INTEGRATING DISCOVERY & DEVELOPMENT. NON STOP ® THE USE OF THE NEW PHYSIOTEL DIGITAL TELEMETRY IMPLANT (DSI) FOR CARDIOVASCULAR ASSESSMENT IN THE GÖTTINGEN MINIPIG. A PRELIMINARY STUDY S. MILANO 1 , C. BORY 1 , C. DUPUIS 1 , E. CHALENCON 1 , P. LEGE 1 , P. FANT 1 , A. WEBB 2 and D. SINGER 3 1 Ricerca Biosciences SAS, Lyon, France; 2 Chiron Bioscience LTD, Mitcham, UK; 3 Data Science International, St Paul, USA INTRODUCTION The Physiotel Digital telemetry implant is available as a replacement for existing large animal transmitters (D70-PCT /PCTP). This new fully digital system accommodates group housing of animals, contains bi-directional communication between implant and acquisition system for remote programming and has an increased transmission range. It also includes a solid tip negative biopotential lead for use in ECG placement. Ventricular Pressure Catheter Systemic Blood Pressure Catheter Biopotential Lead Solid Tip Lead OBJECTIVE The study was conducted to challenge the product capability in Göttingen minipigs using Dofetilide and Nifedipine as oral cardiovascular reference compounds. The testing included several aspects: - surgery implantation /surgery complication - quality of the signals (ECG, systemic blood pressure, left ventricular pressure) - recording of the telemetry signal in group-housed environment - Analysis of the data MATERIAL AND METHODS Test System : Species/strain: Göttingen minipig; Ellegaard Göttingen Minipigs Aps, Denmark. Number of animals in the study: 3 males + 3 females. Weight : 7 to 13 kgs Surgical procedure : Approximately 48 hours before surgical procedures, animals received antibiotherapy with long acting amoxycilline (30 mg/kg; intramuscular (i.m.). Animals were pre-anaesthetised with an intra-muscular injection of Stresnil ® (azaperone) at 0.1 mL/kg, 30 minutes before anesthesia, anesthetized by mask to be intubated (2 to 5 % isoflurane) and maintained under gaseous anesthesia (isoflurane). The implant was inserted into a muscular pocket in the flank. Left ventricular pressure was monitored with a pressure catheter inserted into the apex of the ventricle and secured with a purse-string suture. Systemic blood pressure was monitored via a pressure catheter inserted into the left femoral artery. The positive ECG electrode was placed subcutaneously over the left thorax near the ventral aspect of the thoracotomy incision and the negative solid tip lead was advanced into the cranial vena cava via the external jugular vein. Post-surgical procedures and recovery period : Broad-spectrum antibiotic prophylaxis with amoxicillin (30 mg/kg/day intramuscularly) was given two times at consecutive intervals of 48 hours starting on the day of surgery. Analgesic prophylaxis with flunixine meglumine (2mg kg/day - 0.04 ml/kg/day; im), 5 times at 24 hour intervals (starting on the day of surgery) + Buprenorphine, 3 times (starting on the day of surgery). Local application of povidone iodine for 7 days following surgery. Treatment : Animals were dosed according a cross over design as follows: RESULTS The transmission range of Physiotel Digital allowed collection of telemetry signals from 6 animals housed in a ~16 m 2 room (3 females group housed) with only four transceivers. FIGURE 1 – Animal room arrangement for PhysioTel Digital telemetry recording in the Göttingen minipig FIGURE 2 – Cardiovascular signals from one PhysioTel Digital telemetered Göttingen minipig FIGURE 3 – Cardiovascular effects of Dofetilde in the PhysioTel Digital telemetered Göttingen minipig FIGURE 4 – Cardiovascular effects of Nifedipine in the PhysioTel Digital telemetered Göttingen minipig FIGURE 5 – Example of individual telemetry signals recorded from two group-housed female animals FIGURE 6 – Jugular vein histological changes after chronic indwelling catheterization for solid tip electrode placement An irregular minimal intimal thickening was observed in the vein wall at the level of the tip of the catheter (H&E stained-section). The low severity suggests the catheterisation into the jugular vein for 20 weeks was well tolerated. CONCLUSION These preliminary results are very encouraging, and the Physiotel Digital implant appears to be a suitable alternative to existing large animal telemetry devices. This new implant offers social housing compatibility thereby improving animal well-being with encoded animal identification and a longer transmission range. * * * * $ $ , *+ () ' " #" () % " $ & ( ! !" !$ ! &' ## % ! " # $ & "$ ! # # " ! ! " # $ ! &' ## % ! " # $ & "$ ! # . 0. 2. 4. 6. /.. /0. /2. /4. /6. )/ ).'45 ).'11 .'3 / 0 1 2 4 02 / ( 1 ( % # . 3.. /... /3.. 0... 03.. 1... 13.. 2... 23.. 3... )/ ).'45 ).'11 .'3 / 0 1 2 4 02 ')/ / ( 1 ( # # *(+ . 0. 2. 4. 6. /.. /0. /2. /4. )/ ).'45 ).'11 .'3 / 0 1 2 4 02 ( / ( 1 ( LVP ABP ECG Group Dose level mg/kg Dose volume mL/kg Concentration mg/mL Vehicle 0 5 0 Dofetilide Low dose (LD) 0.1 5 0.02 Dofetilide High dose (HD) 0.3 5 0.06 Nifedipine Low dose (LD) 1 5 0.2 Nifedipine High dose (HD) 3 5 0.6

Upload: others

Post on 19-May-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: INTEGRATING DISCOVERY & THE USE OF THE NEW PHYSIOTEL ... · the use of the new physiotel™ digital telemetry implant (dsi) for cardiovascular assessment in the gÖttingen minipig

INTEGRATING DISCOVERY & DEVELOPMENT. NON STOP ®

THE USE OF THE NEW PHYSIOTEL™ DIGITAL TELEMETRY IMPLANT (DSI) FOR CARDIOVASCULAR

ASSESSMENT IN THE GÖTTINGEN MINIPIG.A PRELIMINARY STUDY

S. MILANO 1, C. BORY 1, C. DUPUIS 1, E. CHALENCON 1, P. LEGE 1, P. FANT 1, A. WEBB 2 and D. SINGER 31 Ricerca Biosciences SAS, Lyon, France; 2 Chiron Bioscience LTD, Mitcham, UK; 3 Data Science International, St Paul, USA

INTRODUCTIONThe Physiotel™ Digital telemetry implant is available as a replacement for existing large animal transmitters (D70-PCT /PCTP). This new fully digital system accommodates group housing of animals, contains bi-directional communication between implant and acquisition system for remote programming and has an increased transmission range. It also includes a solid tip negative biopotential lead for use in ECG placement.

Ventricular Pressure Catheter

Systemic Blood Pressure Catheter

Biopotential Lead

Solid Tip Lead

OBJECTIVEThe study was conducted to challenge the product capability in Göttingen minipigs using Dofetilide and Nifedipine as oral cardiovascular reference compounds.The testing included several aspects:- surgery implantation /surgery complication- quality of the signals (ECG, systemic blood pressure, left ventricular pressure)- recording of the telemetry signal in group-housed environment- Analysis of the data

MATERIAL AND METHODSTest System : Species/strain: Göttingen minipig; Ellegaard Göttingen Minipigs Aps, Denmark. Number of animals in the study: 3 males + 3 females. Weight : 7 to 13 kgsSurgical procedure : Approximately 48 hours before surgical procedures, animals received antibiotherapy with long acting amoxycilline (30 mg/kg; intramuscular (i.m.).Animals were pre-anaesthetised with an intra-muscular injection of Stresnil® (azaperone) at 0.1 mL/kg, 30 minutes before anesthesia, anesthetized by mask to be intubated (2 to 5 % isofl urane) and maintained under gaseous anesthesia (isofl urane). The implant was inserted into a muscular pocket in the fl ank. Left ventricular pressure was monitored with a pressure catheter inserted into the apex of the ventricle and secured with a purse-string suture. Systemic blood pressure was monitored via a pressure catheter inserted into the left femoral artery.

The positive ECG electrode was placed subcutaneously over the left thorax near the ventral aspect of the thoracotomy incision and the negative solid tip lead was advanced into the cranial vena cava via the external jugular vein.

Post-surgical procedures and recovery period : Broad-spectrum antibiotic prophylaxis with amoxicillin (30 mg/kg/day intramuscularly) was given two times at consecutive intervals of 48 hours starting on the day of surgery. Analgesic prophylaxis with fl unixine meglumine (2mg kg/day - 0.04 ml/kg/day; im), 5 times at 24 hour intervals (starting on the day of surgery) + Buprenorphine, 3 times (starting on the day of surgery). Local application of povidone iodine for 7 days following surgery.

Treatment : Animals were dosed according a cross over design as follows:

RESULTSThe transmission range of Physiotel™ Digital allowed collection of telemetry signals from 6 animals housed in a ~16 m2 room (3 females group housed) with only four transceivers.

FIGURE 1 – Animal room arrangement

for PhysioTel™ Digital telemetry

recording in the Göttingen minipig

FIGURE 2 – Cardiovascular signals from one PhysioTel™ Digital telemetered

Göttingen minipig

FIGURE 3 – Cardiovascular effects of Dofetilde in the PhysioTel™ Digital telemetered

Göttingen minipig

FIGURE 4 – Cardiovascular effects of Nifedipine in the PhysioTel™ Digital telemetered

Göttingen minipig

FIGURE 5 – Example

of individual telemetry

signals recorded from

two group-housed female

animals

FIGURE 6 – Jugular vein histological changes after

chronic indwelling catheterization for solid tip

electrode placement

An irregular minimal intimal thickening was observed in the vein wall at the level of the tip of the catheter (H&E stained-section). The low severity suggests the catheterisation into the jugular vein for 20 weeks was well tolerated.

CONCLUSIONThese preliminary results are very encouraging, and the Physiotel™ Digital implant appears to be a suitable alternative to existing large animal telemetry devices. This new implant offers social housing compatibility thereby improving animal well-being with encoded animal identifi cation and a longer transmission range.

LVP

ABP

ECG

Group Dose level mg/kg

Dose volume mL/kg

Concentration mg/mL

Vehicle 0 5 0

Dofetilide Low dose (LD) 0.1 5 0.02

Dofetilide High dose (HD) 0.3 5 0.06

Nifedipine Low dose (LD) 1 5 0.2

Nifedipine High dose (HD) 3 5 0.6