introducing the agilent g6410a triple quadrupole mass...
TRANSCRIPT
WW Seminar Tour
Spring 2006
Introducing the Agilent G6410A Triple QuadrupoleMass Spectrometer
Michael Zumwalt
Quad Mass Filter (Q3)Quad Mass Filter (Q1)
Collision Cell
Ion source
WW Seminar Tour
Spring 2006
Overview of QQQ HardwareQQQ Mechanical Layout
a
Leveraged w/ Q & QTOF
Common w/ QTOF
Quad Mass Filter (Q3)
HED Detector
Quad Mass Filter (Q1)
Collision CellLens 1 and 2
Rough Pump
Turbo Pump
Turbo Pump
Turbo Pump
Quad Mass Filter (Q1)
Collision CellLens 1 and 2
WW Seminar Tour
Spring 2006
MRM (Multiple Reaction Monitoring)
170 210 250 290
210
222
268 280165
Quad Mass Filter (Q3)Quad Mass Filter (Q1)
Collision Cell
Spectrum with backgroundions (from ESI)
Q1 lets onlytarget ion 210 pass through
190 210
210
Collision cell breaks ion 210 apart
150 170 190 210
210158
191
Q3 monitors onlycharacteristic fragments 158 and 191 from ion 210 for quant and qual.
160
158
190
191
no chemical background
WW Seminar Tour
Spring 2006
Autotune and Calibration SystemBuilt-in calibrant delivery system is unique to Agilent
Built-in calibrantdelivery systemused for fastauto-calibration
calibrant solutions
WW Seminar Tour
Spring 2006
Calibrant delivered to MS via the API Interface
Calibrant Delivery System
to waste
from calibrant bottles
nitrogen
to MS spray chamber
from LC pump
Valve can also be utilized to divert
flow from the HPLC via software
WW Seminar Tour
Spring 2006
The analyzer story - minimizing noise
1. Agilent's outstanding Patented orthogonal spray sources maximize ionization while minimizing solvent and matrix noise
2. Counter-Current Drying gas Cold Dielectric capillary...enhances desolvation while minimizing chemical noise.
3. RF Octopole guide……High efficiency ion capture while optimizing wide mass bandwidth ion transmission
4. L2 RF…….greatly enhances high mass ion transmission. Patented!
5. Quad 1….hyperbolic quads optimize ion transmission and spectral resolution
6. RF quadrupole segment…..enhances ion transmission into and out of the collision cell
7. Collision cell….high pressure collision cell with linear acceleration optimizes MS.MS fragmentation while eliminating cross-talk even at very low dwell times. Small diameter high frequency hexapole assembly assists with capturing and focusing fragmented ions.
8. Quad 2 ….hyperbolic quads optimize ion transmission and spectral resolution
9. Off axis matching dual High Energy Dynode detector with Log amp signal compression……..permits high gain with rapid polarity switching, long life and low noise. Off axis design allows neutrals to pass without hitting detector.
10. Multiplier has extremely long life since ions never touch the surface, only electrons.
a
Rough Pump Edwards E2M28 Single 3 stage Edwards
Turbo Pump
2
3 4 5 6 7 810
6
9
1
WW Seminar Tour
Spring 2006
Agilent G1978A Multimode Source
Liquid enters the grounded nebulizer
A charged aerosol is made in the ESI Zone
The aerosol is dried by IR lamps
Neutral analytes and ESI charged analytes pass through the APCI Zone
ESI and APCI ions enter the capillary
Capillary
HPLC inlet
Nebulizer
Drying gas
Corona needle
ESI Zone
APCI ZoneThermalcontainer
WW Seminar Tour
Spring 2006
Agilent Multimode Source - Compound Detection CompoundIon Polarity (+) (-) (+) (-) (+) (-)
Acetazolamide 0 + 0 + 0 +Butyl 4-aminobenzoate + 0 + 0 + 0Cortisone + + + 0 + +Gemfibrozil 0 + 0 + 0 +Hexahydro-...-dione + 0 + 0 + 0Hydroflumethiazide 0 + 0 + 0 +Indole + 0 0 0 + 0Iodipamide 0 0 0 + + +Labetalol + + + + + +Lidocaine + 0 + 0 + 0Morin + + + + + +Paclitaxel 0 0 + 0 + 0Phenylbutazone + + + + + +Procainamide + 0 + 0 + 0Progesterone + 0 + 0 + 0Sulfamethoxazole + + + + + +Tolazamide + + + + + +Uracil 0 + 0 0 0 +Detected by Polarity
67% 56% 67% 50% 78% 61%
Detected by Source
89% 89% 100%
Dedicated APCI Dedicated ESI Simultaneous
WW Seminar Tour
Spring 2006
Environmental Compounds, Alternating Positive/Negative Mixed Mode Ionization
5 ng each analyte; 2.1 x 150 mm Zorbax XDB-C18, 3.5 µ, 60 °C, water/methanol gradient (3-90% methanol) with 1mM ammonium acetate, scan mode m/z 130-330; sample dissolved in 80:20 water/methanol containing 1% acetic acid. * denotes impurity
Positive mixed mode TIC,primarily [M+H]+ and [M+Na]+
min5 10 15 20 25 30 35
DiquatParaquat(respond in ESI only)
Oxamyl Methomyl
Fenuron
Monuron
*
Simazine
Propoxur
AtratonPropham
Fluometuron
Carbaryl
AminocarbCarbofuran
TerbutylazineSiduron
Methiocarb PropazineAmetrynSiduron
Secbumeton
LinuronPrometon
SimetrynAtrazine
Diuron
Prometryn
TerbutrynNeburon
ChlorprophamBarban
Mexacarbate
Negative mixed mode TIC,primarily [M-H]- and [M+acetate]-
min5 10 15 20 25 30 35
4-Nitrophenol(responds best in APCI)
Dicamba2,4-D
Monuron2,4,5-T** * *
*Fluometuron
Diuron
Linuron
NeburonSiduron
WW Seminar Tour
Spring 2006
Variables include:
• Number of poles (i.e. quad, hex, octo)
• Inscribed diameter (R0)
• Drive Frequency
Evaluation included:
• Theoretical modeling
– Calculation, Simulations
• Experimental results
Broad Mass Range Transmission and Transmission Efficiency – Why a Hexapole?
Mass Range Transmission (octo > hex > quad)
0
0.2
0.40.6
0.8
1
1.2
0 500 1000 1500Mass (m/z)
Tran
smis
sion
Quadrupole Hexapole Octopole
Ion focusing: quad > hex > octo
WW Seminar Tour
Spring 2006
Collision Cell Design and Electrical Drive Specifics
Acceleration Potential
to 10 V
RF Voltage = 10 - 550 v
• Hexapole construction
• 2R0 = 4.5 mm
• Length = 150 mm
• Experimental press. range = 0.1 – 20mTorr
• In coming ion energy range = 0 – 250eV
• Accelerating potential range = 0 – 10V
• RF Drive voltage range 10 – 550V
WW Seminar Tour
Spring 2006
Beam Turn-off Characteristics
10
100
1000
10000
100000
-500 0 500 1000 1500 2000
microseconds
Arb
. Uni
ts
mz922 mz118
Collision Cell Clearing Profile
Collision gas = 7mTorr 0 V collision energy 5 V Applied Axial Potential
600µsec
350µsec
WW Seminar Tour
Spring 2006
Fast MRM Selection
Transition 1 Transition 2Q1 = 622m/z Q1 = 600m/z (no signal)CE = 30V CE = 30VAAP = 7V AAP = 7VQ3 = 540m/z Q3 = 540m/z5msec settling 5msec settling5msec sampling 5msec sampling
3x10
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
Abundance vs. Acquisition Time (Min)0.029 0.03 0.031 0.032 0.033 0.034 0.035 0.036 0.037 0.038 0.039 0.04 0.041 0.042 0.043 0.044 0.045
+EIC (540.0000 m/z) MRM (600.0000 m/z); from crosstalk-03.d
Sample:Tuning Soln,Infused
Transition 1
Transition 2
WW Seminar Tour
Spring 2006
Dwell Time Evaluation 200 pg Alprazolam 100, 20, 5 ms dwell times
2.250.740.61% RSD
132021360514860Area
5 ms20 ms100 msDwell
5 ms dwell
WW Seminar Tour
Spring 2006
The Analyzer Story - Detector aspects:
9. Off axis matching dual High Energy Dynode detector with Log amp signal compression……..permits high gain with rapid polarity switching, long life and low noise. Off axis design allows neutrals to pass without hitting detector.
10. Multiplier has extremely long life since ions never touch the surface, only electrons.
a
Rough Pump Edwards E2M28 Single 3 stage Edwards
Turbo Pump
2
3 4 5 6 7 810
6
9
1
WW Seminar Tour
Spring 2006
The Analyzer Story - Pump aspects:
Newly designed manifold with a new single 3-stage turbomolecular pump, backed by a single mechanical (rough) pump.
3-staged partitioning removes the need for two turbomolecular pumps.
a
Rough Pump Edwards E2M28 Single 3 stage Edwards
Turbo Pump
2
3 4 5 6 7 810
6
9
1
WW Seminar Tour
Spring 2006
Reserpine SensitivitySpecification: 500 fg with S/N 20:1, %RSD < 5
100pg
500pg
50pg
10pg5pg1pg0.5pg0.1pgBlank
1pg
0.5pg
0.1pgBlank
WW Seminar Tour
Spring 2006
Reserpine Dilution Series (in mobile phase)Demonstration of Sensitivity via a Standard
min10 20 30 40 50 60
0
200
400
600
800
1000
1200
4.18
2 6.32
0
8.36
6
10.4
65
12.4
69
16.7
18
18.7
43
20.8
31
22.8
79
24.9
78
29.1
42
31.1
96
33.2
98 35.3
59 37.4
28
41.6
28 43.6
85
45.7
64
47.8
34
49.9
07
54.0
91 56.1
60
58.2
29
60.2
91
62.3
59
100fg200fg
500fg
1000fg
2000fg
0
2000
4000
6000
8000
10000
12000
0 500 1000 1500 2000 2500
Sample Amount (fg)
Peak
Are
a (a
rb. u
nits
)
r2 = 0.987
Reserpine (609>195) EIC, acquiring 3 transitions
WW Seminar Tour
Spring 2006
Alprazolam Calibration Results1pg – 10 ng, Positive ESI Mode
1 pg
5 pg
10 pg
5 ng
10 ng
2 ng
1 ng
0
100000
200000
300000
400000
500000
600000
700000
800000
900000
0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000
Abun
danc
e
Amount (pg)
Alprazolam Linearity
Y=77.2x +1593.4R2=0.9974
WW Seminar Tour
Spring 2006
New Agilent MassHunter Workstation Software
Instrument Control
Real-time monitoring
Method set-up
Autotune
Qualitative Analysis
Chromatographic results
Spectral results
Find compounds
Quantitative Analysis
User filters
Compound results
Calibration curve
WW Seminar Tour
Spring 2006
Full Integration: The Lab, Business Applications, OpenLAB, the Company Network
IntranetIntranet
InternetInternet
GC
CE
LC
Lab-on-a Chip
GC/MS
MS WorkstationICP-MS
UV-vis
ELNPDF
Images
DrawingsLIMSResearch
MS Office LegacyData
The Lab Business Applications
Agilent OpenLAB (Laboratory Informatics)
WW Seminar Tour
Spring 2006
Data Acquisition
Instrumentstatus
Data display
Acquisitionsettings
Sequence list
WW Seminar Tour
Spring 2006
Auto or Manual Tune
WW Seminar Tour
Spring 2006
Triple (QQQ) Reporting (Excel) - Tune
WW Seminar Tour
Spring 2006
• Macros in Excel Adding• Add tables and graphics
via tool buttons• Drag and drop
additional columnsfrom XML results
• Formatting of Fontand numbers in Excel
• Print Preview in Excel
Reporting (Excel)
WW Seminar Tour
Spring 2006
Agilent Quantitative Analysis
Batch processing
WW Seminar Tour
Spring 2006
Quantitative View - Method Editor
WW Seminar Tour
Spring 2006
• All peak list level and peak level info available• Visualization of baseline regression• Handling of merged peaks• Peak validation – Quality Factor and ‘Traffic Light Coloring’
Parameter-Free Integrator – Optimal for QQQ data
SD
Baseline Offset
Discarded Peak
Accepted PeaksPeak List (XML)
Info messages
WW Seminar Tour
Spring 2006
New “Parameterless” Integrator
• Handles low level spiky data (statistically sets peak start and end)• Auto adjustment of threshold - Eliminates need for manual re-integration of peaks for low-level MRM signals
WW Seminar Tour
Spring 2006
THC Qualifier ion normalized THC Qualifier ion un-normalized
THC in QuantAnalysisTHC retention time < 4.5 min.
D0 quant ion
D3 quant ion
Overlap of quantand qual ion areas
WW Seminar Tour
Spring 2006
Triple (QQQ) CurveFit Assistant
WW Seminar Tour
Spring 2006
Calibration curve data point information
WW Seminar Tour
Spring 2006
Calibration curve data point zooming
QC (blue triangle) at 2 ng/mL
QC (blue triangle) at 5 ng/mL
WW Seminar Tour
Spring 2006
MRM transitions
Assay # 2, Cocaine, etcCoc, quant = 304.1 > 182.0Coc, qual = 304.1 > 82.0D3 - Coc (ISTD), quant = 307.1 > 185.1D3 - Coc (ISTD), qual = 307.1 > 85.1MDMA, quant = 194.1 > 163.0MDMA, qual = 194.1 > 135.0D5 - MDMA (ISTD), quant = 199.1 > 165.0D5 - MDMA (ISTD), qual = 199.1 > 135.0Meth, quant = 150.1 > 91.0Meth, qual = 150.1 > 119.0D5 - Meth (ISTD), quant = 155.1 > 92.0D5 - Meth (ISTD), qual = 155.1 > 121.0Amp, quant = 136.1 > 93.0Amp, qual = 136.1 > 119.0D5 - Amp (ISTD), quant = 141.1 > 93.0D5 - Amp (ISTD), qual = 141.1 > 124.0
Analytical Conditions – MS
WW Seminar Tour
Spring 2006
Using Agilent Qualitative Analysis for Coc, Meth, Amp, & MDMA
EICs for all 16 MRMs definedand specified for peak integration
Qualitative AnalysisMethod Editor
WW Seminar Tour
Spring 2006
Using Agilent Qualitative Analysis for Coc, Meth, Amp, & MDMA
All 16 MRM transition EICs generated and integrated automatically
All peaks elute< 1.5 min!
Coc
MDMA
Meth
Amp
d3-Coc
d5-MDMA
d5-Meth
d5-Amp
WW Seminar Tour
Spring 2006
G6410A Agilent QQQ
• Outstanding sensitivity for its price class– 0.5 pg RESERPINE on column at 20:1 S/N and 5% RSD with 3 simultaneous MRMs running
concurrently – utilizing Autotune parameters for mass analyzer calibration.
• 1635 amu mass range (m/z 15 – 1650)• Powerful and reliable AutoTune analyzer calibration software• New, Powerful and flexible Quantitation Software• New “Parameterless” integrator• Intelligent Qualitative Data Analysis Software• Flexible reporting based on Excel• Single point LC and MS instrument control• ESI, APCI, and MultiMode sources available