KEN, JERI FU JEN CATHOLIC UNIVERSITY
2008/DEC./11
1
Testing Status of P.P. & APDTesting Status of P.P. & APD
![Page 2: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/2.jpg)
Two kinds of Samples(purpose: enlarge light signals from CsI
crystals)2
(photo pentode tubes)
(avalanche photodiodes)
![Page 3: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/3.jpg)
Schematic Diagrams for APD3
)10&()10&(
)300&()300&(_
VdarkVVlightV
VdarkVVlightVRatioGain APD
We use 400V400V for Serial No. AA0705& 450V450V for Serial No. AA0724
•Light Source: Blue LED•Light Intensity: detected by PIN Photodiode simulate very weak light (very simple scale)
![Page 4: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/4.jpg)
Results: 2008/Nov./254
Temperature, ℃
Time, min Time, min
Time, min
![Page 5: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/5.jpg)
Results: 2008/Nov./265
Temperature, ℃
Time, min Time, min
Time, min
![Page 6: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/6.jpg)
Results: 2008/Nov./276
Temperature, ℃
Time, min Time, min
Time, min
![Page 7: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/7.jpg)
Results: 2008/Dec./17
Temperature, ℃
Time, min Time, min
Time, min
![Page 8: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/8.jpg)
Results: 2008/Dec./28
Temperature, ℃
Time, min Time, min
Time, min
![Page 9: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/9.jpg)
Summary Plots 19
Output Voltage of APD 5 Days
600
700
800
900
1000
0 1 2 3 4 5 6
Days
Out
put
Volt
age
(mV)
APD 0705
APD 0724
HV Mode
2008 Nov.25~Dec./2
2008112520081126
2008112720081201
20081202
![Page 10: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/10.jpg)
Summary Plots 210
GAIN of APD 5 Days
30
40
50
60
70
0 1 2 3 4 5 6
Days
Rati
o
APD 0705
APD 0724
2008 Nov.25~Dec./2
2008112520081126
2008112720081201
20081202
![Page 11: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/11.jpg)
Summary of APD
APD Type: 07240724 is more stable & Gain ratio is higher. Better choose!
When temperaturetemperature is lower, the output voltage becomes higher. But, Gain Ratio is stable.
11
![Page 12: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/12.jpg)
Next Step of APD
We will keep measuring these two APDs and will provide their results on the webpage soon.
Could we have two more APDs? We can tell which type is better.
We are making a magnetic field now. We hope to know how will it affect the gain ratio of APD.
12
![Page 13: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/13.jpg)
Schematic Diagrams for P.P.13
Pentode mode
Diode modediodediode
pentodepentodePP VdarkVVlightV
VdarkVVlightVRatioGain
)100&()100&(
)1000&()1000&(_
![Page 14: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/14.jpg)
Results: 2008/Nov./2514
Temperature, ℃
Time, min Time, min
Time, min
![Page 15: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/15.jpg)
Results: 2008/Nov./2615
Temperature, ℃
Time, min Time, min
Time, min
![Page 16: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/16.jpg)
Results: 2008/Nov./2716
Temperature, ℃
Time, min Time, min
Time, min
![Page 17: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/17.jpg)
Results: 2008/Dec./117
Temperature, ℃
Time, min Time, min
Time, min
![Page 18: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/18.jpg)
Results: 2008/Dec./218
Temperature, ℃
Time, min Time, min
Time, min
![Page 19: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/19.jpg)
Summary Plots 119
Output Voltage of PP 5 Days
4000
4200
4400
4600
4800
5000
5200
5400
0 1 2 3 4 5 6
Days
Out
put
Volt
age
(mV)
PP 2630
PP 2632
PP 2636
HV Mode
2008 Nov.25~Dec./2
2008112520081126
2008112720081201
20081202
![Page 20: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/20.jpg)
Summary Plots 220
GAIN of PP 5 Days
140
160
180
200
220
240
260
0 1 2 3 4 5 6
Days
Rati
o
PP 2630
PP2632
PP 2636
(PP 2630)線性
(PP2632)線性
(PP 2636)線性
2008 Nov.25~Dec./2
2008112520081126
2008112720081201
20081202
![Page 21: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/21.jpg)
Summary of P.P.
Gain-ratio of the three P.P.s are stable. Type 26362636 shows highest gain-ratio.
TemperatureTemperature doesn’t show direct effect on the P.P. output voltage.
21
![Page 22: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/22.jpg)
Next Step of P.P.
We will keep measuring these three P.P.s and will provide their results on the webpage soon.
We are making a magnetic field now. We hope to know how will it affect the gain ratio of P.P.
22
![Page 23: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/23.jpg)
23
Setup: Outside View
Left Box: P.P.
Right Box: APD
Bottom: Box Size
![Page 24: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/24.jpg)
24
Inside the Box: P.P.
The distance between LED and PP samples is 34cm.
The wave length of blue LED is about 450 nm.
20cmPIN PD
![Page 25: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/25.jpg)
25
Inside the Box: APD
The distance between LED and APD samples is 34cm.
The wave length of blue LED is about 450 nm.
20cmPIN PD
![Page 26: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/26.jpg)
26
Output Monitoring
Right: voltage
Bottom: temperature
![Page 27: KEN, JERI FU JEN CATHOLIC UNIVERSITY 2008/DEC./11 1 Testing Status of P.P. & APD](https://reader035.vdocument.in/reader035/viewer/2022062500/5697bfee1a28abf838cb9850/html5/thumbnails/27.jpg)
Conclusion
Comparing the gain-ratio between APD and P.P., we think P.P. is better to enlarge the light from pure CsI crystal.
27