intestinal luminal lactate in shock-marker for altered epithelial cellular metabolism?

1
both in vitro and in vivo, in order to char- acterize the method and its limitations [1, 3, 4, 5]. Furthermore, we have always used it under appropriately controlled experi- mental conditions. It is therefore clear that while we can make no claims concerning the absolute luminal lactate concentrations, the changes observed in cardiac tamponade model are, indeed, still valid. It is highly possible that luminal microdialysate lactate concentrations were lower than the ‘real’ concentration in the surrounding milieu. In other words, the ‘recovery’ was not 100% in the gut lumen as opposed to within blood stream validation where the relative recovery was close to 100% [1]. Low re- covery is now easily overcome with lower dialysate flow rate (0.2 µl/min) and highly developed analyzers which are specifically designed for microdialysis. The sample volumes needed may be as low as 2–4 µl for lactate and still the sampling frequency remains reasonably short for both research and clinical purposes (10–20 min) as op- posed to 4 h of equilibration in the paper by Due and Perner. They quite correctly discuss the problems related to long equili- bration time in a dynamic disease state. We compared our results from the car- diac tamponade study to our earlier experi- ments, where selective occlusion of the su- perior mesenteric artery (SMA) was ap- plied [1, 4]. At comparably low SMA blood flow levels with the identical micro- dialysis method, intestinal luminal lactate was markedly higher in selective intestinal ischemia as opposed to low systemic blood flow. Thereby we suggested that the small intestine is better adapted to low blood flow during low systemic blood flow than isolated, selectively reduced, regional blood flow [3]. In addition, systemic hy- perlactatemia could largely explain the lu- minal lactate release observed [1, 3]. Final- ly, generally spoken, the terminology has to be precise: recovery should be reserved to indicate, in both microdialysis and equi- librium dialysis, the percentage of concen- tration as compared to the surrounding mi- lieu. Sensitivity and specificity for cellular dysoxia, ischemia or hypoperfusion need to be defined by formal receiver operating curve (ROC) analysis. While we find the result of Due and Perner intriguing indeed, we would be interested to see the valida- tion of the method including, for example, occlusive and non-occlusive ischemia and ischemic and non-ischemic hyperlactate- mia under controlled conditions. References 1. Tenhunen JJ, Kosunen H, Heino A, Tuomisto L, Alhava E, Takala J (1999) Intestinal luminal microdialysis- a new approach to assess mucosal dysoxia. Anesthesiology 91:1807–1815 2. Due VL, Bonde J, Espersen K, Jensen TH, Perner A (2002) Lactic acidosis in the rectal lumen of patients with septic shock measured by luminal equilibrium dialysis. Br J Anaesth: (in press) 3. Tenhunen JJ, Jakob SM, Ruokonen E, Takala J (2002) Jejunal luminal microdi- alysate lactate in cardiac tamponade – effect of low systemic blood flow on gut mucosa. Intensive Care Med 28:953–962 4. Tenhunen JJ, Jakob SM, Takala JA (2001) Gut luminal lactate release dur- ing gradual intestinal ischemia. Inten- sive Care Med 27:1916–1922 5. Tenhunen JJ, Uusaro A, Ruokonen E (2001) Mucosal dysoxia in endotoxin shock can be detected by luminal micro- dialysis but not by regional L/P ratios (abstract). Intensive Care Med 27:S182 J.J. Tenhunen ( ) Division of Intensive Care, Department of Anesthesiology and Intensive Care, Kuopio University Hospital, P.O. Box 1777, 70211 Kuopio, Finland e-mail: [email protected] J. Takala Department of Intensive Care Medicine, University Hospital of Bern, 3010 Bern, Switzerland Intensive Care Med (2003) 29:336 DOI 10.1007/s00134-002-1599-5 CORRESPONDENCE Jyrki J. Tenhunen Jukka Takala Intestinal luminal lactate in shock-marker for altered epithelial cellular metabolism? Received: 29 October 2002 Accepted: 5 November 2002 Published online: 15 January 2003 © Springer-Verlag 2003 Sir: We fully agree with Drs. Due and Per- ner that new methods for monitoring (in- testinal) tissue-specific metabolism in criti- cal illness are warranted. Intestinal luminal microdialysis [1] or dialysis membrane [2] based methods may be such approaches. Due and Perner suggest that the rela- tively low luminal lactate concentrations observed by us in experimental cardiac tamponade could be related to the microdi- alysis method [3]. They base their concern on their own observation on high rectal lactate concentrations in patients with sep- tic shock as measured by a different rectal dialysis method [2]. We are well aware that results obtained by microdialysis or other dialysis tech- niques are highly variable depending on the exact method used, since multiple fac- tors affect the actual concentration of the substance measured. These include, for ex- ample: membrane or capillary characteris- tics, dialysate flow rate, equilibration time, the dialysate itself etc. In addition, a highly important factor is the tissue per se and its metabolic and perfusion dynamics. We are therefore very reluctant to draw any com- parison to our results based on the observa- tions by Due and Perner, which were ob- tained under completely different circum- stances with completely different metho- dology. As far as our method is concerned, we have performed a careful validation,

Upload: jukka-takala

Post on 25-Aug-2016

215 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Intestinal luminal lactate in shock-marker for altered epithelial cellular metabolism?

both in vitro and in vivo, in order to char-acterize the method and its limitations [1,3, 4, 5]. Furthermore, we have always usedit under appropriately controlled experi-mental conditions. It is therefore clear thatwhile we can make no claims concerningthe absolute luminal lactate concentrations,the changes observed in cardiac tamponademodel are, indeed, still valid. It is highlypossible that luminal microdialysate lactateconcentrations were lower than the ‘real’concentration in the surrounding milieu. Inother words, the ‘recovery’ was not 100%in the gut lumen as opposed to withinblood stream validation where the relativerecovery was close to 100% [1]. Low re-covery is now easily overcome with lowerdialysate flow rate (0.2 µl/min) and highlydeveloped analyzers which are specificallydesigned for microdialysis. The samplevolumes needed may be as low as 2–4 µlfor lactate and still the sampling frequencyremains reasonably short for both researchand clinical purposes (10–20 min) as op-posed to 4 h of equilibration in the paperby Due and Perner. They quite correctlydiscuss the problems related to long equili-bration time in a dynamic disease state.

We compared our results from the car-diac tamponade study to our earlier experi-ments, where selective occlusion of the su-perior mesenteric artery (SMA) was ap-plied [1, 4]. At comparably low SMAblood flow levels with the identical micro-dialysis method, intestinal luminal lactatewas markedly higher in selective intestinalischemia as opposed to low systemic bloodflow. Thereby we suggested that the smallintestine is better adapted to low bloodflow during low systemic blood flow thanisolated, selectively reduced, regionalblood flow [3]. In addition, systemic hy-perlactatemia could largely explain the lu-minal lactate release observed [1, 3]. Final-ly, generally spoken, the terminology hasto be precise: recovery should be reservedto indicate, in both microdialysis and equi-librium dialysis, the percentage of concen-tration as compared to the surrounding mi-lieu. Sensitivity and specificity for cellulardysoxia, ischemia or hypoperfusion needto be defined by formal receiver operatingcurve (ROC) analysis. While we find the

result of Due and Perner intriguing indeed,we would be interested to see the valida-tion of the method including, for example,occlusive and non-occlusive ischemia andischemic and non-ischemic hyperlactate-mia under controlled conditions.

References

1. Tenhunen JJ, Kosunen H, Heino A, Tuomisto L, Alhava E, Takala J (1999)Intestinal luminal microdialysis- a newapproach to assess mucosal dysoxia.Anesthesiology 91:1807–1815

2. Due VL, Bonde J, Espersen K, JensenTH, Perner A (2002) Lactic acidosis inthe rectal lumen of patients with septicshock measured by luminal equilibriumdialysis. Br J Anaesth: (in press)

3. Tenhunen JJ, Jakob SM, Ruokonen E,Takala J (2002) Jejunal luminal microdi-alysate lactate in cardiac tamponade –effect of low systemic blood flow on gutmucosa. Intensive Care Med28:953–962

4. Tenhunen JJ, Jakob SM, Takala JA(2001) Gut luminal lactate release dur-ing gradual intestinal ischemia. Inten-sive Care Med 27:1916–1922

5. Tenhunen JJ, Uusaro A, Ruokonen E(2001) Mucosal dysoxia in endotoxinshock can be detected by luminal micro-dialysis but not by regional L/P ratios(abstract). Intensive Care Med 27:S182

J.J. Tenhunen (✉)Division of Intensive Care, Department of Anesthesiology and Intensive Care,Kuopio University Hospital,P.O. Box 1777, 70211 Kuopio, Finlande-mail: [email protected]

J. TakalaDepartment of Intensive Care Medicine,University Hospital of Bern,3010 Bern, Switzerland

Intensive Care Med (2003) 29:336DOI 10.1007/s00134-002-1599-5 C O R R E S P O N D E N C E

Jyrki J. TenhunenJukka Takala

Intestinal luminal lactate in shock-marker for alteredepithelial cellular metabolism?

Received: 29 October 2002Accepted: 5 November 2002Published online: 15 January 2003© Springer-Verlag 2003

Sir: We fully agree with Drs. Due and Per-ner that new methods for monitoring (in-testinal) tissue-specific metabolism in criti-cal illness are warranted. Intestinal luminalmicrodialysis [1] or dialysis membrane [2]based methods may be such approaches.

Due and Perner suggest that the rela-tively low luminal lactate concentrationsobserved by us in experimental cardiactamponade could be related to the microdi-alysis method [3]. They base their concernon their own observation on high rectallactate concentrations in patients with sep-tic shock as measured by a different rectaldialysis method [2].

We are well aware that results obtainedby microdialysis or other dialysis tech-niques are highly variable depending onthe exact method used, since multiple fac-tors affect the actual concentration of thesubstance measured. These include, for ex-ample: membrane or capillary characteris-tics, dialysate flow rate, equilibration time,the dialysate itself etc. In addition, a highlyimportant factor is the tissue per se and itsmetabolic and perfusion dynamics. We aretherefore very reluctant to draw any com-parison to our results based on the observa-tions by Due and Perner, which were ob-tained under completely different circum-stances with completely different metho-dology. As far as our method is concerned,we have performed a careful validation,

Verwendete Distiller 5.0.x Joboptions
Dieser Report wurde automatisch mit Hilfe der Adobe Acrobat Distiller Erweiterung "Distiller Secrets v1.0.5" der IMPRESSED GmbH erstellt. Sie koennen diese Startup-Datei für die Distiller Versionen 4.0.5 und 5.0.x kostenlos unter http://www.impressed.de herunterladen. ALLGEMEIN ---------------------------------------- Dateioptionen: Kompatibilität: PDF 1.2 Für schnelle Web-Anzeige optimieren: Ja Piktogramme einbetten: Ja Seiten automatisch drehen: Nein Seiten von: 1 Seiten bis: Alle Seiten Bund: Links Auflösung: [ 600 600 ] dpi Papierformat: [ 595 785 ] Punkt KOMPRIMIERUNG ---------------------------------------- Farbbilder: Downsampling: Ja Berechnungsmethode: Bikubische Neuberechnung Downsample-Auflösung: 150 dpi Downsampling für Bilder über: 225 dpi Komprimieren: Ja Automatische Bestimmung der Komprimierungsart: Ja JPEG-Qualität: Mittel Bitanzahl pro Pixel: Wie Original Bit Graustufenbilder: Downsampling: Ja Berechnungsmethode: Bikubische Neuberechnung Downsample-Auflösung: 150 dpi Downsampling für Bilder über: 225 dpi Komprimieren: Ja Automatische Bestimmung der Komprimierungsart: Ja JPEG-Qualität: Mittel Bitanzahl pro Pixel: Wie Original Bit Schwarzweiß-Bilder: Downsampling: Ja Berechnungsmethode: Bikubische Neuberechnung Downsample-Auflösung: 600 dpi Downsampling für Bilder über: 900 dpi Komprimieren: Ja Komprimierungsart: CCITT CCITT-Gruppe: 4 Graustufen glätten: Nein Text und Vektorgrafiken komprimieren: Ja SCHRIFTEN ---------------------------------------- Alle Schriften einbetten: Ja Untergruppen aller eingebetteten Schriften: Nein Wenn Einbetten fehlschlägt: Warnen und weiter Einbetten: Immer einbetten: [ ] Nie einbetten: [ ] FARBE(N) ---------------------------------------- Farbmanagement: Farbumrechnungsmethode: Alle Farben zu sRGB konvertieren Methode: Standard Arbeitsbereiche: Graustufen ICC-Profil: RGB ICC-Profil: sRGB IEC61966-2.1 CMYK ICC-Profil: U.S. Web Coated (SWOP) v2 Geräteabhängige Daten: Einstellungen für Überdrucken beibehalten: Ja Unterfarbreduktion und Schwarzaufbau beibehalten: Ja Transferfunktionen: Anwenden Rastereinstellungen beibehalten: Ja ERWEITERT ---------------------------------------- Optionen: Prolog/Epilog verwenden: Nein PostScript-Datei darf Einstellungen überschreiben: Ja Level 2 copypage-Semantik beibehalten: Ja Portable Job Ticket in PDF-Datei speichern: Nein Illustrator-Überdruckmodus: Ja Farbverläufe zu weichen Nuancen konvertieren: Nein ASCII-Format: Nein Document Structuring Conventions (DSC): DSC-Kommentare verarbeiten: Nein ANDERE ---------------------------------------- Distiller-Kern Version: 5000 ZIP-Komprimierung verwenden: Ja Optimierungen deaktivieren: Nein Bildspeicher: 524288 Byte Farbbilder glätten: Nein Graustufenbilder glätten: Nein Bilder (< 257 Farben) in indizierten Farbraum konvertieren: Ja sRGB ICC-Profil: sRGB IEC61966-2.1 ENDE DES REPORTS ---------------------------------------- IMPRESSED GmbH Bahrenfelder Chaussee 49 22761 Hamburg, Germany Tel. +49 40 897189-0 Fax +49 40 897189-71 Email: [email protected] Web: www.impressed.de
Adobe Acrobat Distiller 5.0.x Joboption Datei
<< /ColorSettingsFile () /AntiAliasMonoImages false /CannotEmbedFontPolicy /Warning /ParseDSCComments false /DoThumbnails true /CompressPages true /CalRGBProfile (sRGB IEC61966-2.1) /MaxSubsetPct 100 /EncodeColorImages true /GrayImageFilter /DCTEncode /Optimize true /ParseDSCCommentsForDocInfo false /EmitDSCWarnings false /CalGrayProfile () /NeverEmbed [ ] /GrayImageDownsampleThreshold 1.5 /UsePrologue false /GrayImageDict << /QFactor 0.9 /Blend 1 /HSamples [ 2 1 1 2 ] /VSamples [ 2 1 1 2 ] >> /AutoFilterColorImages true /sRGBProfile (sRGB IEC61966-2.1) /ColorImageDepth -1 /PreserveOverprintSettings true /AutoRotatePages /None /UCRandBGInfo /Preserve /EmbedAllFonts true /CompatibilityLevel 1.2 /StartPage 1 /AntiAliasColorImages false /CreateJobTicket false /ConvertImagesToIndexed true /ColorImageDownsampleType /Bicubic /ColorImageDownsampleThreshold 1.5 /MonoImageDownsampleType /Bicubic /DetectBlends false /GrayImageDownsampleType /Bicubic /PreserveEPSInfo false /GrayACSImageDict << /VSamples [ 2 1 1 2 ] /QFactor 0.76 /Blend 1 /HSamples [ 2 1 1 2 ] /ColorTransform 1 >> /ColorACSImageDict << /VSamples [ 2 1 1 2 ] /QFactor 0.76 /Blend 1 /HSamples [ 2 1 1 2 ] /ColorTransform 1 >> /PreserveCopyPage true /EncodeMonoImages true /ColorConversionStrategy /sRGB /PreserveOPIComments false /AntiAliasGrayImages false /GrayImageDepth -1 /ColorImageResolution 150 /EndPage -1 /AutoPositionEPSFiles false /MonoImageDepth -1 /TransferFunctionInfo /Apply /EncodeGrayImages true /DownsampleGrayImages true /DownsampleMonoImages true /DownsampleColorImages true /MonoImageDownsampleThreshold 1.5 /MonoImageDict << /K -1 >> /Binding /Left /CalCMYKProfile (U.S. Web Coated (SWOP) v2) /MonoImageResolution 600 /AutoFilterGrayImages true /AlwaysEmbed [ ] /ImageMemory 524288 /SubsetFonts false /DefaultRenderingIntent /Default /OPM 1 /MonoImageFilter /CCITTFaxEncode /GrayImageResolution 150 /ColorImageFilter /DCTEncode /PreserveHalftoneInfo true /ColorImageDict << /QFactor 0.9 /Blend 1 /HSamples [ 2 1 1 2 ] /VSamples [ 2 1 1 2 ] >> /ASCII85EncodePages false /LockDistillerParams false >> setdistillerparams << /PageSize [ 595.276 841.890 ] /HWResolution [ 600 600 ] >> setpagedevice