
DI_2
DI_4
GND
GND
PH1
PH2
+VS
SYN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
123
EXT.SNSRSELECTINT.SNSR
G5v+V
DI
SRVC
EXT.
SRVC
SNSR
FLASH_WP
+5v
GND
GND
GND
GND
1
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
UTPUT
PWR
ANALOG O
UT
Vsel
1
SW1
RNG
4
123
LED1
JP2
LED
Ain
4
Aout
LED2
1
JP4
2
JP5
4
RNG
1
Tx
mA3
LED3
LED4
JP7
Ain
JP6 Rx
******
***
SW1
LED1D1D1D1
JP3
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
JP2FLAJP2FLA
LED2
1
1
1
LED3LEDLEDLED
LED4LEDLEDLED
1
111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111111
AI_3 ANAL
NT_2
RK OUT
GND
SHLDALO D
SHLDHLDDI_3
NT_2T_2
RK
WiringPower
FMS-1630L INSTALLATION GUIDE
page 33 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Mounting / Wiring Isolated Power Supply
Red / Fused 24 VAC connected to Power (+) on the FMS
1 Amp Fuse
Ground
}
}120 VAC 60 Hz
Black
White
Stepdown Isolation
Transformer
Blue / 24 VAC connected to Power (-) on the FMS
Only Class 2 wiring in this compartment. 24 VAC
60 Hz
*Note:This product should be installed with the manufacturer provided isolated power supply and connected to an electrical circuit protected by a minimum 20A circuit breaker. This circuit breaker should be mounted in an approved electrical enclosure located separately from this product.
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 34
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
CommunicationsWiring
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
123
INT.SNSR
EXT.
SRVC
SNSR
+5v
GND
GND
GND
GND
SHLD
NETWRK IN
NT_2
NT_1
AUX PWR O
UTPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
2
JP5
4
RNG
1mA
3
JP7JP6
JP3
1
1
1
AO_1
AO_2
AO_3
AO_4
AI_1
AI_2
AI_3
AI_4
GND
GND
+10v
+V
ANALOG IN
NT_1
NT_2
NETWRK O
UT
DI_1
DI_2
DI_4
GND
GND
GND
PH1
PH2
+VS
SYN
-
24VIN
SASH SNSRDIGITAL IN
RLY1
RLY2
RLY3
RLY4
C3/4
C2
C1
RELAY OUTPUTS
SHLDDI_3
123
INT.SNSR
EXT.
SRVC
SNSR
+5v
GND
GND
GND
GND
AUX PWR O
UTPUT
ANALOG O
UT
RNG
4
123
Ain
4
Aout
2
JP5
4
RNG
1mA
3
JP7JP6
JP3
1
1
1
FMS1630L
FMS1630L
NC No Connection to Field Wiring
Field Wiring w/space for Number
Internal Wiring
Screw Terminal
Wiring Guide Legend
Symbol Description
Air Flow To and From UnitBetween Room and Corridor
FMS1630LCONTROLLER
LON FTT-10 Free Topology Communications Network
18-22AWG Twisted Pair
Cable Type Maximum Node-to-Node Distance (FT)
Maximum Total Wire Length (FT)
Belden 85102 1640 ft 1640 ft
Belden 8471 1312 ft 1640 ft
Level IV, 22AWG 1312 ft 1640 ft
JY (St) Y x2x0.8 1049 ft 1640 ft
TIA Category 5 820 ft 1476 ft
Free Topology Cable Specifications
Refer to www.lonmark.orgfor more details.
FMS-1630L INSTALLATION GUIDE
page 35 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
CommunicationsGeneral, Wiring
LON WIRING INSTRUCTIONS
Communications connections require that the FMS-1630L units be
connected with twisted pair communication cable to each unit in the
network. The unique network address of each FMS unit is set by a
network management tool like LONMAKER.
All wiring must be done in accordance with the NEC as well as regula-
tions of all authorities having jurisdiction, and must conform to appli-
cable codes. When required by code, communications wiring may be
installed in conduit of a type designed specifically for this purpose.
WIRE TERMINATIONS
The FMS-1630L is provided with a removable connector block with
convenient screw terminals. Make the LON FTT-10 communications
connections as follows:
1 Connect a cable lead to the “NT1 -” terminal (#1).
2 Connect a cable lead to the “NT2” terminal (#2)
GUIDELINES FOR WIRING
Following these guidelines will help to keep wiring-related communica-
tions problems to a minimum:
1. Do not splice communications cable or wire at any point.
2. Avoid “T-tap” technique of routing/connecting communications
cable. Conductor discontinuities produced by such connections may
generate RFI or other electromagnetic interference on the communica-
tions circuit.
3. Do not use wire nut devices for connecting communications cable.
4. Do not route any part of the communications cable through conduit,
junction boxes or other devices containing AC electrical wiring.
5. Do not strap communications cable to any conduit or other device
containing AC electrical wiring, or run communications cable parallel to
(or against) such devices.
NOTE: AC electrical devices such as transformers, disconnects, fluo-
rescent lighting, motor-controllers, variable frequency drives or other
high voltage power sources may generate RF interference which could
cause intermittent problems in the communications network.
Wire the LON network in accordance with LON network standards.
NOTE: Be sure to observe installation instructions regarding possible need for a termi-nation load or other device that may have to be attached on the end of a run.
Typical Network Topologies
Singly Terminated Bus Topology
Doubly Terminated Bus Topology
Loop Topology
Star Topology
Mixed Topology
Termination
Termination Termination
Termination
Termination
Termination
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 36
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
FMS-1630L Series QUICK-START
1- AFTER THE FMS-1630L UNIT HAS BEEN INSTALLED, apply power to the unit. Observe the displayed room pressure and verify that it is responding to room pressure. If it is, proceed to step 2 below. If it is not responding, check the installation, especially the flow tube connections (page 9 or 12 ) and if remote sensor, also check the wiring. - An indication of a maximum negative number (eg, –.10000) indicates the sensor is not connected. - An indication close to zero indicates a missing or blocked air tube to the sensor.
2- TEST THE ALARM STATUS INDICATORS. When the unit is indicating response to room pressure, observe that the status lights on the face of the unit are indicating properly. a- With the room door closed, the displayed room pressure should be .001 “WC or greater (or –.001 or more negative if negatively pressurized). b- If the pressure is positive, select Positive Isolation mode; if it is negative, select Negative Isolation mode. c- The green “NORMAL” light should now be on. d- Open the door to the isolation room. Observe that the displayed pressure drops and the ALARM light comes on. e- After a delay of several seconds, the audible alarm (buzzer) should sound.
FMS-1630L INSTALLATION GUIDE
page 37 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
User Operation SectionTypical Main Display
After power-up the FMS-1630L displays the TRIATEK product description for 5 seconds and then brings up the
main display. A typical main display is shown below. This display contains the following information:
1) Descriptor - This can be up to eight (8) alpha numeric characters such as “RM 1020”.
2) Type of Isolation selected for the room.
3) The room pressure reading.
Typical Main Display(appears upon power-up)
Display Channel 1
Display Channel 2
Display Channel 3
Display Channel 4
RM 1020 Neg Iso-.00190”WC
ISOLATION MONITORFMS-1630L
Four display channel are available on the FMS-1630L. These can be configured as needed but the factory setting is as shown here.
Channel 1: Blank (not used here)
Channel 2: Descriptor - 8 characters and isolation type selected
Channel 3: Room pressure reading
Channel 4: Blank- (not used here)
Display Channels
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 38
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
ROOMSELECT
ALARMSILENCE
TEST
Touchpad Operation for ProgrammingTouchpad Definition and Colors
Touchpad Description
ROOMSELECT
Green Touchpad
TESTYellow
Touchpad
ALARM SILENCE
Red Touchpad
AGrey Touchpad
BGrey Touchpad
CGrey Touchpad
Touchpads are used for special functions during programming the FMS-1630L as described below.
A
B
C
Used for scrolling through menus, increasing numeric values and entering Setup mode.
Used for scrolling through menus, decreasing numeric values and entering Setup mode. Press and release both Green and Yellow touchpads simultaneously to enter the Programming Mode.
Used for advancing thru the Setup menu and storing newly-set data.
Used to enter calibration mode.
Used to accelerate setting values when incrementing or decrementing
Not used in programming mode.
FMS-1630L INSTALLATION GUIDE
page 39 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Touchpad OperationMenu Selections
To enter the menu selection depress both GREEN and YELLOW touchpads simultane-
ously. The LCD will display three menu options at one time. Scroll through the menu using
either GREEN or YELLOW touchpads. Use both touchpads simultaneously to enter a
selected menu.
ROOMSELECT
TEST
ExitField Cal
SetPoints1SetPoints2PID Params
IsoMode Alarms
Lock&Rate
To scroll upward pressGREEN
To scroll downward pressYELLOW
Men
uS
ele
cti
on
Touchpads on the FMS-1630L can be used to
scroll through a set up menu. This set up menu
allows users to define the FMS-1630L personality
to match your facilities needs and requirements.
Press Room Select to scroll in one direction or
Test to scroll in the opposite direction.
To enter the Setup menu simultaneously depress
both the Green and Yellow touchpads simul-
taneously. Three menu items will be displayed
at one time. Press either the Green or Yellow
touchpads to scroll the menu up or down. When
the desired menu item is marked with the arrow
press the Red touchpad to select it.
Exit Field Cal SetPoints1
ISOLATION MONITORFMS-1630L
Typical Menu
Selection Display
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 40
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Use the ROOM SELECT key to advance the value in one digit increments.
Use the TEST key to decrease the value in one digit increments.
Pressing both NEGATIVE ISOLATION and ROOM SELECT advances the valve in 10 digit increments.
Pressing both NEGATIVE ISOLATION and TEST decreases the valve in 10 digit increments.
Pressing both NO ISOLATION and ROOM SELECT advances the valve in 100 digit increments.
Pressing both NO ISOLATION and TEST decreases the valve in 100 digit increments.
Press the ALARM SILENCE key to accept the selected value.
Programming/User GuideSetting Values
Using the Keypad to enter values
0000 0001
0001 0000
0001 0011
0011 0001
0001 0101
0101 0001
TEST
ROOMSELECT
C
B
ALARMSILENCE
ROOMSELECT
B TEST
ROOMSELECT
B TEST
FMS-1630L INSTALLATION GUIDE
page 41 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideSetting Password
LEGEND
B
NA
S
T
E
Press both ROOM SELECT and TEST touchpadssimultaneously
Make selections using ROOM SELECT or TEST
Return to Setup menuPossible Selections
Alarm Silence Key
Not Available
TEST touchpad
Enter value1) Use the keypad to select a numeric value.*2) Press the Alarm Silence key to accept the selection.
*Note: See page 42 for specific instruction on using the keypad to set values
The Setup menu item “Lock&Rate” allows access
to “Password” the user to can setup a password for
protection of the FMS-1630L setup from unauthorized
changes. The owner can select a 5 digit numeric value
using ROOM SELECT and TEST keys to create a
unique password. The password will appear on the
display record the password.
CAREFUL: Be sure the responsible person
keeps a record of the password, as it will be needed
to re-enter the programming mode.
1) From the Main display screen, simultaneously
press both the ROOM SELECT and TEST keys
to display the Setup menu.
2) Press the ROOM SELECT key until the
“Lock&Rate” is marked.
3) Press Alarm Silence to enter the “Lock&Rate”
submenus.
4) Press the ROOM SELECT key until the
“Password” is marked.
5) Press Alarm Silence to enter this set-up
sequence.
6) Use the keypad to set the new password value.*
Record the password for future reference.
7) Press Alarm Silence to accept the password and
return to the Setup menu.
8) Use ROOM SELECT to select the next item you
wish to setup, or select “EXIT” then press Alarm
Silence to return to the normal operating display.
*Note: See page 44 for specific instructions on
using the keypad to set values.
Password
Choose Lock&Rate
Choose Password ❋ Password
❋ 00000
B
S
S
E
S
The Setup menu item “Exit” allows the user to
return to the main operating display when use of the
Setup menu is completed.
1) Press the ROOM SELECT to scroll through the
menu and choose “Exit”.
2) Press Alarm Silence to return to the main operat-
ing display.
Exit
S
Return to Main Display
Exiting Setup
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 42
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
The SetPoints1 selection enables the user to adjusted occupied and unoccupied face velocity setpoints on the
controller. The values used in setting the setpoints are Pa x 10. ie 152 represents 15.2 Pa.
1) From the Main display screen, simultaneously
press both the Room Select and Test touchpads
to display the setup menu.
2) Press the Room Select touchpad until
“Setpoints1” is marked, then press the ALARM
SILENCE touchpad.
3) Use the Room Select or Test touchpad to select
“UnOccP-” then press the ALARM SILENCE
touchpad.
4) Use keypad to set Unoccupied face velocity
setpoint. Use the ALARM SILENCE touchpad
to accept the value. The unit will return to the
SetPoints1 Menu.
5) Use the Room Select or Test touchpad to select
“OccP-” then press the ALARM SILENCE touch-
pad.
6) Use keypad to set Occupied face velocity set-
point. Use the ALARM SILENCE touchpad
to accept the value. The unit will return to the
SetPoints1 Menu.
7) Repeat Steps 3 - 6 for each setpoint value see
note.
7) Select the ALARM SILENCE touchpad to move
back to the main menu.
S
E
SetPoints1
S
Select❋ UnOccP-
Enter Setpoint Value
Programming/User GuideOccupied and Unoccupied SetPoints
Select❋ OccP-
S
E Enter Setpoint Value
S
S
Notes:
The values used in setting the setpoints are in inches Wg.
The following entries relate to the following parameters;
UnOccP+ The Unoccupied Positive Isolation Setpoint
UnOccP- The Unoccupied Negative Isolation Setpoint
UnOccP0 The Unoccupied No Isolation Setpoint
OccP+ The Occupied Positive Isolation Setpoint
OccP- The Occupied Negative Isolation Setpoint
OccP0 The Occupied No Isolation Setpoint
LEGEND
B
NA
S
T
E
Press both ROOM SELECT and TEST touchpadssimultaneously
Make selections using ROOM SELECT or TEST
Return to Setup menuPossible Selections
Alarm Silence Key
Not Available
TEST touchpad
Enter value1) Use the keypad to select a numeric value.*2) Press the Alarm Silence key to accept the selection.
*Note: See page 42 for specific instruction on using the keypad to set values
FMS-1630L INSTALLATION GUIDE
page 43 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Temperature Setpoint
The temperature setpoint is an optional value only used
in applications where the FMS will also be providing tem-
perature control for the isolation room.
1) Press both ROOM SELECT and TEST simltane-
ously to enter programming mode.
2) Enter the Password (factory default 00006) using
keypad and press the Alarm Silence key.
3) Scroll to the “Setpoints2” selection, and press the
Alarm Silence key.
4) Scroll to the “Temp” selection, and press the Alarm
Silence key.
5) The unit will now request the temperature setpoint
value.
6) Enter the value (ie 74 degree fahrenheit) using the
ROOM SELECT and TEST buttons, press the
Alarm Silence key to record the value.
SetPoints2
Choose Setpoints2 ❋ Temp
Set Value ❋ 0.00 oF
S
S
E
S
Programming/User GuideSetPoints2 Temperature (Optional)
LEGEND
B
NA
S
T
E
Press both ROOM SELECT and TEST touchpadssimultaneously
Make selections using ROOM SELECT or TEST
Return to Setup menuPossible Selections
Alarm Silence Key
Not Available
TEST touchpad
Enter value1) Use the keypad to select a numeric value.*2) Press the Alarm Silence key to accept the selection.
*Note: See page 42 for specific instruction on using the keypad to set values
The Dead Band Selection allows the controller to avoid moving the actuator while the sensed flow value is within a certain range from the SetPoint. This reduces wear on the actuator and in some cases may add to the stability of the control loop.
1) From the main display screen, simultaneously press both the ROOM SELECT and TEST touchpads to display the
setup menu.
2) Press the ROOM SELECT touchpad until “Setpoints2” is marked, then press the ALARM SILENCE touchpad.
3) Press the ROOM SELECT touchpad until “DeadBand” is marked, then press the ALARM SILENCE touchpad.
4) The value displayed is the total dead band, with half the value being above the SetPoint and half below the SetPoint.
(i.e., a value of 6 means that the controller will not move the actuator for sensed values of SetPoint +/- 3.). Use the
ROOM SELECT or TEST touchpad to set the value. Use the ALARM SILENCE touchpad to accept the entry.
5) Use the ALARM SILENCE touchpad to confirm the selection and exit to the main menu.
S
S
S
SetPoints2
Select❋ DeadBand
EnterDead Band valueE
Dead Band
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 44
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideSetPoints2 Exaccel
The ExAccel selection enables the controller to Accelerate or super charge its response when the airflow/pres-
sure closely approaches its minimum allowable value. The values used in setting the setpoints are FPM.
1) From the Main display screen, simultaneously
press both the ROOM SELECT and TEST
touchpads to display the setup menu.
2) Press the ROOM SELECT touchpad until
“SetPoints” is marked, then press the ALARM
SILENCE touchpad.
3) Press the ROOM SELECT touchpad until
“ExAccelEnab” is marked, then press the ALARM
SILENCE touchpad.
4) Use the touchpad set the enable value “1” if it is
not already enabled. Use the ALARM SILENCE
touchpad to confirm this selection. If ExAccel
is already enabled used the ALARM SILENCE
touchpad to move to the next setting.
5) Press the ROOM SELECT touchpad until
“ExAccelOn” is marked, then press the ALARM
SILENCE touchpad.
6) Use the touchpad to set the value. \
The unit will display the current setting for the value
at which ExAccel is initiated (‘On’). If the sensed
value ever goes below this value, the controller will
excite the motor and drive it to its maximum open
position as fast as it can. To modify this setting
push the ROOM SELECT or TEST touchpad to
increment or decrement this number accordingly.
Use the ALARM SILENCE touchpad to accept
the value. The unit will display the next menu
entry.
7) Press the ROOM SELECT touchpad until
“ExAccelOff” is marked, then press the ALARM
SILENCE touchpad.
8) Use the the touchpad to set the ‘Off’ value for
ExAccel. This is the value, that when the sensed
velocity figure is greater than this, the ExAccel will
disengage and hand control back to the PID con-
trol loop. Adjust this figure in the same manner
as step 6. Use the ALARM SILENCE touchpad
to accept the value. The unit will display the next
menu entry.
9) The ExAccel settings may require iterative tuning,
as incorrect settings and the fast response of the
system may add some instability to the control
loop. As a guide, set the ON value a little above
the minimum allowed velocity or pressure. Set
the OFF value 10% or so above the ON value.
10) Select the ALARM SILENCE touchpad to move
back to the main menu.
S
S
SetPoints2
Select❋ ExAccelEnab
E Enter Value
S
Select❋ ExAccelOn
S
E
S
Enter Value
Select❋ ExAccelOff
S
E
S
Enter Value
LEGEND
B
NA
S
T
E
Press both ROOM SELECT and TEST touchpadssimultaneously
Make selections using ROOM SELECT or TEST
Return to Setup menuPossible Selections
Alarm Silence Key
Not Available
TEST touchpad
Enter value1) Use the keypad to select a numeric value.*2) Press the Alarm Silence key to accept the selection.
*Note: See page 42 for specific instruction on using the keypad to set values
FMS-1630L INSTALLATION GUIDE
page 45 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuidePID Parameters
1) From the Main display screen press both the
ROOM SELECT and TEST keys to display the
Setup menu.
2) Press the ROOM SELECT key until “PID
Params” is marked with an arrow.
3) Press the Alarm Silence key to enter this setup
sequence.
4) Use the ROOM SELECT key to select “PID1-P”.
5) Press the Alarm Silence key to enter this PID1-P
setup.
6) Enter the “PID1-P” value using the keypad a good
starting value is 100. With proportional control,
the difference between an actual condition and
the desired condition (setpoint) is exaggerated by
the value set for this parameter to produce a more
responsive correction. Values too high will cause
oscillation, and values too low may cause the sys-
tem to respond too slowly. Press Alarm Silence
to accept the displayed value.
7) Use the ROOM SELECT key to select “PID1-I”.
8) Press the Alarm Silence key to enter the PID1-I
setup.
9) Enter the “PID1-I” value using the keypad a good
starting value is 100. This parameter is used to
produce increased correction signal value based
on the time the error condition persists. Press
Alarm Silence to accept the displayed value.
10) Use the ROOM SELECT key to select “PID1-
D”.
11) Press the Alarm Silence key to enter the PID1-D
setup.
12) Enter PID-D value using the keypad a good start-
ing value is 10. Derivative is based on the error
rate of change. This produces a “braking” action
as the desired condition is found to be approach-
ing its setpoint. Press Alarm Silence to accept
the displayed value.
13) Use ROOM SELECT to select the next item you
wish to setup, or select “Exit” then press Alarm
Silence to return to the normal operating display.
Select PID1-P
Enter PID1-P Value
Select PID1-I
Enter PID1-I Value
Select PID1-D
Enter PID1-D Value
PID Params
S
S
S
The Setup menu item “PID Params” allows the FMS-1630L user to select parameters that will match the char-
acteristics of the HMS to the isolation room requirements.
Note:
The values used in setting the PID values are dimensionless and act as gains to the appropriate P I and D
components.
The following entries relate to the following parameters;
PID1-P The Proportional Gain for PID loop 1 (main loop)
PID1-I The Integral Gain for PID loop 1 (main loop)
PID1-D The Differential Gain for PID loop 1 (main loop)
PID3-P The Proportional Gain for PID loop 3 (Temperature loop Optional)
PID3-I The Integral Gain for PID loop 3 (Temperature loop Optional)
PID3-D The Differential Gain for PID loop 3 (Temperature loop Optional)
E
S
S
E
S
S
E
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 46
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideSetting Alarms Parameters
1) From the Main display screen, simultaneously
press both the ROOM SELECT and TEST keys
to display the Setup menu.
2 Press the ROOM SELECT key until “Alarms” is
marked.
3 Press Alarm Silence to enter this setup
sequence.
4) Use the ROOM SELECT key to select
“PosAlarmHi” red led.
5) Press Alarm Silence to accept the displayed
selection.
6) Use the keypad to set the high alarm value.
7) Press Alarm Silence to accept the value and
return to the setup screen.
8) Use the ROOM SELECT key to select
“PosAlarmLo” red LED.
9) Use the keypad to set the low alarm value.
10) Press Alarm Silence to accept the value and
return to the setup screen.
11) Use the ROOM SELECT key to select
“PosWarnHi” yellow LED.
12) Use the keypad to set the high caution value.
13) Press Alarm Silence to accept the value and
return to the setup screen.
14) Use the ROOM SELECT key to select
“PosWarnLo” yellow led.
15) Use the keypad to set the low caution value.
16) Press Alarm Silence to accept the value and
return to the setup screen.
17) Repeat steps 4 -16 for the “NegAlarm” values
17) Use ROOM SELECT to select the next item you
wish to setup, or select “Exit” then press Alarm
Silence to return to the normal operating display.
The room’s pressure status is displayed on the front of the unit by green, yellow, and red LED’s. Set up of these
indicator lights is as described in the following steps. The values used in setting the alarms are in inches Wg.
S
S
S
S
Select PosAlarmHi
Enter Value
Select PosAlarmLo
Enter Value
Select POSWarnHi
Enter Value
Select POSWarnLo
Enter Value
S
Alarms
E
E
E
S
S
E
S
LOWCAUTION
LOWALARM
HIGHCAUTION
HIGHALARM
Min Max
Notes:
The values used in setting the alarms are in inches Wg.
The following entries relate to the following parameters;
PosAlarmHi The High Alarm Limit for Positive Isolation
PosAlarmLo The Low Alarm Limit for Positive Isolation
PosWarnHi The High Warning Limit for Positive Isolation
PosWarnLo The Low Warning Limit for Positive Isolation
NegAlarmHi The High Alarm Limit for Negative Isolation
NegAlarmLo The Low Alarm Limit for Negative Isolation
NegWarnHi The High Warning Limit for Negative Isolation
NegWarnLo The Low Warning Limit for Negative Isolation
FMS-1630L INSTALLATION GUIDE
page 47 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideEnable Door Switch
A room door switch having normally-open or normally-closed contacts may be used with the FMS-1630L
to serve as a Door Interlock for room pressure control. An optional door switch (part number SWD-200)
may be purchased from TRIATEK for this specific purpose.
1) From the main operating display, simultaneously press and then release the ROOM SELECT and
TEST touchpad to display the Setup menu.
2) Use the ROOM SELECT touchpad to select “Lock&Rate”, then press the ALARM SILENCE
touchpad.
3) Use the ROOM SELECT touchpad to select “DoorEnab”, then press the ALARM SILENCE touch-
pad.
4) Use the ROOM SELECT touchpad to enter the value “0” disables and “1” enables the doorswitch,
then press the ALARM SILENCE touchpad to accept the entry.
5) Use the ROOM SELECT touchpad to display “DoorDelay”, then press the ALARM SILENCE
touchpad. Note: DoorDelay is the delay in seconds between door closure and control loor resump-
tion.
6) Use the ROOM SELECT or TEST touchpad to enter the delay value “0 - 3600” seocnds, then
press the ALARM SILENCE touchpad to accept the entry.
7) Press the ALARM SILENCE touchpad to accept this setting and return to the Setup menu.
8) Use the ROOM SELECT touchpad to mark “Exit”, then press the ALARM SILENCE touchpad to
exit the Setup menu and return to the main operating display.
Select ❋ DoorEnab
Enter Value 0 Disables 1 Enables
Select Buzzer Reset ❋ DoorDelay
Enter DoorDelay Value ❋ 0 - 3600 Seconds
Lock&Rate
{
{
S
S
S
S
E
E
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 48
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideBuzzer Calibration
1) From the Main display screen, simultaneously
press both the ROOM SELECT and TEST keys
to display the Setup menu.
2) Press ROOM SELECT key until Lock&Rate is
marked with arrow.
3) Press Alarm Silence to enter this menu.
4) Use ROOM SELECT key to select
BuzzerEnable
5) Press Alarm Silence to enter selection.
6) Use ROOM SELECT key to set value 1 to enable
0 to disable
7) Press Alarm Silence to enter selection.
8) Use ROOM SELECT key to select BuzzDelay
9) Press Alarm Silence to enter selection.
10) Use ROOM SELECT key to set value 0 to 60 sec-
onds before the audible alarm sounds. Adjustable
from 0 to 60 seconds in 1 second increments.
11) Press Alarm Silence to enter selection.
12) Use ROOM SELECT to select the next item you
wish to setup, or select “Exit” then press Alarm
Silence to return to the normal operating display.
In this menu the user can enable or disable the buzzer and set the buzzer delay.
Select ❋ BuzzEnab
Enter Value
Select ❋ BuzzDelay
Enter value ❋ 0-60 seconds
S
Lock&Rate
E
S
S
E
S
The DispRate allows the user to set the rate at which the FMS-1630L display updates its LCD screen (i.e. every 3 seconds).
1) From the main display screen, simultaneously
press both the ROOM SELECT and TEST
touchpads to display the setup menu.
2) Press the ROOM SELECT touchpad until
“Lock&Rate” is marked, then press the ALARM
SILENCE touchpad key.
3) Press the ROOM SELECT touchpad until
“DispRate” is marked, then press the ALARM
SILENCE touchpad key.
3) Use the ROOM SELECT or TEST touchpads to
increment or decrement the display rate. The
value indicated is in seconds. (i.e., 2 = update
the displayed value every 2 seconds)
4) When all of the digits have been set, use the
ALARM SILENCE touchpad to enter this setting
and return to the main menu.
Display Refresh Rate
Lock&Rate
S
S
S
E
Select DispRate
Enter value
FMS-1630L INSTALLATION GUIDE
page 49 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideDamper Position
The MaxDmprPos and MinDmpPos selection allows the user to set the travel limits of the isolation room’s exhaust damper. The factory default values of MaxDmprPos “100” and MinDmpPos “0” should only be changed under the specific instructions of qualified personnel. Please consult with our support staff before changing these values.
1) From the Main display screen, simultaneously
press both the ROOM SELECT and TEST
touchpads to display the setup menu.
2) Press the ROOM SELECT touchpad until
“Lock&Rate” is marked, then press the ALARM
SILENCE touchpad.
3) Press the ROOM SELECT touchpad until
“MaxDmprPos” is marked, then press the
ALARM SILENCE touchpad.
4) Use the touchpad set the value 0-100. Use the
ALARM SILENCE touchpad to confirm this
selection and return to the setup menu.
5) Press the ROOM SELECT touchpad until
“MinDmprPos” is marked, then press the ALARM
SILENCE touchpad.
6) Use the touchpad set the value 0-100. Use the
ALARM SILENCE touchpad to confirm this
selection and return to the setup menu.
7) Select the ALARM SILENCE touchpad to move
back to the main menu.
S
S
Lock&Rate
Select❋ MaxDmprPos
E Enter Value
S
Select❋ MinDmpPos
S
E
S
Enter Value
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 50
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideFilter Speed
The filter speed selection allows the user to set the responsiveness of the flow/pressure sensor. The controller utilizes a third order software digital filter to process the incoming signals from the sensor. The user can adjust the cutoff point of this filter by selecting Slow, Normal or Fast from the menu. For systems that are installed in an environment where there is turbulence in the airflow or there are outside air disturbances (air conditioning vents etc) that are unavoidable, a Slow setting will dampen some of these effects, although system responsive-ness will also be slower than usual.
1) From the main display screen, simultaneously press both the GREEN and YELLOW touchpads to display
the setup menu.
2) Press the GREEN touchpad until “14 Filter Speed” is marked, then press the RED touchpad.
3) Using the GREEN or YELLOW touchpads cycle through the Slow, Normal or Fast settings.
4) Use the RED touchpad to confirm this selection and return to the main menu.
S
S
Filter Speed
Choose Speed❋ Slow❋ Normal❋ Fast
FMS-1630L INSTALLATION GUIDE
page 51 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
Programming/User GuideIsolation Mode
The IsoMode selection enables the FMS-1630 to change the room pressuriztion from Positive, Negative and
Neutral no isolation from the programming menu.
1) From the Main display screen, simultaneously
press both the ROOM SELECT and TEST
touchpads to display the setup menu.
2) Press the ROOM SELECT touchpad until
“IsoMode” is marked, then press the ALARM
SILENCE touchpad.
3) Press the ROOM SELECT touchpad until the
desired isolation mode is marked (Pos, Neg, or
Neut), then press the ALARM SILENCE touch-
pad.
4) Select the ALARM SILENCE touchpad to move
back to the main menu.
Notes:
Pos
Positive Isolation, uses the Positive Isolation SetPoint
Neg
Negative Isolation, uses the Negative Isolation
SetPoint
Neut
No Isolation, use the No Isolation SetPoint.
No Status LEDs will be active.
S
S
IsoMode
Select❋ (Pos, Neg or Nuet)
The isolation mode can also be changed using a keyswitch and the isolation buttons on the keypad avoiding the need to enter the programming menu.
1) From the FMS1630L, using the proper key insert
the key and turn to the UNLOCK position to
change the isolation mode.
2) Press the A key for POSITIVE ISOLATION, B
key for NEGATIVE ISOLATION or C key for
NO ISOLATION button to change the isolation
mode.
3) Return the keyswitch to the LOCK postion and
remove the key.
Notes:
Positive Isolation uses the Positive Isolation
SetPoint.
Negative Isolation uses the Negative Isolation
SetPoint
No Isolation uses the No Isolation SetPoint.
No Status LEDs will be active.
Isolation Mode Alternate Entry
INSTALLATION GUIDE FMS-1630L
FMS1630L - 5.0 - 072100 page 52
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.
FMS-1630L INSTALLATION GUIDE
page 53 FMS1630L - 5.0 - 072110
Due to continuous product improvement, TRIATEK reserves the right to change product specifications without notice.