iso 10993-6: biological evaluation of medical devices - tests for local effects after implantation
TRANSCRIPT
![Page 1: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/1.jpg)
ISO 10993-6
Biological Evaluation of Medical Devices
Tests for local effects after implantation
Presenters: Kelli Young, DVM
Lyn M. Wancket, DVM, PhD, DACVP
![Page 2: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/2.jpg)
2
ISO 10993-6 offers general guidelines for
evaluating local tissue responses to implants
![Page 3: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/3.jpg)
3
What is the purpose of conducting implant
studies?
![Page 4: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/4.jpg)
4
What is the purpose of conducting implant
studies?
Regulatory – address safety/biocompatibility
![Page 5: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/5.jpg)
5
What is the purpose of conducting implant
studies?
Regulatory – address safety/biocompatibility
Screening tool for Research & Development
![Page 6: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/6.jpg)
6
What is the purpose of conducting implant
studies?
Regulatory – address safety/biocompatibility
Screening tool for Research & Development- Example:
- Developing specific implant characteristics
![Page 7: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/7.jpg)
7
What is the purpose of conducting implant
studies?
Regulatory – address safety/biocompatibility
Screening tool for Research & Development- Example:
- Developing specific implant characteristics
- Porous surface to promote ingrowth
![Page 8: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/8.jpg)
8
What is the purpose of conducting implant
studies?
Regulatory – address safety/biocompatibility
Screening tool for Research & Development- Example:
- Developing specific implant characteristics
- Porous surface to promote ingrowth
- Coatings to prevent cell attachments
![Page 9: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/9.jpg)
9
What is the purpose of conducting implant
studies?
Regulatory – address safety/biocompatibility
Screening tool for Research & Development- Example:
- Developing specific implant characteristics
- Porous surface to promote ingrowth
- Coatings to prevent cell attachments
- Anti-microbial coatings
![Page 10: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/10.jpg)
10
What is the purpose of conducting implant
studies?
Regulatory – address safety/biocompatibility
Screening tool for Research & Development- Example:
- Developing specific implant characteristics
- Porous surface to promote ingrowth
- Coatings to prevent cell attachments
- Anti-microbial coatings
- Stimulation of specific cells – osteoinduction
- Etc.
![Page 11: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/11.jpg)
11
Tests for local effects after implantation
![Page 12: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/12.jpg)
12
Tests for local effects after implantation
Evaluates a material’s potential effects on
surrounding living tissue
![Page 13: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/13.jpg)
13
Tests for local effects after implantation
Evaluates a material’s potential effects on
surrounding living tissue
Use of device can determine location of implant
![Page 14: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/14.jpg)
14
Tests for local effects after implantation
Evaluates a material’s potential effects on
surrounding living tissue
Use of device can determine location of implant
Can be conducted in a variety of test models
![Page 15: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/15.jpg)
15
Tests for local effects after implantation
Evaluates a material’s potential effects on
surrounding living tissue
Use of device can determine location of implant
Can be conducted in a variety of test models
Size of the implant may determine test model
![Page 16: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/16.jpg)
16
Tests for local effects after implantation
Evaluates a material’s potential effects on
surrounding living tissue
Use of device can determine location of implant
Can be conducted in a variety of test models
Size of the implant may determine test model
Need multiple evaluation intervals- Short term, 2 up to 12 weeks
- Long term, > 12 weeks
- Degradable materials will have multiple intervals
![Page 17: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/17.jpg)
17
Degradable materials
These materials will result in a response that is
different to non-absorbable materials (negative
control)
![Page 18: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/18.jpg)
18
Degradable materials
These materials will result in a response that is
different to non-absorbable materials (negative
control)
Consider the use of a similar comparative control
![Page 19: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/19.jpg)
19
Degradable materials
These materials will result in a response that is
different to non-absorbable materials (negative
control)
Consider the use of a similar comparative control
Intervals need to be based on absorption profile- Pilot study or utilize other means to determine
![Page 20: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/20.jpg)
20
Degradable materials
These materials will result in a response that is
different to non-absorbable materials (negative
control)
Consider the use of a similar comparative control
Intervals need to be based on absorption profile- Pilot study or utilize other means to determine
Location markers
![Page 21: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/21.jpg)
21
Implant sites
Location varies depending on use of device
![Page 22: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/22.jpg)
22
Implant sites
Location varies depending on use of device
Test and control sites- Negative control- HDPE
![Page 23: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/23.jpg)
23
Implant sites
Location varies depending on use of device
Test and control sites- Negative control- HDPE
- Sponsor provided controls
![Page 24: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/24.jpg)
24
Implant sites
Location varies depending on use of device
Test and control sites- Negative control- HDPE
- Sponsor provided controls
- Test and control(s) implanted in each test model
![Page 25: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/25.jpg)
25
Implant sites
Location varies depending on use of device
Test and control sites- Negative control- HDPE
- Sponsor provided controls
- Test and control(s) implanted in each test model
- Total of at least 10 tests and 10 control sites at termination
![Page 26: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/26.jpg)
26
Evaluation
Macroscopic observations of implant sites
![Page 27: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/27.jpg)
27
Evaluation
Macroscopic observations of implant sites
Microscopic evaluation – primary basis of evaluation- Compared to the negative control and scored
![Page 28: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/28.jpg)
28
Evaluation
Macroscopic observations of implant sites
Microscopic evaluation – primary basis of evaluation- Compared to the negative control and scored
Looking for point at which test article reaches steady
state- Multiple intervals allow for this to be evaluated
![Page 29: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/29.jpg)
29
Evaluation
Macroscopic observations of implant sites
Microscopic evaluation – primary basis of evaluation- Compared to the negative control and scored
Looking for point at which test article reaches steady
state- Multiple intervals allow for this to be evaluated
Sponsor supplied control article help define the
reaction- Absorbable materials will “normally” stimulate an influx of
phagocytic cells to remove the material. Comparisons to a
non-absorbable, inert material will give a misleading
impression
![Page 30: ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effects after implantation](https://reader033.vdocument.in/reader033/viewer/2022042607/55a966fd1a28abb2518b45c0/html5/thumbnails/30.jpg)
30
To view the complete presentation on ISO
10993-6
- Check out NAMSA’s Seminars
- You can view the entire ISO 10993 Series here