isite 3.5: user preferences
DESCRIPTION
iSite 3.5: User Preferences. Learning Objectives. Understand the purpose and configuration of:. General Preferences Window Width/Center Image Processing Filters Display Preferences Hanging Protocols (iSite Radiology only) Canvas Page (iSite Radiology only). Table of Contents. - PowerPoint PPT PresentationTRANSCRIPT
iSite 3.5: User Preferences
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Learning Objectives
• General Preferences
• Window Width/Center
• Image Processing
• Filters
• Display Preferences
• Hanging Protocols (iSite Radiology only)
• Canvas Page (iSite Radiology only)
Understand the purpose and configuration of:
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Table of Contents
• Preferences Dialog
• General Preferences
• Window Width/Center
• Image Processing
• Filters
• Display Preferences
• Hanging Protocols (iSite Radiology only)
• Canvas Page (iSite Radiology only)
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Preferences Dialog
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
User Preferences
• User Preferences configure preferences for the currently logged in user
• System Preferences configure system-wide settings for every user
• Machine Level Preferences configure settings for the machine on which the user is currently logged in
• User Preferences are associated with the Security Code ISTUSRPREF
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Importing and Exporting Preferences
• Importing and Exporting Preferences
• Many preference settings can be imported and exported to facilitate setting up User and System Preferences
• The User Preferences that can be imported and exported are Window Width/Center and Filters
• The System Preferences that can be imported and exported are Window Width/Center, Filters, Annotation, DICOM Sources, Screen Overlays, Paper Printing, Organization Dictionary, Location Dictionary and Body Part Dictionary
• Please note that importing preferences will override any preferences in the system
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Preferences Dialog
• Preferences can be accessed by selecting the ‘P’ icon in the upper-right corner of the iSite Control Strip
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
General Preferences
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
General Preferences
• To configure your General User Preferences, click the ‘General Preferences’ listing and edit the options as desired for your iSite Radiology account
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Main Exam Monitor Layout
• Main Exam Monitor Layout allows users to specify the monitor on which current exams (versus prior exams) will be displayed
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Main Exam Monitor Layout
• ‘Two Monitor’ (dual diagnostic monitor workstations) selection allows current exams to be displayed on the left or right monitor
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Main Exam Monitor Layout
• ‘Four Monitor’ (four diagnostic monitor workstations) allows current exams to be displayed on the left two monitors, the right two monitors or the middle two monitors
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Main Exam Monitor Layout
• For customers who have the ‘Monitor Split Mode’ option enabled
• The ‘Monitor Split Mode’ option will display the active exam as specified in the ‘Four Monitor’ option in the ‘Main Exam Monitor Layout’
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Main Exam Monitor Layout
• The ‘Monitor Split Mode’ is enabled on the fourth (and last) page of the Hanging Protocols Wizard
• The ‘Monitor Split Mode’ is used to emulate two or four diagnostic monitor workstations for one and two diagnostic monitor workstations respectively
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Canvas Page Mode
• “Canvas Page Mode” options specify whether exams displayed in your Canvas Page are organized by patient or by exam
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Canvas Page Mode
• ‘Patient Based Mode’ will allow only one “Main Exam” per patient to be open at a time
• ‘Exam Based Mode’ will allow multiple “Main Exams” per patient to be open at a time
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Automatically Open Next Exam
• If you wish iSite Radiology to automatically open the next exam in Worklist Filters when you mark an exam “Read” click the corresponding checkbox
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Show Tips at Startup
• The ‘Tip of the Day’ may be disabled by un-checking the “Show tips at startup” button
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Tip of the Day
• Alternatively, If you wish to disable the ‘Tip of the Day’ uncheck the “Show tips at startup” checkbox while viewing the ‘Tip of the Day’
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
New Features
• The ‘New Features’ dialog will display upon the first log-in following a system upgrade
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Disable Navigation Rectangle Splitting
• Users who wish to disable Navigation Rectangle Splitting for Canvas Page Navigation Rectangles may select the “Disable Navigation Rectangle Splitting” checkbox
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Disable Navigation Rectangle Splitting
• The red rectangles shown on the Canvas Page indicate which images are displayed on the diagnostic monitors
• The rectangles may be split in order to display nonadjacent series or even series within prior exams
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Disable Navigation Rectangle Splitting
• Demonstrated at left are series from two exams
• The Navigation Rectangles in the upper left corner of each series indicate the placement of each series on the diagnostic monitors
• If the ‘Disable Navigation Rectangle Splitting’ option is active then this is not an option
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Show Measurements Palette
• Users who wish to display the measurements palette automatically when viewing images, may select the “Show Measurement Palette” checkbox
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Display Key Image Stack
• Users who wish to display the Key Image Series at the beginning of the Exam Rack, may select the “Display Key Image Stack” checkbox
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Anatomical Right on Left Side of Image Window
• Please note that the visibility of this option is dependent on the System Preferences set by your institution's iSite System Administrator
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Anatomical Right on Left Side of Image Window
• Users who wish to automatically display axial and oblique CTs with the anatomical right displayed on the left, regardless of the initial orientation of the image, may select the “Axial/Oblique CT:…” checkbox
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Window Width/Center
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Window Width/Center
• Users may select the Window Width and Center dialog to add, modify or delete the default preset window width and center values display
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Add Window Width/Center
• The Add Window Width/Center may be used to create new Window Width/Center presets by establishing a name, indicating the Modality and settings the Width and Center
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Window Width/Center
• To edit a window width and center setting “double click” a listing or highlight a listing and click the ‘Properties’ button
• When you have finished making changes, click ‘Apply’ and then ‘OK’ to save and exit the Dialog box
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Window Width/Center
• Window Width/Center settings are mapped to the number keys in the order displayed on the list
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing
• Edge Enhance: Uses a Laplacian method to increase the visibility of edges in the image
• Edge Detection: Provides an alternative to the Edge Enhance tool, using a Sobel operator to increase the visibility of edges
• Median Filtering: Uses a common filtering method to reduce noise in an image
• CLAHE (Contrast Limited Adaptive Histogram Equalization): Utilizes an adaptive form of histogram equalization that enhances the contrast adaptively across the image
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing
• Image Processing Filters have several configurable parameters which may be configured by each individual user
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing
• Image Processing Filters can be applied manually (via a right mouse click option)
• Any time an Image Processing Filter is applied to an image, iSite displays a “P” (Processed) Icon in the meta data area of that image
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Edge Enhance
• ‘Edge Enhance’ is performed with a Laplacian operator to discriminate the edges; the result is scaled and added back into the original image data
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Edge Enhance
• The intensity of ‘Edge Enhance’ is controlled by the “degree” parameter which can be “Light”, “Medium” or “Strong”
• Edge Enhancement is applicable on any monochrome image
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Edge Enhance
The image on the right side has the ‘Edge Enhance’ filter applied
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Edge Detect
• This class of algorithm accentuates the rising and falling components of edges
• This is done with a Sobel operator; the result is scaled and added back into the original image data
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Edge Detect
• The intensity of ‘Edge Detect’ is controlled by the “degree” parameter which can be “Light”, “Medium” or “Strong”
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Edge Detect
The image on the right side has the ‘Edge Detect’ filter applied
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Median Filter
• ‘Median Filter’ is used as a noise reduction filter
• Neighborhoods around each pixel are sampled and the central pixel is replaced with the median value of the neighborhood
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – Median Filter
The image on the right side has the ‘Median Filter’ applied
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing - CLAHE
• Contrast Limited Adaptive Histogram Equalization (CLAHE) is a variant of histogram equalization
• The draw back of the general histogram equalization method is that the computation is performed across the entire image
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing - CLAHE
• In medical images, the local behavior of the histogram can be radically different than either the global, or non-local behavior
• This class of method performs the histogram equalization on a parameterized region-by-region basis and stitches the results to prevent any boundary edges
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing - CLAHE
• The following parameters can be set via the “Preferences”
• Contextual Region Dimension
• Number Of Bins
• Clip Limit
• Map Level
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Image Processing – CLAHE
The image on the right side has ‘CLAHE’ applied
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Filters
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Filters
• Filters allow you to create custom search criteria for Worklists
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Filters
• Filters can be used to create Reading Worklists to meet specific needs
• Exercise care in creating filters so exams are not accidentally excluded
• To create a filter click ‘Add’
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Filters
• This screen allows the user to generate a filter
• Users are not required to fill in all the fields because the default for an empty field is ‘all’
• When you have finished making changes, click ‘Apply’ and then ‘OK’ to save and exit the Dialog box
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Filters
• The Filter shown will provide a worklist which includes:
– “CARDIAC” subspecialty (as defined in the Exam Dictionary)
– Unread– All Exam Statuses
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Filters
• The Filter shown will provide a worklist which includes:
– “CR” and “DX” (DR) images for patients in
– “Critical Care” Units– For Body Part “Chest”– “Stat” priority– “Unread Only”– “Completed” Exam Status
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Filters
• The Filter shown will provide a worklist which includes:
– “CR” and “DX” (DR) images for patients in
– “Outpatient” Units– For Subspecialty “BONE”– “Non-Stat” priority– All Exam Statuses, except “In
Progress”– For 30 Days– Ordered by Dr.’s Smith,
Jones, or Davis
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Display Preferences
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Display Preferences
• The Display Preferences allow you to set several options
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• Users may select from the Screen Overlays Font Sizes menu in the User Preferences Display Preferences
• The value of each selection is determined by the Machine Preferences Font Size settings
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• ‘Small Font,’ ‘Medium Font,’ and ‘Large Font’ control the font size of the Screen Overlays
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• The Screen Overlay is shown at left with 30 font size
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• ‘Cine Playback Resolution’ controls the matrix size of images displayed during cine playback
• The default setting is “Full Resolution”
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• ‘Default Cine Rate’ controls the frame rate of the cine loop playback
• ‘Default Cine Rate’ will be used if the “Recommended Frame Rate” (DICOM Attribute 0008,2144) or “Cine Rate” (DICOM Attribute 0018,0040) are not present
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• The ‘Scout Line Mode’ defines which scout line method is used
• Single: a scout line showing the location of the slice being reviewed
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• Bracket: a single scout line, showing the location of the slice being reviewed, as well as the two lines showing the limits of the series being reviewed
• All: a line for every slice in the series being reviewed
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Font Sizes
• The ‘Localizer Mode’ option defines whether or not scout lines are displayed when the Localizer Mode is active
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Hanging Protocols
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Hanging Protocols
• The Hanging Protocols preference area allows the user to view the iSite Radiology Hanging Protocols and to delete them if necessary
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Hanging Protocols
• All of the Hanging Protocol properties windows are ‘Read Only’ and cannot be edited from the Preferences page
• Hanging Protocols must be edited from within the Radiology canvas page
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Canvas Page
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Canvas Page
• Canvas Page allows the user to activate the display of information on the iSite Radiology Canvas Page
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Patient Information Area
• Patient Comments(DICOM Tag 0010,4000)
• Patient History(DICOM Tag 0010,21b0)
• Study Comments(DICOM Tag 0032,4000)
CONFIDENTIAL TRN-0035-04 Rev 04/18/06
Exam Information Area
• Reason for Study (DICOM Tag 0032,1030)
• Impressions (parsed from report messages)
CONFIDENTIAL TRN-0035-04 Rev 04/18/06