Page 1: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

INSTRUCTION MANUAL

Page 2: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged
Page 3: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

CONTENTS

1.PREPARATIONBEFOREUSE

1-1Beforeoperatingthepipetty

1-2Nameandfunctionsofkeysanddisplay

2.FUNCTIONANDHOWTOUSE

2-1 S (SingleDispensing)Mode

2-2M (MultipleDispensing)Mode

2-3C (VolumeCalibration) Function

2-4SP (SettingSpeed) Function

2-5EC (Electrophoresis)Function

3.TROUBLESHOOTING

4.FORSAFEUSE

5.CONTACTUS

Page 4: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

1.PREPARATIONBEFOREUSE

1-1Beforeoperatingthepipetty1.Wheninsertthebatteryintothepipette,pleasemakesure

thatthebatteryisfullycharged. Forinitialuse,firstchargethebatteryfully.

Battery:A4Nickelhydridebattery(Recommendedbattery:Panasonic,EVOLTABK-4HLD)

2.Thereisnopowerswitchonthebody.Afterbatteryinsertion,flashingdisplayof"S"inthedisplayistheinitialstate.

3.PowerisautomaticallyturnedOFFinthefollowingcases.・ Selectthemode(Initialstate)・ Notoperateforoneminuteduring"S"or"M"or"START”indicatingindisplay.

4.WhenyouturnONthepower,press"PUSH"buttonor"backbutton".

Page 5: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

1.PREPARATIONBEFOREUSE1-2Nameandfunctionsofkeysanddisplay

“PUSHKey”AspirateandDispense

“UP/DOWNKey”Switchthemode/volume

TipEjector

DisplayIndicatethemode/volume

MODE/ENTERKeySelectthemode/volume

BatteryCover

BACKKeySameasPUSHKey

Batterymark

OPERATIONMODES (Single)Aspiratingonetimeanddispensingonetime

− Contain“MixingMode”:mixingdifferenttypesofliquidsM (Multiple)DispensethesamevolumecontinuouslyC (Calibration) UseforvolumecalibrationSP (Settingspeedin5levels)EC (Electrophoresis)Useforelectrophoresis

DispensingvolumeDispensingnumberoftimesIndicatewhendispensingcanbestarted

Page 6: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

Itinitializesautomaticallywhenyouinsertthebattery.StateSisflashingintheinitialscreen.SwitchthemodebypressUPorDOWNKey,selectbyMODE/ENTERKey.

2-1 S (SingleDispensing)ModeThisisabasicoperationforpipetting.Aspiratingonetimeanddispensingonetime.Thisoperationisthesameasforamanualpipette.

1. Switchdispensevolume bypressUPorDOWNKey.2. Selectthenumber bypressMODE/ENTERKey.3. Settingcomplete whenSTARTisindicatedindisplay.4. Starttheaspirating anddispensingbypressing

PUSHKey.(BackKey)

2.FUNCTIONANDHOWTOUSE

MixingModeThisisausefuloperationwhenuniformlymixingdifferenttypesofliquids.Inthismethod,aspiratinganddispensingarerepeated.Thistypeofrepetitiveoperationoftenresultsinfatigue,butwiththispipetteitisautomaticallycarriedoutwiththetouchofoneswitch.

1. MixingwhilethePUSHorBACKKeyishelddown.2. Itwillstoppedbyreleasethekeyandafterdispensedall

※CannotuseMixingModebylongpresswhen“ECMode”

Page 7: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

[Howtoreturntoinitialscreen]1.SwitchthedispensingvolumebyholdingdownUPorDOWNkeywhenindicatedSTARTindisplay.

2.ReturntomodeselectbyholdingdownMODE/ENTERKeywhenindicatedSTARTindisplay.

2-2 M (MultipleDispensing)ModeThisisasuitablefunctiontodispensethesameliquidcontinuously,suchaswhendispensingaliquidonamicroplate,etc.Theoperationconsistsofaspiratingonetimeanddispensingseveraltimes.

1. PresstheUPorDOWNKeytosetdispensingamountforonetime.PressMODE/ENTERKeytoselect.

2. PresstheUPorDOWNKeytosetdispensingcount.3. Settingcomplete whenSTARTisindicatedindisplay.4. Afterdispensing,pressPUSHKey(BACKKey)twiceto

dispensetheremainingliquid.(IntheMultimode,reverseoperationoccursautomatically.)

2.FUNCTIONANDHOWTOUSE

※PressandholdKeymorethan2seconds tofulldispensing.※Thedispensingamountcanbesetwithinthecapacityrangeofthepipette.※Automaticallycalculatethemaximumdispensingnumberof

timesfromdispensingamountandpipettecapacity.※Cannotdispenseallbylongpresswhen“ECMode”

Page 8: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

2-3 C (VolumeCalibration)Function

2.FUNCTIONANDHOWTOUSE

“pipetty”provideuserwithadispensingvolumecalibrationfunction.Usingthisfunction,itiseasytocorrecterrorsduetodifferencesintipsused,etc.Tocalibratethedispensingvolume,completethefollowingprocedure:

2-4 SP (SettingSpeed)Function

Faster Slower

※”AspirateSpeed”and”DispenseSpeed"cannotbeset separately.Botharethesamespeed.

※Whenusedinalong-termcontinuouslyat"Lo"speed,pleasenotethatthereisacasebodyfeelhot.

2-5 EC (Electrophoresis)Functionpipetty willprovideavaliddispensingfunctionforelectrophoresis.Oneofthefeaturesofpipetty,releasetheKeyandtheoperationwillstart.ItcandispensesmoothlybyreleasingtheKeyintargetsitewithoutshakingtipofchip.

Page 9: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

3.TROUBLESHOOTINGDue to repeatedly aspirating and dispensing various liquids the micropipette easily succumbs to damage and contamination. In the event of device failure consult the following table. Request repair if that still doesn’t solve the problem.

Page 10: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

DANGERThisinstrumentisnotanexplosionproofinstrument.Donotusethepipetteinanenvironmentwherethereisariskofexplosion,oruseitforexplosivechemicalsthatmaycauseexplosion.

1.Donotuseforpurposesotherthanliquidsampling.2.Donotuseradioactivematerial,includingtheradiationliquid.3.Donotusetheliquidofbio-hazard.4.Althoughthisproductisexcellentinchemicalresistance,donotuse5.followingchemicalsthatcauseofbodydamage.

(Strongacid,strongalkaline,acetone,andtrichloroethylene)

CAUTIONDonotattempttodisassembleorrepairthepipettebyyourself.TROUBLESHOOTING”whenitappearsthatthepipettehasamechanicalerror.

1.Donotmodifythisproduct.Itwillbethecauseoftheaccident.2.Useourspecificationofthebattery.Ifyouusetheotherbattery,it

willbecauseoftheleak,overheat,ruptureorignite.Andalsoperformancewilldegrade.

3.Whenstoring,pleaseavoidthelocationofthedustyorhighhumidplace.Itwillbethecauseofthefailure.

4.Byrepeatingthechargeanddischarge,thecontinuousnumberofoperationswillbereduced. Ifnecessary,replacetonewbattery.

4.FORSAFEUSE

Precautionsonpipetteuse andhandlingbattery

Page 11: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

MEMO

Page 12: INSTRUCTION MANUALCONTACT US 1. PREPARATION BEFORE USE 1-1 Before operating the pipetty 1. When insert the battery into the pipette, please make sure that the battery is fully charged

Kitaiioka 1-8-25,Morioka-shi,Iwate 020-0857,JapanTEL:+81-19-601-8228FAX:+81-19-601-8227E-mail:[email protected] URL:http://www.icomes.co.jp/

Icomes Lab Co., Ltd.

© 2016 Icomes Lab Co., Ltd. All rights reserved.No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of Icomes Lab Co., Ltd.

The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.