Keysight 82357B USB/GPIB Interface

User’s Guide

Page 2: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the
Page 3: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

82357B User’s Guide i

Page 4: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

ii 82357B User’s Guide

Notices© Keysight Technologies 2006 - 2014

No part of this manual may be reproduced in any form or by any means (including elec-tronic storage and retrieval or translation into a foreign language) without prior agree-ment and written consent from Keysight Technologies as governed by United States and international copyright laws.

Manual Part Number82357-90003

EditionEdition 9, November 2014

Printed in Malaysia

Keysight Technologies 1400 Fountaingrove Parkway Santa Rosa, CA 95403

Warranty

The material contained in this docu-ment is provided “as is,” and is sub-ject to being changed, without notice, in future editions. Further, to the max-imum extent permitted by applicable law, Keysight disclaims all warran-ties, either express or implied, with regard to this manual and any infor-mation contained herein, including but not limited to the implied warran-ties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for inciden-tal or consequential damages in con-nection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor-dance with the terms of such license.

Restricted Rights LegendU.S. Government Restricted Rights. Soft-ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus-tomers. Keysight provides this customary commercial license in Software and techni-cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com-puter Software Documentation).

Safety Notices

CAUTION

A CAUTION notice denotes a haz-ard. It calls attention to an operat-ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per-formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi-tions are fully understood and met.

Page 5: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

U.S. Government Restricted Rights

The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as “commercial computer software” as defined in DFARS 252.227- 7013 (Oct 1988), DFARS 252.211-7015 (May 1991) or DFARS 252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR 2.101(a), or as “Restricted computer software” as defined in FAR 52.227-19 (Jun 1987) (or any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Keysight standard software agreement for the product involved.

General Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement will control. Duration and conditions of warranty for this product may be superseded when the product is integrated into (becomes a part of) other Keysight products. During the warranty period, Keysight will, at its option, either repair or replace products which prove to be defective. The warranty period begins on the date of delivery or on the date of installation if installed by Keysight.

Warranty Service

For warranty service or repair, this product must be returned to a service facility designated by Keysight. For products returned to Keysight for warranty service, the Buyer shall prepay shipping charges to Keysight and Keysight shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes for products returned to Keysight from another country.

82357B User’s Guide iii

Page 6: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supplied products or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.

The design and implementation of any circuit on this product is the sole responsibility of the Buyer. Keysight does not warrant the Buyer’s circuitry or malfunctions of Keysight products that result from the Buyer’s circuitry. In addition, Keysight does not warrant any damage that occurs as a result of the Buyer’s circuit or any defects that result from Buyer-supplied products.

To the extent allowed by local law, Keysight makes no other warranty, expressed or implied, whether written or oral with respect to this product and specifically disclaims any implied warranty or condition of merchantability, fitness for a particular purpose or satisfactory quality.

Exclusive Remedies

To the extent allowed by local law, the remedies provided herein are the Bu yer’s sole and exclusive remedies. Keysight shall not be liable for any direct, indirect, special, incidental, or consequential damages (including lost profit or data), whether based on warranty, contract, tort, or any other legal theory.

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

iv 82357B User’s Guide

Page 7: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Safety SummaryThe following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.

Safety Notices

WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or loss of life. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

82357B User’s Guide v

Page 8: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Safety SymbolsThe following symbol on the instrument and in the documentation indicates precautions that must be taken to maintain safe operation of the instrument.

Regulatory Markings

The Instruction Documentation Symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied documentation.

The CE mark shows that the product complies with all the relevant European Legal Directives.

ICES/NMB-001 indicates that this ISM device complies with Canadian ICES-001.

The CSA mark is a registered trademark of the Canadian Standards Association. A CSA mark with the indicators “C” and “US” means that the product is certified for both the U.S. and Canadian markets, to the applicable American and Canadian standards.

The C-tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australian EMC Framework regulations under the terms of the Radio Communications Act of 1992.

This product complies with the WEEE Directive (2002/96/EC) marking requirement. The affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.

vi 82357B User’s Guide

Page 9: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

General Safety Information

WARNING • DO NOT use the device if it appears damaged or defective.

• Observe all markings on the device before connecting any wiring to the device.

• DO NOT operate the device in the presence of flammable gases or fumes.

• DO NOT install substitute parts or perform any unauthorized modification to the device.

CAUTION • Use the device with the cables provided.

• Repair or service that is not covered in this manual should only be performed by qualified personnels.

82357B User’s Guide vii

Page 10: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.

Product Category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.

The affixed product label is shown as below:

Do not dispose in domestic household waste

To return this unwanted instrument, contact your nearest Keysight office, or visit:

www.keysight.com/environment/product

for more information.

viii 82357B User’s Guide

Page 11: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Environmental ConditionsThis instrument is designed for indoor use only. The table below shows the general environmental requirements for the product.

General MaintenanceTo remove the dirt or moisture in the enclosure:

• Wipe the case with a damp cloth and mild detergent.

• Do not use abrasives or solvents.

• Wipe the contacts in each terminal with a clean swab dipped in alcohol.

Environmental Conditions Requirements

Temperature 0 °C to 55 °C (Operating)–40 °C to +70 °C (Non-operating)

Humidity Operating up to 90% at 40 °C (Non-condensing)Non-operating up to 90% at 65 °C (Non-condensing)

CAUTION This product is designed for use in compliance with:

• IEC 61010-1:2001/EN 61010-1:2001

• USA: UL61010-1: 2004

• Canada: CSA C22.2 No. 61010-1:2004

82357B User’s Guide ix

Page 12: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Declaration of Conformity (DoC)The Declaration of Conformity (DoC) for this instrument is available on the Keysight Web site. You can search the DoC by its product model or description at the Web address below.

http://www.keysight.com/go/conformity

NOTE If you are unable to search for the respective DoC, please contact your local Keysight representative.

x 82357B User’s Guide

Page 13: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Table of Contents

1 Installing the Keysight 82357B

Step 1: Checking Your Shipment 2

Step 2: Installing Keysight IO Libraries Suite 3

Step 3: Connecting the 82357B 5

82357B hardware description 5Connecting the 82357B to your PC 6

Connecting the 82357B to a USB hub 7

Step 4: Connecting GPIB Instruments 9

Connecting a single GPIB instrument 9Connecting multiple GPIB instruments 10

Step 5: Programming via the 82357B 11

Establishing instrument communication 11Programming GPIB instruments 13

2 Using the Keysight 82357B

Initial 82357B Operating States 18

Introduction to 82357B operating modes 19Single 82357B operation 20Multiple 82357B operation 21SRQ operation 22

Setting Configuration Parameters 23

Changing configuration parameters 23Changing modes of operation 24Setting timeout floor values 25Setting 82357B high-performance operation 26

3 Troubleshooting the Keysight 82357B

Troubleshooting Flowchart 28

82357B User’s Guide xi

Page 14: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Observe the LED states 28

Hardware Checks 30

Check USB cables, USB interface, host PC 30Check Device Manager 31

Software Installation Checks 32

Check suspend/resume operation 32Verify Keysight IO Libraries Suite installation 32

Software Configuration Checks 35

Checking IO Control operation 35Check USB scanner 36

Service and Support Information 37

Contacting Keysight 37

4 Product Specifications

Technical Specifications 40

Supplementary Information 43

xii 82357B User’s Guide

Page 15: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Keysight 82357B USB/GPIB InterfaceUser’s Guide

1Installing the Keysight 82357B

This chapter shows a suggested five- step process to install the 82357B and the Keysight IO Libraries Suite, to connect the 82357B to your PC and to program GPIB instruments via the 82357B.

Page 16: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

Step 1: Checking Your Shipment

Your 82357B USB/GPIB Interface shipment should include the items shown in Figure 1- 1. If any item is missing or damaged, keep the shipping materials and contact Keysight Technologies. See “Contacting Keysight” on page 37 later in this guide for addresses and telephone numbers.

As you check the shipment items, verify that the 82357B serial number at the bottom of the 82357B matches the serial number shown on the serial number label of the 82357B Kit Box and on the 82357B Certificate of Calibration. If the Serial Numbers do not match, contact Keysight. If all Serial Numbers match, you may want to record the Serial Number for future reference.

Figure 1-1 Contents of Keysight 82357B USB/GPIB Interface

2 82357B User’s Guide

Page 17: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Installing the Keysight 82357B 1

Step 2: Installing Keysight IO Libraries Suite

This section describes how to install the Keysight IO Libraries Suite on your PC. The Keysight IO Libraries Suite is a collection of libraries and utilities that gives you the ability to use your instruments from instrument control software.

You should install the Keysight IO Libraries Suite as your first step before installing the hardware because it installs the necessary software and drivers to control your instruments.

If possible, always use the most recent version of the Keysight IO Libraries Suite. This version supports the newest interfaces and operating systems, and has the most advanced features.

1 Verify that your PC meets the minimum system requirements. Refer to the IO Libraries Web page at www.keysight.com/find/iosuite or the IO Libraries Suite Readme.

2 Close all applications on your computer. Insert the Keysight Automation- Ready CD in your CD- ROM drive or download and install the IO Libraries software from www.keysight.com/find/iosuite.

3 Follow the instructions as prompted during the installation. You can select either a Typical or a Custom installation:

• Typical - In most cases, you can select the Typical installation which installs the IO Libraries Suite using the recommended settings.

NOTE You must have Administrator privileges to install Keysight IO Libraries Suite Connection Expert.

82357B User’s Guide 3

Page 18: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

• Custom - Select the Custom installation to:

a Install the IO Libraries Suite in another directory (for 32- bit operating systems only).

b Save disk space by not installing interface manuals.

c Use Keysight 32- bit VISA with another vendor's VISA on the same PC in side- by- side mode. Details on side- by- side mode are available at www.keysight.com/find/side- by- side- install or in the IO Libraries Suite help.

4 After the IO Libraries suite is successfully installed, you will see the IO icon on the Windows taskbar notification area.

Figure 1-2 IO icon on the Windows taskbar notification area

4 82357B User’s Guide

Page 19: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Installing the Keysight 82357B 1

Step 3: Connecting the 82357B

After the Keysight IO Libraries Suite software has been installed, you can connect the 82357B to any USB port on your PC or you can connect the 82357B via standard USB hubs.

82357B hardware description

The Keysight 82357B provides a direct interface connection from your PC USB port to GPIB instruments. The 82357B includes an attached USB cable that is USB 2.0 compliant. This cable is shielded and the connector is specified for up to 1,500 insertions. An 82357B can be directly connected to a single GPIB instrument or up to 14 GPIB instruments via GPIB cables. Up to four 82357B converters may be connected to your PC via standard USB hubs.

Figure 1-3 82357B Hardware Features

NOTE If the Keysight IO Libraries Suite software have not been installed on your PC, STOP. Install the libraries (see Step 2: Installing Keysight IO Libraries Suite).

Green ACCESS LED

USB Cable(Connect to USB port)Red FAIL LED

Green READY LED

GPIB Connector(Connect up to

14 GPIB instruments)

82357B User’s Guide 5

Page 20: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

Connecting the 82357B to your PC

This section shows steps to connect the 82357B to a USB port on your PC or to your PC via a USB Hub.

1 Connect to a USB Port. Make sure the PC is ON and plug the 82357B USB cable into any available USB port on your PC. Do not connect the 82357B to GPIB instruments at this time.

Figure 1-4 Connecting the 82357B to Your PC

2 Observe the LEDs. Observe the LEDs on the 82357B for at least 10 seconds. See Chapter 2, “Using the Keysight 82357B” for a description of the normal LED sequence during an initial installation of the 82357B.

a Initially, only the red FAIL LED should be ON. After a few seconds, all three LEDs should be ON. When all three LEDs are ON, this shows that the 82357B has been successfully installed.

b If all three LEDs are not ON after 10 seconds and all Windows Plug- and- Play Manager activity has ceased, STOP. See Chapter 3, “Troubleshooting the Keysight 82357B” for diagnostics information.

6 82357B User’s Guide

Page 21: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Installing the Keysight 82357B 1

Connecting the 82357B to a USB hub

This section shows steps to connect the 82357B to a USB port on your PC via a standard USB hub.

1 Plug the power adapter into the hub and into an electrical outlet. Make sure the hub is operating in self- powered mode. Figure 1- 5 shows a 4- port self- powered USB hub with two 82357B USB/GPIB Interfaces connected.

2 Make sure your PC is ON. Connect the USB cable of the USB hub to any available USB port on your PC.

3 Plug at least one 82357B USB/GPIB Interface into the port of the USB hub. It is not necessary to connect GPIB instruments to any 82357B at this time.

Figure 1-5 Connecting the 82357B to a USB Hub

NOTE Any USB hub used with the 82357B MUST be self-powered (must not be bus-powered or powered from the USB bus). Also, be sure to check the applicable USB hub documentation for hub operating parameters, such as power requirements and maximum length of USB cables.

82357B User’s Guide 7

Page 22: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

4 Observe the LEDs on the 82357B for at least 10 seconds. See Chapter 2, “Using the Keysight 82357B” for a description of the normal LED sequence during an initial installation of the 82357B.

a Initially, only the red FAIL LED should be ON. After a few seconds, all three LEDs should be ON. When all three LEDs are ON, this shows that the 82357B has been successfully installed.

b If all three LEDs are not ON after 10 seconds and all Windows Plug- and- Play Manager activity has ceased, STOP. See Chapter 3, “Troubleshooting the Keysight 82357B” for diagnostics information.

NOTE For Windows Vista, Windows 7, and Windows 8, there will be no Windows Plug and Play Manager Sequence because once 82357B is initially plugged into a USB port it will automatically install the driver.

8 82357B User’s Guide

Page 23: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Installing the Keysight 82357B 1

Step 4: Connecting GPIB Instruments

After the 82357B has been installed, the next step is to connect GPIB instruments to the 82357B. This step includes:

• Connecting a Single GPIB instrument OR ...

• Connecting Multiple GPIB Instruments

Connecting a single GPIB instrument

Figure 1- 6 shows connection from a single GPIB instrument to the GPIB port on an 82357B. When you have made the connection for your system, go to Step 5: Programming via the 82357B. You may want to record the primary GPIB address of the attached instrument for future programming use.

Figure 1-6 Connection from Single GPIB Instrument to the GPIB Port

CAUTION To avoid damage to the connectors, only finger-tighten the connectors.

82357B User’s Guide 9

Page 24: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

Connecting multiple GPIB instruments

Figure 1- 7 shows a typical way to connect three GPIB instruments to an 82357B. When you have made the connections for your system, go to Step 5: Programming via the 82357B. You may want to record the primary GPIB address of each attached instrument for future programming use.

Figure 1-7 Typical Way of Connecting Three GPIB instruments to an 82357B

NOTE Although Figure 1-7 shows the connection from 82357B to GPIB Instrument 1, the connection can be to any GPIB instrument in the system. Be sure to first connect the GPIB cable to the GPIB instrument and then “piggy-back” the 82357B GPIB connector to the GPIB cable.

*To minimize stress on the connector mountings, stack no more than three cable connector blocks.

10 82357B User’s Guide

Page 25: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Installing the Keysight 82357B 1

Step 5: Programming via the 82357B

After you have connected your GPIB instrument(s) to the 82357B, the next step is to establish communication between your PC and the instruments using Interactive IO. After communication has been established, you can begin programming the instruments using VISA, VISA COM, or SICL.

Establishing instrument communication

When the Keysight IO Libraries Suite is installed on your PC, an IO utility called Interactive IO is also installed. You can use Interactive IO to verify communication between your PC and the connected GPIB instrument(s). This section shows you how to use Interactive IO to verify instrument communication.

1 Click the IO icon on the Windows taskbar notification area.

2 Right- click the instrument and click Send Commands To This Instrument to display the Interactive IO window. For information on Interactive IO, click Help.

3 *IDN? is the default command. Click Send & Read to send the identification query to the instrument and display its reply in the Interactive IO window.

NOTE Once your GPIB interface has been configured in Connection Expert, if you can see the attached GPIB instrument(s) in the Connection Expert explorer and see their IDN string information in the detail pane then communication has been verified. Interactive IO allows you to manually verify communication and send specific commands to your instruments.

82357B User’s Guide 11

Page 26: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

4 To send other commands, click Commands> to select from a list of common commands, or type a command into the Command: field. If you experience timeout errors for some commands, click Options to change the timeout value.

Figure 1-8 Keysight Interactive IO

5 When communication has been established to each GPIB instrument, you can begin to program the instruments using VISA, VISA COM, or SICL. See the next section, Programming GPIB instruments.

12 82357B User’s Guide

Page 27: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Installing the Keysight 82357B 1

Programming GPIB instruments

This section provides an introduction to programming GPIB instruments via the 82357B USB/GPIB interface using the Keysight VISA, VISA COM, and SICL IO Libraries. You can program in various languages/applications, including Visual Basic, Visual C++, Keysight VEE, and National Instruments LabVIEW.

See the applicable user’s guide(s), such as the Visual Basic User’s Guide, for programming guidelines. You can also find additional programming examples using various IO Libraries and instrument drivers in the instrument user’s guide. After the 82357B is successfully installed, the interface acts as a transparent interface for programming GPIB instruments.

For information on programming using Keysight VISA, see the Keysight VISA User’s Guide. For information on VISA COM and for function references for VISA, VISA COM, and SICL, see the IO Libraries Suite Online Help.

Accessing VISA and SICL manuals

You can access .pdf copies of the Keysight VISA User’s Guide and the Keysight SICL User’s Guide for Windows from the IO icon on the Windows taskbar. Adobe Reader is required to view these manuals.

To access the Keysight VISA User’s Guide, click the IO icon then click Documentation > VISA Users Guide. To access the Keysight SICL User’s Guide for Windows, click the IO icon then click Documentation > SICL Users Guide. To access VISA COM information, and function references for VISA, VISA COM, and SICL, click the IO icon, then click Help Topics.

82357B User’s Guide 13

Page 28: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

Introduction to IO interface configuration

An IO interface consists of a hardware interface and a software interface. The Connection Expert utility is used to associate a unique software interface ID with a hardware interface.

The Keysight IO Libraries Suite uses an Interface ID or Logical Unit (LU) Number to identify an interface. This information is passed in the parameter string of the viOpen function call in a VISA program or in the iopen function call in an SICL program.

The Connection Expert assigns an Interface ID and Logical Unit (LU) Number to the interface hardware, as well as other necessary configurations. Typically, the LU Number is automatically assigned and you can ignore its setting. The LU Number is used internally as a unique identifier. When the IO interface is configured, you can use Keysight VISA, VISA COM, or SICL to program assigned instruments.

Example: IO interface configuration

For example, the GPIB interface system in the Figure 1- 9 consists of a Windows PC, an 82357B USB/GPIB interface, and three GPIB instruments with GPIB primary addresses of 3, 4, and 5, respectively. The instruments are connected via GPIB cables.

For this system, the Connection Expert utility has assigned a VISA name of “GPIB1” and a SICL name of “gpib1”. With these names assigned to the interfaces, the VISA/SICL addressing is as shown in Figure 1- 9.

Since unique names have been assigned by the Connection Expert, you can use the VISA viOpen command or the SICL iopen command to open the IO paths to the GPIB instruments as shown in the figure.

14 82357B User’s Guide

Page 29: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Installing the Keysight 82357B 1

Figure 1-9 Typical System Installation - 82357B USB/GPIB Interface

82357B User’s Guide 15

Page 30: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

1 Installing the Keysight 82357B

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

16 82357B User’s Guide

Page 31: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Keysight 82357B USB/GPIB InterfaceUser’s Guide

2Using the Keysight 82357B

This chapter describes normal operating states and modes for the 82357B and provides the guidelines to use the 82357B, including:

• Initial 82357B operating states

• Introduction to 82357B operating modes

• Single 82357B operation

• Multiple 82357B operation

• SRQ operation

Page 32: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

2 Using the Keysight 82357B

Initial 82357B Operating States

Figure 2- 1 shows the sequence of initial operating states when the 82357B is first connected to a USB port on a PC or on a USB hub.

Figure 2-1 Sequence of Initial Operating States

18 82357B User’s Guide

Page 33: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Using the Keysight 82357B 2

Introduction to 82357B operating modes

The 82357B has two modes of operation. When only one 82357B is connected to a USB port within a system, we define the feature as the Single Mode Features. When up to four 82357Bs are connected at the same time to USB ports within a system, we define the feature as the Multiple Mode Features.

Single Mode Features

For single mode operation, the operating parameters (VISA Interface ID, SICL Interface ID, Logical Unit Number, and GPIB Address) are set when the 82357B is first installed.

If this 82357B is unplugged and replugged, or if the 82357B is replaced with a different 82357B, the previous configuration parameters are automatically assigned to the newly attached 82357B. Thus, you can exchange 82357Bs at any time without reconfiguring the interface. This allows exchanging 82357Bs among users, as long as only one 82357B is attached at any one time.

Multiple Mode Features

In contrast, when up to four 82357Bs are connected to the system at the same time, each 82357B must have its own specific set of operating parameters and each 82357B serial number is “bound” to its operating parameters. In multiple mode operation, if you add a new 82357B or if you unplug an 82357B and plug in a new 82357B in its place, the newly installed 82357B, will be assigned a new (unique) set of operating parameters.

NOTE All SICL/VISA applications are notified when their 82357B has been removed from the system by returning VI_ERR_NOINFC (for VISA) or I_ERR_NCIC (for SICL).

82357B User’s Guide 19

Page 34: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

2 Using the Keysight 82357B

Single 82357B operation

When an 82357B is first installed, a default VISA Interface ID, SICL Interface ID, Logical Unit (LU) number and GPIB Address are automatically assigned to the serial number associated with this specific 82357B.

For example, assume an 82357B with serial number MY12345678. When this 82357B is first installed, typical values as shown are automatically assigned to this serial number:

• VISA Interface ID: GPIB0

• SICL Interface ID: gpib0

• Logical Unit: 7

• GPIB Address: 21

The first time an 82357B is attached to a system (assuming the Keysight IO Libraries Suite and 82357B driver are installed), the software recognizes that an 82357B is attached. If the Connection Expert is not running, the software displays a Keysight 82357B USB/GPIB Interface Detected dialog box that allows you to accept the current settings. If the Connection Expert is running, it automatically refreshes, displaying the 82357B as a USB/GPIB interface in its tree view. You can then change the properties of the interface via the Connection Expert.

The VISA and SICL Interface IDs, Logical Unit Number, and GPIB Address may be viewed at any time in the property pane of the Connection Expert. (To view this window, click the IO icon, then select Keysight Connection Expert.)

NOTE You can change the parameter values of the 82357B as required. See “Changing configuration parameters” on page 23 for details.

20 82357B User’s Guide

Page 35: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Using the Keysight 82357B 2

If you disconnect this 82357B and plug in another 82357B (with a different serial number), or if you re- plug the same 82357B, the new 82357B will assume the same VISA Interface ID, SICL Interface ID, LU number, and GPIB Address as the previous 82357B.

Multiple 82357B operation

When two or more 82357Bs are attached to a system at the same time, we define the mode as the multiple mode of operation. In multiple mode of operation, each 82357B is “bound” to its related IO Configuration for that Serial Number. This is a different mode of operation than the single mode of operation in that the configuration is not reused if you replace an 82357B with another 82357B.

As with single mode operation, the first time an 82357B is attached to a system (assuming the Keysight IO Libraries Suite and 82357B driver are installed), the software recognizes that an 82357B is attached. If the Connection Expert is not running, the software displays a Keysight 82357B USB/GPIB Interface Detected dialog box that allows you to accept the current settings. If the Connection Expert is running, it automatically refreshes, displaying the 82357B as a USB/GPIB interface in its tree view; you can then change the properties of the interface in the Connection Expert.

The VISA and SICL Interface IDs, Logical Unit Number, and GPIB Address may be viewed at any time in the property pane of the Connection Expert. (To view this window, click the IO icon, then select Keysight Connection Expert.)

If you plug in another 82357B (with a different Serial Number), the new 82357B will automatically be assigned a unique VISA Interface ID, SICL Interface ID, LU, and GPIB Address.

82357B User’s Guide 21

Page 36: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

2 Using the Keysight 82357B

SRQ operation

If your VISA/SICL application uses SRQ callbacks (viEventHandler() in VISA or ionsrq() in SICL) and your callback does not service the SRQ in a timely manner, your SRQ callback function may be triggered multiple times.

To avoid this possible situation, design your SRQ callback functions to be called only when an SRQ is no longer asserted on the GPIB bus.

NOTE • You can change the parameter values of the 82357B as required. See “Changing configuration parameters” on page 23 for details.

• You can also convert from multiple mode operation to single mode operation. See “Changing modes of operation” on page 24 for details.

22 82357B User’s Guide

Page 37: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Using the Keysight 82357B 2

Setting Configuration Parameters

This section gives guidelines to change or set various configuration parameters for the 82357B, including:

• Changing Configuration Parameters

• Changing Modes of Operation

• Setting Timeout Floor Value

• Setting High- Performance Operation

Changing configuration parameters

To change the VISA or SICL Interface ID, the LU or GPIB Address, or if you want to check the values of these configuration parameters, highlight the USB/GPIB interface in the explorer view (tree view) of the Connection Expert window. Click on the Change Properties... button in the property pane to display the Keysight 82357 Interface dialog box. Choose the settings you want, then click OK. Clicking Cancel will cause the configuration set in the preceding dialog box to be used.

NOTE Although you can change the Logical Unit (LU) Number and GPIB Address values for an 82357B, this is generally not necessary and may cause running applications to fail or stop running.

82357B User’s Guide 23

Page 38: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

2 Using the Keysight 82357B

Figure 2-2 Keysight 82357B (High Speed) Interface - USB/GPIB

Changing modes of operation

If your system has multiple 82357Bs configured (multiple mode operation), the only way to change from multiple mode of operation to single mode operation is to perform the following:

1 Unplug all 82357Bs from the system.

2 Run Connection Expert (click the IO icon, then click Keysight Connection Expert).

3 Delete all 82357B configurations by selecting each USB/GPIB icon in the explorer view, then clicking Delete (or delete all except one configuration).

4 Re- attach and re- configure a single 82357B.

24 82357B User’s Guide

Page 39: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Using the Keysight 82357B 2

Setting timeout floor values

The 82357B has a default timeout “floor” value that is an internal requirement to ensure reliable USB communication. The 82357B will not allow timeouts less than the floor value. (By default, VISA/SICL timeouts are set to infinite time).

To programmatically determine the timeout floor, you can set the timeout to a very small value, such as 1 ms, then query for the actual timeout floor value. VISA and SICL examples are as follows:

Example: Query Timeout Floor (VISA)

tval = 1; // Try to set timeout to 1 msec err = viSetAttribute(id, VI_ATTR_TMO_VALUE, tval_in); ... err = viGetAttribute(id, VI_ATTR_TMO_VALUE, &tval_out); ... printf("Set timeout to [%d], actual timeout that resulted [%d]\n", tval_in, tval_out );

Example: Query Timeout Floor (SICL)

tval = 1; // Try to set timeout to 1 msec err = itimeout(id, tval_in); ... err = igettimeout(id, &tval_out); ... printf("Set timeout to [%d], actual timeout that resulted [%d]\n",tval_in, tval_out );

82357B User’s Guide 25

Page 40: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

2 Using the Keysight 82357B

Setting 82357B high-performance operation

Introduction

The GPIB transfer rate for 82357B writes using large (>1000 bytes) buffer size is affected by the Data Available (T1) delay time. (The transfer rates are not noticeably affected when the buffer size is <1000 bytes). The default delay time used by the 82357B is 800 ns.

The maximum transfer rate for T1 = 357 ns is about 1.15 MB/s as compared to about 714 KB/s for the 82357B default value of 800 ns. Changing the T1 delay affects only the write performance of the 82357B.

Setting T1 Value With VISA

To set the T1 value with VISA, use the VI_AGATTR_GPIB_T1_DELAY attribute. The VI_AGATTR_GPIB_T1_DELAY value is the value of the T1 delay in nanoseconds, and should be no less than VI_AG_GPIB_T1DELAY_MIN and no greater than VI_AG_GPIB_T1DELAY_MAX. This value is defined in Keysight’s ‘visa.h’ header file. To use this value, you must '#define AGVISA_ATTRIBUTES' before the ‘#include ‘visa.h’‘ in your C or C++ source file.

The 82357B supports T1 delays from 357 ns to <max_value> in steps of 40 ns. You can find out the actual value by calling viGetAttribute().

Setting T1 Value With SICL

To set T1 value with SICL, use the igpibsett1delay() command and modify the GPIB environment.

NOTE Changing the T1 delay as described in this section is an advanced feature and also requires attention to cable lengths and other system features.

Attribute Access Privilege

Data Type Range (ns) Used By

VI_AGATTR_GPIB_ T1_DELAY

RW Global ViInt32 VI_AG_GPIB_T1DELAY_MIN to

VI_AG_GPIB_T1DELAY_MAX

GPIB INTFC resources

26 82357B User’s Guide

Page 41: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Keysight 82357B USB/GPIB InterfaceUser’s Guide

3Troubleshooting the Keysight 82357B

This chapter provides troubleshooting guidelines of the 82357B including hardware and software checks and also the service and support information for the 82357B.

Page 42: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

3 Troubleshooting the Keysight 82357B

Troubleshooting Flowchart

Figure 3- 1 shows a suggested sequence of steps to diagnose and troubleshoot 82357B problems, based on the LED states.

You can use the LED states to help diagnose and troubleshoot the 82357B whenever the LED states do not match expected normal states. See Chapter 2, “Using the Keysight 82357B” for the normal LED sequence when the 82357B is initially connected to a USB port.

Observe the LED states

To begin troubleshooting, observe the LED states for at least 10 seconds after the 82357B is connected to a USB port and all Windows Plug- and- Play Manager activity has ceased. Do the following:

• If all LEDs are OFF, start with Hardware Checks

• If the red FAIL LED is ON, start with Software Installation Checks

• If all LEDs are ON, start with Software Configuration Checks

After taking the steps in the check sequence, use the boxes in Figure 3- 1 to determine the next step. For example, if doing a hardware check results in only the red FAIL LED being ON, proceed with Software Installation Checks, and so on.

NOTE You do not have to do all the steps or do the steps in the order shown. If any action results in a change in LED states, go to the applicable check sequence to continue troubleshooting.

28 82357B User’s Guide

Page 43: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Troubleshooting the Keysight 82357B 3

Figure 3-1 Troubleshooting Flowchart

START

All LEDs OFF Red Fail LED ON ALL LEDs ON

READY FAIL ACCESS READY FAIL ACCESS READY FAIL ACCESS

Typical Cause

No power on USB bus or device turned off by Windows Plug and Play Manager.

Typical Cause

Keysight IO Libraries not installed or 82357B USB drivers not installed.

Typical Cause

Improper Keysight IOLibraries configuration.

Hardware Checks Software Installation Checks

Software Configuration Checks

Check Cables, USB Interface, Host PC

Check Suspend / Resume Operation

Check IO Control Operation

Reboot PCVerify Keysight IO

Libraries Installation Check USB Scanner

Check Device Manager Verify Driver Installation

After Doing These Checks:

If the FAIL LED is the only LED ON, go to Software Installation Checks.

If all three LEDs are ON, go to Software Configuration Checks.

If all three LEDs are OFF, contact Keysight.

After Doing These Checks:

If all three LEDs are ON, go to Software Configuration Checks.

If the FAIL LED is the only LED ON, contact Keysight.

After Doing These Checks:

If all three LEDs are ON, contact Keysight.

82357B User’s Guide 29

Page 44: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

3 Troubleshooting the Keysight 82357B

Hardware Checks

If all LEDs are still off for 10 or more seconds after plugging the 82357B USB cable into a USB port, and all Windows Plug- and- Play Manager activity and the Connection Expert refreshes have ceased, start your troubleshooting sequence by performing hardware checks. If any action taken results in a change in the LED status, go to Software Installation Checks or Software Configuration Checks.

Check USB cables, USB interface, host PC

Begin the hardware check by checking connections between the 82357B and PC (plus USB hubs, if used).

1 Check USB cable connections. Check the 82357B USB cable for a good connection to the USB port on the PC or on the USB hub. If you are using a USB hub, verify that the hub is connected to the PC.

2 Unplug/replug the 82357B USB cable. If this does not change the LED status, try plugging the 82357B into another USB port.

3 Check PC USB Port. Verify that the PC USB port is functional and powered (you can check using another USB device).

4 Check PC state. Verify that host computer is not in a suspended power management state.

5 Check USB hub. Try disconnecting the 82357B from the hub and connecting it directly to a USB port on the PC. Some USB hubs are vulnerable to static shock.

6 Check USB cables for damage. Check the USB cable for cuts/crushes. Since the end connectors are somewhat fragile, check for bent/misaligned/crushed connectors.

Reboot PC

If these steps do not change the LED status, reboot the PC.

30 82357B User’s Guide

Page 45: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Troubleshooting the Keysight 82357B 3

Check Device Manager

You can use the Windows Device Manager to reinstall the 82357B, as required.

Go to Device Manager by selecting Start > Control Panel > System > Hardware > Device Manager.

For Windows 8, right- click the bottom- left corner of the Desktop and select Device Manager.

From the Device Manager, select 82357 and then Properties. Tab to Driver and click Reinstall Driver. This will allow the Windows Plug- and- Play Manager to begin searching for a driver for the 82357B. Since the Device Manager may have disabled the 82357B USB device, click Enable to restart the 82357B.

NOTE If you are using a USB scanner, scanner conflicts are possible. See Check USB scanner in the “Software Installation Checks” on page 32.

82357B User’s Guide 31

Page 46: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

3 Troubleshooting the Keysight 82357B

Software Installation Checks

When only the red FAIL LED is still on after 10 seconds, the 82357B has been detected by the host computer.

Check suspend/resume operation

Some Windows operating systems support Power Management which can suspend the PC while the 82357B is in operation. After a Suspend/Resume cycle, the 82357B may not properly resume operation. In this case, you may need to unplug/replug the USB cable to restore 82357B operation. If this does not correct the problem, go to Verify Keysight IO Libraries Suite installation.

Verify Keysight IO Libraries Suite installation

When only the red LED is on after 10 seconds and all Windows Plug and Play Manager activity has ceased, start your troubleshooting sequence by verifying IO Libraries installation. If any action taken results in a change in the LED status, go to Software Configuration Checks or Hardware Checks.

1 Check Keysight IO Libraries Version. If a version of the IO Libraries Suite has been installed, the IO icon is normally displayed on the Windows taskbar notification area.

NOTE If your 82357B applications must not be preempted by a PC Suspend event, we recommend you to disable Power Management on your PC by using the Control Panel > Power Options dialog.

32 82357B User’s Guide

Page 47: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Troubleshooting the Keysight 82357B 3

Figure 3-2 IO icon on the Windows taskbar notification area

a If the IO icon is displayed, click the icon and select About Keysight IO Control to display the version. The version must be 15.0 or greater.

b If the IO icon is not displayed, a version of the IO Libraries Suite may still be installed. To verify this, check for Keysight Connection Expert in the Start menu, or the Start Screen (Windows 8).

c If found, launch the Keysight Connection Expert and select View > Keysight IO Control to place a check mark next to the Keysight IO Control selection.

d If Keysight Connection Expert is not found in the Start menu or the Start Screen (Windows 8), no version of Keysight IO Libraries is installed. In this case, or if the installed version is not 15.0 or greater, you must install the newer version (see Step 2).

82357B User’s Guide 33

Page 48: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

3 Troubleshooting the Keysight 82357B

2 Install Keysight IO Libraries (as required). If Version 15.0 or greater of the Keysight IO Libraries Suite is not installed on your PC, perform this step.

a Remove the 82357B USB cable from the USB port.

b Insert the Automation- Ready CD into your CD- ROM drive and follow the instructions in Chapter 1, “Installing the Keysight 82357B” to install the libraries. If you do not have the Automation- Ready CD, you can download the Keysight IO Libraries Suite from www.keysight.com/find/iolib

c Re- attach the 82357B USB cable to the USB port and observe the LEDs for at least 10 seconds.

If all three LEDs remain ON, go to Software Configuration Checks.

34 82357B User’s Guide

Page 49: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Troubleshooting the Keysight 82357B 3

Software Configuration Checks

If all three LEDs remain on for more than 10 seconds after the 82357B is connected to a USB port, the 82357B has been installed.

Start your troubleshooting sequence by checking IO Control operation. If any action taken results in a change in the LED status, go to Software Installation Checks or Hardware Checks, where applicable.

Checking IO Control operation

When the Keysight IO Libraries Suite is installed, an IO Control is created. When the IO Control is active, it is displayed as an IO icon on the Windows taskbar. By default, the IO Control is always active after the libraries are installed and the IO icon is displayed. However, there may be times when the IO Control can get deactivated. When this happens, SICL/VISA applications that are running with the 82357B will malfunction. Any of the following symptoms may indicate the IO Control is not active:

• Connection Expert is not running, and the Keysight 82357B USB/GPIB Interface Detected dialog box does not appear when an 82357B is first connected to a USB port.

• Connection Expert is running, but does not automatically refresh when an 82357B is first connected to a USB port.

• SICL/VISA applications using the 82357B are unable to open sessions.

• Windows Task Manager shows that iproc82357.exe is not running or is non- responsive.

82357B User’s Guide 35

Page 50: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

3 Troubleshooting the Keysight 82357B

If any of these symptoms occur:

1 Unplug/Replug the 82357B. If unplugging then replugging the 82357B causes the Keysight 82357B USB/GPIB Interface Detected dialog box to appear or the Connection Expert window to refresh, the problem is solved. If not, go to Step 2.

2 Shut down and restart IO Control. Take these steps to shut down and then restart the IO Control. Taking these actions should initialize all attached 82357Bs and display only the green Ready LED.

a If the IO icon is displayed, click the icon and then click Exit. A dialog box explaining the consequences of removing the IO Control appears. Click OK to shut down the IO Control.

b If the IO icon is not displayed, either the icon display has been turned off or the IO Control (and associated iproc82357.exe and iprocsvr.exe) is not active. In this case, launch the Keysight Connection Expert and select View > Keysight IO Control to place a check mark next to the Keysight IO Control selection to restart the IO Control and display the IO icon.

Check USB scanner

In general, USB scanners do not cause problems with the 82357B. However, if you do have problems with 82357B operation and have a scanner installed on your system that uses a USB port, unplug the scanner and then plug the 82357B into the port.

If the 82357B is configured without your scanner attached to your system, the scanner will be locking the 82357B from using the USB bus. In this case, contact your scanner manufacturer to request for software or firmware updates for the scanner.

NOTE Rebooting your PC should ALWAYS restart the IO Control, and re-execute iprocsvr.exe and iproc82357.exe.

36 82357B User’s Guide

Page 51: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Troubleshooting the Keysight 82357B 3

Service and Support Information

There are no user- serviceable parts for the Keysight 82357B USB/GPIB interface. If you suspect a hardware failure for the 82357B, contact Keysight for instructions to return the unit. See the following Contacting Keysight section for telephone numbers/Web site address.

Contacting Keysight

You can reach Keysight Technologies at this telephone number for the Americas:

Americas Call Center: 1- 800- 829- 4444

For other countries, contact your country’s Keysight support organization. A list of contact information for other countries is available on the Keysight Web site:

www.keysight.com/find/assist

82357B User’s Guide 37

Page 52: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

3 Troubleshooting the Keysight 82357B

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

38 82357B User’s Guide

Page 53: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Keysight 82357B USB/GPIB InterfaceUser’s Guide

4Product Specifications

This chapter lists the 82357B technical specifications and supplementary information.

Page 54: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

4 Product Specifications

Technical Specifications

GENERAL REQUIREMENTS

Minimum system requirements Windows XP Service Pack 3

• 600 MHz or higher required (800 MHz recommended)• 256 MB RAM (1 GB or greater is recommended)• Hard Disk Space

- 1.0 GB available for Microsoft .NET Framework 3.5, SP1- 100 MB for Keysight IO Libraries Suite

• Video- Super VGA (800x600) 256 colors or more

• Browser- Microsoft Internet Explorer 6.0 or greater

Windows Vista SP1 and SP2 (32-bit and 64-bit), Business, Ultimate, Enterprise, Home Basic, and Home Premium

• 1 GHz 32-bit (x86), 1 GHz 64-bit (x64), no support for Itanium64• 1GB minimum• Hard Disk Space

- 1.0 GB available for Microsoft .NET Framework 3.5, SP1- 100 MB for Keysight IO Libraries Suite

• Video- Support for DirectX 9 graphics with 128 MB graphics memory recommended (Super VGA graphics is supported)

• Browser- Microsoft Internet Explorer 7 or greater

Windows 7 (with or without SP1) (32- and 64-bit), Home Basic, Home Premium, Professional, Ultimate, and Enterprise, or Windows 8

• 1 GHz 32-bit (x86), 1 GHz 64-bit (x64), no support for Itanium64• 1GB minimum• Hard Disk Space

- 1.0 GB available for Microsoft .NET Framework 3.5, SP1- 100 MB for Keysight IO Libraries Suite

• Video- Support for DirectX 9 graphics with 128 MB graphics memory recommended (Super VGA graphics is supported)

• Browser- Microsoft Internet Explorer 7 or greater

40 82357B User’s Guide

Page 55: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Product Specifications 4

Windows Server 2008 R2 (with or without SP1) (64-bit) Standard and Enterprise

• 1.4 GHz (x64 processor) or 1.3GHz (Dual Core), no support for Itanium64• 1GB minimum• Hard Disk Space

- 1.0 GB available for Microsoft .NET Framework 3.5, SP1- 100 MB for Keysight IO Libraries Suite

• Video- Support for DirectX 9 graphics with 128 MB graphics memory recommended (Super VGA graphics is supported)

• Browser- Microsoft Internet Explorer 7 or greater

Supported standards • Support USB 2.0 high speed and full speed• Standard USB endpoints supported• IEEE-488.1 and IEEE-488.2 compatible• SICL and VISA 2.2

Supported applications (with IntuiLink) • Microsoft Excel 97 and 2000• Microsoft Word 97 and 2000• Check the web for latest supported applications

Supported software development applications

• Visual Basic 6.0 • Visual C++ 6.0• Visual Studio .NET • Keysight VEE 6.0 or greater• BASIC for Windows • LabVIEW 6.0 or greater

GENERAL CHARACTERISTICS

Power USB bus-powered device, +5 V, 500 mA (max), 200 mA (typ)

GPIB transfer rate 1.15 MB/s

Connectors Standard 24-pin IEEE-488, Standard USB A

USB hubs Self-powered hubs

Parallel polling A single parallel poll can easily check up to eight individual devices at once corresponding to the number of data lines on the GPIB.

Dimensions 105 mm (L) x 64 mm (W) x 30 mm (H) (includes connectors)

Weight 215 grams

Cable 2.5 meters, shielded, connector rated for 1500 insertions

LED Indicators READY, ACCESS, FAIL

Warranty 1 year

Maximum connections Maximum 4 converters can be connected to the PC

Instrument connection 14 instruments – daisy chain via GPIB

Configuration Plug-and-play

Pollution degree 2

82357B User’s Guide 41

Page 56: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

4 Product Specifications

ENVIRONMENTAL SPECIFICATIONS

Operating environment 0 °C to 55 °C

Storage environment –40 °C to +70 °C

Operating humidity Up to 90 % at 40 °C non-condensing

Storage humidity Up to 90 % at 65 °C non-condensing

ORDERING INFORMATION

Interface 82357B USB/GPIB Interface

Options Opt 0B1 - Add manual set

Accessories None

42 82357B User’s Guide

Page 57: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Product Specifications 4

Supplementary Information

This section provides supplementary information on the 82357B performance, including supported GPIB modes.The 82357B is defined as a controller as it can be (and is required to be) the system controller.

GPIB Modes of Operation Supported

The 82357B supports standard GPIB modes of operation, except for:

• Passing of Active Controller

• Non- system Controller mode which prevents using SICL Commander sessions or VISA Servant sessions

IEEE-488.1 and IEEE-488.2 Compliance

The 82357B is in full compliance with IEEE 488.1 and IEEE- 488.2 specifications. The 82357B fully supports IEEE- 488.1 subsets AH1, C1, C2, C3, C4, C27, DC0, DT0, LE3, PP0, RL0, SH1, SR0, and TE7.

SRQ Response Time

SRQ response time is slower than with the 82350 PCI GPIB interface as an artifact of the USB implementation. In addition, sharing the USB bus with other devices may impact GPIB performance.

Default T1 Delay

The default T1 delay for the 82357B is 800 nsec. See “Setting 82357B high- performance operation” on page 26 in Chapter 2, “Using the Keysight 82357B” for details.

Maximum 82357B System Configuration

Up to four 82357Bs on a system have been successfully tested.

82357B User’s Guide 43

Page 58: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

4 Product Specifications

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

44 82357B User’s Guide

Page 59: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Index

Index

BBefore You Install, 2

CChanging Configuration Parameters, 23changing configuration parameters, 23Changing Modes of Operation, 24changing modes of operation, 24check device manager, 31Check IO Control Operation, 35Check PC, 30Check Shipment, 2Check USB Cables, 30Check USB Interface, 30Check USB Scanner, 36checking shipment, 2, 30configuration parameters, changing, 23configuration parameters, setting, 23connecting 82357A to USB hub, 7Connecting GPIB Instruments, 9Connection Expert, 9, 14, 20, 21, 23, 24,

35, 36

DDevice Manager checking, 31

EEnvironmental Requirements, ixExample

IO Interface Configuration, 14examples

IO Interface Configuration, 14

HHardware Checks, 30Hardware Description, 5High-Performance Operation, setting, 26

IInitial Operating States, 18initial operating states, 18Installing Keysight IO Libraries Suite, 3interface ID, 14Interface Name, 14IO Control Operation checking, 35IO interface, 14IO Libraries Suite, checking for

installation, 32iopen, 14

KKeysight IO Libraries Suite, verify

installation, 32Keysight telephone number, 37Keysight Web site, 37

LLED States, 18LED states, 18Logical Unit Number, 14

MModes of Operation, changing, 24modes of operation, changing, 24multiple 82357A operation, 21Multiple 82357B Operation, 21Multiple Mode of Operation, 19multiple mode of operation, 19

PParallel Polling, 41PC

power management, 32suspend, 32

PC Checking, 30PC Rebooting, 30

PC, rebooting, 30Programming GPIB Instruments, 13programming GPIB instruments, 13Programming via 82357B, 11

Rrebooting the PC, 30Restricted Rights, iii

Ssetting configuration parameters, 23setting high-performance operation, 26setting timeout floor values, 25single 82357A operation, 20Single 82357B Operation, 20SRQ Operation, 22SRQ operation, 22suspend/resume operation, 32

Ttelephone number, Keysight, 37troubleshooting, 31

suspend/resume operation, 32

VviOpen, 14VISA Assistant, 11

Wwarranty service, iiiWeb site, Keysight, 37

82357B User’s Guide 45

Page 60: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

Index

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

46 82357B User’s Guide

Page 61: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

www.keysight.com

Contact usTo obtain service, warranty, or technical support assistance, contact us at the following phone numbers:

United States:(tel) 800 829 4444 (fax) 800 829 4433Canada:(tel) 877 894 4414 (fax) 800 746 4866China:(tel) 800 810 0189 (fax) 800 820 2816Europe:(tel) 31 20 547 2111Japan:(tel) 0120 (421) 345 (fax) 0120 (421) 678Korea:(tel) (080) 769 0800 (fax) (080) 769 0900Latin America:(tel) (305) 269 7500Taiwan:(tel) 0800 047 866 (fax) 0800 286 331Other Asia Pacific Countries:(tel) (65) 6375 8100 (fax) (65) 6755 0042

Or visit Keysight worldwide Web at:www.keysight.com/find/assist

Product specifications and descriptions in this document are subject to change with-out notice. Always refer to the English version at the Keysight Web site for the lat-est revision.

Page 62: Keysight 82357B USB/GPIB Interface 82357B .pdf · Keysight will, at its option , either repair or replac e products which prove to be defective. The warranty period begins on the

This information is subject to change without notice.© Keysight Technologies 2006 - 2014

Edition 9, November 2014

*82357-90003*82357-90003

www.keysight.com